You are on page 1of 7

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: D8488 − 22

Standard Test Method for


Determination of Hydrogen Sulfide (H2S) in Natural Gas by
Tunable Diode Laser Spectroscopy (TDLAS)1
This standard is issued under the fixed designation D8488; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope D5503 Practice for Natural Gas Sample-Handling and Con-


1.1 This test method is for the online determination of ditioning Systems for Pipeline Instrumentation (With-
hydrogen sulfide (H2S) in natural gas using tunable diode laser drawn 2017)3
absorption spectroscopy (TDLAS) analyzers also known as a D5504 Test Method for Determination of Sulfur Compounds
“TDL analyzers.” The particular wavelength for H2S measure- in Natural Gas and Gaseous Fuels by Gas Chromatogra-
ment varies by manufacturer, typically between 1000 and phy and Chemiluminescence
10 000 nm with an individual laser having a tunable range of D6228 Test Method for Determination of Sulfur Compounds
less than 10 nm. The H2S concentration ranges can be any- in Natural Gas and Gaseous Fuels by Gas Chromatogra-
where from 0-5 ppm(v) to 0-90 % by volume. phy and Flame Photometric Detection
D6968 Test Method for Simultaneous Measurement of Sul-
1.2 Units—The values stated in SI units are to be regarded fur Compounds and Minor Hydrocarbons in Natural Gas
as the standard. No other units of measurement are included in and Gaseous Fuels by Gas Chromatography and Atomic
this standard. TDLAS analyzers inherently output concentra- Emission Detection
tions in unitless molar ratios such as ppm(v). D7551 Test Method for Determination of Total Volatile
NOTE 1—Weight-per-volume units such as milligrams or grains of H2S Sulfur in Gaseous Hydrocarbons and Liquefied Petroleum
per cubic foot or cubic meter can be derived from ppm(v) at “standard Gases and Natural Gas by Ultraviolet Fluorescence
conditions” or standard temperature and pressure. D7833 Test Method for Determination of Hydrocarbons and
1.3 This standard does not purport to address all of the Non-Hydrocarbon Gases in Gaseous Mixtures by Gas
safety concerns, if any, associated with its use. It is the Chromatography
responsibility of the user of this standard to establish appro-
priate safety, health, and environmental practices and deter- 3. Terminology
mine the applicability of regulatory limitations prior to use. 3.1 Definitions—For definitions of general terms used in
1.4 This international standard was developed in accor- D03 Gaseous Fuels standards, refer to Terminology D4150.
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the 3.2 Definitions of Terms Specific to This Standard:
Development of International Standards, Guides and Recom- 3.2.1 absorption spectroscopy, n—refers to spectroscopic
mendations issued by the World Trade Organization Technical techniques that measure the absorption of electromagnetic
Barriers to Trade (TBT) Committee. radiation as a function of frequency or wavelength because of
its interaction with a sample.
2. Referenced Documents 3.2.2 adsorption, n—adhesion of molecules to a solid sur-
2.1 ASTM Standards:2 face forming a molecular or atomic film.
D1945 Test Method for Analysis of Natural Gas by Gas 3.2.3 chemometrics, n—field of science relating measure-
Chromatography ments made on a chemical system or process to the state of the
D4150 Terminology Relating to Gaseous Fuels system via application of mathematical or statistical methods.
3.2.4 desorption, n—phenomenon whereby a substance is
1
This test method is under the jurisdiction of ASTM Committee D03 on Gaseous released from a surface (the opposite of adsorption).
Fuels and is the direct responsibility of Subcommittee D03.12 on On-Line/At-Line 3.2.5 heat trace, n—ribbon-shaped tape that uses electrical
Analysis of Gaseous Fuels.
Current edition approved Nov. 1, 2022. Published November 2022. DOI: resistance or tubing carrying steam to generate heat.
10.1520/D8488-22.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
3
Standards volume information, refer to the standard’s Document Summary page on The last approved version of this historical standard is referenced on
the ASTM website. www.astm.org.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

