Professional Documents
Culture Documents
brief - instruction - contrAA300 - 600 - 700 - en - копія
brief - instruction - contrAA300 - 600 - 700 - en - копія
®
contrAA 300/600/700
Short instruction contrAA® 300/600/700 | update 09.08.2011/ Po 2/ 83
Analytik Jena AG | Konrad-Zuse-Straße 1 | 07745 Jena/Germany | info@analytik-jena.com | www.analytik-jena.com
Index
Index .................................................................................................................................... 3
1 General information ...................................................................................................... 5
2 Flame technique ........................................................................................................... 5
2.1 Start-up of the contrAA 700 for flame-AAS .......................................................... 5
2.2 Creating a simple method ..................................................................................... 7
2.2.1 The cookbook................................................................................................. 7
2.2.2 Processing the method .................................................................................. 9
2.2.2.1 The Lines tab ............................................................................................ 10
2.2.2.2 The Flame tab .......................................................................................... 11
2.2.2.3 The Sample Transport tab ........................................................................ 12
2.2.2.4 Performing automatic flame optimization .................................................. 13
2.2.2.5 The Evaluation tab .................................................................................... 15
2.2.2.6 The Calibration tab ................................................................................... 18
2.2.2.7 Definition of stock and calibration standards ............................................ 19
2.2.2.8 The Statistics tab ...................................................................................... 20
2.2.2.9 The Output tab.......................................................................................... 21
2.3 Creating a measurement sequence .................................................................... 22
2.4 Performing a flame measurement ....................................................................... 24
2.4.1 Results screen ............................................................................................. 26
2.4.2 Verification of the measurement data ........................................................... 27
2.4.2.1 Relative standard deviation ...................................................................... 27
2.4.2.2 Spectrum .................................................................................................. 27
2.4.2.3 Calibration function ................................................................................... 28
2.4.2.4 Recalculation of datasets ......................................................................... 29
2.4.3 Print-out of the measurement protocol ......................................................... 30
2.4.4 Definition of a reslope measurement............................................................ 31
2.4.5 Integration of quality control samples ........................................................... 32
2.4.6 Creating control charts ................................................................................. 33
2.4.7 Performing a standard addition/addition calibration ..................................... 34
2.5 Special applications in Fl-AAS ............................................................................ 36
2.5.1 Optimization of baffle ball (impact bead) and nebulizer ................................ 36
2.5.1.1 Optimization of the nebulizer for the spray chamber from PPS (shipment
since 2010) ............................................................................................................. 38
2.5.1.2 Optimization of impact bead and nebulizer for the PTFE spray chamber
system (Shipment until 2010) ................................................................................. 39
3 Graphite furnace technique ........................................................................................ 40
3.1 Startup of the spectrometer for the graphite tube technique ............................... 40
3.2 Creating a simple method ................................................................................... 41
3.2.1 The cookbook............................................................................................... 41
3.2.2 The Method screen ...................................................................................... 43
3.2.2.1 The Lines tab ............................................................................................ 44
3.2.2.2 The Furnace tab ....................................................................................... 44
3.2.2.3 The Sample transport tab ......................................................................... 46
3.2.2.4 The Evaluation tab .................................................................................... 46
3.2.2.5 The Calib. tab ........................................................................................... 49
3.2.2.6 Definition of stock and calibration standards ............................................ 51
3.2.2.7 The Statistics tab ...................................................................................... 52
Communication between the PC and measuring device is provided by the Aspect CS
application software.
2 Flame technique
2.1 Start-up of the contrAA 300/700 for flame-AAS
Turn on the PC and printer and wait until the computer has finished booting up
Activate the AAS device power switch. The contrAA 700 is initialized. This process takes
several minutes, during which the following processes are carried out automatically
o Initialization of wavelength drive
o Initialization of the burner height adjustment
o Ignition of the xenon short arc lamp
o Focusing the lamp focal spot
o Initialization of the accessories (autosampler and/or hydride system)
The device is ready for measurement when the lamp is ignited and the focal spot is
aligned
The software contains 'cookbook' values for the most important measurement parameters,
such as flame type, fuel gas composition or burner height, for most of the elements to be
determined using the flame AAS. When the elements are selected, these cookbook values
are automatically loaded and a method is created that is used as the starting point for
further method development.
This cookbook data can be viewed as follows:
After closing the cookbook, click on the Method icon or select New Method in the File pull-
down menu.
