Development of U.S. EPA Method 551.1: Daniel P. Hautman
Development of U.S. EPA Method 551.1: Daniel P. Hautman
Downloaded from https://academic.oup.com/chromsci/article/35/5/221/339378 by Hanoi School of Public Health user on 16 February 2024
Daniel P. Hautman*
International Consultants, Inc., 26 W . Martin Luther King Dr., Cincinnati, Ohio 45219
David J . Munch
U.S. Environmental Protection Agency, Office of Water, Cincinnati, Ohio 45268
disinfection byproducts and six halogenated solvents. This method mended as the extraction salt. The change to Na S0 was war
2 4
is significantly revised as U.S. EPA method 551.1 to include 12 ranted due to the influence of trace bromide ion impurities
commonly observed chlorination disinfection byproducts, eight found in NaCl, which were shown to increase sample concen
halogenated organic solvents, and 17 pesticides and herbicides. The trations of brominated DBPs (2,3). Finally, a procedure detailing
applicability of the revised method has increased as a consequence a simultaneous confirmation analysis has been included.
of the larger analyte list. In addition, built into the method is the
The development of U.S. EPA method 551.1 (4) is discussed.
flexibility to choose only those analytes from the list which are
Analyte stability studies in buffered aqueous matrices illustrate
critical to a specific application. Several procedural modifications
are incorporated into the method, such as simplifying sample
the importance of matrix acidification. Volatility experiments
preservation, specifying a different salting agent, allowing the were conducted which indicated the importance of specifying
conditional use of pentane as an alternate extraction solvent, and careful sample handling procedures in order to reduce analyte
recommending simultaneous confirmation analysis. Data are losses.
presented that illustrate the importance of including these method
modifications into U.S. EPA method 551.1.
Experimental
Reproduction (photocopying) of editorial content of this journal is prohibited without publisher's permission. 221
Journal of Chromatographic Science, Vol. 35, May 1997
(J&W Scientific, Folsom, CA). The linear velocity of the as indicated for the primary column. The same temperature
helium carrier gas was 25 cm/s at 35°C. The column oven was program was utilized to allow for simultaneous confirmation
temperature-programmed as follows: held at 35°C for 22 min, analysis.
then increased to 145°C at 10°C/min and held for 2 min, then A splitless mode injector (Hewlett-Packard, Wilmington, DE)
increased to 225°C at 20°C/min and held for 15 min, then with the split delayed to 0.50 min was used at a temperature of
63
increased to 260°C at 10°C/min and held for 30 min or until all 200°C. A linearized nickel electron-capture detector (ECD)
expected compounds had eluted. (Hewlett-Packard) was also used at a temperature of 290°C.
As the confirmation column, a 30-m χ 0.25-mm-i.d. fused- The makeup gas was 95% argon and 5% methane, and the flow
silica capillary column with chemically bonded 6% cyanopropyl- rate was 30mL/min.
Downloaded from https://academic.oup.com/chromsci/article/35/5/221/339378 by Hanoi School of Public Health user on 16 February 2024
phenyl and 94% dimethyl polysiloxane phase was used (Rtx- As a result of including these additional method analytes, the
1301,1.0-μm film thickness or equivalent) (Restek, Bellefonte,
PA). The linear velocity of the helium carrier gas was 25 cm/s at
Table II. Analyte Retention Times
35°C. The column oven was temperature-programmed exactly
Primary Confirmation
Table I. U.S. EPA 551.1 Method Analytes column column
retention time retention time
Analyte CAS no. Analyte (min) (min)
*Analytes added as part of the revised U.S. EPA method 551.1. Atrazine and simazine coeluted on the confirmation column.
