You are on page 1of 14

Chapter 19

Point-of-care testing
James H. Nichols
Department of Pathology, Microbiology and Immunology, Vanderbilt University Medical Center, Nashville, TN, United States

Learning objectives patient-care settings: hospital acute and critical care units,
After reviewing this chapter, the reader should be able to: operating rooms, emergency departments, physician’s
G Define point-of-care testing (POCT). offices, visiting nurses, patient home self-testing, ambu-
G Identify the essentials of quality management for reliable lances, and cruise ships. POCT devices have even trav-
POCT results. eled on the space shuttle.
G Describe POCT limitations. POCT has a number of advantages (Table 19.1) com-
G List the resources available to assist staff in managing their pared with central laboratory testing. POCT devices use a
POCT program. small volume of sample, and the sample does not need to
be processed. POCT is particularly useful in neonates,
Introduction small babies, and those with increased risk of blood loss
Point-of-care testing (POCT) is laboratory testing con- from phlebotomy. POCT devices can analyze whole blood,
ducted close to the site of patient care. There are a num- urine, and other body fluids. Test results are rapidly avail-
ber of portable POCT devices on the market, and the able while the clinician is examining the patient. There is
available menu of analytes is extensive, including glu- no need to transport a specimen to a laboratory and wait
cose, urinalysis, pregnancy, occult blood, electrolytes, for results to come back. POCT thus works with the clini-
blood gases, creatinine, urea, microalbumin, hemoglobin cal flow of patient care, because clinicians can order a test,
A1c, drugs of abuse, therapeutic drug monitoring (lith- receive results, and institute clinical action in the same
ium), as well as testing for infectious diseases like strep- visit. This provides the opportunity for faster therapeutic
tococcus, mononucleosis, influenza, and HIV. Even intervention and potential for improved patient outcomes.
coagulation testing [prothrombin/international normalized However, POCT has some disadvantages (Table 19.1)
ratio (INR)], enzymes [alanine aminotransferase (ALT)/ as well. POCT is more expensive than traditional laboratory
aspartate aminotransferase (AST)], and cardiac markers testing, because POCT reagents are single-use tests, while
[B-type natriuretic peptide (BNP)] are available on the central laboratory reagents are packaged in bulk for high-
Clinical Laboratory Improvement Amendments of 1988 volume analysis. POCT in hospital settings is often not per-
(CLIA) waived devices. CLIA waiver is granted to labo- formed by lab personnel, resulting in challenges in identify-
ratory tests that the US Food and Drug Administration ing deviations in preanalytic, analytic, or postanalytic
(FDA) and Centers for Disease Control and Prevention components of testing. Documentation is an additional con-
(CDC) have determined to be so simple that there is little cern, because POCT may be performed and results not
risk of an erroneous result. CLIA waived tests have mini- interfaced or documented in the patient’s electronic health
mum regulatory requirements, so a majority of POCT is record. Clinical staff may be unfamiliar with the many laws
CLIA waived, allowing the test to be performed in a vari- and compliance requirements that apply to laboratory opera-
ety of settings by a diverse array of staff. However, as tions both within the United States and internationally.
with any test, POCT, even CLIA waived testing, has lim-
itations and poses risk of incorrect results and patient mis-
management if incorrectly performed or poorly managed. Quality point-of-care testing
POCT is also sometimes referred to as near-patient test- Quality POCT is the delivery of results that are safe, reli-
ing, bedside testing, ancillary testing, and satellite testing, able, and trustworthy for patient care. Quality implies the
because the POCT devices can be used in many different appropriate use of technology to meet clinical needs. This

Contemporary Practice in Clinical Chemistry. DOI: https://doi.org/10.1016/B978-0-12-815499-1.00019-3


© 2020 Elsevier Inc. All rights reserved. 323
324 Contemporary Practice in Clinical Chemistry

TABLE 19.1 Point-of-care testing advantages and TABLE 19.2 Example point-of-care testing program for
disadvantages. an academic medical center.
Advantages Method Sites Devices Operators
G Immediate results—no lab transportation Glucose meters 53 213 4500
G Small blood volume
G Wide menu of tests available Hemoglobin A1c 10 21 50
G Whole blood, urine, and other body fluids analyzed without Fecal occult blood 38 1400
processing
G Works within clinical patient flow Gastric occult blood 6 500

Disadvantages Urine dipsticks 43 28 650


G More expensive than traditional laboratory tests Urine pregnancy 41 600
G Quality is questionable as anyone can run the analysis
Serum pregnancy 1 25
G Difficulties with regulatory compliance, billing, and
documentation Rapid strep 20 300
G Patient-focused staff with little formal education/experience
Mononucleosis 1 3
in laboratory testing
Influenza 14 225
pH paper 2 50
Blood gas portable 10 15 400
involves choosing the right method, POCT device or cen- Hemoglobin 3 4 50
tral laboratory analysis, to achieve the best outcome for a
Avoximeter 2 7 30
particular patient at a specific point in their pathway of
care. This might involve selecting between a glucose Prothrombin time/INR 7 7 60
meter, glucose on a blood gas analyzer, or glucose in a Activated clotting time 7 30 150
central laboratory, depending on the clinical situation. In Heparin management 1 3 6
addition, test results are legal documents and all of the
quality control (QC), training, instrument validation, trou- Basic metabolic panel 6 9 275
bleshooting, and other quality assurance practices and
documentation ultimately provide defensible evidence of
regulatory compliance should a test result ever come into
question. Above all, the reputation of the laboratory laboratories in Colorado and Ohio did not follow manu-
depends on the quality of its product, the test result. facturer’s instructions nor did they have package inserts
Unfortunately, there are a number of quality concerns available to staff performing the testing. Other laborato-
with POCT, which have been raised in the scientific liter- ries were cutting occult blood and urine dipsticks into
ature. Complaints about glucose meters for patient self- segments to save costs and extend their use. In an
testing, for instance, represent the largest number of expanded study conducted in eight states, 64% of labora-
complaints filed with the FDA for any medical device, tories (173/270) surveyed did not have manufacturer’s
with over 3200 incidents including 16 deaths [1]. POCT instructions or failed to follow the instructions [5].
devices can be reservoirs for nosocomial and antibiotic- Twenty percent of laboratories (n 5 53) were not perform-
resistant organisms in the hospital, and infections have ing QC as instructed, and 19% of laboratories (n 5 51)
been transmitted by poorly cleaned urinometers and blood did not provide training to staff or evaluate the staff’s
gas devices when carried between patient rooms [2]. ability to perform testing adequately.
Patients at nursing facilities in Mississippi, North POCT is a complex system. In a central laboratory,
Carolina, and California were diagnosed with hepatitis B there is one location and testing is limited to a few analy-
infection transmitted in association with blood glucose zers. Staff is also limited to a pool of medical technolo-
monitoring from shared spring-loaded lancets, glucose gists and medical laboratory scientists who have
meters, insulin vials, and poor hygiene between the significant training and skills in laboratory analysis. Staff
patients [3]. In a two-state pilot study conducted by the in a central laboratory has focused attention on testing.
Department of Health and Human Services (DHHS) on Their primary task is specimen analysis and producing
data from state laboratory inspectors, 77% of the 175,000 quality test results, without the distraction of patient care
CLIA enrolled laboratories had no direct oversight by responsibilities. POCT, on the other hand, is conducted at
someone with laboratory training [4]. Most of these labo- dozens of locations, with hundreds of devices and thou-
ratories were physician office practices, and 50% of sands of operators (Table 19.2). Getting all operators to
Point-of-care testing Chapter | 19 325