If another method has been loaded previously an additional query is displayed. Here
choose basing on standard methods.
The first tab, Lines has different columns with entries that can be edited if necessary.
The Flame tab contains the flame parameters, fuel gas flow rates and burner heights,
which are loaded from the cookbook. These parameters can be manually edited or
determined by an automatic flame optimization (refer to chapter 2.2.2.4).
Flame/Type The displayed flame type corresponds to the last one active. If
necessary, change flame type.
Burner-nebulizer/type Check if the listed burner head is the same as the installed
burner.
Burner angle/Nebulizer The burner angle is generally 0°, but can be changed by
rate manually turning the burner head to reduce the measurement
sensitivity.
Rule of thumb: Rotation by 10°, Sensitivity decreases by a
factor of 2-3. Rotation 90°, Decrease by a factor of 10.
Manual rotation can be entered here for documentation in the
protocol. The nebulizer rate can also be documented in this way.
Both Parameters are not automatically detected by the system.
The accessories installed on the contrAA are defined on the Sample Transport tab.
If an autosampler or the injection module is installed (for installation refer to the manual),
activate the corresponding fields.
The measurement delay time must be adapted depending on the length of the aspiration
capillary. The following delay times are recommended provided if a pre-run measurement is
used in each case under Statistics. If the measurement is performed without a pre-run
measurement, the listed delay times must be increased by approx. 3s.
The Load and Injection time parameters are only active in the injection technique, if peak
area or height is set under Line/Evaluation.
Since the optimum burner height and the gas composition strongly depend on the
respective analyte, the values, preset from the cookbook, should be fine-adjusted once for
each new element. Therefore one or more standard solutions, containing the analyte
elements in a suitable concentration, are needed. The use of commercially available
multielement solutions is recommended.
The suitable standard concentration can be estimated using the information of the
corresponding cookbook screen. Under test concentration there is an analyte concentration
defined that generates an absorption signal of approx. 0.1A. If this value is multiplied by
factors of 2-3, a standard concentration suitable for the optimization is achieved.
The optimization is performed using Automatic Optimization. Depending on your
respective device configuration, this optimization can run as a fully automatic or as a semi-
automatic routine.
In the Control window ignite the flame applying the Ignite Flame command.
If a nitrous oxide/acetylene flame is being used for measurement, after ignition of the
air/acetylene flame, the Air N2O button must be used to manually switch to nitrous oxide.
The flame parameters are now optimized. Now using the Opt. command in the Lines
register, the sequence of the measurement can be optimized according to the flame
parameters.
The method should now be saved to accept the values.
The Evaluation tab contains the line-specific evaluation parameters for determining the
measurement results from the spectra.
The specifications loaded from the cookbook offer good starting conditions for simple
applications, but can be changed if necessary.
Spectral range A maximum of 200 pixels of the CCD array can be selected
for evaluation. Considering the spectral range per pixel this
corresponds to a spectral range of approx. 0.4 nm,
depending on the wavelength range chosen. If 200 pixels are
defined for evaluation, this entire range is displayed in the
spectrum window. As not for every application the full
information over such a large wavelength range is needed
the number of evaluation pixels can be reduced. But in any
case, a minimum of approx. 100 pixels must be selected if a
dynamic BGC fit has been selected (see illustrations).
Spectrum with 200 evaluation pixels Spectrum with 100 evaluation pixels
Note:
An increase in evaluation pixels to 5P leads to a better relative standard deviation and to a
higher measurement sensitivity.
Lowering to 3P mathematically improves the detection limit.
Background
Since the hot flame, and also other atomization units, can emit spectral
correction
emission and absorption spectra that contributes to the analyte signal a
reference measurement with ultrapure water must be performed prior to
measurement of the sample. The spectra recorded in the reference
measurement are then automatically subtracted from the sample
spectrum.
The reference measurement corresponds to the autozero in the
conventional AAS and should be repeated at regular intervals.
Perm. struct.
The AspectCS® software contains a mathematical algorithm, eliminating
permanent absorption structures of the flame, such as O-H molecular
bands etc. Activation of this algorithm can be advantageous for
measurements in the low wavelength range or when the nitrous oxide
flame is applied.
Note:
The columns can be filled and changed by using the autofill function .
Calibration Select the desired type of calibration. In flame AAS this typically is the
techniques Standard calibration , applying aqueous calibration solutions.
