REPORT OF (ON
No.0270-02-BPC/0003/E10-L11/P8/2022
CLIENT : PT. KOTA MINYAK INTERNUSA
JOB TITLE : CALIBRATION
NAME OF EQUIPMENT —:_ DIGITAL PRESSURE GAUGE
DATE OF CALIBRATION: FEBRUARY 26, 2022
LOCATION : LABORATORIUM KALIBRASI
PT BIRO KLASIFIKASI INDONESIA (PERSERO)
CABANG UTAMA KOMERSIL BALIKPAPAN
bki
PT. BIRO KLASIFIKASI INDONESIA (PERSERO)
COMMERCIAL MAIN BRANCH OF BALIKPAPAN
2022SUMMARY REPORT. PAGE iit
REPORT No. : 0270-02-BPC/0003/ REF. NO :
Ee E10-L11/P8/2022 DATED : MARCH 04, 2022
TITLE of REPORT CLIENT / SPONSOR OF PROJECT
CALIBRATION OF DIGITAL PRESSURE PT KOTA MINYAK INTERNUSA
GAUGE
CLIENT / SPONSOR OF PROJECT PT BIRO KLASIFIKASI INDONESIA
(PERSERO)
BALIKPAPAN COMMERCIAL MAIN BRANCH
SM OPERATION & MARKETING
PT KOTA MINYAK INTERNUSA
WORK CARRIED OUT BY
PT BIRO KLASIFIKASI INDONESIA
(PERSERO)
BALIKPAPAN COMMERCIAL MAIN
BRANCH
CUNDUCTED BY
ARDI ARIANTO
Based upon records of calibration there
are summarize as follows :
| Location ‘TYPE OF UNIT TOTAL UNIT | RESULT
a eae ean 4 (one) unit Certificate of Calioration
Batkpapan | ee er tsps 1 (one) unit | Certificate of Calibration
CIENT 2625pst i (one) unit | Certificate of CalibrationPwene
COVER
. SUMMARY REPORT
|. REPORT OF CALIBRATION
. CERTIFICATE OF CALIBRATION
4.1, PRESSURE GAUGE CAPACITY 0 - 40bar / 580psi
4.2, PRESSURE GAUGE CAPACITY 0 - 70bar / 1015psi
4.3. PRESSURE GAUGE CAPACITY 0 - 25MPa / 3625psi
. APPENDIX
5.1. COPY CERTIFICATE OF PERSONNEL
5.2. COPY ACREDITATION OF LABORATORY8482: PT. BIRO KLASIFIKASI INDONESIA (Persero)
Cabang Utama Komersil Balikpapan
1964 T OF CALIBRATION
Report No. : 0270-03-BPC/0003/E10-L11/P8/2022
THIS IS TO CERTIFY THAT, based on the request of PT. Kota Minyak Internusa,
Calibration Laboratory of PT. Biro Klasifikasi Indonesia (Persero) Balikpapan of Commercial
Main Branch Office has been carried out calibration for “Digital Pressure Gauge ” with the
result as follows :
A RAL INI TION
Principle : PT. Kota Minyak Internusa
Job Description 1 Calibration of Digital Pressure Gauge
Date of Calibration : February 26, 2022
Place of Calibration : Balikpapan
Number of Equipment 3 (three) Units
B. RESULT of CALIBRATION
1, Pressure Gauge cap. 0 ~ 40bar / 580psi
Nilai Standar Penunjukan Alat Koreksi
(psi) (osi) (psi) Ketidakpastian
Naik Turun Naik Turun Naik Torun
0 0 0,0 0,0 0,0 0,0
50 50 49,75 49,75 0,25 -0,25
100 100) 99,95 99,95 -0,05 -0,05
160 160 160,86 | 160,86 0,86 0,86
220 220 218,86 | 218,86 | -1,14 =1414
280 280 278,95 | 278,95 =1,05, “1,05 0,8 psi
340) 340 338,66 | 338,66 =1,34 =1,34
400) 400 399,85 | 399,85 “0,15 “0,15
460 460 | 458,85 | 458,85 “1,15 “1,15,
520 520 518,55 | 518,55 “1,45 “1,45,
