Professional Documents
Culture Documents
I-1 Basis of The Oscilloscope
I-1 Basis of The Oscilloscope
I-1 Basis of The Oscilloscope
1. Purpose
For oscilloscope is one of the fundamental observation device, is intended to understand the
operating principle.
2. Theory
Typically, the oscilloscope is a device for observing the cyclically varying voltage waveform.
Here for analog oscilloscope CRT type1), you learn the basic operating principle. Incidentally, in
recent years, as a device similar to an oscilloscope, so-called digital oscilloscope (DSO) is widely
used. But this is exactly because the measurement principle is different, here it is not mentioned, but
there is a need for a variety of care in use.
2.1 Electrostatic deflection type CRT
Figure 1 shows the principle of electrostatic deflection type CRT for use with an oscilloscope.
Inside of the CRT is kept in a vacuum, electrons emitted from the electron gun reach the phosphor
screen without colliding with gas molecules.
The phosphor screen has a fluorescent agent is applied to emit light by electrons collide. In
deflector plate is placed between the electron gun of the fluorescent screen, the electric field
generated by applying a voltage to the plates, it passes through a section of the deflection plate,
electrons are bent path by an electrostatic force. Deviation of the electron trajectory, if the electron
speed when it is driven out from the electron gun is constant, determined by the magnitude of the
voltage applied to the deflection plates. Therefore, it is possible to control the voltage applied to the
deflection plates, it can be seen that it is possible to operate the electron arrival position on the
phosphor screen.
Therefore, by arranging the two axes perpendicular to the deflector (horizontal and vertical) axis,
it is possible to emit light anywhere in the phosphor screen. In addition, it is referred to as a bright
spot that point you have this emission in general. This is the principle of an electrostatic deflection
type CRT.
In addition, the deflection plates instead to provide a pair of electromagnets, there is a method of
controlling the electron trajectories in the Lorentz force to receive a magnetic field in generated
when the electrons pass through. But the CRT magnetic field control type is never used in an
oscilloscope2).
2.2 Horizontal scanning and vertical-axis control
Consider a way to visualize using a CRT when the voltage changes with time. As of now to
Figure 2 (a), it applied a saw-tooth wave VH (t) (A ramp waveform which repeats at a fixed period is
called a saw-tooth wave) on the horizontal axis. If the voltage applied at this time the vertical axis is
0, the inclined portion of the saw-tooth wave, the bright spot on the fluorescent tube surface moves
in a straight line from left to right. It says with a horizontal scanning. Bright spot you have arrived to
the right is again repeating the scan and return to the left. The afterglow and afterimage phenomenon
of the human eye of the fluorescent tube surface, so that the speed of the bright points moving on the
fluorescent tube surface becomes more constant to the human eye are depicted single horizontal line
on a CRT look.
At this time, applying a signal VV (t) varying in time with respect to the vertical axis as shown in
Figure 2 (b), is on the CRT appear trajectory similar bright spot in V V (t). This is the principle of the
oscilloscope 3).
2.3 Trigger
So far, it has been found that it is possible to draw a voltage varying with time on the CRT as the
locus of the bright spot. However, not be the actual waveform measurements in this state. As in
Figure 3 (a), it assumed that VV with periodicity (t) is applied. If the period of the period T0 and VV
(t) of the saw-tooth VH (t) that scans the horizontal axis is different, the locus on the CRT become to
draw a different waveform for each one scanning, CRT can’t normal waveform observed for a
number of trajectories remains as an afterimage in the above. By the way, if the same is the period of
T0 and VV (t), the trajectory to draw each one becomes the same as, in this case, it means that the
apparent waveform overlap all trajectories look stationary 4).
(a) In case of don’t have a trigger function, (b) In case of the trigger function is used
Figure 3: Need for trigger
However, the slope of the VH (t) cannot be moved arbitrarily in order to determine the time the
bright spot moves on the CRT from left to right. (When arbitrarily to vary the inclination, the time
per one horizontal axis scale becomes unclear.)
Therefore, a method for the period and an integer multiple of the period of VV (t) of the VH (t)
by making the time to stop the scanning as shown in Figure 3 (b), instead of changing the slope is
taken. In this case, it is the trigger to signal the start of scan. In a normal oscilloscope, it generates a
trigger signal when the input waveform crosses a constant voltage at the positive slope or negative
slope. Note that positive and negative slope can be selected5).
2.4 Multi phenomenon measurement
The oscilloscope has a normal phenomenon 2 or more, it is possible to observe them at the same
time. However, since generally horizontal scanning system there is only one, has been devised for
drawing a plurality of waveforms at the same time. This way, there are usually two types (Figure 4).
