You are on page 1of 12

European Journal of Orthodontics 15 (1993) 125-135 I 1993 European Orthodontic Society

The tensile bond strength of new and rebonded stainless steel


orthodontic brackets
D. Regan, B. LeMasney, and R. van Noort
School of Clinical Dentistry, University of Sheffield, England

SUMMARY The study investigated the effect on the tensile/peel bond strength of the variables
associated with the bracket base, the enamel surface, and the type of adhesive when both new
and used brackets were rebonded to a previously bonded enamel surface.
The tensile/peel bond strength was firstly evaluated for three different types of stainless steel
orthodontic bracket/base combinations. The cast integral base gave a significantly lower bond
strength than the foil-mesh and photo-etched bases. Following debonding, a group of new
brackets were bonded to the teeth using a chemically-activated or a light-cured adhesive. The old
adhesive had been removed from the enamel by either a hand sealer or a tungsten-carbide bur.
The rebonded new brackets demonstrated a small, but statistically significant fall in bond strength.
No differences were found between the enamel preparations or the adhesives.
A further group of previously debonded brackets were rebonded to the same teeth. The bracket
bases were prepared by either smoothing with a green stone or heating in a bunsen flame followed

Downloaded from by guest on July 2, 2015


by sandblasting and electropolishing. Highly significant falls in bond strength were obtained with
all the bases. No significant differences were found between the two methods of bracket prep-
aration.

Introduction brackets are purchased in 'one-patient' kits


since using a new bracket means breaking into
The prevalence of clinical bond failures should
a new kit. Even if a new bracket is used, the
have been reduced as bracket base design has
bond strength may be affected by residual
progressed. Recent studies have indicated over-
adhesive on the enamel surface. Large aggrega-
all clinical failure rates of 6 per cent for a tions of adhesive, such as may occur around
chemically-cured adhesive (O'Brien et ai, 1989) the periphery of the bracket base, will need to
and 6.5 per cent for a light-cured adhesive be removed from the enamel in order to allow
(Read and O'Brien, 1990). However, wide correct seating of the replacement bracket. Such
variations may be expected depending on the adhesive removal may be performed by hand
bracket type, the adhesive used, the teeth sealers or mechanical instruments (Oliver,
bonded, and a range of operator and patient 1991). However, since routine etching has been
variables. found to penetrate to a mean depth of 28 /*m
Despite these advances, the rebonding of (Legler et ai, 1990), complete removal of all
brackets may still pose a problem due to the adhesive is unlikely.
adoption of pre-adjusted appliances which A further decision to be made is the selection
require inaccurately located brackets to be of an adhesive to be used to rebond the bracket.
repositioned during treatment in order to take Even if the bracket was originally bonded with
full advantage of the archwire slot values and a chemically-cured adhesive, rebonding with a
sliding mechanics (McLaughlin and Bennett, light-cured adhesive would offer the advantage
1991). When rebonding a bracket during treat- of a command set with the ability to tie in the
ment the operator is faced with the choice of archwire immediately.
rebonding the same bracket or using a new When rebonding previously used brackets,
one. these will have varying amounts of adhesive on
Using a new bracket each time increases the the fitting surface. Since the bond at the
cost of providing treatment, particularly if ' bracket/adhesive interface is primarily a
126 D. REGAN ET AL.

mechanical one, and is proportional to the carious, damaged, or malformed specimens


undercuts and surface roughness (Matasa, were rejected. Approximately two-thirds of the
1989), then the amount of remaining adhesive root of the tooth was removed using a band
is likely to affect the bond strength. There are saw in order to allow the teeth to be placed
several possible preparation options regarding centrally in the mounts with the labial surface
the bracket base. The old bracket can be uppermost. Yellow dental stone (Kaffir-D)
rebonded without any preparation, or an poured into prepared silicone moulds, was used
attempt made to remove adhesive remnants as the mounting medium. In order to ensure
either mechanically, or with a simple chairside that the brackets were aligned horizontally to
reconditioning process. Although clinical and the testing assembly, they were bonded before
laboratory studies have been reported on the mounting. This allowed the placement of wires
bond strength of commercially recycled ortho- under the tie wings which could then rest on
dontic brackets (Buchwald, 1989; Regan el ai, the top of the moulds whilst the teeth were
1990), there appears to be little guidance mounted. Before bonding, the enamel surface
regarding the outcome of rebonding brackets was polished for 15 seconds using a rubber
which have become detached during the course cup, and a slurry of pumice and water. Etching
of orthodontic treatment. was performed using a 37 per cent orthophosph-
The purpose of this in vitro study was to oric acid liquid for 45 seconds. The teeth were
investigate the effect on the tensile bond strength then rinsed and dried with oil-free compressed
of the variables associated with the bracket air. The brackets were examined under a
base, the enamel surface, and the type of magnifying lens following bonding and any
adhesive when both new and used brackets are excess adhesive removed using a probe. All
brackets were allowed to bench-cure for 5

