You are on page 1of 9

ORIGINAL RESEARCH

Artificial Intelligence Tool for Detection and Worklist


Prioritization Reduces Time to Diagnosis of Incidental
Pulmonary Embolism at CT
Laurens Topff, MD • Erik R. Ranschaert, MD, PhD • Annemarieke Bartels-Rutten, MD, PhD • Adina Negoita, MD •
Renee Menezes, PhD • Regina G. H. Beets-Tan, MD, PhD • Jacob J. Visser, MD, PhD
From the Department of Radiology (L.T., A.B.R., R.G.H.B.T.) and Department of Psychosocial Research and Epidemiology (R.M.), The Netherlands Cancer Institute, Ples-
manlaan 121, 1066 CX Amsterdam, the Netherlands; GROW School for Oncology and Reproduction, Maastricht University, Maastricht, the Netherlands (L.T., R.G.H.B.T.);
Department of Radiology, St Nikolaus Hospital, Eupen, Belgium (E.R.R.); Faculty of Medicine and Health Sciences, Ghent University, Ghent, Belgium (E.R.R.); Department
of Radiology, La Moraleja Hospital, Madrid, Spain (A.N.); and Department of Radiology and Nuclear Medicine, Erasmus MC, University Medical Center Rotterdam, Rot-
terdam, the Netherlands (J.J.V.). Received July 26, 2022; revision requested September 13; revision received January 13, 2023; accepted February 20. Address correspondence
to L.T. (email: l.topff@nki.nl).

Authors declared no funding for this work.


Conflicts of interest are listed at the end of this article.
See also commentary by Elicker in this issue.

Radiology: Cardiothoracic Imaging 2023; 5(2):e220163 • https://doi.org/10.1148/ryct.220163 • Content codes:

Purpose: To evaluate the diagnostic efficacy of artificial intelligence (AI) software in detecting incidental pulmonary embolism (IPE) at
CT and shorten the time to diagnosis with use of radiologist reading worklist prioritization.

Materials and Methods: In this study with historical controls and prospective evaluation, regulatory-cleared AI software was evaluated to
prioritize IPE on routine chest CT scans with intravenous contrast agent in adult oncology patients. Diagnostic accuracy metrics were
calculated, and temporal end points, including detection and notification times (DNTs), were assessed during three time periods (April
2019 to September 2020): routine workflow without AI, human triage without AI, and worklist prioritization with AI.

Results: In total, 11 736 CT scans in 6447 oncology patients (mean age, 63 years ± 12 [SD]; 3367 men) were included. Prevalence
of IPE was 1.3% (51 of 3837 scans), 1.4% (54 of 3920 scans), and 1.0% (38 of 3979 scans) for the respective time periods. The AI
software detected 131 true-positive, 12 false-negative, 31 false-positive, and 11 559 true-negative results, achieving 91.6% sensitiv-
ity, 99.7% specificity, 99.9% negative predictive value, and 80.9% positive predictive value. During prospective evaluation, AI-based
worklist prioritization reduced the median DNT for IPE-positive examinations to 87 minutes (vs routine workflow of 7714 minutes
and human triage of 4973 minutes). Radiologists’ missed rate of IPE was significantly reduced from 44.8% (47 of 105 scans) without
AI to 2.6% (one of 38 scans) when assisted by the AI tool (P < .001).

Conclusion: AI-assisted workflow prioritization of IPE on routine CT scans in oncology patients showed high diagnostic accuracy and
significantly shortened the time to diagnosis in a setting with a backlog of examinations.

Supplemental material is available for this article.

© RSNA, 2023

I ncidental pulmonary embolism (IPE) is a common co-


morbidity in oncology patients and is related to a high
risk of recurrent venous thromboembolism (1). IPEs can be
occur and in regions where a shortage of radiologists ex-
ists. Consequently, long report turnaround times (TATs)
pose a risk for delayed diagnosis of unsuspected critical
detected at routine contrast-enhanced CT of the chest that findings. Artificial intelligence (AI) applications that can
is performed for indications other than thromboembolic automatically detect critical findings at imaging can be
disease. The pooled frequency of IPE in oncology patients used to prioritize scans in the reading worklist of the radi-
has recently been reported as 3.4% (2). IPEs are associated ologist, with the aim of shortening time to diagnosis and
with poor outcomes, including disease progression and communication with the treating physician (7). AI-based
reduced overall survival (3,4). While an IPE is clinically prioritization tools have been studied for use cases such as
unsuspected, it can be an urgent and life-threatening find- intracranial hemorrhage at CT, acute pathologic abnor-
ing, depending on the patient’s clinical condition and the malities on chest radiographs, and pulmonary embolism
extent of the embolism (4). Timely diagnosis and proper (PE) on dedicated CT pulmonary angiograms (CTPAs),
management of IPEs are therefore essential. with varying results (8–11).
The workload of radiologists has increased substantially To the best of our knowledge, the clinical benefit of
during the past decades, mainly the result of higher vol- a worklist triage tool for IPE has not been investigated.
umes and complexity of imaging examinations (5,6). It Detection of IPEs can be challenging because filling de-
is therefore not unusual for radiology departments to be fects caused by thrombus need to be identified on CT
confronted with a backlog of unreported examinations, scans acquired in the venous phase with suboptimal con-
especially when unexpected surges in imaging requests trast enhancement of the pulmonary arteries. The relative
This copy is for personal use only. To order printed copies, contact reprints@rsna.org
AI Tool for Detection and Worklist Prioritization in Incidental Pulmonary Embolism at CT