Figure 4: Two phenomenon sweeps
By the way, the impedance at the time the probe end is turned 10 MΩ, it is to 10 times the
impedance of the probe when 1:1 (Ri becomes itself when the 1:1 probe). Since the oscilloscope can
also be considered as a kind of voltmeter effects towards the internal resistance is high has on the
non-measuring circuit is reduced. Also, since even less equivalent input capacitance can be seen
from the Figure, it is also advantageous for measurements at high frequencies.
However, if not satisfied Equation (1) is, it becomes the attenuation has a frequency
characteristic, it would be observed distorted waveform. Therefore, when using the 10:1 probe, it is
necessary to perform the capacity adjustment prior to use. This varying the Cp or is performed by
varying the capacitance is inserted in parallel with Cc.
2.7 Notes on the ground terminal of the probe
In a typical oscilloscope, grounding of all input channels (GND) is connected to the common
housing. That is, all of the input channels and the housing is the same potential is grounded7).
For this reason, it is very dangerous to the connection as shown in Figure. 7, since the cause of the
accident be careful. Following points to prevent accidents it is to be checked, particularly stringent.
3. Experimental Methods
3.1 Initial adjustment of the oscilloscope
To perform the oscilloscope of adjustment in the following steps before you begin the
experiment.
3.1.1 Adjustment and confirmation of focus, brightness and rotation
First, GND input selector switch, and a trigger to AUTO to power. Next, the XY sweep switches
from the level of luminance. Gradually increase the brightness from this state, to display a bright
spot on the tube surface. You do focus adjustment so that the magnitude of the bright spots is
minimized.
Once focus adjustment is completed, immediately subjected to sweeping by switching the sweep
time switch, bright line to appear on the tube surface. When a long time to display a bright spot,
because the seizure occurs in the cathode-ray tube, work up here quickly performed.
Then, to adjust the vertical position adjustment knob as bright line coincides with the horizontal
axis of the center of the screen. Here, you will confirm whether the emission line is tilted. To adjusts
the rotation when the horizontal axis is tilted.
3.1.2 Inspection of probe
Here, you will confirm if the probe to work properly. CH. 1 of input switch what is the
changeover switch of the probe to the DC is 1:1, and further the voltage range of about 0.1V / div,
connecting the probe to the CAL pin oscilloscope. Adjust the sweep rate, so that a square wave is
displayed on the tube surface. In this case, be sure that the fine adjustment knob such as sweep speed
and vertical range is in the position of CAL. In case there is no specific reason, it must be always
kept to the position of the CAL. Here, it is confirmed that the peak value of the square wave that is
displayed on the tube surface is consistent with the voltage being displayed near CAL pin
oscilloscope. If or when a rectangular wave with a large wave differ is not displayed, such as the
probe abnormality can be considered.
(a) Normal, (b) Essential adjustments, (C) Essential adjustments
Figure 8: (10: 1) observation example when using probe
Next, when there is a changeover switch of the probe, which is switched to10:1, by a voltage
range of about 10mV / div, similarly to observe the square wave from the CAL pin. In this case,
Figure. 8 (b), it is necessary to adjust the probe when it is a waveform such as (c). You do set in the
same manner also CH.2.
3.2 Basis of waveform observation by the oscilloscope
The block diagram and appearance of the arbitrary waveform generator are shown in Figure 9(a),
(b). Connect the grounding clip of the probe to the GND terminal, and performs observation of the
waveform by connecting the probe end to the output terminal of the waveform generator.
Relationship between the signal waveform output to the selected waveform and the terminal 1 of
the waveform generator is shown in Table 1.
Figure 12: When input of oscilloscope is set to DC coupling and AC coupling experiments to
observe waveform difference
(b) AC coupling
The sinusoidal wave is generated in the generator, to observe the output terminal 3. The output
terminal 3 is designed to a small signal components in a relatively large DC component
waveforms that are superimposed is output. If you want to observe for the signal component,
and when measuring a DC coupling, consider the difference in the case of measuring an AC
coupling.
3.2.3 Probe 10:1
Connect the probe to the terminal 2, consider the peak value Vp-p to be measured if the probe is
set to 10:1 and 1:1 for the difference determined, respectively. It will note that the output impedance
of the terminal 2 is about 1 MΩ.
