Professional Documents
Culture Documents
Research 1
Research 1
Measuring inter-arm differences (IADs) in blood pressure interpretation with undesirable consequences. For instance,
(BP) is recommended by numerous guidelines to be per- overestimation of the IAD could lead to unnecessary referral
formed at each first visit.1–6 This is because significant IAD to a specialist and cause an unnecessary burden for healthcare.
in BP may indicate the presence of congenital heart disease, The present review therefore aimed at investigating the influ-
peripheral vascular disease, unilateral neurological, muscu- ence of methodology of BP measurement on the IAD value.
loskeletal abnormalities, or aortic dissection.7 However, even
when the IAD has seemingly no pathological background, Methods
relevant IADs (i.e., ≥10 mm Hg) are still important to know, as Identification of papers. For performing the present meta-
office measurements consequently performed at the arm with analysis the Prisma guidelines were used as a reference.10 We
the lowest BP can lead to a wrong diagnosis and undertreat- performed systematic searches for IADs in PUBMED, Medline,
ment of hypertension.8 In addition, to verify the effectiveness EMBASE, and the Cochrane databases using the following
of antihypertensive therapy it is of clinical importance that BP keywords: inter-arm measurements, double-arm measure-
is measured in the same arm on all sequential occasions.9 ments, IAD, peripheral vascular disease, subclavian stenosis,
Despite the large-scale recommendation of performing double- subclavian steal syndrome. Additional studies were found from
arm measurements the procedure is not clearly described in the reference lists of identified articles and reviews. Two investiga-
guidelines. The lack of a clear description about the preferred tors (W.J.V. and T.T.) independently screened the full text of all
method has led to a wide variation in IAD assessment and potentially relevant articles. In case of disagreement the article
was discussed to reach consensus.
1Microlife Corporation, Taipei, Taiwan; 2Department of Clinical Epidemiology
and Technology Assessment, University Hospital Maastricht, Maastricht, Study selection. We included only those papers that fulfilled
The Netherlands; 3Radboud University Nijmegen, Medical Center,
Department of General Internal Medicine, Nijmegen, The Netherlands.
the following criteria:
Correspondence: Willem J. Verberk (Willem.verberk@microlife.ch)
Received 21 October 2010; first decision 13 November 2010; accepted 11 June 2011. 1. Papers that were published in either the English or Dutch
© 2011 American Journal of Hypertension, Ltd. language.
2. The procedure of assessing IAD was described in suffi- 688 Citations identified in PubMed,
cient detail. Medline, Cochrane and Embase
Statistical analysis. Analyses were performed with meta-anal- 35 Full-text articles considered for
inclusion
ysis random-effects meta-regression using aggregate-level data
(“metareg”) in Stata version 9.2 Texas. Results were pooled 11 Articles identified from references
weighted with inverse variances (direct pooling).11 Because
age, gender, and BP value of the right arm might influence the 46 Full-text articles considered for
inclusion
difference in BP, these variables were also entered in the regres-
sion model to reduce heterogeneity. When standard deviations
present analysis, this study compared different methods and investigated reproducibility; c, measurements not patients.
Table 2 | Differences in prevalences inter-arm difference (IAD) between groups divided by measurements, device and number of
readings
Overall mean Measurements Device Readings
Prevalences (%) (95% CI) RR: Seq/Sim P RR: Man/Auto P RR: 1/≥2 P
Systolic
IAD ≥10 mm Hg (%) 13.6 (10.1–18.4) 2.2 (1.4–3.6) <0.01 2.1 (1.1–3.9) <0.05 2.0 (1.1–3.8) <0.05
IAD ≥20 mm Hg (%) 4.2 (2.4–7.5) 4.8 (1.1–21.9) <0.05 4.4 (1.8–10.8) <0.01 4.3 (1.6–11.4) <0.01
Diastolic (mm Hg)
IAD ≥10 mm Hg (%) 6.5 (3.8–11.2) 2.5 (1.0–6.3) <0.05 3.7 (1.6–8.6) <0.01 4.4 (1.7–11.4) <0.01
IAD ≥20 mm Hg (%) 0.7 (0.2–2.4) 0.4 (0–18.4) 0.64 17.4 (0.2–1,445.1) 0.21 — —
Table 3 | Differences in average and absolute inter-arm difference (IAD) values between groups divided by measurements, device and
number of readings
Overall Measurements Device Readings
average Seq–Sim P Man–Auto P 1–≥2 P
Systolic
IAD absolute 5.4 ± 1.7 0.8 ± 3.4 0.81 4.0 ± 6.4 0.55 1.2 ± 4.1 0.78
(mm Hg)
IAD (mm Hg) 0.57 ± 1.9 0.5 ± 3.2 0.87 0.4 ± 6.0 0.94 0.6 ± 4.9 0.91
Diastolic (mm Hg)
IAD absolute 3.6 ± 1.2 –0.6 ± 2.5 0.81 1.2 ± 6.7 0.86 0.4 ± 3.0 0.91
(mm Hg)
IAD (mm Hg) 0.5 ± 1.3 0.6 ± 2.3 0.79 −0.7 ± 3.8 0.86 0.1 ± 3.3 0.98
All values are provided as mean ± s.e.m.