1
D8488 − 22
3.2.5.1 Discussion—Heat trace tape is attached to sample laser temperature. The laser generates a near-infrared beam of
tubing and other sample-conditioning components to avoid light that passes through the cell window, which is typically
condensation and stabilize the temperature of the wetted reflected using a mirror (or mirrors) within the cell, and then
components and the gas stream. returns through the window and into a photodiode detector.
3.2.6 selectivity, n—refers to the extent to which a measure- The photodiode signal is used to determine how much light is
ment of concentration can be made in a mixture of various absorbed at specific wavelengths.
chemicals without significant influence from other components 4.3 Fig. 2 is a graph of typical regions in the near-infrared
in the mixture. spectrum where H2S will be absorbed. In the graph, the x-axis
3.2.7 tunable diode laser absorption spectroscopy (TDLAS), indicates the wavelength. The y-axis indicates the “absorption”
n—technique for measuring the concentration of a specific of light in which 1.0 (or 100 %) is the maximum. Where the
component such as hydrogen sulfide (H2S) in a gaseous sample absorption is more than 0, H2S is indicated. The vertical lines
by absorption spectrometry using tunable diode lasers. within the graph indicate the magnitude of absorption at
specific wavelengths. Each individual absorption line can be
3.3 Abbreviations:
potentially used for TDLAS H2S measurement. The actual
3.3.1 CO2—carbon dioxide
wavelength used will vary based on manufacturer, background
3.3.2 H2O—water composition, measurement specification requirements, and la-
3.3.3 H2S—hydrogen sulfide ser availability.
3.3.4 TDLAS—tunable diode laser absorption spectroscopy 4.4 The sensitivity of the measurement is determined by the
absorption as well as the length of the laser beam path (path
4. Summary of Test Method length) within the sample cell. The path length is fixed and can
4.1 A representative sample of the gas is extracted from a range from about 30 cm to 30 m depending on the measure-
process pipe or pipeline and is transferred by a sample ment range and the wavelength used. By optimizing the path
transport line through an appropriately designed sample- length and wavelength, linearity less than 0.1 % can be readily
handling system to the inlet of an analyzer. The sample shall be achieved. The TDLAS manufacturer must be consulted for
conditioned with a minimum, preferably negligible, impact on actual linearity specifications.
the H2S concentration. The gas flows continuously through the 4.5 This test method can be used as a guideline for produc-
analyzer and is vented to atmosphere, or a flare, or back to the ing H2S measurement data of a known and defensible quality
process stream depending on application and regulatory re- when using a TDLAS analyzer. A procedure is outlined below
quirements. for validating measurement integrity.
4.2 The gas sample stream flows through the measurement
portion of the analyzer. An overall diagram of a simple TDLAS 5. Significance and Use
system is shown in Fig. 1. A solid-state laser with a narrow 5.1 H2S measurements in natural gas are performed to
wavelength range is used as a light source. Electronics drive ensure concentrations satisfy gas purchase contract criteria and
the laser and a thermoelectric cooler precisely stabilizes the to prevent pipeline and associated component corrosion.

FIG. 1 Main Components of the TDLAS System

2
D8488 − 22

FIG. 2 H2S Absorption in the Near Infrared (NIR) Spectrum (Source: Pacific Northwest National Laboratory Northwest Infrared Database)

