Professional Documents
Culture Documents
Electric Bed
User Manual
(Apply to:C-4DS, C-4DL, C-4BL, C-2DL, C-2BL, M420, G420, W420, G220, B402)
Page | 1
1. Introduction
Note: Please charge the battery for 24 hours before first use
Electric bed is an electrically actuated acute care bed, which has a range of functions and
positional movements that assist caregivers and provide for patient comfort. Ease of use,
functionality, serviceability and cleaning practicality are key features of electric bed. The bed
has a fabricated steel structure with a powder coated finish. The plastic components are made
from urethane, PP and POM nylon and HDPE and ABS plastic.
The bed is appropriate for adult patient diagnosis, cure or care under the medical monitoring,
to support the patient to adjust their own position to meet clinical required position. See
details in “4.ELECTRONIC FUNCTIONS”.
Statement: For model C-4DS, the weighting function is only used for
measuring the weight of patient for basic reference. This weighting
function shall not be used for any clinical diagnosis or therapy purpose. If
necessary, please use other independent weighting measurement medical
devices for clinical purpose!!
Application Scope
C-4DS, C-4DL, C-4BL, M420 and G420 are intended to be used in APPLICATION
ENVIRONMENT 1, 2, 3, and 5;
W420 is class II ME equipment which is intended to be used in APPLICATION ENVIRONMENT
1, 2, 3, 4 and 5;
C-2DL, C-2BL, G220 and B402 are intended to be used in APPLICATION ENVIRONMENT 3 and
5.
Note:
APPLICATION ENVIRONMENT 1:
intensive/critical care provided in a hospital where 24 h medical supervision and constant
monitoring is required and provision of life support system/equipment used in medical
procedures is essential to maintain or improve the vital functions of the PATIENT
APPLICATION ENVIRONMENT 2:
acute care provided in a hospital or other medical facility where medical supervision and
monitoring is required and ME EQUIPMENT used in medical procedures is often provided to
help maintain or improve the condition of the PATIENT
APPLICATION ENVIRONMENT 3:
long-term care in a medical area where medical supervision is required and monitoring is
provided if necessary and ME EQUIPMENT used in medical procedures may be provided to
help maintain or improve the condition of the PATIENT
APPLICATION ENVIRONMENT 4:
care provided in a domestic area where ME EQUIPMENT is used to alleviate or compensate for
an injury, disability or disease
APPLICATION ENVIRONMENT 5:
outpatient (ambulatory) care, which is provided in a hospital or other medical facility, under
medical supervision where ME EQUIPMENT, is provided for the need of persons with illness,
Page | 2
injury or disability for treatment, diagnosis or monitoring
Statement: The patient is not the intend operator. If the patients have a certain
self-care ability, they can use the remote control to adjust their position in nursing
care (or in family member care).
Key Features
Page | 3
Mattress Support Platform from ground(mm)
Electric bed model
Low error±20 High error±20
C-4DS 380 840
C-4DL 380 840
C-4BL 380 840
C-2DL 380 840
C-2BL 380 840
M420 400 750
G420 285 690
G220 285 690
W420 380 840
B402 490 490
• Left and right tilting function, can be tilted to the left or right(15°)。( include this
function)
• (Single or Twin) castors with central operation for braking and steering or without
central operation. (optional)
• Equip with split side rails、¾ length folding side rails、Flip side rails and hidden side
rails (optional)
Page | 4
2. Model Classification
□ □ □ □
Product according to function and parameters are divided into 12 models (see sheet 1)
Page | 5
2.2 Different models function comparison sheet
Electric bed models C-4DS C-4DL C-4BL C-2DL C-2BL M420 G420 W420 G220 B402
Classification Class I Class I Class I Class I Class I Class I Class I Class II Class I Class I
Application Environment 1, 2, 3, 5 1, 2, 3, 5 1, 2, 3, 5 3, 5 3, 5 1, 2, 3, 5 1, 2, 3, 5 1, 2, 3, 4, 5 3, 5 3, 5
Mass of Electric Bed 196 kg 174 kg 142 kg 140 kg 140 kg 115 kg 141kg 143 kg 138 kg 101 kg
Max. Patient Weight 135 kg
Safe Working Load 250 kg
Input voltage AC 230 V 50 Hz
Power Input 350 VA 350 VA 350 VA 300 VA 300 VA 250 VA 350 VA 350 VA 300 VA 300 VA
Mattress Support Platform length (error±20 mm) 2030 2030 2030 2030 2030 2030 2030 2070 2030 2030
