You are on page 1of 3

Designation: F 2203 – 02 (Reapproved 2007)

Standard Test Method for


Linear Measurement Using Precision Steel Rule1
This standard is issued under the fixed designation F 2203; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

1. Scope TABLE 1 Description of Materials Measured in Round Robin


Studies
1.1 This test method covers the measurement of linear
Measurement Measurement Ruler Scaling
dimension of flexible packages and packaging materials. It is Set
Material Type
Characteristic Applied
recommended for use with an allowable tolerance range of 3
A Foil Web Web Width in.
mm (1⁄8 in.) or greater based on gage repeatability and B Paper Web Web Width in.
reproducibility presented in the Precision and Bias section. C Spunbonded Olefin Web Web Width mm
D Printed Foil Web Print Repeat Length in.
2. Referenced Documents E Printed Paper Web Print Repeat Length in.
F Printed Spunbonded Olefin Print Repeat Length mm
2.1 ASTM Standards: 2 Web
D 1898 Practice for Sampling of Plastics3 G Forming Film Web Web Width mm
H Finished Pouch Dimension A in.
D 4332 Practice for Conditioning Containers, Packages, or I Finished Pouch Dimension B in.
Packaging Components for Testing J Finished Pouch Dimension C in.
K Finished Pouch Dimension D in.
E 691 Practice for Conducting an Interlaboratory Study to L Finished Pouch Dimension E in.
Determine the Precision of a Test Method M Finished Pouch Dimension F in.

3. Terminology
3.1 Definitions: 4.3 This test method does not address acceptability criteria.
3.1.1 linear dimension—the measurement of length, width These need to be jointly determined by the user and producer
or relative positions. of the product.
3.1.2 parallax error—the error resulting from a change in
observational relationship to a fixed position. For example, 5. Apparatus
looking at a measurement gage from different angles can cause 5.1 Precision Steel Rule:
variation in the reporting of that measurement. 5.1.1 Tempered steel rule in increments of 1⁄64 in., 100ths, or
1⁄2 mm.
4. Significance and Use 5.1.2 Steel rule should be of sufficient length to measure full
4.1 This test method provides a means for measuring linear dimension of interest.
dimensions. Accurate measurement of dimensions can be 5.1.3 It is recommended that a calibration be performed on
critical to meeting specifications and characterizing process the apparatus used and it is certified to a recognized industry
performance. standard.
4.2 This test method should not be applied to tolerance
ranges of less than 3 mm (1⁄8 in.) when it is preferable that test 6. Sampling
error does not exceed 30 % of tolerance range. See Precision 6.1 The number of samples tested should be adequate to be
and Bias Section for gage repeatability and reproducibility predictive of performance. Caution should be taken when
results. eliminating samples with defects as this can bias results.
6.2 See Practice D 1898 for guidance on sampling practices.
1
This test method is under the jurisdiction of ASTM Committee F02 on Flexible 7. Conditioning
Barrier Packaging and is the direct responsibility of Subcommittee F02.20 on 7.1 Conditioning of the samples will depend on the material
Physical Properties.
Current edition approved April 1, 2007. Published May 2007. Originally under evaluation. If conditioning before testing is appropriate,
approved in 2002. Last previous edition approved in 2002 as F 2203 – 02e1. normal, and desirable, then condition the test specimens at 23
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or 6 2°C (73.4 6 3.6°F) and 50 6 5 % relative humidity for not
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
less than 24 h prior to test.
the ASTM website. 7.2 See Practice D 4332 for guidance on conditioning prac-
3
Withdrawn. tices.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

1
F 2203 – 02 (2007)
TABLE 2 Summary of Interlaboratory Results by Measurement Set
Repeatability Reproducibility 95 % 95 %
Measurement Average Standard Standard Repeatability Reproducibility
Set mm Deviation Deviation Limit Limit
sr sR r R
A 266.85 0.162 0.204 0.459 0.576
B 777.30 0.324 0.424 0.917 1.200
C 826.25 0.167 0.267 0.472 0.756
D 82.42 0.132 0.199 0.374 0.564
E 279.47 0.000 0.162 0.000 0.459
F 379.78 0.204 0.319 0.578 0.903
G 326.78 0.204 0.387 0.578 1.094
H 133.57 0.162 0.253 0.459 0.717
I 253.65 0.094 0.284 0.265 0.803
J 9.97 0.162 0.280 0.459 0.791
K 18.10 0.162 0.204 0.459 0.576
L 114.41 0.229 0.339 0.648 0.959
M 225.91 0.229 0.298 0.648 0.843

