© Unfors 5000104-A
February 2010
Base Unit version: 8221010-A Detector version: 8222010-A
8221021-A 8222011-A
8222012-A
8222013-A
8222014-A
Contents
The Unfors Solo. . . . . . . . . . . . . . . . 3 Warranty, Service and Support. . . 33
The Unfors Solo Detector options. . . . 3 Service. . . . . . . . . . . . . . . . . . . . . . . . . . 33
Easy to use . . . . . . . . . . . . . . . . . . . . . . . 4 Return Procedure for Service and War-
ranty . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Getting Started. . . . . . . . . . . . . . . . . 5
Contacts. . . . . . . . . . . . . . . . . . . . . . . . 34
Keys on the Unfors Solo Base Unit. . . . 5
Battery and Charger. . . . . . . . . . . . . . . . 6 Disposal. . . . . . . . . . . . . . . . . . . . . . 35
R/F Measurement. . . . . . . . . . . . . . . 7 Specifications . . . . . . . . . . . . . . . . . 36
Unfors Solo General. . . . . . . . . . . . . . . 36
CT Measurement. . . . . . . . . . . . . . 10
Unfors Solo R/F . . . . . . . . . . . . . . . . . . 36
How to measure CT dose. . . . . . . . . . 10
Unfors Solo R/F w/mAs. . . . . . . . . . . . 37
mA/mAs Measurement . . . . . . . . . 13 Unfors Solo RAD. . . . . . . . . . . . . . . . . 38
Unfors Solo Dose. . . . . . . . . . . . . . . . . 39
Measurement definitions. . . . . . . . 16
Unfors Solo DENT. . . . . . . . . . . . . . . . 40
Dose . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Unfors Solo CT. . . . . . . . . . . . . . . . . . . 41
Dose Rate. . . . . . . . . . . . . . . . . . . . . . . 18
Unfors Solo PC kit. . . . . . . . . . . . . . . . 42
kV/kVp. . . . . . . . . . . . . . . . . . . . . . . . . . 18
Unfors Solo HVL option. . . . . . . . . . . . 42
Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Pulse. . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
HVL (Half Value Layer) – optional. . . . 21
mA and mAs. . . . . . . . . . . . . . . . . . . . . 21
Unfors Solo PC kit. . . . . . . . . . . . 22
Unfors Xi View. . . . . . . . . . . . . . . . . . . 22
Transfer data to a computer. . . . . . . . 23
Installing Unfors Xi View. . . . . . . . . . . 23
Setup of Bluetooth communication. . . 23
Getting help . . . . . . . . . . . . . . . . . . . . . 24
Accessories. . . . . . . . . . . . . . . . . . 25
The Unfors Flexi stand. . . . . . . . . . . . . 25
The Unfors OPG holder . . . . . . . . . . . 25
The Unfors Cassette holder . . . . . . . . 26
Tips and Tricks. . . . . . . . . . . . . . . . 27
Arrangement of Measurement. . . . . . . 27
Sensor Position. . . . . . . . . . . . . . . . . . . 27
Reset Switch. . . . . . . . . . . . . . . . . . . . . 28
kVp on AMX. . . . . . . . . . . . . . . . . . . . . 28
Display Codes. . . . . . . . . . . . . . . . . . . . 28
Frequently Asked Questions. . . . . . . . 29
Troubleshooting. . . . . . . . . . . . . . . . . . 30
2
The Unfors Solo
The Unfors Solo consists of a base unit and a detector.
Communication between detector and base unit is purely digital, thereby
minimizing sensitivity to mechanical or electrical stress.
ON/OFF : Turns on the Unfors Solo and off when in SENSOR MENU.
EXIT: Exit to the previous menu.
STEP: A short press steps through available options.
SELECT: A long press selects an option.
Note!
The sensor fields must not be irradiated when “Zero adjust” is displayed or when
values on display are blinking. If this happens, EXIT to the SENSOR MENU
and re-enter the MEASURE MODE for automatic Zero adjust.
Getting Started
5
Battery and Charger
The Unfors Solo Base Unit has a standard chargeable NiMH, 9 V battery,
providing approximately 10 – 20 hours of operation (depending on selected
detector and whether Bluetooth is used). Since chargeable batteries have
a self discharge current, Unfors strongly recommends fully charging the
battery before first use. Battery voltage is displayed, together with setup
information, when the instrument is switched on.
