You are on page 1of 6

Magnetic resonance imaging with optical preamplification and detection

A. Simonsen1 , J. D. Sanchez2 , S. A. Saarinen1,3 , J. H. Ardenkjær-Larsen2 , A. Schliesser1,3 , and E.


S. Polzik1
1
Niels Bohr Institute, University of Copenhagen
2
Center for Hyperpolarization In Magnetic Resonance (HYPERMAG), Technical University of Denmark
3
Center for Hybrid Quantum Networks (Hy-Q), University of Copenhagen

Laser Analysis
Abstract
MR scanner
Optical fiber
arXiv:1906.12260v1 [physics.med-ph] 19 Jun 2019

Magnetic resonance (MR) imaging relies on conventional


electronics that is increasingly challenged by the push for
stronger magnetic fields and higher channel count. These Coil
problems can be avoided by utilizing optical technologies. RF
bias
As a replacement for the standard low-noise preamplifier,
we have implemented a new transduction principle that Transducer Ca
pac
Pre
am
upconverts an MR signal to the optical domain and im- itor plif

Analysis
2 cm s ier
aged a phantom in a clinical 3 T scanner with signal-to-
+
noise comparable to classical induction detection. -
DC source
1.3 cm Standard detection circuit

1 Main Figure 1: Simplified schematic of electro-mechano-optical


transduction. The main photograph shows the MR scan-
Magnetic resonance (MR) imaging is a well-established
ner with the vacuum chamber inside. In the sketch of the
and non-invasive tool for routine clinical diagnostics and
scheme, an external frequency bias powers the detection
basic research. Its utility would benefit from enhanced
scheme to light and optical fibers routes light from laser to
signal-to-noise and acquisition speed—the key perfor-
transducer (inset lower left), and from transducer to anal-
mance parameters for MR imaging—and is the focus of
ysis. The coil can be any standard RF coil used for MR
considerable research efforts. Two trending developments
detection. In the standard MR detection scheme (lower
are higher magnetic fields and more detection coils in ar-
right), the RF coil is matched to a low-noise preamplifier.
rays. Yet most technical innovation relies on the same
A DC voltage source powers the amplifier and electrical
principle of matching a detection coil to a low-noise pream-
cables carry the amplified signal out of the scanner for
plifier in the receiver chain of the scanner [1] similar to
further processing.
the diagram in Fig. 1 (lower right), a common practice
that testifies to the extraordinary performance of con-
ventional electronics obtained through decades of opti-
mization. However, the electronics is increasingly chal- low-loss optical fiber which moves the receiver electronics
lenged by the ever-increasing magnetic field strengths in out and away from the coils and high magnetic field in
MR imaging [2]. And with the push for more coils, the the scanner. The idea is sketched in Fig. 1 where a signal
standard approach adds problems like cross-talk and elec- induced in an MR coil is combined with a radio-frequency
trical interference associated with the individual pream- (RF) bias, and together they vibrate a micro-mechanical
plifiers and cables that each coil needs, not to mention element which, in turn, modulates the amplitude of re-
that dense arrays demands a miniaturization of the cor- flected laser light. The transduction scheme can have a
responding circuits [3]. These issues may be alleviated by low noise-temperature [6], can target any signal frequency,
leveraging optical communication and sensing. In partic- is compatible with telecom optical wavelengths, and has
ular, optical fibers are insensitive to high magnetic fields, been successfully used to detect a nuclear magnetic res-
immune to electrical interference, and considered MR safe onance signal [7, 9]. And with proper circuit design and
[4]. Converting the MR signal into an optical modula- protective steps, we here show for the first time that it
tion has been done [5], and is even commercially available can be integrated into a commercial, clinical MR scanner
(Phillips dStream), but those approaches still use the stan- to acquire and reconstruct a complete MR image.
dard electrical preamplifier first and then convert the am- Fig. 1 shows a simplified version of our detection setup.
plified signal. In contrast, this work replaces the conven- A detailed diagram is in the supplementary information.
tional electrical preamplifier with a transducer [6–8] that Inside an MR imaging scanner (3 T GE MR750), a typ-
up-converts the MR signal onto an optical carrier. The ical MR RF coil formed a resonant circuit together with
signal can then be analyzed at the other end of a long and a mechanically compliant capacitor—the transducer. The

