Professional Documents
Culture Documents
^ K
Gas Analysis
'A. M. Buswell and C. S. Boruff, The Relation between the Chemical Composition of Organic Matter
and the Quality and Quantity of Gas Produced during Sludge Digestion, Sewage Works J., 4: 454
(1932).
700 P A R T 2 Water and Wastewater Analysis
mon classes of organic materials produces carbon dioxide and methane in the fol-
lowing ratios:
2
D. P. Smith and P. L. McCarty, Energetic and Rate Effects on Methanogenesis of Ethanol and
Propionate in Perturbed CSTRs. Biotechnology and Bioengineering, 34: 39 (1989).
C H A P T E R 33 Gas Analysis 701
calibration before use, and therefore is particularly suitable when analyses are con-
ducted on an infrequent basis.
Gas chromatographic analysis has the distinct advantage of speed; a gas analy-
sis can be completed in only a few minutes. For this advantage to be utilized, how-
ever, the instrument must have been previously calibrated for each gas of interest,
the oven must have reached a constant temperature, and the detector must be giving
a stable response. Because of these requirements, this method of analysis is more
suited to routine work where gas analyses are conducted several times each week,
making it permissible to keep the instrument always in operating condition. This
method is sufficiently accurate for most practical purposes, although not quite as
good as volumetric analysis. Although it is possible to separate hydrogen, oxygen,
nitrogen, methane, carbon dioxide, and hydrogen sulfide by gas chromatography,
this is not presently practical on a routine basis, using one instrument witii a single
column and detector. In addition, the quantities of hydrogen or hydrogen sulfide oc-
curring in digester gas are usually too small in concentration to be measured with
any accuracy. Gas chromatography is mainly suited to routine analysis to determine
the relative proportions of methane, carbon dioxide, and "air" (nitrogen plus oxy-
gen) occurring in a sample of anaerobic reactor gas. Normally, these are the main
gaseous components of interest.
Carbon Dioxide
Carbon dioxide is measured by bringing a sample of known volume, usually 100
mL, into contact with a solution of potassium hydroxide. The carbon dioxide reacts
with the hydroxide to form potassium carbonate, as shown in the following equation:
C0 2 + 2KOH -» K2C03 + H 2 0 (33.1)
I vol. 0 vol.
-ill
mL
;-»4-íi.íi l . ^ - f S i f t
, six, •. ,.;' ¡ASMS
# Hi ill
":•••::?•• W l l
:i:,xx
Al :
>f;
j- - j . ~ ! - £ f , > •:¿
^ « ' ^ « »
î • ;_£ -•; i .'•--) l';
• I'M
.s ' -f.
í,\; :sJ¡
P mm
^¿-^G^KA¿-:_^
Figure 33.1
Gas-analysis apparatus equipped for separate low-
temperature oxidation of hydrogen and catalytic oxidation
of methane. (Courtesy of Burrell Scientific, Inc.)
In the reaction, the carbon dioxide disappears from the gaseous phase, and the
potassium carbonate formed remains in the liquid phase; therefore, the loss in vol-
ume of the gas is equal to the carbon dioxide content. Additionally, any hydrogen
sulfide present in the gas also combines with the potassium hydroxide,
H2S + 2KOH -» K2S + 2H 2 0 (33.2)
1 vol. 0 vol.
but the volume of hydrogen sulfide is usually so small that its effect may be ignored.
Potassium hydroxide is used instead of sodium hydroxide to absorb carbon
dioxide because of the greater solubility of potassium carbonate. If sodium hydrox-
ide is used, sodium carbonate tends to precipitate. Some of it usually floats and
clogs the capillary passages.
CHAPTER 33 Gas Analysis 703
Oxygen
Theoretically, there is very little possibility of oxygen being present in the gas pro-
duced by anaerobic treatment. However, small amounts may gain entrance to the
sample during the sampling procedure and during charging of the gas-analysis ap-
paratus. It is therefore good practice to analyze for oxygen. The presence of more
than 0.1 to 0.2 percent usually indicates poor technique in sampling and in transfer-
ring the sample to the gas-analysis apparatus.