marked and returned to the reading worklist. During the third


Abbreviations period (June to September 2020), radiologists were assisted with
AI = artificial intelligence, CTPA = CT pulmonary angiogram, an AI-based worklist prioritization tool. The 15-week intervals
DNT = detection and notification time, IPE = incidental PE, PE =
pulmonary embolism, TAT = turnaround time, TTR = time to read between periods allowed radiologists to get accustomed to the
new instructions and implementation of the AI software.
Summary
Artificial intelligence–based prioritization of radiologist reading Imaging Data
worklists significantly reduced the time to diagnosis of incidental
pulmonary embolism at CT in patients with cancer. Eligible chest CT scans acquired with intravenous contrast
agent were selected using the study description. CT scans were
Key Points acquired in the venous phase, 60–70 seconds after contrast me-
■ Artificial intelligence (AI) software for detecting incidental pul- dium administration. Dedicated CTPAs were not selected and
monary embolism (IPE) at chest CT in patients with cancer were excluded (n = 3) after data collection. Sixty-four scans
showed high diagnostic accuracy in a large sample of 11 736 scans
(sensitivity, 91.6%; specificity, 99.7%; negative predictive value, were excluded because of incomplete or missing image files.
99.9%).
■ In a practice with a backlog of unreported examinations, AI-based AI Software
worklist prioritization reduced the median detection and notifica- Aidoc Medical provided commercially available AI-based im-
tion time of IPE in flagged scans from several days to 1.0 hour.
age analysis software that has been cleared by Conformité Eu-
■ The missed rate of IPE was significantly reduced from 44.8% to
2.6% when radiologists were assisted by the AI tool (P < .001).
ropéenne and the U.S. Food and Drug Administration. The
intended use of the software was specifically the prioritization
Keywords of IPE. The tool was connected with the picture archiving and
CT, Computer Applications, Detection, Diagnosis, Embolism, communication system to immediately analyze eligible imag-
Thorax, Thrombosis ing studies after acquisition. After processing, a positive or neg-
ative result was transferred to the radiologist’s reading worklist
frequency of CT scans that are positive for IPE is also consider- in the radiology information system. Positive examinations
ably lower when compared with the relative frequency of posi- were marked with a bright color in the worklist. Additionally,
tive CTPAs. Therefore, a clinically useful AI tool must perform a heatmap highlighting the suspected abnormality was sent to
with high sensitivity and an acceptable false-positive rate. the picture archiving and communication system.
The purpose of this study was to evaluate the diagnostic ef- Examinations that were not analyzed by the AI software due
ficacy of AI software for the detection and prioritization of IPE to failed data orchestration or technical validation were excluded
at chest CT in terms of diagnostic accuracy and its effect on time (n = 378). Failure of AI analysis was possibly related to image
to diagnosis in a real-world clinical environment. quality (eg, insufficient contrast enhancement, incompatible ac-
quisition parameters).
Materials and Methods
Reference Standard
Study Design and Patients The radiology report was used to categorize all CT scans for
We performed a study to evaluate the diagnostic efficacy of the presence of IPE. Subsequently, two board-certified radi-
AI software to analyze IPE on chest CT scans in clinical prac- ologists (A.N., general radiologist with 3 years of experience;
tice by using data from historical controls and prospectively and L.T., subspecialized in oncologic imaging with 5 years of
analyzed data. The institutional review board approved this experience) independently reviewed CT scans that were cat-
study and waived the requirement for study-specific informed egorized as positive by the radiology report or AI software. The
consent. The AI software was provided by Aidoc Medical. The readers reviewed only imaging data, including heatmap results,
authors are not affiliated with the industry and had control of without any clinical information to label CT scans as positive,
the data and information submitted. negative, or inconclusive. In cases of disagreement between
The study was conducted in the radiology department of a readers, consensus was reached by a third dedicated chest ra-
comprehensive cancer center during three periods of 15 weeks diologist (A.B.R., with 8 years of experience). CT scans that
each (Fig 1). All adult (≥18 years) patients with cancer who un- were categorized as negative by both the radiology report and
derwent CT of the chest with intravenous contrast agent were AI software were not additionally reviewed for the presence or
included. In total, 11 736 chest CT scans were assessed across absence of IPE.
the three periods. During the first period (April to July 2019), The location of the most proximal embolism was identified
radiologists received no AI assistance and no special instructions for each positive CT scan. False-positive detections were classi-
while reporting. During the second period (November 2019 to fied as technical artifact, flow artifact, or abnormalities adjacent
February 2020), radiologists received no AI assistance but were to or within a pulmonary artery.
instructed to screen all newly acquired CT scans for the presence
of IPE. The task of human triage was performed by staff radiolo- Diagnostic Accuracy
gists during working hours. Examinations with IPE were imme- The performance of the AI software in the detection of IPE
diately reported, and examinations without acute findings were was evaluated on both prospectively analyzed and historically