1
circuit was tuned to resonate at the Larmor precession width at half max) away from the resonance of the op-
frequency of the nuclear spins aligned along a strong mag- tical cavity. Laser light was routed to the sample by an
netic field. In our case, the setup specifically involved optical network and coupled directly from fiber into the
13
C nuclear spins and a 3 T magnetic field, i.e. a preces- cavity through focusing optics. The same fiber also col-
sion of 32 MHz. The scanner applied an RF pulse to ex- lected the reflection, and the optical network then routed
cite the nuclear precession—the transmit pulse. Further- the signal to a detector. To electrically connect the trans-
more, it applied gradient magnetic fields that enable selec- ducer and the circuit, we mounted and wirebonded the
tive excitation of atoms and encodes their phase through membrane chips to an 8-pin integrated circuit socket. Be-
spatially-dependent time evolution. As the spins precess, cause the transducer needs vacuum to operate, we placed
they create an oscillating magnetic flux through the coil the assembly inside a vacuum chamber made of glass-fiber
thus inducing a voltage in the RF coil. That voltage is that could hold a pressure below 1 × 10−3 mbar through-
usually electronically amplified and processed to gener- out the measurement series—see the method section. The
ate the MR image. Our new development consists of re- chip and the circuit were then connected with a short RF
placing the standard preamplifier with an optomechanical cable via a vacuum feedthrough. We protected the trans-
transducer that converts the MR signal to an amplitude ducer from electro-static discharges with a switch (see Fig.
modulation of laser light. The transducer was connected 1, lower left) that could short-circuit the capacitor pins.
directly in parallel to the coil and consisted of a freely sus- This was particularly important whenever we connected
pended membrane—equivalent to a tiny drum-skin—that the transducer to circuitry or transported it. In addi-
simultaneously constitutes one side of a capacitor and one tion, we took several protective steps beyond [6–9, 12] to
mirror in an optical cavity. Charges on the capacitor exert prevent the membrane from collapsing during the trans-
a force on the membrane and cause it to move, and this mit pulse: first, two crossed diodes were added in parallel
motion then changes the capacitance and thus also the to the transducer to short it if the induced voltage ex-
charges that affects the motion. Hence, the mechanical ceed their threshold. Such protection is a common way
and electrical resonance are parametrically coupled [10]. to protect standard preamplifiers, but did not sufficiently
As the MR signal induces charges on the capacitor, the protect the transducer on its own. In addition, it unfor-
resulting motion proportionally changes the reflection of tunately limits the RF bias that can be applied, but we
light from the optical cavity. This transduction is most could anyway get satisfactory transduction performance.
sensitive when the membrane is driven near resonance, Second, the detection resonance was detuned to minimize
which a signal at any frequency can do in unison with an the voltage and current induced in the coil by the transmit
RF bias if the beatnote between the signal and the bias pulse. This is another standard technique in MR imaging
is at the membrane’s resonance—see [11] for the detailed [1] and the scanner supplied the trigger signal that con-
theory. In our case, that meant the bias frequency had to trolled the detuning. For this purpose, the coil loop had
be the difference between the Larmor precession and the a segmenting capacitor and additional components that
membrane resonance, 32 MHz and 1.4 MHz respectively. together created a tank circuit when the trigger activated
Note that because the bias and signal are non-degenerate, a PIN diode’s forward-voltage threshold. The full circuit
interference and cross-talk between them can be removed diagram is described in the methods section and shown
through filtering and the mechanical response helps with in supplementary information. Third, the RF bias was
that. After the transduction, we downconverted the de- switched off by pulse-modulating the bias amplitude with
tected signal from optical modulation to the RF domain the same trigger that detuned the circuit. See the full
simply by measuring the reflected optical power with a setup diagram in supplementary information. Note that
detector. Although this output was frequency-shifted to the RF bias and spin-flip pulse overlap a little because the
the mechanical frequency, it corresponded to the voltage long (8 m) RF cables to and from the scanner add a time
induced in the coil by the MR sequence and was conse- delay (∼100 ns), but this was not a problem with every-
quently sufficient to reconstruct the MR image through thing else in place.
post-processing. However, we chose, out of convenience, to
leverage the scanner’s dedicated imaging software and fed In its first implementation, the transduction scheme had
problems with excess noise which we addressed in the fol-
the transduced signal back to the scanner’s receiver chan-
lowing ways: first, the RF bias was filtered twice to re-
nel. That required the signal to be frequency-shifted back
to the signal frequency expected by the scanner, which weduce its sideband noise at the detection frequency. Once
did by mixing the transducer signal with the RF bias (seethrough an external filter and then through a filter built
the methods section and figure in supplementary informa- into the detection circuit. Such sideband noise can limit
tion). sensitivity [7, 9]. Second, there was considerable added
noise when flexible cables connected the setup inside the
The transducer’s membrane was circular and made out scanner room with the setup outside. We believe the poor
of alumina and aluminum under tensile stress attached shielding of the cables allowed ambient noise to leak into
on top of a partially reflective mirror. See the methods the setup. Using instead semi-rigid cables with better
section and [8] for details. Because aluminum reflects, shielding reduced the noise significantly. Third, electrical
the membrane and mirror formed an optical cavity with noise at the membrane-frequency could drive the mechan-
a length determined by the cleanroom fabrication proce- ical motion because our sample had trapped charges [8].
dure. We designed the stack such that the fixed wave- We cancelled these out with a DC bias on the transducer
length of our laser was about one cavity linewidth (half which gave significant reduction in the overall transducer

