Professional Documents
Culture Documents
www.elsevier.com/locate/talanta
Received 17 April 2002; received in revised form 16 July 2002; accepted 19 August 2002
Abstract
The determination of chlorine with o -tolidine in greywater has been studied, and a batch method and a sequential
injection method have been proposed. It was found that the reaction of o -tolidine with chlorine was slower in a
greywater matrix. Grey water samples must be filtered before analysis, or alternatively, a blank of sample must be
measured. The samples are very unstable, and after 2 h of storage in dark conditions at 4 8C the chlorine concentration
can diminish. The standard addition method, in some samples gave unsatisfactory results as a consequence of matrix
effects. It was concluded that samples with a total organic carbon (TOC) higher than 60 mgC l1 can not be accurately
analysed using these methods. However, samples with a TOC lower than 60 mgC l 1 gave 1009/3% recoveries. The
linear range of the methods was 0 /3.0 and 0 /5.0 mgCl2 l 1, and the limit of quantification 0.2 and 0.5 mgCl2 l 1, for
the batch method and the sequential injection method, respectively. This is the first paper devoted to grey water from
the analytical chemistry point of view.
# 2002 Elsevier Science B.V. All rights reserved.
Greywater frequently contains levels of faecal strategies in Analytical Chemistry [10 /12]. Basi-
coliforms and other bacteria, which make it cally, FIA allows a higher sampling frequency
unsuitable for use as toilet flushing due to a than SIA, and SIA is less reagent consuming than
certain risk of human contact (from splashing FIA. Considering that the required frequency of
and aerosols) [3]. For these reasons disinfection is chlorine analysis for a satisfactory control of the
a key step, either from the sanitary point of view, chlorination process is a few samples per hour at
and also to prevent the appearance of undesirable 3/4 sampling points in a small scale greywater
odour. An efficient and commonly used waste- recycling system for flushing toilets, and also
water disinfection technology is the chlorination bearing in mind that o-tolidine is classified as a
with hypochlorite [4]. According to preliminary toxic reagent, and consequently the minimization
experiments carried out in a 200-bed hotel during of its consumption is advisable, a SI procedure was
summer 2001, and considering recently published considered adequate for the automatic determina-
results on risk assessment in small scale water tion of chlorine. Many flow injection procedures
recycling systems [5], the safety and quality of the for chlorine suitable for drinking water analysis
reused water depends, mainly, on the correct with good analytical characteristics with photo-
chlorination of the greywater circuit. Greywater metric [13 /16], chemiluminescent [17 /19], electro-
varies greatly in composition and generation flow chemical [20], and potentiometric [21] detection
over a 24-h period, and consequently the chlorine have been published, but this is the first paper
demand is also variable. Then, a constant chlor- devoted to a sequential injection procedure for
ination can cause problems due to excess (disin- chlorine determination.
fectant odour and waste of hypochlorite) or to
defect (storage of undisinfected greywater for
more than 24 h can result in offensive odours).
2. Experimental
Adequate chlorination requires a frequent analy-
tical control of samples from different parts of the
circuit. The aim of this paper is to study an 2.1. Apparatus
analytical determination of hypochlorite in grey-
water. According to reviewed literature, such a The manifold components were an automatic
matrix (greywater), in spite of its potential, non- burette Crison 2031 and an eight port multiposi-
potable, reuse to flush toilets, and others, such as tion valve supplied by Crison Instruments, Barce-
garden irrigation [6], remains unstudied from the lona, Spain. The burette was equipped with a 2.5-
analytical point of view. Normally, greywater ml syringe, which permitted a 1 ml minimum piston
parameters are determined using methods for stroke, and with a two way valve that permitted
wastewater. However, greywater matrix shows a the aspiration of carrier stream when necessary.
higher content of detergents and lower content of 0.8 mm i.d. PTFE tubing was used in the
organic matter, and this can lead to errors due to construction of the sequential injection system.
matrix effects. Greywater should be studied in The burette and the multiselection valve were
Analytical Chemistry as a specific sample, differ- connected to a personal computer via an RS 232
ent to blackwater. interface, and controlled by home-made software.
The analytical reaction selected to carry out the Total organic carbon was determined using a
determination reported here is the well-known TOC-5000-A Shimadzu analyzer.
chlorination of o -tolidine [7 /9] due to its extensive A Shimadzu UV-120-02 spectrophotometer
use in potable water control. equipped with a flow through cell of 10 mm light
Besides, the automation of this analytical deter- pathlength and 18 ml volume was used as a
mination was also an objective of this paper, in detector. Peaks were obtained using a pen recor-
order to facilitate the routine analysis. der. The sequential injection system is depicted in
Flow injection analysis (FIA) and sequential Fig. 1. The cell used in the batch procedure had
injection analysis (SIA) are two widely used also a pathlength of 10 mm.
J.G. March et al. / Talanta 58 (2002) 995 /1001 997
Fig. 1. Scheme of the sequential injection system. Standard or samples were arranged around the valve at valve position from four to
eight. AB, autoburette; MV, multiselection valve.