Getting Started
6
R/F Measurement
Using R/F, RAD, Dose or DENT version.
For best accuracy, center the selected sensor field (R/F low, R/F high, RAD,
DENT low or DENT high) and position the long axis of the sensor field
perpendicular to the anode-cathode axis of the tube.
To adjust for temperature changes in the detector, the Unfors Solo auto-
matically zero adjusts at start up and after each exposure (the values on the
display are blinking during zero adjust). The Unfors Solo also monitors the
detector temperature between exposures to initiate a zero adjust if neces-
sary.
1. SENSOR MENU
R/F low Sensor for conventional low dose rate measurements
(R/F and Dose) < 1 mGy/s (7 R/min), normally after a phantom.
R/F high Sensor for conventional high dose rate measurements
(R/F and Dose) > 1 mGy/s (7 R/min), normally before a phantom.
RAD Sensor for conventional high dose rate measurements
> 0.1 mGy/s (0.7 R/min), normally before a phantom.
DENT low Sensor for low dose rate measurements < 1 mGy/s
(7 R/min), normally after a phantom or for applica-
tions with high (> 10 mm Al or equivalent) inherent
filtration.
DENT high Sensor for high dose rate measurements > 1 mGy/s
(7 R/min), normally before a phantom.
R/F Measurement
7
Press STEP to step between sensors and SELECT to select a sensor.
2. MEASURE MODE
After a zero adjust, the instrument is in MEASURE MODE . The displayed
values are updated after each exposure, or continuously after 4 seconds of
fluoroscopy.
kVp kVp (for the R/F high, RAD or DENT high sensor at
adequate signal levels) or kV (R/F low or DENT low).
(Not available on Unfors Solo Dose.)
Dose Gy (Air kerma, free in air) or R (Exposure)
Dose rate Gy or R per s, m or h
Time ms, s
HVL mm Al (only with optional HVL kit)
Pulses pulses
mAs mAs (requires Unfors Solo R/F with mAs)
mA mA (requires Unfors Solo R/F with mAs)
3. SETUP MENU
Press SELECT to enter the SETUP MENU (from MEASURE MODE ) and
STEP to step between setup parameters.
All values are stored in a non-volatile memory and are valid until manually
changed. At start up (after the battery status information), valid Trig delay,
kVp delay and Calc. delay are displayed in sequence.
(Factory settings in bold.)
R/F Measurement
8
Trig delay A delay in ms after the normal trig of the instrument. Utilize when
an unwanted part of an exposure, such as a pre-pulse, should be
excluded from the measurement. No measurements are per-
formed during the Trig delay. (0, 5, 10, 50, 100, 200, 500, 1000,
2000 ms)
Trig level Use Trig level to measure correct exposure time on waveforms
with slowly increasing output, such as with single phase dental.
The default setting is a low value, that depends on the selected
sensor; see Specifications. Can be set to 25, 50 or 75 % of the
peak value of the previous exposure. Use it with the same genera-
tor settings and FDD. Not selectable before the first exposure or
at low dose rates.
kVp delay Defined as the delay in ms, after the trig delay, but before the kVp
measurement (and waveform) window begins. Use a kVp delay
on machines with slow rising output, such as single phase intra
oral machines and fluoroscopy systems. (0, 2, 5, 10, 50, 150, 300,
1000, 1200, 1500, 1700, 2000 ms)
Calc. delay Defined as the dead time after end trig, before data is calculated.
Default is 0.5 s but it is recommended to use a longer delay when
measuring on pulsed fluoroscopy where the time between pulses
may exceed 0.5 s. (0.5, 2, 4, 6 s)
Dose unit Change the default units and/or use this feature to recalcu-
late a value to a different unit. (Gy or R per s, m or h, where 1
Gy = 114.1 R.)
Backlight Display lighting option. (off, 2, 5, 10 s)
Auto Power Off After 5, 20 or 60 minutes of inactivity.
Info Displays the serial number (S/N) and firmware version of the con-
nected detector and the Unfors Solo Base Unit and the calibration
date for the selected sensor. There is also a possibility to reset
the instrument settings to factory defaults.
kVp Mode “kV only” prevents the instrument from calculating kVp, which
otherwise occurs automatically when R/F high is selected and the
signal level is high enough (“kV/kVp”). Useful for example when
measuring near that signal level.
