You are on page 1of 8

nature publishing group original contributions

See reviewer commentary page 1189

Blood Pressure Measurement Method and Inter-Arm


Differences: A Meta-Analysis
Willem J. Verberk1, Alfons G.H. Kessels2 and Theo Thien3

Downloaded from https://academic.oup.com/ajh/article/24/11/1201/2730264 by guest on 01 November 2022


Background respectively), when using a manual instead of an automated device
Screening for inter-arm difference (IAD) of blood pressure (BP) at (2.1 (95% CI: 1.1–3.9), P < 0.05; 4.4 (95% CI: 1.8–10.8), P < 0.01 and
each first visit is recommended by numerous guidelines whereas 3.7 (95% CI: 1.6–8.6), P < 0.01, respectively) and when performing
it is unclear whether the method by which IAD is measured has only one BP measurement instead of multiple (2.0 (95% CI: 1.1–3.8),
significant influence on the IAD value. P < 0.05; 4.3 (95% CI: 1.6–11.4), P < 0.01 and 4.4 (95% CI: 1.7–11.4),
P < 0.01, respectively).
Methods
A systematic review is made of the studies reporting on double- Conclusion
arm measurements and the association of IAD with procedure Screening for IAD of BP is important but the measurement
characteristics (Medline/PubMed, Embase, and Cochrane Library). methodology has a major influence on IAD results. To prevent
overestimation and observer bias IAD should be assessed
Results simultaneously at both arms, with one or two automatic devices and
The mean absolute IAD was 5.4 ± 1.7 and 3.6 ± 1.2 mm Hg for systolic multiple readings should be taken.
and diastolic BP, respectively. Of all subjects 14% had a systolic IAD
Keywords: automatic oscillometric device; blood pressure;
≥10 mm Hg, 4% a systolic IAD ≥20 mm Hg, and 7% a diastolic IAD
blood pressure measurement; double-arm measurements;
≥10 mm Hg. The relative risk (RR) of obtaining a systolic IAD ≥10 and
hypertension; inter-arm difference; simultaneous measurements
20 mm Hg and a diastolic IAD ≥10 mm Hg is higher when measuring
sequentially instead of simultaneously (2.2 (95% CI: 1.4–3.6), P < 0.01; American Journal of Hypertension, advance online publication 21 July 2011;
4.8 (95% CI: 1.1–21.9), P < 0.05 and 2.5 (95% CI: 1.0–6.3) P < 0.05, doi:10.1038/ajh.2011.125

Measuring inter-arm differences (IADs) in blood pressure interpretation with undesirable consequences. For instance,
(BP) is recommended by numerous guidelines to be per- overestimation of the IAD could lead to unnecessary referral
formed at each first visit.1–6 This is because significant IAD to a specialist and cause an unnecessary burden for healthcare.
in BP may indicate the presence of congenital heart disease, The present review therefore aimed at investigating the influ-
peripheral vascular disease, unilateral neurological, muscu- ence of methodology of BP measurement on the IAD value.
loskeletal abnormalities, or aortic dissection.7 However, even
when the IAD has seemingly no pathological background, Methods
relevant IADs (i.e., ≥10 mm Hg) are still important to know, as Identification of papers. For performing the present meta-
office measurements consequently performed at the arm with ­analysis the Prisma guidelines were used as a reference.10 We
the lowest BP can lead to a wrong diagnosis and undertreat- performed systematic searches for IADs in PUBMED, Medline,
ment of hypertension.8 In addition, to verify the effectiveness EMBASE, and the Cochrane databases using the following
of antihypertensive therapy it is of clinical importance that BP keywords: inter-arm measurements, double-arm measure-
is measured in the same arm on all sequential occasions.9 ments, IAD, peripheral vascular disease, subclavian stenosis,
Despite the large-scale recommendation of performing double- subclavian steal syndrome. Additional studies were found from
arm measurements the procedure is not clearly described in the reference lists of identified articles and reviews. Two investiga-
guidelines. The lack of a clear description about the preferred tors (W.J.V. and T.T.) independently screened the full text of all
method has led to a wide variation in IAD assessment and potentially relevant articles. In case of disagreement the article
was discussed to reach consensus.
1Microlife Corporation, Taipei, Taiwan; 2Department of Clinical Epidemiology

and Technology Assessment, University Hospital Maastricht, Maastricht, Study selection. We included only those papers that fulfilled
The Netherlands; 3Radboud University Nijmegen, Medical Center,
Department of General Internal Medicine, Nijmegen, The Netherlands.
the following criteria:
Correspondence: Willem J. Verberk (Willem.verberk@microlife.ch)
Received 21 October 2010; first decision 13 November 2010; accepted 11 June 2011. 1. Papers that were published in either the English or Dutch
© 2011 American Journal of Hypertension, Ltd. language.

AMERICAN JOURNAL OF HYPERTENSION | VOLUME 24 NUMBER 11 | 1201-1208 | November 2011 1201


original contributions Inter-arm blood pressure differences

2. The procedure of assessing IAD was described in suffi- 688 Citations identified in PubMed,
cient detail. Medline, Cochrane and Embase

3. Data of the average left and right BP values or the IAD


653 Excluded because abstract title and
value or number of relevant IADs were provided. abstract did not fulfil inclusion criteria

Statistical analysis. Analyses were performed with meta-anal- 35 Full-text articles considered for
inclusion
ysis random-effects meta-regression using aggregate-level data
(“metareg”) in Stata version 9.2 Texas. Results were pooled 11 Articles identified from references
weighted with inverse variances (direct pooling).11 Because
age, gender, and BP value of the right arm might influence the 46 Full-text articles considered for
inclusion
difference in BP, these variables were also entered in the regres-
sion model to reduce heterogeneity. When standard deviations

Downloaded from https://academic.oup.com/ajh/article/24/11/1201/2730264 by guest on 01 November 2022