2.2. Reagents and solutions the reaction coil. (5) Stop for 3 min. (6) Propulsion
of 2000 ml through the reaction coil to the
o-Tolidine (3,3?-dimethyl bencidine), purchased measuring cell. The absorbance at 437 nm was
from Probus (Spain), and sodium hypochlorite 5% recorded. Aspiration rate and propulsion rate were
solution, sodium thiosulphate, potassium iodate, 16.7 and 2.0 ml min 1, respectively. 1.1 mol l1
potassium iodide and starch, purchased from HCl was used as carrier. As in the batch method,
Panreac (Spain), were used. All reagents were of measurements must be obtained against a blank of
analytical-reagent grade. A 6.3 10 3 mol l 1 o- sample or, alternatively, the sample must be
tolidine reagent solution was prepared in 1.8 filtered through 0.45 mm cellulose acetate mem-
mol l1 HCl. The carrier used was 1.1 mol l 1 brane.
HCl. Sodium hypoclorite was titrated before use
[9] and its concentration labelled as chlorine, an
hypochlorite working solutions were prepared by
3. Results and discussion
dilution with water.
Table 1
Study of the sandwich composition
A 3.2 mg Cl2 ll standard was used. Other variables were indicated in the procedure.
a
Mean of three measurements.
Fig. 5. Relationship between % recovery and TOC for several greywater samples. The chlorine concentration in the samples in this
experiment were in the range 0.6 /1.2 mgCl2 l 1. Each result is an average of four experiments.
J.G. March et al. / Talanta 58 (2002) 995 /1001 1001
samples with a TOC B/60 mgC l1 which led to and the collaboration and interest of Mr Domingo
satisfactory recovery values (in the order of 1009/ Muñoz (manager of the hotel) in this project is
3%) and samples with TOC /60 mgC l1 that gratefully acknowledged. M. Gual thanks the
showed, in most cases, deficient recovery values. town council of Palma de Mallorca for a fellow-
These facts were, in part, due to the reaction of ship.
added chlorine with (organic)compounds present
in the sample during the application of the
method. This affirmation is supported on the
diminution of chlorine concentration with time
References
that was observed from chlorinated samples in the
laboratory. In fact, several days after chlorination [1] D. Christova-Boal, R.E. Eden, S. McFarlane, Desalination
of a sample, the chlorine concentration fell to 106 (1996) 391.
undetectable values. In samples with low TOC [2] F. Kiya, H. Aya, Water Sci. Technol. 23 (1991) 2189.
these reaction can take place in a non detectable [3] B. Jeppesen, Desalination 106 (1996) 311.
extension (during the application of the method). [4] Metcalf-Eddy, Tratamiento y depuración de aguas resi-
duales, Ed Diaz de Santos, Madrid, 1995.
Taking these facts into account, it is recommend- [5] C. Diaper, A. Dixon, D. Butler, A. Fewkes, S.A. Parsons,
able, when doing routine work, to know the TOC M. Strathern, T. Stephenson, J. Strutt, Water Sci. Technol.
range of analysed samples. Then, samples with 43 (2001) 83.
TOC /60 mgC l 1 should be diluted before [6] J.A. Andersen, Water Sci. Technol. 33 (1996) 37.
analysis. It is worth mentioning that grey water [7] APHA, AWWA, WCCF, Standard Methods for the
examination of Water and Wastewater, 16th ed., 1985.
from the Hotel involved in this project gave, in all [8] D.J. Leggett, N.H. Chen, D.S. Mahadevappa, Analyst 107
cases, TOC in the 309/20 range (grey water (1982) 433.
samples with higher TOC used in this work were [9] D.F. Boltz, W.J. Holland, J.A. Howell, in: D.F. Boltz, J.A.
obtained from residential buildings). Howell (Eds.), Colorimetric Determination of Nonmetals,
Interscience, New York, 1978.
The SI method and the batch method were also
[10] M. Valcárcel, M.D. Luque-de-Castro, Flow Injection
compared using 10 different samples. The correla- Analysis: Principles and Applications, Ellis Horwood,
tion line found for the mean values obtained by the Chichester, 1987.
automatic method (Y) versus the batch method [11] J. Ruzicka, E.H. Hansen, Trends Anal. Chem. 17 (1998)
(X), corresponding to a 95% confidence level and 8 69.
degree of freedom, was [12] J. Ruzicka, G.D. Marshall, Anal. Chim. Acta 237 (1990)
329.
Y /(/0.099/0.02)/(1.0089/0.007)X, R2 / [13] M.C. Cheregi, A.F. Danet, Revista de Chimie 51 (2000)
1
0.9998, (X and Y are mgCl2 l ). This line 566.
demonstrated the agreement between both meth- [14] F. Cañete, A. Rios, M.D. Luque-de-Castro, M. Valcárcel,
odologies. Analyst 113 (1988) 739.
[15] K.K. Verma, A. Jain, A. Townshend, Anal. Chim. Acta
261 (1992) 233.
[16] M. Catalá Icardo, J.V. Garcı́a Mateo, J. Martı́nez
Acknowledgements Calatayud, Analyst 126 (2001) 2087.
[17] W. Quin, Z.J. Zhang, S.N. Liu, Anal. Lett. 30 (1997) 11.
The authors acknowledge the financial support [18] G.P. Irons, G.M. Greenway, Analyst 120 (1995) 477.
[19] G.P. Irons, G.M. Greenway, Anal. Proc. 31 (1994) 31.
from the Spanish Ministry of Science and Tech-
[20] H. Iketake, A. Yamada, Bunseki Kagaku 48 (1999) 1123.
nology for the project on grey water reuse (REN [21] A. Sakai, A. Hemmi, H. Hachiya, F. Kobayashi, S. Ito, Y.
2000-0336-P4-02). Also, the availability of Hotel Asano, T. Imato, Y. Fushinuki, I. Taniguchi, Talanta 45
Brasilia installation (Palma de Mallorca, Spain) (1998) 575.