580) 580, 579,25 | 579,25 “0,75 “0,75
2. Pressure Gauge cap. 0 - 70bar / 1015psi
Nilai Standar Penunjukan Alat Koreksi
(psi) (psi) (psi) Ketidakpastian
Naik Turun Naik Torun Naik Torun
0 0 0,0 0,0 0,0 0,0
90 90 89,85 80,85 0,15 -0,15
180 180 179,85 |" 179,85 “0,15 0,15
360) 360 360,95 | 360,95 0,95 0,95
450 450, 450,95 | 450,95 0,95 0,95
540 540 540,95 [540,95 0,95 0,95 1,4 psi
640 | 640 640,95 | 640,95 0,95 0,95
780 780, 779,95 | 779,95 -0,05) -0,05
920 920 920,25 | 920,25 0,25 0,25
950) 950) 955,48__| 955,48 5,48 5,48
995) 995 994,85 | 994,85 -0,15 -0,15B. 1 PT. BIRO KLASIFIKASI INDONESIA (Persero) —___
Cabang Utama Komersil Balikpapan
1964
3. Pressure Gauge cap. 0 ~ 25MPa / 3625psi
Nilai Standar Penunjukan Alat Koreksi
(psi) (psi) (psi) Ketidakpastian
Naik Turun Naik Turun Naik Turun
0 0 0,0 0,0 0,0 0.0
400) 400 400,95 | 400,95 0,95 0,95
800) ‘800 800,95 | 800,95 0,95 0,95
1200 1200 | 119995 |" 1199,95 | -0,05 0,05
1600 1600 | 1601,95 | 1601,95_|~_1,95 1,95
2000 2000 | 200,95 [201,95 _|_1,95 1,95 4,8 psi
2300 2300 | 2301,95 | 2301,95 1,95 1,95
2600 2600 | 2601,95 | 2601,95 1,95 1,95
3000 3000 | 299,95 | 2999,95 |" -0,05 -0,05
3300 3300__ [302,95 | 3302,95 [2,95 2,95
3600 3600 | 3601,95 | 3601,95 | 1,95 195
NOTE : Final result of calibration see on the certificate as attached.
Balikpapan, March 04, 2022
Head of Balikpapan Commercial
h Operator
Ardi Arianto {+PRESSURE GAUGE S/N. SDI2 5-L-3-1-020 |]
CAP. 0 — 40bar / 580psiBALIKPAPAN COMMERCIAL MAIN BRANCH
bki PT. BIRO KLASIFIKASI INDONESIA (Persero) Vi ‘KAN
JLMTHaryono No.8 Rng Road BALIKPAPAN ~76326 LK-254408
Phone: (462-542) 876642, 576643,
‘omal: oe@bidco.id
'so/tec 17025:2017
SERTIFIKAT KALIBRASI
iLIBRATION CERTIFICATE
Nomor —: 010-PG-LK254-BPC-22
Number
ALAT Hal 12
Equipment
1 Nama Pressure Gauge 4 Kapasitas 40bar / 580 psi
Name Capacity
2 Type/Model Digital / Gi92-001 5 Resolusi i 1 psi
‘iypecmoaet ‘Kesotutton
3 MerkiBuatan : Brothoterm 6 Nomor Seri SDI2.5-L-3-1-020
‘Manufacturer Dental Number
ILIK
Owner
1 Nama PT Kota Minyak intermusa
Name
2 Alamat A. Jend, Suditman No.12 Kel. Damai - Balikpapan
Adeiress
STANDAR
Standard
1 Nama DWT - Budenberg / CPBS800
Name
2 Ketelusuran — : Hasilklibrai yang dilaporkan tertelusur ke Satu Pengulran SI melalui LK-057-IDN
Traccabiniy
TANGGALDITERIMA Ta 26 Februari 2022
Date ofaccepiance — 22.¥eb, 2021 ‘Late of calibration
KONDISL Suhu/Temp, Kelembaban
-Exwrronment condition of calioration 20°C 7%
LOKASIKALIBRASI Laboratorium Kalibrasi - PT BKI Cabang Utama Komersil Balikpapan
‘Location of calibration
‘ODA, SL Pressure Gauge in dikalibrasi dengan menggunakan alt Standar Dead Weight Tester