24 Studies excluded because not
were missing in the study, we imputed the highest standard fulfilling inclusion criteria (appendix I)

deviations from all studies. In case of “zero-counts” 0.5 was


added to the specific cell.12 22 Studies included in analysis

There were two studies13,14 that performed both the simul-


taneous and the sequential method. When relevant for overall Figure 1 | Flow chart of article selection.
comparison between the simultaneous and sequential method,
we used data from both methods. The Kolmogorov–Smirnov determine IAD measurements left and right arm BP measure-
test was performed to test whether distribution deviated from ments were performed sequentially or simultaneously. When
normal and when this was the case log-transformation was performed simultaneously all studies but one24 used auto-
performed for analysis and the exponentiated coefficients were mated devices. On average BP was similar in the right and left
interpreted as the relative risk (RR). arm for systolic (136 ± 5 and 135 ± 5 mm Hg; P = 0.68) and
If studies performed inter-arm measurements in multiple diastolic BP (78 ± 3 and 77 ± 3 mm Hg; P = 0.61), respectively.
visits we used only the first visit. Heterogeneity was tested
using I2 statistics and I2 >0.5 was arbitrarily chosen as indicat- IADs
ing notable heterogeneity.15 Prevalences of subjects with a systolic IAD ≥10 and 20 mm Hg
were provided in 19 (refs. 8,9,13,16–20,23–33) (n = 14,047)
Results and 13 studies9,16,17,19–22,25,27,29,30,32,33 (n = 6,508), respectively.
Data selection Seventeen studies8,9,16,17,19–22,24,25,27–33 (n = 12,956) showed
As shown in Figure 1 literature research identified 688 the number of subjects with a diastolic IAD ≥10 mm Hg and
abstracts from which 46 papers were selected by two authors six studies9,17,20,27,32,33 (n = 4,447) a diastolic IAD ≥20 mm Hg.
(W.J.V. and T.T.). Of these 24 papers were excluded after care- Systolic IAD values (right–left BP values) were provided or
ful consideration. These papers, with reason for exclusion, are could be calculated in 11 studies9,14,17,21,25,28–33 (n = 8,109).
provided in Supplementary Appendix S1 online. Diastolic IAD values in eight studies14,17,18,21,28,29,32,33
(n = 2,696). Absolute systolic and diastolic IADs were given in
Description of studies seven studies9,18,19,22,26,27,29 (n = 5,528). Kolmogorov–Smirnov
The 22 cross-sectional studies that were selected for analysis analysis showed that the IAD prevalences were not normally
(Table 1) entailed 14,540 patients who had a median age of 56 distributed
(range 31–79) years, a median male % of 36 (range 0–48) and
covered a wide diverse population (hypertensive, normoten- Methods of BP measurements
sive, elderly, diabetics, HIV patients, and pregnant women). Tables 2 and 3 provide differences between groups separated
The heterogeneity test showed that the proportion variation for measurement methods (simultaneous vs. sequential meas-
due to heterogeneity (I2) ranged from 0.81 to 0.96. BP meas- urements), device (manual vs. automatic) and the number
urement was predominantly performed in a sitting posi- of readings (1 vs. 2 or more readings) used to determine
tion but four studies reported supine position.16–19 BP was IAD prevalences (Table 2), average and absolute IAD val-
measured with a standard mercury device,14,20–25 an aneroid ues (Table 3). There were 12 studies8,9,13,14,17,18,24,26,28,29,31,32
device,16 an automated oscillometric device8,9,13,14,17–19,26–32 (n = 8,853) in which simultaneous measurements and 12
and an automated auscultatory device.33 Of the automated studies13,14,16,19–23,25,27,30,33 (n = 6,001) in which sequential
devices used in the included studies in all but 4 (refs. 17– measurements were performed, which means that in two stud-
19,27) a validated device was used as listed on the websites of ies13,14 both methods were assessed. In 14 studies an auto-
www.dableducational.org and/or www.bhsoc.org indicating to mated oscillometric device8,9,13,14,17–19,26–32 (n = 12,644) was
have passed one of the recognized protocols for validation of used for IAD determination, in seven studies9,17,20,27,32,33
BP devices.34–36 Most studies reported to have measured arm- (n = 2,063) a manual sphygmomanometer (aneroid or mer-
circumference for selecting the appropriate sized cuff and to cury) and in one study an automated auscultatory device.33
have waited for 5 min before starting BP measurement. To There were seven studies in which only one measurement

1202 november 2011 | VOLUME 24 NUMBER 11 | AMERICAN JOURNAL OF HYPERTENSION


Inter-arm blood pressure differences original contributions

Table 1 | Studies to inter-arm differences of blood pressure


Age Men Readings n (%) ≥10 mm Hg n (%) ≥20 mm Hg IAD (R–L) (mm Hg)
Reference Population n (years) (%) (n) Device Method Systolic Diastolic Systolic Diastolic Systolic Diastolic
16 NT (29%) 447 48 ± 15 59 1a Aneroid Seqa 119 (26.6)a 67 (15.0)a 32 (7.2)a −2.3 ± 12a −0.8 ± 7a
HT (71%) 3b Sim, (5.3)b,c (4.0)b,c (0.1)b,c 0.5 ± 5b −1.3 ± 5b
2 observersb
20 NT (32%) 174 (24–70) 56 3 MS Seq 84 (48) 48 (28) 23 (13) 11 (6)
HT (68%)
20 NT (100%) 23 (18–89) 30 1 MS Seq 14 (61) 10 (43) 6 (26) 2 (9)
33 Elderly 174 76 ± 9 16 6 AAD Seq 2 (1.4) 0 2 (1) 0 0.93 ± 4.0 0.72 ± 2.8

Downloaded from https://academic.oup.com/ajh/article/24/11/1201/2730264 by guest on 01 November 2022