Downloaded from by guest on July 2, 2015


rebonded to a previously bonded enamel
surface. minutes and then mounted in the moulds which
were then stored in water at 37°C for 24 hours
before bond testing.
Materials and methods
Bond strength testing
Brackets and adhesives
Tensile/peel bond strength testing was carried
Three different stainless steel bracket/base com- out using a Lloyd Mk 5 machine (Lloyd
binations were chosen to represent types com- Instruments pic, Fareham, Hampshire, Eng-
monly used in clinical practice. The bases land). The tensile load was applied using a
selected were a conventional foil-mesh base (A- 0.020" x 0.025" stainless steel wire assembly
Company, San Diego, California), a cast integ- consisting of two 40 x 10 mm rectangles of wire
ral base (Edgeway, Ortho Organizers Inc., San which were soldered at one end as described
Marcos, California), and a photo-etched base previously (Regan and van Noort, 1989). A
(Microlok, GAC International Inc., Comnak, cross-head speed of 2 mm per minute was used
New York). Pre-adjusted Edgewise premolar and the force required to debond the brackets
brackets with a slot size of 0.022" by 0.028" was recorded in Newtons.
were used. Two different adhesives were
employed in the tests. These were a two-paste, Initial bond strength
highly-filled orthodontic bonding system (Max- The first part of the investigation consisted of
imum, Ortho Organizers Inc., San Marcos, bonding 180 new brackets to the teeth using
California) and a light-cured, highly-filled the two-paste adhesive (Group I). This group
orthodontic adhesive (Transbond, Unitek Cor- acted as a control and provided the brackets
poration/3M, Monrovia, California). for rebonding to the same enamel surfaces.
Following debonding, the bond strength was
Bonding recorded, and all teeth and bracket bases were
The bases were bonded to extracted premolar examined under a magnifying lens in order to
teeth which had been stored in 70 per cent assess the amount of adhesive remaining on
ethyl alcohol. All brackets and teeth were the tooth and the mode of bond failure. The
allocated a number so that a bracket could be enamel surfaces were classified according to the
identified and rebonded to the original tooth. adhesive remnant index of Artun and Bergland
The teeth were carefully examined and any (1984). -
REBONDING STAINLESS STEEL BRACKETS 127

Score 0 = no adhesive left on the tooth. Two main methods of bracket base preparation
Score 1 = less than half of the adhesive left on were used before rebonding. Before rebonding,
the tooth. 20 of each type of base were prepared by
removing adhesive from the base using a green
Score 2 = more than half of the adhesive left
stone in a slow speed handpiece. A further 20
on the tooth.
of each type were prepared according to the
Score 3= all adhesive left on the tooth, with method described by Buchman (1980). First,
a distinct impression of the bracket the bracket was held in a bunsen flame for 3
base. seconds in order to burn off the acrylic. This
produces a temperature of approximately
Rebonding new brackets 1200°C. The bracket was then quenched in
After bond testing, 40 new Microlok brackets room temperature water and cleaned in a
were bonded to the teeth from which brackets laboratory sandblaster for 5 seconds in order
had been debonded (Group 2). For 20 of these to remove inorganic filler particles. Finally, the
teeth, the adhesive was removed from the bracket was electropolished for 20 seconds by
previously bonded enamel surface with a tung- connecting the alligator clip of a Rocky Moun-
sten-carbide bur in a slow speed handpiece. tain 700 electropolisher (Rocky Mountain
The remainder were prepared by using a hand Orthodontics, Denver, Colorado) to a wire
sealer (Ash osteo No. 4, Dentsply Ltd., Glouces- ligature holding the bracket. In order to
ter). Equal numbers of the new brackets were differentiate the effects of flaming and sandblast-
bonded with either the same two-paste adhesive ing from the electropolishing process, a group
or a light-cured adhesive. This gave four groups of 20 Microlok brackets were rebonded follow-
of tooth preparation/adhesive combinations.