1–≥2 readings, differences between studies that performed 1 and 2 or more readings to determine IAD. Man–Auto, differences between a manual and automatic device; Seq–Sim
indicates differences between average and absolute IAD values obtained with sequential and simultaneous measurements.
was performed13,16,18,20–23 (n = 2,515) to determine IAD and 2.0 to 4.4). For average and absolute IAD values there were no
16 studies8,9,14,17,19,20,24–33 (n = 12,339) provided two or more s ignificant differences (Table 3).
measurements. One study was divided in subpopulations There was a significant (P < 0.01) and negative correlation
which were measured one and three times.20 of the systolic IAD ≥10 mm Hg and the number of readings
As shown in Table 2 the method of IAD performance had a (Figure 3).
significant influence on systolic (IAD ≥10 and 20 mm Hg) and Overall, the method (simultaneous or sequential), the
diastolic (IAD ≥10) prevalence: sequential measurements led device, and the number of measurements had no significant
to a significant higher prevalence than simultaneous meas- effect on mean and mean absolute IADs.
urements (RR ranges from 2.2 to 4.8), the use of a manual BP
device led to a significant higher prevalence than when using Patient characteristics
an automated device (RR ranges from 2.1 to 4.4). There was no correlation with systolic mean absolute and
However, there was a strong relationship between devices mean IAD and BP values, age, and gender or for diastolic IADs
and measurement methods: all, except one, simultaneous and IAD prevalences (data not shown).
measurements were performed with automated devices
(Figure 2). For this reason, we performed a post hoc multi- Discussion
variate analysis using both method (simultaneous or sub- The findings in the present paper showed that the average
sequent) and devices (automated or manual) as covariates. mean absolute IAD is 5.4 and 3.6 mm Hg for systolic and
Results for systolic IAD prevalence showed that the method diastolic BP, respectively and 14% of all subjects had a systo-
(RR 1.9 (95% CI: 1.2–3.2), P < 0.05) but not the device (RR 1.5 lic IAD of 10 mm Hg or more. The number of subjects with
(95% CI: 0.9–2.7), P = 0.13) had a significant influence on the a systolic and diastolic IAD ≥10 mm Hg and systolic IAD
prevalence. ≥20 mm Hg was significantly lower when BP was measured
With regard to the number of readings it was shown that with an automatic instead of a manual device, when meas-
one reading resulted in a significant higher prevalence than urements were performed simultaneously instead of sequen-
when two or more readings were performed (RR ranges from tially and with two or more measurements than with only one
Clark, 2007 29
0
Prevalence systolic IAD > 10 mm Hg
Figure 2 | Forest plot of the prevalences of subjects with a systolic inter-arm difference (IAD) of ≥10 mm Hg. The studies are grouped by device (automated or
manual) and measurement method (simultaneous or subsequent).
70
Systolic IAD ≥10 mm Hg significant differences shown when groups were separated
according to different methods. There were no significant dif-
60
ferences between BP values on the right and left arm.
50
Prevalence (%)
40
Strength and weakness of the paper
30 r 2 = 0.11; P < 0.01 This study should be interpreted within the context of its limi-
20 tations. Studies showed a substantial heterogeneity, there was a
10 wide disparity among several methods and subjects and it was
0 not always clear how the patient population had to be inter-
0 2 4 6 8
preted and what the main reason was (method, population
Readings (n)
device) for finding differences in value and prevalence of IAD.
Figure 3 | The prevalence of subjects with a systolic inter-arm difference (IAD) For this reason it might be questioned whether these studies
of ≥10 mm Hg related to the number of readings, r2 values are calculated with should be pooled for analysis at all. However, the authors are
log transformation. of the opinion that despite the large heterogeneity the pooled
analysis is a useful method as this heterogeneity is mainly
r eading. In addition, there was a significant trend toward a related to the different measurement procedures used (device,
lower prevalence of systolic IAD ≥10 with each extra measure- method, readings) and thus reflects clinical practice. Seen
ment. Because in all except one study simultaneous measure- within the limitations, the analysis therefore provides use-
ments were performed with automated devices, it appeared ful clinical information. In some papers standard deviations
that the method but not the device had a significant influ- or variances were not provided so that the highest standard
ence on outcome. For mean and absolute IADs there were no deviations from all studies was imputed. Studies that reported
more pairs of readings are more likely to have used automated coat effect which causes the first measurement to be higher
rather than manual methods. The same studies are likely to than the next.27,42,43 In addition, BP is a variable hemody-
have contributed to each relevant section of the results. These namic phenomenon that constantly fluctuates over time, mak-
items cause a limit to the conclusions that can be drawn inde- ing sequential measurements difficult to compare.
