You are on page 1of 6

772771

research-article2018
SRIXXX10.1177/1553350618772771Surgical InnovationYamashita et al

Original Article
Surgical Innovation

Evaluation of Surgical Instruments


1­–6
© The Author(s) 2018
Reprints and permissions:
With Radiofrequency Identification sagepub.com/journalsPermissions.nav
DOI: 10.1177/1553350618772771
https://doi.org/10.1177/1553350618772771

Tags in the Operating Room journals.sagepub.com/home/sri

Kazuhiko Yamashita, DEng1 , Kaori Kusuda, DSci2, Yoshitomo Ito, MD3,


Masaru Komino, RN3, Kiyohito Tanaka, MD4, Satoru Kurokawa, DEng5,
Michitaka Ameya, DEng5, Daiji Eba, BA6, Ken Masamune, DEng2,
Yoshihiro Muragaki, MD2, Yuji Ohta, DEng7, Chugo Rinoie, DPM, ABPO8,
Kenji Yamada, DEng1, and Yoshiki Sawa, MD1

Abstract
Background. Surgical instrument retention and instrument breakage compromise surgery quality and lead to medical
malpractice. We developed an instrument tracking system that could alert surgeons to instrument retention during
surgery and monitor instrument use to reduce the risk of breakage. Methods. This prospective, experimental clinical
trial included 15 patients undergoing inguinal hernia surgery or lumpectomy under general anesthesia at Saiseikai
Kurihashi Hospital. Radiofrequency identification (RFID)-tagged surgical instruments were used, and a detection
antenna was placed on a mayo stand during the operation. We analyzed the 1-loop detection ratio (OLDR)—that is,
the capability of the antenna to detect devices in a single reading—and the total detection rate (TDR)—that is, the
data accumulated for the duration of the operation—of the RFID-tagged instruments. Results. Data analysis revealed
that the OLDR was 95% accurate, whereas the TDR was 100% accurate. The antenna could not detect the RFID tag
when there was interference from electrocautery noise radiation, and 6% of instrument movement was undetected by
the antenna; however, the TDR and instrument use were detected at all times. Conclusions. Surgical instruments can be
tracked during surgery, and this tracking can clarify the usage rate of each instrument and serve as a backup method of
instrument counting. However, this study was conducted on a small scale, and RFID tags cannot be attached to small
surgical instruments used in complex operations such as neurosurgery. Further efforts to develop a tracking system
for these instruments are warranted.

Keywords
surgical instruments with RFID tags, operating room, tracking system

Introduction used to screen for retained instruments after surgery is


rather simple and has remained unchanged: counting the
Retention of surgical instruments inside the body cavity
continues to occur despite preventive measures such as 1
Osaka University, Osaka, Japan
routine manual counting of surgical instruments by oper- 2
Tokyo Women’s Medical University, Tokyo, Japan
ating room nurses. According to incident reports, instru- 3
Saiseikai Kurihashi Hospital, Saitama, Japan
ments or sponges are retained in approximately 1 in every 4
Japanese Red Cross Kyoto Daini Hospital, Kyoto, Japan
10 000 surgical procedures that involve an open cavity.1-4 5
National Institute of Advanced Industrial Science and Technology,
According to Gawande et al,1 one-third of the reported Ibaraki, Japan
6
EBA Corporation, Aichi, Japan
cases had retention of surgical instruments. 7
Ochanomizu University, Tokyo, Japan
Screening systems that detect and prevent potentially 8
Methodist Hospital of Southern California, Arcadia, CA, USA
harmful events are critical for ensuring patient safety. A
Corresponding Author:
guideline published by the World Health Organization Kazuhiko Yamashita, Graduate School of Medicine, Osaka University,
recommends manual counting of surgical instruments by 1-3 Yamadaoka, Suita, Osaka 565-0871, Japan.
2 nurses after the operation.5 The system most widely Email: k-yamashita@obd.med.osaka-u.ac.jp
2 Surgical Innovation 00(0)