Std. prod. From the pull-down menu select whether the reference solutions will
be manually prepared or should be automatically diluted by the
autosampler from a standard stock solution. The latter option is only
accessible if an autosampler has been defined.
If the preparation of standard solutions has been set by autosampler, open the Stocks
window to define the stock solutions.
An empty screen is displayed. Click on New, to add stock solutions.
A screen is displayed in which the number of calibration blank values and calibration
standards can be defined.
In the Statistics page the statistical methods are selected to be used for the calibration and
sample measurement. The settings do not depend onto the selected calibration method and
remain unchanged with anyother changes of the method.
Type of statistics Select the favorized type of statistics. In AAS this typically is the
mean value statistics.
Confidence interval Select the required option. Details on this topic can be found in
calculation/probability the AspectCS® help function.
Number of repeat Define the number of repetitive measurements which should be
measurements performed for each calibration standard, for each quality control
sample or analysis sample.
Note:
The options for the automatic saving of Spectra and Methods are already activated under
Output. It is recommended to stay with this setting.
Then the desired number of decimal places and the printing sequence can be set for each
element.
A complete flame method has been created, which contains the optimized flame and
calibration parameters. Now the method should be saved under a unique name. For this
click on Save and enter a suitable method name. The method can also include a remark.
Close the window with OK.
Note:
When storing the methods the method version is automatically indicated with a version
number that automatically increases with every storage process.
The integration of quality control samples is described in chapter 2.4.5.
Click on the Sequence icon in the main menu or select New sequence from the File pull-
down menu.
When the last point for selection is reached, e.g., number x of samples, confirm with OK.
The selection is accepted and the selection window closes.
Note:
There also can be special actions added to the sequence, such as a shut-down of the
flame at the end of the measurements.
If you want to store the samples with information of additional parameters, for example, a
predilution, special weights, filling volumes or calculation into special units, do not enter the
sample names directly into the sequence screen, but select the button Samples.
Using the Samples →Sequence command, the sample data are copied into the
corresponding fields of the sequence row. Vice versa the content of the sequence can be
transferred into the sample ID page.
If manual measurements, follow the software prompts for aspirating reference solution and
standards, etc. or let the autosampler process the started measurement sequence.
During the measurement, the results screen fills with the measurement values and if not
otherwise set by service the following results windows are displayed,. The results windows
can be adjusted if necessary using the Options icon.
The results screen displays the overview of the measured data in chronological order.
After the measurement has been completed, the following parameters in the results screen
should be examined.
For absorbance values higher than 0.05 the RSD values should be below 2.0% for air-
acetylene measurements.
2.4.2.2 Spectrum
To check the calibration function, either open the Calibration icon or double-click
on the Calculate calibration line on the results screen.
The Calibration window opens.
Note:
A recalculation of datasets is not only recommended if the calibration parameters have to
be changed, but also if the correction mode (static/dynamic) or the number of evaluation
pixels is to be changed. Prerequisite for such a recalculation is, that the analysis spectra
are saved with the dataset. Refer to chapter 2.2.2.9.
To recalculate the data set perform the desired modifications in the calibration graph or in
the method and save the method again.
The print-out of measurement reports can be performed in different manners. The following
description shows one of the possibilities.
A reslope measurement is useful if the complete calibration has been measured some time
ago and the flame conditions have changed meanwhile. Applying a blank value and a
standard measurement, a slope adjustment can be performed.
First, in the method, define the standard that should be applied as the recalibration standard.
Therfore open the Method screen Calibration, Concentrations.
Note:
Select a standard from the
upper third of the
calibration range as the
recalibration standard.
Quality control samples (QC samples) are used for evaluation/test of a method and its
stability. Therefore there are samples or reference materials used whose analyte contents
are known and should be determined within specified limits. Optionally the results of such
QC samples can be stored automatically in QC/control charts.
Click on New/Modify
and select QC-Sample
from the pull-down
menu in the Add QC-
Sample type window.
If there is no database
for QC samples stored,
select enter manually
and confirm with OK.
Set a + sign under QC-Chart, as long as the measured data should be entered directly into
a QC/control chart.
Set a + sign under Reaction, if a measured sample, that is outside of the tolerance range,
should lead to the selected reaction.
An optimization of the baffle ball and nebulizer is necessary if the mixing chamber has been
removed and disassembled, for example, for cleaning purposes.