26 HP 40 74 35 8 AOD Sim 4 (10) 4.2 (3.1–3.5) (A) 3.6 (2.8–4.4) (A)
26 HP 40 31 20 8 AOD Sim 1 (3) 3.3 (2.6–4.1) (A) 2.7 (2.0–3.3) (A)
17 HP (–CVD) 364 49 57 4 AOD Sim 19 (5.2) 23 (6.3) 0 (0) 1 (0.3) 1.1 ± 4.9 0 ± 5.1
(0.6–1.6) (–0.5 to 0.5)
21 GP 237 >16 years — 1 MS Seq 95 (40) 55 (23) 4.8 (3–6.5) 3.7 (2.4–5)
(routine)
22 HP 280 60 — 1 MS Seq 65 (23.2) 38 (13.6) — —
9 HP + staff 400 56 ± 20 50 2 AOD Sim 80 (20) 45 (11) 14 (3.5) 15 (3.8) 1.81 ± 8.6 –0.23 ± 8.3
6.32 ± 6.12 (A) 5.06 ± 6.57 (A)
23 HT 100 55 ± 10 33 1 MS Seq (blinded, 18 (18)
2 observers)
18 Gen pop 1,090 62 ± 11 36 1 AOD Sim 99 (9.1) –0.6 ± 6.6 1.1 ± 4.7
1 device 4.9 ± 4.4 (A) 3.7 ± 3.0 (A)
24 Elderly 528 79 ±10 32 2 MS Sim 70 (14) 28 (5) — —
2 observers
27 Volunteers 854 56 49 3 AOD Seq 79 (9.2) 14 (1.6) 9 (1.1) 1 (0.1) 4.61 ± 4.10 (A) 2.96 ± 2.51(A)
27 HT Siblings 2,395 56 39 3 AOD Seq (14.2) (2.8) (1.8) (0.1) 5.35 ± 4.98 (A) 3.09 ± 2.73 (A)
28 HP + staff 384 54 ± 18 49 4 AOD Sim 13 (3.4) 4 (1.4) 1.2 ± 5.0 0.4 ±4.2
29 HT 94 70 ± 10 40 4 AOD Sim 18 (19) 7 (7) 3.1± 7.0 1.6 ± 5.6
5.9 ± 4.9 (A) 4.6 ± 3.6 (A)
25 HT 247 69 45 2 MS Seq 57 (23) 15 (6) 8 (3) 1.6 (0.4–2.9) –1.4 (0.8–2.0)
14 HT ® 145 58 ± 16 48 3b aAOD aSim (1&3) 1.5 ± 3.3 (1&3) 0.9 ± 2.5 (1&3)
(2 devices)
bMS bSeq (2) 3.1 ± 6.4 (2) 1.4 ± 3.8 (2)
13 DM ® 169 67 ± 10 50 1A AOD SeqA 55 (33) — —
1B 2 devices SimB 27 (16)
2C Sim C 16 (9)
30 HP ± CKD® 421 63 ± 13 95 3 AOD Seq 158 (38) 60 (15) 40 (10) 5.1
Seq 83 (20) 29 (7) 12 (4) 2.9
(+ 1 week)
19 HIV ♀ (+) 335 — 0 2 AOD Seq 87 (26) 50 (15) 20 (6) 6 ± 5 (A) 4 ± 3 (A)
controls (–) 238+
97–
31 Pregnant 5,435 32 0 2–6a AOD Sim 450 (8.3) 124 (2.3) 0.8 –0.6
(HT + NT) (–10.3 to 11.9) (–8.3 to 7.0)
32 HP 63 61 ± 13 54 3 AOD Sim 0 0 0 0 0.04 ± 5.1 0.4 ± 3.2
1 device
8 DM 101 66 ± 13 59 4 AOD Sim 10 (10) 3 (3) — —
Data provide average ± s.d. of all readings.
!, average of both populations together; (A), absolute difference; (R–L), blood pressure values of the right arm—blood pressure values of the left arm; ®, reproducibility study; AAD,
automatic ausculatory device; AOD, automatic oscillometric device; CAD, coronary artery disease; CKD, chronic kidney disease; CVD, cardiovascular disease; D, diastolic blood pressure;
dia, diastolic; DM, diabetes mellitus; Gen pop, general population; GP, general practitioner; HP indicates hospital referred patients; HT, hypertensives; IAD, inter-arm difference; MS, manual
sphygmomanometer; n(%) Δ, number and percentage of patients with an inter-arm difference greater than or similar to 10 or 20 mm Hg; n, number; NT, normotensives; PVD, peripheral
vascular disease; Ran, randomized; S, systolic blood pressure; seq, sequential measurements; Sim, simultaneous measurements, Sup, supine position; sys, systolic; X-linked, crosslinked.
aPatients were measured until variations between consecutive readings fell within 10mm Hg in systolic pressure and 6 mm Hg in diastolic pressure. bOnly the first method is used for the

present analysis, this study compared different methods and investigated reproducibility; c, measurements not patients.

AMERICAN JOURNAL OF HYPERTENSION | VOLUME 24 NUMBER 11 | november 2011 1203


original contributions Inter-arm blood pressure differences

Table 2 | Differences in prevalences inter-arm difference (IAD) between groups divided by measurements, device and number of
readings
Overall mean Measurements Device Readings
Prevalences (%) (95% CI) RR: Seq/Sim P RR: Man/Auto P RR: 1/≥2 P
Systolic
IAD ≥10 mm Hg (%) 13.6 (10.1–18.4) 2.2 (1.4–3.6) <0.01 2.1 (1.1–3.9) <0.05 2.0 (1.1–3.8) <0.05
IAD ≥20 mm Hg (%) 4.2 (2.4–7.5) 4.8 (1.1–21.9) <0.05 4.4 (1.8–10.8) <0.01 4.3 (1.6–11.4) <0.01
Diastolic (mm Hg)
IAD ≥10 mm Hg (%) 6.5 (3.8–11.2) 2.5 (1.0–6.3) <0.05 3.7 (1.6–8.6) <0.01 4.4 (1.7–11.4) <0.01
IAD ≥20 mm Hg (%) 0.7 (0.2–2.4) 0.4 (0–18.4) 0.64 17.4 (0.2–1,445.1) 0.21 — —

Downloaded from https://academic.oup.com/ajh/article/24/11/1201/2730264 by guest on 01 November 2022


All values are provided as RR and 95% confidence intervals (CIs).
1/≥2 readings, 1 reading divided by 2 or more readings; Man/Auto, manual divided by automatic; Seq/Sim, relative risk (RR) values were calculated by dividing the IAD prevalences
obtained from sequential by those from simultaneous measurements.