perform a test consistently, in the same way every time, is centers for medicare and medicaid services (CMS) and
difficult. POCT is performed by clinical staff whose pri- the state departments of public health. CLIA sets the min-
mary focus is patient care. Nurses and physicians are only imum requirements for laboratory testing based on the
secondarily focused on performing the necessary steps for complexity of analysis, instrument maintenance, and
QC, calibration, maintenance, cleaning, and quality assur- result interpretation. Tests are categorized as waived,
ance of POCT devices. There is the assumption that if a moderate, or high complexity with increasingly higher
POCT device generates a result, then that result must be levels of requirements for operator education, training,
correct; otherwise, the device would not have generated a method validation and quality documentation. The Joint
value. However, there are many factors that can affect the Commission, the College of American Pathologists
quality of the POCT results. (CAP), and the Commission on Office Laboratory
The DHHS studies from state inspectors uncovered Accreditation are private accreditation agencies that are
quality concerns with the performance and staff training deemed under CLIA to inspect and accredit labs that meet
in a significant percentage of sites performing POCT, the minimum CLIA standards in addition to their individ-
mostly physician offices [4,5]. Most physicians and facili- ual agency recommendations. Although each of the
ties were eager to change once the issues were brought to accreditation agencies varies in the details and stringency
their attention but had no resources or background to of their recommendations, all of the agencies require
design the training and ongoing quality assurance pro- some level of equipment validation before use of the test
gram. Laboratory professional consultation was needed to on patients, written procedures, operator training and
improve the regulatory compliance of these sites and competency on the specific test, as well as ongoing man-
guarantee the quality of POCT results. For many hospitals agement of patient results (with appropriate normal refer-
and health systems, this consultation is available through ence ranges or interpretive comments) and safe handling
a medical technologist who specializes in POCT quality and disposal of biohazardous patient specimens. CLIA
assurance, called the “point-of-care coordinator.” The compliance is enforced by blocking Medicare and
point-of-care coordinator is the laboratory liaison to the Medicaid payments for noncompliant laboratories and
nursing unit to assist with operator training, instrument levying fines against the organization and laboratory
validation, documentation, and quality improvement. For director, while accreditation agencies can pull a labora-
some health systems, the point-of-care coordinator is tory’s accreditation, institute fines, and publicize citations
extended to physician office laboratories and home nurs- for laboratories with quality concerns. Despite the sale of
ing networks under the health system umbrella, whereas many POCT devices in local grocery stores and pharma-
other private physician offices hire pathologists, doctoral- cies, a physician cannot simply walk into a store, pur-
level laboratory directors, or point-of-care coordinators as chase a device, and start testing patients without first
paid consultants to assist with their POCT regulatory ensuring licensure of the site performing testing and docu-
compliance. The point-of-care coordinator is thus a spe- mentation of quality assurance for regulatory compliance.
cialized medical technologist who has the additional regu- POCT regulations cover laboratory testing conducted
latory knowledge, communication, and management skills for patient care decisions and medical management. The
to help POCT sites improve the quality of their testing portability and ease of use of POCT devices make them
and meet regulatory guidelines. convenient for use in research protocols; however,
research testing is exempt from the CLIA regulations, but
only if no patient care decisions are made from the test
Point-of-care testing regulations results. Staff should be warned that the CLIA guidelines
The challenge of managing POCT is the complexity of still apply to the use of a POCT device as part of a
ensuring regulatory compliance at dozens of sites, with research protocol if a medical decision is made or patient
hundreds of devices and thousands of operators. Each site care is changed based on the test result. For example, the
must supervise equipment, personnel, and safety practices use of a pregnancy test to determine whether a patient can
comparable with performing the testing at a single labora- enroll in a research study and potentially receive a study
tory location; therefore POCT multiplies the many labora- drug is a medical, patient care decision, so all of the
tory issues, like validation and monitoring of devices, CLIA and accreditation guidelines would apply to the
reagent lots, expiration dates, and temperatures across pregnancy test. Alternatively, if the pregnancy test was
dozens of sites. Inspectors use the phrase “site neutral” to performed as part of the study and none of the test results
describe the test requirements that are independent of were communicated to anyone caring for the patient, no
where the test is conducted, and POCT sites are treated change in medical management could occur based on the
like independent laboratories regardless of location. test. This testing would be considered purely research.
The quality of POCT, like other laboratory tests, is When results are recorded in a log to be reviewed after
regulated by the federal CLIA law and inspected through the study completion, the results cannot be utilized for
326 Contemporary Practice in Clinical Chemistry

