CM-350C
User’s Manual
1
CM-350C User’s Manual 1282
Statement
Our company has the right to modify this User manual without previous noticing the
end users.
No copy or translating version is legal without our company’s permission.
Notice: Please disconnect the power before open the machine casing.
2
CM-350C User’s Manual 1282
1.Product features:
Class that equipment belongs to: Class I
Type that equipment belongs to: Type BF
Power supply: AC 220V±10% or AC 110V±10%
Power capacity: 1100VA±10%
Equipment application condition: Disconnect with the ground when working at high
frequency.
The Equipment is designed to defense from defibrillation.
General type machine
Working mode: Intermittent loading, working continuously
Non-perpetual installation equipment
2. Symbols
3
CM-350C User’s Manual 1282
3. Technical Parameters
3.1 Environment requirement.
a) Temperature :5℃~40℃;
b) Relative humidity :≤80;
c) Barometric pressure :86.0~106.0kPa;
e) Power supply :220V±22V/110V±11V, 50/60Hz±1Hz。
Working frequency :360~512kHz
3.2 Output Rating:
a) CUT: 1W~350W(500Ω)
b)Blend 1 : 1W~250W(500Ω);
c)Blend 2 : 1W~200W(500Ω);
d)COAG 1: 1W~150W (500Ω);
e)COAG 2: 1W~80W (500Ω);
f) BIOPOLAR: 1W~ 70W(100Ω)
3.3 Power supply:5A 220V/110V∽±10%,50/60Hz。
3.4 Machine power dissipation:≤1000VA。(CUT 400W、BIOPOLAR 70W output)
3.5 Transportation and storage
Environment temperature:一 40℃~55℃;
Relative humidity:≤95%;
Barometric pressure:500hPa~1060hPa。
Glossary explanation
4
CM-350C User’s Manual 1282
Neutral electrode is connected with the patient's body relatively large area of the
electrodes to provide a low density of high-frequency current loop, in order to avoid
the human tissues burns and such harmful physical effects.
Other names of Neutral electrode are Plate electrode, or deposited electrode.
The difference between high-frequency output parameters and the current settings of
the Electrosurgical unit is only caused by the load resistance is too low, for example,
coagulation electrode is too large, the short circuit between the electrode in use and
the neutral electrodes, or equipment failure. The Electrosurgical unit has a neutral
electrode safety system (NESSY), to monitor equipment and electrical connection
between the neutral electrodes and neutral electrode sticks on the patient in right way.
5
CM-350C User’s Manual 1282
6
CM-350C User’s Manual 1282
When use the neutral electrode with two contact surfaces, not only to monitor the
electrical connections between the equipment and the neutral electrode automatically,
but also to monitor the neutral electrode on the patient's right applicator.
In addition, it can do the automatic continuous measurement to the electrical
conductivity of two contact surfaces on the neutral electrode and the skin of patients,
and automatically compare the measurement results and the high frequency current
strength flowing through the neutral electrode.
If the high-frequency current intensity greater than the corresponding measured
conductivity of the permitted value NESSY will be issued the visual and audible
alarm signal, if the electrical conductivity between the measured neutral electrodes
and the patient is too weak, then you cannot start a unipolar mode of operation.
There are many various neutral electrodes in the high-frequency surgery, but so far
the best ideal neutral electrodes for all applications do not exist. In order to ensure that
when users select the most suitable neutral electrodes with greater freedom, NESSY
provides a flexible security system. It can be in the best way to apply for the users’
neutral electrode.
Warning
For security reasons, only if properly inform all the equipment user about the revision
that users could modify NESSY settings. In addition, modifications of NESSY must
be made appropriate records and archives.
d) If some key on the electrode is on a short circuit or low resistance due to some fault.
(such as the moisture inside the electrode handle )., Or is pressed when connect with
power. The machine will alarm and display the error code when connect with power.
e) If the foot switch happen to a short circuit or Pedal stuck or was pressed in the
process of connecting with power. The machine will sound alarm and display a error
code on the screen when connect with power.
Warning
7
CM-350C User’s Manual 1282
f) the machine will test the performance automatically after connecting with the
power. If some function is error, machine will alarm, and display the corresponding
the error code. The machine could identify the below error of the device and relevant
accessories.