Table 3 | Differences in average and absolute inter-arm difference (IAD) values between groups divided by measurements, device and
number of readings
Overall Measurements Device Readings
average Seq–Sim P Man–Auto P 1–≥2 P
Systolic
IAD absolute 5.4 ± 1.7 0.8 ± 3.4 0.81 4.0 ± 6.4 0.55 1.2 ± 4.1 0.78
(mm Hg)
IAD (mm Hg) 0.57 ± 1.9 0.5 ± 3.2 0.87 0.4 ± 6.0 0.94 0.6 ± 4.9 0.91
Diastolic (mm Hg)
IAD absolute 3.6 ± 1.2 –0.6 ± 2.5 0.81 1.2 ± 6.7 0.86 0.4 ± 3.0 0.91
(mm Hg)
IAD (mm Hg) 0.5 ± 1.3 0.6 ± 2.3 0.79 −0.7 ± 3.8 0.86 0.1 ± 3.3 0.98
All values are provided as mean ± s.e.m.
1–≥2 readings, differences between studies that performed 1 and 2 or more readings to determine IAD. Man–Auto, differences between a manual and automatic device; Seq–Sim
indicates differences between average and absolute IAD values obtained with sequential and simultaneous measurements.

was performed13,16,18,20–23 (n = 2,515) to determine IAD and 2.0 to 4.4). For average and absolute IAD values there were no
16 studies8,9,14,17,19,20,24–33 (n = 12,339) provided two or more s­ ignificant differences (Table 3).
measurements. One study was divided in subpopulations There was a significant (P < 0.01) and negative correlation
which were measured one and three times.20 of the systolic IAD ≥10 mm Hg and the number of readings
As shown in Table 2 the method of IAD performance had a (Figure 3).
significant influence on systolic (IAD ≥10 and 20 mm Hg) and Overall, the method (simultaneous or sequential), the
diastolic (IAD ≥10) prevalence: sequential measurements led device, and the number of measurements had no significant
to a significant higher prevalence than simultaneous meas- effect on mean and mean absolute IADs.
urements (RR ranges from 2.2 to 4.8), the use of a manual BP
device led to a significant higher prevalence than when using Patient characteristics
an automated device (RR ranges from 2.1 to 4.4). There was no correlation with systolic mean absolute and
However, there was a strong relationship between devices mean IAD and BP values, age, and gender or for diastolic IADs
and measurement methods: all, except one, ­simultaneous and IAD prevalences (data not shown).
measurements were performed with automated devices
(Figure 2). For this reason, we performed a post hoc multi- Discussion
variate analysis using both method (simultaneous or sub- The findings in the present paper showed that the average
sequent) and devices (automated or manual) as covariates. mean absolute IAD is 5.4 and 3.6 mm Hg for systolic and
Results for systolic IAD prevalence showed that the method diastolic BP, respectively and 14% of all subjects had a systo-
(RR 1.9 (95% CI: 1.2–3.2), P < 0.05) but not the device (RR 1.5 lic IAD of 10 mm Hg or more. The number of subjects with
(95% CI: 0.9–2.7), P = 0.13) had a significant influence on the a systolic and diastolic IAD ≥10 mm Hg and systolic IAD
prevalence. ≥20 mm Hg was significantly lower when BP was measured
With regard to the number of readings it was shown that with an automatic instead of a manual device, when meas-
one reading resulted in a significant higher prevalence than urements were performed simultaneously instead of sequen-
when two or more readings were performed (RR ranges from tially and with two or more measurements than with only one

1204 november 2011 | VOLUME 24 NUMBER 11 | AMERICAN JOURNAL OF HYPERTENSION


Inter-arm blood pressure differences original contributions

Reference Device Method ES (95% Cl)

Stergiou, 2008 32 Auto Simultaneous 0.01 (−0.01, 0.03)

Orme, 1999 17 Auto Simultaneous 0.05 (0.03, 0.08)

Kimura, 2004 18 Auto Simultaneous 0.09 (0.07, 0.11)

Fotherby, 1993 26 Auto Simultaneous 0.10 (0.01, 0.19)


9
Lane, 2002 Auto Simultaneous 0.20 (0.16, 0.24)

Clark, 2009 8 Auto Simultaneous 0.10 (0.04, 0.16)

Kleefstra, 2007 13 Auto Simultaneous 0.16 (0.10, 0.22)

Clark, 2007 29

Downloaded from https://academic.oup.com/ajh/article/24/11/1201/2730264 by guest on 01 November 2022


Auto Simultaneous 0.19 (0.11, 0.27)

Fotherby, 1993 26 Auto Simultaneous 0.03 (−0.02, 0.07)

Poon, 2008 31 Auto Simultaneous 0.08 (0.08, 0.09)

Karagiannis, 2005 28 Auto Simultaneous 0.03 (0.02, 0.05)

Mendelson, 2004 24 Manual Simultaneous 0.13 (0.10, 0.16)

Hashimoto, 1984 33 Auto Subsequent 0.01 (−0.00, 0.03)

Lazar, 2008 19 Auto Subsequent 0.26 (0.21, 0.31)


13
Kleefstra, 2007 Auto Subsequent 0.33 (0.25, 0.40)

Arnett, 2005 27 Auto Subsequent 0.09 (0.07, 0.11)

Arnett, 2005 27 Auto Subsequent 0.14 (0.13, 0.16)

Agarwal, 2008 30 Auto Subsequent 0.38 (0.33, 0.42)

Kristensen, 1982 20 Manual Subsequent 0.50 (0.43, 0.57)

Pesola, 2002 23 Manual Subsequent 0.18 (0.10, 0.26)

Harrison, 1960 16 Manual Subsequent 0.27 (0.23, 0.31)


29
Clark, 2007 Manual Subsequent 0.23 (0.18, 0.28)

0
Prevalence systolic IAD > 10 mm Hg

Figure 2 | Forest plot of the prevalences of subjects with a systolic inter-arm difference (IAD) of ≥10 mm Hg. The studies are grouped by device (automated or
manual) and measurement method (simultaneous or subsequent).