patient care and the test is classified as research; thus at the time of testing and prompting the operator through
CLIA does not apply in this case. A good rule of thumb the correct sequence of steps to perform the test consis-
to determine if the regulations apply is to answer the tently. Data stored within the device can be downloaded
question, “Will a change in treatment or management to a computer database or POCT data manager, by plac-
occur based on the result of this point-of-care test?” For ing the device into a docking station that connects to the
the pregnancy test, a decision may be made to include the Internet, via modem, or wirelessly to a remote POCT data
subject in the study if the test is negative and to exclude manager. The POCT data manager can also interface with
the subject if the test is positive. This constitutes medical a laboratory information system (LIS) and a clinical data
management, in which a clinical decision (to enroll in the repository (CDR), such as a hospital information system
study) is made based on a test result. CLIA regulations (HIS) or electronic patient medical record, and other clini-
therefore apply in this case, but do not apply when the cal information system (CIS) for permanent storage of the
test results are protected in such a way as to prevent test results.
the use of the result by any clinicians actively caring for The POCT data manager has multiple functions. The
the patient. data manager communicates bidirectionally with devices.
Documentation is fundamental to all of the accredita- From the device, the data manager collects test informa-
tion requirements, and data records and logs are consid- tion and associated date, time, patient identification (ID),
ered the supporting documentation for regulatory operator ID, lot and serial number fields, as well as any
compliance. Some devices provide automated electronic associated result flags and comments. The data manager
capture of the necessary data (patient and operator identi- can also send current lists of competent operators and
fication, date, time, device serial number, reagent lot update any reagent and control lots, expiration dates, and
number, expiration dates, etc.) at the time of analysis. other information to the device remotely. The data man-
However, over 50% of POCT devices are manual, visu- ager acquires control results to document successful per-
ally interpreted dipsticks and kits that require handwritten formance of QC at the required frequency of testing, as
records or manual input into an electronic medical record. well as any corrective actions for failed QC. POCT data
This has led to an increase in the utilization of automated managers use patient identification to search against
readers for manual dipsticks and lateral flow tests that can active patients in the hospital or LIS to transfer patient
automate the interpretation of results and be interfaced results to those systems for permanent storage. Newer
with the patient’s electronic medical record. The amount POCT devices have the capability of positive patient iden-
of data requiring review from POCT can be overwhelm- tification. These devices acquire admissions/discharge/
ing, and computerized POCT data management can assist transfer information on patients through the data manager
personnel in regulatory compliance, billing, and review of and can display the patient’s name on the POCT device
data for quality improvement. Point-of-care coordinators before analysis. The operator then confirms the patient
can assist personnel with interpretation of the regulatory identification by entering a second patient identifier (like
guidelines, development of a quality program, and prepa- birth date) before the device will allow testing. Positive
ration and review of the necessary documentation to meet patient identification reduces the chance of ID entry errors
the specific accreditation guidelines. Some laboratories that can prevent results from being transferred to the cor-
may be accredited by multiple agencies and must prepare rect patient’s medical record. The data manager thus acts
documentation to support inspection against multiple sets as a central processor for information coming from POCT
of recommendations. Federal agencies like CLIA may devices to the patient’s medical record and for informa-
deem equivalence to another agency with stricter recom- tion required by a device for continued operation. POCT
mendations such as New York State, the Joint data managers automate the review of data and reduce the
Commission, or CAP, for instance. Agencies are continu- amount of labor required to manage POCT operations.
ously updating their recommendations, so point-of-care Historically, each manufacturer developed their data
coordinators and staff must keep current on the most manager systems in isolation, meaning that different
recent changes that may affect their POCT operations. POCT devices required different physical cables to con-
nect to the same data manager, and even language and
communication protocols for data were not shared
Point-of-care testing connectivity and between different devices and manufacturers. This com-
plicated and significantly increased the cost of imple-
interfacing menting and changing devices because of the need to
Computerized data management automates the collection purchase additional computer systems and interfaces to
and review of the volumes of data acquired by POCT. accommodate the data transfer. The lack of common com-
Newer POCT devices have computerization that ensures munication standards led to the development of the
regulatory compliance by collecting pertinent information POCT1 standard for point-of-care connectivity by the
Point-of-care testing Chapter | 19 327

FIGURE 19.1 The POCT1 device connectivity standard describes two interfaces. The device interface allows bidirectional communication between
the point-of-care testing devices and the point-of care data managers through docking station access points using the Internet, wireless, or serial com-
munication. The observation reporting interface is an electronic data interface that transfers patient test results and order information between the
point-of-care data managers and the laboratory information system, clinical data repository (an electronic medical record), or other clinical information
system. Reproduced with permission from Clinical Laboratory Standards Institute. Point-of-Care Connectivity: Approved Standard POCT1-A2. CLSI,
Wayne, PA, 2006, p. 306.

