Ultrasonic Flaw Detector & Thickness GaugeVersion 1: Ultrasonic Testing v \
=, -_ \ .
Version 2: Thickness Gauge
SD card port(166)
a VGA port(mini HDMD
\ Probe port
Encoder port
DC power in port
‘Wrist strap Back strap Mini USB port(reserved)
‘Thickness zero-test block(4mm)
y
‘Dynamic beam tracking to visualize the flaw location. ‘Up to 14 types of weld simulation.
Guide you step by step for the first time operation.
2-8
Set wp & probe info Calibration Make DAG/ AVG curve Detection
IP 66 with compact size: 198 (W)* 128 (H) *52 (L) mm
Light weight: 0.9kg only, including battery
7" LCD with high resolution 640x480 pixels
Adjustable pulse width, negative square wave transmission up to 350V.
Operating frequency range: 0.5-20MHz, multiple steps of wide broadband and narrow-band for selection.
Easy operation: only a few buttons, intuitive interface and logical menu, support right/left-hand operation, outdoor mode.
Conventional UT functions
Standard functions: Weld, plate and forging test wizards, DAC, Peak memory, AGC (Auto gain contro, video record.
Optional functions: B-Scan, TCG, FFT (probe spectral analysis), CSC (curved surface correction), flat weld simulation,
crack height measurement, BEA(backwall echo attenuator), AWS, API SUE.
Thickness measurement functions
Standard functions: A-scan thickness measurement (echo to echo mode), Auto-search, velocity measurement, alarm
and dataset management.
‘Optional functions: CoatMode, B-Scan, MultiLayers, Vpath, TDG and temperature compensation.
* Specific functions are subject to final order.Bring easier and more Auto created by takinga known flat-bottom — @Up to 6 (curves/ lines), each one
convenient flaw evaluatio hole or large flat-bottom echo for reference. max, 10 reference points.
GE/Olympus probe listed.
2
‘The probe waveform, spectrum and For depth and horizontal distance Display A-scan echo in imaging
center frequency of the probe can be correction when testing circumference mode based on time or encoder, 50 as
measured precisely by capturing echoes. with an angle probe. to achieve more intuitive test result for
easy observation and analysis.
‘The crack height is automatically ‘This function is to help seta gate over an area and adjust
calculated with this function. the gain for this area regardless of the global gain. Its very
useful for inspection of Forgings and Castings with allowing
independent gain control of the area under the gate with the
BEA for backwall echo monitoring.It can be used for compensating the Based on time interal or encoder,
Joss of echo amplitude due to propagation display the measurement readings
of sound path. in B-mode image.
Alter setting the coating velocity,
coat and base material thickness can
be displayed at the same time.
Up to4 layers can be measured When there is temperature difference Ail the original dual element probes,
at the same time. between the calibration block and the have a set of default V-path calibration
detected workpiece, it can be used for curves. Users can make a set of
temperature compensation. Adjustable UserVpath curves fora specific probe.
range is -10 to 400 C.
— CS
Port dock thickness TG data sets MultiLayers-thickness With PES-02D
measurement measurement for encoded B-Scan
Checking data file, Screen capture, Measuring data analysis, Playback. Generating test reports in word or excel format.
Abundant report samples are available.
Itcan be installed in every operator's laptop without extra cost.Technical Specification
No.of Channel 1
Probe Connector Type _ | LEMO 00 pes)
Standard (RBI, measurement from transmit pulse to the fist
‘bottom wave) All measurement using Zero Crossing.
Work Mode — eho to Eeho (bi-82 measurement by auto-tracking the
‘Second bottom wave according to the ist bottom Wave)
‘Through coating measurement.
i I a gt ee
Ta ar Sa
nova RSE Sao rere
tegen su inane ace,
~ ESSE SST |e
a ci Ee
a =
aa
‘Auto adjustable faut-search or auo-gan)
Fine Gain oo =
Surface Compensation | Fullgain range =
dis 05-20nix-34B)
Operation Frequency | 1-4/0.5-10/2-20/1/25/4/5/10/13/15/20MH, 1 levels
Rely ‘Neva Posie Ful RF Fier Negative Posie Pal RF
ijec -80R step =
‘Detection ange | 0-15000mm, min dloplyrnge 2.5mm lea bes probe, tri tempera eed
indication Resolution | — 0.010.070 mm (.00170.017001 inet)
00-9198 #0.05mm
Indication Precision 10.00-88 8% = (ll + 0.04)mm
Error 100.0-400.0mm #30 mm
‘Tested with TGMS-T0L probe; His the measured thickness.
Tube Wall Thickness ‘With TGMS-10L probe it can measure stel tubo with diameisr
‘Measurement fo less than 20mm and wal thickness no les than 2mm.
Material Velocity 100-20000%/, mi, sep IIs 1200-20000 m/e
—_— 5-1000mun
10-1000rm, min, sep 0.1mm “10-600mm
0-200us min step 001us 0-200 us
(0-100mm, step 0.1mm —
For weld plate and orgog esting =
Peak’ Flank/ Frank
Gate: amplitude, amplitude dB diference, sound path,
orontal distance, vera! distance, south path diferenee
a Detween Gate a and By _
Cursor: 2 ross cursors, measuring hortzontl and vertical
ions on'B scan Image. and dstance between cursors
facvated far optional B san)
‘ate Mode Standard (Gate Ais selected in wandard measurement mode,
‘No.of Gate 2 —
GateStart Full range Gate star -10-1000%mm, min. step 0.
Gate Wah Pallrange ‘Gate width: 1-2000rum mn, step Om
(Gate Thresh 10-50% step 1% (Gate thresh: 10-80% or 10--90%(for RY, step 1%
Oiffon: if enabled, auto adjuss tothe proper display range.
‘Auto Search = fain and gate position based on the measured wave signals 50
‘510 improve measurement efciency.
‘Velociy Measurement | — Velocity dynam measurement
‘Measure the known reference Block Tor fat ero point
Calibration zero polnt/20 point + weocty! probe angle 2dB =
Submis APISUE, TCG, AWS, C86, CrackMeas, FFT, BEA FlaWeldSim, | ¢aesfoge, path, TDG, B-Scan TempComp, Mul
Besean
SIuI
TES
TP Code
‘Wage
Temal storage
Ccerications|
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Add: #77, Jinsha Road, Shantou 515041, Guangdong, China
‘Tel: +86-754-88250150 Fax: +86-754-88603664
E-mail: siul@siul.com Website: http://www.siul.com
ce
“Operation Temperanae |
‘Storage Temperamare
Dimension wa |
“Encoder Comnecrer |
TOI
57” high brightness TFT LCD, 610-480 pixels
Tachi mm
] Si USE, Scare TEC asa UGK Pas TSG A Sa
Suny HON wah Ge - =
Gi ean! hush? Read japanese
DC 12V (ememal power suppiy:74V banery)
ah (under facony default mode)
10C~ #150
20010
iP
‘Approx Og including 202g baton
198128 52 om)
pe core)
cd
‘UT ERIPGEETPOTO R ENTSO 2PP3E=1 Seanad)
ASIMESI: (Optional
TOL ENISS17 200? seandard)