Not applicable for Unfors Solo Dose.
R/F Measurement
9
CT Measurement
The Unfors Solo CT Detector is an ion chamber with a 10 cm active length
and a built-in bias voltage. The center and the edges (+ 5 cm and - 5 cm) of
the active volume are marked on the phantom adapter.
CT Measurement
10
For positioning of the Unfors Solo CT Detector free in air, Unfors recom-
mends the use of the Unfors Flexi stand.
1. SENSOR MENU
CT Dose Ion chamber sensor
Press STEP to step between sensors and SELECT to select a sensor.
Note! When turning on the Base Unit with the Unfors Solo CT Detector con-
nected, “Stabilizing” will be displayed for a few seconds while the electronics are
stabilizing.
2. MEASURE MODE
After a zero adjust, the instrument is in MEASURE MODE . The displayed
value is updated after each exposure.
Dose Gy (Air kerma, free in air) or R (Exposure)
Dose Length Product Gycm or Rcm
Note! The Unfors Solo CT Detector can be used in two dose modes (dose or dose
length product (DLP)), selectable in the SETUP MENU. Since the Unfors Solo
CT chamber is 10.0 cm long, DLP readings (in Gy•cm or R•cm) will always be
exactly 10 times higher than dose readings.
3. SETUP MENU
Press SELECT to enter the SETUP MENU (from MEASURE MODE ) and
STEP to step between setup parameters.
CT Measurement
11
All values are stored in a non-volatile memory and are valid until manually
changed.
(Factory settings in bold.)
Dose unit Change the default units and/or use this feature to recalculate a
value to a different unit. (Dose: Gy, Gycm, R or Rcm, Tempera-
ture: °F or °C, Pressure: kPa or mmHg)
Backlight Display lighting option. (off, 2, 5, 10 s)
Auto Power Off After 5, 20, 60 minutes of inactivity.
Info Displays the serial number (S/N) and firmware version of the con-
nected detector and the Unfors Solo Base Unit and the calibration
date for the sensor. There is also a possibility to reset the instru-
ment settings to factory defaults.
Temperature and pressure values are shown only in the Unfors Xi View,
not on the Unfors Solo Base Unit display.
CT Measurement
12
mA/mAs Measurement
Using the Unfors Solo R/F with mAs.
Note! It is possible to measure mAs simultaneously with other measurements
with Unfors Solo R/F. The instrument will trig on radiation, not on the current in
the Unfors mAs cable. Any initial rush current will thus be excluded. The measured
mAs value automatically appears in MEASURE MODE on the display.
1. Turn off the X-ray generator.
2. Remove the jumper of the mA/mAs port in the generator.
3. Connect the black and red banana plugs of the Unfors mAs cable to the
mA/mAs port of the generator. (Note! The mA/mAs instrument is de-
pendent on the polarity. Using the wrong polarity will result in no trig.)
4. Switch on the X-ray generator.
5. Turn on the Unfors Solo, with no charger connected, and select mA/
mAs in the SENSOR MENU. Make an exposure and read the measure-
ments in the display.
Caution!
Do not forget to replace the mA/mAs jumper when removing the Unfors mAs
cable.
Caution!
Users of the Unfors mA/mAs meter must be aware of the potential damage to
generators and electrical human hazards in case of improper connection or failure
of any part of the meter circuit. This feature is intended for use only by personnel
authorized in performing calibration and repair of X-ray equipment.
mA/mAs Measurement
13
1. SENSOR MENU
mA/mAs Circuitry for invasive measurements of mA, mAs, time, pulses.
2. MEASURE MODE
After a zero adjust, the instrument is in MEASURE MODE . The displayed
values are updated after each exposure or continuously after 4 seconds of
fluoroscopy.
mAs mAs
mA mA
Time ms, s
Pulses pulses
3. SETUP MENU
Press SELECT to enter the SETUP MENU (from MEASURE MODE ) and
STEP to step between setup parameters.
When in the SETUP menu, the active selection will blink. All values are
stored in a non-volatile memory and are valid until manually changed. At
start up (after the battery status information), valid Trig delay and Calc.
delay are displayed in sequence.
(Factory settings in bold.)