Mattress Support Platform Width (exclude side rails)
(error±20 mm) 830 830 830 830 830 830 920 920 920 920
Mattress Support Low (error±20 mm) 380 380 380 380 380 400 285 380 285 490
Platform from ground High (error±20 mm) 840 840 840 840 840 750 690 840 690 490
Head Boad maximum break angle (degree) ≥65 ≥65 ≥65 ≥65 ≥65 ≥65 68 70 ≥68 ≥65
Upper foot frame maximum angle (degree) ≥35 ≥35 ≥35 × × ≥35 ≥35 ≥35 ≥35 ≥35
Head and foot overall slope angle (error±1 degree) 0-14 0-14 0-14 0-14 0-14 × 0-14 0-14 × ×
Motor-driven Qty. 4 4 4 2 2 3 4 4 3 2
Break Qty. 3 3 3 2 2 3 3 3 3 2
Weighting function √ × × × × × × × × ×
side rails ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Electronic Patient Positioning (EPP) √ √ × × × × × × × ×
Nurse Control Panel ACL
quick release CPR √ √ √ √ √ ☼ ☼ ☼ ☼ ☼
Battery Backup √ √ √ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Central castors ☼ ☼ ☼ × × × × × × ×
Page | 6
Bed head /bed foot panel ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Under bed lighting ☼ ☼ × √ × × × × × ×
Extension frame (error±20 mm) 195 195 × × × × 195 × × ×
Table board ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Solid wood frame × × × × × × ☼ ☼ ☼ ×
Self pick up rod ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Drip rod ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Debris basket ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Note:
1、“√”is “have”; “☼”is optional (different models or “no”);“×”is “no”。
2、M420、G220、W420、B402 can’t be used for shock emergency treatment.
3、Bed board material can be offered ABS or steel sheet.
4、Frame and bed head/foot and colour are optional.
5、In addition to ACL (Attendant Control Lock), ACC (Attendant Control Compact) is an optional control for all models.
5、For model M420, G420, W420, G220 and B402, at least battery or CPR is equipped.
Sheet 2 Basic Parameter
Page | 7
3. WARNINGS & CAUTIONS
It is important that this User Manual is read in its entirety prior to the operation of the bed.
The manual is designed to ensure that users are equipped with the information required to
competently and responsibly operate the bed and its accessories.
It is recommended that the User Manual be kept in an easily assessable place for quick
reference. If required CARETEK product specialists or your local distributor can provide
additional training.
All warnings and cautions within this manual should be strictly adhered to. Failure to do so
could result in harm to the patient or operator.
GENERAL WARNINGS
• Electrical equipment can be hazardous if used incorrectly.
• Use the bed to support one patient only. The Electric bed is intended for one
person use and may become unstable if used incorrectly.
• Residual current devices (RCD) are not supplied with the Electric bed. Please
consult with your Biomedical Engineer/advisor concerning RCD requirements.
• The safe working load of the Electric bed is 250Kg. This includes the mass of
the patient (Max. Patient Weight is 135 kg), mattress and any accessories
fitted to the bed.
• Do not place the handset or power cord across the mattress surface of the
bed. When the bed is not connected to mains power. Always store the power
cord within the power cord fixtures fitted to the bed.
• Do not use accessories that are not designed or approved for use with the
Electric bed.
Page | 8
GENERAL SAFETY WARNINGS
• Always use a mattress of the correct size and type. A range of suitable
pressure reducing and pressure relieving mattresses are available from
CARETEK.
• Before operating the bed make sure that the patient is safely positioned to
avoid entrapment.
• Before transporting the bed make sure that the bed is at a suitable height to
move without stooping, or straining.
• When the patient's condition could lead to patient entrapment the mattress
support platform should be left in the flat position whilst unattended.
• The clinically qualified person responsible should consider the age, size and
condition of the patient before using the bed and its accessories.
• Always ensure that all hospital lines and power cables are free from possible
entanglement with mechanical parts of the bed.
• Always ensure that attendant and patient limbs are free from potential
entrapment areas before operating any functions.