Downloaded from by guest on July 2, 2015


ing flaming only.
These brackets were again debonded on the
The groups of brackets were randomly
Lloyd testing machine and the tensile/peel bond
divided so that half of the brackets were bonded
strength recorded. The main aim of this part
with a two-paste adhesive and the rest with a
of the investigation was to compare two
light-cured adhesive. Tensile/peel bond testing
common methods of tooth preparation used
clinically. If no significant difference was found was performed as for the previous groups.
then the simplest method of tooth preparation Samples of teeth and bases were selected for
could be used for the remaining rebond tests, examination under the scanning electron micro-
thus reducing the number of specimens required. scope. The area of each base was estimated
from electron micrographs by tracing the photo-
Rebonding used brackets micrograph on a Bit-Pad Two digitizer (Summa-
graphics Corporation, Fairfield, Connecticut)
The final part of the study was concerned with
connected to an AT IBM compatible personal
the variables associated with the rebonded
computer. In addition to tracing the peripheries
bracket bases. Bracket failure could occur at
of the bases, an estimate was made of the
one of several interfaces, each of which would
leave a different amount of adhesive remaining available cohesive bonding area of each base
on the bracket base and tooth surface. by tracing the undercut features.
In order to ensure a similar distribution of The data were analysed statistically using an
sites of bond failure in each group, both the analysis of variance (ANOVA) and any signi-
brackets and teeth were examined for adhesive ficant differences revealed by this procedure
remnants. Brackets were selected for rebonding were further investigated using the Scheffe test
in order to ensure that the distribution of the with a 95 per cent confidence limit (/ ) <0.05).
adhesive remnant index values seen in the group This phased experimental design explored the
as a whole was reflected in each of the sub- effect on the tensile bond strength of the
groups. Group 3 consisted of 40 each of the important variables associated with the original
A-Company and Edgeway brackets together bracket bases, the preparation of the formerly
with 60 Microlok brackets. bonded enamel surface, and the rebonding of
In all cases, the enamel was prepared by both new and previously used brackets. At the
removal of adhesive with a hand sealer before same time, it avoided needless duplication of
final polishing with a rubber cup and pumice. testing procedures.
128 D. REGAN ET AL.

Results Table 1 Means and standard deviations of rebonded


brackets (Newtons).
Initial bond strength (Group 1)
The initial mean tensile/peel bond strengths of Bracket Tooth Bracket Adhesive Mean SD
prep prep
the three brackets together with their standard
deviations are illustrated graphically in Fig. 1. A-Co sealer gs lc 57.71 25.20
The highest bond strength was obtained with 2p 77.12 27.11
the foil-mesh A-Company bracket and the f/sb lc 73.25 52.01
lowest with the cast integral Edgeway bracket. 2p 63.25 42.72
The results of the initial analysis of variance, Edgeway sealer lc 33.68 19.10
gs
using the bond strength as the dependant 2p 28.87 16.55
variable, showed a significant difference between f/sb lc 25.23 13.27
the groups (P< 0.0001). The Scheffe multiple 2p 33.41 9.97
range test demonstrated that there was no
Microlok tc bur new lc 66.70 40.81
significant difference between the bond strength 98.34 22.82
2p
of the A-Company and Microlok brackets, but sealer new lc 100.86 34.92
that the bond strength of the Edgeway bracket 2p 87.03 31.54
was significantly poorer than both of the other gs lc 26.67 13.53
two brackets (P<0.05). 2p 49.54 25.29
f lc 21.69 14.30
2p 32.98 25.80
Rebonding new brackets (Group 2) f/sb lc 39.38 9.82
The mean tensile/peel rebond strengths of all 2p 57.99 36.32

