Professional Documents
Culture Documents
Dissolution 2020
Dissolution 2020
18
Dissolution
Vivian A. Gray1, Thomas W. Rosanske2
1
V. A. Gray Consulting, Inc., Hockessin, DE, United States; 2T.W. Rosanske Consulting, Overland Park,
KS, United States
O U T L I N E
conditions are one of the main experimental pa- 18.2.1.2 USP Apparatus 2 (paddle)
rameters to be controlled in dissolution testing. The USP Apparatus 2 is currently the most
Sink conditions can be achieved by the appro- frequently used apparatus for the dissolution
priate selection of the dissolution apparatus testing of solid dosage forms. The dissolution
and dissolution medium. vessel used with this apparatus is the same as
for the USP Apparatus 1. However, the basket
assembly is replaced by a paddle of specified di-
18.2.1 Apparatus mensions as shown in Fig. 18.2. With this appa-
ratus, the dosage form is dropped directly into
The USP General Chapter <711> Dissolution5 the vessel containing the dissolution medium
describes several apparatus types used in disso- and allowed to sink to the bottom; the paddle
lution testing. Other compendia such as the is then rotated at a specified speed but can be
Pharmacopoeia of Japan (JP),6 the British Phar- immersed in the vessel when the dosage form
macopoeia (BP),7 and the European Pharmaco- is dropped as long as the rotation device is
poeia (Ph. Eur.)8 all include the Apparatus 1 switched off; this is the common industry prac-
and 2 dissolution equipment as described in tice. The USP specifies placing the dosage form
the USP. The USP Apparatus 1 (basket) and in the apparatus and immediately operating at
Apparatus 2 (paddle) are by far the most a specified rotational speed. The USP does not
frequently used for immediate release and state whether the paddle can or cannot be
most extended-release dosage forms. The USP immersed prior to addition of the dosage form,
Apparatus 1, 2, 3, and 4 are described in detail only that the rotation be started after the dosage
below. The USP Apparatus 3, 4, 5, 6, and 7 are form has been added. However, experience has
used less frequently than Apparatus 1 and 2, shown that care should be taken to ensure that
but are being used more frequently for charac- the dosage form sinks to the bottom of the vessel
terizing the release of the active ingredient and does not stick to the sides, which can result
from novel dosage forms.9 Descriptions of USP in different dissolution profiles. Dosage units
Apparatus 5, 6, and 7 are found in the USP Gen- that float should be weighed down with a sinker
eral Chapter <724> Drug Release10; these are (see Section 18.3.3. Sinkers) or tested with USP
devoted to transdermal patches. In vitro release Apparatus 1.
of transdermals is described in more detail in
Chapter 19. 18.2.1.3 USP Apparatus 3 (Reciprocating
Cylinder)
The Reciprocating Cylinder, shown in
18.2.1.1 USP Apparatus 1 (basket) Fig. 18.3, is often used as a research tool where
The USP Apparatus 1, shown schematically in a change of pH in the media during the test is
Fig. 18.1, consists of a covered vessel of specified necessary, for example, when testing enteric
shape and dimensions and capacity of 1000 mL, coated tablets. The dosage unit can be moved
a metallic shaft one end of which attaches to a from row to row, with the vessels in each row
motor, and a cylindrical metallic mesh basket containing media of different pH or composi-
that attaches to the opposite end of the shaft. tion. The equipment has a special configuration
The dosage form is placed inside of the basket for beaded products; the beads are contained
and the basket assembly is immersed in the by the screens in the upper and lower parts of
dissolution vessel containing a minimal volume the cell, yet the reciprocating motion allows for
of dissolution medium and rotated at a specified good mixing. The typical media volume is
speed. 200e250 mL.
FIGURE 18.1 USP Apparatus 1, basket. Reproduced with permission from The United States Pharmacopeial Convention (USP)
©2017.
18.3.1 Solubility
The solubility of the drug substance is an
important, though not the only, driving force
in the dissolution test (Eqs. 18.1 and 18.2). It
is, however, important first to know the solubi-
lity in three physiologically relevant media: FIGURE 18.3 USP Apparatus 3, Reciprocating Cylinder.
0.1 M HCl, 4.5 pH buffer, and 6.8 pH buffer. Reproduced with permission from The United States Pharmaco-
peial convention (USP) ©2017.
