Type D4700
NANOpure Bioresearch
Deionization System
OPERATION MANUAL
AND PARTS LIST
SERIES 747
Control Panel
Water Inlet
Ultrafiltered
NANOpure
ultrapure water system
Water
Drawoff
Valve
Remote
Dispenser
Ultrafilter
Flush Valve
Canisters
2
Table of Contents
3
Safety Information
4
SAFETY INFORMATION
5
Introduction
Introduction
Warning This manual contains the information you will
This device is to be used need to install, operate, and maintain the
with water feeds only. NANOpure Bioresearch Grade, Series 747,
Sanitizing/cleaning agents ultrapure water system manufactured by
must be used in compliance Barnstead|Thermolyne Corporation. It is your
with instructions in this responsibility to read and understand the contents
manual. Failure to comply of this manual prior to installation and use of this
with the above could result in equipment.
explosion and personal The NANOpure Bioresearch is designed to
injury. produce Type I Reagent Grade Water equal to or
exceeding standards established by ASTM, CAP,
and NCCLS with bacterial endotoxin levels below
0.005 EU/ml. Careful attention to the following
instructions will assure that the NANOpure
Bioresearch runs properly and produces water to
specification. Illustrated parts lists are attached
inside. Take a few minutes to familiarize yourself
with the hardware before installation.
The unit includes a constant bleed of the
ultrafiltered reject water. This bleed can be
discontinued if you feel it is not required. See the
appropriate section of this manual if you want to
stop the constant bleed.
Dimensions
Wall mounted models
Width 22" (559 mm)
Depth 14 1/4" (362 mm)
Height 29 1/2" (749 mm)
Bench mounted models
Width 22" (559 mm)
Depth 16 1/4" (413 mm)
Height 32 1/2" (826 mm)
Plumbing Connections
Feed water Inlet 3/8" OD tubing or 1/4" NPTF
Ultrafilter Reject 3/8" hose-barb
Notes:
1
NANOpure Bioresearch will produce Type I water using pretreated water (RO, DI, Distilled) or high quality tap water provided feedwater
suitability is qualified by laboratory analysis and recommended feed flowrate is maintained.
2
Flowrates are dependent on operating conditions and filter usage. Flowrates will also depend on filter compaction.
7
Installation
Choosing a Site
The NANOpure Bioresearch system features
remote controls and dispenser allowing system to
be mounted almost anywhere within the laboratory.
11
6. Hang the NANOpure Bioresearch unit on
Warning the bracket by sliding the mounting pins
Wall composition and into the bracket grooves. Reinstall the hold
construction, as well as down bolt.
fastener type, must be
considered when mounting this 7. Install the top cover by sliding forward.
unit. The mounting surface and Ensure the pins on the cover align with the
fasteners selected must be holes on the forward cabinet section.
capable of supporting a Install the top cover screws and the
minimum of 275 lbs. interlock plug removed in step 1. (See
Inadequate support and/or Figure B, page 9).
fasteners may result in
damage to mounting surface 8. Remove the Remote Dispenser from
and/or equipment. If you are packaging.
unsure of mounting surface
composition, condition and 9. Remove the tubing connector from the
construction, or correct ends of the Remote Dispenser tubing.
fasteners, consult your building Retain for use as replacement parts. See
maintenance group or Figure C, page 10, for identification of the
contractor. connector components.
Do not mount the NANOpure 10. Remove the tubing nut, grab ring, backup
Bioresearch directly over ring and O-ring from Remote Dispenser
equipment that requires connection on NANOpure Bioresearch.
electrical service. Routine Save as replacement parts.
maintenance of this unit may
involve water spillage and 11. Connect the Remote Dispenser tubing to
subsequent electrical shock NANOpure Bioresearch at elbow
hazard if improperly located. connections on NANOpure Bioresearch
(Figure B, page 9). It is not important
which nut on the dispenser is attached to
which connector on the NANOpure
Bioresearch.
12
INSTALLATION
13
INSTALLATION
15
INSTALLATION
Feed
Reject
Product
17
Initial Operation
Initial Operation
Install and rinse cartridges and filters as follows:
Caution 1. Install a D0836 pretreatment cartridge,
with the small opening up, into canister
Ensure O-ring is in place in position one (Figure F). Replace canisters
the canister groove; wet two, three and four.
O-ring prior to securing
canister on the head. Locking pin on canister handle ring must be fully
engaged into hole on head before system is
Do not tighten canister operated (see Figure D, page 14, for proper
handle ring beyond point positioning).
where locking pin and pin
hole line up.
2. Connect electrical service by installing
Secure locking pin before electrical cord to power entry module on
operating. Pin must be fully the right hand portion of the NANOpure
engaged before operating. Bioresearch. (Figure B, page 9.)