Clinical and Laboratory Standards Institute (CLSI) [6]. committees develop high-level policies that can cross
POCT1 was developed by a group of vendors, consumers, departmental and institutional boundaries in hospitals and
and government representatives to enable the seamless health systems with departmental or multihospital admin-
information exchange between POCT devices, electronic istration silos.
medical records, and LIS. This standard defines the physi- A key function of the POCT committee is to review
cal connections as well as the communication protocols and approve requests for new tests and to address compli-
for data transfer between the POCT devices and the data ance concerns with existing testing. Due to its interdisci-
managers (the device interface), as well as between data plinary composition, the committee can balance opinions,
managers, and the LIS, CDR, or CIS through the elec- depersonalize judgments, and deflect the political heat
tronic data interface (Fig. 19.1). The POCT1 standard is that may arise from decisions to remove testing or decline
based on the Institute of Electrical and Electronics the implementation of new testing on specific nursing
Engineers and Health Level Seven standards for medical units or locations. The committee can take the blame for
information transfer, and representatives from these a decision rather than an individual like the POCT coordi-
groups worked with CLSI in the development process. nator or laboratory director. Representation from purchas-
There are currently many devices that comply with the ing and distribution can alert the committee to requests
POCT1 standard, and several POCT data manager sys- for tests outside of the committee approval process if staff
tems currently use this standard to connect to multiple try to go around the committee. Other ancillary staff, like
POCT devices. nutritionists who may change diet based on glucose or
electrolyte test results, may also be useful to have repre-
sented on the POCT committee.
Point-of-care testing quality assurance Pharmacists are becoming more involved in POCT,
not just because of drug dosing based on POCT results,
programs but because pharmacies are now stocking POCT devices
Compliance with regulatory guidelines requires an under- and pharmacists in some hospitals are starting to perform
standing of the regulations, good communication among POCT as an alternative to nursing staff. Pharmacy-based
all the staff involved in POCT, organization, and plan- clinics, in Britain and America, such as minute clinics
ning. The formation of a POCT committee, an interdisci- and urgent care centers, are starting to offer POCT as part
plinary committee to oversee POCT approvals and quality of a rapid care pathway designed to facilitate same day
concerns, can assist in the development of a quality assur- visits when patients cannot get a same-day appointment
ance program and provide a discussion forum for issues with their primary care physician. Patients are often seen
as they arise. It is important to have representation from by nurse practitioners, and POCT is offered to make a
all groups involved in POCT. Administration, laboratory, diagnosis (like rapid strep testing) and get treatment (such
nursing, and physicians are key individuals to have on the as an antibiotic prescription) in the same visit. POCT in
POCT committee, but representation from purchasing, these clinics is conducted at the request of the clinician or
distribution, pharmacy, and nutritional services may pharmacist, but in some locations, patients can even order
also be important when specific issues arise. POCT their own tests and have them conducted by the
328 Contemporary Practice in Clinical Chemistry

pharmacist (direct-access testing). Pharmacists, because with recurrent quality issues should implement additional
of their training and experience with POCT, can provide a follow-up, closer supervision, or other actions to improve
quality result that can be trusted more than patient self- performance. Accreditation inspectors like to see trend
testing at home, and pharmacy testing may be more con- graphs, demonstrating the identification of a problem,
venient for patients than having to wait to schedule a phy- intervention, and improvement of performance over time.
sician visit, get a test order, travel to a phlebotomy station Accreditation agencies used to schedule periodic inspec-
that will then transport a specimen to a central laboratory, tions with advance notification; however, the Joint
and further wait for a test result to be communicated back Commission and CAP now utilize unannounced inspec-
to the physician where a call can be made to the patient tions. Staff must always be ready for an inspection and
to institute treatment. Think of the urine pregnancy test: if will no longer have time to prepare records and quality
the patient performs the test at home, they must then go documentation in advance. Use of quantitative perfor-
to a physician for repeat testing by a central laboratory to mance monitors and integration of the POCT program
verify a positive result. Visiting a pharmacy creates a reli- into the institution quality improvement plan are a good
able test result that can be trusted for ongoing care, means of identifying and continuously improving on pro-
because the pharmacy must follow laboratory quality reg- gram weaknesses. Development of an individual quality
ulations when performing POCT compared with home control plan (IQCP) can identify risk of errors and define
self-testing, which has no quality requirements. a quality assessment for tracking ongoing performance.
A POCT quality assurance program encompasses all Baseline performance assessment and monitoring trends
of the device standardization, validation, training, compe- of improvement over time also provide excellent docu-
tency, and operational aspects of performing POCT. mentation whenever a POCT result is questioned, regard-
POCT should be considered a part of patient care within less of who questions a result—a clinician, a patient, a
the institution rather than an ancillary laboratory service, staff member, or an outside inspector.
so POCT should be incorporated into the overall institu-
tional quality improvement plan. Quantitative measures of
POCT performance, such as compliance errors with moni-
Interdisciplinary communication
toring temperatures, failure to recognize reagent expira- The management of POCT requires the development of a
tions, or corrective actions after control failure, can be quality assurance program and clear communication of
tracked and sorted by site to determine persistent pro- that program to everyone involved. The objectives of
blems and trends (Table 19.3). Nursing units and sites improved patient care through POCT and the path to
achieving this goal need to be clearly defined. Likewise,
the fundamentals of regulatory compliance and quality
TABLE 19.3 List of possible quality improvement improvement must also be spelled out. Unfortunately,
performance monitors for point-of-care testing. clinicians perceive POCT from a different perspective
Successful QC than the laboratory. Technology and nursing has a kind of
G QC documentation “love hate” relationship. Taking the optimistic view,
G Number of errors where wrong QC analyzed (e.g., high technology is intricately linked to the science of nursing
control analyzed as low) and physician practice. POCT provides more rapid clini-
G Percentage of QC that fail cal decision-making, moving patients through the system
G QC outliers with comment
(the hospital or physician’s office) more quickly; improv-
G Failed QC with appropriate action (patients not tested)
ing the efficiency of delivering care; and decreasing
Utilization (number of tests per site or device) patient wait times. However, from the pessimistic view-
G Tests billed versus tests purchased point, technology detracts from the art of nursing and
G Single lots of test and QC in use at any time does not integrate well into the daily patient care respon-
Compliance sibilities. Compliance with regulatory demands and qual-
ity assurance protocols is time- and labor-intensive for
G Untrained operators performing testing
G Clerical errors or data entry errors during testing or result clinical staff and takes the focus of nurses and physicians
reporting away from the patient. Physicians are also experiencing
G Medical record has results reported with reference ranges increased pressure to examine more patients while con-
G Expired reagents and QC/reagent bottles and kits dated ducting successful research programs, maintaining resi-
appropriately
dent education, and developing unique programs in their
G Refrigerator temperature monitored
G Proficiency testing successful discipline for academic advancement. Nurses are encoun-
G Action plan in response to site compliance deficiencies tering increasing pressure to cross-train and diversify,
take on more roles, multitask, and increase the use of
unlicensed personnel for some aspects of patient care.
Point-of-care testing Chapter | 19 329