The optimization is performed manually, i.e. without any autosampler.
!! Note:
Always turn off the contrAA before the autosampler is removed. If the autosampler cable is
removed when the device is still turned on or the connection of the autosampler to the
device is interrupted, the electronics can be damaged.
Before you start the optimization procedure, please make sur that the nebulizer has well
been fixed in the mixing chamber.
Prepare a solution of 2 mg/L Cu for flame optimization.
corresponding icon
Now hold the aspiration hose briefly in pure water and check if water is aspirated or if air
bubbles occur in the water container.
For the following description of the flame optimization, a differentiation between two
possible nebulizer/spray chamber systems is necessary.
The currently shipped nebulizer /spray chamber system is built from brown or black poly
propylene sulfide and uses a plastic nebulizer. Flame optimization for this system is
described in the next chapter 2.5.1.1
Chapter2.5.1.2 describes the optimization for the whitespray chambers (teflon with stainless
steel nebulizer) that have been shipped until 2010.
Start the PC as described in chapter 2.1. and then the contrAA 600/700.
A basic configuration has already been performed by service during device installation.
For most of the elements that are determined by graphite furnace AAS the software
contains 'cookbook' values for the most important method parameters, such as matrix
modifiers, pyrolysis or atomization temperature,. When the elements to be measured are
selected, these cookbook values are automatically called up and a method is created that is
used as the starting point for further method development.
This cookbook data can be viewed as follows:
After closing the cookbook, click on the Method icon or alternatively open the File pull-
down menu and select the New Method command.
If a different method was already previously loaded, an additional query appears. Select
Based on default parameters.
The first tab, Lines has different columns with entries that can be edited if necessary.
Test time With graphite tube AAS, typically 3-8s; can be freely edited.
The Furnace tab contains the graphite furnace parameters that were loaded from the
cookbook.
The tab shows the most important parameters of the graphite furnace program, as well as
the matrix modifier to be used.
The first column shows the name of each temperature step. This can be edited if required.
Column 2 shows the actual temperature, column 3 shows the respective ramp rate in
°C/sec and column 4 the respective temperature hold time.
The program includes three drying steps, in which the sample is slowly evaporated. The
internal gas flow (Purge gas) is set to a maximum, thereby ensuring a controlled transport
of the resulting water and solvent vapor out of the graphite tube..
In a first pyrolysis step at 350°C, any remaining amounts of moisture are removed of the
graphite furnace. This generally prolongs the life-time of the graphite tube.
In the second pyrolysis step, the sample is decomposed and interfering sample matrix is
partially evaporated. Directly before atomization, an additional program step is carried out
at a constant pyrolysis temperature. In this step the internal gas flow is stopped in order to
create good conditions for the subsequent atomization. This program step is set by the
software itself and cannot be edited.
In the atomization step, the analyte is then evaporated at high ramp rates and high
temperature and is thus able to absorb the radiated light. This step triggers the start of the
actual absorption measurement.
In the cleaning step, the graphite tube is finally heated-out at high temperatures in order to
eliminate any residual sample matrix and to prepare the graphite tube for the next injection.
During the subsequent cool-down of the graphite tube, the system independently performs
a reference measurement.
If the program was edited, you can click on the Check program button to activate a
program logic check. A green LED next to the button indicates that the program does not
contain any inconsistencies.
Note:
The temperature program loaded from the cookbook is based on the physiochemical
properties of the analyte and follows empirical values. It offers a good starting program for
further method development, but must be further optimized based on a standard solution
and the sample to be determined.
The Sample transport tab is used to define the accessories installed with the contrAA.
The graphite furnace is equipped with the autosampler MPE60. This is already activated in
the method.
After each measurement, 2 wash cycles are preset for the autosampler capillary. This can
be changed if required.
Under Dilution you can define whether samples whose measured absorbance values are
calculated above the calibrated range are to be diluted in the graphite tube. With this option,
the sample volume is reduced, but the total volume is kept by the addition of an appropriate
amount of blank solution. Alternatively, the volume can be reduced without the addition of a
blank.
The Evaluation tab contains the line-specific evaluation parameters for the calculation of
the measurement results from the spectrum.
The specifications loaded from the cookbook offer good starting conditions for simple
applications in graphite furnsce AAS, but can be changed if necessary.