Downloaded from by guest on July 2, 2015


the rebonded brackets following the different Key: new = new, unused bracket; gs = green stone; f/sb =
tooth preparations, bracket preparations, and flame and then sandblasted; f= flamed only; lc = light-cured
adhesives are given in Table 1 together with adhesive; 2p = two-paste adhesive.
their standard deviations.
The new Microlok brackets rebonded to a between the two groups prepared by hand
previously bonded enamel surface showed a scaling or a tungsten-carbide bur. In view of
reduction in bond strength (Fig. 2). This this finding, the other types of brackets were
amounted to a mean 19.6 per cent reduction only rebonded to teeth prepared with a sealer
in tensile/peel bond strength and was statistic- to remove excess adhesive from the enamel
ally significant at the P<0.05 level. surface.
The teeth to be rebonded with these new
Microlok brackets had received two types of Rebonding used brackets (Group 3)
preparation. No significant difference was found The results in Table 1 also showed that the
differences between the light-cured and the two-
paste adhesives were not consistent, and that
"» 160-,
c each performed better than the other with
o
I 140 certain bracket and tooth preparation combina-
o tions. No statistical difference was found
S- 120^ 113 .6 109 .7 between the two adhesives in any case.
O)
c 100 Because the brackets and teeth were num-
£ bered, it was possible to compare the perform-
« 80
•o ance of the same bracket/tooth combinations
c 60. 1
o 60 when bonded and then rebonded. The fall in
n
«
bond strength of the different groups is shown
a 40 graphically in Fig. 3. Since there was no signi-
Q.
ficant difference between them, the results have
20
C been combined for the two adhesives. This
o 0 shows that all brackets gave a lower mean
A-Company Edgeway Microlok tensile/peel bond strength on rebonding. The
Figure 1 Mean tensile/peel bond strengths of the original percentage reduction was 41.4 per cent for the
new brackets (in Newtons). A-Company bases, 47.6 per cent for the Edge-
REBONDING STAINLESS STEEL BRACKETS 129

120
109.7

rength (1Newtoi
98.3 100.9
100
87 2-paste
80 light-cured
66.7
w 60
•a
ensi le/peel bor

40

20

0
Original Tungsten-carbide Sealer
Figure 2 Mean tensile/peel bond strength (in Newtons) of the new Microlok brackets rebonded to a debonded enamel
surface. The original (i.e the control) bond strength is shown for comparison.

Downloaded from by guest on July 2, 2015


Original
Rebond

200 150 100 50 0 50 100 150 200

Tensile/peel bond strength (Newtons)


Figure 3 Mean tensile/peel bond strengths (in Newtons) of brackets debonded and rebonded to the same teeth. Key: gs,
prepared with green stone; f/sb, flamed and then sandblasted; f, flamed only.

way bases and 64.8 per cent for the Microlok strength of new Microlok brackets bonded to
bases. These differences between original and previously bonded enamel was significantly
rebond tensile/peel bond strength were tested higher than any of the used brackets (P<0.05).
statistically using a paired /-test and found to
be significant at the P< 0.001 level. It is Adhesive remnant index (ARI) distribution
interesting to note that despite the reduction The main purpose of recording the adhesive
in their bond strength, the used A-Company remnant index following the initial debonding
bases were still stronger than the original new procedure was to allow the selection of compar-
Edgeway bases. The differences between the able sub-groups for the subsequent tests. The
types of bracket preparation were not statistic- percentage frequency distribution of the ARI
ally significant, although it can be seen that is given in Table 2. This shows a broadly similar
just flaming the Microlok brackets results in a distribution amongst the three types of bracket
very low rebond strength. The tensile/peel bond base with most of the bases leaving over 50
130 D. REGAN ET AL.

Table 2 Percentage frequency distribution of the between the various types (Fig. 4). The bases
adhesive remnant index (ARI) for each type of new were classified into two groups; those with less
bracket following initial debonding. than 50 per cent of the undercuts filled with
adhesive and those with more than 50 per cent
Bracket ARI = 0 ARI=l ARI = 2 ARI = 3 of the undercuts filled with adhesive. It was
A-Company 2.4 19.0 27.4 51.2
found that 49.2 per cent of the Edgeway bases
Edgeway 0 18.8 34.4 46.8 had less than 50 per cent of the undercut areas
Microlok 1.8 15.5 48.2 34.5 filled with fractured adhesive whereas no
A-Company and only 4.6 per cent of Microlok
bases failed in such a manner. It appeared that
per cent of the adhesive on the enamel surface the mesiodistal horizontal channels provided
following removal in the tensile/peel mode. the main undercut areas in these premolar
Since all the teeth and brackets could be Edgeway bases, thus reducing the effective
identified at each stage of the experiment, it undercut to 2.98 mm2 (i.e. the area of these
channels as measured on the electron micro-
was possible to investigate the effect of the graphs). A particular problem noticed with the
ARI on both the initial and rebond strength Microlok bases was the presence of voids due
using the analysis of variance procedure. No to air entrapment. This would also tend to
significant differences were found between the reduce the efficiency of the undercuts.
ARI value and either the initial bond strength
or the rebond strength for each of the bracket The difference in these results compared with
bases. earlier data illustrates the danger of basing
recommendations on the outcome of testing
one type of bracket only. It is difficult to