--If some key on the front of the panel is short circuit or pressed incorrectly. The
alarm and error code will show where the error is after connecting with the power,
--If some key on the electrode is on a short circuit or low resistance due to some fault.
(such as the moisture inside the electrode handle )., Or is pressed when connect with
power. The machine will alarm and display the error code when connect with power
--e) If the foot switch happen to a short circuit or Pedal stuck or was pressed in the
process of connecting with power. The machine will sound alarm and display a error
code on the screen when connect with power.
--If the connection of the neutral electrode is disconnected, or contact resistance (only
two-sided contact with the surface of the neutral electrode is concerned), the device
will cut off the high-frequency generator, sound the alarm, and shows the error wrong
number
Thermal damage caused by the improperly used of the the neutral electrode or non use
the neutral electrode.
If the neutral electrode used improperly, or not even use, there will e dangerous of
8
CM-350C User’s Manual 1282
causing the patients’ body great danger of thermal damage on the role point of the
neutral point. Neutral electrode must be close to the surface, and reliably attached to
the patient's body muscle fullness.
Warning
Effective contact surface-Conductivity between the neutral electrode and patient must
be same to the intensity of high-frequency power capacity(high-frequency current).
Effective contact surface means the surface of neutral electrode which can have
conductive contact with the patient's skin during the operation.
The intensity of cutting and coagulation should be setting according to the function of
the device. Meanwhile, the time should be just enough. For example, the effect is not
good when the device is setting according to the standard. The reason may be the poor
neutral electrode attachment, poor contact with electrical connectors,faulty cable or
vestigial electrical insulation tissues in the application of electrodes.
9
CM-350C User’s Manual 1282
Monopolar cutting and / or coagulation must use the appropriate neutral electrode.
The neutral electrode should be not only connected with the device, but also sticked to
the patients carefully.
The Electrosurgical Unit is equipped with a neutral electrode safety system (NESSY).
It can automatically monitor the connected condition between the neutral electrode
and the device, and sticking states of patients
Warning
If you use the neutral electrode which has a single contact with the surface, NESSY
only can monitor the electrical connections between devices and neutral electrode, not
monitoring sticking states of patients.
NESSY: neutral socket is a neutral pole pad socket. Its mark means the following
item:
Normal Neutral Electrode
This Electrosurgical Unit suits demand of Chapter 19.101b of GB9706.4-2009,
Isolated from the ground when the neutral pole pad stays in high frequency status.
Based on GB9706.4-2009, our unit suits the demand of BF. Besides, NESSY also
indicates that the neutral pad could cling to the patient’s body during the no-fiber
shake.
NESSY = Neutral Electrode Security System
Neutral Electrode Security System .
Electrosurgical Unit has been set with NESSY, which could monitor the electrical
links between unit and neutral electrode.
When using a whole piece of neutral pad, it could only automatically monitor the
electronic link with pole pad. If the pad is connect well, then green color diode
Neutral indicator light, we could start all of the operation modes.; If the connection
break off, then red color diode neutral indicator light, we cannot start the monopole
mode.
When using two pieces of neutral pad, it could not only automatically monitor the
electronic link with pole pad, but also the cling situation with the patient’s body.
Besides, it could automatically test the electric conductivity between 2 pads and
patient’s body, and makes a automatically compare between the test result and the
10
CM-350C User’s Manual 1282
If the high frequency currency is much stronger than the permission of the tested
electric conduction, then NESSY will give a visual and audio alarm, then we cannot
start the Monopole operation mode.
There are various kinds of neutral pad, but there is no ideal model which could suit all
users’ demand. In order to ensure each user could use our machine freely, NESSY
supply a flexible neutral pad security system, which could give us the best choice to
use the neutral pole pad.
Security Note: Revising NESSY could only be done after informing all the users.
Meanwhile, NESSY Revising must be recorded and saved.
Before cling the two pieces of neutral pad to the patients’ body, we must first check if
the green color diode Neutral indicator light or not.
If the connection is break off after start, or the contact resistance is too big (only for
the divided pad), then the unit will give audio alarm of cutting the current, and give us
the mistake codes.
