You are on page 1of 30

...

with people in mind

Electric Bed
User Manual

Manufacture’s Name: Guangzhou GT Electronics Limited


Manufacture’s Address: Xiguakeng, Guanqiao, Shilou Town,Panyu
District,Guangzhou, CHINA

(Apply to:C-4DS, C-4DL, C-4BL, C-2DL, C-2BL, M420, G420, W420, G220, B402)

Audit date:May 6, 2014 Revised:October 9, 2014 Version: 2014/10


Guangzhou GT Electronics Limited

Page | 1
1. Introduction
Note: Please charge the battery for 24 hours before first use

Electric bed is an electrically actuated acute care bed, which has a range of functions and
positional movements that assist caregivers and provide for patient comfort. Ease of use,
functionality, serviceability and cleaning practicality are key features of electric bed. The bed
has a fabricated steel structure with a powder coated finish. The plastic components are made
from urethane, PP and POM nylon and HDPE and ABS plastic.

Clinical Applications and Area of Use

The bed is appropriate for adult patient diagnosis, cure or care under the medical monitoring,
to support the patient to adjust their own position to meet clinical required position. See
details in “4.ELECTRONIC FUNCTIONS”.

Statement: For model C-4DS, the weighting function is only used for
measuring the weight of patient for basic reference. This weighting
function shall not be used for any clinical diagnosis or therapy purpose. If
necessary, please use other independent weighting measurement medical
devices for clinical purpose!!

Application Scope
C-4DS, C-4DL, C-4BL, M420 and G420 are intended to be used in APPLICATION
ENVIRONMENT 1, 2, 3, and 5;
W420 is class II ME equipment which is intended to be used in APPLICATION ENVIRONMENT
1, 2, 3, 4 and 5;
C-2DL, C-2BL, G220 and B402 are intended to be used in APPLICATION ENVIRONMENT 3 and
5.

Note:
APPLICATION ENVIRONMENT 1:
intensive/critical care provided in a hospital where 24 h medical supervision and constant
monitoring is required and provision of life support system/equipment used in medical
procedures is essential to maintain or improve the vital functions of the PATIENT
APPLICATION ENVIRONMENT 2:
acute care provided in a hospital or other medical facility where medical supervision and
monitoring is required and ME EQUIPMENT used in medical procedures is often provided to
help maintain or improve the condition of the PATIENT
APPLICATION ENVIRONMENT 3:
long-term care in a medical area where medical supervision is required and monitoring is
provided if necessary and ME EQUIPMENT used in medical procedures may be provided to
help maintain or improve the condition of the PATIENT
APPLICATION ENVIRONMENT 4:
care provided in a domestic area where ME EQUIPMENT is used to alleviate or compensate for
an injury, disability or disease
APPLICATION ENVIRONMENT 5:
outpatient (ambulatory) care, which is provided in a hospital or other medical facility, under
medical supervision where ME EQUIPMENT, is provided for the need of persons with illness,

Page | 2
injury or disability for treatment, diagnosis or monitoring

Statement: The patient is not the intend operator. If the patients have a certain
self-care ability, they can use the remote control to adjust their position in nursing
care (or in family member care).

Product structure and composition


The registration involving total 10 models electric bed (C-4DS, C-4DL, C-4BL, C-2DL, C-
2BL, M420, G420, W420, G220, B402), the material , structure, principle. Function, use,
electrical parameters and others are basically the same. The structure and function of C-4DS
is the most complex in the series. Therefore, C-4DS as an example to show its function and
working principle, others are similar.
Electric bed consists of Mattress Support Platform, base frame, electric SYSTEM, other
accessories are optional.
Mattress Support Platform consist with many different function support plate, such as back
section, seat section, upper leg section, leg section, Head Board, Foot Board, both side rails,
and so on. Undercarriage consists with castors, bottom frame, lifting frame, etc. Electric
operation part mainly by linear propulsion motor, the control box, the patient controller, lock
control panel , weighing sensors ( Whether equip and QTY. find Sheet 2 Basic Parameters)
Linear propulsion motor make of drive motor, reduction, screws, nuts, bushings, putting, slide,
spring, shell, turbine, vortex rods, mocro control switch, ect. Accessories mainly include: Drip
rod, self pick up rod, table and so on.

Key Features

• Total safe working load (SWL) of the bed is 250Kg.


• Weighing function,accuracy 0.1kg/0.5kg(C-4DS include this function、M420 and
optional)
• Out of bed alarm function.(C-4DS、M420 and include this function)

• Electrically adjusted height range,adjusted height see sheet 1:


Sheet 1 Electrically adjusted height range

Page | 3
Mattress Support Platform from ground(mm)
Electric bed model
Low error±20 High error±20
C-4DS 380 840
C-4DL 380 840
C-4BL 380 840
C-2DL 380 840
C-2BL 380 840
M420 400 750
G420 285 690
G220 285 690
W420 380 840
B402 490 490

• Electrically actuated mattress platform Trendelenburg tilt (14°) and reverse


Trendelenburg tilt (14°), which includes an electronic horizontal stop function.
(M420、W420、G220 and B402 exclude this function)
• Four sectioned electrically operated profiling steel mattress platform with independent
adjustment of backrest angle (Maximum 65°), knee section angle (Maximum 35°) (、
G220、C-2Bl and C-2DL exclude knee panel adjustment function)

• Left and right tilting function, can be tilted to the left or right(15°)。( include this
function)

• Retracting backrest to reduce abdominal squeezing.