21
Normal Operation
Indicator Lights
Display
Membrane Switches
Modes of Operation
The NANOpure Bioresearch has one membrane
panel with three switches which control its
functions.
• Mode Allows you to choose from
resistivity, set point, or temperature.
23
SMART PURITY METER
24
SMART PURITY METER
25
Inlet Cell
Note Installation
To disrupt the flow of water
to the NANOpure The NANOpure Bioresearch system offers as an
Bioresearch, remove the optional feature the ability to monitor the resistivity
inlet valve by depressing the at the NANOpure Bioresearch inlet water. To install
stainless steel thumb pad. the inlet cell:
The insert with the valve is 1. Disconnect feedwater supply. Open the
easily removed. To resupply Remote Dispenser valve to depressurize
the NANOpure Bioresearch unit.
with water, push the insert
into the coupling body and 2. Remove the power cord. Remove the
listen for the click that pump interlock plug or the pump protector.
indicates full closure. Remove the cover screws. Pull the top
cover straight out and remove.
Path of
Inlet Cell
Inlet Cell Well Cable to
Printed
Circuit
Board
Terminal Block
Reference/Inlet Cell
Toggle Switch
8. Operate normally.
Filler panel
29
Remote Dispenser
30
Installing Float or Pressure
Switch
Accessories D0603, D0606 (float switches) and
D2706 (pressure switch) are designed to protect
the NANOpure Bioresearch pump by alerting the
NANOpure Bioresearch of an inadequate feedwater
condition so that the pump can be shut down. Use
the following instructions for installation.
1. Disconnect from electrical service.
32
Maintenance and Servicing
33
MAINTENANCE AND SERVICING
Replacing Fuses
35
MAINTENANCE AND SERVICING
System Sanitization
Frequency of cleaning is difficult to determine
because of the wide variety of feedwater supplies
which can be used, but the need for cleaning can
be easily determined. Whenever cartridges are
replaced, perform a sanitization procedure utilizing
the D50223 sanitization cartridge, a new, easy-to-
use sanitization cartridge.
1. Turn the unit off by depressing the ON/
OFF/Standby button on the control panel
until the display is blank. Disconnect the
power from electrical service.
36
MAINTENANCE AND SERVICING
37
Warning 12. Allow the system to depressurize.
Depressurize system prior to Carefully remove all canisters from the
attempting to remove system by depressing the thumb lever and
canisters. rotating 1/4 turn to your left. Discard the
remaining solutions within the canister. Do
not rinse the canisters.
Caution
Ensure O-ring is in place in 13. Install fresh cartridges in the system as
the groove and wet O-ring indicated in Initial Operation under
prior to securing the canister "Replacing Cartridges." Do not reinstall
on the head. used cartridges (they may contain large
amounts of bacteria).
Secure locking pin before
operating. Pin must be fully 14. Press the control panel “on/off” button to
engaged before operating. start the pump and fill the system. Run
water through the system to drain any
remaining disinfecting solution. A flush of
10 liters is sufficient.
38
MAINTENANCE AND SERVICING
Shutdown
If NANOpure Bioresearch is to be shut down for
an extended period of time, the system should be
completely drained and the cartridges, the ultrafilter
and final filters removed to prevent the growth of
bacteria.
If the system has remained inactive and full of
water, then the system should be drained, sanitized
and new cartridges and final filters installed prior to
use.
39
MAINTENANCE AND SERVICING
40
Troubleshooting Guide
Pump does not run. Display Pump protector (in reservoir), Connect the pump protector or
light. feedwater line pressure switch pressure switch cord to the
or jumper plug not connected receptacle located inside top
to pump interlock. cover. If a Barnstead pressure
switch is installed in the
feedwater line, the pump will
not start until the line pressure
rises to 0.35 kg/cm2 (5 psi).
41
TROUBLESHOOTING GUIDE
Recirculated water will not Exhausted cartridge. Replace all the cartridges as
rinse up to desired purity level. indicated in the REPLACING
CARTRIDGES section (page
21).
Cartridges out of order. Install the cartridges in the
proper order as indicated in
the INITIAL OPERATION
section (page 13).
Cartridges upside down. Install the cartridges right side
up as indicated In the INITIAL
OPERATION section (page
13).
Feed water bypassing Be sure that small O-ring
cartridge(s). inside head is not damaged
and is properly installed.
Check valve malfunctioning. Remove tubing from check
valve; turn unit on. If water
leaks from end of check valve,
wash to remove any
particulates.
Display reads "Err" when Resistivity cell disconnected or Check resistivity cell wiring.
checking resistivity. wired improperly.