5.2 Using TDLAS for the measurement of H2S in natural scrubber) may be used. Since hundreds of possible wave-
gas enables a high degree of selectivity with minimal interfer- lengths are available in the near-infrared band for measuring
ence from common constituents in natural gas streams. The H2S, it is not practical to list the potential interferences. The
TDLAS analyzer can detect changes in concentration with a manufacturer should indicate the intended stream composition
relatively rapid response compared to other methods so that for the analyzer and the allowable variations in the stream to
operators may take swift action when designated H2S concen- the end user.
trations are exceeded.
6.2 Protection from liquid carryover such as from heavy
5.3 Primary applications covered in this test method are hydrocarbons, amines, and water in the sample lines is neces-
listed in 5.3.1 and 5.3.2. Each application may have differing sary to prevent liquid pooling in the cell or the sample system
requirements and methods for gas sampling. Additionally, components. Liquid carryover is especially a concern on the
different natural gas applications may require unique spectro- outlet of gas-sweetening processes that use aqueous solutions
scopic considerations. such as amines.
5.3.1 Raw natural gas is found in production, gathering
sites, and inlets to gas-processing plants characterized by 7. Apparatus
potentially high levels of water (H2O), carbon dioxide (CO2),
H2S, and heavy hydrocarbons. Gas-conditioning plants and 7.1 A TDLAS analyzer system includes the following sub-
skids are normally used to remove H2O, CO2, H2S, and other systems: (1) sample extraction, (2) sample transport, (3)
contaminants. sample-conditioning system, (4) TDLAS analyzer, and (5) vent
5.3.2 High-quality “sales gas” is found in transportation line.
pipelines, natural gas distribution (utilities), and natural gas 7.1.1 Sample Extraction Hardware—Sample extraction is
power plant inlets. The gas is characterized by a very high required to obtain a representative sample from the pipeline. To
percentage of methane (90 to 100 %) with small quantities of maximize the response speed, it is recommended to reduce the
other hydrocarbons and trace levels of contaminants. pressure at the sample point. To avoid condensation that may
occur from expanding the gas when it is depressurized (espe-
6. Interferences cially when the pipeline pressure is high), it is important to
6.1 TDLAS analyzers can be highly selective. They can understand the phase diagram of all the components in the gas
measure target component with very little interference from (for example, hydrocarbons, alcohols, and water). Use an
background composition with some limitations. There may be extraction probe and a regulator as shown in Fig. 3 mounted so
some interference from background components, for example, that the tip of the probe is in the center third of the pipe
at some wavelengths, methane, ethane, and carbon dioxide diameter. If the dew point of the gas is lower than the ambient
may absorb at the same wavelength as H2S. If interferences temperature after consideration for temperature reduction be-
exist at a particular wavelength, a different wavelength can be cause of gas expansion through the regulator (approximately
used and other techniques such as chemometrics, compensa- 3 °C per 6 bar), all sampling apparatus such as the probe and
tion algorithms, vacuum pumps (to separate the spectroscopy regulator may need to be heat traced or enclosed in a heated
peaks), or differential measurements (a technique using an H2S chamber, or both. According to Practice D5503, “vapor sample

3
D8488 − 22

FIG. 3 Sample Probe with Integrated Pressure Regulator (Left) and Sample Probe with No Regulator (External Pressure Regulator to be
Used) (Right)

must be kept at least 10 °C above the hydrocarbon dew point desorption affects. Typically, if heat tracing is required, a
temperature to prevent condensation of the sample.” sample panel will be installed in a heated building or a heated
7.1.2 Sample Transport Hardware—The sample transport enclosure. Components such as valves, regulators, and fittings
line carries the sample from the sample extraction point to the that are used in the sample extraction, sample transport, and
analyzer. The length of the sample transport tubing should be sample conditioning shall be designated as “instrument or
as short as possible. Heat trace is necessary if the environmen- analytical grade.” If the sample gas may contain an unsafe
tal temperature is close to the dew point of the sample gas. level of H2S, the sample system and the heated enclosure shall
Heat trace prevents liquid condensation and adsorption and have the appropriate purge ports and warning system to prevent
desorption of H2S from the walls of the tubing caused by accidental exposure during repair and maintenance.
ambient temperature changes (Fig. 4). When heat trace is used, 7.1.3.2 The sample vent line transports the sample from the
the entire length of tubing shall be heated and insulated with no analyzer to a safe location or back into the process stream. It
gaps. Electropolished tubing is recommended for use in ana-
should be an unrestricted line. Many users vent the analyzer to
lytical sample transport to reduce adsorption and desorption
a high point. If so, appropriate equipment may be used to
effects and optimize speed of response in the sample transport.
protect against lightning strikes at the vent. Also, a 180° bend
Silica-lined tubing may also be used.
with at least a 30 mm radius should be applied to the end of the
7.1.3 Sample-Conditioning System Hardware—The sample-
vent to minimize rain and wind intrusion. Additionally, a
conditioning system is typically installed immediately before
the analyzer and contains a regulator to perform a final pressure screen of some kind on the end should be used to discourage
drop, a means to control and measure flow, and filters to insects that might plug the vent. Never vent the analyzer inside
remove particulates and trap liquids. Additionally, the sample- a building or enclosure.
conditioning system may provide a means for bypassing the 7.1.4 Reference Standard Hardware—A binary gas mixture
analyzer to connect a reference gas for validating or challeng- (for example, H2S in methane) can be used to validate a
ing the analyzer and an outlet for venting the gas. Sample TDLAS analyzer. In Fig. 5, a simple setup is shown. The
system recommendations are described in detail in Practice analyzer may have a validation gas inlet built into the sample-
D5503. conditioning system, so the three-way ball valve is optional.
7.1.3.1 Similar to the sample transport line, the sample- All the components used must minimize the retention of the
conditioning system may require heat tracing to eliminate analyte of interest between the condensate or filter material and
condensation and reduce erratic readings caused by adsorption/ the effluent gas. Finer surface finish, electropolished stainless