2
noise. With those noise-eliminating steps in place, we ob- brane [9]. Bandwidth can be increased by using multiple
tained the data shown in Fig. 2 with the optomechanical mechanical modes [12]. Improvements of the optical cav-
transduction. ity will also benefit performance. Incidentally, our sample
had a suboptimal cavity length for the specific bias volt-
We imaged a phantom (Fig. 2a) consisting of a bottle
age, so we can expect to reach better sensitivity using
filled with ethylene glycol (purity 99.8 %; natural abun-
the present platform but with some fabrication optimiza-
dance, 1.1 %, 13 C; triplet with JCH of 142 Hz). The
tion. Note that we deliberately reduced signal-to-noise
recorded image (Fig. 2c) shows the spatial density of
13 slightly by upconverting the detected signal in order to
C atoms in a cross-section through the bottle. As we
send it back to the scanner. This step was not necessary,
expected, the image does not have a uniform signal-to-
but convenient. Our future work will aim to address the
noise throughout the volume because we used a surface coil
trapped charges in transducer fabrication in order to elim-
where the detected signal decays with the increased dis-
inate the DC bias, and to address the transducer’s need
tance to the coil (Fig. 2b). This effect is not corrected in
for vacuum with cleanroom packaging. The required vac-
the image processing. To each image voxel corresponds a
uum should be achievable with available techniques [13].
spectrum obtained from the spin-precession (Fig. 2d). For
Furthermore, while the RF bias is necessary to power the
ethylene glycol, that spectrum has a characteristic triplet
transduction scheme, supplying it with cables is not a ne-
as shown in the reference measurement (Fig. 2d), obtained
cessity. The bias could instead be delivered wirelessly [14],
with a commercial coil but still using the same phantom
potentially delivered by the same coil that generates the
and scanner. And the spectrum obtained with the trans-
transmit pulses, and collected by the RF coil. Such a
ducer (Fig. 2e) also shows the same triplet. Additionally,
scheme would eliminate all electrical connections to the
the two detection schemes have different background noise;
RF coil, leaving the optical fiber as the only physical con-
for the standard electronic amplifier the noise is flat in a
nection to the circuitry. This would enable a great increase
broad window, but for the transducer it has a distinctive
in the density of elements in MR array coils, avoiding all-
narrow Lorentzian lineshape plus an offset (Fig. 2e). We
together the performance and safety issues that stems from
believe the Lorentzian feature comes from the membrane’s
the large number of electrical connections present in MR
spectral response because the Lorentzian peak frequency
arrays based on current technology.
and linewidth changed with the bias power as we expect
from the electromechanical interaction [6, 11]. Addition-