Trig delay A delay in ms after the normal trig of the Unfors Solo. Utilize
when an unwanted part of an exposure, such as a pre-pulse,
should be excluded from the measurement. No measurements are
performed during the Trig delay. ( 0, 5, 10, 50, 100, 200, 500, 1000,
2000 ms)
Calc. delay Defined as the dead time after end trig, before data is calculated.
Default is 0.5 s but it is recommended to use a longer delay when
measuring on pulsed fluoroscopy where the time between pulses
may exceed 0.5 s. (0.5, 2, 4, 6 s)
Backlight Display lighting option. (off, 2, 5, 10 s)
Auto Power Off After 5, 20 or 60 minutes of inactivity.
mA/mAs Measurement
14
Info Displays the serial number (S/N) and firmware version of the
detector and the Unfors Solo Base Unit and the calibration date
for the selected sensor. There is also a possibility to reset the
instrument settings to factory defaults.
Note!
Avoid connection of the Unfors battery charger when measuring mA/mAs, as it
may create ground currents affecting the measurement. It is recommended to use
a laptop operating on battery or a Bluetooth connection when capturing wave-
forms or transferring data to a computer running Unfors Xi View. Connection of a
PC (mains supplied) via a serial cable to the Unfors Solo Base Unit may create
ground currents affecting the measurement.
mA/mAs Measurement
15
Measurement definitions
Figure 1 illustrates the Unfors Solo parameter definitions available on a
typical waveform. In most situations the default settings are suitable for ac-
curate measurements. In certain situations advanced users can change the
measuring definitions of the Unfors Solo by activating Trig delay and/or Trig
level settings. Because the measured parameters will be affected by the set-
tings of delays and trig levels, it is recommended to note the setup informa-
tion shown on the Base Unit display at power on.
Dose
The Unfors Solo R/F, RAD, Dose and DENT Detectors measure dose with
a multi-segment sensor, and the Active compensation feature automatically
corrects the displayed dose (and dose rate) for beam qualities with an HVL
of 1 – 14 mm Al (for example: a filtration of 45 mm Al at 140 kVp gives an
HVL of about 13 mm Al).
Measurement definitions
16
The CT meter is an ionization chamber with carbon fibre housing. It has
a flat energy dependence curve and automatic pressure and temperature
correction.
The dose is accumulated from the start trig point plus Trig delay until the
end trig.
Note!
The trig level setting (25, 50 or 75 %) affects the dose measurement.
Note!
The backscatter protection of the Unfors Solo sensor may cause differences in
readings compared to ion chambers. If both are placed directly against an object,
i.e. image receptor or lead apron, the Unfors Solo will read lower compared to an
ion chamber. However the Unfors Solo Detector will be reading True Input Dose
while the ion chamber is reading the input dose plus back scatter from the object
directly behind it. If using the Unfors Solo Detector and an ion chamber in free air
they should be comparable ± both meter tolerances.
Measurement definitions
17
Dose Rate
The Unfors Solo calculates dose rate as dose/exposure time if the exposure
time is less than 6 seconds.
After an exposure longer than 6 seconds, a dose rate value recorded ap-
proximately 2 seconds before end trig will be displayed.
kV/kVp
The Unfors Solo calculates kVp on the R/F high, RAD or DENT high sensor
if the signal level is high enough, otherwise kV average will be displayed. If
no kVp value is displayed, try decreasing the FDD or increasing mA.
kVp is calculated within the kVp measuring window. The kVp measuring
window begins after Trig delay and kVp delay and is approximately 160 ms
wide.
If the exposure time is less than 6 seconds, kV average is the integrated
average of the kV values during the whole measurement.
After an exposure longer than 6 seconds, a kV value recorded approximate-
ly 2 seconds before end trig will be displayed.
Measurement definitions
18
After the exposure, kV (or kVp if available) is displayed and a waveform is
transmitted to the Unfors Xi View. The Unfors Xi View also gives a possi-
bility to increase the length of the waveform window.
kVp matches kV on high frequency or DC like waveforms. If there is a ripple
on the waveform, a kV value will be lower than kVp.
Time
The Unfors Solo measures time from the start trig (may be adjusted with
the Trig level setting) until the signal falls below 25 % of the peak (50 or 75
% if adjusted with the Trig level setting). At low dose rates (about 1 % of
the max dose rate for the active sensor), the 25 % end level is changed to a
low level (about the lowest measurable dose rate for the active sensor).