Therefore, the analyst should be familiar with
the BCS classification system4 and characterize
the drug according to the criteria provided substance in the formulation becomes the focus
(solubility and permeability). Poorly soluble of the method development. Proteolytic en-
drugs may require using alternative media, zymes may also be needed to dissolve pellicles
typically ones containing surfactants. Following formed during the dissolution testing of gelatin
an understanding the solubility of the drug sub- capsules, which can impede the release of the
stance, the dissolution rate of the drug drug.
18.3.3 Sinkers
Dosage forms that tend to float or move
around during the course of a dissolution test
can result in variability and bias. Sinkers are
often used to hold such dosage forms in place
during a dissolution test. There are many types
of sinkers that may be used in dissolution
testing, including custom-made sinkers, and
commercially manufactured sinkers. However,
it is important that the construction of the sinker
is well-defined because it can contribute to vari-
ability of the results. Therefore, when sinkers
are necessary, detailed descriptions and an
explanation of why a sinker is necessary must
be stated in the method. When comparing
different sinkers (or sinker vs. no sinker), tests
FIGURE 18.4 USP Apparatus 4, flow-through-cell.
must be run concurrently with each sinker.
Reproduced with permission from The United States Pharmaco- Each sinker type must be evaluated based on
peial Convention (USP) ©2017. its ability to maintain the dosage at the bottom
of the vessel without inhibiting drug release.
18.3.2 Choice of media A comparison of the different sinker types (or
no sinker) is performed using the same criteria
The evaluation of dissolution media starts for intermediate precision described in USP
with the medium or media in which the drug General Chapter <1092>.16 The variability at
is most readily soluble. The type of formulation all time points should be evaluated when the
then is considered. For fast dissolving drugs, method calls for the generation of a dissolution
the current FDA Guidance18 recommends profile rather than a single point determination.
When transferring a method, the sinkers must profiles of formulations that are intentionally
be duplicated as closely as possible in any sub- manufactured with deliberate, meaningful varia-
sequent testing facility. tions of the most relevant critical manufacturing
parameters (e.g., 10%e20% change to the
ranges of selected parameters).19,20
18.3.4 The “infinity point” In the past, dissolution tests have generally
been conducted at a single time point resulting
The final time point selected for a dissolution
in fast dissolution rates. However, unless the
test does not necessarily correspond to complete
product is a very rapidly dissolving BCS Class
dissolution of drug. However, it is important to
1 or BCS Class 3, a slower profile test is typically
establish that, at some point, or under some con-
needed to give pertinent information about any
ditions, all of the drug material can be accounted
change in the product. Regulatory agencies pre-
for. This can be particularly useful when investi-
fer a method that is clinically relevant, that is,
gating out-of-specification results obtained by
one that shows an IVIVC or certain discrimina-
Assay or Uniformity of Dosage Units (see Chap-
tory power. Since clinically relevant specifica-
ter 11 for further discussion). Performing an “in-
tions are the becoming a requirement for the
finity point,” or fast-stir, test in the early product
characterization and regulation of drug prod-
development phases or routinely on samples in
ucts, laboratories are putting more resources
testing can give an indication of the recovery/ac-
into method development and the establishment
curacy, assuming that the drug is present at or
of an IVIVC or at least some relationship be-
around 100% of the label claim in the dosage
tween dissolution and in vivo performance.
unit. To obtain an infinity point, during the
normal test after the last time point is pulled,
the paddle or basket speed is increased to at least
150 rpm for 30e60 min without stopping the 18.3.6 Design of experiments and quality
test, after which time a further sample is taken. by design concepts and variations
Although there is no requirement for 100%
Development of a dissolution method should
dissolution in the profile, the infinity point can
be a team effort. Formulators provide informa-
provide supportive data when compared to the
tion on the critical quality attributes (CQAs)
content uniformity data. The infinity point data
and provide the prototype dosage forms that
for all six vessels can be compared to the Assay
reflect important formulation and processes var-
and the Uniformity of Dosage Units (weight uni-
iables studied by DoE studies. The principles of
formity or content uniformity) values in terms of
quality by design (QbD) can also be applied to
the mean and the variability. This may also pro-
assessment of dissolution methods. The DoE
vide useful information as to any artifacts from
studies may be useful to establish the critical
the dissolution method and/or the dosage
factors that influence the dissolution rate. The
form interacting with the media.