ally, moving the bias frequency also shifted the detected References
signal with respect to the noise peak and scaled its am-
plitude following the peak shape. This is an expected be- 1. Edelstein, W., Hardy, C. & Mueller, O. Electronic
haviour if the MR signal drives the membrane’s motion decoupling of surface-coil receivers for NMR imag-
and therefore further corroborates that the spectral shape ing and spectroscopy. Journal of Magnetic Resonance
of the noise is related to the mechanical motion. Further- 67, 156–161. issn: 0022-2364 (Mar. 1986).
more, the transducer’s flat noise background scales with
2. Sobol, W. T. Recent advances in MRI technology:
optical power which suggests that it originates from am-
Implications for image quality and patient safety.
plitude noise of the laser, most likely due to shot-noise
Saudi Journal of Ophthalmology 26, 393–399. issn:
of light. Finally, we have not found any other spectral
1319-4534 (Oct. 2012).
feature in the circuit response, nor any spurious noise in
the setup, that is consistent with the observed Lorentzian 3. Wiggins, G. C. et al. 96-Channel receive-only head
peak. coil for 3 Tesla: design optimization and evaluation.
Magnetic resonance in medicine 62, 754–62. issn:
In summary, we have implemented MR imaging with di- 1522-2594 (Sept. 2009).
rect optical detection and amplification of the MR induc-
tion signal thus bringing the technical benefits of optical 4. Taffoni, F., Formica, D., Saccomandi, P., Di Pino,
signal processing to the receiver chain of an MR scanner. G. & Schena, E. Optical fiber-based MR-compatible
To the best of our knowledge, this work represents the sensors for medical applications: an overview. Sen-
first successful implementation of direct electro-mechano- sors (Basel) 13, 14105–14120. issn: 1424-8220 (Oct.
optical transduction in an MR system. We clearly see the 2013).
expected image and spectrum for our phantom and coil, 5. Memis, O. G., Eryaman, Y., Aytur, O. & Atalar, E.
and a noise-background compatible with the mechanical Miniaturized Fiber-Optic Transmission System for
motion’s Lorentzian linewidth plus an offset from optical MRI Signals. Magnetic Resonance in Medicind, 165–
noise. The transducer’s spectrum shows a signal-to-noise 173 (2008).
comparable with a commercial system, although still not 6. Bagci, T. et al. Optical detection of radio waves
as high, and a transduction bandwidth that is narrow com- through a nanomechanical transducer. Nature 507,
pared to standard preamplifiers. However, both features 81–85. issn: 1476-4687 (Mar. 2014).
can be improved greatly with straightforward improve-
ments such as increasing the RF bias amplitude, reducing 7. Takeda, K. et al. Electro-mechano-optical detection
the capacitance in parallel with the transducer, and espe- of nuclear magnetic resonance. Optica 5, 152–158.
cially by decreasing the membrane-capacitor gap [7]. The issn: 2334-2536 (Feb. 2017).
noise can be reduced by cooling the system, increasing
the mechanical quality factor, and using a lighter mem-