5.3 Cut
5.3.1
3 Cut Modes: Pure Cut, Blend Cut 1, Blend Cut 2. Every once pressing the button, the
modes changed.
5.3.2
11
CM-350C User’s Manual 1282
Switch on the power between 1-40w, pressing once to add or reduce 1w.
Switch on the power between 40-100w, pressing once to add or reduce 5w.
Switch on the power above 100w, pressing once to add or reduce 10w.
Switch on the power above for a long time, it will be add or reduce power quickly.
Note: In general type surgery, cutting mini watt mode should be employed, and
enlarge the power gradually if necessary. A depth 2~3cm knife edge should be cut by
a steel knife firstly, and then dip in the blood with gauze, at the same time keep
cutting and Coagulation. When the cut organization shows yellow, it indicates that the
output power is over than the required, you can improve the cutting speed or reduce
the output power. The best effect after cutting is that the organization shows white
with slight yellow, at this time, the hemostasis at good effect and body tissue burnt
slightly.
12
CM-350C User’s Manual 1282
function, so you don't have to set the power for the cutting quality. Adjustable power
limit is to ensure patient safety, prevent unintended thermal damage to tissue. At the
same time, it can prevent the damage of precise cutting devices, such as needle
electrodes, because if they are contacted with other metal parts when the device has
been initiated and cause high-frequency current, the devices will be easily broken.
This damage may occur in laparoscopic surgery. Don’t mistaken adjustable power
limit as the power setting of conventional high-frequency surgical device. The cutting
quality of the conventional device directly depends on the setting of power.
Cutting is particularly suitable for the human tissue for its poor conductivity.
However, the high frequency voltage of applied electrode, under the cutting model,
can achieve very high value, so cut with good electric insulation instruments, when
operate on poor conductive body tissue. We don’t recommend the cutting function for
Bipolar or quasi bipolar equipment, for their insulation is weak generally and can not
be connected with high voltage. You can adopt the other functions for the bipolar or
quasi bipolar devices.
During the surgical procedures, the doctor use cut or coagulation according to
different human tissue. Press the yellow button for cut, press the blue button for
coagulation. The output power is adjustable, and accompanied by an audible alert.
The machine can also work with the endoscope.
13
CM-350C User’s Manual 1282
When the displayed value is between 1-40, pressing Increase or decrease button each
time the value will plus or minus 1. When the displayed value is between 40-100,
pressing Increase or decrease button each time the value will plus or minus 5. When
the displayed value is above 100, pressing the buttons will plus or minus 10. If you
press the button continuously, the value will be increase or decrease quickly.
Since the endoscopic surgery has very strict rule about power output control, it is also
called low-voltage coagulation. Therefore it is mainly used in endoscopic surgery.
Generally been used for vascular coagulation. When cutting off the blood vessels,
which are about 1-2mm or less, the blood vessels can be points by the plane blade in
order to avoiding multiple use of gauze dipping in blood. When the white with a little
yellow color occurs, cut off the vessel from the middle.、
It’s a kind of automatic stop function of blood coagulation,the computer monitors the
bleeding tissue to reduce the adhesion. It turns off the output when seeing the blood is
stopped.
14
CM-350C User’s Manual 1282
It’s applied in the Orthopaedic, Microsurgery, Brain surgery, etc. Clamping tissue
with the bipolar forceps as less as possible in order to ensure the coagulation effect.
The rear panel of the computer high-frequency electrosurgical unit includes: outlet,
fuse block, foot switch jack, protective earth terminal, the product nameplate and fans.
5.5.1 Power Insurance 5 A (BGXP5 × 20) glass insurance, 4A (BGXP5 × 20) glass
insurance. Because when the maximum power output of the electrosurgical unit begin,
there will be the impulse current, 5 A insurance should be used.
5.5.2 AC outlet, rated 5A/220V ~, 50/60 Hz power frequency
5.5.3, The product nameplate.
5.5.4 Footswitch socket: Five pole function foot switch and hand control handle
simultaneously. Bipolar function only under the control of the foot switch.
