MC - 240

12.1 MTH12A Board Fuses

■ Removal Procedures
(1) Remove the MTH12A board.
“11.7 MTH12A Board”
(2) Replace the fuse.
“■ Fuse Replacement Procedures”

■ Fuse Locations
MTH12A board
FUSE5 FUSE4
FUSE7
FUSE6
FUSE12 FUSE8

FUSE1 FUSE10

FUSE11 FUSE9

FUSE2

FUSE3
FUSER6H4C03.EPS

FR6H4C03.EPS

■ Fuse Information

Board indications Type Rating (A) Fuse number ✼


FUSE1 Slow blow (T) 5 A31
FUSE2 Slow blow (T) 5 A21
FUSE3 Slow blow (T) 5 A11
FUSE4 Slow blow (T) 5 A41
FUSE5 Slow blow (T) 4 B11
FUSE6 Slow blow (T) 4 B21
FUSE7 Slow blow (T) 4 C11
FUSE8 Slow blow (T) 3 C21
FUSE9 Slow blow (T) 5 H11
FUSE10 Slow blow (T) 5 J11
FUSE11 Slow blow (T) 2 K11
FUSE12 Slow blow (T) 3 C31
✼ Fuse number: Number indicated in the fuse block diagram.
Tbl_MC214_LB.EPS

■ Reinstallation Procedures
For reinstallation, reverse the removal steps.

010-051-05 CR-IR346RU Service Manual MC - 240