Downloaded from by guest on July 2, 2015


Discussion extrapolate from the laboratory to the clinical
The original bond strength results demonstrate situation on the basis of these results since the
a significantly lower tensile/peel bond strength brackets were removed in the tensile/peel mode
for the Edgeway cast base when compared with rather than a shear/peel mode as would occur
both the A-Company mesh base and the with occlusal forces. Both the A-Company and
Microlok brackets should perform adequately,
Microlok photo-etched base. This contradicts but there may be some reservations about the
two previous investigations involving the Edge- Edgeway base.
way base (Regan and van Noort, 1989; Regan
et al., 1990). The reasons for this could be Following removal of a bracket, several
related to the bases, the substrates or the methods have been described for eliminating
operators. In the earlier studies, flat upper the residual adhesive from the enamel surface.
central incisor brackets were bonded to a flat Hand sealers are useful for large areas of
adhesive but may gouge the tooth surface
composite substrate. The present research used (Burapavong et al., 1978). Rotary instruments
curved premolar brackets bonded to extracted such as green stones and diamond burs may
teeth. It is possible that the adaptation to the also produce serious enamel destruction (Retief
substrate and the undercut pattern is more and Denys, 1979). Ultrasonic sealers are efficient
efficient on the flat brackets. The mean area of at composite removal, but the amount of severe
the Edgeway base (9.83 mm2) was found to be tooth surface damage is unacceptable (Walmsley
slightly smaller than the Microlok base et al., 1989). Zachrisson and Artun (1979)
(10.36 mm2) and the A-Company base found that no method resulted in complete
(10.29 mm2). A more important factor is the adhesive removal but that a tungsten-carbide
area of the available undercut, since this will bur used at a low speed produced the best
determine the surface area of the adhesive results. Other methods suggested have been a
which may fail in a cohesive mode. Again, the mechanical band driver with a chisel (Dragiff,
differences between the bases were minimal. 1979; Gait, 1984) or a reciprocating handpiece
The Microlok base has a slightly larger undercut designed for finishing composite restorations
area (5.62 mm2) than the Edgeway (5.50 mm2) (Oliver, 1991).
and A-Company (5.12 mm2) bases. However, Whatever method is employed, it would
when the debonded bases were examined, a appear that complete removal of all adhesive
difference was found in the mode of failure is unlikely and a significant number of ortho-
REBONDING STAINLESS STEEL BRACKETS 131

dontic patients require the removal of stain This retention of adhesive on the enamel
which has formed on resin still present even surface was confirmed by this study. Figure 5
years after orthodontic treatment (Brown, reveals an enamel surface following bracket
1988). debonding. Etching had produced the typical
Type I etching pattern where the enamel rod
centres are preferentially etched (Ten Cate,
1989). The centres of the rods are now filled
with resin tags. Figure 6 shows an area next to
one of these islands following the application
of an etching liquid. It appears that the etching
procedure has no effect on the areas with
adhesive still present. This may explain the
reduction in bond strength which occurs even
when new brackets are used since there will be
a reduction in the amount of mechanical
retention available on the enamel surface.
Although this fall in bond strength with new
brackets was statistically significant it is unlikely
to be of clinical importance since the rebond
strength still falls within the value regarded as
necessary to withstand orthodontic and occlusal
forces (Reynolds, 1975). In similar studies with
new brackets, Faust et al. (1978) found that

Downloaded from by guest on July 2, 2015


Figure 5 Enamel surface following debonding. The etched
enamel prism centres are now filled with adhesive. (Original
magnification x 640.)