• After any adjustments, repair or service and before use, make sure all
attaching hardware is tightened securely
• DO NOT make anybody underneath the bed or in between the raised bed
frame components at anytime
• Side Rails must only be used with a mattress of the correct size which is
approved for use with the bed.
• Seek and utilize input about the patient from the patient and family (or
Page | 9
authorized representative) to assist in identifying nursing and medical risk
interventions.
• Although bed rails are not rated to any specific weight limitation, the bed rails
may become deformed or broken if excessive side pressure is exerted on the
bed rails. The bed rail is not an assist rail for getting into or out of bed. DO
NOT use the bed rails as push handles when moving the bed.
Page | 10
4. ELECTRICTRONIC FUNCTIONS
Warning
• Ensure that the patient, caregivers and or other equipment is not positioned to
become either trapped or crushed during bed adjustment.
• Ensure that handset cable and electrical supply cable cannot become entangled
with moving parts of the bed.
• DISCONNECT the electrical power cable from the mains and store before
transporting the bed.
Caution
• Do not operate the bed controls for more than the control systems prescribed
duty cycle (i.e. ON:2 MIN. OFF/18 MIN).
• The mattress frame height adjustment uses two actuators working in conjunction
and separate to each other creating the vertical lift and enabling the bed to tilt in
opposite directions.
• The backrest adjustment actuator includes an inbuilt manual quick release CPR
function.
• Knee raise actuator
• The actuators are controlled by the buttons on the handset as well as the option of
being controlled by the Nurse Control Panel (ACL) at the foot of the bed, which
also enables the handset buttons to be disabled when required.
Handset Location
The handset is permanently attached at the centre of the bed through a flexible coiled cable.
With minimal manipulation the handset can be positioned at either side of the bed for easy
access by the patient or caregiver. The handset can be stored by hooking it on to the drainage
bag rail either side of the bed.
Adjustment occurs only while a button is pressed. Adjustment will stop when the button is
released, or the movement selection reaches the end of its range of adjustment.
Page | 11
Handset
To move the backrest up or down, press and hold the appropriate button on the handset. The
backrest can be adjusted through a range of between 0° and 65°.
The retracting backrest provides additional comfort for the patient by allowing the mattress to
expand. When operating the backrest or cardiac positioning handset buttons the backrest slides
progressively towards the head end of the bed. This action enlarges the seat section of the bed
and helps prevent compression of the abdomen reducing shearing and risk of tissue damage.
Warning
Care should be taken whilst operating the backrest function to ensure hands
and fingers are away from potential squeeze hazards.
Page | 12
4.3 Mattress Platform Height Adjustment
To move the mattress platform up or down, press and hold the appropriate button on the
handset. The beds platform height can be adjusted through a range (refer to chart1) (depending
on castor choice)
Warning
It is recommended that the bed be positioned at its lowest height when
unattended by caregivers to minimise the risk of patient injury from falls when
getting in and out of the bed.
To move the thigh positioning section up and down, press and hold the appropriate button on
the handset. The thigh section is adjustable through a range of 0° to 35°, which provides
adjustable support to the patient’s legs.
Auto Contour Positioning provides synchronised movement of the backrest and knee
positioning in one operation. The position helps provide a more natural resting environment for
the patient.
The tilt button allows the bed to be positioned in a foot or head down tilt position. Tilt
positioning achievable is between 0° and 14° for both head and foot down tilting.
Warning
The Trendelenburg function should only be used under professional supervision.
• In use the mattress platform may initially rise before tilting. The bed will automatically
adjust to ensure there is sufficient clearance between the mattress platform and the
mechanical structure of the bed before tilting.
• When reversing the tilt angle of the bed with the tilt button remaining depressed the
mattress platform will pause at horizontal for two seconds before continuing on with its
tilting movement
The under bed lighting feature provides a soft LED illumination of the floor beneath the bed
carriage. This safety feature provides floor illumination for patients at night when getting in
and out of bed. In the event the bed is operating on battery power (e.g. during transport) the
lights will power down after approximately 2-3 minutes.
Page | 13
4.8 Nurse Control Panel (ACL)
1 Unlock
2 Lock
The Attendant Control Lock (ACL) is used to prevent unauthorised operation of the beds main
functions. The ACL is located at the foot end of the bed or side of the bed.
1. If the knob is pointed to , the pendant control is locked, all movements of bed are
disabled.