DoE can evaluate the dissolution operational
parameters21,22 or the product formulation
release attributes.23,24 They may provide under-
18.3.5 Determination of discriminatory
standing of the release mechanisms and deter-
power mine if the dissolution method can show
The regulatory agencies often ask sponsors to change to the CQAs. As an application of a
support the discriminatory ability of the selected QbD approach to dissolution, Li, et al. were
dissolution method. This discriminatory ability able to demonstrate the ability to predict changes
may be demonstrated by comparing dissolution in drug dissolution rate over time as a function of
resource for dissolution method validation. Vali- the drug and can be characteristics of the drug
dation of a dissolution test method consists of substance, the formulation, the manufacturing
two parts. The first part, and the part that will process, or stability.
be given the most emphasis in this section, is Once the appropriate dissolution conditions
the validation of the dissolution test method, have been established, the analytical method
that is, the actual dissolution run and the should be suitably validated. The validation pa-
removal of the samples for analysis. The second rameters may vary depending on the intended
part is the analytical method used to determine use, but will typically include, at a minimum,
the concentration of the drug, which is usually linearity and range, accuracy, precision, speci-
UV spectrophotometry or HPLC with UV ficity, solution stability, and robustness. Each
detection. of these analytical parameters is discussed in
There are numerous aspects of the dissolu- detail elsewhere in this book in various con-
tion procedure that require validation and texts, and the general principles are the same
there are different levels of validation depend- for dissolution. This section will discuss the
ing on the phase of development. In early validation of parameters unique to dissolution
development, filtration, deaeration, linearity, testing. All dissolution testing must be per-
precision, solution stability, selectivity, and ac- formed on a qualified dissolution apparatus
curacy/recovery should be considered as part meeting the specified mechanical and perfor-
of the validation. During later development mance standards and with qualified associated
(Phase III and beyond) full validation studies, analytical instrumentation for the determina-
to include intermediate precision, automation tion of sample concentrations.
and robustness are added. This chapter focuses
on validation of methods used in later stages of
development, when full method validation is
necessary.
18.4.1 Linearity and range
It is presumed that full validation will be con- Linearity and range demonstrate the ability of
ducted on the final dissolution test method the dissolution method to obtain test results over
established for a new drug application or a mar- the range of expected concentrations that are pro-
keting authorization application. The ideal portional to the concentration of the analyte in
dissolution method should exhibit a low vari- the sample. Analytical detector linearity should
ability in the dissolution results, the solutions be established over the entire range of concentra-
should be stable in the dissolution media, and tions expected during the procedure. For imme-
there should be an established dissolution pro- diate release formulations, a concentration
file, using a minimum of three time points. The range of at least 50% of the lowest concentration
dissolution profile should be gradual with at expected in the dissolution vessel to 120% of the
least two points at or below 85% of the active dis- highest concentration is sufficient. For extended-
solved to satisfy the rules for an f2 analysis.26 The release products, the concentration range should
f2, or similarity factor, is a critical tool for demon- extend from approximately 10%e120% of that
strating bioequivalence, although it is not expected from dissolution of the entire dose. If
needed for BCS Class I drugs where the product an extended-release product is formulated in
dissolves rapidly, e.g., 85% in 15 min within the multiple strengths, the detector linearity should
relevant range of pH. The dissolution test be confirmed from 10% of the lowest strength
method should be discriminating and capable concentration to 120% of the highest strength
of detecting changes in the product CQAs, as concentration. When justified, it may be accept-
these attributes would influence the release of able to use different dissolution conditions to
18.4.8 Other validation parameters pulling the sample from the vessel simulta-
neously by manual and automated sampling
Other aspects of validation may include methods for each time point. Note that the
carryover of residual drug, effect of an in- correction for the volume withdrawn from the
residence probe (e.g., auto-sipper or fiber optic medium is doubled in the latter case.
probe), adsorption of drug onto the vessel,
tubing and filters, and cleaning and/or rinse cy-
cles. Evaluation of the filter generally includes 18.6 Sources of error in dissolution testing
preparation of a suitable standard solution
(lowest and highest profile concentrations are Several factors can contribute to errors and
recommended) and a completely dissolved bias in a dissolution test, arising from the equip-
sample solution. For the standard solutions, re- ment (apparatus and dissolution vessels), dis-
sults of the filtered solution (appropriate solved gases, standard solutions, and reference
discard volumes should be determined before- standards, vibration and other mechanical aber-
hand) are compared to those of the unfiltered rations. Critical factors related to running the
standard. For the filtered sample solution, re- method that are not well documented in the
sults should be compared to a fully dissolved written procedure can be a common cause of
and centrifuged sample solution. The accept- failed method transfers.
able range for standard and sample filtration ef-
ficiency is generally between 98% and 102% of
the unfiltered standards solutions and unfil- 18.6.1 The dissolution apparatus
tered but centrifuged sample solution.