Nurses and physicians must be responsible for the physi- Successful POCT is based on self-management.
cal, emotional, and spiritual care of the patient. These car- Clinical units make a decision that faster testing will
ing and nurturing roles are in direct contrast to the improve patient care and must allocate the resources nec-
scientific rigors of the laboratory that include accuracy, essary to support the testing, while the laboratory provides
precision, and quality focus. all of the policies, procedures, and information required
Successful POCT management requires an apprecia- for the staff to succeed and continuously improve their
tion of different viewpoints, mutual respect, and compro- performance. One means of distributing POCT informa-
mise. There are always multiple ways to achieve a goal. tion to many sites within an institution is through a dedi-
Conflict with POCT frequently arises from the laboratory, cated website for POCT. A website is easily accessible
dictating how clinicians should act in response to a prob- from anywhere within a health system with only a com-
lem without considering all perspectives. As compliance puter, web browser, and internet connection. An institu-
issues or other differences of opinion arise, participants tional POCT website may contain contact information for
need to look beyond the paradigms of their discipline and the lab director, point-of-care coordinator, and support
think outside the box. In any field, there are shared sets of staff as well as training checklists, policies and proce-
assumptions or paradigms. Paradigms determine what is dures, compliance reports, minutes of the POCT commit-
taught, the methods, and how findings are interpreted. tee, current POCT projects, and links to the FDA
Nursing has one set of paradigms, the laboratory has laboratory safety tips or current CAP/Joint Commission
another, and physicians have a third set of paradigms. accreditation checklists. A website could also contain a
When a compliance issue is noted, like an unapproved list of POCT contacts on each of the clinical units and fre-
POCT device, the laboratory may interpret this as a will- quently asked questions written in a language that the
ful breach of policies and failure to go through the POCT clinical staff can understand and relate. The website could
committee approval process. However, the physicians contain compliance trends by unit of their performance
may perceive this only as an attempt to improve the care monitors, so units can compare their performance to
of their patients and improve turnaround time of results. others and help prioritize their action plans. Self-
They may not be aware of the stringency of laboratory management encourages clinical staff to take charge of
regulations and the need for committee approval or device their own testing.
validation before using a POCT device. Nursing, on the Many institutions have adopted electronic document
other hand, was conducting the test, because the physician control systems that can be an alternative to creating a
ordered it and was only following instructions. The prob- dedicated POCT website. Electronic document control
lem here is a lack of understanding about the regulations systems allow the procedures to be available on the nurs-
governing POCT and a failure to communicate effectively ing units from any of the clinical workstation computers,
the issues. A clear discussion of the POCT committee pol- so staff does not have to hunt for a hardcopy of a proce-
icies and the reasoning behind the policies will be more dure when conducting testing. Electronic document sys-
successful than dictating compliance. When staff under- tems remind staff for required annual review of policies
stand why they must follow policies, they will be more and provide an easy means of distributing policy revisions
likely to comply with QC performance, refrigerator tem- to all the sites performing POCT simultaneously. Most
perature monitoring, and seeking approval before purchas- importantly, POCT policies and procedures can be inte-
ing new POCT devices. grated with other nursing policies in format and wording,
emphasizing the role of POCT in patient care rather than
as a “laboratory” service.
Self-management Institutions with a large number of staff have also
POCT is a decentralized process, and the laboratory can- adopted electronic training and competency systems.
not be at every site 24 hours a day to monitor test perfor- These programs can track operator competency and send
mance. The laboratory has to delegate responsibility for messages to those pending expiration of their training.
POCT to the clinical staff and trust that nursing and phy- Training/competency systems can utilize slides, video
sicians will properly manage the testing process. In this with sound, exams, and other tools to refresh staff on key
model, the laboratory takes a central oversight role for points to remember when performing POCT and ensure
POCT, but the clinical staff takes responsibility for the that these points were received and understood. CLIA and
day-to-day operations such as purchasing reagents, dis- accreditation agencies require six elements of competency
carding outdated tests, validating staff competencies, and to be evaluated annually including:
performing QC and other functions needed to maintain
1. Direct observation of routine test performance
testing on the medical unit. This laboratory clinical part-
2. Monitoring the recording and reporting of test results
nership is based on mutual respect and shared responsibil-
(including critical results)
ity for POCT quality.
330 Contemporary Practice in Clinical Chemistry