2. If the knob is pointed to , the pendant control is unlocked, all movements of bed are
enabled.
Warning
ACL is used as an emergency stopping device to interrupt the movement of
medical bed, when the medical bed moves accidentally.
1 2
Up
3 4
Down
The Attendant Control Panel (ACP) is used to prevent unauthorised operation of the beds main
functions. The ACP is located at the foot end of the bed.
Page | 14
3. Pressing “BED LEVEL & UP” will raise the deck until it is horizontal while
simultaneously lowering the backrest and knee break back to a flat deck position.
Pressing “BED LEVEL & DOWN” will lower the deck until it is horizontal while
simultaneously lowering the backrest and knee break back to a flat deck position.
4. Pressing “CHAIR POSITION & UP” will raise the head and lower the foot of the bed
while simultaneously raising the backrest and knee break to create a seated position.
Pressing “CHAIR POSITION & DOWN” will lower the head and raise the foot of
the bed while simultaneously lowering the backrest and knee break back to a flat deck
position.
5. Pressing the Electronic Patient Positioning “EPP” button will automatically level all
functions of the bed back to a flat deck position whilst lowering the bed to a safe
working height.
Warning
When the emergency patient positioning (EPP) function is activated care should be
taken when there is potential for moving parts of the bed to either trap or crush the
patient, caregivers and or other equipment.
1. Pressing the “TRENDELENBURG” button will lower the head and raise the foot of
the bed to the maximum tilt of 14°, while simultaneously lowering the back rest and
knee break back to a flat deck position – NO pause if passing through horizontal.
Note: It is NOT necessary to push the ‘up/down’ buttons when operating the EPP or
TRENDELENBURG functions as these are programmed to operate directly from the
associated button.
This facility allows users to selectively disable some, or all of the electrical features of the bed.
In normal operation the key functions of the bed should be in an (unlocked) state of use. This is
indicated on the ACC panel by LED lights above the function buttons.
To lock and unlock each function hold the “key” button and at the same time press the function
button that is required to be locked or unlocked. An LED light will illuminate when the
function is in the locked position.
The bed is fitted with a battery backup system which allows the electrically operated functions
to be used for short periods when the bed is disconnected from mains power. The bed switches
between mains and battery power automatically.
Page | 15
The battery should be charged regularly and not allowed to discharge fully. Fully discharging
the battery will cause a reduction in battery life.
To charge the battery, connect the power cord to the mains supply. This should be done as
often as possible to maximise battery life. If the bed is not used for a long period of time it is
recommended that the batteries are charged in accordance with the section on Transportation
and Storage within this manual.
Warning
Do not expose the battery to open flame or immerse in liquid.
NOTE: The beds batteries should be charged for 24 hours prior to first use
NOTE: Keep the bed plugged to a mains power source for optimum performance.
NOTE: Bed adjustment will be slower when the power cord is unplugged and the bed is
powered from its batteries.
NOTE: Power cord storage is located at the head end of the bed. The cable should be stowed
when the bed is unplugged or in transport by wrapping around the “cord minder” hooks.
A. Buttons operation (All functions will be enabled by pressing combination button and a
Page | 16
selected function button simultaneously)
1.Enable
Press Combination button + On/Off button to enable the panel or turn off the panel. Any
function can be used when the panel is enabled.
2.Scale
Press the Scale button to measure the weight, then display will show the weight and unit. After
20seconds, display will show company’s logo.
Range: 10 kg to 250 kg
4.Zero button
Press Zero button before patient lay on the bed to clear the bed ,mattress and other article’s
weight, then only measure the patient’s weight.
6. Unit set
Two units can be selected, they are kg and lbs, by default the unit is kg.
4.11 Control panel on side rail(Optional)
KNEE BREAK UP
BACK
REST UP MAIN
LIFT UP
BACK REST
DOWN KNEE BREAK
DOWN
MAIN
LIFT
The way of enable the function button please refer to page 12.
Page | 17
5. Assembly and general operation
5.1 Side Rails
The fitted ¾ length side rails have two positions, either fully raised or folded down closed.
When the side rail is raised to its full height position it will automatically lock in the
positioning hole on the rails clamp plate. The release/lock knob feature can be both visually
and audibly confirmed.
The side rails are fixed (bolted) in place, but can be removed if required.