2 rcti.rsna.org ■ Radiology: Cardiothoracic Imaging Volume 5: Number 2—2023


Topff et al

Figure 1: Flowchart shows study design and data selection per time period. AI = artificial intelligence, CTPA = CT pulmonary
angiogram, TP1 = time period 1, TP2 = time period 2, TP3 = time period 3.

collected data to increase the sample size. Sensitivity, specific- rejected for P < .05. Statistical analyses were performed by
ity, negative predictive value, positive predictive value, and ac- a statistician (R.M.) using the R environment for statistical
curacy were calculated. computing (version 3.6.3; https://www.r-project.org) (12).

Time-related End Points Results


Timestamps were registered electronically to calculate the fol-
lowing time intervals: (a) time to process, defined as the time Patient Demographics
interval between forwarding of the study to the AI software A total of 11 736 chest CT scans from 6447 unique patients
and availability of the AI analysis result; (b) detection and no- (mean age, 63 years ± 12 [SD]; 3367 men [52.2%]) were in-
tification time (DNT), defined as the time interval between cluded. Patient demographics per time period are shown in Ta-
availability of the study in the radiologist’s reading worklist and ble 1. Most imaging studies (11 333 of 11 736 [96.6%]) were
the opening of the study by the radiologist; (c) TAT, defined performed in outpatients.
as the time interval between availability of the study in the
worklist and finalization of the report; and (d) time to read Imaging Data
(TTR), defined as the time interval between opening of the Chest CT scans were acquired with seven models from three
study and finalization of the report. To avoid changes in report- manufacturers: Toshiba (5836 scans of 11 736 [49.7%]),
ing behavior, radiologists were not informed about the registra- Siemens (5140 of 11 736 [43.8%]), and Philips (760 of
tion of time-related end points. 11 736 [6.5%]) (Table S1). CT series reconstructed with a
soft-tissue kernel were obtained with a section thickness of
Statistical Analysis 1.0–2.0 mm. In the majority of CT scans (9923 of 11 736
The prevalence of IPE in different time periods was compared [84.6%]), the chest was imaged in combination with the
using a χ2 test. CIs for diagnostic accuracy metrics, such as abdomen and/or neck.
sensitivity and specificity, were computed using a binomial
distribution. A Fisher exact test was used to investigate the Diagnostic Accuracy
reduction in missed IPEs by radiologists with AI versus with- The prevalence of IPE in the sample was 1.2% (143 of 11 736
out AI assistance. scans). The prevalence of IPE was similar across the three
Per time variable (DNT, TAT, and TTR), we were inter- time periods (χ2 test, P = .17). Overall, the AI software re-
ested in comparing values between IPE-positive and -negative turned 131 true-positive, 12 false-negative, and 31 false-pos-
CT scans for every time period. A Student t test was used itive results (Table 2). Three were inconclusive, and the other
to compare each given log-transformed time variable be- 11 559 examinations were categorized as negative by both the
tween groups of positive and negative CT scans. CIs were radiology report and AI software. The AI software showed
determined for these differences between positive and nega- a sensitivity of 91.6% (131 of 143 scans; 95% CI: 86.7,
tive CT scans, per time period. Subsequently, we compared 95.8), specificity of 99.7% (11 559 of 11 590 scans; 95% CI:
CIs among the three time periods. Given that different time 99.6, 99.8), negative predictive value of 99.9% (11 559 of
periods involved different and independent examinations, 11 571 scans; 95% CI: 99.8, 99.9), positive predictive value
nonoverlapping CIs led to the conclusion that differences of 80.9% (131 of 162 scans; 95% CI: 74.7, 86.4), and accu-
were statistically distinct. In all cases, the null hypothesis was racy of 99.6% (11 690 of 11 733 scans; 95% CI: 99.5, 99.7).

Radiology: Cardiothoracic Imaging Volume 5: Number 2—2023 ■ rcti.rsna.org 3


AI Tool for Detection and Worklist Prioritization in Incidental Pulmonary Embolism at CT