6.1 Disinfection
Before using this machine, please connect the grounding wire. And also, please
disinfect the head, shank, bipolar forceps, cable and other parts of the application.
Disinfection methods: wipe off dirt with alcohol gauze, then placed in formalin
fumigation liquid container at room temperature, not less than 10 hours.
15
CM-350C User’s Manual 1282
be able to protect patients, operators and third-party from life and health hazards.
Prior to delivery of equipments, each device's functionality and security shall be
tested by the manufacturer. This is in order to ensure the reliability and security of
equipment performance after transportation and field installation. An operator can
operate the equipment only after field performance test by manufacturer or supplier
performance and introductions of how to operate according to the instructions by the
party responsible for equipment operations.
6.3 Output
6.3.1 Monopole Mode: Using handle pencil to make output. Handle output sockets:
Handle switches and foot switch could be used at the same time effectively.
6.3.2 Bipolar Mode: when using Bipolar output, only Foot Switch could be used.
Note: Handle pencil, Bipolar Needle and other accessories is prohibited to contact
with the liquid directly, nor soaked with medical liquid and other liquids.
Note: During working, negative pad must be inserted, otherwise, device cannot work
normally. And give or can not work, and give audio and video alarm.
6.5 The machine has pure cut, no blood cut(blend 1), cautery(blend 2), coagulation1,
coagulation2, high efficient bipolar coagulation modes for the efficient operation:
Note: General surgery should be used started from low-power output state, gradually
increase the output power if in need. Using steel knife to open a incision with depth of
2 ~ 3mm, Dipping blood with medical gauze, and at the same time ,cutting along the
incision and stanch bleeding. Yellow color of the incision indicates the power is too
large, can increase the cutting speed or reduce power output. White and light yellow
color indicates the best state of incision, stanch bleeding and less burn.
16
CM-350C User’s Manual 1282
Electrosurgical unit is equipped with an automatic open and close loop control system
to control and adjust the cut to ensure that the selected cutting quality and stability.
The electrosurgical unit is with high voltage automatic control, thus we do not need to
set the output power of cutting quality. In the cutting operation mode with high
voltage automatic control, so do not set the power of cutting quality. Adjustable
power limit is to ensure patient safety, prevent unintended thermal damage to human
tissue, while preventing the needle electrodes and the like, such as precision cutting
instruments have been activated in case of equipment and other metal equipment in
contact due to high frequency current is too high and cause damage, such damage
may occur in laparoscopic surgery. Adjustable power limit must not be confused with
conventional high-frequency surgical equipment, conventional cutting device is
directly dependent on the quality of the power settings.
Cut especially suit for human tissue with poor conductive. However, the application
of cutting mode on the high voltage electrode can reach very high values, therefore,
performance in the conductive tissues when doing the cutting, we must make sure to
use use good electrical insulation equipment. For the bipolar or quasi-bipolar devices,
cutting function is not recommended, because the devices are generally relatively
small insulated channel, cannot connect with high pressure voltage. For the bipolar or
quasi-bipolar devices, other functions recommended
17
CM-350C User’s Manual 1282
If we cut off the blood vessels of 1 ~ 2mm or less and to avoid multiple dipping the
blood with Gaze, you can use the plane blade ends click both ends of the
vessels. When the color turns white and slightly yellow, please cut off from
the middle. Press this button to choose one of the following coagulation modes.
7.2 Do not open the cabinet and don’t insert the metal objects into the pore when
using, if there are high-voltage inside the machine. Don’t let the liquid flow into the
machine, so as to avoid accidents.
7.3 Don’t use the knife handle and the electrodes for the "fight the plate experiment",
because the plate is thin or disposable negative plate.
18
CM-350C User’s Manual 1282
This requirement is not only suitable for application electrodes including cables and
plugs, but also suitable for neutral electrodes including cables and plugs. (6.8.2 aa in
GB9706.4)
All the isolations of electrode, handle of electrode, cable and plug etc. should be kept
in good conditions.
The whole area of polar plate (electrode plate) must closely attach to the body of
patient (beneath buttocks), and should be as close as possible to operation area.[ 6.8.2
bb)1)in GB9706.4]
Effective contact surface i.e. electric conductivity between neutral electrode and
patient should be equal to high frequency capacity applied namely the intensity of
high frequency current. Here, effective contacting surface means neutral electrode
surface contacting electric conduction contacting with patient’s skin during
operations.