Warning
Do not operate side rails when there is potential of entrapment or crushing to the
patient, caregivers, visitors and/or other equipment.
5.1.1 Side Rail Operation 2
Warning
• Care should be taken whilst operating the side rails to ensure hands and fingers
are kept away from potential squeeze hazards.
• Side rails should only be used after clinical assessment and within
organisational policy and guidelines.
The side rail locating pin has a yellow coating that allows for visual inspection of the locking
device. When the side rail is in the raised position, viewing the release knob from the side there
should be no yellow shaft visible. If there is yellow shaft visible the safety side is NOT
securely locked in the upright position and needs to be adjusted.
The yellow coating should NOT be visible from above when the safety side is correctly locked
in position.
Unlocked Locked
Image1 Iamge2
Page | 18
Side rail release and lock knob shown in unlocked and locked positions
Warning
• After adjustment of the side rail visually check the locking pin location.
There should be no yellow shaft on the locking pin visible if locked
correctly.
Warning
• Care should be taken whilst operating the side rails to ensure hands and fingers
are kept away from potential squeeze hazards.
• Side rails should only be used after clinical assessment and within
organisational policy and guidelines.
Warning
• Care should be taken whilst operating the side rails to ensure hands and fingers
are kept away from potential squeeze hazards.
• Side rails should only be used after clinical assessment and within
organisational policy and guidelines.
Page | 19
5.2 Manual Cardiopulmonary Resuscitation (CPR) Function
A clearly marked emergency CPR release handle is available on both sides of the bed. The
manual release function returns the beds backrest to a level position in the event of emergency
resuscitation being required.
Emergency Lowering: To lower the backrest in an emergency, pull up on the red CPR release
handle (1). The handles are positioned at the side and towards the head of the bed. The
lowering movement is assisted by a gas damper, but CAUTION is required as the backrest
may lower quickly under patient weight.
In addition to the CPR lever there is an Electronic Patient Positioning (EPP) button on the
Nurse Control Panel (ACL). This button will drive the bed deck into a low flat horizontal
position automatically, regardless of the beds configuration before starting the movement.
Warning
The backrest can fall quickly - keep hands clear to avoid trapping.
Page | 20
Central Lock Pedal Positions
Caution
• Disconnect the bed power cord before moving the bed.
• Do not remove the power cord from the source by pulling on the cord.
• Do not use the beds side rails to manoeuvre the bed.
• Do not pull on the power cord, or handset cord to manoeuvre the bed.
• Always apply the brakes when the bed is in a stationary position.
Panel Aligned
5.5 Bed Extension Panel Fitted
Release handle
Page | 21
The integral bed extension is located at the foot of the bed and is extendable to 19.5cm from its
normal closed position.
To Extend: Apply the central locking brakes. Pull up on the Release Handle and applying a
central force pull out from the mattress platform. The Bed Extension automatically locks when
fully extended and the latch lever is released.
Reverse to Close
Warning
• The total Linen Rack load should not exceed 10Kg
• Do not leave the Linen Rack in the extended position when not in use.
Page | 22
6. Parameter list
Warning
The mattress height should not exceed 150mm ,otherwise it may cause a risk of fall
down
z Check for compression of the mattress’ outside perimeter. Easily compressed
perimeters can increase the gaps between the mattress and the bed rail.
z The mattress to bed rail interface should prevent an individual from falling between
the mattress and bed rails and possibly smothering.
z Care should be taken that the mattress does not shrink over time or after cleaning.
Such shrinkage increases the potential space between the rails and the mattress.
z Ensure that the mattress is appropriately sized for the selected bed frame, as not all
beds and mattresses are interchangeable.
z The space between the bed rails and the mattress and the headboard and the mattress
should be filled either by an added firm inlay or a mattress that creates an interface
with the bed rail that prevents an individual from falling between the mattress and
bed rails.
z Maintenance and monitoring of the bed, mattress, and accessories such as
patient/caregiver assist items should be ongoing.
Standard
Page | 23
7. Cleaning
Warning
• Before cleaning always disconnect the bed from the mains power source.
Caution: It is strongly recommended that protective clothing be used when carrying out
cleaning processes on the bed.
Caution: The following procedure also applies to the beds accessories, but does not include
mattresses.
• Wipe the beds surfaces with a soft cloth moistened with hot water and mild
detergent (or the hospital’s recommended cleaning solution). It is recommended that
the cloth be a white material to prevent bleeding of die colours to the bed. Take extra
care in areas that can harbour dirt or dust.