Three examinations were marked as


Table 1: Patient Demographics per Time Period
inconclusive by the review panel. The
diagnostic accuracy per time period is Routine Workflow Human Triage Worklist Prioritization
available in Tables S2–S5. Variable without AI (TP1) without AI (TP2) with AI (TP3)
Table 3 lists the location of the most No. of CT scans 3837 3920 3979
proximal filling defect in pulmonary No. of unique patients 3309 3387 3441
arteries of IPE-positive examinations. Age (y)* 62 ± 12 62 ± 12 63 ± 12
In total, 54 of 143 clots (37.8%) were
Sex
located in the main or lobar pulmonary
M 1725 (52.1) 1773 (52.3) 1787 (51.9)
arteries. Fifty-two of these (96.3%) were
detected by the AI software, including F 1584 (47.9) 1614 (47.7) 1654 (48.1)
all main pulmonary clots (n = 9). Two No. of outpatient scans 3712 (96.7) 3786 (96.6) 3835 (96.4)
examples of true-positive detections are No. of inpatient scans 125 (3.3) 134 (3.4) 144 (3.6)
shown in Figure 2. Note.—Unless otherwise noted, data are numbers, with percentages in parentheses. AI =
False-negative findings were lo- artificial intelligence, TP1 = time period 1, TP2 = time period 2, TP3 = time period 3.
cated in the segmental or subsegmen- * Data are means ± SDs.
tal arteries in 10 of 12 examinations
(83.3%). The other two false-negative
findings (16.7%) were located in lobar
Table 2: Diagnostic Accuracy in Detection of IPE by the AI
arteries; however, the clots were small and most likely Software Alone
chronic (Fig 3).
False-positive findings (n = 31) were consistent with Variable IPE Present IPE Absent Inconclusive Total
known mimickers of PE (Table 4). Most false-positive AI positive 131 31 3 165
findings were categorized as flow artifacts (13 of 31 AI negative 12 11 559 0 11 571
[41.9%]) (Fig 4A, 4B). Technical artifacts (seven of 31 Total 143 11 590 3 11 736
findings [22.6%]) were predominantly caused by respira-
Note.—Data are numbers of scans. AI = artificial intelligence, IPE =
tory motion. In 11 of 31 examinations (35.5%), extravas-
incidental pulmonary embolism.
cular abnormalities (eg, lymphadenopathy [Fig 4C, 4D])
or intravascular abnormalities (eg, stump thrombus) led
to false-positive detections. The number of false-negative
and false-positive findings per time period is available in Table 3: Location of the Most Proximal Filling Defect in Pul-
Tables S2–S4. monary Arteries on Positive Incidental Pulmonary Embolism
In 44.8% (47 of 105) of IPE-positive CT scans from the Scans
first two time periods (historically collected data), the find- No. of CT Scans
ing was missed in the radiology report but correctly identi-
fied by the AI software. In comparison, in the third time pe- True-Positive False-Negative Total
riod when radiologists were assisted by the AI software, only Artery Location Scans (n = 131) Scans (n = 12) (n = 143)
one of 38 studies positive for IPE (2.6%) was missed by the Main 9 (6.9) 0 (0) 9 (6.3)
radiologists. This resulted in a 94% reduction (44.8% vs Lobar 43 (32.8) 2 (16.7) 45 (31.5)
2.6%) in missed IPE with use of AI assistance (Fisher exact Segmental 62 (47.3) 6 (50) 68 (47.6)
test, P < .001). Categorization of IPEs missed by the radi- Subsegmental 17 (13.0) 4 (33.3) 21 (14.7)
ologists according to the location of the most proximal clot
Note.—Data in parentheses are percentages.
was as follows: lobar in eight of 48 examinations (16.7%),
segmental in 28 of 48 examinations (58.3%), and subseg-
mental in 12 of 48 examinations (25%). No main IPE was
missed by the radiologists. Time-related End Points
The study design excluded CT scans that were not success- The AI analysis of CT scans was performed without human
fully analyzed by the AI software due to failed data orchestration intervention. The median processing time of CT scans by the
or technical validation. The scans from the first two time periods AI software was 3 minutes, with a maximum of 20 minutes. In
in which the AI analysis failed (63 of 7820 [0.8%]) were all re- 99.5% (3958 of 3979) of studies, the AI result was available to
ported as negative for IPE. However, during the third time pe- the radiologist at the time of opening the study.
riod, the AI software was deployed in the clinical environment, The median DNTs for IPE-positive examinations in all
and scans were analyzed in real time. As a result, a larger propor- patients were 7714, 4973, and 87 minutes for the respective
tion of CT scans (315 of 4294 [7.3%]) was not processed and time periods of routine workflow without AI, human triage
thus excluded. Nevertheless, two of the 315 excluded CT scans without AI, and worklist prioritization with AI. The median
were reported as positive for IPE, with a segmental and subseg- DNTs for IPE-positive examinations in outpatients only were
mental location, respectively. 8950, 5454, and 80 minutes for the respective time periods.

4 rcti.rsna.org ■ Radiology: Cardiothoracic Imaging Volume 5: Number 2—2023


Topff et al

Figure 2: True-positive detection of incidental pulmonary embolism (PE) by the artificial intelligence (AI) software. (A, B) Im-
ages in a 68-year-old woman who underwent routine CT with intravenous contrast agent for outpatient follow-up of melanoma.
(A) Axial CT image shows a large filling defect straddling the bifurcation of the pulmonary trunk (arrow) and extending into both
pulmonary arteries, compatible with an incidental saddle PE. (B) Corresponding AI heatmap highlights the detected abnormality
(red), thereby prioritizing the case in the radiologists’ worklist. (C, D) Images in a 58-year-old woman with a history of rectal cancer
undergoing outpatient follow-up. (C) Axial restaging CT image with intravenous contrast agent shows a small incidental subsegmen-
tal PE in the right lower lung lobe (arrow). (D) Corresponding AI heatmap enables the radiologist to localize the finding (red).