Figure 4 Scanning electron micrographs of the bracket


bases following initial debonding. (a) A-Company base.
Most of the available undercuts are filled with adhesive and
the fracture interface is at the level of the wire mesh
intersections. (Original magnification x 20, Scale bars =
100 /im.) (b) Edgeway base. The bulk of the adhesive is
restricted to the deeper horizontal channels. (Original mag-
nification x40. Scale bars= 100 jim.) (c) Microlok base. In
the centre of the base, the undercuts have only been partially
filled with adhesive. (Original magnification x 20, Scale
bars = 100 /mi.)
132 D. REGAN ET AL.

Figure 6 Enamel surface following debonding and re-etching. Next to the retained adhesive (A) are resin tags (RT) from
the original bonding. The acid-etching procedure has no effect on composite.

Downloaded from by guest on July 2, 2015


the differences among adhesives or between reduced when compared to a single polymerized
brackets were more dramatic than bond and unit (Reisbeck and Brodsky, 1971). The factors
rebond strengths. Jassem et al. (1981) did not which may affect the bond between an old and
obtain any significant difference when rebond- new composite include the ability to wet the
ing new foil-mesh brackets with a chemically- surface of the material, the amount and chem-
activated adhesive. ical activity of available resin for bonding and
The amount of adhesive remaining on the the effect of surface treatment prior to repair
enamel surface following debonding was not (Saunders, 1990). Polishing the surface reduces
found to be significantly related to either the the bond strength by reducing chemically
initial bond strength or the rebond strength. reactive groups (Davidson et al., 1981) and
Although the quantity of adhesive was estim- producing surface debris (Miranda et al., 1984).
ated in this study by using a subjective index The age of the repaired composite also affects
(the ARI), the lack of relationship confirms the the bond strength, with the greatest reduction
findings of O'Brien et al. (1988) who performed in reactivity occurring during the first 24 hours
a more rigorous analysis of the residual debris (Soderholm and Roberts, 1991). It has also
following debonding. been found that composite stored in water has
When a used bracket is to be rebonded, there a lower bonding potential than when stored
is the additional problem of adhesive on the dry (Causton, 1975). This would indicate that
bracket base. This can vary from a reduction a bracket which had been removed for reposi-
in the available undercuts to complete oblitera- tioning and then rebonded immediately after
tion if the fracture has occurred at the enamel/ could be expected to have a better bond strength
adhesive interface. than one which had been in the oral environ-
The base may simply be prepared by ment for some time.
smoothing any projections with a green stone Commercial bracket reconditioning processes
without attempting to remove the adhesive use either heat or chemical solvents. Neither of
from the undercuts. In some cases this may these are practical to perform fully at the
leave a composite surface devoid of undercuts chairside, although a limited heating procedure
for rebonding to the enamel (Fig. 7). The may be used in an attempt to burn off the
strength of old composite restorations repaired adhesive. Buchman (1980) found evidence of
with fresh material has been found to be greatly significant annealing and carbide separation
REBONDING STAINLESS STEEL BRACKETS 133

when using a bunsen flame to remove adhesive experienced (Matasa, 1989). A stainless steel
from stainless steel bracket bases. In addition, bracket subjected to high temperatures also
further problems such as clogging of the base takes on an unacceptable brown appearance
and obstruction of undercuts and slots may be and requires repolishing using an electropol-
ishing process. This may result in significant
loss of metal and alteration in the undercuts
(Fig. 8). The present study found no difference
between any of the bracket preparation pro-
cesses and the use of heat cannot be recom-
mended.
Zachrisson (1985) recommended a 3-minute
rebonding procedure involving removal of
adhesive from the enamel surface using a sealer
or tungsten-carbide bur. If a new bracket is
not to be used, the adhesive remaining on the
loose bracket is removed with a bur or merely
roughened. Following a 60-second etch, the
bracket is then bonded with a chemically-cured
heavily-filled adhesive. From the present study,
(a) the suitability of this technique has been
confirmed for the foil-mesh brackets. However,
with both the photo-etched and cast bases, the

Downloaded from by guest on July 2, 2015


resulting bond strength may well be unac-
ceptable. In these cases it would be advisable
to use a new bracket. Since no difference was
found between the types of adhesive used in
this study, there would appear to be some
advantage in employing a light-cured adhesive
for rebonding since it offers a generous working
time and a command set which allows brackets
to be positioned and excess material to be
easily removed (Read, 1984).
Extrapolation of laboratory data to the
clinical situation should always be done with
care. However, the findings of this investigation
do seem to be borne out by a clinical investi-
gation of Kinch et al. (1988). First time bonds
were found to have a failure rate of 12.6 per
cent, but subsequent bonds failed at a greatly
increased rate of 23.1 per cent. In addition,
they tended to fail fairly rapidly within the first
14 weeks following rebonding.