Air in system. Purge air from system by
opening Remote Dispenser
and/or Ultrafiltered Water
Drawoff.
System electronics or cell out Check resistivity of reference
of calibration. cell. If resistivity displayed is
not between 9.7 and 10.3
electronics need recalibration.
If resistivity reading is proper,
clean cell and reinstall. If
problem persists, replace cell.
No constant reject out of Blocked reject flow device. Remove flow restrictive device
Ultrafilter reject. and back flush.
Display reads E_A. System electronics failure. Replace main PC board.
42
TROUBLESHOOTING GUIDE
Short cartridge life. Cartridges being used are Check the expiration date.
beyond expiration date. Cartridges begin to lose
capacity after being stored two
years from the date of
manufacture. Replace the
cartridges with unexpired
ones.
Change in feedwater If a Barnstead ROpure is the
characteristics. feedwater source, check that
the membrane is functioning
properly.
If a Barnstead Still is the
feedwater source, ensure that
the distillate temperature to
the NANOpure Bioresearch
does not exceed 49°C
(120°F).
If tap water is the feedwater
source, check the quality of
the water. In some cases the
quality of the water will change
with the seasons. Changing
the source (city water to well
water, or well water to city
water) will result in a water
quality change.
If feedwater is from a central
water purification system,
verify water quality and proper
functioning of the system.
43
Replacement Parts Listing
Recommended Spares
Consumables
Consumable parts are those required to support the day-to-day operation of this
equipment.
Barnstead |Thermolyne establishes two types of consumables; those items that must
periodically be replaced to maintain performance (filters, resin cartridges, etc.) and other
items of limited life (indicator lights, fuses, etc.) that you can expect to replace on a more or
less random basis. Where practical, Barnstead |Thermolyne recommends the frequency of
replacement, or provides information on life expectancy from which you may calculate a
replacement interval compatible with your usage pattern.
The replacement of consumable parts is discussed in the Maintenance and Servicing
section to assist you in accomplishing your own service.
Consumables may be ordered separately and in some cases, as an expendables kit.
Check with your Barnstead/Thermolyne representative for additional information on the
expendables kit.
Description Catalog No. Recommended Quantity
Ultrafilter Water Drawoff Final Filter FL703X2 2
Remote Dispenser Final Filter D3751 2
Pretreatment Cartridge D0836 1
Ultrapure SG Cartridge D5027 2
OrganicFree Cartridge D5021 1
3.0 Ampere Slow Blow Fuse, 115 Volt 04455 1
2.0 Ampere Slow Blow Fuse, 230 Volt 04420 1
Teflon® Tape, Roll 06078 1
Printed Circuit Board Fuse, 100, 120V FZX43 1
Printed Circuit Board Fuse, 240V FZX48 1
Sanitization Cartridges (pkg of 3) D50223 1
44
REPLACEMENT PARTS LISTING
Safety Stock
For critical applications where performance with minimum downtime is required,
Barnstead |Thermolyne recommends that you maintain a local stock of those parts listed in
the General Maintenance parts and Safety Stock sections.
46
Exploded View
100 & 120 Volt Models
Key Part # Description
1 DL687X1A Top cover
2 05766 1/4" OD x 1/4 NPT Elbow
3 PU733X1A Pump assembly, 100/120V
4 CRX72 Cord set 100/120V
5 04247 Pump interlock plug
6 CUX8 Quick Disconnect
7 CUX9 Quick Disconnect Insert
8 PC550X3A PC board, reference call
9 SW550X1A Display
10 WH550X2 Interconnect cable 10' (WH550X1 6")
11 PC550X1A Main PC board, 100/120V
12 02280 Pressure reducing valve
13 02214 Stainless steel check valve
14 BR550X2 Head end fitting
15 06440 O-ring between heads
16 FP550X1 Head connector pins
17 BR550X4 Head connector
18 BK550X2 Head
19 GSX27 Head o-ring
20 GSX28 Head to canister o-ring
21 HN550X1A Canister handle ring
22 CS550X1 Canister
23 CV747X1A Auxiliary Drawoff
24 PM703X3 Luer Connector
25 DL630X18 Ultrafilter housing
26 E550X1A Resistivity cell
27 CS895X3 Remote Dispenser
30 GSX29 O-ring
31 D4740 Floor stand
32 CEX172 Power Entry Module
33 CAX44 Filter, Power Supply
34 RY747X1A 100/120V Solenoid Valve
Not Shown
TN687X1A Transformer
02273 Ultrafilter Flush Valve
PMX99 Constant Bleed Restrictor
FL703X2 Filter
48
Wiring Diagrams