70
Systolic IAD ≥10 mm Hg significant differences shown when groups were separated
according to different methods. There were no significant dif-
60
ferences between BP values on the right and left arm.
50
Prevalence (%)

40
Strength and weakness of the paper
30 r 2 = 0.11; P < 0.01 This study should be interpreted within the context of its limi-
20 tations. Studies showed a substantial heterogeneity, there was a
10 wide disparity among several methods and subjects and it was
0 not always clear how the patient population had to be inter-
0 2 4 6 8
preted and what the main reason was (method, population
Readings (n)
device) for finding differences in value and prevalence of IAD.
Figure 3 | The prevalence of subjects with a systolic inter-arm difference (IAD) For this reason it might be questioned whether these studies
of ≥10 mm Hg related to the number of readings, r2 values are calculated with should be pooled for analysis at all. However, the authors are
log transformation. of the opinion that despite the large heterogeneity the pooled
analysis is a useful method as this heterogeneity is mainly
r­ eading. In ­addition, there was a significant trend toward a related to the different measurement procedures used (device,
lower prevalence of systolic IAD ≥10 with each extra measure- method, readings) and thus reflects clinical practice. Seen
ment. Because in all except one study simultaneous measure- within the limitations, the analysis therefore provides use-
ments were performed with automated devices, it appeared ful clinical information. In some papers standard deviations
that the method but not the device had a significant influ- or variances were not provided so that the highest standard
ence on outcome. For mean and absolute IADs there were no deviations from all studies was imputed. Studies that reported

AMERICAN JOURNAL OF HYPERTENSION | VOLUME 24 NUMBER 11 | november 2011 1205


original contributions Inter-arm blood pressure differences

more pairs of readings are more likely to have used automated coat effect which causes the first measurement to be higher
rather than manual methods. The same studies are likely to than the next.27,42,43 In addition, BP is a variable hemody-
have contributed to each relevant section of the results. These namic phenomenon that constantly fluctuates over time, mak-
items cause a limit to the conclusions that can be drawn inde- ing sequential measurements difficult to compare.
pendently about each facet of the measurement technique. On In order to exclude the possibility of erroneous IAD
the other hand the use of automated devices seems strongly determination due to sequential performance some inves-
related to simultaneous measurements as automated devices tigators have performed simultaneous double-arm meas-
facilitate the simultaneous measurements, whereas simulta- urements with two observers44,45 or with two automatic
neous measurements with manual devices are very difficult monitors.8,9,13,14,17,26,28,29,31,46 Although the latter seems a
to accomplish in clinical practice. In addition, two observers good idea a minor bias can still be introduced due to a delay
are less likely to perform measurements simultaneously than between readings and inter-device differences even when
when one automated device with two cuffs or two automated devices are from the same brand.9 Changing the devices

Downloaded from https://academic.oup.com/ajh/article/24/11/1201/2730264 by guest on 01 November 2022


devices are used. halfway through the performance might reduce the influ-
Papers published before the year 1960 were excluded from ence of the inter-device difference8,9,17,26,28,29 but compli-
the present review. Although this line is arbitrary it may be cates the measurement procedure. For this reason, automatic
justified by the following arguments. First, although the mer- devices with two identical cuffs may give the opportunity
cury sphygmomanometer could be of sufficient quality in the of performing the most accurate simultaneous double-arm
early years stethoscopes were of lesser quality and a poor qual- measurements. Momentarily, automatic oscillometric double-
ity stethoscope has a major influence on accuracy.37 Second, cuff devices have been developed (e.g., Microlife WatchBP
most of the studies from before 1960 lack a detailed method Office32) for diagnosis in regular healthcare and thus elimi-
description which makes it complicated or impossible to nate the practical concern that a device with the possibility to
know how to interpret the values. Third, at that time the 4th perform simultaneous measurements could not be widely dis-
Korotkoff sound was commonly used to determine the diasto- tributed among clinics or general practitioner practices.47
lic BP value. Fourth, as can be seen from the review of Clark et
al. many studies from before 1960 show very high prevalences Device
of patients with relevant IADs; these values seem not compa- Data from the present paper showed a significant lower preva-
rable with the values from younger papers and imply that this lence of relevant IADs (≥10 mm Hg) between studies that used
is the result of methods or patient selection.38 Several large- an automatic instead of a manual device. This might be due
scale studies have been published that were excluded for the to observer bias related to the latter and digit preference that
present paper because of the methods that were used.39–41 could increase the BP difference between the arms.1 In addi-
Although these papers contain interesting information, for tion, most studies that used a manual device also measured
the present review we chose only to compare those studies in both arms sequentially instead of simultaneously, which lead to
which double-arm measurements were performed with regu- a higher number of relevant IADs. However, it is unlikely that
lar BP monitors. Although, Doppler devices are commonly this is the only cause of the significant higher IAD ≥10 mm Hg
used to measure systolic upper-arm pressure when assessing prevalence obtained with a manual device because sequential
ankle brachial index it is not recognized as routine tool to per- measurements could also lead to a smaller difference in IAD as
form upper-arm BP measurements. the second measurement is taken with knowledge of the first.
There were significant differences for prevalences (IAD ≥10
or 20 mm Hg) but not for mean and absolute mean IAD val- Multiple measurements
ues. This, is most likely, related to the cutoff values; in other The present paper showed a significant trend toward fewer sub-
words: although the value is not significantly higher there is a jects with a relevant IAD when more readings were performed.
greater chance that it is above the threshold value. Finally, in This finding is underscored by the study of Arnett et al. who
the present study we could not find a relationship with aver- showed that increasing the number of BP measurements
age IAD values and IAD prevalences (IAD ≥10 or 20 mm Hg) from one to three resulted in a decrease in the mean absolute
and patient characteristics such as age, gender, and BP values. systolic IAD (from 6.3 to 4.6 mm Hg) and a 30% reduction in
However, this does not mean that there is not such a relation- variance in the difference between the right arm and the left
ship but the possibilities for analyzing were limited by the lack arm.27 From other studies it is known that the reproducibility
of access to full data. increases with more BP readings.48 From studies that investi-
gated the reproducibility of IAD13,14,30 only one study showed
Sequential and simultaneous measurements that IADs are reproducible and carry prognostic formation.30
In the present analysis, simultaneous measurements resulted In that study three measurements were assessed to determine
in much fewer patients with a systolic IAD ≥10 mm Hg as the IAD value which might be the reason for the high repro-
compared to sequential measurements. This finding was in ducibility. Eguchi et al. also used three measurements but their
agreement with the finding of others who compared both opposite conclusion might be influenced by the low IAD of
methods.13,14 A reason for the higher IAD with sequential ≥5 mm Hg that was chosen as endpoint, which overall is not
measurements could be related to a cuff-response or white- considered a relevant IAD value.14