pendently about each facet of the measurement technique. On In order to exclude the possibility of erroneous IAD
the other hand the use of automated devices seems strongly determination due to sequential performance some inves-
related to simultaneous measurements as automated devices tigators have performed simultaneous double-arm meas-
facilitate the simultaneous measurements, whereas simulta- urements with two observers44,45 or with two automatic
neous measurements with manual devices are very difficult monitors.8,9,13,14,17,26,28,29,31,46 Although the latter seems a
to accomplish in clinical practice. In addition, two observers good idea a minor bias can still be introduced due to a delay
are less likely to perform measurements simultaneously than between readings and inter-device differences even when
when one automated device with two cuffs or two automated devices are from the same brand.9 Changing the devices
Clinical significance of inter-arm measurements Disclosure: W.J.V. is an official employee of Microlife Corporation (Taipei,
The present data show that 14% of the subjects had an IAD Taiwan). A.G.H.K. and T.T. declared no conflict of interest.
of 10 mm Hg or more for systolic BP. This seems to justify the 1. Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, Jones DW, Kurtz T,
recommendation in the guidelines and from many investiga- Sheps SG, Roccella EJ. Recommendations for blood pressure measurement
in humans and experimental animals: part 1: blood pressure measurement in
tors9,17,18,21,24,26,28,30 to perform BP measurements on both humans: a statement for professionals from the Subcommittee of Professional
arms at each first visit. The importance of determining the and Public Education of the American Heart Association Council on High Blood
IAD is confirmed by the finding that the mean absolute IAD Pressure Research. Circulation 2005; 111:697–716.
2. 1999 World Health Organization-International Society of Hypertension
was 5 and 4 mm Hg for systolic and diastolic BP, respectively. Guidelines for the Management of Hypertension. Guidelines Subcommittee.
Although an IAD of 5 mm Hg may not be clinically relevant J Hypertens 1999; 17:151–183.
for an individual patient on a population basis this difference 3. Mancia G, De Backer G, Dominiczak A, Cifkova R, Fagard R, Germano G, Grassi G,
Heagerty AM, Kjeldsen SE, Laurent S, Narkiewicz K, Ruilope L, Rynkiewicz A,
has been associated with e.g., a 34% difference in stroke.49 In Schmieder RE, Boudier HA, Zanchetti A, Vahanian A, Camm J, De Caterina R,
addition, results from a health survey led to the estimation
19. Lazar J, Holman S, Minkoff HL, Dehovitz JA, Sharma A. Interarm blood pressure 37. O’Brien E, Petrie J, Littler W, de Swiet M, Padfield PL, O’Malley K, Jamieson M,
differences in the women’s interagency HIV study. AIDS Res Hum Retroviruses 2008; Altman D, Bland M, Atkins N. The British Hypertension Society protocol for the
24:695–700. evaluation of automated and semi-automated blood pressure measuring
20. Kristensen BO, Kornerup HJ. Which arm to measure the blood pressure? Acta Med devices with special reference to ambulatory systems. J Hypertens 1990; 8:
Scand Suppl 1982; 670:69–73. 607–619.
21. Cassidy P, Jones K. A study of inter-arm blood pressure differences in primary care. 38. Clark CE, Campbell JL, Evans PH, Millward A. Prevalence and clinical implications
J Hum Hypertens 2001; 15:519–522. of the inter-arm blood pressure difference: A systematic review. J Hum Hypertens
22. Clark CE, Powell RJ. The differential blood pressure sign in general practice: 2006; 20:923–931.
prevalence and prognostic value. Fam Pract 2002; 19:439–441. 39. Shadman R, Criqui MH, Bundens WP, Fronek A, Denenberg JO, Gamst AC,
23. Pesola GR, Pesola HR, Lin M, Nelson MJ, Westfal RE. The normal difference in McDermott MM. Subclavian artery stenosis: prevalence, risk factors, and
bilateral indirect blood pressure recordings in hypertensive individuals. Acad association with cardiovascular diseases. J Am Coll Cardiol 2004; 44:618–623.
Emerg Med 2002; 9:342–345. 40. Aboyans V, Kamineni A, Allison MA, McDermott MM, Crouse JR, Ni H, Szklo M,
24. Mendelson G, Nassimiha D, Aronow WS. Simultaneous measurements of blood Criqui MH. The epidemiology of subclavian stenosis and its association with
pressures in right and left brachial arteries. Cardiol Rev 2004; 12:276–278. markers of subclinical atherosclerosis: the Multi-Ethnic Study of Atherosclerosis
25. Clark CE, Campbell JL, Powell RJ. The interarm blood pressure difference as (MESA). Atherosclerosis 2010; 211:266–270.