If it is misused or even not used, then there would be a great risk to tissues of human
body by accidental heat trauma at point of neutral electrode as well as on the other
parts of patient.
There are two types of plate, metal polar plate, one-off skin application plate.
The metal polar plate must be packed in two layers. It should be wet by saline before
using. The whole area of polar plate must closely attach to the body of patient
(beneath buttocks), and should be as close as possible to operation area, in order to get
the good return circuit. One-off skin application plate electric conduction should be
gumming to the body of the patient (beneath buttocks).
7.6 Patient should not touch any metal parts (such as: operating-table, bracket etc.)
that are connected with grounding or the metal parts with considerable capacitance of
19
CM-350C User’s Manual 1282
7.7 To avoid skin-to-skin contacting (i.e. arm to body of patient), put a piece of dry
gauze in between. [6.8.2 bb)in 3GB9706.4]
Between the contacting points of four limbs or skin while lying, there should be put a
piece of dry gauze in between to isolate them each other.
7.8 When the high frequency surgical equipment and the physiological monitor are
used for the same patient, all monitoring poles shall be kept as far as possible from the
surgical pole. It is not recommended to use the needle-style monitoring pole. In any
case, it is recommended to use the monitoring system with high frequency current
limiter.
In any case it is recommended to use with high-frequency current limiter monitoring
system. [6.8.2 bb)4)in GB9706.4]
Do not use the electrocardiogram (ECG) electrodes within 15 com away from the
surgical area
7.9 The cables of the surgical pole shall be kept away from the patient or other cables.
For those surgical poles are not used temporarily, they shall be placed away from the
patient. [6.8.2 bb)5)in GB9706.4]
Like the scalpel, high-frequency surgery is a potential risk if Negligence, it should
always be cautious when using the cutting electrode and coagulation electrode, put it
far away from the patient or other staff during intermittent.
It is forbidden to make the hilt, bipolar forceps and other accessories contact the
solution directly, especially to immerge them in the solution or other liquids.
7.10 In the surgery process, when the high frequency current flows through the
smaller part of the body, the bipolar technology is preferred to avoid unnecessary
concretion.
20
CM-350C User’s Manual 1282
human tissue, smallest diameter area always feel the heat firstly. If the high frequency
current flows through the same diameter in a long distance (a), the tissue will be
coagulated in the entire distance. If the diameter of tissue which is near by the
coagulation point is smaller than the diameter of the coagulation point, then, in the
area nearby point (b) can also be coagulated.
Any time the doctor needs to to ensure that high-frequency current does not flow
through a small organization or small-diameter vascular
7.11 The target shall be achieved at a power output as low as possible. [6.8.2 bb)7)
in GB9706.4]
7.12 In the normal setting mode, if the output is lowered obviously or the bistoury
can’t work normally, it indicates that poor contact or improper use occurs on the pole
pads. [6.8.2 bb)8)in GB9706.4]
7.13 If the surgery is made on the chest or head, the flammable anesthetic, giggle gas
and oxygen shall not be used unless the anesthetic gas is exhausted or the
anti-anesthetic equipment is employed. Before the high frequency surgery is carried
out, the solvent of the flammable detergent, disinfector or bond shall be vaporized out.
Before using the bistoury, wipe the flammable fluids remained on the patient’s lower
body, body concave (e.g. bellybutton) and body cavity (vagina). Great attention shall
be paid on the fire risk due to the inside gas. Some materials (e.g. absorbent cotton
and gauze filled with oxygen) may be ignited by the spark produced by the device.
[6.8.2 bb)9)in GB9706.4]
21
CM-350C User’s Manual 1282
of explosion
Therefore, before surgery, We must remove flammable gases and / or use CO2-blown
spray the affected body cavity, to remove flammable gas.
During the surgery of Transurethral resection (TUR), in the electric arc between
the removal ring and the wash liquid, the H2O molecules crack into H2
and O2. These gases will be gathered at the top of the bladder, making
a high-explosive gas mixture. If we do the surgery in the mixture, there may be a risky
explosion.