222
Journal of Chromatographic Science, Vol. 35, May 1997
chromatographic analysis time was increased to approximately may, however, result in poorer detection limits because the
1.5 h. Consequently, a simultaneous confirmation analysis is sample deposited on the precolumn is split into two columns.
recommended as a means to process samples efficiently. In By performing a simultaneous confirmation, a complete pri
order to perform simultaneous confirmation, the GC must be mary quantitative and secondary analyte confirmation anal
equipped with a second ECD. Using a two-hole ferrule, both the ysis is conducted from a single injection.
primary and confirmation capillary columns are inserted into
a single injection port. Alternatively, an uncoated 1-m length of Matrix preservation and acidification
deactivated capillary column can be installed into the injection The haloacetonitriles included in this method are susceptible
port and subsequently attached to the primary and confirma to base-catalyzed degradation; therefore, field sample acidifi
Downloaded from https://academic.oup.com/chromsci/article/35/5/221/339378 by Hanoi School of Public Health user on 16 February 2024
tion columns using a Y-connector. This alternate procedure cation is required. Method 551 includes a procedure which
Fortified Observed
‡
concentration concentration* Average MDL †
EDL
Analyte (μg/L) (μg/L) % recovery RSD (%) (μg/L) (μg/L)
223
Journal of Chromatographic Science, Vol. 35, May 1997
involved dropwise addition of concentrated HC1 to the sample pling team, concentrated HC1 may need to be shipped with the
matrix until pH 4.5 is attained. Due to the presence of car sampling kit. This would require special packaging conforming
bonate and other dissolved salts in field sample matrices which to Department of Transportation (DOT) shipping regulations as
create a natural buffering effect, the volume of concentrated well as field handling precautions, due to the corrosive nature
HC1 required to adjust this pH was site-specific and seasonally of HC1.A preferred alternative to this HCl procedure would
dependent. Consequently, a trained field sampling team would include the addition of a dry buffering agent placed in the sam
need to carefully monitor the matrix pH after each dropwise ad pling vial in the laboratory prior to shipping the kit to the field
dition of HC1. If the carbonic acid endpoint at pH 4.2 was sur sampling site. This procedure would remove many of the
passed, the sample pH would fall rapidly, and the integrity of the potential variables that would be introduced by the sampling
Downloaded from https://academic.oup.com/chromsci/article/35/5/221/339378 by Hanoi School of Public Health user on 16 February 2024
sample would be compromised. In addition, to ensure that a team and surmount the shipping concerns pertaining to con
consistent high-quality concentrated HC1 is used by the sam- centrated HCI.
MtBEextracted,NH4Cl
NH Clpreserved
preservedreagent
4 reagentwater
water
Fortified Observed
†
concentration concentration* Average MDL EDL‡
Analyte (μg/L) (μg/U recovery (%) RSD (%) (Hg/L) (Mg/L)
224
Journal of Chromatographic Science, Vol. 35, May 1997
Initial experiments were conducted using a pH 6.0 phos phosphate salts to 2.5% Na HP0 and 97.5% KH P0 , a pH of
2 4 2 4
phate buffer. The pH 6.0 dry phosphate buffer was a homoge 5.3 was achieved. By using this modified phosphate buffer, a
neous mixture of 13% sodium phosphate, dibasic (Na HP0 ) 2 4 second aqueous matrix stability study was conducted again
and 87% potassium phosphate, monobasic ( K H P 0 ) . An 2 4 using reagent water and local tap water which were dechlori
aqueous buffered pH of 6.0 was attained by dissolving 1.0 g of nated and fortified with method analytes. Triplicate analyses
this dry buffer in 100 mL of water. By using this buffer salt, an were conducted on samples extracted on days 0, 7,14, and 21.
aqueous matrix stability study was conducted using reagent During this storage study, the aqueous samples were again
water and local tap water which were dechlorinated and forti held in a refrigerator at 4°C and were protected from light.