• Auto Contour Positioning - Provides synchronised movement of the backrest and knee
positioning in one operation.(B402、、G220、C-2BL and C-2DL exclude this
function)

• (Single or Twin) castors with central operation for braking and steering or without
central operation. (optional)

• Integral mattress platform extension.(G220、W420、、B402、C-2BL and C-4BL


exclude this function)
• Towel rack(C-2DL、C-2BL、M420、W420、、G420、G220 and B402 exclude this
function)

• Equip with split side rails、¾ length folding side rails、Flip side rails and hidden side
rails (optional)

• Light quantity head and foot panels, interchangeable.(optional)

• Two under bed LED white lighting (optional)

Page | 4
2. Model Classification

2.1 Model and encoding rules


□ - □ □ □

Mattress Support Platform specification code (S is high


configuration type ;L is standard type)
Type code (B is basic; D is Luxury)
Model code (2:2 section bed;4:3 section bed)
Series Code (C:C series)

□ □ □ □

Type code (20:adjustable height electric bed;


02: fixed height electric bed)
Model code (2:2 section bed;4:3 section bed)
Series Code (M、W、G、L、B series)

Product according to function and parameters are divided into 12 models (see sheet 1)

Page | 5
2.2 Different models function comparison sheet
Electric bed models C-4DS C-4DL C-4BL C-2DL C-2BL M420 G420 W420 G220 B402
Classification Class I Class I Class I Class I Class I Class I Class I Class II Class I Class I
Application Environment 1, 2, 3, 5 1, 2, 3, 5 1, 2, 3, 5 3, 5 3, 5 1, 2, 3, 5 1, 2, 3, 5 1, 2, 3, 4, 5 3, 5 3, 5
Mass of Electric Bed 196 kg 174 kg 142 kg 140 kg 140 kg 115 kg 141kg 143 kg 138 kg 101 kg
Max. Patient Weight 135 kg
Safe Working Load 250 kg
Input voltage AC 230 V 50 Hz
Power Input 350 VA 350 VA 350 VA 300 VA 300 VA 250 VA 350 VA 350 VA 300 VA 300 VA
Mattress Support Platform length (error±20 mm) 2030 2030 2030 2030 2030 2030 2030 2070 2030 2030
Mattress Support Platform Width (exclude side rails)
(error±20 mm) 830 830 830 830 830 830 920 920 920 920

Mattress Support Low (error±20 mm) 380 380 380 380 380 400 285 380 285 490

Platform from ground High (error±20 mm) 840 840 840 840 840 750 690 840 690 490
Head Boad maximum break angle (degree) ≥65 ≥65 ≥65 ≥65 ≥65 ≥65 68 70 ≥68 ≥65

Upper foot frame maximum angle (degree) ≥35 ≥35 ≥35 × × ≥35 ≥35 ≥35 ≥35 ≥35
Head and foot overall slope angle (error±1 degree) 0-14 0-14 0-14 0-14 0-14 × 0-14 0-14 × ×
Motor-driven Qty. 4 4 4 2 2 3 4 4 3 2
Break Qty. 3 3 3 2 2 3 3 3 3 2
Weighting function √ × × × × × × × × ×
side rails ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Electronic Patient Positioning (EPP) √ √ × × × × × × × ×
Nurse Control Panel ACL
quick release CPR √ √ √ √ √ ☼ ☼ ☼ ☼ ☼
Battery Backup √ √ √ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Central castors ☼ ☼ ☼ × × × × × × ×

Page | 6
Bed head /bed foot panel ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Under bed lighting ☼ ☼ × √ × × × × × ×
Extension frame (error±20 mm) 195 195 × × × × 195 × × ×
Table board ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Solid wood frame × × × × × × ☼ ☼ ☼ ×
Self pick up rod ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Drip rod ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Debris basket ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼ ☼
Note:
1、“√”is “have”; “☼”is optional (different models or “no”);“×”is “no”。
2、M420、G220、W420、B402 can’t be used for shock emergency treatment.
3、Bed board material can be offered ABS or steel sheet.
4、Frame and bed head/foot and colour are optional.
5、In addition to ACL (Attendant Control Lock), ACC (Attendant Control Compact) is an optional control for all models.
5、For model M420, G420, W420, G220 and B402, at least battery or CPR is equipped.
Sheet 2 Basic Parameter

Page | 7
3. WARNINGS & CAUTIONS

It is important that this User Manual is read in its entirety prior to the operation of the bed.
The manual is designed to ensure that users are equipped with the information required to
competently and responsibly operate the bed and its accessories.

