Operating instructions

The unit complies with


the corresponding EC
guidelines.
Fig. 1: 12632.00 Cobra4 Sensor-Unit Conductivity+

CONTENTS 1 SAEFTY PRECAUTIONS

1 SAFETY PRECAUTIONS

2 PURPOSE AND CHARACTERISTICS


• Carefully read these operating instructions completely
before operating this instrument. This is necessary to
3 FUNCTIONAL AND OPERATING ELEMENTS avoid damage to it, as well as for user-safety.
3.1 Operating elements • Only use the instrument for the purpose for which it was
3.2 Sensor interface designed.
• Only use the instrument in dry rooms in which there is no
4 NOTES ON OPERATION risk of explosion.
• Protect Cobra4 Sensor-Unit Conductivity+ from dust,
5 HANDLING moisture and vapours. Use a slightly moist lint-free cloth
5.1 Putting into operation to clean the instrument. Do not use aggressive cleaning
agents or solvents.
5.2 Recording of conductivity values
• Take care that no liquid penetrates in through the hous-
5.3 Cleaning and care of conductivity measuring cells
ing openings, as such penetration would result in dam-
5.4 Recording of temperatures age to Cobra4 Sensor-Unit Conductivity+.
• Only use the experimental set-up for the purpose for
6 TECHNICAL SPECIFICATIONS which it is intended.
• Do not open up the instrument.
7 PARTS SUPPLIED • Only connect the data output of Cobra4 Sensor-Unit
Conductivity+ to the measured data recording instru-
8 ACCESSORIES ments specifically intended for it.
8.1 Probes that can be directly connected • The present-day state of software development makes it
8.2 Probes needing an adapter for connection impossible to guarantee that a product is free of faults.
8.3 Calibration solution for conductivity measuring cells PHYWE Systeme GmbH & Co. KG therefore does not
8.4 Storage of conductivity measuring cells take on any liability for damages that occur during the
installation or the use of the instrument.
9 NOTES ON THE GUARANTEE