Figure 7 Debonded brackets prepared with' a green stone,


(a) A-Company base. This has left a uniformly flat surface
of adhesive. (Original magnification x20, Scale bars =
100 ftm.) (b) Edgeway base. The adhesive has been ground
flat producing a mechanically non-retentive surface. The
studs have also been flattened leaving sharp edges which
could cause stress concentration. (Original magnification
x20, Scale bars= 100/im.) (c) Microlok base. The sur-
face is deeply scratched and few undercuts remain patent.
(Original magnification x 20, Scale bars= 100 /«n.)
134 D. REGAN ET AL.

Conclusions than A-Company foil-mesh and Microlok


photo-etched stainless steel bases.
1. The integral cast base (Edgeway) gave a 2. Rebonding new brackets to an enamel
significantly lower tensile/peel bond strength surface which had been previously etched
and bonded resulted in a statistically signi-
ficant fall in tensile/peel bond strength. This
reduction was not considered sufficient to
be important clinically since the brackets
should still have an adequate bond strength
to withstand normal orthodontic and occlu-
sal forces.
3. There was no difference in tensile/peel bond
strength following removal of the adhesive
from the enamel surface with either a hand
sealer or a tungsten-carbide bur prior to
rebonding.
4. Rebonding previously used brackets resulted
in a significant fall in tensile/bond strength.
There was no difference between the results
obtained following bracket base preparation
with a green stone and a more complicated

Downloaded from by guest on July 2, 2015


process involving heating in a bunsen flame,
sandblasting and electropolishing. The fall
in bond strength was considered to be
important clinically and certain types of
bracket base should not be reused.
5. No difference was found in rebond strength
when using a chemically-cured adhesive and
a light-cured adhesive.

Address for correspondence


Mr D Regan
Department of Child Dental Health
School of Clinical Dentistry
University of Sheffield
Sheffield S10 2SZ
England

Figure 8 Debonded brackets prepared by flaming, sand-


blasting and electropolishing. (a) A-Company base. There
has been an appreciable loss of metal from the foil-mesh.
(Original magnification x20, Scale bars = 100/im.) (b)
Edgeway base. The base has lost the rough cast appearance
and the studs have been abraded so that they have become
wedge-shaped. (Original magnification x40, Scale bars =
100 urn.) (c) Microlok base. The undercut features have
become more rounded and some contain acrylic ash and
silica particles. (Original magnification x 20, Scale bars =
100 /mi.)
REBONDING STAINLESS STEEL BRACKETS 135

Acknowledgements McLaughlin R P, Bennett J C 1991 Finishing and detailing


with a preadjusted appliance system. Journal of Clinical
The authors would like to thank Mr Orthodontics 25: 251-264
M. Debenham of Precision Orthodontics who Miranda F J, Duncanson M G, Dilts W E 1984 Interfacial
very kindly supplied the brackets and adhesives bonding strengths of paired composite systems. Journal
used in this study. of Prosthetic Dentistry 51: 29-32
O'Brien K D, Read M J F, Sandison R J, Roberts C T
1989 A visible light-activated direct-bonding material: an
References in vivo comparative study. American Journal of Ortho-
Artun J, Bergland S 1984 Clinical trials with crystal growth dontics and Dentofacial Orthopedics 95: 348-351
conditioning as an alternative to acid-etch enamel pre- O'Brien K D, Watts D C, Read M J F 1988 Residual debris
treatment. American Journal of Orthodontics 85: and bond strength—Is there a relationship? American
333-340 Journal of Orthodontics and Dentofacial Orthopedics 94:
Brown D 1988 Orthodontic materials update: 1. Ortho- 222-230
dontic bonding adhesives. British Journal of Orthodontics Oliver R G 1991 A new instrument for debonding clean-
15: 277-278 up. Journal of Clinical Orthodontics 25: 407-410
Buchman D J L 1980 Effects of recycling on metallic direct- Read M J F 1984 The bonding of orthodontic attachments
bond orthodontic brackets. American Journal of Ortho- using a visible light cured adhesive. British Journal of
dontics 77: 654-668 Orthodontics 11: 16-20
Buchwald A 1989 A three-cycle in vivo evaluation of recon-
ditioned direct-bonding brackets. American Journal of Read M J F, O'Brien K D 1990 A clinical trial of an
Orthodontics and Dentofacial Orthopedics 95: 352-354 indirect bonding technique with a visible light-cured
adhesive. American Journal of Orthodontics and Dento-
Burapavong V, Marshall C W, Apfel D A, Perry H T 1978 facial Orthopedics 98: 259-262
Enamel surface characteristics on removal of bonded
orthodontic brackets. American Journal of Orthodontics Regan D, van Noort R 1989 Bond strengths of two integral