fied with method analytes. Triplicate analyses were conducted
Downloaded from https://academic.oup.com/chromsci/article/35/5/221/339378 by Hanoi School of Public Health user on 16 February 2024
on samples extracted on days 0, 7, 14, and 20. During this Sample handling and analyte volatility
storage study, the aqueous samples were held in a refrigerator As part of the development of method 551.1, analyte volatility
at 4°C and were protected from light. experiments were conducted. These volatility experiments were
Most analytes were stable; however, significant losses were designed to examine the impact that sample transfers from
observed in trichloroacetonitrile (TCAN) by day 7. TCAN losses one sampling vial to a second vial would have on analyte
by day 20 are illustrated in Figure 3. As a result of this pH 6.0 recoveries. These transfers seem logical because samples are
stability study, it became apparent that further pH reduction of received in 60-mL vials and 50 mL of sample is required for
the matrix was required to stabilize TCAN from base-catalyzed extraction. Typically, an analyst may be inclined to zero a top
degradation. A buffering salt combination that would yield a loading balance with a clean, empty 60-mL vial and then
matrix pH of 4.5-5.0 would have been ideal. Experiments were transfer the sample from the original vial used for sample col
initiated with buffering systems containing succinic acid and lection until a weight of 50 g (nearly equivalent to 50 mL) is
sodium succinate or oxalic acid and sodium oxalate, but both of attained. Another possibility would involve using a graduated
these proved to be problematic. In each case, the nondissociated cylinder to measure the 50 mL volume and then pouring the
acid component of the buffer was an organic compound that measured sample into a clean 60-mL vial, actually accounting
could be extracted with the solvent and subsequently became a for two sample transfers. The original method 551 outlined a
chromatographic interferant. It was concluded that the procedure of extracting the sample in the original sampling vial
buffering salt needed to be an inorganic salt, and attention was without performing such sample transfers, but the text did not
refocused on the phosphate buffer. By adjusting the ratio of the explain the rationale behind this procedure. Several laboratories
225
Journal of Chromatographic Science, Vol. 35, May 1997
inquired about the acceptability of performing such sample icant problems using NaCl as the extraction salt in the analysis
transfers as a means to accurately measure sample volume, of chlorinated water samples that have utilized ammonium
and therefore specific experiments were conducted to evaluate chloride to reduce free chlorine. Chlorinated drinking water
analyte losses when these transfers were performed. must be dechlorinated at the time of sampling as a means to
A set of 12 identical fortified reagent water samples were eliminate additional reactions of free chlorine with humic and
prepared and divided into four experimental groups; each group fulvic organic matter from the source water. Free chlorine is the
contained three samples for subsequent triplicate extraction general term used to define the equilibrium of hypochlorous
–
and analysis. Only the more volatile DBPs and halogenated sol acid (HOC1) and hypochlorite ion (OCl ) introduced to the
vents were included in this study because the pesticides and water as a result of chlorination. If free chlorine is not reduced
Downloaded from https://academic.oup.com/chromsci/article/35/5/221/339378 by Hanoi School of Public Health user on 16 February 2024
herbicides are not volatile. The first group of three samples at the time of sampling, additional DBPs will form in the sam
was used for calibration with no sample transfers performed pling vial from these reactions, and the integrity of the sample
once the matrices had been fortified. This first group acted as at the time of sampling will not be preserved. Ammonium chlo
the control because it was presumed that minimal loss would ride is used as a reducing agent because it has no negative
occur if no sample transfers occurred. The second group of effects on any of the method analytes examined, specifically the
three samples was transferred from the vial in which they were haloacetonitriles, which are attacked by sodium sulfite or
originally fortified to a second vial and extracted immediately. sodium thiosulfate. Ammonium chloride converts free chlorine
A third group of three samples was transferred just like the to chloramine, which is much less reactive than free chlorine.
second but was subsequently set on the lab bench, capped but At the time of extraction, when sodium chloride is added to the
with approximately 10 mL headspace, for one half hour prior to sample as the extraction salt, trace impurities of bromide ions
–
extraction. The fourth group was first set for half an hour on the (Br ) can react with this combined chlorine species, and it is
lab bench, capped but with approximately 10 mL headspace, theorized that bromamines, which are believed to be more
then was transferred, recapped, and set for an additional half an reactive than chloramines, are formed. These bromamines can,
hour prior to extraction. in turn, react with organic matter to potentially increase the
amount of brominated organic DBPs (2).