1206 november 2011 | VOLUME 24 NUMBER 11 | AMERICAN JOURNAL OF HYPERTENSION


Inter-arm blood pressure differences original contributions

Clinical significance of inter-arm measurements Disclosure: W.J.V. is an official employee of Microlife Corporation (Taipei,
The present data show that 14% of the subjects had an IAD Taiwan). A.G.H.K. and T.T. declared no conflict of interest.
of 10 mm Hg or more for systolic BP. This seems to justify the 1. Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, Jones DW, Kurtz T,
recommendation in the guidelines and from many investiga- Sheps SG, Roccella EJ. Recommendations for blood pressure measurement
in humans and experimental animals: part 1: blood pressure measurement in
tors9,17,18,21,24,26,28,30 to perform BP measurements on both humans: a statement for professionals from the Subcommittee of Professional
arms at each first visit. The importance of determining the and Public Education of the American Heart Association Council on High Blood
IAD is confirmed by the finding that the mean absolute IAD Pressure Research. Circulation 2005; 111:697–716.
2. 1999 World Health Organization-International Society of Hypertension
was 5 and 4 mm Hg for systolic and diastolic BP, respectively. Guidelines for the Management of Hypertension. Guidelines Subcommittee.
Although an IAD of 5 mm Hg may not be clinically relevant J Hypertens 1999; 17:151–183.
for an individual patient on a population basis this difference 3. Mancia G, De Backer G, Dominiczak A, Cifkova R, Fagard R, Germano G, Grassi G,
Heagerty AM, Kjeldsen SE, Laurent S, Narkiewicz K, Ruilope L, Rynkiewicz A,
has been associated with e.g., a 34% difference in stroke.49 In Schmieder RE, Boudier HA, Zanchetti A, Vahanian A, Camm J, De Caterina R,
addition, results from a health survey led to the estimation

Downloaded from https://academic.oup.com/ajh/article/24/11/1201/2730264 by guest on 01 November 2022