Calibration method dengan nomor sei CPBS800, ketertuusaren ke Satan Laernasinal (SI) mela
Laboratorium Kalibrasi Global Quality Indonesia LK-0S7-IDN
‘Metode kerja yang digunakan sesuai dengan Prosedur Mutu Laboratorium Kalibrasi
PT Bito Klasifikasi indonesia (Persero) No. : WI-LAB-18 issued 04
ACUAN BS/EN 837-1 : 1998, ISOMEC Guide 98-3:2008, DKD-R 6-1:2014
‘Kejerence
LD, "ASTIAN. RAST it
‘Kesulf of calbration and uncertainty of caloranon (attached
D (04 Maret 2022
‘Date of sssuebki PT. BIRO KLASIFIKAS! INDONESIA (Persero) KAN
BALIKPAPAN COMMERCIAL MAIN BRANCH
JL MT Haryono No. Ring Road BALIKPAPAN —76214
Phone: (462-542) 875582, 876643,
mal: boc@bkicoid
ISO/IEC 170252017
LAMPIRAN SERTIFIKAT KALIBRASI
Attachment of calibration certificate
Hal22
Noor 010-PG-LK254-BPC-22 Tanggal 04 Maret 2022
Number Date
1. Hasil Kalibrasi
osull of Calioraton
‘Nilai Standar Penunjukan Alat Koreksi Ketidakpastian
(psi) (psi) (psi) | Uncertainty *)
Naik | Turun Naik Turun Up_| Down (psi)
0 0 0,00 0,00 0,00 | 0,00
50 50 49,75 49,75 0.25 | -0,25
100 | 100 99,95 99.95 -0,05 | -0,05
160 | 160 160,86 160,86 0,86 | 0,86
220 | 220 218,86 218,86 Aid | 114
280 | 280 278,95 278,95 -1,05 | -1,05 08
340 | 340 338,66 338,66 -134 | 134
400 | 400 399,85 399,85 -0,15 | -015
460 | 460 458,85 458,85 LAs | -115
520 | 520 518,55 518,55 “145 | -1,45
580 | seo 519,25 579,25 0,75 | 0,75
‘Note: *) Keiidakpastian yang dilaporkan adalah Ketdakpastian bentangan dengan tngkat Kepercayaan 95% dan
faktor cakupan ke2
DILARANG mengutip/memperbanyak
SERTIFICAT ita tlh inaBALIKPAPAN COMMERCIAL MAIN BRANCH
JL.MIF Haryona No.8 ng Road BALIKPAPAN ~76124 UG264108
Phone: (462-542) 875642, 876543
emai: be@blico.d
bki PT, BIRO KLASIFIKASI INDONESIA (Persero) Vi ‘KAN
ISO/IEC 17025:2017
SERTIFIKAT KALIBRASI
CALIBRATION CERTIFICATE
Nomor — : 011-PG-LK254-BPC-22
Number
ALAT Hal 12
Equipment
1 Nama 1: Pressure Gauge 4 Kepesitas ‘TObar / 1015 psi
Name Capacity
2 TypeModel —: Digital /G192-5001 5 Resolusi 1 psi
‘DypetModet ‘xesolution
3 Merk/Buatan : Brothoterm 6 Nomor Seri. SDI2.5-L-3-1-070
Manufacturer ‘Serial Number
MILIK
Owner
1 Nama PT Kota Minyak Intemusa
Name
2 Alamat Il Jend, Sudirman No.12 Kel. Damai - Balikpapan
Address
STANDAR
‘Standard
1 Nama DWT - Budenberg / CPB5800
Name
2 Ketelusuran —: Hasilkalibrasi yang dilaporkan tertelusur ke Satuan Pengukuran SI melalui LK-057-IDN
Traceaoility
TANGGAL DI TANGGAL KALIBRAST 26 Februari 2022
Dale of acceprance 22Feb, 2022 Date of canoration
KONDISL UNGANKALIBRASI: Suhu/Temp. Kelembaban,
Environment condition of calibration 19 61.5%
LOKASIKALIBRASI Laboratorium Kalibrasi - PT BKI Cabang Utama Komersil Balikpapan
‘Locanon of cauibration