3
a d
Ethylene glycole

Coil

Normalized
sensitivity
b

6 Distance (cm)

c 1

e
0.8

0.6
6 cm

0.4

0.2

6 cm Frequency offset (kHz)

Figure 2: MR image data obtained with the electro-mechano-optical transduction a, Sketch of the coil and phantom
setup. The image plane intersects the center of the coil and is parallel to the bottle circumference. b, Simulated
sensitivity versus distance to the coil. c, Surface plot showing the normalized peak amplitude of the detected signal.
The dashed line indicates where the plane of the coil intersects the image. d, Spectrum of the MR signal measured
with classical induction detection. e, Same spectrum measured with the transducer. Both spectra shows data with
and without an induced MR signal, normalized to the signal-peak. With a signal, the spectrum shows the highest
peak among the image voxels; without, the spectrum shows the average over the entire image and is equivalent to the
noise background of detection.

4
8. Simonsen, A. et al. Sensitive optomechanical trans- 2 Methods
duction of electric and magnetic signals to the opti-
cal domain. Optics Express 27, 18561–18578. issn: 2.1 Transducer fabrication
1094-4087 (June 2019).
We fabricated the transducer in a cleanroom using stan-
9. Tominaga, Y., Nagasaka, K., Usami, K. & Takeda, dard techniques—see [8] for a step-by-step account. The
K. Studies on NMR-signal up-conversion from radio- starting substrate was a fused silica wafer, 100 mm in di-
frequency to optical regimes using a lightweight ameter, with a partially reflective dielectric mirror on top.
nanomembrane transducer. Journal of Magnetic Res- Alumina layers both protected the mirror and supported
onance 298, 6–15. issn: 10907807 (2018). the membrane while aluminum defined the top and bottom
10. Dougherty, W. M., Bruland, K. J., Garbini, J. L. & electrodes of the membrane-capacitor. The membrane
Sidles, J. A. Detection of AC magnetic signals by consisted of ∼70 nm alumina and ∼90 nm aluminum. We
parametric mode coupling in a mechanical oscillator. added tensile stress to the aluminum by annealing the
Measurement Science and Technology 7, 1733–1739. wafers before releasing the membrane. Alumina got its
issn: 0957-0233 (Dec. 1996). tensile stress during deposition. A sacrificial silicon ni-
11. Zeuthen, E., Schliesser, A., Taylor, J. M. & Sørensen, tride layer was between the electrodes and it had a thick-
A. S. Electro-optomechanical equivalent circuits for ness designed to the desired cavity length. The layer also
quantum transduction. Physical Review Applied 10, determines the gap between the capacitor electrodes. At
044036. issn: 03029743 (Oct. 2017). the end of the fabrication, the layer is etched away to re-
lease the membrane.
12. Haghighi, I. M. et al. Sensitivity-bandwidth limit in a The bottom electrode contained a hole, aligned to the
multimode optoelectromechanical transducer. Physi- center of the membrane, where light could pass through.
cal Review Applied 9, 034031. issn: 2331-7019 (Mar. Laser light was coupled into the cavity through this hole,
2018). using a gradient refractive-index lens attached to the back-
13. Rushton, J. A., Aldous, M. & Himsworth, M. D. side of the transducer chip. We achieved optical alignment
Contributed Review: The feasibility of a fully minia- by first centering the membrane with respect to a sili-
turized magneto-optical trap for portable ultracold con chip attached to the backside of the transducer chip.
quantum technology. Review of Scientific Instru- The silicon die had a large, square, and centered hole that
ments 85, 121501. issn: 0034-6748 (Dec. 2014). guided the lens and ensured alignment to the cavity. Light
14. Menke, T. et al. Reconfigurable re-entrant cavity for was supplied through a fiber-pigtail terminated in a glass
wireless coupling to an electro-optomechanical de- ferrule, with the ferrule and lens aligned by a glass tube
vice. Review of Scientific Instruments 88, 094701. that snugly fitted both. By changing the distance between
issn: 0034-6748 (Sept. 2017). ferrule and lens, we maximized the light delivered by the
fiber, focused through the lens, reflected from the sam-
15. Rispoli, J. V. et al. Trap Design and Construction for ple, and collected back into the input fiber. Finally, all
High-Power Multinuclear Magnetic Resonance Ex- components were fixed permanently with glue.
periments. Concepts in magnetic resonance. Part B,
Magnetic resonance engineering 46B, 162–168. issn:
1552-5031 (Oct. 2016). 2.2 Circuit
16. Abraham, E. R. I. & Cornell, E. A. Teflon The full circuit used to detect the imaging signal is shown
feedthrough for coupling optical fibers into ultra- in the supplementary information. The RF coil was a flat
high vacuum systems. Applied Optics 37, 1762–1763. spiral coil with four windings and an outer diameter of
issn: 0003-6935 (1998). 50 mm, wound with a 1.6 mm diameter silver wire. It had
an inductance estimated to be 490 nH from simulation.
A segmenting capacitor was inserted [15] which formed
a trap circuit together with parallel circuitry. The trap
detuned the detection coil when activated by a PIN diode,
i.e. when the diode’s forward-bias threshold was exceeded
by transmit trigger voltage. The main circuit board also
included a bandpass filter at the RF bias frequency. Its
purpose was to filter the RF bias’ sideband noise at the
MR signal frequency and prevent the RF bias from loading
the detection resonance. Notably, the filter was designed
to allow a DC offset on the bias to pass through.