Figure 3: False-negative findings of two chronic lobar incidental pulmonary embolisms (PEs) that were not detected by the
artificial intelligence (AI) software. (A) Axial CT image with intravenous contrast agent in a 70-year-old man (an outpatient) with
urothelial carcinoma shows a small incidental PE (IPE) located against the vessel wall in the right pulmonary artery bifurcation (ar-
row), compatible with a small chronic IPE. (B) Contrast-enhanced coronal CT image in a 62-year-old man (an inpatient) with lung
cancer shows a small eccentric filling defect in the pulmonary artery of the left lower lobe (arrow). These findings were not detected
by the AI software.

When only considering true-positive examinations flagged by overlap between the time period with AI assistance and those
the AI software in the third time period (n = 34), the median without AI assistance, differences were statistically significant
DNT was 62 minutes. The DNT of 29 of the 34 true-positive between the time periods. In contrast, we found no evidence
examinations (85.3%) was less than 6 hours. In comparison, of a difference between routine workflow without AI and hu-
the DNT of the four nonprioritized false-negative examina- man triage.
tions in the third time period ranged from 1280 minutes to Report TAT showed similar results as DNT. The median
12 684 minutes. Figure 5 shows the DNTs of all IPE-negative TATs for IPE-positive examinations were 7772, 4983, and 148
versus -positive studies for each time period. CIs were calcu- minutes for the three respective time periods. When only consid-
lated for the time differences between negative and positive ering true-positive examinations flagged by the AI software, the
studies in each time period (Fig 6). Given that the CIs did not median TAT was 91 minutes.

Radiology: Cardiothoracic Imaging Volume 5: Number 2—2023 ■ rcti.rsna.org 5


AI Tool for Detection and Worklist Prioritization in Incidental Pulmonary Embolism at CT

The median TTR for all examinations was 16 minutes. The prioritization resulted in a significantly reduced median DNT
median TTR for positive examinations was 21 minutes. We and TAT for flagged scans with IPE, from several days to 1.0 and
found no evidence of a difference in TTR between time periods. 1.5 hours, respectively. In contrast, unassisted triage of CT scans
by radiologists did not have a significant effect on the reduction
Discussion of DNT or TAT when compared with the routine workflow.
We evaluated the clinical value of AI software for the analysis of This is likely a result of the time-consuming nature of this task,
IPE on a large sample of chest CT scans (n = 11 736) in oncol- contributing to low yield.
ogy patients. The AI tool accurately detected IPE on chest CT To the best of our knowledge, no other published study has
scans with intravenous contrast agent, with a high sensitivity investigated the diagnostic performance of AI software for the
of 91.6% (131 of 143 scans), specificity of 99.7% (11 559 of detection and prioritization of IPE. Previous studies have as-
11 590 scans), and negative predictive value of 99.9% (11 559 sessed the diagnostic accuracy of deep learning algorithms in the
of 11 571 scans). False-negative classification occurred in 12 of detection of PE on dedicated CTPAs (13–18). For this task, sen-
143 examinations (8.4%) but was limited to segmental, sub- sitivities and specificities ranged from 73% to 96% and 77% to
segmental, and small chronic lobar clots. No IPE in the main 96%, respectively. Although our study focused on the detection
pulmonary arteries was missed by the software. The number
of false-positive detections by the software, 31 of 165 flagged Table 4: Causes of False-Positive Detections by the Arti-
examinations (18.8%), can be considered acceptable because ficial Intelligence Software
radiologists could easily identify false-positive findings by us-
ing the heatmap. In total, only 0.3% of all analyzed examina- No. of CT
tions (31 of 11 736) were falsely positive. The impact of false- Cause Scans (n = 31)
positive alerts on radiologist workflow was therefore limited. Technical artifact 7 (22.6)
During prospective evaluation, the AI software was de- Flow artifact 13 (41.9)
ployed in a clinical environment with a backlog of unreported Abnormality adjacent to a pulmonary artery 9 (29)
examinations to prioritize IPE on routinely acquired chest CT Abnormality within a pulmonary artery 2 (6.5)
scans, mostly obtained in outpatients with known primary ma-
Note.—Data in parentheses are percentages.
lignancy for posttreatment follow-up. The AI-based worklist

Figure 4: False-positive detections by the artificial intelligence (AI) software. (A, B) Images in a 59-year-old woman (an out-
patient) with melanoma who underwent CT with intravenous contrast agent. (A) Axial CT image shows slightly decreased contrast
opacification in a segmental pulmonary artery in the right lower lobe (arrow). This finding was compatible with a flow artifact without
any clinical significance. (B) Corresponding AI software heatmap misclassified the finding as a possible incidental pulmonary em-
bolism (IPE), as highlighted in red. (C, D) Images in a 36-year-old woman (an outpatient) with cervical cancer who underwent CT
with intravenous contrast agent. (C) Axial CT image shows hilar and mediastinal lymphadenopathy. An enlarged right hilar lymph
node shows impression on the right pulmonary arteries (arrow). The finding was misclassified by the AI software as IPE, as shown on
(D) the corresponding axial heatmap in red.