• To remove potentially infectious materials such as body fluids, or when the bed has
been used by someone with a known infection. Clean with NaDCC (e.g. Presept,
Actichlor) at 10,000 ppm of available chlorine. Rinse and dry.
Caution
z Do not use abrasive Hypocarbonate or Phenolic compounds or pads.
z According to the difference of the disinfector, there may make the metal parts
corrupted or the plastic parts turned colors, quality deteriorative change, split, so
please use the above recommended disinfectors.
z Please confirm well the ingredients of the disinfector before disinfection.
z The washing machine and automatic cleaning is not suitable for the equipment.
Page | 24
8. Repair and Maintain
The following preventative maintenance procedures should be carried out every 12 months.
Caution:Other than general maintenance activities all other service or repair work must be
performed by qualified and trained persons who are approved by CARETEK.
Follow local governing ordinances and recycling instructions regarding disposal or recycling of
device components, including batteries.
A. When maintenance, it is suggested that a use of lubricate of actuators is necessary by
service personnel or trained technician.
When do the daily examination on the connection joints of movable parts of the bed, please
add proper lubricate oil on them. If find serious breakages, please immediately replace it to
make sure the security
When components of the bed system (e.g., new bed rails or mattresses) are changed or replaced,
the reassessment of hospital bed is needed.
B. When using an extension cord, use only a flexible mains cables and cords shall be type
HD22.10 H05-BQ-F or equivalent quality for mechanical robusiness. Use of improper
extension cord could result in a risk of fire and electric shock.
Warning
The bed must be disconnected from the mains power supply before starting any
maintenance activity.
General maintain(Daily)
Set the Attendant Control Panel (ACL) lockout switches to locked. Refer to ACL
1
instructions in this manual (section 1.3.1.3).
Examine the bed for obvious signs of damage. All aspects of the equipment
2 should operate as intended. Check that all nuts, bolts and other fasteners are tight
and are not missing.
Examine flexible cables and conduits for cuts, cracks, abrasions or other
3
deterioration.
Check that the power supply plug is not damaged. If either the power cable or
4 plug is damaged then both the cable and plug must be replaced as a complete
assembly by an approved service agent.
Apply the brakes and push the bed forwards. If any of the four braking castors
1
rotate the brake is not fully effective.
If either the brakes or the steering locks are not working correctly contact an
3
approved service agent.
Page | 25
Battery
A beeping sound heard during normal bed operation means that the battery is at
1 low charge. Reconnect the bed to the mains power supply for a minimum of 24
hours. If the problem reoccurs use step 2 below for verification.
Check the batteries condition every six months using the following procedure:-
1. Make sure that the bed has been connected to the mains power supply for
an uninterrupted 24 hour period.
2. Disconnect the bed from the mains power supply.
3. Apply a load of approximately 80Kg to the mattress platform.
4. Raise and lower the mattress platform from minimum low height to
2
maximum high height three times.
5. If the bed does not operate correctly in step 4, perform steps 1 to 4 again.
If the bed continues to operate incorrectly the batteries may need
replacing. In this situation consult an approved service agent.
6. It is suggested that the battery should be optimized when used in used or
in storage for two months, otherwise the running time of battery is
significantly shortened.
Side rails
If the red shaft is visible when the side rail is in the raised and locked position,
4
contact CARETEK for service advice.
Check that the rails pivot axles are securely fitted to the bed. The pivot axles are
locked into place by the manufacturer by a 10mm lock nut. It is rare that re-
5
tightening of the axles would be required, but in the case that it is ensure that the
nut is not over tightened during the tightening procedure.
Page | 26
9. Transportation and storage
The following limits apply to normal operating conditions and or periods of storage
longer than 15 weeks:
When storing the bed, or when the bed is not in use for prolonged periods of time it is
recommended that the beds battery system be maintained periodically using the following
guidelines:-
Page | 27
10. Labels, signs and warnings instruction on bed
Mass
Safe working load of SHP (Lifting pole including the Grip) 76,5 kg
:CAUTION
CE certification
Page | 28
Authorised Representative In The European Community
Manufacturer
Date of manufacture
serial number
This way up
Keep dry
Page | 29
11. Trouble shooting
Page | 30