METODA KALIBRASL ‘Pressure Gauge ini dikalibrasi dengan menggunakan alat Standar Dead Weight Tester
Calibration method {dengan nomor seri CPBS800, ketertlusuran ke Satuan interasional (SI) melalui
Laborstorium Kalibrasi Giobal Quality Indonesia LK-057-IDN
‘Metode kerja yang digunakan sesuai dengan Prosedur Mutu Laboratorium Kalibresi
PT Biro Klasifikasi indonesia (Persero) No. : WL-LAB-18 issucd 04
ACUAN ‘BS/EN 837-1 : 1998, ISOMEC Guide 98-3:2008, DKD-R 6-1:2014
Kefjerence
RASI DA KALI ‘Clerlampir)
‘Result of calibrattion and uncertainty of calibration (Attached)
DITERBITKAN TANGGAL: 04 Maret 2022
Date of rssuebki PT. BIRO KLASIFIKASI INDONESIA (Persero) VKAN
BALIKPAPAN COMMERCIAL MAIN BRANCH
LMF Haryono No. ing Road BALIKPRPAN ~76114 UC25HI0N
Phone: (+62:582) #76082, 875643
‘email: oobi coig
ISO/IEC 17025:2017
LAMPIRAN SERTIFIKAT KALIBRASI
Attachment of calibration certificate
Hal 2/2
‘Nomor 011-PG-LK254-BPC-22 ‘Tangeal 04 Maret 2022
‘Number bate
1. Hail Kalibrasi
Result of Cabioraton
Nilai Standar Penunjukan Alat Koreksi__| Ketidakpastian
(psi) (psi) si) J Uncertainty *)
Naik _|Turun Naik ‘Turun Up_| Down (psi)
0 0 0,00 0,00 0,00 | 0,00
90 Ei) 89,85 89,85 0,15 | -0,15
180, 180 179,85 179,85 0,15 | -0,15
360 360 360,95 360,95 0,95 0,95
450 450 450,95 450,95 0,95 | 0,95
540 540 540,95 540,95 0,95 0,95 14
640 640 640,95 640,95 0,95 0,95
780 780 779,95 779,95 0,05 | -0,05
920 | 920 920,25 920,25 0,25 | 0,25
950 | 950 955,48 955,48 5,48 | 548
995 | 995 994,85 994,85 0,15 | 0,15
Note: *) Ketidakpastian yang dilsporkan adalah Ketidakpastian bentangan deogan tingkat Kepercayaan 95% dan
faktor cakupan k=2| PRESSURE GAUGE S/N. 2002 254056204
CAP. 0 — 25MPa / 3625psiPT. BIRO KLASIFIKASI INDONESIA (Persero) VKAN
b
BALIKPAPAN COMMERCIAL MAIN BRANCH —
JI. MT Haryeno No Ring Road BALKPAPAN ~76114 U25440N
Phone: (362-542) 876642, 576543
ema: boc@bKicoid
ISO/IEC 17025:2017
SERTIFIKAT KALIBRASI
‘CALIBRATION CERTIFICATE
Nomor : 012-PG-LK254-BPC-22
Number
ALAT Hal 12
Equipment
1 Name Pressure Gauge 4 Kapasitas 25MPa//3625 psi
Name Capacity
‘2 TypelModel Analog /G192-5001 5 Resolusi | psi
‘Typetmodet Kesolution
3 Merk/Buatan —: Brothoterm 6 Nomor Seri 2002 254056204
‘Marugacturer Serial Number
PEMILIK
Owner
1 Nama PT Kota Minyak Internusa
fame
2 Alamat IL Jend. Sudirman No.12 Kel, Damai - Balikpspan
Aaaress
STANDAR
‘Standard
1 Name : DWT -Budenberg / CPBS800
2 Ketelusuran = Hasitkalibrasi yang dlaporkan tertelusur ke Setuan Penguikwan SI melalui LK-057.1DN
draceaninty
‘TANGGAL DITERIMA TANGGAL KALIBRAST 26 Februari 2022
Dae of aecepiance 22 Keb. 2022 ‘Date af calibration
KONDISLLINGKUNGANKALIBRASI: Suhu/Temp. Kelembaban