7.14 For patients who are with pacemakers or pacemaker electrode, there may be a
danger, because it may interfere with the work of the pacemaker or the pacemaker
may be damaged,. If anything is abnormal, we should go and find
a cardiologist to get help. [6.8.2 bb) 10) in GB9706.4]
7.15 The interference caused by High-frequency surgical equipment may have bad
effect on the work of other medical electrical equipment. [6.8.2 bb) 11) in GB9706.4]
High-frequency surgical equipment usually have a high frequency voltage and current
which make interference with other electronic devices. When installed sensitive
electronic equipment in the operating room, we need to consider this issue. In
principle, sensitive electronic equipment should be placed as far as possible away
from the high-frequency surgical equipment, especially high-frequency
current transmission
cables. In addition, the high-frequency current cable works as
the radio antenna cable, so its length should not exceed the actual needs, and
certainly not parallel to the sensitive electronic equipment, or be too close.
Taking into account its interference for the sensitive electronic devices, this device
has been equipped with a dedicated generator. Compared with
conventional high-frequency surgical equipment, it produces a relatively low level of
interference.
22
CM-350C User’s Manual 1282
7.16 Users should check the accessories before using the machine every time,
especially check the electrode cable insulation. [6.8.2 dd in GB9706.4]
7.17.1 Pencil shank has no fixed number of uses. It may be damaged once it exceeds
its life. Pencil head can be use repeatedly.
7.17.2 Metal negative plate can be used repeatedly while disposal negative plate can
be used only once. When the wire of the negative plate is used repeatedly, please do
not use it more than 3 months. Once the plate surface damages, it should
be immediately replaced. It may be damaged at any time once it exceeds the limit of
usage.
Please make sure to clean the connector usually which connects the plate wire and
plate.
Make sure that we clean the conducting resin or liquid which flows into the connector
on time to avoid the damage to the machine.
7.17.3 There is no fixed times of uses for the cable of the bipolar forceps and cable
plugs. They may be damaged once they exceed the usage life.
Once the insulation of the bipolar drops, it should be replaced immediately.
7.17.5 Once there is any damage for the power cable of the main unit, it should be
replaced immediately.
Make sure that the main unit and attachment are working well and effective before
each surgery.
7.19 Disassemble parts and its manual for assemble and disassemble
The replacement manual for consumable material
23
CM-350C User’s Manual 1282
24
CM-350C User’s Manual 1282
There are 4 positioning locks within dustproof cover, hold one side of dustproof cover
with hand, take off it with force, wash and clean it with sponge, put it under natural
condition for drying after cleaning, and replace it within dustproof cover, relocate it
into the original place parallelly
7.20 Safety regular maintenance The main unit shall be run in unloaded status at the
normal temperature not less than 20 hours every month, check if the accessories run
normally and effectively.
7.21 The main unit shall be checked by professional technician at least four times
every year. mainly including removing dust in the machine, checking if the machine
works normally, safety inspection, condition of isolation, and checking if the
accessories are correct and effective. . .
7.22 When there are sound prompts such as, no power output or audio and light alarm
during operation of this machine, the machine cannot operate normally, please stop
the machine and check if high voltage fuse is damaged. If the high voltage fuse is not
damaged, please check if skin application plate and cable are in good condition;
replace them if they are damaged.
7.23 ★ Note that the product’s plug shall be kept in connection with the power
supply in a good status when it runs with power output.
★ To avoid accident, never leave the product unloaded running in the surgery
process.
7.24 In case that the power supply doesn’t comply the requirements (220V±10%,
supply frequency 50Hz ), please use regulated power supply. Note that when it runs
with power output, the product’s plug shall be kept in connection with the power
supply in a good status .In case that the power supply doesn’t comply the
requirements (220V±10%, supply frequency 50/60±1Hz ), please use regulated
power supply. To avoid accident, never leave the product unloaded running in the
surgery process.
7.25 It is crucial to check the ground wire of the power is reliable before start up the
product. Before using the product, please make sterilization on its head, hilt, bipolar
forceps, cables and other working parts.
