You are on page 1of 1

SKAN WirelessGT configuration report

Configuration number confE0003-20230823


User Skan

Process
PP1 PP2 PP3 PP4 PP5 PP6 PP7 PP8 PP9 PP10 PP11 PP12 PP13
Process1 SKAN PD 1500Pa 450Sec. 180Sec. 200Pa 100 300Sec. 50Pa No Yes No 0Pa 100Pa
Process2 Process 2 1000Pa 300Sec. 600Sec. 100Pa 25 420Sec. 100Pa No No No 5Pa 100Pa
Process3 Process 3 1000Pa 300Sec. 600Sec. 100Pa 25 420Sec. 100Pa No No No 5Pa 100Pa
Process4 Process 4 1000Pa 300Sec. 600Sec. 100Pa 25 420Sec. 100Pa No No No 5Pa 100Pa
Process5 Process 5 1000Pa 300Sec. 600Sec. 100Pa 25 420Sec. 100Pa No Yes No 5Pa 100Pa
Process6 Process 6 1000Pa 300Sec. 600Sec. 100Pa 25 420Sec. 100Pa No No No 5Pa 100Pa
Process7 Process 7 1000Pa 300Sec. 600Sec. 100Pa 25 420Sec. 100Pa No No No 5Pa 100Pa
Process8 Process 8 1000Pa 300Sec. 600Sec. 100Pa 25 420Sec. 100Pa No No No 5Pa 100Pa
Process9 Process 9 1000Pa 300Sec. 600Sec. 100Pa 25 420Sec. 100Pa No No No 5Pa 100Pa

PP1=Process name PP2=Test start pressure [p1] PP3=Inflate timeout [t1] PP4=Test time [t3] PP5=Pressure drop limit [p3] PP6=Pulse count [n]
PP7=Stabilisation timeout [t2] PP8=Hysteresis level [p2] PP9=Early abort PP10=Process enabled PP11=Small volume process
PP12=Minimum pressure loss[p6] PP13=Maximum pressure rise[p7]

System parameters Security configuration


Auto start test No Guest Operator Administrator Service Supervisor SKAN
Network ID 0 DISPLAY_WGT_STATUS 1 1 1 1 1 1
Report path C:\Users\Siddharth Gurav\Desktop\PDPQ WGT REPORT\ AUTO_START_WITH_RFID 1 1 1 1 1 1
Use Adobe reader as PDF printer No MANUAL_START 0 1 1 1 1 1
Page format A4 DISPLAY_COMPLETED_TESTS 0 1 1 1 1 1
Ethernet bridge support No CREATE_BATCH_REPORT 0 1 1 1 1 1
Window always ontop No PRINT_REPORT 0 1 1 1 1 1
Export process report to "csv" No ABORT_TEST 0 1 1 1 1 1
Include process comment No ALARM_ACK 0 1 1 1 1 1
Enable campaign mode No REVISION_MODE 0 0 0 1 0 1
Use Windows security Yes VALVE_ON_OFF 0 0 0 1 0 1
Domain server name Local computer CALIBRATE_SENSOR 0 0 0 1 0 1
Windows group name PROCESS_SETTINGS 0 0 1 0 1 1
Operator WGT-OPERATOR SYSTEM_SETTINGS 0 0 1 0 0 1
Administrator WGT-ADMINISTRATOR SECURITY_SETTINGS 0 0 1 0 0 1
Service WGT-SERVICE SPARE1 0 0 0 0 0 1
Supervisor WGT-SUPERVISOR SPARE2 0 0 0 0 0 1
Auto logout time (minutes) 0 DEBUG_INTERFACE 0 0 0 0 0 1
Always fail test on Communication error No
Disable SKAN access No
Activate OPC Server No
Window Size 100%
Disable Chart in Process report No

RFID Short name Long name Process WirelessGT ID Label


01135F1C98 GP02 Glove Port 02 1 000195000000509E 211017.005
01135F3335 GP03 Glove Port 03 1
01135EEA48 GP04 Glove Port 04 1
010C5767BA TB01 Test Bucket 01 5
01118A5DE2 TB02 Test Buccket 02 5
010C57530C GP01 9800 Y 9 E4 5 1

Date/signature ______________________________________________________________________
Page 1/1
Filename confE0003-20230823.pdf Created 23/08/2023 14:26:04 Software version 1.0.7