Measurement definitions
19
If the radiation has a pulsed characteristic, the time is measured until the
last pulse ends. The dead time interval between pulses must, however, be
less than the Calc. Delay time (0.5, 2, 4 or 6 s). The R/F low or DENT
low sensor has an electrical bandwidth of 0.1 kHz (slower rising and fall-
ing slopes), causing the displayed exposure time to be a few milliseconds
longer than for the R/F high or DENT high sensor, which has a bandwidth
of 2.5 kHz.
Figure 5. 75 % trig level activated (figure also applies to 25 and 50 % trig level
settings). Please note that the trig level setting affects the dose measurement.
Pulse
The Unfors Solo counts pulses from 1 to 9999. The trigger to increment
the pulse counter occurs when the dose rate has a negative slope and the
amplitude falls below 25 % of the peak amplitude for the exposure. At peak
dose rates of approximately < 3 µGy/s (for R/F low) the signal to noise
Measurement definitions
20
ratio is too low to count pulses and “---” will be displayed. If possible, in-
crease the mA and/or kVp to increase the radiation output or decrease the
FDD.
Pulse is available for the R/F, RAD, Dose, DENT and mA/mAs Detectors.
mA and mAs
The Unfors Solo R/F Base Unit with mAs is capable of measuring mA, mAs
and time. If the peak mA is > 8 mA, it can also measure pulses, frame rate
and mAs per pulse.
When an exposure longer than 6 seconds has ended, a mA value recorded
approximately 2 s before end trig is displayed.
Measurement definitions
21
Unfors Solo PC kit
The optional Unfors Solo PC kit contains:
• Unfors Xi View software
• serial cable
• serial to usb converter
• optional: bluetooth adapter
Unfors Xi View
Unfors Xi View is suitable for
• viewing waveforms
• storing measurements
• transferring data to Excel or other software
Note! The latest version of Unfors Xi View is available on http://www.unfors.com
Getting help
A complete manual for Unfors Xi View can be found in the Help menu.
Accessories
25
4. Expose and align the fluorescent line with the preferred black line
(selected sensor) on the alignment lid. Soften the lighting if needed and
comply with your local regulations of radiation protection when adjust-
ing the place of the holder.
Put the detector in the holder and place the holder with the detector
perpendicular to the anode-cathode axis of the tube and expose. (You may
need to turn the holder to cover up the right area below the AEC cham-
bers.)
Accessories
26
Tips and Tricks
Arrangement of Measurement
Make sure nothing is in the X-ray field when measuring with the Unfors
Solo.
The fact that the dose is proportional to 1/d2 (where d is the FDD – Focus
to Detector Distance) may be used when positioning the sensor.
Sensor Position
The solid state sensors in the Unfors Solo Detector are positioned approxi-
mately 7 mm beneath the detector surface. The position is marked with a
line on the aluminium housing. This position is used as a calibration refer-
ence point for dose.
If the detector hangs up when the battery is down and does not respond
even when a charger is connected, reset as described above and charge for
some minutes before turning the instrument on again.
kVp on AMX
When measuring on AMX machines with R/F high, RAD or DENT high,
keep a short distance to get kVp instead of kV.
Display Codes
Display Meaning
*** Measurement has started.
--- No value can be calculated.
< 1 ms Less than 1 ms valid samples.
LoSignal The dose rate is too low to calculate a value (increase
mA or decrease FDD).
HiSignal The dose rate is too high to calculate a value (increase
FDD or decrease mA).
LoRange Calculated kV is less than specified range.
HiRange Calculated kV has exceeded specified range.
t<delay The exposure time is not long enough to calculate a
value.
Ofl. Overflow, i.e. calculated value is numerically higher than
can be displayed.
Auto range out At least one of the measuring ranges is out of order.
of order
Zero adjust out Hardware offset levels may be out of range due to ex-
of range treme temperatures.
What is the difference between DENT high and DENT low, and how do I
use them?
DENT low is a sensor for low dose rate measurements, normally after a
phantom or for applications with high (> 10 mm Al or equivalent) inherent
filtration. It has a high sensitivity at the cost of a lower bandwidth (0.1 kHz)
and should be used at dose rates below approximately 1 mGy/s.
DENT high is optimized for higher signal levels and features a higher band-
width (2.5 kHz). It is typically used before a phantom, and is required for
measuring kVp (as opposed to kV).