Method of Sterilization: First, wipe off the filth on all parts with the alcohol gauze;
then fumigate them in formalin solution at the normal temperature not less than 10
hours.
25
CM-350C User’s Manual 1282
surgical operations. To avoid the risks during the operation, surgeons and their
assistants shall be aware of the risks and follow the rules so that can prevent. The
following explains the risks and the rules in detail.
Warning
In the high frequency surgery, it shall isolate the patient from conductive objects. The
black cover on operation table has conductivity which can split partial electronic
charge, so during the operation, to be safety, it shall sheet an insulating layer between
patient and black cover, like died cloth.
If the insulating layer is getting wet during the operation due to sweating, irrigation,
urine and etc., it shall adapt water-proof plastic layer and use catheter.
If you turn on the machine accidentally, and electrode conductor contact with the
patient directly or a damp cloth contact with patients indirectly, then it may result in
burning patients.
For instance, the following reasons may lead to turn on the machine inadvertently:
● Press the footswitch inadvertently
● Press the manual control switch inadvertently
● Fault in footswitch, manual control switch or cable
26
CM-350C User’s Manual 1282
●Conductive liquids (such as blood, amniotic fluid, urine, saline, wash fluid, etc.) into
the footswitch or manual control switch
● internal fault in high frequency unit
In order to prevent patients from thermal damage due to turning on the unit
inadvertently and pay more attention to the following usage rules:
●Never put electrode on patient’s body or aside them casually and never let electrical
conductor, or wet cloth contact with the patient indirectly.
●The cables of the surgical pole shall be kept away from the patient or other cables.
●Set sound signal can be recognized easily, which could prompt working condition of
high-frequency unit.
●In some operations, such as endoscope surgery, even if they do not work, cutting
electrode or forceps have to contact with the patient inevitably. At that time, surgical
shall use this instrument with great care.
The instruments shall be turn off when be removed from the body. Otherwise, all the
body parts that contact the instruments will be burned. Therefore, when that kind of
mistake happen surgical shall immediately cut off the power of high-frequency unit
and try to remove electrodes from human body.
27
CM-350C User’s Manual 1282
8. output data
8.2.1 neutral electrode (deposited plates) should be isolated from ground at high
frequencies, and application part must be isolated from ground both in the high and
low frequency , to ensure that the high-frequency leakage current passing thru 200Ω
resistor from each electrode to ground will not exceed 150mA when you do
experiment as per the following the requirements.
Keep the electrode away from the cable by 0.5m, place them on the insulated surface
which is 1m above the ground or any conductor surface , output port unloaded, and all
metal housing of Ⅱ type equipment and internal power supply device must be
grounded.
Device with insulated enclosure must be placed at a grounded metal plate whose area
at least equal to the device's basel area (see Figure 104 below).
Equipment operates at each work's maximum output setting, measuring the
28
CM-350C User’s Manual 1282
Any application part specially designed for the use of bipolar must be isolated from
ground and other application part, no matter at high and low frequencies.
≤0.1mA ≤0.5mA
Enclosure leakage
current
29
CM-350C User’s Manual 1282
10、The measures which must be taken when the equipment comes into use
10.1 please connect the additional ground wire. The end of the power cable which is
connected to the network power supply has three cores ,and the long one in the middle
should be connected to earth when the product is used.
10.2 Before using the product ,please make sterilization on its head, hilt, bipolar
forceps, and the working part of the high frequency cable.
Method of Sterilization: First, wipe off the filth on all parts with the alcohol gauze;
then fumigate them in formalin solution at the normal temperature not less than 10
hours.
11、Transport & Storage
11.1 Transport: As per the purchase contract.
11.2 Storage: The packed unite should be kept in a ventilated room without any
corrosive gas (Temperature: -40℃~ 55℃, RH≤95%).
ERR-4: Make sure that the current mode (Bipolar bipolar light for Bipolar bipolar
mode, light off for Unipolar mode)
30
CM-350C User’s Manual 1282
NOTE: circuit plate is not contact with the patient. Shows ERR-8
SHOWS ERR-1,ERR-2,ERR-6, please Return factory to repair, non-professionals
prohibit to repair!
31