Extraction salt investigations To confirm this effect, a study was conducted investigating
–
In U.S. EPA method 551, NaCl is recommended as the extrac the extraction salts NaCl, Na SO , and Br -doped NaCl. These
2 4
tion salt. An extraction salt is required as a means to salts were used in the extraction of chlorinated tap water that
increase the ionic strength of the aqueous matrix, thus increas had been preserved with ammonium chloride. The objective of
ing the extraction efficiency. Xie et al. (2,3) have found signif- this study was to compare the observed DBPs from samples that
226
Journal of Chromatographic Science, Vol. 35, May 1997
Downloaded from https://academic.oup.com/chromsci/article/35/5/221/339378 by Hanoi School of Public Health user on 16 February 2024
–
the maximum allowable trace Br ions in American Chemical types. These sample types were characterized as shown in Table
Society (ACS)-grade NaCl. ACS-grade NaCl is certified to con VII. Extractions were performed by first adding MtBE as the
tain less than 0.01% Br"; therefore, the 10 g of NaCl added to 50 extraction solvent. The first and second types of samples that
mL of sample could potentially contribute 1.0 mg Br". By employed either 10 g NaCl or 12 g Na S0 , respectively, were
2 4
Analyte peak identification and concentrations using MtBE Table VI. Referenced Data for Figure 2
for extraction of fortified reagent water on the primary column*
Peak Concentration (μg/L) Analyte peak identification and concentrationis using MtBE
Analyte
for extraction of fortified reagent water on the corconfirmation column*
1 Chloroform 5.00 Peak Analyte Concentration (μg/L)
2 1,1,1-Trichloroethane 5.00
3 Carbon tetrachloride 5.00 1 Chloroform 5.00
4 Trichloroacetonitrile 5.00 2 1,1,1-Trichloroethane 5.00
5 Dichloroacetonitrile 5.00 3 Carbon tetrachloride 5.00
6 Bromodichloromethane 5.00 4 Trichloroacetonitrile 5.00
7 Trichloroethylene 5.00 5 Trichloroethylene 5.00
8 Chloral hydrate 5.00 6 Bromodichloromethane 5.00
9 1,1 -Dichloro-2-propanone 5.00 7 1,1 -Dichloro-2-propanone 5.00
10 1,1,2-Trichloroethane 44.8 Chloropicrin 5.00
11 Chloropicrin 5.00 9 Tetrachloroethylene 5.00
12 Dibromochloromethane 5.00 10 1,1,2-Trichloroethane 44.8
13 Bromochloroacetonitrile 5.00 11 Dichloroacetonitrile 5.00
14 1,2-Dibromoethane (EDB) 5.00 12 Dibromochloromethane 5.00
15 Tetrachloroethylene 5.00 13 1,2-Dibromoethane (EDB) 5.00
16 1,1,1 -Trichloro-2-propanone 5.00 14 1,1,1 -Trichloro-2-propanone 5.00
17 Bromoform 5.00 15 Bromochloroacetonitrile 5.00
18 Dibromoacetonitrile 5.00 16 Bromoform 5.00
19 1,2,3-Trichloropropane 50.0 17 1,2,3-Trichloropropane 50.0
20 1,2-Dibromo-3-chloropropane (DBCP) 5.00 18 Dibromoacetonitrile 5.00
21 Surrogate: decafluorobiphenyl 10.0 19 1,2-Dibromo-3-chloropropane (DBCP) 5.00
22 Hexachlorocyclopentadiene 28.0 20 Surrogate: decafluorobiphenyl 10.0
23 Trifluralin 7.04 21 Hexachlorocyclopentadiene 28.0
24 Simazine 200 22 Trifluralin 7.04
25 Atrazine 200 23 Hexachlorobenzene 1.98
26 Hexachlorobenzene 1.98 24 Atrazine/simazine 400
27 Lindane (γ-BHC) 30.1 25 Lindane (gamma-BHC) 30.1
28 Metribuzin 19.9 26 Heptachlor 5.00
29 Bromacil 30.1 27 Metribuzin 19.9
30 Alachlor 34.9 28 Alachlor 34.9
31 Cyanazine 60.4 29 Metolachlor 70.0
32 Heptachlor 5.00 30 Heptachlor epoxide 14.0
33 Metolachlor 70.0 31 Bromacil 30.1
34 Heptachlor epoxide 14.0 32 Cyanazine 60.4
35 Endrin 5.00 33 Endrin 5.00
36 Endrin aldehyde 7.00 34 Endrin aldehyde 7.00
37 Endrin ketone 4.96 35 Methoxychlor 20.1
38 Methoxychlor 20.1 36 Endrin ketone 4.96
* Bromofluorobenzene as the internal standard was not included in this chromatogram. * Bromofluorobenzene as the internal standard was not included in this chromatogram.