instruments before and after use to check for discrepan-


cies.6,7 However, counting is known to be highly prone to
errors because it relies on human consistency and accu-
racy, and the operative setting has time constraints, dis-
tractions, and unexpected interruptions. The rate of
retained instruments has been reported to be 1 per 70 dis-
crepancy cases.3
In addition, malfunctioning of surgical instruments
has led to serious medical accidents during surgery. These
accidents include direct tissue damage and retained pieces
of broken instruments. In a previous study, the breakage
rate of surgical instruments in 2 years was 9.1%, and 2
near-miss incidents (10 in 100 000 operations) were
potentially critical.8 Most practitioners have not paid suf-
ficient attention to defective surgical instruments during
and after the operation. Figure 1. Radiofrequency identification–tagged instruments.
Because of human limitations, human error is highly
possible when manually counting different surgical size, frequently used surgical instruments can be placed
instruments with a similar appearance within a limited on it during surgery. Therefore, we considered that it
time frame, even if double-checking is done. If a sensing would be suitable for tracking the surgical instruments.
function is added to surgical instruments, double-check-
ing by a human and a tracking system can be performed,
and improved reliability can be expected. Another advan- Methods
tage of the sensing function is that the frequency of the Participants
use of surgical instruments during an operation can be
automatically calculated. If the type and number of surgi- The RFID tracing rate was studied in 15 patients who
cal instruments required for each operation are clarified, either underwent unilateral inguinal hernia surgery (2
the choice of surgical instruments can be optimized, and cases) or lumpectomy with axillary lymph node dissec-
the frequency of counting errors of surgical instruments tion and sentinel lymph node biopsy (13 cases) while
can be expected to decrease. The inventory quantity of under general anesthesia. The Human Studies Ethics
surgical instruments can also be optimized in hospitals. Committees at Tokyo Healthcare University and Saiseikai
In a previous study, we developed a radiofrequency Kurihashi Hospital approved this study. The operation
identification (RFID) tag for recognition and traceability method was in accordance with the ethical guidelines of
of surgical instruments.9 Next, we developed a manage- the hospital. All participants provided written informed
ment system composed of a scanning device that func- consent. The participants did not receive any stipend.
tioned as an antenna, which can detect RFID tags and
software to facilitate surgical tray assembly in central Protocols for RFID-Tagged Surgical
sterile processing departments. This management system
Instruments and Research
has been successfully tested in a 27-month study con-
ducted in central sterile processing departments.10 Figure 1 shows surgical instruments tagged with an RFID
However, devices capable of detecting surgical instru- (diameter, 6 mm; width, 2 mm; radiofrequency, 13.56
ments with RFID tags that can be used in the operating MHz). Table 1 lists the surgical instruments with an RFID
room have not yet been developed. tag used for inguinal hernia surgery (61 items) and
In this study, we developed an antenna that can be lumpectomy (46 items). Any practitioner who requested
used in the operating room to detect commonly used sur- (Figure 1) RFID-tagged instruments not listed in Table 1
gical instruments with an RFID tag. The goal of this study was provided with individually prepared untagged instru-
was to determine the tracing rate of the antenna in cor- ments. In this study, an antenna was developed to read the
rectly detecting the RFID attached to the surgical instru- information from the tag in the operating room. Figure 2
ments used in the operating room. The purpose of this shows the placement of the antenna on the mayo stand
feasibility study was to verify the effectiveness of RFID- during the operation.
tagged surgical instruments. For this reason, we decided RFID-tagged surgical instruments can be repeatedly
to use the mayo stand in the detection of RFID-tagged cleaned in the usual manner with disinfectant wash and
surgical instruments in the operating room. Because the sterilized using the same protocol (autoclave sterilization
mayo stand is used in various surgeries and is small in in the central sterile processing department) implemented
Yamashita et al 3