6 rcti.rsna.org ■ Radiology: Cardiothoracic Imaging Volume 5: Number 2—2023


Topff et al

Figure 5: Box plot shows detection and notification times (DNTs) of incidental pulmonary embolism (IPE)–negative versus IPE-positive CT scans
per time period: routine workflow without artificial intelligence (AI), human triage without AI, and worklist prioritization with AI. DNT was markedly
reduced for positive CT scans during the third time period with AI assistance (median DNT, 87 minutes vs routine workflow DNT of 7714 minutes [5
days]). The horizontal line in each box plot indicates the median, and the box corresponds to the IQR. The whiskers indicate minimum and maximum
values in the data. Circles represent outliers. TP1 = time period 1, TP2 = time period 2, TP3 = time period 3.

Figure 6: Graph shows 95% CIs of detection and notification time (DNT) differences between incidental pulmonary
embolism–positive and –negative CT scans per time period: routine workflow without artificial intelligence (AI) (TP1), human
triage without AI (TP2), and worklist prioritization with AI (TP3). The DNT difference was largest for the third time period.
Given that the CIs of the third time period versus the first and second periods did not overlap, differences were significant
between these periods.

of IPE on venous CT scans, which can be considered more chal- and 95.0% specificity. However, there was no significant reduc-
lenging, our sensitivity and specificity were comparable with or tion in report communication times. This is likely related to the
higher than studies identifying PE on CTPAs. Furthermore, ret- overall short TATs of examinations in an emergency department.
rospective studies on diagnostic accuracy might not determine Tools to prioritize the reading worklist would provide the most
the real clinical impact on patient care (18). Schmuelling et al benefit in clinical settings with a high workload and a backlog of
(19) evaluated the clinical implementation of AI software for unreported examinations, as in our situation. The shorter time
prioritization of positive CTPAs in the emergency setting; the to detection of IPE in our study has limited generalizability to
authors showed good diagnostic accuracy, with 79.6% sensitivity practices with short report TATs.

Radiology: Cardiothoracic Imaging Volume 5: Number 2—2023 ■ rcti.rsna.org 7


AI Tool for Detection and Worklist Prioritization in Incidental Pulmonary Embolism at CT

PE is one of the diagnoses that is most commonly missed or consisted solely of oncology patients. The frequency of IPE is
delayed by physicians (20). Detection of IPE on routine contrast- likely lower in the general patient population, which might im-
enhanced chest CT scans can be especially challenging when the pact the clinical relevance of the AI software. Third, the study
IPE is small and isolated. When analyzing historical CT data, we design was not randomized. Although patient volumes in all
found that 44.8% of IPEs were missed by radiologists. Low detec- time periods were similar and unaffected by the COVID-19
tion rates of IPE by radiologists have also been reported in other pandemic, TATs in the radiology department can be affected
studies (21,22). Wildman-Tobriner et al (23) applied a different by many factors, such as staffing levels. To account for varia-
AI algorithm to retrospectively analyze 11 913 CT examinations tions between periods, we calculated the time differences be-
for undiagnosed IPE and found 49 missed IPEs (0.41%), lead- tween positive and negative IPE examinations within each
ing to a missed rate of 38% (49 of 128 IPEs). In our study, we period separately and compared CIs of the difference among
prospectively evaluated the effect of AI assistance on missed IPEs. the periods. Fourth, statistical analysis of the time variables
The number of missed IPEs was reduced to one scan of 38 (2.6%), assumed independence of examinations; however, the study
thereby demonstrating that AI software can assist radiologists to included multiple examinations per patient. Fifth, the study
significantly improve the detection rate of IPE. focused on diagnostic efficacy; we did not evaluate the value on
The clinical relevance and proper management of IPE remain patient outcomes and cost-effectiveness. Future studies should
a subject of debate. It is well known that venous thromboembo- investigate the effect of early diagnosis of IPE on morbidity
lism in oncology patients is associated with high morbidity and and mortality.
mortality (24). Observational studies suggest that the prognosis In conclusion, we demonstrated that commercially available
of IPE is similar to that of symptomatic PE with regard to the AI software had high diagnostic accuracy in the detection of IPE
risk of recurrence and mortality (25). Consequently, treatment on chest CT scans in patients with cancer and was effective in
guidelines for IPE are similar to those in symptomatic PE (26). significantly reducing the time to diagnosis of positive examina-
Radiologic findings, such as thrombus load and central location, tions compared with the routine workflow in a setting with a
have been associated with adverse clinical outcomes in acute PE backlog of unreported scans.
(27). These findings can also help determine IPE severity (4).
Author contributions: Guarantor of integrity of entire study, L.T.; study concepts/
In our study, 37.8% (54 of 143) of IPE-positive scans showed
study design or data acquisition or data analysis/interpretation, all authors; manu-
emboli in the main or lobar pulmonary arteries. Therefore, the script drafting or manuscript revision for important intellectual content, all authors;
benefit of AI-based worklist prioritization for timely assessment approval of final version of submitted manuscript, all authors; agrees to ensure any
questions related to the work are appropriately resolved, all authors; literature re-
and treatment is most evident in these patients. The majority of
search, L.T., E.R.R., J.J.V.; clinical studies, L.T., A.B.R., A.N., J.J.V.; statistical
physicians also treat smaller, more distal incidental emboli in pa- analysis, L.T., R.M., J.J.V.; and manuscript editing, L.T., E.R.R., A.B.R., R.M.,
tients with cancer (28). In our study, most IPEs that were missed R.G.H.B.T., J.J.V.
by radiologists but detected by the AI software were segmental
Data sharing: Data generated or analyzed during the study are available from the
(28 of 48 [58.3%]) or subsegmental (12 of 48 [25%]). We must corresponding author by request.
therefore consider the risk of overdiagnosis, specifically of iso-
lated subsegmental IPE, which if left untreated would cause no Disclosures of conflicts of interest: L.T. No relevant relationships. E.R.R. No
more harm than treatment complications (29). To our knowl- relevant relationships. A.B.R. No relevant relationships. A.N. No relevant relation-
ships. R.M. No relevant relationships. R.G.H.B.T. No relevant relationships. J.J.V.
edge, no randomized controlled trials have assessed the effective- Grant to institution from Qure.ai; consulting fees from Tegus; payment to institu-
ness of anticoagulation therapy in patients with subsegmental tion for lectures from Roche; travel grant from Qure.ai; participation on a data
PE (30). However, recent studies support the use of anticoagula- safety monitoring board or advisory board from Contextflow, Noaber Foundation,
and NLC Ventures; leadership or fiduciary role on the steering committee of the
tion therapy for subsegmental PE in oncology patients (1,31). PINPOINT Project (payment to institution from AstraZeneca) and RSNA Com-
Further studies are needed to assess the relevance of diagnosing mon Data Elements Steering Committee (unpaid); phantom shares in Contextflow
and treating small incidental emboli. and Quibim.