227
Journal of Chromatographic Science, Vol. 35, May 1997
Downloaded from https://academic.oup.com/chromsci/article/35/5/221/339378 by Hanoi School of Public Health user on 16 February 2024
improvement when compared to the results obtained at the ini
tial pH of 6.0, as shown in Figure 3.
At day 21, TCAN recoveries had fallen to 74% and 72% for
reagent and tap water, respectively. Consequently, sample
Figure 3. TCAN degradation (buffered reagent water).
holding times were set at 14 days in method 551.1. All other
method analytes were stable through 21 days
at pH 5.3. Additionally, extract storage
studies were conducted using the day 0
extracts. All analytes reflected recoveries
between 80 and 120% when stored as
extracts in a freezer held at -26°C for 21
days, but a conservative approach was
adopted in the method, and a 14-day holding
time for sample extracts was also defined.
Due to site-specific and seasonal variability
of natural buffering in field samples, the dry
phosphate buffer recommendation in U.S.
EPA method 551.1 is 1.0 g of 1.0% Na HP02 4
228
Journal of Chromatographic Science, Vol. 35, May 1997
These results clearly indicate that samples should not be vial should be weighed, and then, following sample extraction,
transferred from the original sampling vial in which they were the vial should be reweighed. The extracted sample volume is
shipped. The results also show the need to extract the sample determined by the weight difference.
immediately after reducing the sample volume to the 50 mL As an additional precaution to prevent volatile analyte losses,
that is required for extraction. The best procedure for accurately the extraction procedure of method 551 has been improved in
reducing the sample volume to 50 mL is by using precalibrated method 551.1. Method 551 specifies that, after adding NaCl
sampling vials. Vials should be precalibrated in the laboratory to the sample, the salt should be dissolved by "inverting and
before they are used for sampling by accurately measuring 50 shaking the vial vigorously" prior to the addition of the
mL of reagent water with a TD graduated cylinder, pouring extraction solvent methyl-t-butyl ether. It was believed that
Downloaded from https://academic.oup.com/chromsci/article/35/5/221/339378 by Hanoi School of Public Health user on 16 February 2024
this volume into the sampling vial, and scoring the vial at the this could affect the analysis results, due to the high volatility
meniscus. When a sample is received in the vial, the volume is of these method analytes. Therefore, method 551.1 has
reduced using a pipet to the precalibrated mark. If samples are reversed these steps so that the extraction solvent is added prior
received in 60-mL vials that have not been precalibrated, then to the addition of the extraction salt. Then, as the sample is
10 mL of sample should be removed, the remaining sample and extracted, the extraction salt dissolves. The extraction time has
been extended to allow the salt sufficient time to dissolve, which
Table VII. Extraction Salt Comparison Study enables the efficient extraction of all the method analytes.