Spectrum with 200 evaluation pixels Spectrum with 100 evaluation pixels
Eval.Pixels The maximum of the absorption line is expected on pixel no. 101. Since the
absorption profile is larger than the range of an evaluation pixel, adjacent
pixels can also be included for evaluation, i.e. the absorption signal is
integrated over several measurement pixels. Generally, 3 or 5 pixels are
used for evaluation in graphite tube AAS.
Note:
An increase in evaluation pixels to 5P leads to a better relative standard
deviation and to a higher measurement sensitivity.
Lowering to 3P improves the calculation of the detection limit.
Before processing the next Calibration tab, it is recommended to optimize the time-
temperature program of the graphite furnace. A corresponding description can be found in
chapter 3.3.
Std. prep. From the pull-down menu select whether the reference solutions were
prepared manually or should be automatically diluted by the
autosampler from a standard stock solution. In this case, a defined
aliquot of the standard stock solution is pipetted into the graphite tube,
together with the volume of a diluent needed to make up the entire
volume. This usually is the blank solution.
If by variation of volume is selected, the addition of a diluent solution
is omitted.
Volumes The sample and calibration volume used for determination. The
default setting is 20µL. The sample volume can be changed if
required.
Note:
Increasing the pipetting volume will significantly lengthen the required
drying time, i.e., the time-temperature program will have to be
adjusted accordingly (see also chapter 3.3). For platform tubes the
maximum total pipetting volume is 40µL, for wall-type tubes it is 50µL.
Calib. func. Select the calculation type of the calibration function. Select between
linear, non-linear or automatic calculation. Details can be found in
the software help function.
Note:
The calculation type can be changed and the data can be recalculated
later at any time.
Intercept These parameters can be changed if required. With the basic setting,
and Weighting the intercept is calculated and the calibration data is not weighted.
Unit Select the unit of the calibration function In graphite furnace AAS this
generally is µg/L.
.
If the preparation of the standard solutions was defined by autosampler, open the Stocks
window to define the prepared stock solutions.
An empty screen is displayed. Click on New to add stock solutions.
Note:
For the calibration standards, a pipetting volume of less than 4µL is not recommended.
On the Statistics tab select the statistical methods to be used for calibration and sample
measurement. The settings are independent of the selected calibration mode and remain
unchanged with any other change of the method.
Statistics Select the desired type of statistics. In AAS this is generally the
mean value statistics.
Confidence Select the desired option here. Details can be found in the
interval AspectCS® help function.
calc./Confidence
level
Note:
In graphite furnace AAS, generally 2–3 replicate measurements
are used. For the measurement of highly refractory elements
such as vanadium or boron, it is recommended to select a Pre-
run. Then an initial measurement is performed, but the measured
value is not used for calculation of the mean value. Thus memory
effects can be reduced.
Under Output, some options for the automatical storage of Spectra and Methods are
already activated. It is recommended to keep these settings.
The desired number of decimal places and the printing sequence can now be defined for
each element.
First check the drying based on an aqueous standard as well as the sensitivity of the
measurement. For this, prepare a test standard. For the optimum test concentration, refer
to the cookbook.
icon.
If the use of a matrix modifier is recommended in the cookbook, then prepare the modifier
as described in the appendix.
Set the matrix modifier, as specified in the method, to position 89 or change the modifier
position in the method.
Set the test standard to any desired autosampler position.
Now program a simple measurement sequence.
To do so, click on the Sequence icon in the main menu or open the File pull-down menu
and select the New Sequence command.
The placement of the drop in the graphite tube can be observed using the furnace camera.
During the drying phase control the evaporation of the solvent.
Now optimize the sample drying in accordance with the following points:
The drop should spread out in the first two drying steps in the platform/on the tube wall and
slowly evaporate, without reaching boiling point. This can be achieved by optimizing the
temperatures, ramp and hold times.
Note:
Too rapid drying will lead to condensation of water vapor in the graphite contacts and the
furnace shroud, which on further heating can lead to corrosion and thus reduces the life-
time of the graphite parts.
The first pyrolysis step at 350°C serves to eliminate any residual water vapor from the
graphite tube before higher temperatures are reached. This step requires no further
optimization.
The pyrolysis temperature, which is loaded from the cookbook, is the maximum
temperature for the analyte when using the specified matrix modifier. If necessary, this
specified pyrolysis temperature must be lowered significantly.
