You are on page 1of 5

E ORIGINAL ARTICLE

Shoulder & Elbow


0(0) 1–5
! The Author(s) 2018
Neuroanatomical distribution of Reprints and permissions:
sagepub.co.uk/journalsPermissions.nav
sensory receptors in the human DOI: 10.1177/1758573218760245
journals.sagepub.com/home/sel
elbow joint capsule

Erica Kholinne1,3, Hyun-Joo Lee2, Maria F Deslivia1,


Kim Ga Yeong3, Suk-Joong Lee4, Sungjoon Lim3
and In-Ho Jeon3

Abstract
Background: The topographic arrangement of sensory receptors in the human elbow joint capsule is pertinent to their
role in the transmission of neural signals. The signals from stimuli in the joint are concisely delivered via afferent pathways
to allow recognition of pain and proprioception. Sensory receptors in the elbow joint include mechanoreceptors and
free nerve endings acting as nociceptors, although the distribution of each of the structures has not been determined,
despite their importance for the integrity of the joint. We therefore aimed to investigate the neuroanatomical distri-
bution and densities of mechanoreceptors and free nerve endings in the capsule of the elbow, at the same time as
considering surgical approaches that would result in the minimum insult to them.
Methods: Four elbow joint capsules were harvested from fresh cadavers. The specimens were carefully separated from
adjacent osteoligamentous attachments and the capsular complex was stained with a modified gold chloride method.
Evaluations of free nerve endings, and Golgi, Ruffini and Pacinian corpuscles were performed under an inverted light
microscope. The number and density of each structure were recorded.
Results: Ruffini corpuscles observed to be the dominant mechanoreceptor type. No Golgi corpuscle was observed.
Free nerve endings were found at the highest density at posterodistal sites, whereas mechanoreceptors were most
frequent at bony attachment sites.
Conclusions: A consistent distribution pattern of articular sensory receptors was observed, which allows further
understanding of elbow pathology. An awareness of the neuroanatomical distribution of sensory receptors in the
elbow joint capsule may allow their preservation during surgical procedures for elbow joint pathology.

Keywords
capsule, elbow, distribution, neuroanatomical, sensory receptors
Date received: 4th October 2017; revised: 6th December 2017; accepted: 19th January 2018

Introduction 1
Department of Orthopedic Surgery, St Carolus Hospital, Jakarta,
The elbow joint is stable as a result of its native gyn- Indonesia
2
Department of Orthopedic Surgery, Kyungpook National University
glymoid shape.1 However, the elbow capsule and its
Hospital, Daegu, Republic of Korea
ligamentous structure have been described as contribut- 3
Department of Orthopedic Surgery, Asan Medical Center, University of
ing to elbow joint stability.2 Mechanoreceptors in tissue Ulsan, Seoul, Republic of Korea
4
sense proprioception and transmit afferent signals as a Department of Orthopedic Surgery, Keimyung University, Dongsan
protective reflex against instability.3–6 The role of col- Medical Center, Daegu, Korea
lateral ligaments in the stability of the elbow has been
Corresponding author:
well studied;7 however, there nevertheless remains a In-Ho Jeon, Department of Orthopedic Surgery, University of Ulsan,
paucity of research on the elbow capsule. A previous Asan Medical Center, 86 Asanbyeongwon-gil, Songpa-gu, Seoul, Korea.
anatomical study has demonstrated that oblique and Email: jeonchoi@gmail.com
2 E Shoulder & Elbow 0(0)