Table 1. Surgical Instruments Included in the Sets. table, where the instruments are counted. During the
operation, a scrub nurse moves each instrument from the
Inguinal Hernia Lumpectomy
Surgical Instruments (n = 61) (n = 46) instrument table onto the mayo stand, handing the instru-
ments to the surgeon as needed. Therefore, the instru-
1 Saddle-shaped hook 1 ments used change as the operation progresses.
2 Langenbeck retractor 2 2 Under our protocol, the antenna is placed on the mayo
10 × 25 mm stand, with a sterile cloth covering the device. This allows
3 Langenbeck retractor 2 2
practitioners to perform the operation as usual without
13 × 40 mm
distraction while allowing automatic detection of RFID-
4 Langenbeck retractor 2
13 × 60 mm tagged instruments. A video camera was placed above the
5 Long tweezers 2 mayo stand to obtain a visual record of instrument usage.
6 Tweezers 2 2 The data obtained through the automatic detection of
7 Tweezers with hook 2 2 RFID-tagged instruments were compared with the visual
8 Adson tissue forceps 1 2 record to determine accuracy. At the same time, the usage
9 Cloth clamp 2 4 rate of RFID-tagged instruments per operation was calcu-
10 Mosquito pean, curved 10 5 lated using the data obtained.
11 Mosquito Kocher 5
12 Pean, straight 3 1
Characteristics of the Developed Antenna and
13 Pean, curved 10 5
14 Kocher artery forceps 10 5 Analysis Method
15 Kelly forceps, 16 cm, 1 1 The antenna shown in Figure 2 allows the detection of
strong bending RFID tags regardless of its direction. The antenna com-
16 Kelly forceps, 18 cm, 1
bines several small detecting parts and requires 8 to 13 s
strong bending
per detection. The variance in detection time is caused by
17 Mikulicz’s forceps 4
18 Needle holder for gold, 2 2
the number of RFID-tagged instruments placed on top of
16 cm the antenna at a given time. After the antenna records the
19 Mathieu needle holder 2 1 RFID tag information, the ID and time log are exported in
20 Cooper blunt tongs, 1 1 a CSV file.
14 cm The RFID tag information collected within 13 s of
21 Cooper bluff direct, 1 placement was compared with the data obtained through
14 cm video recording, and the detection ratio per instrument
22 Metzenbaum, 16 cm 1 1 was calculated (1-loop detection ratio [OLDR]). The
23 Scalpel handle, No. 3 2 2 detection range of the system was defined by the distance
between the antenna and the RFID tag. When the RFID
tag was positioned face-to-face, or parallel with the
antenna, the detection range was 8 cm. When the tag was
positioned perpendicular to the antenna, the detection
range was 5 cm. The detection range was set at 8 cm for
parallel-positioned tags to ensure the detection of the
RFID-tagged instruments placed above the mayo stand.
RFID-tagged instruments placed on the instrument table
or on the patient’s bed as well as instruments currently in
use are not detected by the antenna. If the detection range
had been very long, the antenna might have erroneously
detected instruments that were not on the antenna, which
would cause data disruption.
The detection range extended vertically from the
antenna; however, the lateral detection range was less
Figure 2. Placement of the antenna on the mayo stand than 1 cm from the outer edge of the antenna. During
(arrow).
operation, instruments were placed at the edge of the
mayo stand, and the instruments could extend past the
for all other devices and then organized in a container. edge, as shown in Figure 2. This placement may affect
Before the operation, the surgical instruments are data detection, which are recorded once every 13 s.
removed from the container and placed on the instrument Accurate detection can be further ensured by educating
4 Surgical Innovation 00(0)