It is recommended that the User Manual be kept in an easily assessable place for quick
reference. If required CARETEK product specialists or your local distributor can provide
additional training.

All warnings and cautions within this manual should be strictly adhered to. Failure to do so
could result in harm to the patient or operator.

Warnings highlighted throughout this manual identify possible hazards in procedures or


conditions which if not followed correctly could result in serious injury or fatality.

Cautions highlighted throughout this manual identify possible hazards in procedures or


conditions, which if not followed correctly could result in equipment damage or failure.

GENERAL WARNINGS
• Electrical equipment can be hazardous if used incorrectly.

• Do not use electrically operated beds in the presence of flammable gases.

• Do not use electrically operated beds in operating theatres.

• Use the bed to support one patient only. The Electric bed is intended for one
person use and may become unstable if used incorrectly.

• It is recommended that the bed be positioned at its lowest height when


unattended by caregivers to minimise the risk of patient injury from falls when
getting in and out of the bed.

• Residual current devices (RCD) are not supplied with the Electric bed. Please
consult with your Biomedical Engineer/advisor concerning RCD requirements.

• The safe working load of the Electric bed is 250Kg. This includes the mass of
the patient (Max. Patient Weight is 135 kg), mattress and any accessories
fitted to the bed.

• Do not place the handset or power cord across the mattress surface of the
bed. When the bed is not connected to mains power. Always store the power
cord within the power cord fixtures fitted to the bed.

• Do not use accessories that are not designed or approved for use with the
Electric bed.

• Unauthorized modification of the ME equipment or its part is not allowed

• Strangling, suffocating, bodily injury, or death may occur when patients or


parts of their bodies are caught between rails or between the bed rails and
mattresses.

Page | 8
GENERAL SAFETY WARNINGS
• Always use a mattress of the correct size and type. A range of suitable
pressure reducing and pressure relieving mattresses are available from
CARETEK.

• Before operating the bed make sure that the patient is safely positioned to
avoid entrapment.

• Before transporting the bed make sure that the bed is at a suitable height to
move without stooping, or straining.

• Always apply the brakes when the bed is stationary.

• When the patient's condition could lead to patient entrapment the mattress
support platform should be left in the flat position whilst unattended.

• The clinically qualified person responsible should consider the age, size and
condition of the patient before using the bed and its accessories.

• Always ensure that all hospital lines and power cables are free from possible
entanglement with mechanical parts of the bed.

• Always ensure that attendant and patient limbs are free from potential
entrapment areas before operating any functions.

• After any adjustments, repair or service and before use, make sure all
attaching hardware is tightened securely

• DO NOT make anybody underneath the bed or in between the raised bed
frame components at anytime

• Side Rails must only be used with a mattress of the correct size which is
approved for use with the bed.

• Maximum mattress thickness should be 150mm.

• Incompatible mattresses can create hazards. Read instructions for use.

• Chose the mattress for the medical bed by below size.

• The manual/electric bed is NOT designed to be used as a patient transport.


When transporting a patient, use an approved patient transport. Otherwise,
injury or damage may result.

• Re-evaluate and revise patient’s treatment program as needed if an episode of


entrapment or near-entrapment occurs with or without serious injury.

• Seek and utilize input about the patient from the patient and family (or

Page | 9
authorized representative) to assist in identifying nursing and medical risk
interventions.

• Although bed rails are not rated to any specific weight limitation, the bed rails
may become deformed or broken if excessive side pressure is exerted on the
bed rails. The bed rail is not an assist rail for getting into or out of bed. DO
NOT use the bed rails as push handles when moving the bed.

• The mains plug is considered as isolation device from supply mains.

• Do not to position the equipment to make it difficult to disconnection.

• The temperature of pendant control may up to 41,2°C. Do not position the


pendant control contacting with the patient to protect against heat injury to
the patient.

Page | 10
4. ELECTRICTRONIC FUNCTIONS

Warning
• Ensure that the patient, caregivers and or other equipment is not positioned to
become either trapped or crushed during bed adjustment.

• Ensure that the coiled power lead is not stretched excessively

• Ensure that handset cable and electrical supply cable cannot become entangled
with moving parts of the bed.

• DISCONNECT the electrical power cable from the mains and store before
transporting the bed.

• If the power cable shows any sign of damage, replace immediately.

• Some functions may cause injury to the patient if unintentional movement


occurs. See Nurse Control Panel (ACL) instructions for handset lockout to
avoid unintentional movement.