3.2.4 Measurements of the phase difference
As shown in Figure 13, CH. 1 side of the probe to the terminal 1 is connected to CH. 2 side of
the probe to the terminal 4. When the output signal of the terminal 1 V1 (t) = Asin (ωt) + A0, the
output signal of the terminal 4 is set to V4 (t) = Bsin (ωt +φ ) + B0, phase difference between the
output signal is obtained in the following manner.
4. Results and Discussion (This section after it is created to suit each situation is shown an
example.)
4.1 Use equipment
・Oscilloscopes: IWATSU SS-5702
・DC-power: KENWOOD PR18-1.2A
・Test any signal generator
4.2 Oscilloscope initial adjustment
It was performed to adjust the focus brightness rotation following the experimental procedure. In
addition, it was confirmed that there is no problem in the operation showed that the operation check
of the probe.
4.3 Basis of waveform observation by the oscilloscope
The result of the waveform observed by using the output of the arbitrary waveform -generator, is
shown below.
4.3.1 Waveform observation basis
(a) Trapezoid (observation of pulse waveform)
And the waveform selection number 3 (trapezoidal wave), the results of measurement of the
output terminal 1 on the oscilloscope, the following results were obtained. Trapezoidal wave
output to terminal 1, became pulse width td = 8.5 ms, the rise time tr = 1.2 ms, fall time tf = 1.1
ms, the period T = 20.0 ms and pulse amplitude Vp-p = 2.3 V, respectively.
(b) Sinusoidal wave of observation
In the waveform selection number 1 (sinusoidal wave), the results of measurement using an
oscilloscope output of the terminal 1 and the amplitude A = 1.2 V, ω = 1.3 ×10 3 rad / s, A0 =
2.1 V in the Equation (2) became.
This means that it can be expressed by the Equation (4).
(b) AC coupling
When the waveform selection number 1 (sinusoidal wave), where the output terminal 3
following the experimental procedure was measured as a DC coupling to a (...omission ...).
Then, where it was re-measured as the AC coupling, it has become (... stands ...). Therefore, it
was confirmed that it is (... stands ...).
4.3.3 10: 1 probe
The waveform selection number is set to 1 (sinusoidal wave) following the experimental method,
a high impedance output peak value was measured by the (terminal 2) the probe 1:1 was measured to
be 1.2 mV.
From the results of Figure.9 and 4.3.1 (b), the peak value of the output waveform of the terminal
2 should be approximately 5.6 mV. Therefore, it is considered to be the (... omitted ...).
4.3.4 Measurements of the phase difference
(a) Direct measurement of the phase difference
The waveform selection number is set to 1 (sinusoidal wave), the experimental method was
observed output waveform V4 (t) of the output waveform V1 (t) and pin 4 of the terminal 1 in
accordance with, V4 (t) and V1 (t) is the time difference Δ t=0.1ms, it has been found that it is
delayed. Since the period V1 (t) and also V4 (t) is T = 2.3 ms, the phase difference between V1
(t) and V4 (t) were asked Φ= 0.1 and rad = 16◦ from the Equation (5).
φ= 2πΔt/T (5)
(b) Measurements using Lissajous
In accordance with experimental method, where it drew a Lissajous as a horizontal axis V4 (t)
the vertical axis V1 (t), Va = 1.2 mV, is Vb = 0.9 mV was obtained. Therefore, the phase
difference was determined and was at Φ = 0.8, rad = 49◦ that determined by the Equation (3).
Comparing the results of this (a), (hereinafter omitted)
4.4 Report challenges
·To indicate the reason why the required phase difference between the two signals from the
Equation (3) by observing the Lissajous.
5. Conclusion (example)
... My results were obtained for the purpose ... was ... that. Therefore, it has been found to be.....
References
[1] State the literature that was in reference always.
Annotation
1) In honor of the inventor and is also referred to as the cathode-ray tube (Braun tube). The
general name cathode-ray tube (cathode-ray tube: CRT).
2) Television receiver or computer monitor is a magnetic field control.
3) This principle is similar to "Fundamentals of Automatic measurement". Section of the basis for
automatic measurement also may be helpful.
4) Strictly speaking it looks stopped if an integral multiple of the period of T0 is VV (t).
5) Usually, to the moment you return to the left from the case and the right end that stops the scan,
you are off the bright spot for CRT is not drawn to prevent and extra line seizure. This is called
blanking.
6) Depending on the oscilloscope models, such as CH.1 is that is free to switch things and the user
of the vertical axis in the horizontal axis is CH.2, there are many different types. Therefore, it is
necessary to confirm a sufficient display of manuals and operating surface.
7) Those grounded input is independent called "Floating". Generally oscilloscope with a floating
input is expensive.