4 –
10 g NaCl doped with Br as extracting salt with tion salt had the lowest observed brominated organic DBPs.
delayed MtBE extraction; NaCl is fortified with 6.44 mg Slight increases of 12.6 and 6.9% for bromoform and dibro
NaBr(5.0 mg B r ) –
moacetonitrile, respectively, were measured in type-1 samples
when compared to the type-2 samples. However, for the third
†
Chloroform 68.7 1.62 62.4 2.50 68.5 2.06 69.4 0.22
1,1,1-Trichloroethane <0.05 <0.05 - <0.05 - <0.05
- -
Carbontetrachloride <0.05 <0.05 - <0.05 - <0.05
- -
Trichloroacetonitrile 0.053 2.16 0.051 1.96 0.052 3.84 0.059 2.60
Dichloroacetonitrile 6.87 3.08 6.57 1.84 7.25 2.63 7.18 0.81
Bromodichloromethane 21.9 1.82 21.7 2.39 22.1 2.32 22.4 0.68
Trichloroethylene <0.05
- <0.05 - <0.05 - <0.05 -
1,1 -Dichloro-2-propanone 0.385 5.47 0.453 8.13 0.461 7.11 0.476 28.40
Chloropicrin 1.14 2.63 1.24 1.85 1.21 4.25 1.25 0.92
Dibromochloromethane 5.37 2.42 5.50 2.26 5.48 3.22 5.82 0.71
Bromochloroacetonitrile 2.50 4.21 2.39 1.58 2.79 4.13 2.85 1.60
1,2-Dibromoethane <0.05 - <0.05 - <0.05 <0.05 -
-
Tetrachloroethylene <0.05 <0.05 - <0.05 - <0.05 -
-
1,1,1 -Trichloro-2-propanone 4.05 2.24 3.92 0.96 4.09 2.72 4.05 1.40
Bromoform 0.493 4.17 0.438 1.97 0.494 4.04 1.07 8.48
Dibromoacetonitrile 0.449 2.28 0.420 0.72 0.734 7.33 1.97 3.55
1,2-Dibromo-3-chloropropane <0.05 <0.05 - <0.05 <0.05 -
- -
Surrogate recovery (%) 108 1.418 102 0.98 111 1.80 117 3.22
Decafluorobiphenyl
229
Journal of Chromatographic Science, Vol. 35, May 1997
type of samples, in which NaCl was allowed to dissolve and the analysis. By installing a new DB-1 column, these performance
extraction was delayed, bromoform and dibromoacetonitrile problems were overcome. If the columns to be used for
concentrations increased by 12.8 and 74.8%, respectively, over method 551.1 have been used for several years or have had ex
what was observed in the type-2 N a S 0 samples. Type 4 sam
2 4 tended use with extracts from harsh sample matrices (e.g.,
ples, for which Br" was intentionally fortified into the NaCl wastewater, acidified sample extracts, hazardous waste sam
and extraction was delayed, displayed the most dramatic bromi- ples), it may be difficult to meet the criteria established for
nated DBP increases over the type-2 N a S 0 samples; bromo 2 4 this LPC standard, and column replacement may be the best
form and dibromoacetonitrile increased 144 and 369%, alternative. If a laboratory is not conducting analyses for pes
respectively. ticides and herbicides, a modified LPC solution may be pre
Downloaded from https://academic.oup.com/chromsci/article/35/5/221/339378 by Hanoi School of Public Health user on 16 February 2024
These findings support the arguments of Xie and confirm the pared. This modified LPC solution can omit the endrin analyte
potential for increased concentrations of brominated organic breakdown component as well as the resolution requirement
DBPs when NaCl is used as the extraction salt in ammonium- for bromacil and alachlor under column performance. In
chloride-quenched samples. As a result, the use of N a S 0 as 2 4 addition, substitute analytes in place of lindane for the sensi
the extraction salt is recommended in method 551.1. tivity check and hexachlorocyclopentadiene for chromato
graphic performance can be selected. These substitute com
Additional changes in U.S. EPA 551.1 pounds must meet the same acceptance criteria as listed in
To insure the proper performance of the GC system that is Table IX with a concentration for the substitute sensitivity
used to analyze samples, a Laboratory Performance Check check analyte near the minimum detection limit for that
solution (LPC) has been included in method 551.1. This check analyte and the concentration for the substitute chromatog
solution is designed to monitor instrument performance raphic performance analyte at a middle- to high-calibration
based on the parameters of instrument sensitivity, chro level.
