Professional Documents
Culture Documents
1. Principle:
Formation of Atomization, hollow cathode lamp, mono chromator and detector are the main
components in atomic absorption spectroscopy instrumentation.
2. Applications:
The AAS is used for identification, detection, determination of metal elements
1. Ag(Silver) 17. Eu (Europium) 33. Mo(Molybdinum)
2. Al(Aluminium) 18. Fe(Iron) 34. Na(Sodium)
3. As (Arsenic) 19. Ga (Gallium) 35. Nb (Niobium)
4. Au (Gold) 20. Gd (Gadolinium) 36. Nd (Neodymium)
5. B (Boron) 21. Ge (Germanium) 37. Ni(Nickel)
6. Ba(Brium) 22. Hf(Hafnium) 38. Pb(Lead)
7. Be(Berelium) 23. Hg(Mercury) 39. S(Sulpher)
8. Bi (Bismuth) 24. Ho(Holmium) 40. Sb (Antimony )
9. Ca(Calcium) 25. In (Indium) 41. Se(selenium)
10. Cd(Cadmium) 26. Ir (Iridium) 42. Si(silica)
11. Co (Cobalt) 27. K (Potassium) 43. Sn(tin)
12. Cr (Chromium) 28. La (Lanthanum) 44. Sr(strontium)
13. Cs (Cesium) 29. Li (Lithium) 45. V(vanadium)
14. Cu(Copper) 30. Lu (Lutetium) 46. Ti(titanium)
15. Dy (Dysprosium) 31. Mg (Magnesium) 47. Zn(zinc)
16. Er (Erbium) 32. Mn(Manganese) 48. Zr(zirconium)
Procedure instrumentation and reporting of Inorganic sample By AAS
3. Safety:
Laboratory should provide the following personnel protective equipment’s to the testing
personnel’s
Cartridge mask
Gloves
Heat resistant gloves
Tongs
Eye washer
Safety shower
4. Machine Operation:
Steps to be followed to analysed Sample:
a. Switch on the fume hood
b. Switch on the air Pump System(330 to 370kpa) 2.4 l/min
c. Switch on the computer
d. Turn on the Acetylene gas(55 to 96 kpa),1.2 L/min
e. Switch on the AAS System
f. Ignite the burner and leave the system for 30 min for optimization
g. The working area must be completely clear of all hazardous materials including
corrosive liquids and flammable solvents.
h. The liquid trap is filled with the same solvent as will be used in the sample matrix.
i. The drain tubing is positioned sloping downwards and is free of kinks or sharp bends
so that the waste liquid can drain freely into the vessel. The end of the tube must be
above the expected level of waste liquid in the vessel.
j. Vent tube is installed, if necessary..
k. The exhaust system is operating.
l. Gasses are turned on and set to the correct pressures as per the specifications.
m. If fitted, the heater on the N2O regulator is turned on.
n. The correct burner is fitted.
o. The flame shield, front panel and chimney have been fitted.
p. The pressure relief bung has been lubricated with distilled water and installed
correctly.
q. All O-rings have been inspected. Immediately replace any O-rings that are cracked or
deformed.
5. Preparation of Standard:
6. Preparation of samples
Prepare the samples as per the test method along with method blank, reagent blank, QC
and Spike
7. Operational Software :
Index Page:
When start the SpectrAA software, the first screen to appear is the introductory screen. The
Index page contains four buttons: Worksheet, Reports, Administration, and Exit.
The function of each button is as follows: Worksheet Opens the Worksheet window
Reports opens the reports window allows to generate a report for the current worksheet or
any of the worksheets saved in the system. The Reports window consists of four tabbed
pages: Worksheet, Select, Settings, and Report. To select a particular
Administration Opens the Administration window The Administration window allows to:
migrate worksheets into the system (such as old worksheets on CD-ROM, etc.); delete
worksheets and library methods; and activate and modify password protection of the system.
The Administration window also allows to configure system.
Use the Develop page to add, delete, review and modify methods, modify sequence
parameters, change the order of methods and copy methods to the Method Library.
Refer page no 4
Use the Labels page to set up solution labels, weights, volumes, and dilution
factors and to configure SPS or PSD samplers.
Use the Analysis page to control the AA and initiate analysis. Data is presented in
the form of a spread sheet, with one row per sample. The sample labels are
displayed in the left-hand column with concentration results for each element
across the remaining columns.