TABLE 3 Summary of Interlaboratory Averaged Results 8.2 Lay sample to be measured on flat surface with suffi-
Average Test Value 284.2 mm (11.19 in.) cient color contrast to distinguish the edge of the sample.
Repeatability (within a laboratory) standard deviation (sr) 0.187 mm (0.0074 in.) Sample should lay flat and smooth without wrinkles, creases or
95 % repeatability limit (r) 0.528 mm (0.0208 in.)
Reproducibility (between laboratories) standard 0.288 mm (0.0113 in.) folds. Material should not be under tension when measured.
deviation (sR) 8.3 Lay ruler over product or place product on top of ruler
95 % reproducibility limit (R) 0.815 mm (0.0321 in.)
so that the characteristic to be measured can be clearly viewed
and referenced to the ruler. Care should be taken to properly
align sample to avoid skewing errors. A ruler may be set on
8. Procedure edge along the surface rather than laid flat if the risk of parallax
8.1 Review applicable specifications, drawings, or proce- error is apparent.
dures. Specify unit of measure to be used and directions related 8.4 Carefully align the starting point of the measurement to
to precision requirements (for example, measure to nearest 0.5 the leading edge of fixed scale division on rule (Fig. 1,
mm, round up or down to nearest 0.5 mm, to 1⁄64 in., etc.). Example A). Then measure to the end point of the character-
Depending upon the level of accuracy and visual acuity, image istic. Note the scale division on the ruler that corresponds to the
magnification may be an aid. end point of the measurement. Determine the measurement by

NOTE—Example A and Example B begin at the leading edge of fixed scale division on rule at a point within the scale, at 1 in. or 10 mm, for example.
FIG. 1 Proper Alignment to Ruler Scale Divisions

2
F 2203 – 02 (2007)
subtracting the start point measurement from the end point. for test results that are averages from testing 3 specimens for
Record the number in the unit of measure and precision each test result, then the following applies:
required by specification. 10.3 Repeatability “r” is the interval representing the criti-
cal difference between two test results for the same material,
9. Report
obtained by the same operator using the same equipment on the
9.1 Report the following information: same day in the same laboratory. Two test results shall be
9.1.1 Lot number and source of material, date, time, loca- judged to be not equivalent if they differ by more than the “r”
tion and operator of test and complete identification of mate- value for that material, in this instance approximately 0.528
rials being tested, mm. That is, it can be expected that the same operator
9.1.2 Any conditioning of the materials, measuring the same sample may obtain a reading that varies by
9.1.3 Any and all deviations from standard, and up to 0.528 mm (0.0208 or approximately 1⁄64 in.).
9.1.4 The sampling plan and number of specimens tested
10.4 Reproducibility “R” is the interval representing the
along with test results.
critical difference between two test results for the same
10. Precision and Bias material, obtained by different operators using different equip-
ment in different laboratories, not necessarily on the same day.
10.1 Precision—A research report4 describes a round robin
Two test results shall be judged to be not equivalent if they
conducted in 2001 in accordance with Practice E 691, involv-
ing seven laboratories measuring 13 dimensional attributes of differ by more than the “R” value for that material, in this
five packaging materials. Materials and measurement attributes instance approximately 0.815 mm. That is, it can be expected
are listed in Table 1. Measurements taken included use of steel that different operators using different equipment in different
rules scaled in inch and millimetre divisions. Statistical sum- laboratories may obtain readings that vary by up to 0.815 mm
maries of repeatability (within a laboratory) and reproducibility (0.0320 or approximately 1⁄32 in.).
(between laboratories) are listed in Tables 2 and 3. All test 10.5 Any judgment in accordance with 10.3 or 10.4 will
results are expressed in SI millimetre (mm) units of measure. have approximately 95 % (0.95) probability of being correct.
10.2 Concept of “r” and “R” in Tables 1 and 2—If Sr and 10.6 Bias—There are no recognized standards by which to
SR have been calculated from a large enough body of data, and estimate the bias of this test method.

4
11. Keywords
A research report is available from ASTM International Headquarters. Request
RR:F02–1017. 11.1 dimension; linear measurement; rule; scale

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

You might also like