10 WASTE DISPOSAL

1
www.phywe.com, © All rights reserved 12632.00 / 4808
2 PURPOSE AND CHARACTERISTICS etc. are to be used in the immediate vicinity. The individual
The Cobra4 Sensor-Unit Conductivity+, Conductivity / Tem- connecting leads are each not to be longer than 2 m.
perature Pt1000 serves to measure conductivity values in The instrument can be so influenced by electrostatic charges
aqueous solutions as well as to measure their temperatures. and other electromagnetic phenomena (HF, bursts, indirect
According to the type of application, this Cobra4 Sensor-Unit lightning discharges) that it no longer works within the given
can be connected to the following measured value recording specifications. Carry out the following measures to reduce or
eliminate the effect of such disturbance:
instruments for transmittance of measured values to them:
Ensure potential equalization at the PC (especially with Lap-
tops). Use screening. Do not operate high frequency emitters
• a Cobra4 Wireless-Link (12601.00) for data transfer to a
(e.g. radio equipment or mobile radiotelephones) in the im-
computer by wireless,
mediate vicinity. When a total failure of the instrument occurs,
• a Cobra4 USB-Link (12610.00) for data transfer to a com- unplug it and plug it back in again for a reset.
puter via a USB-cable,
• a handheld Cobra4 Mobile-Link instrument (12620.00).
5 HANDLING
The interface to a measured data recording instrument is at This section describes how to use the Sensor-Unit and re-
the front of the Sensor-Unit, whereby a mechanically secure cord force data. To avoid failure or improper operation,
click connection is ensured by the mushroom-shaped click-on please read carefully through this section.
connector and a hole.
5.1 Putting into operation
The Sensor-Unit is equipped with a five pin DIN socket Please make sure that the current version of measure is in-
("MAB socket") for the connection of sensing units. A com- stalled on your PC.
bined conductivity-temperature electrode Pt1000 (13701.01) Download free of charge via:
with DIN plug ("MAB plug") is usually used for the measure-
www.phywe.com … software … measure
ment of conductivity values. With this, the conductivity and
temperature of the test solution can be simultaneously meas- measure main program
ured. measure module Cobra4
An adapter for conductivity measuring cells with 4 mm Alternatively, the software update can be performed with the
socket / DIN plug (13701.02) enables a conductivity measur- measure DVD.
ing cell with 4 mm plug (18151.02) or a Pt1000 temperature
probe with 4 mm plug (13702.01) to be connected to the Connect your Cobra4 Sensor-Unit Conductivity+, Conductiv-
Sensor-Unit. ity / Temperature Pt1000 to one of the three possible meas-
Should only temperature measurement be required, then a ured value recording instruments via the 15-pin plug connec-
temperature immersion probe of type Pt1000 with DIN plug can tor. Connect a conductivity-temperature probe Pt1000
be connected (12123.00). A list of sensors (conductivity, (13701.01) or a temperature probe (12123.00) to the six pin
temperature), that can be connected to this Sensor-Unit is DIN socket of the Sensor-Unit. The Cobra4 Sensor-Unit is
given in section 8 (Accessories). then immediately ready to start as the connected measure-
ment recording instrument supplies the required voltage.
More detailed information on the steering of measured value
3 FUNCTIONAL AND OPERATING ELEMENTS recording is given in the operating Instructions supplied with
3.1 Operating elements the measured value recording instruments Mobile-Link, USB-
Cobra4 Sensor-Unit Conductivity+, Conductivity / Tempera- Link and Wireless-Link.
ture Pt1000 has no elements for manual operation. Operation
is carried out via the hand-held Cobra4-Mobile-Link or via the 5.2 Recording of conductivity values
computer when Cobra4-USB-Link or Cobra4-Wireless-Link To measure the conductivity of a test solution, dip the meas-
are used. Operation can be started as soon as connection uring head of the conductivity measuring cell completely into
has been made between Cobra4 Sensor-Unit Conductivity+, the solution. For measurements on a series of solutions, start
Conductivity / Temperature Pt1000 and one of the three with the one that has the lowest conductivity if possible. The-
measured value recording instruments. re is always a certain carry-over of solution from one test
solution to the next, so that this procedure keeps possible
errors to an acceptable amount, as the carrying over of a
3.2 Sensor interface trace of a weakly conducting solution to a more strongly con-
The measured values for conductivity and temperature can ducting solution leads to a smaller relative error than the
be read out from the Sensor-Unit at a maximum speed of 1 other way round.
measurement per second (1 Hz). As for all other settings, In each case, on changing from one solution to the next, the
setting of the measurement frequency wanted for the re- measuring head and the electrodes must be rinsed with pure
cording of measured values is made via the particular instru- water (demineralized or distilled).
ment connected to record measured values.
When absolute conductivity values are to be recorded, then
the cell constant of the probe must be previously determined
using a calibration solution of known conductivity (e.g. with
4 NOTES ON OPERATION calibration solution 1413 µS/cm (25°C), (47070.02)). The
This high-quality instrument fulfils all of the technical re- value of this cell constant is then to be entered in the appro-
quirements that are compiled in current EC guidelines. The priate position in the measurement recording instrument. The
characteristics of this product qualify it for the CE mark. value of the temperature coefficient of the solution to be
This instrument is only to be put into operation under special- measured should also to be entered there and temperature
ist supervision in a controlled electromagnetic environment in compensation is to be triggered there. If a conductivity meas-
research, educational and training facilities (schools, universi- uring cell with integrated temperature sensor (e.g. 13701.01)
ties, institutes and laboratories). is used, temperature compensation of the measured values
This means that in such an environment, no mobile phones can be made automatically by activating the appropriate func-