If no matrix modifier is used, apply the maximum pyrolysis temperatures specified in the
cookbook under Remarks.
Optimization is based on the following points:
Note:
For the measurement of real samples, the pyrolysis temperature, ramp rate and hold time
are very important. Therefore, the ramp and hold times specified by the cookbook often
have to be adjusted to the samples. However, they are sufficient for a preoptimization of the
temperature program based on a simple aqueous standard.
3.3.3 Atomization
The atomization temperature is mainly optimized using the received signal profile. The
following points should be used for orientation:
The clean out temperature is preset from the cookbook so that the analyte is completely
evaporated out of the graphite tube. With high analyte contents or low volatile sample
matrix it may, however, be necessary to increase this temperature.
Note:
Avoid clean-out temperatures of over 2500°C
.Exception: the determination of highly refractory elements, such as V, Ti, Mo.
After this preoptimization of the temperature program, check the precision of the
determination for three replicate measurements. Ideally, this should not exceed a value of
3% for a signal of 0.1 A*s.
The actual optimization of the graphite furnace method is performed using a characteristic
sample.
Depending on the sample composition/sample matrix, the graphite furnace program has to
be modified. A high acid concentration of the sample can lead to a significantly different
drying behavior than an aqueous standard or also matrix components of the sample may
affect the volatility of the analyte.
The aim of the method development is to eliminate the sample matrix as well as possible,
before atomization is performed, however without evaporating the analyte itself. The
automatic optimization routine therfore provides good assistance.
For such an automated optimization, open the Furnace, Optimization screen.
Note:
Before starting the optimization, the method should be given a name and stored.
It is recommended to run the optimization applying a characteristic sample. If the analyte
content is very low (absorbance < 0.05),however, you should spike some standard to the
sample in order to obtain a reasonable signal value for optimization
After completion of the optimization run, a graph is shown which displays the analyte signal
as a function of the applied temperature.
Note:
The integration of quality control samples is described in chapter 2.4.5.
In the main menu click on the Sequence icon or open File from the pull-down menu and
select the New sequence command.
Note:
In graphite furnace technique a
reference measurement does
not need to be programmed. In
this technique it is auto-
matically performed by the
system.
When the selection has been performed, e.g., the number x of samples, confirm with OK.
The selection is accepted and the window closes.
The sequence is now displayed as follows:
If you like to store the samples with any corresponding information such as a predilution
factor, special weights and filling volumes or a certain sample unit do not enter the sample
name directly in the sequence screen, but use the Samples command in the window..
Alternatively you also may click on the ID samples button in the main menu.
This opens an empty Sample ID screen. Using the New/Append function individual
samples can be added and be stored with individual information.
Using the Samples →Sequence command, the sample data is stored or updated from the
specified sequence row in the sequence. Conversely, you can also load the content of the
sequence into the sample table.
Now you have created a simple sequence. Confirm it with OK.
The sequence screen is now displayed.
The autosampler now starts the measurement sequence. During this process, the results
screen is filled with the measurement values and the following results windows are
displayed, if not otherwise defined. Report window, spectra display, signal plot, bargraph
window and the image of the furnace camera.
The display of the signal plot is especially important.
If necessary the results windows can be adjusted via the Options, Analysis sequence
button.
The Results screen displays the overview of the measured data in chronological order.
QC Res. Displays the results for QC samples, recovery rates, detection limits etc. if
previously defined.
After the measurement is completed, the following parameters in the results screen should be
checked:
Relative standard deviation
For absorbances above 0.1 the RSD values should be below 5.0%.
Spectrum
Did you get a clear spectrum with an appropriatel selection of evaluation pixels for standard
solutions?
To check this, open the respective measurement row with a double-click on the line.
The wavelength-resolved
spectrum is displayed.
The connection pixels are
highlighted in grey.
To check the calibration function, double-click on the Calculate calibration line on the
results screen.
The Calibration window appears.
If you find that the type of calculation of one or more elements has to be changed or you
want to eliminate standards, the dataset will have to be recalculated.
Note:
In graphite furnace technique a spike recovery rate of 90-110% is considered to be ideal
but recovery rates of 85-115% are still acceptable.
Poorer recovery rates indicate the existence of a non-spectral interferences. This normally
cannot be corrected by repeating the measurement, but requires a standard addition
calibration to be performed (see chapter 3.5.4).