I sometimes get readings which are far off expected when measuring with
Unfors Solo DENT on panoramic X-ray machines, why?
Most likely the detector’s sensor area (2 x 22 mm) is not fully aligned with,
or not completely centralized in the beam. Use the Unfors OPG holder
(see page 25 for instructions) which has a fluorescent screen cap, which will
confirm the position of the beam in relation to the sensor.
What happens if the mA/mAs cable is connected with the wrong polarity?
The Unfors Solo will not show any mA/mAs readings. As long as the mA/
mAs cable is securely connected to both generator and base unit (in any
polarity), there is no increased risk of damaging the generator or the instru-
ment.
Troubleshooting
When turned on, the Unfors Solo briefly shows “Battery down” and re-
fuses to start.
Please charge the battery, see chapter Getting Started.
When the Unfors Solo CT is turned on, the instrument displays “Offset out
of range” after displaying “Stabilizing...” for a long time.
This means that the ionization chamber current fails to reach a stable level
when the instrument is turned on. Please make sure that:
• the chamber is not irradiated during stabilization or zero adjust.
• you are not operating the instrument outside its specified temperature
range.
If the problem persists, your instrument might be in need of a repair. Please
contact Unfors Service department.
I get strange mA/mAs readings when I’m using the Unfors Solo’s AC adap-
tor or when my computer is connected to mains power.
Make sure that the instrument
• was connected to the generator during zero adjust.
• is operating on battery (no AC adaptor).
• is not connected to a grounded computer (use Bluetooth or a laptop
running on batteries with no Ethernet cable connected).
Service
Unfors Instruments offers you service which includes calibration, repair,
warranty and accessories. In case your instrument fails, our service staff
will assist in the expeditious repair and return of your product. Our normal
service handling time is 14 days.
The Unfors Service Program allows you to minimize the cost of your instru-
ment over its product life cycle and to keep the instrument in top function-
ing condition. Unfors Instruments offers an annual, full service repair and
calibration facility at our two service centers, one located in Sweden for
international customers and one in the United States for North and South
American customers.
The Unfors Service Program offers a 12 month extended warranty after
annual calibration and service. After 5 years from the date of purchase, the
extended warranty is no longer offered.
A participant of the Unfors Service Program is obligated to return the
instrument annually, with a 12 month cycle (± 1 month) based on the origi-
nal delivery date. Failure to do so will void any part of the Unfors Service
Program.
Contacts
Do not hesitate to contact the Unfors Help desk.
For support issues:
International: +46 31 719 97 80
support@unfors.se
Germany: +49 731 175 492–20
support@unfors.de
North and South America: +1 860 355–2588
support@unfors.com
For service issues:
International: +46 31 719 97 10
Fax: +46 31 910 950
service@unfors.se
North and South America: +1 860 355 2588
Fax: +1 860 350 7650
service@unfors.com
This symbol, found on the product or on its box, indicates that this product
must not be disposed of or discarded with normal household waste. All
electronic or electrical waste equipment must be sent to a recycling center
designated for such hazardous waste. Proper disposal of your electronic
and electrical waste equipment will help us to conserve our natural re-
sources. Moreover, proper recycling of electronic and electrical waste
equipment will help to ensure human safety and the health of the environ-
ment. For more information about electronic and electrical waste equip-
ment disposal, recovery and collection points, please contact your local city
government, your household waste disposal service, the place of purchase,
or the manufacturer of the equipment.
Disposal
35
Specifications
Unfors Solo General
EMC tested According to EN 61000-6-1:2001 and EN 61000-6-3:2001
Exposure needed One
Reset Automatic
Temp. range 15 – 35 °C (59 – 95 °F)
Detector cable 2 and 10 m (6.5 and 33 ft)
length
Patent Germany DE69430268.6-08, UK 0758522
Japan 3449721, Sweden 9302909-8
France 075822, USA 5761270
Size Base Unit 28 x 74 x 142 mm (1.1 x 2.9 x 5.6 in)
Weight 250 g (9 oz)
Power off Automatic after 5, 20 or 60 min of inactivity
Power source Rechargeable NiMH standard 9 V
Battery time 10 – 20 hours (depending on detector and if Bluetooth is used)
Read out Three row alphanumerical backlit display with four digits
numerical resolution
Specifications
36
Range 45 – 150 kV/kVp
Uncertainty 2%
(for total filtrations from 2.5 mm Al up to 15 mm Al or
equivalent, Active Compensation)
Sensitivity 0.4 mA, 40 kV, 40 cm (15.7 in), no added filtration
(R/F low)
Sensitivity 0.8 mA, 70 kV, 50 cm (19.7 in), no added filtration
(R/F high)
Exposure time
Range 1 ms – 999 s
Uncertainty 0.5 % or 0.2 ms
Pulse
Range 1 – 9999 pulses
Peak trig level > 3 µGy/s
(R/F low)
Peak trig level > 1 mGy/s
(R/F high)
(1) 45 mm Al added filtration at 145 kVp gives a HVL of ~13 mm Al.