USP Apparatus 1 and 2 can be sources of er-
ror if not closely inspected before using. Obvi-
ously, dimensions should be as specified.5 In
18.5 Automation cases of both baskets and paddles, shafts must
be straight and true. The paddles are sometimes
Automated sample collection has become partially coated with Teflon. This coating can
much more widely used in recent years. If peel and partially shed from the paddle,
possible, when conducting a validation of an causing flow disturbance of hydrodynamics
automated method, there should be a compari- within the vessel. Paddles can rust and become
son to a manual sampling method for the same nicked or dented; this can adversely affect
dissolution conditions. All profile time points dissolution hydrodynamics and also be a source
should be evaluated. This validation can be of contamination. Thorough cleaning of the
done in one of two ways: (1) when the drug paddles is important to preclude carryover of
dissolution results are not highly variable, and drug or medium.
understanding the effect of an in-residence probe The baskets need special care and examina-
is desired, two concurrent runs (same sampling tion. The basket mesh/screen can become
intervals, n ¼ 6) using manual and automated frayed, misshapen, or warped with use. Screen
sampling methods are compared using the mesh size may change over time, especially
criteria established for intermediate precision, when used with acidic medium. There are
or (2) if the dissolution results are highly variable different designs for attaching baskets to shafts.
(i.e., the RSD is above 20% at time points of The attachment can be with clips or with O-
10 min or earlier and 10% RSD or above at later rings. These attachment variations can affect
time points), the analysis can be performed by dissolution results, depending upon the product;
screens or particles can cling to bubbles on the areas of testing are especially troublesome. Sam-
glass surface of the vessel or shafts. ple introduction can be difficult and sometimes
The dissolution test should be performed as impossible to control and some products have
soon as practical after deaeration of the medium. dissolution profiles that are “position depen-
It is common practice to rotate the paddle to dent.” For example, if the tablet is off-center,
equilibrate the temperature of the dissolution the dissolution rate may be higher due to shear
medium. However, excessive stirring prior to forces. Conversely, placing the sample close to
starting the experiment can result in redissolu- the center can result in coning and a decrease
tion of atmospheric gasses. The paddle should in dissolution rate. Film-coated tablets can be
always be stopped prior to adding the test sticky when wetted, resulting in inconsistent
article, and great care should be taken to ensure positioning of the test article in the vessel.
that the test article falls in the same place every Replacement of the paddle with a basket or the
time at the bottom of the vessel and does not use of a sinker is the best way to minimize the ef-
stick to the sides. fects of position provided build-up of gelatinous
There are several methods for deaeration of matter does not impair the dissolution.
medium, some manual and some automated. Suspensions can be introduced into the disso-
The method described in USP General lution vessel in a variety of ways: manually, us-
Chapter <711>5 uses heat, filtration, and vac- ing syringes or pipettes, pouring from a tared
uum. Helium sparging is also frequently used beaker, or automated delivery using calibrated
for deaeration of the dissolution medium; how- pipettes. Each method has its own set of limita-
ever, recent shortages of helium have made this tions, although automated sample introduction
approach less practical and commonplace. tends to reduce variability. Mixing of a suspen-
sion sample can generate air bubbles; therefore,
the mixing time of suspension samples must be
18.6.5 Standard solutions strictly controlled to reduce erroneous or biased
Great care should be taken when preparing results.
standard solutions, especially if the standard 18.6.6.2 Dissolution media
must be dried before weighing, as well as
ensuring that the drug powder is completely dis- The dissolution medium is a critical compo-
solved. Notably, prednisone USP RS becomes nent of the test that can cause problems if not
very hard upon drying and difficult to dissolve carefully controlled. One cause of inaccurate re-
in water. Dissolving the powder first in a small sults is the volume of medium withdrawn
amount of alcohol often helps to eliminate this through multiple sampling if not replaced care-
problem. A history of the typical absorptivity fully to maintain sink conditions.