transverse fibrous bands exist in the anterior side of the through a filtration process. The gold chloride solution
elbow capsule.8 In the clinical perspective, the capsule was subsequently removed, and the specimens were
of the elbow joint plays a major part in arthrofibrosis soaked with 2.5% formic acid solution and shaken
cases.9 The intrinsic contractures present in such cases for at least another 12 hours. Following this shaking,
have led to the identification of the elbow capsule as the the specimens were washed three times with running
main site of pathology. Recently, the elbow capsule was distilled water for 5 minutes to remove the solution.
observed to contribute to painful elbow conditions such Each specimen was then transported to a conical tube
as tendinopathy of the common extensor tendon and, prefilled with 30% sucrose solution and stored in a
in addition, innervation of the undersurface of the refrigerator at 4 C for 1 day to 2 days. After a specimen
elbow capsule may generate pain in common extensor had sunk to the bottom of the conical tube, it was
tendon malady.10 In the present pilot study, which transferred to a new vial containing 30% sucrose and
aimed to provide an increased understanding of the optimum cutting temperature compound (OCT). These
elbow capsule, we observed the microanatomy of four vials were then shaken with the penetron shaker for 2
fresh-frozen cadaveric elbows for the existence and dis- hours. Once this process was completed, the stained
tribution of mechanoreceptors and free nerve endings. specimens were embedded in 30% sucrose and OCT
Our hypotheses were: (i) sensory receptors exist in the compound in a 3 : 2 ratio and then frozen, in accord-
human elbow capsule and (ii) the density of sensory ance with a previously reported technique.12 A cryo-
receptors varies according to location within the sectioning machine (Leica CM3050-S Research
elbow capsule. Cryostat; Leica Biosystems, Nussloch, Germany) was
then used to section the frozen specimens parallel to
their longitudinal (horizontal) axis and perpendicular
Materials and methods to their vertical axis, with 30 -mm thick slices being
attached to glass microscope slides.
Cadaver dissection and specimen preparation
The capsules of four elbow joints were harvested from
fresh-frozen cadavers. All were males and had a mean
Microscopic examination
age of 69 years. No history of previous elbow injury An inverted light microscope was used for observation
was found on the medical records. The elbow capsules of mechanoreceptors. The Freeman and Wyke classifi-
were harvested within 12 hours of the cadavers being cation with the modification of Hagert et al. was used
thawed at room temperature. Dissection was carefully as a reference for evaluating the types and numbers of
performed, taking care to obtain the entire capsule por- Golgi, Ruffini, and Pacinian corpuscles.6,13,14 The slides
tions, without causing damage to them. The attachment were first visualized under low power magnification
to the bone was peeled off. After dissection, the speci- ( 100) and then at higher magnification ( 200) to
mens were immersed in 4% paraformaldehyde solution identify each receptor. During inspection of the mech-
with a neutral pH. After fixation in the solution for at anoreceptor structures, checks were made on previous
least 24 hours, the specimens were divided into regions and lateral serial slides to determine whether the struc-
of interest according to their location (anterior and pos- ture was continuously visible. If a structure was con-
terior) and bony attachment site (proximal, distal bony firmed, it was counted as one unit. Discontinued objects
attachment and mid-substance). with uncertain or doubtful morphology were not
counted. We reconfirmed each structure using  400
magnification to minimize any misreading or poten-
Modified gold chloride staining tial bias that might alter the quantification.
The O’Connor and Gonzales modified gold chloride Mechanoreceptors were evaluated and recorded in
staining method was applied to the specimens.11 Vials accordance with their location (anterior and posterior)
consisting of fresh lemon juice and 98% formic acid and bony attachment site (proximal, distal bony attach-
solution in a 3 : 1 ratio were prepared and shaken ment and mid-substance).
with a penetron shaker (Penetron Swirling Shaker
Model Mark IV; SPI Supplies, Sunkay Laboratories,
Tokyo, Japan) inside a fume hood for 30 minutes.
Density calculation
Gold chloride solution (Gold chloride solution To measure the volume of region of interest, custo-
200 mg/dL in deionized water, HT1004; Sigma- mized software was used to automatically measure the
Aldrich, St Louis, MO, USA) was then added to the area on the slide. This customized software was design
vials and they were shaken for a further 90 minutes. based on object detection and recognition. Software
This process was repeated for later batches after recy- was written in OpenCV-Python, version 2.7 (Python
cling the gold chloride solution from the previous batch Software Foundation, Reston, VA, USA).15 The area
E Kholinne et al. 3