(2) 1000 mGy/s up to 70 kVp, 400 mGy/s at 100 kVp, 250 mGy/s at 140 kVp.
Specifications
37
Unfors Solo RAD
Size Detector 12 x 22 x 117 mm (0.5 x 0.9 x 4.6 in)
Weight 50 g (2 oz)
Dose
Range 10 µGy – 9999 Gy (1 mR – 9999 R)
Trig level 100 µGy/s (0.7 R/min)
Uncertainty 5 % or ± 10 µGy (1 mR)
(40 – 150 kVp, HVL: 1.5 – 14 mm Al (1) , Active Compensation)
Dose rate
Range 20 µGy/s – 1000 mGy/s (2) (140 mR/min – 7000 R/min)
Min. peak trig level 100 µGy/s (0.7 R/min)
Uncertainty 5 % or ± 10 µGy/s (70 mR/min)
(40 – 150 kVp, HVL: 1.5 – 14 mm Al (1) , Active Compensation)
kV/kVp
Range 45 – 150 kV/kVp
Uncertainty 2%
(for total filtrations from 2.5 mm Al up to 15 mm Al or
equivalent, Active Compensation)
Sensitivity 0.8 mA, 70 kV, 50 cm (19.7 in), no added filtration
Exposure time
Range 1 ms – 999 s
Uncertainty 0.5 % or 0.2 ms
Pulse
Range 1 – 9999 pulses
Peak trig level > 1 mGy/s
(1) 45 mm Al added filtration at 145 kVp gives a HVL of ~13 mm Al.
(2) 1000 mGy/s up to 70 kVp, 400 mGy/s at 100 kVp, 250 mGy/s at 140 kVp.
Specifications
38
Unfors Solo Dose
Size Detector 12 x 22 x 117 mm (0.5 x 0.9 x 4.6 in)
Weight 50 g (2 oz)
Dose (R/F low)
Range 10 nGy – 9999 Gy
(1 µR – 9999 R)
Trig level 100 nGy/s (0.7 mR/min)
Uncertainty 5 % or ± 10 nGy (1 µR)
(40 – 150 kVp, HVL: 1.5 – 14 mm Al (1), Active Compensation)
Dose (R/F high)
Range 10 µGy – 9999 Gy (1 mR – 9999 R)
Trig level 100 µGy/s (0.7 R/min)
Uncertainty 5 % or ± 10 µGy (1 mR)
(40 – 150 kVp, HVL: 1.5 – 14 mm Al (1) , Active Compensation)
Dose rate (R/F low)
Range 10 nGy/s –1 mGy/s (70 µR/min – 7 R/min)
Min. peak trig level 100 nGy/s (0.7 mR/min)
Uncertainty 5%
or ± 10 nGy/s (70 µR/min)
(40 – 150 kVp, HVL: 1.5 – 14 mm Al (1) , Active Compensation)
Dose rate (R/F high)
Range 20 µGy/s – 1000 mGy/s (2) (140 mR/min – 7000 R/min)
Min. peak trig level 100 µGy/s (0.7 R/min)
Uncertainty 5%
or ± 10 µGy/s (70 mR/min)
(40 – 150 kVp, HVL: 1.5 – 14 mm Al (1) , Active Compensation)
Exposure time
Range 1 ms – 999 s
Uncertainty 0.5 % or 0.2 ms
Pulse
Range 1 – 9999 pulses
Peak trig level > 3 µGy/s
(R/F low)
Peak trig level > 1 mGy/s
(R/F high)
(1) 45 mm Al added filtration at 145 kVp gives a HVL of ~13 mm Al.