range of the standard can also be very useful to Surfactants can be difficult to clean from the
determine if the standard has been prepared surface of the vessel, especially if the concentra-
properly. tion is high (>0.5%). Surfactants, such as SLS,
can accumulate on the surfaces of the sampling
lines, carboys, other glass containers, and plastic
18.6.6 Method considerations bottles, which may require extensive rinsing to
assure complete removal. Batch-to-batch vari-
18.6.6.1 Sample introduction ability of surfactants, SLS in particular, in terms
The best way to avoid errors and data “sur- of grade, age, impurities (especially electrolytes)
prises” is to put a great deal of effort into the can significantly and affect solubility.34 The
development and validation of methods. Some foaming properties of surfactants can make it
to be cleaned frequently to avoid buildup of dissolution test. Video recording of the dissolu-
drug, excipients, surfactants, or buffer salts tion process can be an invaluable approach in
from the dissolution medium. diagnosing anomalous or poor reproducibility
A common source of error can derive from a of dissolution results. Accurate, meaningful
different filter type being used in the automated dissolution occurs when the product dissolves
dissolution test versus the manual dissolution without disturbance from barriers to dissolution,
test. If the same filter type and pore size cannot or disturbance of vessel hydrodynamics from
be used, then filter validation is essential. any source. The particle disintegration pattern
must show freely dispersed particles. Anoma-
18.6.6.6 Cleaning lous dissolution usually involves one or more
The analyst should take special care to of the following observations:
examine this aspect when validating the method.
• floating chunks of tablet
In many laboratories, where different products
• spinning
are tested on the same equipment, this is a crit-
• coning
ical issue that, if inadequately monitored, may
• mounding
be a cause of inspection failures and erroneous
• gumming
results.
• swelling
18.6.6.7 Method transfer • capping
• “clam shell” erosion
Problems occurring during transfer of
• off-center position
methods can often be traced to not having
• sticking of particles adhering to apparatus or
used exactly the same type of equipment, such
vessel walls
as baskets/shafts, sinkers, dispensing appa-
• sacs
ratus, sampling method, or difference in the
• swollen/rubbery mass
testing environment. The sampling technique
• clear pellicles
(manual vs. automated), and sample introduc-
tion, should be uniform. A precise description Along with good documentation, familiarity
of medium and standard preparation, including with the dissolution behavior of a product is
grade of reagents, in the method is essential and essential in quickly identifying changes in stabil-
should go well beyond simply stating the ity or changes associated with a modification of
method should be conducted according the the formulation. One may notice a change in
relevant pharmacopeial general chapter. Video the size of the dissolving particles, excipients
recording may be useful to detect differences floating upward, or a slower erosion pattern.
between sites. Changes in the formulation or an increase in
strength may produce previously unobserved
basket screen clogging. If the contents of the bas-
18.7 Visual observations ket immediately fall out and settle to the bottom
of the vessel, a spindle assembly surge might be
One of the most useful tools for identifying indicated. If the medium has not been properly
sources of error is close visual observation of deaerated, the analyst may see particles clinging
the dissolution test. Trained analysts can to vessel walls. The presence of bubbles always
pinpoint many problems because they have indicates that deaeration is necessary. Lastly,
developed a knowledge and understanding of the water bath should contain clean water so vi-
the cause and effect relationships of well- sual observations of the dissolution test can be
known issues that can occur during the made clearly and easily.
classification considerations.4,26 A recent FDA drug products are rare, the specifications are usu-
guidance shows a further breaking down of im- ally based on dissolution data from pivotal
mediate release formulations to cover specif- batches and other important lots, including stabil-
ically immediate release drug products ity data.
containing high solubility drug substances.18 It
should be noted that the FDA provides only
guidance and the recommendations posed in
the various guidances are not legally binding.