nurses to avoid this practice; however, in this study, no


additional training or orders were given to the nurses,
allowing them to perform their usual operative roles, and
they had no knowledge of the antenna or the RFID tags.
Throughout an operation, the position of the surgical
instruments on the mayo stand changes. Instruments are
often repositioned or even stacked above each other on
the mayo stand after being used in the surgical field or
when preparing new instruments retrieved from the
instrument table. Even if a single detection loop per-
formed every 13 s cannot detect an instrument, it is likely
that the instrument will be detected in a subsequent detec-
tion loop. To evaluate whether the instruments were being Figure 3. Sample detection log of radiofrequency
accurately detected, the total detection rate (TDR) was identification–tagged instruments.
calculated from the data. The OLDR and TDR were cal-
culated and compared with the video image record.
Table 2. Usage Rate of Surgical Instruments Based on
Operative Method.a
Results
Rate of Instrument
Each operation was performed by a surgeon, an assistant Operative Method Usage (%) Number of Items
surgeon, and a scrub nurse, with a total of 4 surgeons and
Case 1 47.5 29
8 scrub nurses. Surgeons included experienced instruc-
Case 2 41.0 25
tors and medical interns, and the nurses also had various Case 3 56.5 26
levels of experience. Case 4 56.5 26
Figure 3 shows the transition in the number of RFID- Case 5 50.0 23
tagged instruments that were detected by the antenna Case 6 60.9 28
placed on the mayo stand. The horizontal axis shows the Case 7 39.1 18
progression of operation by time, and the vertical axis Case 8 41.3 19
shows the number of RFID-tagged instruments. Throughout Case 9 37.0 17
the course of an operation, 6 to 14 RFID-tagged instru- Case 10 37.0 17
ments were placed on the mayo stand. At the end of the Case 11 47.8 22
operation, the instruments decreased from 4 to 0. Case 12 47.8 22
Table 2 shows the operations performed, the usage Case 13 63.0 29
rate, and the number of instruments used. The average Case 14 34.8 16
usage rates of the surgical instruments were 44.3% ± Case 15 37.0 17
3.3% and 47.6% ± 10.9% for inguinal hernia surgery and a
Operative method: cases 1-2, inguinal hernia; cases 3-12, mastectomy
lumpectomy, respectively. The mean operation durations, with lymph node dissection; and cases 13-15, partial or total
measured as the difference between the time when an mastectomy.
RFID-tagged instrument was first placed on the mayo
stand and the time when the last detection was recorded,
were 4151 ± 795 and 5752 ± 2466 s for inguinal hernia the mayo stand. Scrub nurses were not instructed to place
surgery and lumpectomy, respectively. instruments in a specific manner, nor were they required
After verification of the video record and the RFID to perform a specific maneuver for detection. The mayo
detection log, the TDR of the surgical instruments placed stand is usually used only for the instruments that are
at least once on the mayo stand was found to be 100% immediately necessary for the operation. By tracking the
(Table 3). The mean OLDRs were 97.4% ± 1.8% and usage of the surgical instruments, the time of initiation
95.9% ± 1.2% for inguinal hernia surgery and lumpec- and completion of the operation can be approximated.
tomy, respectively. In a previous study, 32% of instruments packaged
for an operation were packaged erroneously.11 The
number of instruments packaged before an operation
Discussion could be optimized by evaluating instrument usage.
The system we developed was able to detect 100% of the Thus, it may be possible to reduce human error when
surgical instruments placed on the mayo stand. The sys- preparing the container of manually counted instru-
tem also collected the usage rate of each instrument, ments. This study revealed that precise tracking of
along with the types and number of instruments placed on instruments during surgery is possible. Through our
Yamashita et al 5

Table 3. OLDR and TDR Between the Video Record and the RFID System.

Instruments Seen in Instruments Detected by OLDR Recognition TDR Recognition


No. Video (n) the System (n) Accuracy (%) Accuracy (%)
1 4665 4515 96.8 100
2 3345 3324 99.4 100
3 7672 7324 95.5 100
4 3771 3524 93.5 100
5 3949 3812 96.5 100
6 4703 4554 96.8 100
7 8309 7916 95.3 100
8 1667 1627 97,6 100
9 3032 2923 96.4 100
10 1423 1346 94.6 100
11 1624 1557 95.6 100
12 3159 3064 97.0 100
13 2956 2788 94.3 100
14 3141 2998 95.5 100
15 1176 1145 97.4 100

Abbreviations: OLDR, 1-loop detection ratio; RFID, radiofrequency identification; TDR, total detection rate.