Oxygen in gas from anaerobic treatment is generally measured by the use of alka-
line pyrogallol. Under alkaline conditions, pyrogallol (1,2,3-trihydroxybenzene) is ox-
idized by oxygen. The end products of the reactions, such as carbon dioxide and or-
ganic acids, are held as potassium salts in the absorbing solution. If all carbon dioxide
has been removed from the sample before bringing it into contact with the alkaline
pyrogallol, any decrease in sample volume will be due to removal of oxygen.
Hydrogen
Hydrogen may be determined separately in the presence of methane by passing the
gas mixture over cupric oxide maintained at a temperature in the range of 290 to
300°C. Under such conditions hydrogen is oxidized to water, but methane is not ox-
idized. The water vapor formed condenses at the temperatures to which the sample
must be reduced for subsequent volume measurements, and therefore loss in vol-
ume after contact with heated cupric oxide is a measure of the hydrogen present.
Methane
After removal of hydrogen, methane may be determined in the residual gas by slow
combustion or by catalytic oxidation. In either case oxygen is required, but the tech-
nique for each is very different,
Catalytic Oxidation A mixture of the gas containing methane and oxygen is used
in this method, and the oxidation is performed catalytically attemperaturesbelow
the kindling point; consequently, an explosion does not occur. The volume of oxy-
gen required is determined from the equation
CH, + 20 2 -» C0 2 + 2H20 (I) (33.3)
I vol. 2 vol. 1 vo!. 0 vol.
(excess)
It will be noted that 2 volumes of oxygen are required to oxidize 1 volume of
methane and that 1 volume of carbon dioxide is produced. The 2 volumes of water
vapor produced are reduced to zero volume at the temperature at which final gas
volumes are measured. Since 2 volumes of oxygen are required for each volume of
methane, at least 2 \ volumes should be used to provide a sufficient excess to carry
the reaction to completion.
After the analyses for carbon dioxide, oxygen, and hydrogen arefinished,the
volume of residual gas containing methane is about 60 to 70 mL if a 100-mL sam-
ple is used initially. It is impossible to mix 60 to 70 mL of gas with 21timesits vol-
ume of oxygen in the equipment provided. The maximum-size sample of methane
ro4 P A R T 2 Water and Wastewater Analysis
that may be used in a 100-mL buret and be mixed with 21timesits volume of oxy-
gen is about 28 mL. The usual procedure is to waste sufficient residual gas to leave
a sample ranging in size from 20 to 25 mL. Some analysts prefer not to waste the
gas but store the excess in the oxygen absorption pipet as a reserve supply in case it
is needed. Oxygen is next admitted to the sample in proper amount, and then the
mixture is brought into contact with the catalyst.
The methane content of the gas used for the combustion may be determined in
two ways. Inspection of Eq. (33.3) will show that 2 volumes of oxygen are used for
each volume of methane oxidized and 1 volume of carbon dioxide results. A con-
traction in gas volume occurs that is equal to the amount of oxygen used. Since this
is equal to two times the methane.
Volume of methane = £ total contraction
The residual gas, after measurement to determine total contraction, still contains
carbon dioxide in a volume equal to the methane originally present. It may be mea-
sured by bringing it into contact with potassium hydroxide and measuring the resid-
ual gas to get the volume absorbed.
Final calculations of methane content must be based upon an adjustment of the
values obtained on the test portion of the methane-bearing sample to the total vol-
ume. For example, if 65 mL remained after carbon dioxide, oxygen, and hydrogen
were removed and 25 mL was used for the methane determination, the values
should be multiplied by a factor of 65/25 to obtain the percentage of methane.
Slow Combustion The separate determination of methane by the slow-combustion
method is not common. The procedure is essentially the same as for a combina-
tion of hydrogen and methane. Analysis of the data is exactly as described under
catalytic oxidation.
Nitrogen
Nitrogen is a relatively inert gas and remains unchanged at the end of the usual gas-
analysis procedure. It is assumed that it is the only component of any significance
that behaves in this manner. It is customary to total the carbon dioxide, oxygen, hy-
drogen, and methane percentages and subtract them from 100. The difference repre-
sents inert gases and is reported as nitrogen.