Caution
• Do not operate the bed controls for more than the control systems prescribed
duty cycle (i.e. ON:2 MIN. OFF/18 MIN).
• The mattress frame height adjustment uses two actuators working in conjunction
and separate to each other creating the vertical lift and enabling the bed to tilt in
opposite directions.
• The backrest adjustment actuator includes an inbuilt manual quick release CPR
function.
• Knee raise actuator
• The actuators are controlled by the buttons on the handset as well as the option of
being controlled by the Nurse Control Panel (ACL) at the foot of the bed, which
also enables the handset buttons to be disabled when required.

Handset Location

The handset is permanently attached at the centre of the bed through a flexible coiled cable.
With minimal manipulation the handset can be positioned at either side of the bed for easy
access by the patient or caregiver. The handset can be stored by hooking it on to the drainage
bag rail either side of the bed.

4.1 Handset Button Operation

Adjustment occurs only while a button is pressed. Adjustment will stop when the button is
released, or the movement selection reaches the end of its range of adjustment.

Page | 11
Handset

BACK REST UP BACK REST DOWN

MAIN LIFT UP MAIN LIFT DOWN

KNEE BREAK UP KNEE BREAK DOWN

AUTO CONTOUR UP AUTO CONTOUR DOWN

TILT – ANTI TREND TILT - TREND

UNDER BED LIGHT


ON/OFF

4.2 Backrest Adjustment

To move the backrest up or down, press and hold the appropriate button on the handset. The
backrest can be adjusted through a range of between 0° and 65°.

The retracting backrest provides additional comfort for the patient by allowing the mattress to
expand. When operating the backrest or cardiac positioning handset buttons the backrest slides
progressively towards the head end of the bed. This action enlarges the seat section of the bed
and helps prevent compression of the abdomen reducing shearing and risk of tissue damage.

Warning
Care should be taken whilst operating the backrest function to ensure hands
and fingers are away from potential squeeze hazards.

Page | 12
4.3 Mattress Platform Height Adjustment
To move the mattress platform up or down, press and hold the appropriate button on the
handset. The beds platform height can be adjusted through a range (refer to chart1) (depending
on castor choice)

Warning
It is recommended that the bed be positioned at its lowest height when
unattended by caregivers to minimise the risk of patient injury from falls when
getting in and out of the bed.

4.4 Knee Positioning Adjustment

To move the thigh positioning section up and down, press and hold the appropriate button on
the handset. The thigh section is adjustable through a range of 0° to 35°, which provides
adjustable support to the patient’s legs.

4.5 Auto Contour Positioning

Auto Contour Positioning provides synchronised movement of the backrest and knee
positioning in one operation. The position helps provide a more natural resting environment for
the patient.

4.6 Tilt Positioning - Trendelenburg / Reverse Trendelenburg

The tilt button allows the bed to be positioned in a foot or head down tilt position. Tilt
positioning achievable is between 0° and 14° for both head and foot down tilting.
Warning
The Trendelenburg function should only be used under professional supervision.

• In use the mattress platform may initially rise before tilting. The bed will automatically
adjust to ensure there is sufficient clearance between the mattress platform and the
mechanical structure of the bed before tilting.

• When reversing the tilt angle of the bed with the tilt button remaining depressed the
mattress platform will pause at horizontal for two seconds before continuing on with its
tilting movement

4.7 Under Bed Lighting(Optional)

The under bed lighting feature provides a soft LED illumination of the floor beneath the bed
carriage. This safety feature provides floor illumination for patients at night when getting in
and out of bed. In the event the bed is operating on battery power (e.g. during transport) the
lights will power down after approximately 2-3 minutes.

Page | 13
4.8 Nurse Control Panel (ACL)

1 Unlock

2 Lock

The Attendant Control Lock (ACL) is used to prevent unauthorised operation of the beds main
functions. The ACL is located at the foot end of the bed or side of the bed.

1. If the knob is pointed to , the pendant control is locked, all movements of bed are
disabled.

2. If the knob is pointed to , the pendant control is unlocked, all movements of bed are
enabled.

Warning
ACL is used as an emergency stopping device to interrupt the movement of
medical bed, when the medical bed moves accidentally.

4.9 Nurse Control Panel (ACC) (Optional)

1 2
Up

3 4
Down

The Attendant Control Panel (ACP) is used to prevent unauthorised operation of the beds main
functions. The ACP is located at the foot end of the bed.

Page | 14
3. Pressing “BED LEVEL & UP” will raise the deck until it is horizontal while
simultaneously lowering the backrest and knee break back to a flat deck position.

Pressing “BED LEVEL & DOWN” will lower the deck until it is horizontal while
simultaneously lowering the backrest and knee break back to a flat deck position.

4. Pressing “CHAIR POSITION & UP” will raise the head and lower the foot of the bed
while simultaneously raising the backrest and knee break to create a seated position.

Pressing “CHAIR POSITION & DOWN” will lower the head and raise the foot of
the bed while simultaneously lowering the backrest and knee break back to a flat deck
position.
5. Pressing the Electronic Patient Positioning “EPP” button will automatically level all
functions of the bed back to a flat deck position whilst lowering the bed to a safe
working height.
Warning
When the emergency patient positioning (EPP) function is activated care should be
taken when there is potential for moving parts of the bed to either trap or crush the
patient, caregivers and or other equipment.