18.9.2 Extended-release drug products
That said, it is recommended that proposed Extended-release formulations are, in essence,
specifications be discussed with FDA prior to designed such that release from the dosage form,
presenting in a formal submission. Specifica- and hence the dissolution of the drug substance,
tions that are within the FDA guidance for a is controlled and occurs over an extended period
specific dosage form are likely to receive less of time. As such, single point, or in most cases
resistance than those which are outside the two-point, specifications are insufficient to
guidance, but in any case, the specifications adequately characterize the quality and batch-
must be justified with appropriate supportive to-batch consistency of the drug product. In
data. All dissolution specifications must addition, because of the designed drug release
apply to a given drug product over its entire characteristics, it is often possible to establish
shelf life. an IVIVC for these formulations. Whether or
not an IVIVC can be established, the specifica-
tion should reflect results from biobatches
designed to determine acceptable clinical
18.9.1 Immediate release drug products
behavior. The dissolution specification for
Specifications for immediate release dosage extended-release formulations should include a
forms are intended to assure a quality measure minimum of three time points, one early time
of batch-to-batch consistency and in some cases point to establish the absence of dose dumping,
an indicator of acceptable bioavailability. For one time point in the middle of the release pro-
drug products containing drug substances with file, and one at the end where at least 80% of
high solubility over the physiological pH range, the dose has been released.
dissolution is primarily used as a quality control Approaches to IVIVC for extended-release
check for batch-to-batch consistency, and a sin- products are discussed in detail in the FDA
gle point specification is most often considered guidance: Extended-release solid oral dosage
sufficient. In such cases, a generally acceptable forms: development, evaluation, application
specification is not less than 85% (Q ¼ 80%) dis- of in vitro/in vivo correlations,45 and USP
solved in 30 min. Acceptance criteria other that General Chapter <1088>, In Vitro and In Vivo
this can be proposed but must be accompanied Evaluation of Dosage Forms46 A well-
by appropriate supportive data.26 established correlation will allow for a reason-
For poorly soluble drugs, a two-point specifi- able prediction of the in vivo behavior of
cation is sometimes needed. It is also possible formulations without performing a bioavail-
that in some cases an IVIVC can be established. ability study. Finding the appropriate correla-
If that is the case, specification limits should be tion has been the focus of numerous
set based on the IVIVC and the dissolution test studies.47e49 The establishment of an IVIVC
can be used to distinguish between bioequivalent has been the subject of much discussion
and bioinequivalent formulations or batches. recently in the arena of QbD and setting clini-
However, since IVIVC for immediate release cally relevant specifications.50
14. Martin, G.; Gray, V. General Considerations for Dissolu- Health and Human Services, Food and Drug Adminis-
tion Methods: Development, Validation, and Transfer. tration, Center for Drug Evaluation and Research
JVT 2011, 17 (1), 8e11. (CDER), U.S Government Printing Office: Rockville,
15. Rohrs, B. R. Dissolution Method Development for MD, USA, 1997.
Poorly Soluble Compounds. Dissolution Technol. 2001, 27. Rohrs, B.; Thamann, T.; Gao, P.; Stelzer, D.; Bergren, M.;
8 (3), 6e12. Chao, R. Tablet Dissolution Affected by a Moisture
16. USP 42eNF 37, the Dissolution Procedure: Development Mediated Solid-State Interaction between Drug and
and Validation <1092> Rockville; USP: MD, USA, 2018. Disintegrant. Pharm. Res. 1999, 16, 1850e1856.
17. Published online: https://www.ich.org/products/ 28. Gray, V.; Beggy, M.; Brockson, R.; Corrigan, N.;
guidelines.html. Verified on 9 September 2019. Mullen, J. A Comparison of Dissolution Results Using
18. Dissolution Testing and Acceptance Criteria for Immediate- O-Ring versus Clipped Basket Shafts. Dissolution Tech-
Release Solid Oral Dosage Form Drug Products Containing nol. 2001, 8 (4), 8e11.
High Solubility Drug Substances; FDA Guidance for In- 29. Tanaka, M.; Fujiwara, H.; Fujiwara, M. Effect of the Irreg-
dustry; U.S. Department of Health and Human Services, ular Inner Shape of a Glass Vessel on Prednisone Dissolu-
Food and Drug Administration, Center for Drug Evalu- tion Results. Dissolution Technol. 2005, 12 (4), 6e14.
ation and Research (CDER), U.S Government Printing 30. Scott, P. Geometric Irregularities Common to the Disso-
Office: Rockville, MD, USA, 2018. lution Vessel. Dissolution Technol. 2005, 12 (1), 18e21.
19. Gray, V. A. Power of the Dissolution Test in Distinguishing 31. Vangani, S.; Flick, T.; Tamayo, G.; Chiu, R.; Cauchon, N.
a Change in Dosage Form Critical Quality Attributes; AAPS Vibration Measurements on Dissolution Systems and
PharmSciTech, 2018. Published online: https://doi.org/ Effects on Dissolution Prednisone Tablets RS. Dissolu-
10.1208/s12249-018-1197-7. tion Technol. 2007, 14 (1), 6e14.