The SpectrAA software enables to have up to a maximum of 10 worksheets open at the same
time. The Worksheet tab at the bottom of the Analysis page enables to switch between
worksheets.
Worksheet menus
When viewing the worksheet, the following menu items are available:
File
This menu allows to perform filing tasks such as opening, saving and renaming files.
Edit
The Edit menu allows to copy and paste information, Edit Replicates and Go To certain parts
of a worksheet.
View
The View menu enables to nominate what is displayed on the Analysis page.
Instrument This menu enables to perform many Instrument functions such as Optimization,
Reslope, and Zero.
Options
The Options menu enables to alter the displays on the Analysis page.
If working with a worksheet created from a template, you can also add a method to your
worksheet.
Note: (a) wait until a stable flame is established before measurement. (b) Press the red
button next to the flame button will turn off the flame.(c) The flame can be lit up only when
open a worksheet.
(b) Next, place sample solution for test. If the concentration of sample solution is below
the standard, the result is obtained and shown in the Analysis window. If sample is in
the over range, the SIPS will perform a dilution to make the concentration of the
testing analyte within the standard range, then a test is done and the concentration of
sample is recorded. However, if the concentration of sample is beyond the largest
dilution factor that the SIPS can do, the calibration will fail. In this case, an error flag
will be displayed in the data cell, for example, “1366.89D”.
After measurement of samples, make sure to save results. The software will
automatically save worksheet with results when close the worksheet or exit the
SpectrAA.
Inject distill water after sequence completion as a maintenance procedure. It just
rinses the capillary tube, the liquid trap, the spray chamber and the burner to remove
any residual of previous sample.
Turn off the flame by press the red button next to the flame button. When turn off
the flame, a warning message on the screen telling that the SIPS is stopped because
the flame is off. This is normal.
view or print result. To do so, go to the Window menu and select the Report, then
select worksheet and Report, window like this will generates
Procedure instrumentation and reporting of Inorganic sample By AAS
There is no software control to turn off the lamps except switch the lamp. The lamp is off
when exit SpectrAA.
8. Exit SpectrAA.
9. Turn off the power of the instrument.
10. Close the Acetylene valve and turn off the Air valve.
11. Clean up all testing solutions.
12. After 10 min turn off fume hood.
The detail information about the use of the software and some explanations about the use of
this instrument can be found in the on-line help in the SpectrAA.
Calculation
Total lead (Pb) content, mg/kg = (C/W) × V × F
% = (C/W) × V × F × (1/10000)
where C = concentration of Pb in the sample solution ( mg/L)
V = volume of sample solution (mL)
F = dilution factor if necessary
W = sample weight (g)
(Note: * The Consumer Product Safety Improvement Act of 1008 (CPSIA) requires the
regulatory limit for lead in paint and other similar surface coatings to be reduced from
0.06% to 0.009% on August 14, 2009.)
Remarks:
1.If total lead content is greater than the requirement, the result shall be verified by another
technique (e.g. ICP-MS, see Appendix 2).
2.Any sample possibly having greater than 80% of the limit for lead in paint in any of the
composited paints should be retested individually.
(c) On Flame systems, also clean the front panel and flame shield. If necessary, use a mild
detergent to clean the cover. Do NOT use an abrasive cleaner. When analyzing biological
samples, wipe the instrument with chlorhexidine in 95% ethanol.
Windows Clean the sample compartment windows and the lamp compartment window
with a gentle stream of filtered air, or a puffer brush, like those used to clean camera
lenses.
If the windows are dirty, wipe them using an optical tissue held between tweezers and
dampened with a 50% ethanol/water solution.
Filters Check the filter in the air compressor supply line and clean if necessary
Gas cylinders
Change gas cylinders according to the instructions supplied by the manufacturer. When
changing gas cylinders:
Procedure instrumentation and reporting of Inorganic sample By AAS
Inspect all gas supply pipes and hoses. Replace any hose showing signs of damage or
deterioration.
Test all connections for leaks using a commercial leak testing solution, a mild dish washing
soap solution, or an electronic leak detector.
Test the operation of all regulators.
Ensure that shut off valves are correctly functioning.
Weekly
Check for O-ring deterioration:
Three O-rings inside the burner, spray chamber, and the bung
Two o-rings that seal the gas inlets to the spray chamber
Procedure instrumentation and reporting of Inorganic sample By AAS