2.3 MR setup
Supplementary information shows the full setup diagram
explained here. We mounted the RF coil and detection
circuit right outside a cryostat made from glass-fiber. Al-
though the cryostat can cool the transducer and circuit,

5
we only used it as a vacuum chamber. It could maintain Supplementary information
vacuum below 1 × 10−3 mbar up to five hours after discon-
necting from its pump because it contained two molecular
sieves (activated charcoal and sodium aluminum silicate), Bias-tee Semi-rigid Bias-tee
Scanner cable (7 m) Modulation input
both cooled to liquid nitrogen temperature (77 K). When receiver
mounting the transducer in the cryostat, we tried to align RF bias
Vacuum sd ucer Mixer
the membrane perpendicular to the main magnetic field. Transmit Tran Detector
50/50
Without this alignment, the mechanical linewidth broad- trigger
splitter
ens, likely due to the Lorentz force on the charges in the 90/10 splitter Optical
membrane. Laser light came to the transducer through a Detuning fibers Laser DC bias
Filter Amplifier
custom fiber-feedthrough [16] for the cryostat, an optical +

-
90/10 splitter, and a long single-mode fiber connected di- Semi-rigid
rectly to the fiber-coupled 1064 nm laser outside the scan- Coil cable (7 m) Bias-tee
Circuit board Filter
ner room. The splitter only routed 10 % of input light Inside scanner
to the chip and dumped the remaining power. Light re-
flected from the transducer’s cavity went back into the Figure 3: Detail setup schematic.
same splitter which then distributed 90 % into another
long single-mode fiber that lead back outside and con-
nected to a custom-built detector.
Inside the scanner room, we connected the trigger sig- Trigger signal input
To transducer
nal directly to the detuning trap on the circuit and also
sent the trigger outside to modulate the RF bias. That Detuning Semi-rigid
same cable also carried the transduced signal back into trap circuit coaxial cable
the scanner. Two bias-tees handled the routing by fre-
~ 5 cm
quency discriminating the low-frequency trigger and the
Bandpass
high-frequency signal. A separate cable carried the RF filter
bias. Outside the scanner room, the RF bias had its pulse-
Bias input
modulation input connected to the trigger and its output
split 50/50. One part became the bias drive and we am-
plified it before sending it to the detection circuit through
and external filter. The other part became the local oscil-
Tuning capacitor and crossed diodes
lator in a mixer that frequency-shift the optically detected Segmented pick-up coil
signal back to the MR signal frequency. The output of this
mixer was the input of the scanner’s receiver channel. We Laser 390 nH 390 nH
added a DC bias onto the RF bias drive with another bias- 22 µH 560 pF 490 nH

tee, right after the first bandpass filter for the RF bias and 220 nH
90 pF
before the long cable going into the scanner room. Trigger 100 pF 67 pF 62 pF Bias
10 pF
22 µH 560 pF

Trap circuit Resonance circuit Bandpass filter

Figure 4: Photo (top) and detailed diagram (bottom) of


the detection circuit

You might also like