3. Review of intermediate test results, worksheets, QC American Association for Clinical Chemistry (AACC)
records, proficiency test results, and preventive main- has a community forum, the AACC Artery, to discuss
tenance records POCT issues (www.aacc.org/community/forums). AACC
4. Direct observation of preventive maintenance and also offers a POC Bootcamp and certification program for
function check performance coordinators who are managing POCT programs. There
5. Assessment of test performance using previously ana- are also a number of regional point-of-care coordinator
lyzed specimens or proficiency testing samples groups throughout the country, and their meetings, activi-
6. Assessment of problem-solving skills ties, and original articles are available on a POCT website
(www.pointofcare.net). Point-of-care coordinators and
Many of these elements can be assessed through an
staff involved in POCT thus have a variety of resources
electronic training/competency system with an exam that
and ways to contact others involved in POCT to get
includes troubleshooting questions, but direct observation
answers to their questions, seek advice, and get feedback
of test performance and comparison of test results
from staff who are experiencing similar challenges.
requires one-on-one time of a staff trainer with each
POCT operator. Observing test performance cannot be
automated through an electronic training system.
Staff members who perform testing are wholly respon-
Analytical performance
sible for their success as well as their failure. By position- Although regulatory compliance and management is a sig-
ing the laboratory as a resource for POCT and distancing nificant aspect of POCT, the analytical performance of
the laboratory from the performance aspects of the POCT devices also contributes to the overall quality of results
program, staff members tend to help each other, commu- and clinical applicability of a test. Poor testing will result
nicate more about their issues, and work more as a team when staff fail to follow manufacturer’s instructions on a
than when the laboratory is mandating compliance and device with excellent analytical performance; it is the
dictating practice. Each unit can find unique solutions to same as if staff do everything right, but picks a device
issues that work best within the unit’s workflow. Issues with poor analytical performance. Technical performance
are brought to staff attention by the laboratory and staff is and management practices both contribute to quality
encouraged to inspect their own practice and determine POCT. Therefore institutions must be concerned about
the solution that operationally works best for them. For picking the right test for the specific patient as well as
instance, two levels of QC must be analyzed each day of ensuring that the test is conducted properly, and QC and
testing, but this can be accomplished in a number of quality assurance practices are followed.
ways. Nursing units should rotate the responsibility The required performance characteristics of a device
among different staff, but a nursing unit might choose to must match the clinical need of the patient and the ques-
have the first shift run QC, because it works better into tion being asked by the physician. POCT devices are dif-
their patient rounds, while another unit might find that ferent methods than core laboratory instrumentation. Each
night shift works best when staff are not as busy. By facil- method has unique bias, imprecision, and analytical inter-
itating communication between the POCT sites and the ferences that may limit its clinical utility. Glucose meters,
encouraging staff to solve their own problems in a regula- for instance, have greater imprecision and are susceptible
tory compliant manner, the laboratory becomes a resource to different interferences than central laboratory methods.
of information to help staff improve their own program. Despite recommendations by the American Diabetes
Staff can then find the best ways to integrate POCT into Association for agreement of glucose meters within 5% of
their current workflow without the laboratory interfering. a central laboratory method, none of the currently mar-
This improves interdisciplinary relations between the clin- keted glucose meters can meet these performance stan-
ical sites and the laboratory and promotes collegiality dards [9]. Glucose meters are thus not recommended for
rather than animosity. use in screening or diagnosis of diabetes [10,11]. The var-
There are many POCT resources that are available to iability of glucose meters limits their use to management
assist staff with managing POCT. A journal, Point of of patients already diagnosed with diabetes. CLSI recom-
Care: The Journal of Near-Patient Testing & Technology mends that glucose meters agree with a comparative cen-
(www.poctjournal.com), is dedicated to basic research, tral laboratory method within 6 12 mg/dL for results
method validation data, regulations, and other topics ,100 mg/dL and 6 12.5% for results $ 100 mg/dL [12].
related to POCT management and device technology. The Glucose meters are too variable for use in analyzing spe-
Journal of Analytical Laboratory Medicine also publishes cimens collected after a glucose tolerance test for diagno-
peer-reviewed research and offers case studies, focused sis of diabetes, and meter variability also limits the utility
studies, and laboratory reflections involving POCT. of meters in health fair and shopping mall screening pro-
Several textbooks are also devoted to POCT [7,8], and the grams. Device performance requirements can change with
Point-of-care testing Chapter | 19 331

the concentration of analyte and the clinical application provide a faster glucose measurement, but the variability
(screening, diagnosis, or management). of the method may not allow the device to be used for
Operators are a key part of device performance, and diagnosis or screening of general patients, while central
there are several studies indicating poorer performance by laboratory glucose results can be utilized for these situa-
staff without laboratory training than operators with train- tions. This is why clinical validation of a device on the
ing. The CDC compared performance on proficiency test- intended population using routine staff operators at that
ing samples sent to laboratories performing POCT [13]. location is so important before implementation of a test
Proficiency samples are specimens of unknown concen- for patient care.
tration analyzed like patient specimens where the results POCT devices are portable, so reagents and test strips
are graded and compared with all other laboratories per- may be exposed to varying environmental conditions
forming the same test. The CDC noted significantly better compared with testing in a well-controlled laboratory.
performance on proficiency samples for medical technolo- Visiting nurses travel to patient’s homes, and tests can be
gists working in accredited hospital laboratories than non- exposed to cold in the winter and heat in the summer.
laboratory staff performing testing in physician office Humidity can damage colorimetric test strips, like urine
laboratories. Intermediary performance was seen at those dipsticks. Altitude may also affect some tests that are sen-
sites performing testing with laboratory-trained operators. sitive to oxygen, like glucose and blood gas tests. Storage
Training improves performance, and the Oregon Health and environmental conditions are factors that need to be
Sciences University noted significant improvement in controlled to ensure quality test results. Tests and devices
both the performance of glucose testing and interpretation should be removed from vehicles when not in use to pre-
of results after a standardized training program [14]. vent freezing and overheating. Strip vials also need to be
Performance not only improved immediately after training tightly recapped and tests stored in cool, dry places away
but continued to improve consistently for the following from radiators and windows where the tests may be dam-
several months that were monitored by the study. aged from heat, light, and moisture.
Operators and the training of operators can thus affect the Drugs, disease metabolites, and diet can affect POCT
quality of POCT results as well as the interpretation and results in different ways compared with traditional labora-
treatment decisions based on POCT results. tory methods. Some glucose strips are sensitive to salicy-
Standardization is an important aspect of successful late, acetaminophen, and urea. Vitamin C in high doses
POCT programs. Selecting one manufacturer for glucose can act as an electron carrier in POCT glucose reactions.
meters and one kit for pregnancy testing simplifies quality Dietary constituents affect other POCT methods. Meat
management. With dozens of POCT sites and thousands and peroxidase-containing vegetables, for instance, may
of operators, increasing the number of different devices give false-positive reactions in guaiac-based occult blood
complicates the ability to manage the program and stres- tests. Hydration status and drinking excess water can
ses available staff and resources. Validation, procedures, dilute urine, generating false-negative drug and pregnancy
and training are required for each device for regulatory tests. Hemolysis of blood elevates potassium levels and
compliance. CAP also requires subscription to a may not be detected in a whole blood sample used for
proficiency-testing program. As the number of different POCT compared with serum with laboratory analysis
devices multiplies, there are more procedures and training where hemolysis is evident as red color in the sample; so
to maintain, and staff who rotate or test patients infre- hemolysis can be overlooked with POCT, leading to
quently may not remember how to operate the test when falsely increased potassium results and the potential for
required. For POCT, simplicity is fundamental. Fewer misinterpretation and inappropriate clinical management.
devices, limited testing sites, and reducing the number of Critical illness is another factor to consider with
operators simplify the management of testing and enhance POCT. Patients presenting with extremes of physiology
compliance. With fewer devices and a smaller group of challenge the performance of POCT devices. Glucose
key POCT operators, it is easier to train staff and more meters originally developed for use in patient self-testing
economical to manage the program. may perform differently when used in a hospital environ-
ment. Diabetic outpatients are ambulant and generally
healthy compared with critical inpatients on oxygen ther-
Method limitations apy and various medications. Critically ill patients also
POCT methods are different from clinical laboratory have low hematocrits that can interfere with glucose
methods and have their own biases and interferences. methodologies. Manufacturers have been encouraged by
Clinical treatment protocols and pathways of care based the FDA to include a limitation in the meter package
on central laboratory methods may not necessarily be insert that “the performance of the test system has not
transferable to the same patient populations when using been evaluated in the critically ill.” While newer meters
POCT results. As noted already, glucose meters may have received FDA approval for use in critically ill
332 Contemporary Practice in Clinical Chemistry