was then multiplied by 30 mm (the slice thickness) to by loose connective synovial tissue and medium to large
estimate the total volume. In this manner, the total sized vessels.
volume of each compartment was calculated. The dens-
ity of mechanoreceptors was defined as the number of
Density of mechanoreceptors and free nerve endings
mechanoreceptors divided by the unit volume (number/
cm3) of tissue. Density comparisons were made between The mean densities of mechanoreceptors and free nerve
the anterior and posterior sites, as well as between the endings grouped according to their location and attach-
proximal, distal bony attachment and mid-substance ment site are provided in Table 1.
sites.
Discussion
Results
As a gynglymoid hinged joint, the elbow is stable
We observed Ruffini and Pacinian corpuscles (Figures 1 according to its native structure. Bony and soft tissue
and 2) and free nerve endings (Figure 3) within the constraints form the primary elbow joint stabilizers,
capsules. These sensory receptors were located mostly with the capsule and muscle complex acting as second-
in the subsynovial layer. They were mostly surrounded ary stabilizers. Elbow instability may deteriorate fol-
lowing insult to secondary stabilizers.16 A previous
study showed that elbow capsule lesions were almost
always associated with complex elbow instability.17 The
functions of mechanoreceptors in relation to capsular
structures are unclear. The proprioceptive organs in a
normal intact elbow joint provide active elbow joint
stability. The maximal stimulation of afferent signals
from an articular site occurs when the joint is max-
imally loaded.18 This supports not only the idea of
the capsule as a mechanical stabilizer, but also as it
having a protective reflexogenic function. Hence,
there is an urgent need to understand the mechano-
receptors as a protective reflexogenic organ for the
joint. The role of mechanoreceptors has been studied
widely in the knee joint. The existence of
Figure 1. A coil-shaped Ruffini corpuscle (yellow arrow) with
partial encapsulation. This low threshold and slow adapting
mechanoreceptor has arborizing nerve branches with bulbous
terminals (magnification  200).

Figure 2. A Pacinian corpuscle (yellow arrow) observed as an


ovular corpuscle with a thick lamellar capsule; it was embedded
in ligament fibres. Pacinian corpuscles have a low threshold and
rapidly adapting mechanoreceptor characteristics Figure 3. Large nerve fibres observed under synovial tissue
(magnification  200). (magnification  200).
4 E Shoulder & Elbow 0(0)

Table 1. Mean (SD) densities of mechanoreceptors and free nerve endings according to location (per cm3 of tissue).

Mechanoreceptors

Location Golgi corpuscle Ruffini corpuscle Pacinian corpuscle Total Free nerve endings

Anterior Proximal 0 1.60 (0.33) 0.80 (0.16) 2.4 (0.49) 1.94 (0.15)

Mid-substance 0 1.84 (0.87) 0.27 (0.55) 2.12 (1.16) 2.40 (0.33)

Distal 0 1.94 (0.52) 0.13 (0.29) 2.07 (0.40) 1.49 (0.75)

Posterior Proximal 0 1.40 (0.45) 0.54 (0.77) 1.95 (0.31) 2.27 (1.67)

Mid-substance 0 0.82 (0.11) 0 0.82 (0.11) 2.11 (0.88)

Distal 0 1.23 (0.28) 0 1.23 (0.28) 2.77 (1.89)

