ADAMIITM-CD34 has a high correlation (R2 > 0.95) with flow cytometry instruments, and has low site-to-site variation and
high reproducibility. ADAMIITM-CD34 is cleared for use with MPB and HPC-A by FDA 510(k) and CE.
Add sample and Add RBC lysis buffer Load the prepared Insert the slide into Get result
reagents to a test and incubate for sample into ADAMIITM ADAMIITM- CD34.
tube and incubate 10 minutes. assay slide.
for 20 minutes.
1500 125
Site1
Site2
ADAMII viable CD34 (cells/µL)
100 Site3
CD34 count (cells/µL)
1000
75
50
500
Allowable difference
±10%
25
Passing-Bablok fit
(y = 0.2715 + 1.002 x)
0 0
0 500 1000 1500 Low Medium High
Specification
Light source 1 brightfield, 3 fluorescences Loading volume 25 µL
Camera High-sensitivity monochrome CCD Stage Automated X-Y-Z stage
Sample type HPC-A, Mobilized PB, Control, *Cord blood Dimension 300 x 420 x 370 mm (L x W x H)
Total frame 75 frames Weight 19.3 kg
*CE certified