FIG. 4 Heat-Trace Tubing with Self-Regulating Heat Tape Bundled with Insulation and Protective Jacket

4
D8488 − 22

FIG. 5 Example Setup for Introducing a Binary Gas into the H2S Analyzer

steel, elimination of dead legs, and passivation to prevent H2S 9.1.1 Compressed gas standard regulators, gas lines, and
losses, such as silica-lined stainless steel, should be considered. fittings must have a fine surface finish, be electropolished
stainless steel or silica lined, and be well passivated for
8. Hazards transport and delivery of H2S.
8.1 The process pipeline may contain very high-pressure 9.1.2 All compressed gas standards must be re-certified as
gas (tens of MPa). The sample from the process line shall be recommended by the manufacturer or as needed to ensure a
dropped close to atmospheric pressure before going into the documented and known accuracy.
TDLAS measurement cell. The recommended practice is to use
a 0 to 500 kPa pressure regulator set at 150 to 300 kPa at the 10. Calibration and Standardization
extraction point plus a 0 to 150 kPa regulator set at 70 kPa at
the analyzer. This allows for longer sample runs, better 10.1 TDLAS analyzers are factory calibrated by the manu-
pressure control, and protection against the failure of one facturer. The manufacturer may allow for reading offsets to be
regulator. applied for external calibration to a trusted reference. However,
it is generally not necessary to calibrate the analyzer in the
8.2 It is recommended to install a shutoff valve between the
field.
sample point and the analyzer so that the analyzer may be
isolated during maintenance and other similar events. 10.2 Sample stream composition must be determined ac-
8.3 Observe all safety regulations related to the use of cording to Test Methods D1945, D7833, or other suitable
equipment in hazardous area locations and in the presence of method for hydrocarbon content and Test Methods D5504,
combustible and poisonous gases. D6228, D6968, D7551, or other suitable method for sulfur gas
content and results submitted to the manufacturer prior to
9. Reagents and Materials calibration. Other analyses may be needed to determine con-
9.1 Compressed Gas Standards—Blended gaseous H2S centration of species other than hydrocarbons when they are
standards in nitrogen or methane base gas may be used per present on significant amounts. This compositional information
manufacturers recommendation. Care must be exercised in the better ensures that the analyzer will produce accurate measure-
use of compressed gas standards since they can introduce ments for the particular stream when installed in the field.
errors in measurement due to lack of uniformity in their 10.2.1 The stream composition must not change signifi-
manufacture or instability in their storage and use. Compressed cantly after the TDLAS analyzer has been installed or the
gas standards should be stable, NIST or similarly traceable, of accuracy of the measurement may drift outside data quality
high purity, and consumed prior to the expiration date. objectives for the analysis.

