Trigger
► what starts the breath – pressure, flow, time
Inspiratory limit/control variable
► Pressure, flow, volume
Cycling mechanism
► what terminates the breath – pressure, volume, flow,
time
Baseline
AND
Other controls
Pressure-Regulated Volume Control
4
4
5
3 Flow
5
1 6
2
3
1 6 4 Volume
5
2
3
1 6
1. PEEP 2
2. Triggering
3. Rate of rise of pressure
4. Pressure support level
5. Termination of breath
6. PEEP
How it works
Decreased compliance
MAINTAIN
Normal compliance DELIVERED
4
5 TIDAL
Increased compliance VOLUME
3
1 6
2
Indications for PRVC
4
5
3 Flow
5
1 6
2
3
1 6 4 Volume
5
2
3
1 6
1. PEEP 2
2. Triggering
3. Rate of rise of pressure
4. Pressure support level
5. Termination of breath
6. PEEP
Volume-support ventilation
4
4
5
3 Flow
5
1 6
2
3
1 6 4 Volume
5
2
3
1 6
1. PEEP 2
2. Triggering
3. Rate of rise of pressure
4. Pressure support level
5. Termination of breath
6. PEEP
Volume-support
Dual control breath-to-breath
Closed-loop control of Pressure-Support ventilation
ƒ Uses tidal volume as the feedback control for continuously
adjusting the pressure support level
Trigger
ƒ Flow or pressure
Inspiratory limit variable
ƒ Pressure and volume
Cycling
ƒ Flow cycling
Baseline is PEEP
Logic for Volume Support
Volume-Assured Pressure Support
Dual control within a breath
Called VAPS (Bird 8400Sti and TBird)
Called Pressure Augmentation (Bear 1000)
Amato et al (Chest 1992) Haas et al (Respir Care 1995)
ƒ VAPS reduced work of breathing by 50% over volume-control
ventilation
ƒ Improved dynamic compliance
ƒ Reduced Auto-PEEP
Volume-assured Pressure Support
Combines desired aspects of volume and pressure-
targeted ventilation
Trigger
ƒ Pressure or flow
Inspiratory limit variable
ƒ Dual control within a breath
ƒ At first, it is pressure-limited (IT IS PRESSURE-SUPPORT)
ƒ If the computer senses that the desired tidal volume will not be
delivered, it switches to constant-flow volume control
ƒ The flow and duration is determined by the remaining tidal
volume to be delivered
Cycling
ƒ Flow-cycling or time-cycled
Baseline is PEEP
Ideal ventilatory support in ALI/ARDS
1. Venturi valve
2. Exiratory valve
3. Timing switch
► Respiratory rate
► PaO2
► Compliance
John Down’s Editorial
► Institute APRV
John Downs
Phigh
► Above the mean airway pressure on conventional mechanical ventilation
► Below the pause pressure (alveolar pressure)
Plow
► Should be set at a level
• Does not allow alveolar collapse
• Does not produce auto-PEEP
Thigh
► No specific recommendations
► Start with an I:E 1.5-2:1 and then increase if tolerated and necessary
Tlow
► No specifc recommendations
► Should not be so low that it produces auto-PEEP
► Should not be set too high that it encourages alveolar collapse
Things to monitor
Respiratory
► Work of breathing
► RR
► Comfort of breathing
► Asynchrony/Dyssynchrony
► Gas exchange
• Oxygenation
• Ventilation
Cardiovascular
► HR
► BP
► Perfusion
► CVP