mechanoreceptors in the elbow joint has also been could result in a proprioceptive deficit leading to recur-
reported; however, there has been a lack of information rent instability.22 Awareness of the mechanoreceptor
on their spatial arrangement within the capsule.19 The distribution in the elbow capsule may allow the surgeon
present study now provides a description of their dis- to preserve proprioceptive integrity during surgery and
tributions and densities. improve postoperative joint stability.
We did not observe the presence of Golgi corpuscles Free nerve endings will detect pain stimuli and trans-
within the elbow capsule. This suggests that not all mit them to the central nervous system; they have been
types of mechanoreceptors are present in the elbow studied widely in respect to their association with pain
capsule, a finding that is supported by a study that generation in lateral epicondylitis pathology.10,23 The
only found Golgi corpuscles in joint ligaments.20 The present study showed that free nerve endings were at
density of mechanoreceptors was higher in the anterior a higher density in the posterior distal sites. An anat-
part of the capsule than in the posterior part, with omy study on cadavers showed that the attachment of
Ruffini corpuscles being found to be the major receptor the joint capsule at the posterior site, under the poster-
type. As a slow adapting receptor, Ruffini corpuscles odistal of the extensor carpi radialis brevis origin,
are always active, even when a joint is static. They sense merges with the supinator muscle complex.23 At this
changes in intra-articular pressure, as well as the direc- site, the joint capsule forms an enthesis structure. By
tion and amplitude of joint movements. These cor- taking both this and our findings into consideration,
puscles also regulate muscle reflex and tone under this attachment site of the joint capsule could be a
continuous tonic conditions. Any extreme torque pain generator in lateral epicondylitis pathology.
within the joint caused by extension, flexion or rotation
will be signalled by Ruffini corpuscles.21 Pacinian cor-
Limitations
puscles were also shown to co-exist with Ruffini cor-
puscles. As a fast adapting receptor, Pacinian The present study was solely an anatomical study and
corpuscles control acceleration, quick joint movement involved a low number of samples; therefore, no statis-
and vibration. Pacinian corpuscles will signal mechan- tical analysis was performed in this pilot study. The
ical stimuli at the onset or cessation of a movement.20 staining method used has not been tested on multiple
The concurrent equilibrium action between these two specimens in different age groups; hence, we were
sensory receptors is important for joint stabilization. unable to draw any conclusions regarding the morpho-
We were not able to identify the existence of Pacinian logical differences of sensory receptors in respect to the
corpuscles in the posterior capsule (mid-substance and age of tissue donors.
distal site) and the paucity of this region might indicate
that this area of the capsule does not have a major
contribution to fast acceleration in joint movements.
Conclusions
This is also supported by the scarcity of other mech- A consistent distribution pattern for articular sensory
anoreceptors in the mid-substance site. Therefore, we receptors was observed; this allows further understand-
suggest that surgical approaches might be best con- ing of elbow pathology. Awareness of the neuroana-
ducted through this site, rather than bony attachment tomical distribution of the receptors in the elbow
sites. Approaches through bony attachment sites would joint capsule may allow their preservation during sur-
likely insult the articular mechanoreceptors, which gical procedures for elbow joint pathology.
E Kholinne et al. 5