populations, consumers are still warned of interferences fluoride and oxalate, can interfere with some POCT meth-
with meter results when utilizing capillary fingerstick ods, so correlations of glucose meters against laboratory
samples in patients with poor peripheral circulation. This methods must quickly analyze a whole blood specimen on
limitation warns the consumer of the potential for interfer- the glucose meter and either simultaneously collect a sep-
ence in subpopulations of the critically ill that may show arate sample for laboratory analysis, or centrifuge and
variations in results with certain sample types and condi- separate the remainder of the POCT specimen. Once sep-
tions. Consumers should restrict use of glucose meters in arated from cells, glucose is stable in plasma/serum for
order to adhere to package insert limitations, such as several days refrigerated and longer if frozen. While use
acceptable hematocrit and oxygen ranges, and prohibit of whole blood for POCT and serum/plasma for labora-
use of capillary fingerstick samples from patients with tory analysis ensures appropriate specimen types for each
poor peripheral circulation (i.e., diabetic ketoacidosis, methodology, the glucose concentration may not match
nonketotic/hyperosmolar conditions, congestive heart fail- between the two methods because of physiologic differ-
ure stage IV, shock, and peripheral vascular disease) ences between whole blood and serum/plasma (Fig. 19.2).
[15,16]. Blood can pool in the periphery and capillary Glucose distributes within the aqueous portion of
samples in such patients may not reflect the physiology of blood. Since there is less water inside erythrocytes (73%
central circulation. Currently, only one glucose meter has water/27% lipids and protein) compared with the plasma
been FDA-cleared for use in critically ill patients using portion of blood (93% water/7% lipids and protein), the
capillary fingerstick samples, but the analytical perfor- glucose concentration will be lower inside red blood cells
mance of this meter is significantly more variable than than plasma from the same patient specimen. The actual
laboratory methods. Clinicians are cautioned to interpret difference will depend on the hematocrit (concentration
POCT results with care in patients with rapidly changing of erythrocytes in a sample). At a normal hematocrit of
physiologic conditions, particularly when analyzing capil- 45%, plasma is B11% higher than whole blood. Patients
lary samples in critically ill patients despite FDA approval with higher hematocrits will have greater difference
for use in these populations. between whole blood and plasma/serum glucose, while
Use of venous or arterial samples may be a more patients with lower hematocrits will have less difference.
acceptable alternative in patients with poor peripheral cir- Method comparisons and validation of POCT devices
culation, and capillary sampling is affected. Alternatively, must take into account these physiologic differences in
hospitals may choose to use a different glucose methodol- specimen matrix when interpreting the agreement between
ogy such as a blood gas analyzer or send a sample to the different methods. To simplify interpretation, most
central laboratory [15]. However, delays in test turn-
around may prohibit sending a sample to a central labora-
tory. Hospitals can also choose to use a glucose meter
off-label by analyzing capillary samples despite the manu-
facturer warnings for critically ill patients. Caution should
be taken, however, since off-label use shifts the CLIA-
waived glucose meter to a high-complexity category
requiring additional validation and documentation. CLIA
high-complexity may further limit the staff who can per-
form testing, as hospitals frequently use patient care assis-
tants, staff with only a high school education and on the
job training, to perform glucose testing [15,17]. Under
CLIA high complexity, staff with only a high school
diploma would not be eligible to conduct glucose meter
testing, and more advanced clinical staff, such as regis-
tered nurses or nurse practitioners, with additional labora-
tory training, experience, and licensure (in some states)
may be required.
POCT employs unprocessed specimens. Whole blood
is analyzed on POCT devices, whereas most laboratory
methods require serum or plasma. Glucose, however, is FIGURE 19.2 Whole blood to plasma differences. The physiologic
unstable in whole blood. The specimen must be analyzed water content in whole blood is a mixture of the plasma and erythrocyte
red blood cell water and varies with the hematocrit or volume of red
quickly after collection, or inhibitors of cellular glycoly- blood cells in blood. Analytes that follow water, like glucose, electro-
sis, like fluoride or oxalate, must be added to stabilize the lytes, and some drugs, will show similar differences in concentration
glucose levels in the sample. Specimen additives, like when measured in whole blood versus plasma.
Point-of-care testing Chapter | 19 333