‘Environment conaiiuon of calioration ZC 66%
LOKAS! KALIBRASI Laboratorium Kalibrasi - PT BKI Cabang Utama Komersil Balikpapan
‘Location of calrbranon
METODA KALIBRASL Pressure Gauge ini dikalibrasi dengan menggunekanalat Standar Dead Weight Tester
Calibration method éengan nomor seri CPBS80O, ketertulusuran ke SatuanInterasional (1) melahi
Laboratorium Kalibrasi Global Quality Indonesia LK-0S7-IDN
‘Metode kerja yang digunakan sesusi dengan Prosedur Mutu Laboratorium Kalibrai
PY Biro Kiasifkasi indonesia (Persero) No. : WILAB-S8 issued O&
ACUAN BS/EN 837-1 : 1998, ISO/IEC Guide 98-3:2008, DKD-R 6-1:2014
Kejerence
HASIL KALIBRASI DAN KETIDAKPASTIAN KALIBRAST (Teslampir)
‘Kesuit calibration and uncertainty oj calioranon (anacnea)
(04 Maret 2022
Date of issue
inn abortorian aoa PT Bo Kia! agonal (eres)
SERTIFICAT isa itt bun anéetangen plat yng berwenang dan cp abortariam alas PT re Kailas done (ers)bki PT. BIRO KLASIFIKAS! INDONESIA (Persero) VKAN
BALIKPAPAN COMMERCIAL MAIN BRANCH
LMT Haryone No. Ring Read BAUKPAPAN ~76116 Lk.254:0N
Phone: 462-542) 875582, 576643,
‘emai: bpe@iskicoid
's0/tEC 17025:2017
LAMPIRAN SERTIFIKAT KALIBRASI
~~ Attachment of calibration certificate
Hal 22
Nomar 012-PG-LK254-BPC-22 Tanggel 04 Maret 2022
‘umber Date
1. Hasil Kalibrasi
‘Result of Cattoranon
Nilai Standar Penunjukan Alat Koreksi | Ketidakpastian
(psi) (psi) (psi)__|/ Uncertainty *)
Naik [Turan Naik Turun Up_| Down (psi)
0 0 0,00 0,00 0,00 | 0,00
400 400 400,95 400,95 0,95 0,95
800 800 800,95 800,95 0,95 0,95
1.200 1.200 1199.95, 1,199,95 -0,05 | -0,05
1.600 1,600 1601,95 1.601,95 1,95 1,95
2.000 | 2.000 2001,95 2.001,95 195 | 1,95 48
2300 | 2.300 2301,95 2.301,95 195 | 1,95
2.600 | 2.600 2601,95 2.601,95 195 | 1,95
3.000 | 3.000 2999.95 2.999,95 -0,05 | -0,05
3.300 3.300 3302,95, 3.302,95 2,95 2,95
3.00 | 3,600 3601,95 3.601,95 195 | 1,95
"Note: *) Ketidakpastian yang dilaporkan adalah ketidakpastian bentangan dengan tingkat kepercayaan 9596 dan
faktor cakupan k=2CERTIFICATE OF PERSONNEL & ACREDITATION
CERTIFICATE OF LABORATORY1201 40BEAC @ TUN, TUEVard TW we ested ners. ents eenton eins le ao
CERTIFICATE
This is to certify that
Ardi Arianto
has attended the training on
Calibration Method
at PT Biro Klasifikasi Indonesia
The training covered the following subjects:
Calibration Work Instruction of Test Gauge, Pressure Gauge,
Ultrasonic Thickness, Hardness Tester
Estimation of Uncertainty in Calibration
Quality Assurance of Calibration Results
Training Date : 2016 July 25 - 27
Certificate Registration No, 1303212504/01
fs
(on behalf of TUV Rheinland Group
Jakarta/indonesia, 2017 January 26
es aie ZX TOvrheiniand®
Precisely Right.