You are on page 1of 1

IDT ADM

Z-Axis Detector Segment Layout 16 Detectors x 24 Segments MUX

1.5 1.5 1.5 1.5 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 1.5 1.5 1.5 1.5

4 x 1.5mm 8 x .75mm C
8 x .75mm 4 x 1.5mm
L

Basic Slice Widths


16 x .75mm
1.5 1.5 1.5 1.5 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 1.5 1.5 1.5 1.5

C
L

16 x 1.5mm
1.5 1.5 1.5 1.5 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 1.5 1.5 1.5 1.5

C
L

8 x 3mm MUX
1.5 1.5 1.5 1.5 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 .75 1.5 1.5 1.5 1.5

C
L

672 x 24D in 42 ADM's

1 42
ADMC’s

Ribbon Cable
(ADMC to M/B)

Tmin=28C - Minimum Speed Tmin=28C - Minimum Speed


DMS Fan Speed Control Trange=10C - Variable Speed (max speed @38C) DMBL (ADM’s 1-21) DMBR (ADM’s 22-42) Trange=10C - Variable Speed (max speed @38C) DMS Fan Speed Control
Fan OFF @ 23C (Tmin -5C hysteresis) Fan OFF @ 23C (Tmin -5C hysteresis)

To RCOM
From Ref Det. LMB Temp Sensor Thermal Switch OFF @42C +/- 2.8C 672 Dets x 16 Rings = 10572 values per IP RMB Temp Sensor
ON @ 32C +/- 2.8C Temp data
Converter DMS Temp = 34C +/- 1C Reference Detector DMS Temp = 34C +/- 1C

800mbs
Control Lines
Fiber Optic
Transmitter Receiver IRS/ACQ Mx8000IDT/BrillianceCT
To ADM’s
Data output
DMS Layout Block Diagram

RCOM ADU
Philips Training Center
Control Signals (ACU + DOM)
Revision: Drawing No.
2/3/04 PEK

You might also like