Sources of Error
There are four major sources of error in gas analysis, as follows:
Collection, Storage, and Handling of Samples Unless special care is taken in
the collection of samples, contamination by air occurs. Samples should always be
collected in glass or metal tubes, particularly if an appreciable time elapses before
analysis can be made. Gum-rubber balloons are not suitable because they are per-
meable to hydrogen and methane. Transfer of gas from the sample tube to the gas-
analysis apparatus requires the use of a displacing fluid, and some modification of
the sample is apt to occur. Also, some air may gain entrance.
C H A P T E R 33 Gas Analysis 70S
Confining Fluid Mercury is the ideal confining fluid because of the insolubility
of all gases in it but, because of its great density, cost, and potential for human
health hazard, it is not usedtoday.For ordinary purposes, a high degree of accuracy
is not needed and less ideal confiningfluidscan be used. Water has much too great a
solvent power for all the gases involved to serve satisfactorily. However, it has been
found that an aqueous solution containing 20 percent sodium sulfate and 5 percent
sulfuric acid has markedly reduced solvent powers. This is the mixture normally
used in portable equipment but does introduce some error in analysis.
Incomplete Combustion of Methane During the combustion of methane, a high
concentration of oxygen is present at the start of the combustion, but as the combus-
tion proceeds, the oxygen concentration decreases markedly, owing to use and to
dilution by the carbon dioxide formed. Unless a volume of oxygen at least 2 \ times
the size of the gas sample is used, an oxygen deficiency may occur. Incomplete
combustion is a common cause of high nitrogen values and of low methane values.
Temperature Changes In gas analysis, small changes intemperaturecan cause
serious errors. This is a special problem during the measurement of hydrogen and
methane, where the reactions are conducted at high temperatures. There is always a
tendency to measure the volume remaining after combustion before the temperature
of the gas has returned to the original value. This may lead to positive errors in
some instances and negative errors in others.
Chromatograph, a curve of peak height versus gas volume may be prepared. The
percentage by volume of a given gas in a sample is then equal to its measured vol-
ume divided by the total of the sample injected. Peak area is also widely used and
can be determined automatically with an integrater.
Thermal-conductivity detectors tend to lose sensitivity with time, with the re-
sult that the peak height for a given volume of gas tends to decrease. This can be
corrected by frequent standardization. However, it has been found that the decrease
in peak height is proportionally the same for methane, carbon dioxide, and air.
Since gas from anaerobic treatment is composed almost entirely of these gases, the
summation of their respective concentrations should total almost 100 percent. Any
significant decrease below this total will indicate a decrease in detector sensitivity,
provided that other analytical errors have not been made. If this is true, then the cor-
rect percentage for each component can be determined by proportioning each up to
give a total of 100 percent. For example, if the measured percentages of the three
components total 90 and the measured methane percentage is 54 percent, its actual
percentage would be 54(100/90) or 60 percent.
Sources of Error
Errors in collection, storage, and handling of samples will be the same as with the
volumetric analysis method. In addition, changes in instrument temperature or car-
rier gas flow rate, changes in detector sensitivity, degeneration of column packing,
and inaccuracies in measurement of volume of injected gas samples will all result in
errors. With gas from anaerobic treatment the summation of the methane, carbon
dioxide, and air percentages should total close to 100 percent. If it does not, then
one of the discussed sources of en-ors might be the cause.
I ! -
•
•r
¡y
: \ ' .
•
-
:
'•¿•m
• •ir_
y i
Xyl Ï
•í .
• -ï-
.?/ -ym : :•
:
ft I
•
< , ;
€• •'.
-£-«.-•
:
.i
.-,• Í? T S
Í
-¿"
II •
Figure 33.2
Tutweiler apparatus used to determine hydrogen sulfide in sludge
gas. (Courtesy of Burretl Scientific, Inc.)
PROBLEMS
33.1 How is gas analysis used in the control of the anaerobic waste treatment process?
33.2 What principle is involved in the separation of gases by chromatography?
33.3 (a) How is the fraction of nitrogen determined in a gas sample using volumetric
analysis?
(b) What sources of error would be of particular concern in determining nitrogen by
this method?
REFERENCE
Clesceri, L. S., A. E. Greenberg, and A. D. Eaton (eds.): "Standard Methods for the
Examination of Water and Wastewater," 20th ed., American Public Health Association,
Washington, DC, 1998.