1. Pressing the “TRENDELENBURG” button will lower the head and raise the foot of
the bed to the maximum tilt of 14°, while simultaneously lowering the back rest and
knee break back to a flat deck position – NO pause if passing through horizontal.

Note: It is NOT necessary to push the ‘up/down’ buttons when operating the EPP or
TRENDELENBURG functions as these are programmed to operate directly from the
associated button.

4.10 ACC Disable Function(Optional)

This facility allows users to selectively disable some, or all of the electrical features of the bed.
In normal operation the key functions of the bed should be in an (unlocked) state of use. This is
indicated on the ACC panel by LED lights above the function buttons.

To lock and unlock each function hold the “key” button and at the same time press the function
button that is required to be locked or unlocked. An LED light will illuminate when the
function is in the locked position.

4.11 Battery Backup Function

The bed is fitted with a battery backup system which allows the electrically operated functions
to be used for short periods when the bed is disconnected from mains power. The bed switches
between mains and battery power automatically.

The battery Features are as following:


Rated Capacity is 1.3Ah 24V
Charging Current is max. 500mA.

Page | 15
The battery should be charged regularly and not allowed to discharge fully. Fully discharging
the battery will cause a reduction in battery life.

To charge the battery, connect the power cord to the mains supply. This should be done as
often as possible to maximise battery life. If the bed is not used for a long period of time it is
recommended that the batteries are charged in accordance with the section on Transportation
and Storage within this manual.

Warning
Do not expose the battery to open flame or immerse in liquid.

NOTE: The beds batteries should be charged for 24 hours prior to first use

NOTE: Keep the bed plugged to a mains power source for optimum performance.

NOTE: Bed adjustment will be slower when the power cord is unplugged and the bed is
powered from its batteries.

NOTE: Power cord storage is located at the head end of the bed. The cable should be stowed
when the bed is unplugged or in transport by wrapping around the “cord minder” hooks.

NOTE: The replacement of the battery must be operated by professional personal.

4.12 Functionality buttons on the SCO display(Optional)


Statement: The weighting function is only used for measuring the weight
of patient for basic reference. It shall not be used for any clinical
diagnosis or therapy purpose.

A. Buttons operation (All functions will be enabled by pressing combination button and a

Page | 16
selected function button simultaneously)
1.Enable
Press Combination button + On/Off button to enable the panel or turn off the panel. Any
function can be used when the panel is enabled.

2.Scale
Press the Scale button to measure the weight, then display will show the weight and unit. After
20seconds, display will show company’s logo.
Range: 10 kg to 250 kg

3.Accuracy( change from 0.5% to 0.1%)


Generally 0.5% will be applied while 0.1% can be applied just for reference.

4.Zero button
Press Zero button before patient lay on the bed to clear the bed ,mattress and other article’s
weight, then only measure the patient’s weight.

5.Temporary compensation button


This function is used to Add or Remove articles from bed temporarily, while the system only
show the patient weight.
Operation:
Press Compensation button+ combination button before add articles to bed, then display shows:
Please + or – things, then add or remove articles from bed temporarily, like clothes, towels,
books, pillows and so on, then press compensation button+ combination button again, now
display shows “OK”, now it is shows the original weight of the patient.

6. Unit set
Two units can be selected, they are kg and lbs, by default the unit is kg.
4.11 Control panel on side rail(Optional)
KNEE BREAK UP
BACK
REST UP MAIN
LIFT UP

BACK REST
DOWN KNEE BREAK
DOWN

MAIN
LIFT

■ The LED of the function will be light when it is disabled.

The way of enable the function button please refer to page 12.

Page | 17
5. Assembly and general operation
5.1 Side Rails
The fitted ¾ length side rails have two positions, either fully raised or folded down closed.
When the side rail is raised to its full height position it will automatically lock in the
positioning hole on the rails clamp plate. The release/lock knob feature can be both visually
and audibly confirmed.
The side rails are fixed (bolted) in place, but can be removed if required.
Warning
Do not operate side rails when there is potential of entrapment or crushing to the
patient, caregivers, visitors and/or other equipment.
5.1.1 Side Rail Operation 2

To raise the side: Lift the top


rail until the locating pin/knob
automatically clicks into its
locked position at maximum
height.
To lower the side: Hold the 1
top rail near the foot end of the
bed (1). Pull the operating knob
(2) and lower the side towards
the foot of the bed controlling
the descent speed. As the side
begins to be lowered the
operating knob can be released.