20. Suarez, S.; Marroum, P.; Hughes, M. Chapter 15, Bio- 32. Gao, Z.; Moore, T. W.; Doub, W. H. Vibration Effects on
pharmaceutic Considerations in Drug Product Design Dissolution Tests with USP Apparatus 1 and 2. J. Pharm.
and In Vitro Drug Product Performance. In Applied Bio- Sci. 2008, 97 (8), 3335e3343.
pharmaceutics and Pharmacokinetics; Shargel, L., Yu, A., 33. Collins, C. C. Vibration: What Is it and How Might it Ef-
Eds., 7th ed.; McGraw-Hill Education, 2016. fect Dissolution Testing. Dissolution Technol. 1998, 5 (4),
21. Eaton, J.; Deng, G.; Hauck, W.; Brown, W.; Manning, R.; 16e18.
Wahab, S. Perturbation Study of Dissolution Apparatus 34. Crison, J. R.; Weiner, N. D.; Amidon, G. L. Dissolution
Variables-A Design of Experiment Approach. Dissolu- Media for In Vitro Testing of Water-Soluble Drugs, Effect
tion Technol. 2007, 14 (1), 20e26. of Surfactant Purity and Electrolyte on In Vitro Dissolu-
22. Kikawai, L.; Tran, D.; Hauck, W.; Shah, V.; Stippler, E. tion of Carbamazepine in Aqueous Solutions of Sodium
Effect of Various Operational Parameters on Drug Lauryl Sulfate. J. Pharm. Sci. 1997, 86 (3), 384e388.
Release from a 1 % Hydrocortisone Semisolid Dosage 35. Wuelfing, W.; Kosuda, K.; Templeton, A.; Harman, A.;
Form Using the Vertical Diffusion Cell Apparatus. Mowry, M.; Reed, R. Polysorbate 80 UV/vis Spectral
Dissolution Technol. 2012, 19 (3), 6e12. and Chromatographic CharacteristicsdDefining Bound-
23. Villalobos, R.; Viquez, H.; Hernandez, B.; Ganem, A.; ary Conditions for Use of the Surfactant in Dissolution
Melgoza, L. M.; Young, P. M. Parameters Affection Analysis. J. Pharm. Biomed. Anal. 2006, 41, 774e782.
Drug Release from Inert Matrices. 1: Monte Carlo 36. Soltero, R. A.; Hoover, J. M.; Jones, T. F.; Standish, M.
Simulation. Pharm. Dev. Technol. 2012, 17 (3), 344e352. Effects of Sinker Shapes on Dissolution Profiles.
24. Chowdary, K. P.; Rao, R.; Surya Prakasa, K. A Factorial J. Pharm. Sci. 1989, 78 (1), 35e39.
Study on the Effects of b-cyclodextrin and Poloxamer 37. Lindenberg, M.; Wiegand, C.; Dressman, J. Comparison
407 on the Dissolution Rate of Valsartan from CD Com- of Adsorption of Several Drugs to Typical Filter
plexes and Their Tablets. J. Pharm. Biol.Chem. Sci. 2011, Materials. Dissolution Technol. 2005, 12 (1), 22e25.
2 (3), 438e444. 38. Hauck, W. W.; Manning, R. G.; Cecil, T. L.;
25. Li, H.; Nadig, D.; Kuzmission, A.; Riley, C. Prediction of Brown, W. E.; Williams, R. L. Proposed Change to
the Changes in Drug Dissolution from an Immediate- Acceptance Criteria for Dissolution Performance Verifi-
Release Tablet Containing Two Active Pharmaceutical cation Testing. Pharm. Forum 2007, 33 (3), 574e579.
Ingredients Using an Accelerated Stability Assessment 39. Published online: https://www.usp.org/resources/
Program (ASAPprimeÒ). AAPS Open 2016, 2, 7. compendial-tools. Verified on 16 December 2018.
https://doi.org/10.1186/s41120-016-0010-5. 40. Martin, G. P.; Gray, V. A. Overview of Dissolution In-
26. Guidance for Industry; Dissolution Testing of Immediate- strument Qualification, Including Common Pitfalls.
Release Solid Oral Dosage Forms; U.S. Department of Dissolution Technol. 2011, 18 (2), 6e10.