(2) 1000 mGy/s up to 70 kVp, 400 mGy/s at 100 kVp, 250 mGy/s at 140 kVp.
Specifications
39
Unfors Solo DENT
Size Detector 12 x 22 x 117 mm (0.5 x 0.9 x 4.6 in)
Weight 50 g (2 oz)
Dose (DENT low)
Range 10 nGy – 9999 Gy (1 µR – 9999 R)
Trig level 1 µGy/s (7 mR/min)
Uncertainty 5 % or ± 10 nGy (1 µR)
(40 – 150 kVp, HVL: 1.5 – 14 mm Al (1), Active Compensation)
Dose (DENT high)
Range 10 µGy – 9999 Gy (1 mR – 9999 R)
Trig level 100 µGy/s (0.7 R/min)
Uncertainty 5 % or ± 10 µGy (1 mR)
(40 – 150 kVp, HVL: 1.5 – 14 mm Al (1) , Active Compensation)
Dose rate (DENT low)
Range 10 nGy/s –1 mGy/s (70 µR/min – 7 R/min)
Min. peak trig level 1 µGy/s (7 mR/min)
Uncertainty 5 % or ± 10 nGy/s (70 µR/min)
(40 – 150 kVp, HVL: 1.5 – 14 mm Al (1) , Active Compensation)
Dose rate (DENT high)
Range 20 µGy/s – 1000 mGy/s (2) (140 mR/min – 7000 R/min)
Min. peak trig level 100 µGy/s (0.7 R/min)
Uncertainty 5 % or ± 10 µGy/s (70 mR/min)
(40 – 150 kVp, HVL: 1.5 – 14 mm Al (1) , Active Compensation)
kV/kVp
Range 45 – 150 kV/kVp
Uncertainty 2%
(for total filtrations from 2.5 mm Al up to 1 mm Cu or
equivalent, Active Compensation)
Sensitivity 0.4 mA, 40 kV, 40 cm (15.7 in), no added filtration
(DENT low)
Sensitivity 0.8 mA, 70 kV, 50 cm (19.7 in), no added filtration
(DENT high)
Exposure time
Range 1 ms – 999 s
Uncertainty 0.5 % or 0.2 ms
Pulse
Range 1 – 9999 pulses
Peak trig level > 3 µGy/s
(DENT low)
Peak trig level > 1 mGy/s
(DENT high)
(1) 45 mm Al added filtration at 145 kVp gives a HVL of ~13 mm Al.
(2) 1000 mGy/s up to 70 kVp, 400 mGy/s at 100 kVp, 250 mGy/s at 140 kVp.
Specifications
40
Unfors Solo CT
Size detector 200 x 20 x 12 mm (7.9 x 0.8 x 0.5 in)
Size Ø detector 7.5 mm (0.30 in)
Size Ø phantom 12.5 mm (0.49 in)
adapter
Effective length 100 mm (3.94 in)
Weight 50 g (1.75 oz)
Range 10 µGy – 9999 Gy (1 mR – 9999 R)
20 µGy/s – 100 mGy/s (140 mR/min – 680 R/min)
Uncertainty 5 % (at reference point RQA9; 120 kV, 42.5 mm Al)
Energy dependence < 5 % (at 80 kV to 150 kV; RQR and RQA qualities)
Radial uniformity ±2%
Axial uniformity ± 3 %, within rated length
Influence of relative < 0.3 % (for RH < 80 %)
humidity
Uncertainty in 2%
temp. and pressure
correction
Pressure range 80.0 – 106.0 kPa
International Fulfills requirements in IEC 61674
Standard
Unfors Solo CT detector comes with a phantom adapter to fit a standard head and/
or body phantom.
Specifications
41
Unfors Solo PC kit
System Windows 98 (Second Edition, OSR2), Windows 2000,
requirements Windows XP, Windows Vista
File format XML
Communication RS-232 (115200/8-N-1) or Bluetooth
Bluetooth option
Connector 9-pin D-SUB, pre-configured for communication with Unfors
Xi View
Operating distance 100 m nominal (actual performance depends on environment
and receiving Bluetooth module)
Waveform
Bandwidth R/F low 0.1 kHz
Bandwidth 2.5 kHz
R/F high & RAD
Memory depth 1 200 ms
Specifications