o 2. INSTALLING CytoSorb INTO A RRT- SYSTEM FOR USAGE POST - DIALYZER P. 2-4
• The arterial tubing system should be pre-filled with priming solution before connecting the Adsorber.
Fig. 1
• Anticoagulation needs to occur before treatment. aPTT target time is 60 - 80 seconds, ACT > 180% of
standard value. aPTT or ACT should be checked during therapy to ensure proper anticoagulation.
• Effectiveness of the removal of molecules is dependent on the blood flow rate through the Adsorber. The
blood flow rate should be between 200ml/min and 400ml/min. Lower blood flow rates may be compensated
for by extending the treatment time. The maximum running time of the Adsorber should not exceed 24 h.
• Materials:
• CytoSorb
• 2 Liters of isotone saline (priming-) solution (with Luer-Lock-connection)
• CytoSorb Priming Adapter 2
• CytoSorb Adapter 2
• Adsorber-holder
• tubing-clamps
• IV pole
Fig. 2
Saline
CytoSorb Prim-
ing adapter2
CytoSorb
Waste
Fig. 3
Fig. 4
dializer/hemofilter
Return line
bubble trap
CytoSorb
Dialyzer
CytoSorb
Adapter 2
Arterial line