system, multiple instruments can be accurately detected lower detection rate in such instances, however, does not
with their RFID tags, enabling double-checking by reduce the quality of the instruments or the operation
both the system and humans. This can allow the staff to itself. This study shows that the system can be used in the
focus more on inspecting the instruments for defects operating room to accurately detect instruments.
and patient monitoring. Thus, the system can contribute Retention of surgical instruments or breakage of instru-
to improving the operation quality and patient safety. ments within the body cavity substantially lowers patient
Furthermore, the system can count the total usage of safety. To prevent retention of surgical instruments within
surgical instruments, which can alert staff when a spe- the body cavity, a reliable method to count and monitor
cific instrument has reached a usage rate where defects instruments must be developed. Furthermore, to prevent
can be expected or when devices requiring sterilization breakage of instruments during surgery and to maintain
or other maintenance reach their maximum usage. the quality of the operation, instruments must be managed
The antenna requires 13 s to detect the tagged instru- so that the rate of disinfection, sterilization, and usage
ments placed above it. There is a possibility that if an may be tracked. This detailed management of the instru-
instrument does not remain on the mayo stand for a dura- ments cannot be performed by scrub nurses, who are
tion of 13 s, the instrument will not be detected, thus already required to physically handle the instruments
affecting the OLDR. However, in the 15 operations per- while simultaneously counting the instruments. A system
formed, this did not occur. Thus, a detection time of 13 s is necessary to support these nurses, and the system tested
may still allow accurate detection. in this study may provide a solution to these challenges.
There are several factors that may lower the OLDR, The system enabled automated tracing of individual surgi-
such as close stacking of the tags of several instruments, cal instruments while scrub nurses were performing their
interference from digital noise during electrocautery, or usual operative roles, without additional distractions.
placement of the RFID tag outside the lateral detection
range of the antenna. The video record showed that elec-
Limitations
trocautery was frequently performed during breast sur-
geries, and the RFID detection log showed no record This feasibility study was carried out during 15 inguinal
during electrocautery. However, as soon as electrocautery hernia and lumpectomy surgeries. The risk of surgical
was stopped, the RFID tag was detected. The RFID tag instrument retention in these 2 types of surgeries is not
detection rate was 100% in these cases as well. known to be high. However, other surgeries performed in
When electrocautery was performed for 300 s, the the same way as these procedures are performed deep in
antenna was unable to detect for 23 loops. A total of 10 the body cavity. Therefore, this system is considered to be
RFID-tagged instruments placed on the mayo stand would suitable to track RFID-tagged surgical instruments in
equate to 230 counts, which may be the reason why the these higher-risk surgeries. We plan to apply it to other
OLDR decreased in some instances during the study. The types of operations in the future.
6 Surgical Innovation 00(0)