Dean V, Dickstein K, Filippatos G, Funck-Brentano C, Hellemans I, Kristensen SD,
that underestimation of systolic BP by >10 mm Hg could fail to McGregor K, Sechtem U, Silber S, Tendera M, Widimsky P, Zamorano JL,
Erdine S, Kiowski W, Agabiti-Rosei E, Ambrosioni E, Lindholm LH, Viigimaa M,
prevent >4,000 fatal and nonfatal coronary events and >6,000 Adamopoulos S, Agabiti-Rosei E, Ambrosioni E, Bertomeu V, Clement D,
fatal and nonfatal strokes each year in England.50 Erdine S, Farsang C, Gaita D, Lip G, Mallion JM, Manolis AJ, Nilsson PM, O’Brien E,
Studies that provided the relationship between IAD and car- Ponikowski P, Redon J, Ruschitzka F, Tamargo J, van Zwieten P, Waeber B,
Williams B; Management of Arterial Hypertension of the European Society
diovascular risk in the long-term22,25,30 showed that a systolic of Hypertension; European Society of Cardiology. 2007 Guidelines for the
and diastolic IAD ≥10 mm Hg and a systolic IAD ≥20 mm Hg Management of Arterial Hypertension: The Task Force for the Management of
led to significant shorter event free survival time as com- Arterial Hypertension of the European Society of Hypertension (ESH) and of the
European Society of Cardiology (ESC). J Hypertens 2007; 25:1105–1187.
pared to those with lower systolic IAD values.22,25 In addition, 4. Williams B, Poulter NR, Brown MJ, Davis M, McInnes GT, Potter JF, Sever PS,
Agarwal et al. showed among patients with renal disease that McG Thom S; British Hypertension Society. Guidelines for management of
every 10 mm Hg difference in systolic BP between the arms hypertension: report of the fourth working party of the British Hypertension
Society, 2004-BHS IV. J Hum Hypertens 2004; 18:139–185.
conferred a mortality hazard of 1.24 (95% CI: 1.01–1.52).30 5. Seedat YK, Croasdale MA, Milne FJ, Opie LH, Pinkney-Atkinson VJ, Rayner BL,
Due to reasons as provided above it is alarming that IAD is Veriava Y; Guideline Committee, Southern African Hypertension Society;
determined by only 23% of the physicians51 and that very few Directorate: Chronic Diseases, Disabilities and Geriatrics, National Department of
Health. South African hypertension guideline 2006. S Afr Med J 2006; 96:337–362.
general practitioners and hospital physicians follow “even the 6. Chobanian AV, Bakris GL, Black HR, Cushman WC, Green LA, Izzo JL Jr, Jones DW,
most critical aspects of BP measurement guidelines.”52 Materson BJ, Oparil S, Wright JT Jr, Roccella EJ; National Heart, Lung, and Blood
Institute Joint National Committee on Prevention, Detection, Evaluation, and
Treatment of High Blood Pressure; National High Blood Pressure Education
Patient characteristics
Program Coordinating Committee. The Seventh Report of the Joint National
The present paper did not see any relationship with IAD Committee on Prevention, Detection, Evaluation, and Treatment of High Blood
value and/or prevalence and patient characteristics such as Pressure: the JNC 7 report. JAMA 2003; 289:2560–2572.
7. Perloff D, Grim C, Flack J, Frohlich ED, Hill M, McDonald M, Morgenstern BZ.
age, gender, and BP. An overview from other studies that Human blood pressure determination by sphygmomanometry. Circulation 1993;
investigated the relationship of IAD with certain patient 88:2460–2470.
characteristics is provided in Supplementary Appendix S2 8. Clark CE, Greaves CJ, Evans PH, Dickens A, Campbell JL. Inter-arm blood pressure
difference in type 2 diabetes: a barrier to effective management? Br J Gen Pract
online. Overall, there is a wide heterogeneity among studies 2009; 59:428–432.
with regard to most patient characteristics except for obesity 9. Lane D, Beevers M, Barnes N, Bourne J, John A, Malins S, Beevers DG. Inter-arm
and ankle brachial index (seems related to IAD value) and differences in blood pressure: when are they clinically significant? J Hypertens
2002; 20:1089–1095.
age, gender, handedness, and diabetes (seems not related to 10. Moher D, Liberati A, Tetzlaff J, Altman DG; PRISMA Group. Reprint–preferred
IAD value). reporting items for systematic reviews and meta-analyses: the PRISMA statement.
Results of the present analysis showed that the mean abso- Phys Ther 2009; 89:873–880.
11. Rothman J. Modern Epidemiology. Epidemiology Resources: Chestnut Hill, MA,
lute IAD value is 5 and 4 mm Hg for systolic and diastolic BP 1986, pp. 183–185.
value and 14% of all subjects had a systolic IAD ≥10 mm Hg. 12. Jüni P, Altman DG, Egger M. Systematic reviews in health care: Assessing the
This together with the findings that the IAD value is related to quality of controlled clinical trials. BMJ 2001; 323:42–46.
13. Kleefstra N, Houweling ST, Meyboom-de Jong B, Bilo HJ. [Measuring the blood
cardiovascular risk underscores the importance of performing pressure in both arms is of little use; longitudinal study into blood pressure
double-arm measurements at each first visit. differences between both arms and its reproducibility in patients with diabetes
The method of how IAD screening is performed seems to mellitus type 2]. Ned Tijdschr Geneeskd 2007; 151:1509–1514.
14. Eguchi K, Yacoub M, Jhalani J, Gerin W, Schwartz JE, Pickering TG. Consistency of
have a major influence (commonly overestimation) on the blood pressure differences between the left and right arms. Arch Intern Med 2007;
value of IAD and should therefore be clearly described in the 167:388–393.
guidelines. Results of the present study indicate that inter-arm 15. Higgins JP, Thompson SG. Quantifying heterogeneity in a meta-analysis. Stat Med
2002; 21:1539–1558.
BP measurements should be determined by simultaneous BP 16. Harrison EG Jr, Roth GM, Hines EA Jr. Bilateral indirect and direct arterial pressures.
measurements with one automatic device with two cuffs or Circulation 1960; 22:419–436.
two automated devices and a diagnosis of an IAD of 10 mm Hg 17. Orme S, Ralph SG, Birchall A, Lawson-Matthew P, McLean K, Channer KS. The normal
range for inter-arm differences in blood pressure. Age Ageing 1999; 28:537–542.
or more should be based on multiple readings. 18. Kimura A, Hashimoto J, Watabe D, Takahashi H, Ohkubo T, Kikuya M, Imai Y. Patient
characteristics and factors associated with inter-arm difference of blood pressure
Supplementary material is linked to the online version of the paper at http:// measurements in a general population in Ohasama, Japan. J Hypertens 2004;
www.nature.com/ajh 22:2277–2283.

AMERICAN JOURNAL OF HYPERTENSION | VOLUME 24 NUMBER 11 | november 2011 1207


original contributions Inter-arm blood pressure differences

19. Lazar J, Holman S, Minkoff HL, Dehovitz JA, Sharma A. Interarm blood pressure 37. O’Brien E, Petrie J, Littler W, de Swiet M, Padfield PL, O’Malley K, Jamieson M,
differences in the women’s interagency HIV study. AIDS Res Hum Retroviruses 2008; Altman D, Bland M, Atkins N. The British Hypertension Society protocol for the
24:695–700. evaluation of automated and semi-automated blood pressure measuring
20. Kristensen BO, Kornerup HJ. Which arm to measure the blood pressure? Acta Med devices with special reference to ambulatory systems. J Hypertens 1990; 8:
Scand Suppl 1982; 670:69–73. 607–619.
21. Cassidy P, Jones K. A study of inter-arm blood pressure differences in primary care. 38. Clark CE, Campbell JL, Evans PH, Millward A. Prevalence and clinical implications
J Hum Hypertens 2001; 15:519–522. of the inter-arm blood pressure difference: A systematic review. J Hum Hypertens
22. Clark CE, Powell RJ. The differential blood pressure sign in general practice: 2006; 20:923–931.
prevalence and prognostic value. Fam Pract 2002; 19:439–441. 39. Shadman R, Criqui MH, Bundens WP, Fronek A, Denenberg JO, Gamst AC,
23. Pesola GR, Pesola HR, Lin M, Nelson MJ, Westfal RE. The normal difference in McDermott MM. Subclavian artery stenosis: prevalence, risk factors, and
bilateral indirect blood pressure recordings in hypertensive individuals. Acad association with cardiovascular diseases. J Am Coll Cardiol 2004; 44:618–623.
Emerg Med 2002; 9:342–345. 40. Aboyans V, Kamineni A, Allison MA, McDermott MM, Crouse JR, Ni H, Szklo M,
24. Mendelson G, Nassimiha D, Aronow WS. Simultaneous measurements of blood Criqui MH. The epidemiology of subclavian stenosis and its association with
pressures in right and left brachial arteries. Cardiol Rev 2004; 12:276–278. markers of subclinical atherosclerosis: the Multi-Ethnic Study of Atherosclerosis
25. Clark CE, Campbell JL, Powell RJ. The interarm blood pressure difference as (MESA). Atherosclerosis 2010; 211:266–270.