Warning
• Care should be taken whilst operating the side rails to ensure hands and fingers
are kept away from potential squeeze hazards.
• Side rails should only be used after clinical assessment and within
organisational policy and guidelines.
The side rail locating pin has a yellow coating that allows for visual inspection of the locking
device. When the side rail is in the raised position, viewing the release knob from the side there
should be no yellow shaft visible. If there is yellow shaft visible the safety side is NOT
securely locked in the upright position and needs to be adjusted.
The yellow coating should NOT be visible from above when the safety side is correctly locked
in position.
Unlocked Locked

Image1 Iamge2

Page | 18
Side rail release and lock knob shown in unlocked and locked positions

Warning
• After adjustment of the side rail visually check the locking pin location.
There should be no yellow shaft on the locking pin visible if locked
correctly.

• DO NOT use other manufacturer’s side rails on the bed.

5.1.2 Separate side rails


operation

To raise the side rails:


Lift the top rail until the locating
pin/knob automatically clicks into its
locked position at maximum height.

To lower the side rails:Pull the


handle (1), side rails will lower to its 1
bottom automatically. Then adjust
the backrest to be flat position.
1

Warning
• Care should be taken whilst operating the side rails to ensure hands and fingers
are kept away from potential squeeze hazards.

• Side rails should only be used after clinical assessment and within
organisational policy and guidelines.

5.1.3 Turn over side rails


operation
2
To raise the side rails: Lift the
top rail (1) until the locating
1
pin/knob automatically clicks into
its locked position at maximum
height.
To raise the side rails: Pull 1
handle (2) automatically.

Warning
• Care should be taken whilst operating the side rails to ensure hands and fingers
are kept away from potential squeeze hazards.

• Side rails should only be used after clinical assessment and within
organisational policy and guidelines.

Page | 19
5.2 Manual Cardiopulmonary Resuscitation (CPR) Function
A clearly marked emergency CPR release handle is available on both sides of the bed. The
manual release function returns the beds backrest to a level position in the event of emergency
resuscitation being required.

Emergency Lowering: To lower the backrest in an emergency, pull up on the red CPR release
handle (1). The handles are positioned at the side and towards the head of the bed. The
lowering movement is assisted by a gas damper, but CAUTION is required as the backrest
may lower quickly under patient weight.

In addition to the CPR lever there is an Electronic Patient Positioning (EPP) button on the
Nurse Control Panel (ACL). This button will drive the bed deck into a low flat horizontal
position automatically, regardless of the beds configuration before starting the movement.

(1) CPR Release Handle

Warning
The backrest can fall quickly - keep hands clear to avoid trapping.

5.3 Castor System - Brakes & Steering


The bed central locking castor system is operated by the use of foot pedals found at the foot
end of the bed.
• With the pedals in the Red - Down Brake position the brakes are applied to all four
castors of the bed simultaneously and the bed is immobilised.
• With the pedals in the Horizontal Free position the brakes are released and the bed can
be moved in any direction.
• With the pedals in the Green - Down Steer/Track position, two steering castors at the
head end of the bed are engaged to provide better control when pushing the bed in a
straight line.
• When moving the bed, push from the foot end holding the panels top push rail.

Page | 20
Central Lock Pedal Positions

Brake Free Steer

Caution
• Disconnect the bed power cord before moving the bed.
• Do not remove the power cord from the source by pulling on the cord.
• Do not use the beds side rails to manoeuvre the bed.
• Do not pull on the power cord, or handset cord to manoeuvre the bed.
• Always apply the brakes when the bed is in a stationary position.

5.4 Head and Foot Panels


Installation & Removal
The panel includes a decorative infill to one side. There is no specific way in which the panels
are to be fitted as they are both interchangeable and therefore cannot be placed incorrectly.
Generally the infill face of the panel is fitted to face towards the foot of the bed providing a
more pleasing aesthetic appearance.
To Install:Using two hands align the panels bottom tubes with the beds corner sockets as
shown in the below diagrams. Lower the panel gently until it seats at the bottom of the beds
plastic corner sockets.
To Remove: Reverse the above procedure.

Panel Aligned
5.5 Bed Extension Panel Fitted

Release handle

Page | 21
The integral bed extension is located at the foot of the bed and is extendable to 19.5cm from its
normal closed position.

To Extend: Apply the central locking brakes. Pull up on the Release Handle and applying a
central force pull out from the mattress platform. The Bed Extension automatically locks when
fully extended and the latch lever is released.

To Close: Reverse the above

5.6 Linen Rack


The Linen Rack is used to rest clean linen on
whilst bed sheets are being changed. The
Linen Rack is located at the foot end of the
bed and extends by 20cm.

To extend: Using two hands on the soft


coated handle simply pull the friction held
stripper rail towards you until it stops in its
fully extended position.

Reverse to Close

Warning
• The total Linen Rack load should not exceed 10Kg
• Do not leave the Linen Rack in the extended position when not in use.

5.7 Self Help pole (lifting pole)(Optional) -- A device suspended above


the bed intended to allow the patient to change position by gripping it. Safe working load
of 76.5 kg.
5.8 IV Pole (Optional)--Hanging infusion bottles bags. Safe working load
of 10 kg.
5.9 X-Ray Cassette Holder(Optional)--A device for putting the X-ray
film on it.
5.10 “C” Size Oxygen Bottle Holder(Optional)--A device Clamps
to head or foot end of Profile beds for fixing the Oxygen Bottle, Safe working load of 15
kg.