matographic performance, column performance, and break In addition, due to concerns over the safety of MtBE as the
down of endrin. These parameters are listed in Table IX along extraction solvent, method 551.1 has specified pentane as an
with the method analytes utilized to perform this evaluation, alternate solvent. This solvent does not extract chloral hydrate
their concentration in MTBE or pentane, and the acceptance from aqueous samples and therefore cannot be used if chloral
criteria. If endrin breakdown exceeds 20%, the problem can hydrate analysis is required. Also, 5.0 mL of pentane rather
most likely be solved by performing routine maintenance on than 3.0 mL of MtBE is used for extraction. This larger volume
the injection port, including replacing the injection port is required as a means to increase the ratio of pentane to water.
sleeve and all associated seals and septa. If column or chro This was required because pentane does not emulsify into the
matographic performance criteria cannot be met, new water as efficiently as MtBE. In addition, due to the high
columns may need to be installed, column flows may need to volatility of pentane, an internal standard must be used with
be corrected, or modifications may need to be adapted to the this extraction solvent. Bromofluorobenzene is recommended,
oven temperature program. During early method develop due to the fact that it is highly unlikely to be observed in any
ment work, significant chromatographic and column perfor field samples.
mance problems were observed while using a DB-1 column Finally, method 551 specified an extracted sample size of 35
that had been used for several years in drinking water extract mL collected in 40-mL vials and a 2.0-mL volume of MtBE
Concentration (μg/mL)
Parameter Analyte in MtBE or pentane Acceptance criteria
where is the peak width at half height, and W(1/10) is the peak width at one tenth height.
†
Resolution between the two peaks as defined by the equation:
where t is the difference in retention times between the two peaks, and W is the average peak width, at the baseline, of the two peaks.
‡
% BD = Percent breakdown. Endrin breakdown calculated using the equation.
230
Journal of Chromatographic Science, Vol. 35, May 1997
extraction solvent. Often there was insufficient space in the been widened. Flexibility has been built into the method in
40-mL sampling vial to add the required 8 g of NaCl and 2.0 mL several ways. For instance, even though the analyte list has
of MtBE once 5.0 mL of aqueous sample was removed. In addi expanded, laboratories are not required to perform analysis for
tion, with this small volume of solvent, it was difficult tofilltwo the total analyte list if their project requirements include only
autosampler vials with the 1.5 mL of solvent that remained select analytes. In addition, a laboratory may choose to use
after sample extraction. Only 1.5 mL of solvent can be recovered pentane as the extraction solvent over MtBE if chloral hydrate
from the extracted sample because a thin layer of extraction sol analysis is not required. A potential problem with NaCl as the
vent remains at the solvent-aqueous layer interface, and MtBE extraction salt has been overcome through the use of Na SO as 2 4
exhibits a small degree of water solubility. Consequently, the the extraction salt. Analyte volatility has been addressed and
Downloaded from https://academic.oup.com/chromsci/article/35/5/221/339378 by Hanoi School of Public Health user on 16 February 2024
volume of sample to be extracted was increased to 50 mL fol precautions to avoid analyte losses have been specified.
lowing collection in a 60 mL vial. This 60-mL vial allowed for
approximately 10 mL of headspace, which permitted sufficient
room for the 20 g of Na SO specified in method 551.1 as the
2 4
231