5
D8488 − 22
10.3 The procedure in Section 11 includes methods for Alternatively, a value close to the normal operating concentra-
validating the original factory calibration of a TDLAS ana- tion or the concentration at which critical decisions are made
lyzer. may be used.
11. Procedure 11.6 Validation Frequency—Validation should be performed
when the analyzer is first commissioned. After that, the initial
11.1 Connect the sample extraction, sample transport, and
frequency should be defined by company policy or regulatory
sample-conditioning equipment as described in Section 7.
requirement as applicable. As a standard recommendation, the
Observe the manufacturer’s installation recommendations.
analyzer should be validated quarterly or no less than once per
11.2 Connect the electrical systems (power and signals) per year.
the manufacturer’s recommendations. The method for taking
concentration readings may vary by manufacturer or user; in 11.7 Validation Blends—As explained in Section 6, the
most cases, the analyzer will have a display. Additionally, the TDLAS analyzer’s readings may be affected by the back-
analyzer may have a digital or analog interface. ground gas. For example, if the analyzer was calibrated at the
factory for H2S in methane, the analyzer may not be accurate
11.3 Start the analyzer according to the manufacturer’s when measuring H2S in pure carbon dioxide. For this reason, it
instructions. Observe recommendations regarding electronics is important to validate the analyzer using the appropriate
and sample system startup and allow for warm-up and purge background gas that the manufacturer prescribes based upon an
times. Follow the manufacturer’s recommendations, if any, analysis of the sample stream or a similar sample stream. Do
regarding external influences such as electromagnetic interfer- not attempt to validate the analyzer using an alternative
ence (for example, high-power radios), weather and tempera- background gas without manufacturer confirmation. In general,
ture conditions, or heavy vibration (for example, compressor this information should be provided by the manufacturer in the
pump). calibration certificate or the operator’s manual.
11.4 Analyzer Validation—Once the analyzer has been
11.8 Field Calibration—The analyzer may be adjusted to
initialized, the analyzer performance can be evaluated by
match a reference standard calibration gas, if necessary. Refer
comparing to a known reference. The TDL analyzer’s mea-
to the manufacturer’s instruction manual as procedures vary.
surement is inherently more reliable than a reference gas
because of the difficulty in handling, transporting, storing, and 11.9 Zero-check—Although contamination of H2S is not
sampling such mixtures. After a successful validation, it is typically encountered, an analyzer may be zero-checked with a
assumed that the analyzer’s original factory certification re- gas containing no H2S to ensure that it is reading zero and that
mains valid. Generally, it is not necessary to adjust the readings there is no H2S background. If the analyzer fails the zero-
of the analyzer to match the reference gas unless dictated by check, refer to the manufacturer’s troubleshooting guide.
policy or by the manufacturer. 12. Precision and Bias
11.4.1 After setting up the validation test equipment, test for
leaks by blocking the inlet to the analyzer and pressurizing the 12.1 Precision—An interlaboratory study of this test method
sample transport line to around 100 to 150 kPa. Do not block is being conducted and a complete precision statement is
the outlet of the analyzer as high pressure may damage the expected to be available prior to 2026.
analyzer. Inspect for leaks using soapy water or an electronic 12.1.1 Repeatability—The repeatability standard deviation
leak detector, if available. If no leaks are found, the analyzer from a single operator has been determined to be approxi-
port can be opened. mately 60.1 to 60.5 ppm(v) on an average reading of
11.4.2 Monitor the readings for several minutes until they 4 ppm(v).
stabilize and then record the concentration reading. If the 12.1.2 Precision will vary depending on the model and
readings are within the tolerance of the reference standard’s manufacturer of the TDLAS analyzer. The operator shall refer
stated uncertainty plus the analyzer’s stated uncertainty, the to the manufacturer’s published repeatability specifications as
analyzer’s performance has been validated and the factory well as the associated repeatability conditions for the apparatus
certification remains valid. under test.
11.4.3 With some analyzers, this process may be automated 12.2 Accuracy—The test method does not dictate the accu-
using switching valves and software whereby the auto- racy for particular applications; quality objectives of the
validation check can be programmed to occur periodically or measurement are determined by the buyer and seller of the gas
by electronic command. (per the contract).
11.4.4 A single-point validation is sufficient for most 12.3 Bias—The procedure in this test method for the analy-
applications to verify that the analyzer is working properly sis of hydrogen sulfide in natural gas by tunable diode laser has
(given the inherent linearity of TDLAS response). However, a no known bias.
two-point verification may be required by regulation or for
some applications and instruments. 13. Keywords
11.5 Concentration—The concentration for performing 13.1 hydrogen sulfide; H2S; natural gas; TDL; TDLAS;
validation may be recommended by the manufacturer. tunable diode laser absorption spectroscopy

6
D8488 − 22
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

You might also like