Page | 22
6. Parameter list

Recommend mattress size

Max mattress Width =840-860mm, Length =1940mm, Height =100 mm-


size 120mm

Warning
The mattress height should not exceed 150mm ,otherwise it may cause a risk of fall
down
z Check for compression of the mattress’ outside perimeter. Easily compressed
perimeters can increase the gaps between the mattress and the bed rail.
z The mattress to bed rail interface should prevent an individual from falling between
the mattress and bed rails and possibly smothering.
z Care should be taken that the mattress does not shrink over time or after cleaning.
Such shrinkage increases the potential space between the rails and the mattress.
z Ensure that the mattress is appropriately sized for the selected bed frame, as not all
beds and mattresses are interchangeable.
z The space between the bed rails and the mattress and the headboard and the mattress
should be filled either by an added firm inlay or a mattress that creates an interface
with the bed rail that prevents an individual from falling between the mattress and
bed rails.
z Maintenance and monitoring of the bed, mattress, and accessories such as
patient/caregiver assist items should be ongoing.

Electrical Data (230V)

See 2.2 Different models function


Power In
comparison sheet
Duty Rating Interrupt 10%, Max 2 minutes / 18minutes
Class I (except model W420 is Class II),
Electric Shock Protection Type B applied part (The whole bed is
applied part)
Liquid Ingress Protection IPX4
2 x 12V Series Connected Batteries
Internally Powered Equipment
Sealed, Rechargeable Lead/Acid Gel
(Battery Backup Only)
1.2 amp hrs.
C-2BL, C-2DL, C-4BL, C-4DL, C-4DS:
0078281V1-0, V-1.0
Software version:
B-402, G220, G420, M420, W420: no
software configured

Standard

The production of the bed is compliance with IEC 60601-2-52 international


standard and passed EMC/EMI test.

Page | 23
7. Cleaning

Warning
• Before cleaning always disconnect the bed from the mains power source.

Caution: It is strongly recommended that protective clothing be used when carrying out
cleaning processes on the bed.
Caution: The following procedure also applies to the beds accessories, but does not include
mattresses.

• Wipe the beds surfaces with a soft cloth moistened with hot water and mild
detergent (or the hospital’s recommended cleaning solution). It is recommended that
the cloth be a white material to prevent bleeding of die colours to the bed. Take extra
care in areas that can harbour dirt or dust.

• Rinse with clean water and a dry clean cloth.

• To remove potentially infectious materials such as body fluids, or when the bed has
been used by someone with a known infection. Clean with NaDCC (e.g. Presept,
Actichlor) at 10,000 ppm of available chlorine. Rinse and dry.

• Allow the cleaned parts to dry before replacing the mattress.

Caution
z Do not use abrasive Hypocarbonate or Phenolic compounds or pads.
z According to the difference of the disinfector, there may make the metal parts
corrupted or the plastic parts turned colors, quality deteriorative change, split, so
please use the above recommended disinfectors.
z Please confirm well the ingredients of the disinfector before disinfection.
z The washing machine and automatic cleaning is not suitable for the equipment.

Page | 24
8. Repair and Maintain

The following preventative maintenance procedures should be carried out every 12 months.
Caution:Other than general maintenance activities all other service or repair work must be
performed by qualified and trained persons who are approved by CARETEK.
Follow local governing ordinances and recycling instructions regarding disposal or recycling of
device components, including batteries.
A. When maintenance, it is suggested that a use of lubricate of actuators is necessary by
service personnel or trained technician.
When do the daily examination on the connection joints of movable parts of the bed, please
add proper lubricate oil on them. If find serious breakages, please immediately replace it to
make sure the security
When components of the bed system (e.g., new bed rails or mattresses) are changed or replaced,
the reassessment of hospital bed is needed.
B. When using an extension cord, use only a flexible mains cables and cords shall be type
HD22.10 H05-BQ-F or equivalent quality for mechanical robusiness. Use of improper
extension cord could result in a risk of fire and electric shock.

Warning
The bed must be disconnected from the mains power supply before starting any
maintenance activity.

General maintain(Daily)

Set the Attendant Control Panel (ACL) lockout switches to locked. Refer to ACL
1
instructions in this manual (section 1.3.1.3).
Examine the bed for obvious signs of damage. All aspects of the equipment
2 should operate as intended. Check that all nuts, bolts and other fasteners are tight
and are not missing.
Examine flexible cables and conduits for cuts, cracks, abrasions or other
3
deterioration.
Check that the power supply plug is not damaged. If either the power cable or
4 plug is damaged then both the cable and plug must be replaced as a complete
assembly by an approved service agent.