Downloaded from https://academic.oup.com/ajh/article/24/11/1201/2730264 by guest on 01 November 2022


predictor of cardiovascular events in patients with hypertension in primary care: 41. Aboyans V, Criqui MH, McDermott MM, Allison MA, Denenberg JO, Shadman R,
cohort study. J Hum Hypertens 2007; 21:633–638. Fronek A. The vital prognosis of subclavian stenosis. J Am Coll Cardiol 2007;
26. Fotherby MD, Panayiotou B, Potter JF. Age-related differences in simultaneous 49:1540–1545.
interarm blood pressure measurements. Postgrad Med J 1993; 69:194–196. 42. Verberk WJ, Kroon AA, Lenders JW, Kessels AG, van Montfrans GA, Smit AJ,
27. Arnett DK, Tang W, Province MA, Oberman A, Ellison RC, Morgan D, Eckfeldt JH, van der Kuy PH, Nelemans PJ, Rennenberg RJ, Grobbee DE, Beltman FW,
Hunt SC. Interarm differences in seated systolic and diastolic blood pressure: the Joore MA, Brunenberg DE, Dirksen C, Thien T, de Leeuw PW; Home Versus Office
Hypertension Genetic Epidemiology Network study. J Hypertens 2005; 23: Measurement, Reduction of Unnecessary Treatment Study Investigators. Self-
1141–1147. measurement of blood pressure at home reduces the need for antihypertensive
28. Karagiannis A, Tziomalos K, Krikis N, Sfikas G, Dona K, Zamboulis C. The unilateral drugs: a randomized, controlled trial. Hypertension 2007; 50:1019–1025.
measurement of blood pressure may mask the diagnosis or delay the effective 43. Verberk WJ, Kroon AA, Thien T, Lenders JW, van Montfrans GA, Smit AJ,
treatment of hypertension. Angiology 2005; 56:565–569. de Leeuw PW. Prevalence of the white-coat effect at multiple visits before and
29. Clark CE, Campbell JL, Powell RJ, Thompson JF. The inter-arm blood pressure during treatment. J Hypertens 2006; 24:2357–2363.
difference and peripheral vascular disease: cross-sectional study. Fam Pract 2007; 44. Pesola GR, Pesola HR, Nelson MJ, Westfal RE. The normal difference in bilateral
24:420–426. indirect blood pressure recordings in normotensive individuals. Am J Emerg Med
30. Agarwal R, Bunaye Z, Bekele DM. Prognostic significance of between-arm blood 2001; 19:43–45.
pressure differences. Hypertension 2008; 51:657–662. 45. Gould BA, Hornung RS, Kieso HA, Altman DG, Raftery EB. Is the blood pressure the
31. Poon LC, Kametas N, Strobl I, Pachoumi C, Nicolaides KH. Inter-arm blood same in both arms? Clin Cardiol 1985; 8:423–426.
pressure differences in pregnant women. BJOG 2008; 115:1122–1130. 46. Singer AJ, Hollander JE. Blood pressure. Assessment of interarm differences. Arch
32. Stergiou GS, Lin CW, Lin CM, Chang SL, Protogerou AD, Tzamouranis D, Intern Med 1996; 156:2005–2008.
Nasothimiou E, Tan TM. Automated device that complies with current guidelines 47. Materson BJ. Inter-arm blood pressure differences. J Hypertens 2004; 22:
for office blood pressure measurement: design and pilot application study of the 2267–2268.
Microlife WatchBP Office device. Blood Press Monit 2008; 13:231–235. 48. Verberk WJ, Kroon AA, Kessels AG, de Leeuw PW. Home blood pressure
33. Hashimoto F, Hunt WC, Hardy L. Differences between right and left arm blood measurement: a systematic review. J Am Coll Cardiol 2005; 46:743–751.
pressures in the elderly. West J Med 1984; 141:189–192. 49. MacMahon S, Peto R, Cutler J, Collins R, Sorlie P, Neaton J, Abbott R, Godwin
34. O’Brien E, Petrie J, Littler W, de Swiet M, Padfield PL, Altman DG, Bland M, Coats A, J, Dyer A, Stamler J. Blood pressure, stroke, and coronary heart disease.
Atkins N. An outline of the revised British Hypertension Society protocol for the Part 1, Prolonged differences in blood pressure: prospective observational
evaluation of blood pressure measuring devices. J Hypertens 1993; 11:677–679. studies corrected for the regression dilution bias. Lancet 1990; 335:
35. Association for the Advancement of Medical Instrumentation. American National 765–774.
Standard for Electronic or Automated Sphygmomanometers. ANSI/AAMI SP-10-1992. 50. Primatesta P, Brookes M, Poulter NR. Improved hypertension management and
AAMI: Arlington, VA, 1993. control: results from the health survey for England 1998. Hypertension 2001;
36. O’Brien E, Pickering T, Asmar R, Myers M, Parati G, Staessen J, Mengden T, Imai Y, 38:827–832.
Waeber B, Palatini P, Gerin W; Working Group on Blood Pressure Monitoring of the 51. McKay DW, Campbell NR, Parab LS, Chockalingam A, Fodor JG. Clinical
European Society of Hypertension. Working Group on Blood Pressure Monitoring assessment of blood pressure. J Hum Hypertens 1990; 4:639–645.
of the European Society of Hypertension International Protocol for validation of 52. Cushman WC. A century of indirect blood pressure measurement. Back to basics.
blood pressure measuring devices in adults. Blood Press Monit 2002; 7:3–17. Arch Intern Med 1996; 156:1922–1923.

1208 november 2011 | VOLUME 24 NUMBER 11 | AMERICAN JOURNAL OF HYPERTENSION

You might also like