2
www.phywe.com, © All rights reserved 12632.00 / 4808
tion on the measured data recording instrument Resolution 0.1 µS/cm
The solution to be measured should always be moving a little Measurement accuracy± 4% of measured value ± 0.15 µS/cm
during measurement. Keep it in motion by stirring it with a
Measuring range 2 0,0...45000 µS/cm
magnetic stirrer or similar, or by carefully moving the meas-
Resolution 1 µS/cm
uring probe to and fro.
If a measuring cell other than the conductivity-temperature Measurement accuracy ± 4% of measured value ± 3 µS/cm
probe Pt1000 (13701.01) is used, then this can be connected
to the Sensor-Unit by means of an adapter for conductivity
measuring cells with DIN plug and 4 mm sockets (13701.02). Measuring range 3 0,0...1000 mS/cm
The adapter has two 4 mm sockets for connecting conductiv- Resolution 0.01 mS/cm
ity measuring cells and two 4 mm sockets for connecting Measurement accuracy± 4% of measured value ± 0.05 mS/cm
Pt1000 temperature sensors. This assumes, however, that Data rate 1 Hz
the conductivity measuring cell and the temperature probe
are each equipped with two 4 mm plugs.
Temperature:
5.3 Cleaning and care of conductivity measuring cells
Measuring range -20...+150°C
Conductivity measuring cells do not normally age but,
according to the electrode material, the measurement elec- Resolution 0,1 K
trodes could possibly be attacked during measurement in Measurement accuracy ± 0.5 K
certain media (e.g. strong acids, and bases) and this could (in the range 0...+100°C)
lead to damage to them. All measuring cells should therefore Data rate 1 Hz
be thoroughly rinsed with pure water (demineralized or dis- Dimensions:
tilled) after each usage. Should the measurement electrodes Housing dimensions (mm) approx. 60 x 70 x 30
become contaminated with grease or oil, then the measuring Weight 65 g
cell is to be cleaned with warm water and a household clean-
ing agent.
Measuring cells with platinum-clad platinum electrodes, such 7 PARTS SUPPLIED
as conductivity-temperature probe Pt1000 (13701.01), are The extent of delivery is as follows:
supplied with a protective cap that is filled with distilled water.
These measuring cells must not be allowed to dry out, as it • Cobra4 Senor-Unit Conductivity+ 12632.00
could then come to a crumbling off platinum black, which • Operating Instructions
would lead to a change in the cell constant. They must there-
fore always be stored in clean water, e.g. in a protection
sleeve for electrodes, GL 25/12 (37651.15). Do not expose 8 ACCESSORIES
the measurement electrodes to any mechanical measures 8.1 Probes that can be directly connected
during cleaning of the platinized measuring cell. Conductivity-temperature electrode Pt1000,
DIN plug 13701.01
5.4 Recording of temperatures Temperature immersion probe Pt1000,
This Sensor-Unit can measure temperature as well as con- stainless steel, DIN plug 12123.00
ductivity.
Conductivity-temperature probe Pt1000 (13701.01) has a 8.2 Probes needing an adapter for connection
Pt1000 temperature sensor integrated in the probe. As the Adapter for conductivity measuring cells
temperature sensor is inside the measuring head, there is a
short wait until temperature equalization has occurred be- 4 mm socket / DIN plug 13701.02
tween the medium and the temperature sensor, after which Conductivity measuring cell, l = 120 mm,
the latter gives the correct temperature of the medium. 4 mm plug 18151.02
Should you want to only measure temperatures, without si- Temperature probe Pt1000, glass, 4 mm plug 13702.01
multaneous recording of the conductivity, then a temperature
sensor with a DIN plug (12123.00), which has a stainless
8.3 Calibration solution for conductivity measuring
steel jacket, can be used Because of its construction, this
cells
sensor responds more quickly to temperature changes than
the conductivity-temperature probe Pt1000 (13701.01). The following solution can be used for conductivity measuring
The selection of the measurement channel for temperature cell calibration:
is always made via the Mobile-Link, USB-Link or Wireless- Calibration solution 1413 µS/cm (25°C), 460 ml 47070.02
Link measured value recording instrument.
8.4 Storage of conductivity measuring cells
6 TECHNICAL SPECIFICATIONS Protection sleeve for pH electrode, GL 25/12 37651.15
(Typical for 25°C)
Operating temperature range 5... 40°C
Relative humidity < 80%

Conductivity:
The Sensor-Unit has 3 measuring ranges for conductivity.
Selection of a range is made via the instrument connected for
measured value recording:
Measuring range 1 0.0...2500 µS/cm

3
www.phywe.com, © All rights reserved 12632.00 / 4808
9 NOTES ON THE GUARANTEE
We guarantee the instrument supplied by us for a period of
24 months within the EU, or for 12 months outside of the EU.
Excepted from the guarantee are damages that result from
disregarding the Operating Instructions, from improper han-
dling of the instrument or from natural wear.
The manufacturer can only be held responsible for the func-
tion and technical safety characteristics of the instrument,
when maintenance, repairs and alterations to the instrument
are only carried out by the manufacturer or by personnel who
have been explicitly authorized by him to do so.

10 WASTE DISPOSAL
The packaging consists predominately of environmentally
compatible materials that can be passed on for disposal by
the local recycling service.

Should you no longer require this product,


do not dispose of it with the household
refuse.
Please return it to the address below for
proper waste disposal.

PHYWE Systeme GmbH & Co. KG


Abteilung Kundendienst (Customer Service)
Robert-Bosch-Breite 10
D-37079 Göttingen

Phone +49 (0) 551 604-274


Fax +49 (0) 551 604-246