The intended use of the investigated AI software is limited to


workflow triage and not diagnostics (32). Automated worklist References
1. Kraaijpoel N, Bleker SM, Meyer G, et al. Treatment and long-term clinical
prioritization can assist radiologists, who remain responsible for outcomes of incidental pulmonary embolism in patients with cancer: an
verifying flagged examinations and must be aware of possible international prospective cohort study. J Clin Oncol 2019;37(20):1713–1720.
false-negative findings. Our study showed that a considerable 2. Meyer HJ, Wienke A, Surov A. Incidental pulmonary embolism in onco-
logic patients—a systematic review and meta-analysis. Support Care Cancer
number of scans (315 of 4294 [7.3%]) were not analyzed by 2021;29(3):1293–1302.
the deployed AI software due to issues with data retrieval and 3. Lawson P, Raskin S, Soffer S, et al. Incidental pulmonary embolism in CT
technical validation, resulting in delayed diagnosis of IPE in scans of oncological patients with metastatic disease undergoing clinical trials:
frequency and linkage with onset of disease progression (PE-PD association).
these patients. This demonstrates the importance of monitoring Br J Radiol 2020;93(1115):20200591.
and improving the yield of scans analyzed by AI software after 4. Qdaisat A, Kamal M, Al-Breiki A, et al. Clinical characteristics, management,
deployment. and outcome of incidental pulmonary embolism in cancer patients. Blood
Adv 2020;4(8):1606–1614.
The study had limitations. First, the number of false-neg- 5. Kwee TC, Kwee RM. Workload of diagnostic radiologists in the foreseeable
ative IPEs may be underestimated because examinations that future based on recent scientific advances: growth expectations and role of
were categorized as negative by both the radiology report and artificial intelligence. Insights Imaging 2021;12(1):88.
6. McDonald RJ, Schwartz KM, Eckel LJ, et al. The effects of changes in
AI software were not reviewed—this was not feasible due to utilization and technological advancements of cross-sectional imaging on
the high number of included scans. Second, the study sample radiologist workload. Acad Radiol 2015;22(9):1191–1198.