To add the spike to the sequence, select Sequence, Append, QC samples then add the
previously defined spike (refer to the figure on the following page))
The system always relates the measurement of the spike to the previously measured
sample and shows the result of the spike on the results screen under QC Res.
On the same screen you can define whether you want to prepare the addition standards
manually or use the automatic dilution option of the autosampler.
The Blank correction is preset to Absorbance corrected. More information on blank
correction set to Concentration corrected can be found in the Manual or the help function
in the AspectCS® software.
In the table, define the sample volume and the addition volume of the addition standards.
Note:
The blank is not defined in this screen. It is automatically inserted in the sequence (see
flame technique, chapter 2.4.7).
An addition calibration must be performed within the linear working range, since a non-
linear calculation is not possible. Exceeding the linear working range would result in large
measurement errors.
Typically, 3-4 addition standards should be used.
Now create a measurement sequence, as already described for the standard calibration.
After calibration has been performed, the typical calibration function of an addition will be
available (see chapter 2.4.7.)
To adjust the autosampler, first open the Autosampler symbol and go to the
Technical parameters screen.
Click on the Align MPE to furnace field.
1 Adjusting screw 1
2 Lock nut
3 Adjusting screw 2
4 Adjusting aid
5 Clamp nut
6 Adjusting screw 3
7 Lock nut
The screens that are now displayed contain an assistant function that will guide you through
the complete adjustment procedure. Carry out the following adjustment steps:
- Open the clamp-nut and push-out the dosing tube by approx. 8 mm from the canula
of the autosampler arm.
- Replace the dosing funnel in the graphite furnace by the adjusting aid with crosshair.
- Lower the autosampler arm via the software to the adjusting aid.
- Use the adjustment screws 2 or 3 as necessary to perform minor readjustment of x-
direction.
- Note:
Only adjust the x direction using the arrow keys when the autosampler capillary is
very far from the target position. In most cases, the fine adjustment described above
is sufficient.
- Adjust the y direction using adjusting screw 1.
- Readjust the x direction using adjusting screws 2 and 3.
- Check the adjustment by lowering the autosampler arm via the software. The dosing
tube must dip into the whole of the adjusting aid without any contact.
- Adjust the z direction via the software:
Lower the autosampler arm to the upper edge of the adjusting aid and then lower it
by two steps more (approx. 0.5 mm).
- The adjustments for the x and z direction are saved internally.
- Secure the positions of the adjusting screws with the lock nuts (2 and 7).
- Remove the adjusting aid and insert the funnel.
Energy spectrum of the reference measurement Energy spectrum of the sample measurement
The illustration shows the energy distribution of the reference and sample measurement
over the 200 measurement pixels of the detector.
In the sample spectrum, besides the energy drop on the central measurement pixel, which
is caused by the absorption of the analyte, a general energy decrease over the entire
wavelength range can be observed. This is caused by a broadband background absorption,
such as light scattering, or by a fluctuation of the lamp intensity.
By selecting connection pixels (dynamic or static), the detectable baseline is set to zero.
Then, the difference between baseline and the absorption maximum is calculated, which
provides a simultaneous and automatic compensation of the broadband background.
Since especially in graphite furnace technique it is important to know whether any high
broadband background is present, the amount of broadband background is calculated from the
reference and sample spectrum and shown as a red line in the signal plot.
Note:
This broadband background is calculated very roughly as the exact background absorption
value is not applied to calculate the specific signal. For this reason, the red line may show
very large differences between individual replicate measurements and a background signal
of less than 0.1A and below the baseline can therefore generally be neglected. The latter is
mainly caused by a fluctuation of the lamp intensity.
In flame and hydride techniques, the the broadband background absorption is not displayed
in the spectrum plot..
Structured background absorption is caused by molecules or matrix atoms that also absorb
within the considered spectral wavelength range.
If there is no direct overlay of this external absorption with the actual measurement
wavelength, this has no effect on the measured value of the analyte. It is recognized by the
dynamic correction algorithm as a real spectral structure and not included into the
calculation of the baseline.
Example: The absorption of nickel in the vicinity of the bismuth absorption line
The nickel line is clearly separated from the analysis wavelength and is recognized by the
dynamic background correction as a real absorption structure.
Now select the line to be identified by a mouse click. The spectra screen and the line table
now display the respective interfering element or molecule.
