Ablation Zone

Reference Guide

Ablation Zone Reference Guide 1


Width (cm) x Depth (cm)

Width (cm)
*Note: Ex vivo bovine liver data may differ from human clinical outcomes. Depth (cm)

18G 1:1 Real Size 16G 1:1 Real Size


12W 45W 110W 135W 75W 120W 140W 170W
2 min 5 min 6 min 8 min 6 min 8 min 10 min 12 min
0.66 x 0.73 1.30 x 1.50 1.80 x 2.10 2.30 x 2.98 1.70 x 1.65 2.20 x 2.40 2.70 x 3.10 3.25 x 3.50

5 mm 10 mm 15 mm 20 mm 10 mm 15 mm 20 mm 25 mm

165W 175W
10 min 12 min 180W 195W
2.70 x 3.48 3.01 x 3.90 12 min 15 min
3.50 x 4.20 4.00 x 5.30

25 mm 30 mm
30 mm 40 mm

17G 1:1 Real Size 15G 1:1 Real Size


55W 115W 145W 175W 125W 140W 175W
5 min 6 min 8 min 10 min 8 min 10 min 12 min
1.60 x 1.63 2.00 x 2.30 2.43 x 3.00 2.85 x 3.50 2.50 x 2.50 2.90 x 3.10 3.40 x 3.50

10 mm 15 mm 20 mm 25 mm 15 mm 20 mm 25 mm

180W 195W
180W 195W 12 min 18 min
12 min 15 min 3.80 x 4.25 4.40 x 5.30
3.20 x 4.10 3.87 x 5.20

30 mm 40 mm
30 mm 40 mm

2 Ablation Zone Reference Guide Ablation Zone Reference Guide 3


Width (cm) x Depth (cm)

Width (cm)
*Note: Ex vivo bovine liver data may differ from human clinical outcomes. Depth (cm)

19G 1:1 Real Size


5W 10W 15W 20W 50W 90W
20 sec 1 min 2 min 3 min 5 min 6 min
0.48 x 0.47 0.61 x 0.74 0.95 x 0.99 1.22 x 1.32 1.65 x 1.95 2.20 x 2.90

4 mm 5 mm 7 mm 10 mm 15 mm 20 mm

18G 1:1 Real Size


5W 12W 25W 45W 110W 135W 165W 175W
20 sec 2 min 3 min 5 min 6 min 8 min 10 min 12 min
0.52 x 0.53 0.66 x 0.73 1.01 x 1.05 1.30 x 1.50 1.80 x 2.10 2.30 x 2.98 2.70 x 3.48 3.01 x 3.90

4 mm 5 mm 7 mm 10 mm 15 mm 20 mm 25 mm 30 mm

17G 1:1 Real Size 15G 1:1 Real Size


55W 115W 145W 170W 125W 140W 175W
5 min 6 min 8 min 10 min 8 min 10 min 12 min
1.60 x 1.63 2.00 x 2.30 2.43 x 3.00 2.85 x 3.50 2.50 x 2.50 2.90 x 3.10 3.40 x 3.50

10 mm 15 mm 20 mm 25 mm 15 mm 20 mm 25 mm

180W 195W 180W 195W


12 min 15 min 12 min 18 min
3.20 x 4.10 3.87 x 5.20 3.80 x 4.25 4.40 x 5.30

30 mm 40 mm 30 mm 40 mm

4 Ablation Zone Reference Guide Ablation Zone Reference Guide 5


*Note: Ex vivo bovine liver data may differ from human clinical outcomes.

Width (cm) x Depth (cm)

Width (cm)
Depth (cm)

16G 1:1 Real Size


130W 145W 165W 200W
10 min 12 min 15 min 18 min
2.76 x 2.58 3.61 x 3.60 4.43 x 4.85 4.74 x 5.84

15 mm 40 mm
20 mm 30 mm

• Saline infusion functionality for larger ablations

• All-in-one solution for aspiration, infusion, and ablation

*Injection of 0.5 cc / min of normal saline


17G 1:1 Real Size 15G 1:1 Real Size
70W 110W 135W 160W 150W 180W 200W
6 min 8 min 10 min 15 min 12min 18 min 24 min
2.10 x 2.07 2.40 x 2.30 3.17 x 3.50 4.21 x 4.81 3.95 x 3.88 4.83 x 4.95 5.25 x 6.09

15 mm
10 mm
20 mm 20 mm
30 mm 40 mm
30 mm

6 Ablation Zone Reference Guide Ablation Zone Reference Guide 7


*Note: Ex vivo bovine liver data may differ from human clinical outcomes.

Figure Width (cm) x Depth (cm)

Width (cm)
Depth (cm)
*Inter-electrode distance = 25 mm *Inter-electrode distance = 25 mm
Octopus RF Electrode Octopus RF Electrode
with VIVA multi RF Generator (Single Switching Mode) 1:1 Real Size with VIVA multi RF Generator (Dual Switching Mode) 1:1 Real Size
17G 17G
200W 200W 400W 400W
12 min 15 min 12 min 15 min
5.37 x 5.80 5.81 x 6.12 6.26 x 6.62 6.92 x 7.38

40 mm
40 mm

30 mm 30 mm

15G 15G
200W 200W 400W 200W
12 min 15 min 12 min 15 min
5.60 x 5.96 6.15 x 6.33 6.67 x 7.27 7.67 x 7.97

40 mm

30 mm 40 mm

30 mm

8 Ablation Zone Reference Guide Ablation Zone Reference Guide 9


Duo
Duo
*Note: Ex vivo bovine liver data may differ from human clinical outcomes.

Advanced Technology in RFA


STARmed Co., Ltd. is a young and fast-growing company from South Korea, specializing in
Width (cm) x Depth (cm) radiofrequency technology. Passionately working on the best possible RF treatment,
STARmed has collaborated with a number of Key Opinion Leaders in developing the most
innovative electrodes and leading-edge RF technology for a variety of clinical applications.
Width (cm)
Depth (cm)

Figure To find out more detail, please contact us at inquiry@STARmed4u.com.

*200W, 12 min, +10W/min after roll-off / Inter-electrode distance = 15 mm


Octopus RF Electrode-Duo
with VIVA multi RF Generator (Dual Switching Mode) 1:1 Real Size
17G 15G
50W 50W
12 min 12 min
4.95 x 5.75 5.54 x 6.05

30 mm 30 mm

10 Ablation Zone Reference Guide Ablation Zone Reference Guide 11


www.STARmed4u.com

Ablation Zone
Reference Guide

#B-1403, Daebang Triplaon, 158 Haneulmaeul-ro, Ilsandong-gu, Goyang-si, Gyeonggi-do, 10355 Korea
Tel: +82-31-816-3546 Fax: +82-31-816-4546 E-mail: inquiry@STARmed4u.com
www.STARmed4u.com ST-SD-AZ Rev.1 Mar, 2022
Ablation Zone Reference Guide 12