Professional Documents
Culture Documents
G6PDH Mak015bul
G6PDH Mak015bul
TECHNICAL BULLETIN
Storage/Stability
The kit is shipped on wet ice and storage at –20 °C,
protected from light, is recommended.
2
Results
Calculations The G6PDH activity of a sample may be determined by
Correct for the background by subtracting the final the following equation:
measurement [(A450)final] obtained for the 0 (blank)
NADH standard from the final measurement [(A450)final] G6PDH Activity = B × Sample Dilution Factor
of the standards. Background values can be significant (Reaction Time) × V
and must be subtracted from all readings. Plot the
NADH standard curve. B = Amount (nmole) of NADH generated between Tinitial
Note: A new standard curve must be set up each time and Tfinal.
the assay is run. Reaction Time = Tfinal – Tinitial (minutes)
V = sample volume (mL) added to well
Calculate the change in measurement from Tinitial to Tfinal
for the samples. G6PDH activity reported as nmole/min/mL (milliunit/mL)
One unit is the amount of enzyme that catalyzes the
∆A450 = (A450)final – (A450)initial conversion of 1.0 µmole of glucose-6-phosphate to
6-phosphoglucono-δ-lactone and generates 1.0 µmole
Compare the ∆A450 of each sample to the standard of NADH per minute at 37 °C.
curve to determine the amount of NADH generated by
the kinase assay between Tinitial and Tfinal (B). Example:
NADH amount (B) = 5.84 nmole
First reading (Tinitial) = 3 minute
Second reading (Tfinal) = 32 minutes
Sample volume (V) is 0.01 mL
Sample dilution is 1
Troubleshooting Guide
Problem Possible Cause Suggested Solution
Assay Buffer Ice Cold Assay Buffer must be at room temperature
Omission of step in procedure Refer and follow Technical Bulletin precisely
Plate reader at incorrect wavelength Check filter settings of instrument
Assay Not Working
For fluorescence assays, use black plates
Type of 96 well plate used with clear bottoms. For colorimetric assays,
use clear plates
Use the Assay Buffer provided or refer to
Samples prepared in different buffer
Technical Bulletin for instructions
Repeat the sample homogenization,
Cell/Tissue culture samples were
increasing the length and extent of
incompletely homogenized
homogenization step.
Samples with erratic
Samples used after multiple freeze-thaw Aliquot and freeze samples if needed to use
readings
cycles multiple times
Presence of interfering substance in the
If possible, dilute sample further
sample
Use of old or inappropriately stored Use fresh samples and store correctly until
samples use
Thaw all components completely and mix
Improperly thawed components
gently before use
Use of expired kit or improperly stored Always check the expiration date and store
Lower/Higher reagents the components appropriately
Readings in Samples Allowing the reagents to sit for extended Always prepare fresh Master Reaction Mix
and Standards times on ice before use
Refer to Technical Bulletin and verify correct
Incorrect incubation times or temperatures
incubation times and temperatures
Incorrect volumes used Use calibrated pipettes and aliquot correctly
Thaw and resuspend all components before
Use of partially thawed components
preparing the reaction mix
Pipetting errors in preparation of standards Avoid pipetting small volumes
Prepare a Master Reaction Mix whenever
Pipetting errors in the reaction mix
possible
Non-linear Standard
Air bubbles formed in well Pipette gently against the wall of the tubes
Curve
Standard stock is at incorrect Always refer to the dilutions in the Technical
concentration Bulletin
Recheck calculations after referring to
Calculation errors
Technical Bulletin
Substituting reagents from older kits/lots Use fresh components from the same kit
Samples measured at incorrect
Check the equipment and filter settings
wavelength
If possible, dilute sample further
Unanticipated results Samples contain interfering substances
Deproteinize samples
Sample readings above/below the linear Concentrate or dilute samples so readings
range are in the linear range
LS,MF,MAM 02/13-1
2013 Sigma-Aldrich Co. LLC. All rights reserved. SIGMA-ALDRICH is a trademark of Sigma-Aldrich Co. LLC, registered in the US and other
countries. Sigma brand products are sold through Sigma-Aldrich, Inc. Purchaser must determine the suitability of the product(s) for their
particular use. Additional terms and conditions may apply. Please see product information on the Sigma-Aldrich website at
www.sigmaaldrich.com and/or on the reverse side of the invoice or packing slip.