Note:
If there is no direct overlay of element lines or molecular structures with the analysis
wavelength, the interference line in the spectrum is automatically corrected, i.e., it is not
included in the calculation of the baseline. A User action is not necessary.
In graphite furnace technique, yellow to orange-red bars in the signal plot indicate the
presence of structured background absorption. For such signals, check whether there is a
direct overlay of these structures with the measurement wavelength or whether they are
clearly separated from the analyte line.
In the signal plots, the red line indicates the presence of broadband background absorption,
the yellow and orange bars indicate structured background absorption.
In the right signal plot, an incorrect compensation is also visible. In this case, it is absolutely
necessary that the sample spectrum is checked for interferences.
Therefore, various auxiliary tools are offered in the spectra display.
Open the signal window of the measurement by means of a double-click on the result line
and select Spectrum. The time-averaged spectrum is shown.
While it is not directly possible to evaluate the interference in the averaged spectrum, the
three-dimensional spectrum plot indicates that the interference only occurs at the end of the
graphite furnace measurement.
For the identification of interfering spectra, the spectra database can be used as described
above.
The list offers molecule and element spectra, which can be simply selected and compared
with the structures in the spectrum. In the example a direct overlay of the Zn line by NO
molecule bands is shown, caused by a high concentration of nitric acid. In a similar manner,
a high content of iron will interfere by an overlaying iron line.
To compensate for the direct overlay of element and molecule absorption lines, it is
necessary to perform a spectrum correction. This is described in the following chapter.
First measure the individual matrix components that were found to cause the spectral
overlay.
Note:
The concentrations of the matrix components do not have to exactly match the
concentration in the samples. However, the concentrations should be high enough to
generate clear absorbance values and, if possible, be in a similar concentration range than
in the sample.
Example:
For the overlay of the zinc line by HNO3, a nitric acid solution with a higher concentration
(e.g., 5%) is measured as a sample.
After the measurements, open the measurement line with a double-click and open the
spectrum display via the Show spectrum button.
Save the spectrum of the matrix component as a correction spectrum by using the Corr.
spectrum button. Define a spectrum name and close the process with Save.
Now open again the measurement values and the spectrum display of the sample via the
Show spectrum button.
Activate the Model button. Using the New/modify model option open the Spectral
corrections window. The sample spectrum is displayed.
Note:
Using the Mask. button and holding down the left mouse button, certain ranges in the
spectrum can be masked that should not be included into the calculation of the correction
model. The range of the analyte line (pixel 92-110) is already masked by default. Masking
of more ranges can be necessary if no correction spectrum was recorded for individual
components of the sample or if additional permanent structures exist.
After the method has been saved, the future measurements with this method will be
performed with the created correction model. Measurements already completed can also be
recalculated with this method, so that a new measurement is not necessary.
4.2.1 General
IBC stands for Iterative Background Correction. It was developed and patented by the
Institute for Analytical Spectroscopy (ISAS) in Berlin.
The method is similar to using a high-pass filter in electronics. Here, lower frequencies are
attenuated or eliminated completely, but high frequencies are not affected. This results in a
reduction of noise or a smoothing of the baseline.
Using this method is advantageous
- for measurements in the lower wavelength range (<200nm)
- with the presence of strong molecule bands within the detected range
- and with broadband background absorption which is directly overlays the
measurement wavelength.
A noisy baseline can be observed especially in the wavelength range below 200 nm caused
by the absorption. This complicates the selection of suitable connection pixels for baseline
correction. The IBC algorithm can be used here to obtain a smoothened baseline (refer to
the figure).
Dynamic
mode
The IBC option is also advantageous if the broadband absorption of a matrix component is
directly overaying the analysis line and the dynamic background compensation cannot
calculate any suitable connection pixels.
This occurs, for example, through the broadband absorption of aluminum in the wavelength
range of 193 nm. The arsenic line is located directly in the flank of aluminum absorption.
While the dynamic background compensation fails here, the IBC option results in an
evaluable spectrum.
In the Method, Evaluation window, under Background correction activate the IBC
option.
This deactivates the BGC fit column.
Notes:
When using the IBC mode, measurement sensitivity is decreased by about 30%. However,
depending on the application, improved detection limits can be achieved.
When using spectrum corrections, the spectrum to be corrected and the correction spectra
must be recorded with the same correction method (IBC or reference).
It is recommended to test by means of recalculations whether or not the IBC method is
advantageous for the respective application.