You are on page 1of 1

AS138-005 Rev.

11/15
®
CBC-3D
for MINDRAY
HEMATOLOGY CONTROLS
CONTROL

ASSAY VALUES AND EXPECTED RANGES


LOT B0823
2023-11-05
Instrument Parameter Low Normal High
LOT B0823L LOT B0823N LOT B0823H
3
Mindray BC-3600 WBC 10 /uL 1.9 ± 0.5 7.5 ± 1.0 19.7 ± 2.5
6
CD version RBC 10 /uL 2.15 ± 0.18 4.56 ± 0.24 5.58 ± 0.30
1.8(FDA) or greater HGB g/dL 5.5 ± 0.4 13.5 ± 0.6 18.5 ± 0.8
HCT % 16.6 ± 1.5 38.8 ± 2.0 52.2 ± 2.4
MCV fL 77.0 ± 5.0 85.0 ± 5.0 93.5 ± 5.0
MCH pg 25.6 ± 2.5 29.6 ± 2.5 33.2 ± 2.5
MCHC g/dL 33.2 ± 3.0 34.8 ± 3.0 35.5 ± 3.0
RDW-CV 15.0 ± 3.0 14.5 ± 3.0 14.0 ± 3.0
3
PLT 10 /uL 60 ± 20 248 ± 40 525 ± 60
MPV fL 7.3 ± 3.0 7.3 ± 3.0 7.7 ± 3.0
LYM % 58.5 ± 9.0 30.0 ± 8.0 14.0 ± 8.0
MONO % 8.5 ± 8.0 7.0 ± 6.0 3.5 ± 3.0
GR % 33.0 ± 9.0 63.0 ± 8.0 82.5 ± 9.0
3
LYM x 10 /uL 1.1 ± 0.2 2.3 ± 0.6 2.8 ± 1.6
3
MONO x 10 /uL 0.2 ± 0.2 0.5 ± 0.5 0.7 ± 0.6
3
GR x 10 /uL 0.6 ± 0.2 4.7 ± 0.6 16.3 ± 1.8
MINDRAY BC-3200 WBC 103/uL 2.1 ± 0.5 7.9 ± 1.0 20.7 ± 2.5
6
RBC 10 /uL 2.18 ± 0.18 4.59 ± 0.24 5.60 ± 0.30
HGB g/dL 5.6 ± 0.4 13.4 ± 0.6 18.2 ± 0.8
HCT % 16.7 ± 1.5 39.3 ± 2.0 53.5 ± 2.4
MCV fL 76.6 ± 5.0 85.7 ± 5.0 95.5 ± 5.0
MCH pg 25.7 ± 2.5 29.2 ± 2.5 32.5 ± 2.5
MCHC g/dL 33.5 ± 3.0 34.1 ± 3.0 34.0 ± 3.0
*RDW-CV 15.4 ± 3.0 14.4 ± 3.0 13.7 ± 3.0
3
PLT 10 /uL 72 ± 20 253 ± 40 506 ± 60
*MPV fL 8.6 ± 3.0 7.6 ± 3.0 7.8 ± 3.0
LYM % 50.5 ± 12.0 25.5 ± 8.0 14.0 ± 6.0
*MONO % 16.5 ± 10.0 11.0 ± 8.0 7.0 ± 6.0
GR % 33.0 ± 9.0 63.5 ± 8.5 79.0 ± 8.0
3
LYM x 10 /uL 1.1 ± 0.3 2.0 ± 0.7 2.9 ± 1.3
3
*MONO x 10 /uL 0.3 ± 0.3 0.9 ± 0.7 1.4 ± 1.3
3
GR x 10 /uL 0.7 ± 0.2 5.0 ± 0.7 16.4 ± 1.7
*For new software version 01.07.00.
For technical assistance in the USA and Canada call Technical Service at (800) 523-3395.
All brands and products are trademarks or registered trademarks of their respective companies.

R&D Systems, Inc.


614 McKinley Place NE
Minneapolis, MN 55413 USA
AS138-005 Rev. 11/15

You might also like