The results of this research are only applicable to surgi- Analysis and interpretation: Kazuhiko Yamashita, Kaori
cal tools with RFID tags attached. Because RFID tags were Kusuda, Satoru Kurokawa, Mitsutaka Ameya, Daiji Eba, Chugo
not attached to individually prepared instruments, these Rinoie, Yuji Ohta, Kenji Yamada, Ken Masamune, Yoshihiro
were not evaluated. Moreover, RFID tags are not designed Muragaki, Yoshiki Sawa
Study supervision: Kazuhiko Yamashita, Kaori Kusuda, Ken
to be attached to small instruments used in complex opera-
Masamune, Yoshihiro Muragaki, Yoshiki Sawa
tions such as neurosurgery. Hence, tracking of surgical
instruments used in those surgeries cannot be accom-
Declaration of Conflicting Interests
plished with the system developed in this study. Developing
a more compact RFID tag may be a solution; however, a The author(s) declared no potential conflicts of interest with
smaller RFID tag will have shorter detection range in rela- respect to the research, authorship, and/or publication of this
article.
tion to the antenna, which may adversely affect the TDR.
However, for the majority of surgical instruments used in
Funding
orthopedics or neurosurgery, the RFID tags can be attached
easily, which could help decrease instances of surgical The author(s) disclosed receipt of the following financial sup-
instrument retention within open cavities. port for the research, authorship, and/or publication of this arti-
cle: This study was performed as an exploratory research
This research is one of the first to utilize RFID-tagged
performed by KazY (15K12608) and KK (974682).
surgical instruments in real operations, tracking the usage
of each instrument. The OLDR was shown to be greater
ORCID iD
than 96%, whereas the TDR was 100% accurate. Because
the RFID uses electromagnetic waves to communicate, it Kazuhiko Yamashita, https://orcid.org/0000-0003-1268-7260.
becomes undetectable in the presence of interference from
noise radiated during electrocautery. Because of this, the References
OLDR decreases as the duration of noise lengthens. 1. Gawande AA, Studdert DM, Orav EJ, Brennan TA, Zinner
Complete tracking of OLDR remains challenging, even MJ. Risk factors for retained instruments and sponges after
with the developed system. However, because the TDR surgery. N Engl J Med. 2003;348:229-235.
2. Bani-Hani KE, Gharaibeh KA, Yaghan RJ. Retained surgi-
values are more important in tracking the usage rate and
cal sponges (gossypiboma). Asian J Surg. 2005;28:109-115.
preventing residual instruments, the implications of the
3. Egorova NN, Moskowitz A, Gelijns A, et al. Managing
results achieved through this research are relevant. Further the prevention of retained surgical instruments: what is the
research is warranted to evaluate the effectiveness of this value of counting? Ann Surg. 2008;247:13-18.
system in reducing cases of instrument breakage or acci- 4. Gonzalez-Ojeda A, Roriguez-Alcantar DA, Arenas-
dental residual instruments through the tracking of surgi- Marquez H, et al. Retained foreign bodies following intra-
cal instrument usage and counts. abdominal surgery. Hepatogastroenterology. 1999;46:
808-812.
5. World Health Organization. WHO Guidelines for Safety
Conclusion Surgery: Safe Surgery Saves Lives. Geneva, Switzerland:
World Health Organization; 2009.
An antenna that can detect an RFID-tagged surgical instru- 6. Beyea SC. Counting instruments and sponges. AORN J.
ment was developed and tested in the operating room. We 2003;78:290, 293-294.
found that surgical instruments can be tracked during sur- 7. Van Wicklin SA. Implementing AORN recommended
gery, and their usage rate can be clarified. The results of this practices for prevention of deep vein thrombosis. AORN J.
study suggest that the use of an instrument-tracking system 2011;94:443-451.
may contribute to improved accuracy of the instrument 8. Yasuhara H, Fukatsu K, Komatsu T, Obayashi T, Saito Y,
count, thus avoiding instrument retention in the body cav- Uetera Y. Prevention of medical accidents caused by defec-
ity, and the prevention of instrument breakage during oper- tive surgical instruments. Surgery. 2012;151:153-161.
ation, substantially improving the quality of the operation. 9. Yamashita K, Kusuda K, Tokuda Y, et al. Validation of
cleaning evaluation of surgical instruments with RFID tags
attached based on cleaning appraisal judgement guidelines.
Acknowledgments Conf Proc IEEE Eng Med Biol Soc. 2013;2013:926-929.
We thank Hiroshi Honda, MD, Junya Tamaki, RN, and Michiko 10. Kusuda K, Yamashita K, Ohnishi A, et al. Management
Suzue, RN, for their contributions to this study. of surgical instruments with radio frequency identification
tags. Int J Health Care Qual Assur. 2016;29:236-247.
11. Wubben I, van Manen JG, van den Akker BJ, Vaartjes SR,
Author Contributions van Harten WH. Equipment-related incidents in the oper-
Study concept and design: Kazuhiko Yamashita, Kaori Kusuda ating room: an analysis of occurrence, underlying causes
Acquisition of data: Kazuhiko Yamashita, Kaori Kusuda, and consequences for the clinical process. Qual Saf Health
Yoshitomo Ito, Masaru Komino, Kiyohito Tanaka Care. 2010;19:e64.

You might also like