CENTRAL LOCKING SYSTEM

Apply the brakes and push the bed forwards. If any of the four braking castors
1
rotate the brake is not fully effective.

2 Check the two steering lock castors for correct operation.

If either the brakes or the steering locks are not working correctly contact an
3
approved service agent.

Page | 25
Battery

A beeping sound heard during normal bed operation means that the battery is at
1 low charge. Reconnect the bed to the mains power supply for a minimum of 24
hours. If the problem reoccurs use step 2 below for verification.
Check the batteries condition every six months using the following procedure:-

1. Make sure that the bed has been connected to the mains power supply for
an uninterrupted 24 hour period.
2. Disconnect the bed from the mains power supply.
3. Apply a load of approximately 80Kg to the mattress platform.
4. Raise and lower the mattress platform from minimum low height to
2
maximum high height three times.
5. If the bed does not operate correctly in step 4, perform steps 1 to 4 again.
If the bed continues to operate incorrectly the batteries may need
replacing. In this situation consult an approved service agent.
6. It is suggested that the battery should be optimized when used in used or
in storage for two months, otherwise the running time of battery is
significantly shortened.
Side rails

• The locking pin mechanism and striker plate arrangement should be


checked every three months.
• Inspect the side rails for general wear and tear on fittings and pivot joints
1 every six months.
• Check for rust, structural integrity of welds and paint finish every twelve
months.
• Check the bed rail for entrapment gaps
Raise the side rail to the upright position until the locking mechanism clicks into
2 place. Refer to the operating procedures on section 5.1. of this manual for further
instruction.
Check that the there is no yellow shaft visible. In the locked position the red on
3
the locking pin shaft should not be visible from above

If the red shaft is visible when the side rail is in the raised and locked position,
4
contact CARETEK for service advice.

Check that the rails pivot axles are securely fitted to the bed. The pivot axles are
locked into place by the manufacturer by a 10mm lock nut. It is rare that re-
5
tightening of the axles would be required, but in the case that it is ensure that the
nut is not over tightened during the tightening procedure.

Page | 26
9. Transportation and storage

Handle with care


The bed should be stored in a clean, dry well ventilated area.
Do not drop and avoid shock or violent impact when transporting.
The following limits apply during transport and or a storage period of up to 15 weeks
duration:
Ambient temperature -25°C to +70°C
Relative humidity ≤ 93%
Air pressure 50 kPa to 106 kPa

The following limits apply to normal operating conditions and or periods of storage
longer than 15 weeks:

Ambient temperature +10°C to +40°C


Relative humidity 30% to 75%
Air pressure 80 kPa to 106 kPa

When storing the bed, or when the bed is not in use for prolonged periods of time it is
recommended that the beds battery system be maintained periodically using the following
guidelines:-

Storage temperature Charging Interval

Charge batteries for 24


20°C or below
hours every 9 months.

Charge batteries for 24


20 to 30°C
hours every 6 months.

Charge batteries for 24


30 to 40°C
hours every 3 months.

Page | 27
10. Labels, signs and warnings instruction on bed

Side rail locked warning

Max. patient weight 135 kg; Safe working load 250 kg

Mass

Safe working load of Extended Rack 20KG

Safe working load of Linen Rack 10 kg

Safe working load of SHP (Lifting pole including the Grip) 76,5 kg

Safe working load of IV pole total 10 kg

Potential Shearing Hazard

Potential Squeezing Hazard

: Refer to instruction manual/booklet

:CAUTION

: Protection Against Ingress Of Water

CE certification

Page | 28
Authorised Representative In The European Community

Class II ME equipment (only apply to W420)

Type B Applied part

Manufacturer

Date of manufacture

serial number

potential equalization conductor

Correct Disposal of This Product (Waste Electrical & Electronic Equipment)


Statement: The LAY RESPONSIBLE ORGANIZATION must contact its local
authorities to determine the proper method of disposal of potentially bio-hazardous
parts and ACCESSORIES.
Temperature limitation
The product package should be transported or stored at a temperature between -25 and
70 degrees (centigrade).
Upper limit of humidity
The product package should be transported or stored at a humidity ≤ 93%.
Atmospheric pressure limitation
The product package should be transported or stored at an atmospheric pressure
between 50 kPa and 106 kPa.

Fragile, handle with care

This way up

Keep dry

Page | 29
11. Trouble shooting

12.Warranty and Service


Product Name Electric bed Model/Specification See cover page
Manufacturer GUANGZHOU GT ELECTRONICS LIMITED
Address/ zip Shilou Town, Panyu District, Guangzhou
code Guangdong Province, China 511447
Tel +86 20 848 46868 Fax +86 20 848 47888
Manufacture Manufacture date
See label Expiration date
date +10year
Warranty Warran
Warranty of Bed 3year of 1year ty of 2year
battery castor
After-sale
GUANGZHOU GT ELECTRONICS LIMITED
service company
Website www.chinabusinesslink.com

Page | 30

You might also like