Author Contributions 9. Germscheid NM and Hildebrand KA. Regional variation


EK contributed to study design, data analysis and is present in elbow capsules after injury. Clin Orthop Rel
drafting of the manuscript. HJL, MFD, KGY, SJL Res 2006; 450: 219–224.
and SJL contributed to study design and drafting of 10. Sasaki K, Ohki G, Iba K, Kokai Y, Yamashita T and
Wada T. Innervation pattern at the undersurface of the
the manuscript. IHJ contributed to study design and
extensor carpi radialis brevis tendon in recalcitrant tennis
critical revision of the manuscript. elbow. J Orthop Sci 2013; 18: 528–535.
11. O’Connor BL and Gonzales J. Mechanoreceptors of the
Declaration of Conflicting Interests medial collateral ligament of the cat knee joint. J Anat
1979; 129: 719–729.
The author(s) declared no potential conflicts of interest 12. Witherspoon JW, Smirnova IV and McIff TE. Improved
with respect to the research, authorship, and/or publi- gold chloride staining method for anatomical analysis of
cation of this article. sensory nerve endings in the shoulder capsule and labrum
as examples of loose and dense fibrous tissues. Biotech
Histochem 2014; 89: 355–370.
Funding 13. Hackl M, Wegmann K, Ries C, Lappen S, Scaal M and
The author(s) disclosed receipt of the following financial Muller LP. Annular ligament reconstruction with the
support for the research, authorship, and/or publication of superficial head of the brachialis: surgical technique and
this article: This research was supported by the convergence biomechanical evaluation. Surg Radiol Anat 2017; 39:
technology development program for bionic arm through the 585–591.
National Research Foundation of Korea(NRF) funded by 14. Hagert E, Lee J and Ladd AL. Innervation patterns of
the Ministry of Science, ICT & Future Planning (No. thumb trapeziometacarpal joint ligaments. J Hand Surg
2014M3C1B2048422). 2012; 37: 706–714.
15. Marciniak T, Chmielewska A, Weychan R, Parzych M
and Dabrowski A. Influence of low resolution of images
Ethical review and patient consent on reliability of face detection and recognition. Multimed
The present study was approved by our Institutional Review Tools Appl 2013; 74: 4329–4349.
Board. 16. O’Driscoll SW, Jupiter JB, King GJ, Hotchkiss RN and
Morrey BF. The unstable elbow. Instr Course Lect 2001;
50: 89–102.
References 17. Giannicola G, Polimanti D, Sacchetti FM, et al. Soft tissue
1. Morrey BF. Applied anatomy and biomechanics of the constraint injuries in complex elbow instability: prevalence,
elbow joint. Instr Course Lect 1986; 35: 59–68. pathoanatomy, and classification. Orthopedics 2012; 35:
2. Reichel LM and Morales OA. Gross anatomy of the elbow e1738–e1745.
capsule: a cadaveric study. J Hand Surg 2013; 38: 110–116. 18. Ferrell WR. The adequacy of stretch receptors in the cat
3. Aydog ST, Korkusuz P, Doral MN, Tetik O and Demirel knee joint for signalling joint angle throughout a full
HA. Decrease in the numbers of mechanoreceptors in range of movement. J Physiol 1980; 299: 85–99.
rabbit ACL: the effects of ageing. Knee Surg Sports 19. Phillips D, Petrie S, Solomonow M, Zhou BH, Guanche
Traumatol Arthrosc 2006; 14: 325–329. C and D’Ambrosia R. Ligamentomuscular protective
4. Dunning CE, Zarzour ZD, Patterson SD, Johnson JA and reflex in the elbow. J Hand Surg 1997; 22: 473–48.
King GJ. Ligamentous stabilizers against posterolateral 20. Zimny ML. Mechanoreceptors in articular tissues. Am J
rotatory instability of the elbow. J Bone Joint Surg 2001; Anat 1988; 182: 16–32.
83A: 1823–1828. 21. Zimny ML and Wink CS. Neuroreceptors in the tissues
5. Ekholm J, Eklund G and Skoglund S. On the reflex effects of the knee joint. J Electromyogr Kinesiol 1991; 1:
from the knee joint of the cat. Acta Physiol Scand 1960; 50: 148–157.
167–174. 22. Witherspoon JW, Smirnova IV and McIff TE.
6. Freeman MA and Wyke B. The innervation of the knee Neuroanatomical distribution of mechanoreceptors in
joint. An anatomical and histological study in the cat. the human cadaveric shoulder capsule and labrum.
J Anat 1967; 101: 505–532. J Anat 2014; 225: 337–345.
7. Petrie S, Collins JG, Solomonow M, Wink C, Chuinard R 23. Nimura A, Fujishiro H, Wakabayashi Y, Imatani J,
and D’Ambrosia R. Mechanoreceptors in the human Sugaya H and Akita K. Joint capsule attachment to the
elbow ligaments. J Hand Surg 1998; 23: 512–518. extensor carpi radialis brevis origin: an anatomical study
8. Morrey BF and An KN. Functional anatomy of the liga- with possible implications regarding the etiology of lat-
ments of the elbow. Clin Orthop Rel Res 1985; 201: 84–90. eral epicondylitis. J Hand Surgery 2014; 39: 219–225.

You might also like