Downloaded from by guest on July 2, 2015


74: 176-187 bracket-base combinations: an in vitro comparison with
foil-mesh. European Journal of Orthodontics 11: 144-153
Causton B E 1975 Repair of abraded composite fillings.
An in vitro study. British Dental Journal 139: 286-288 Regan D, van Noort R, O'Keeffe C 1990 The effects of
recycling on the tensile bond strength of new and clinically
Davidson C L, Duysters P P E, De Lange C, Bausch J R used stainless steel orthodontic brackets: an in vitro study.
1981 Structural changes in composite surface material British Journal of Orthodontics 17: 137-145
after dry polishing. Journal of Oral Rehabilitation 8:
Reisbick M H, Brodsky J F 1971 Strength parameters of
431-439 composite resins. Journal of Prosthetic Dentistry 26:
Dragiff D A 1979 A new debonding procedure. Journal of 178-185
Clinical Orthodontics 13: 107-111
Faust J B, Grego G N, Fan P L, Poers J M 1978 Penetration Retief D H, Denys F R 1979 Finishing of enamel surfaces
coefficient, tensile strength and bond strength of thirteen after debonding orthodontic attachments. Angle Ortho-
direct bonding cements. American Journal of Orthodont- dontist 49: 1-10
ics 73: 512-525 Reynolds I R 1975 A review of direct orthodontic bonding.
Gait C 1984 Use of an automatic handpiece for debonding British Journal of Orthodontics 2: 171-178
of composite resin. European Journal of Orthodontics 6:
254-256 Saunders W P 1990 Effect of fatigue upon the interfacial
bond strength of repaired composite resins. Journal of
Jassem H A, Retief D H, Jamison H C 1981 Tensile and Dentistry 18: 158-162
shear strengths of bonded and rebonded orthodontic
attachments. American Journal of Orthodontics 79: Soderholm K J, Roberts M J 1991 Variables influencing
661-668 the repair strength of dental composites. Scandinavian
Kinch A P, Taylor H, Warltier R, Oliver R G, Newcombe Journal of Dental Research 99: 173-180
R G 1988 A clinical trial comparing the failure rates of Ten Cate A R 1989 Oral histology: development structure
directly bonded brackets using etch times of 15 or 60 and function. C V Mosby Company, St Louis
seconds. American Journal of Orthodontics and Dento-
facial Orthopedics 94: 476-483 Walmsley A D, Jones P A, Hullah W 1989 Ultrasonic
debonding of composite-retained restorations. British
Legler L R, Reteif, D H, Bradley, E L 1990 Effects of Dental Journal 166: 290-294
phosphoric acid concentration and etch duration on
enamel depth of etch: An in vitro study. American Journal Zachrisson B U 1985 Bonding in orthodontics. In: Graber
of Orthodontics and Dentofacial Orthopedics 98: T M, Swain B F (eds) Orthodontics, current principles
154-160 and techniques. C V Mosby Company, St Louis
Matasa C G 1989 Pros and cons of the reuse of direct- Zachrisson B U, Artun J 1979 Enamel surface appearance
bonded appliances. American Journal of Orthodontics after various debonding techniques American Journal of
and Dentofacial Orthopedics 96: 72-76 Orthodontics 75: 121-137
Downloaded from by guest on July 2, 2015

You might also like