8 rcti.rsna.org ■ Radiology: Cardiothoracic Imaging Volume 5: Number 2—2023


Topff et al

7. Ranschaert E, Topff L, Pianykh O. Optimization of radiology workflow significant effects on report communication times and patient turnaround
with artificial intelligence. Radiol Clin North Am 2021;59(6):955–966. in the emergency department nine months after technical implementation.
8. O’Neill TJ, Xi Y, Stehel E, et al. Active reprioritization of the reading worklist Eur J Radiol 2021;141:109816.
using artificial intelligence has a beneficial effect on the turnaround time for 20. Schiff GD, Hasan O, Kim S, et al. Diagnostic error in medicine: analysis of
interpretation of head CT with intracranial hemorrhage. Radiol Artif Intell 583 physician-reported errors. Arch Intern Med 2009;169(20):1881–1887.
2020;3(2):e200024. 21. Gladish GW, Choe DH, Marom EM, Sabloff BS, Broemeling LD, Munden
9. Baltruschat I, Steinmeister L, Nickisch H, et al. Smart chest x-ray worklist RF. Incidental pulmonary emboli in oncology patients: prevalence, CT
prioritization using artificial intelligence: a clinical workflow simulation. Eur evaluation, and natural history. Radiology 2006;240(1):246–255.
Radiol 2021;31(6):3837–3845. 22. Deniz MA, Deniz ZT, Adin ME, et al. Detection of incidental pulmonary
10. Annarumma M, Withey SJ, Bakewell RJ, Pesce E, Goh V, Montana G. embolism with multi-slice computed tomography in cancer patients. Clin
Automated triaging of adult chest radiographs with deep artificial neural Imaging 2017;41:106–111.
networks. Radiology 2019;291(1):196–202. 23. Wildman-Tobriner B, Ngo L, Mammarappallil JG, Konkel B, Johnson JM,
11. Arbabshirani MR, Fornwalt BK, Mongelluzzo GJ, et al. Advanced machine Bashir MR. Missed incidental pulmonary embolism: harnessing artificial intel-
learning in action: identification of intracranial hemorrhage on computed ligence to assess prevalence and improve quality improvement opportunities.
tomography scans of the head with clinical workflow integration. NPJ Digit J Am Coll Radiol 2021;18(7):992–999.
Med 2018;1(1):9. 24. Schmaier AA, Ambesh P, Campia U. Venous thromboembolism and cancer.
12. R Core Team. R: a language and environment for statistical computing. Curr Cardiol Rep 2018;20(10):89.
Vienna, Austria: R Foundation for Statistical Computing; 2020. https:// 25. Klok FA, Huisman MV. Management of incidental pulmonary embolism.
www.R-project.org. Eur Respir J 2017;49(6):1700275.
13. Huang SC, Kothari T, Banerjee I, et al. PENet—a scalable deep-learning 26. Mulder FI, Di Nisio M, Ay C, et al. Clinical implications of incidental venous
model for automated diagnosis of pulmonary embolism using volumetric thromboembolism in cancer patients. Eur Respir J 2020;55(2):1901697.
CT imaging. NPJ Digit Med 2020;3(1):61. [Published correction appears 27. Meinel FG, Nance JW Jr, Schoepf UJ, et al. Predictive value of computed
in NPJ Digit Med 2020;3:102.] tomography in acute pulmonary embolism: systematic review and meta-
14. Weikert T, Winkel DJ, Bremerich J, et al. Automated detection of pulmonary analysis. Am J Med 2015;128(7):747–59.e2.
embolism in CT pulmonary angiograms using an AI-powered algorithm. 28. den Exter PL, van Roosmalen MJG, van den Hoven P, et al. Physicians’
Eur Radiol 2020;30(12):6545–6553. management approach to an incidental pulmonary embolism: an international
15. Buls N, Watté N, Nieboer K, Ilsen B, de Mey J. Performance of an artificial survey. J Thromb Haemost 2013;11(1):208–213.
intelligence tool with real-time clinical workflow integration—detection of 29. Dobler CC. Overdiagnosis of pulmonary embolism: definition, causes and
intracranial hemorrhage and pulmonary embolism. Phys Med 2021;83:154– implications. Breathe (Sheff) 2019;15(1):46–53.
160. 30. Yoo HH, Nunes-Nogueira VS, Fortes Villas Boas PJ. Anticoagulant treat-
16. Liu W, Liu M, Guo X, et al. Evaluation of acute pulmonary embolism and ment for subsegmental pulmonary embolism. Cochrane Database Syst Rev
clot burden on CTPA with deep learning. Eur Radiol 2020;30(6):3567–3575. 2020;2(2):CD010222.
17. Cheikh AB, Gorincour G, Nivet H, et al. How artificial intelligence improves 31. Yan M, Kieser R, Wu CC, Qiao W, Rojas-Hernandez CM. Clinical factors
radiological interpretation in suspected pulmonary embolism. Eur Radiol and outcomes of subsegmental pulmonary embolism in cancer patients.
2022;32(9):5831–5842. Blood Adv 2021;5(4):1050–1058.
18. Soffer S, Klang E, Shimon O, et al. Deep learning for pulmonary embolism 32. U.S. Food & Drug Administration. K201020. https://www.accessdata.fda.
detection on computed tomography pulmonary angiogram: a systematic gov/cdrh_docs/pdf20/K201020.pdf. Published August 26, 2020. Accessed
review and meta-analysis. Sci Rep 2021;11(1):15814. May 6, 2022.
19. Schmuelling L, Franzeck FC, Nickel CH, et al. Deep learning-based auto-
mated detection of pulmonary embolism on CT pulmonary angiograms: no

Radiology: Cardiothoracic Imaging Volume 5: Number 2—2023 ■ rcti.rsna.org 9

You might also like