manufacturers calibrate their glucose meters to read a concentration and can be biased in surgical and trauma
plasma equivalent result despite accepting whole blood patients who receive transfusions and volume replace-
samples. Glucose meters should theoretically be compara- ment. So POCT results for hemoglobin and hematocrit
ble with central laboratory methods in patients with nor- should specify not only that result was performed point-
mal hematocrits. Differences in hematocrit are one reason of-care, but whether the result was calculated or mea-
why critically ill patients and neonatal populations with sured. These are only a few of the interface challenges to
low and high hematocrits may show biases between the consider because of the unique method differences and
two methods, and why validation of POCT in the intended biases that can occur with POCT in some patient
patient population is so important. populations.
There are a variety of factors that can affect POCT
differently from laboratory methods. These factors must
be considered when implementing POCT or when inter-
Risk management
preting the POCT literature. A device that works well on POCT systems are not perfect. All devices have limita-
one medical unit and patient population may show differ- tions. When used outside manufacturer recommendations
ent results in another patient population due to medica- or used improperly, incorrect results can be generated.
tion, diet, or operational differences between the two Errors can be more likely with POCT because of the num-
sites. Validation of POCT devices by staff operators on ber of operators involved, the volume of tests being con-
the intended patient populations will most likely uncover ducted, and the various locations where testing is
interferences and other potential problems before imple- performed. POCT is often delegated to the lowest level of
mentation and routine use of a device. When environmen- staff who have minimal training and experience. Staff
tal effects or unexpected interferences are discovered, shortcuts, differences in operator technique, and failure to
practices can be changed to control against test exposure appreciate patient and preanalytic factors can all compro-
or alternative devices can be selected, which do not expe- mise test quality. POCT is chosen over laboratory testing,
rience the same interferences. because it is rapid and convenient. POCT can be per-
The potential for interference, operator, and environ- formed at a variety of locations with the opportunity for
mental effects sets POCT apart from central laboratory the device and the reagents to be exposed to heat, cold,
methods. Results from POCT should be clearly separated and humidity. The number of operators, tests, and loca-
from central laboratory test results in the patient’s chart. tions multiplies the complexity of managing the testing
As hospitals move toward electronic records, POCT result process in a quality manner. Environmental exposure,
interfaces should distinguish the source of the test result sample, and operator errors can all lead to incorrect
for the ordering clinician. Clinical systems have the abil- results.
ity to overlay test results, so all glucose values or sodium Risk management can reduce errors. Manufacturers
values, for instance, can be viewed and trended together. employ risk management when developing new methods
From the clinician’s perspective, seeing more information in order to document that common hazards or the poten-
together on a computer screen saves time and prevents tial for errors have been addressed by the device engineer-
scrolling through result spreadsheets and multiple mouse ing and that the limitations have been outlined in the
clicks. However, differences in methodologies between package insert and instructions for use. The FDA consid-
POCT and central laboratory will cause results to jump ers the manufacturer risk management plan when review-
around when plotted or trended together. This can lead to ing the device for market approval. This plan outlines
clinical confusion and chasing results where treatment is how the device is intended to be used (use-case scenario)
based on one result, but the next result comes back too and how the manufacturer addresses or detects potential
low. Based on this, therapy is readjusted; then the next misuse. With POCT, a device may be brought to the
result is too high, so therapy is adjusted again. Combining patient’s bedside for testing, or alternatively a sample
results from multiple methods can lead to such clinically may be collected and brought to a device in a spare utility
confusing trends, overtreatment, and frequent changes in room on the nursing unit for testing. Moving a sample
treatment that would not occur if the individual methods away from the patient presents the potential to mix-up
were separated in the patient’s medical record. samples with other patients tested in that utility room,
Other distinctions that should be identified in the med- unless the sample is labeled appropriately. These two use-
ical record are calculated and measured results. Some case scenarios present different opportunities for particu-
blood gas analyzers measure hematocrit by conductivity lar errors, such as specimen mix-up. Risk management
and calculate hemoglobin by a common conversion examines each step in the testing process for weaknesses
(hemoglobin 5 hematocrit/3), while other analyzers mea- where errors may occur and defines actions that may be
sure hemoglobin spectrophotometrically and calculate required to reduce, prevent, or detect those errors before a
hematocrit. Conductivity hematocrit is sensitive to protein result is released and action is taken by clinicians. So
334 Contemporary Practice in Clinical Chemistry

laboratories can learn to reduce their chance for test errors devices and menu of tests expands, the POCT market can
by adopting industrial risk management principles. be expected to grow. The major concern with POCT is
Testing liquid QC specimens in a manner like patient providing quality test results with multiple devices, at
specimens have historically been utilized to prove the dozens of sites by thousands of operators. Management of
ability of a test system to achieve a test result with the POCT, therefore, requires organization. POCT manage-
expected level of quality; however, liquid controls have ment can best be organized by forming a POCT commit-
limitations. With POCT, analyzing a QC sample con- tee to supervise new test requests and set policies,
sumes a single-use test kit, and there is no guarantee that standardizing and minimizing the number of devices,
the next kit will perform comparably. Based on this, man- planning implementation through validation of devices
ufacturers have developed a number of control processes and training of operators, and providing consistent labora-
engineered into the test or actions required to be per- tory oversight to comply with regulatory requirements.
formed by the laboratory to achieve quality results. POCT is a different method and may not be freely inter-
Newer devices have clot and bubble detection, volume changeable with laboratory results in clinical protocols.
detection to ensure adequate sample was applied, and Each method has unique interferences, biases, and impre-
internal controls to prove the viability of reagents such as cision that can lead to disparate results in certain patient
internal controls on pregnancy, HIV, and drug tests. populations. The potential for method differences stresses
Reagents are being shipped with a temperature monitor to the need to validate devices at the intended clinical sites
ensure the package has not been exposed to heat or cold by using clinical staff operators to test typical patients
temperatures outside of manufacturer specifications dur- before the decision to implement a device for patient
ing shipping. Staff are expected to check the temperature care. Newer devices are computerized and can automati-
monitor on receipt of each shipment and verify accept- cally collect and transfer the test results and QC data nec-
ability before attempting to place the kits in use for essary to document regulatory compliance. Interfaces of
patient testing. Each POCT device and test kit is unique POCT results should distinguish POCT results from labo-
and may present different opportunities for error depend- ratory results in order to prevent clinical confusion that
ing on how the device is used by the consumer. Applying can occur from overlaying results with unique biases in
risk management to trace the sample through the testing certain patient populations and clinical situations. Many
process can reveal hazards or potential sources for error newer POCT devices also offer improvements in detect-
and the actions that are required to prevent and detect ing common errors through internal control processes.
those errors. CLSI has published a guideline for assessing Risk management offers a means for laboratory directors
the risk of errors from implementing unit use testing to determine the required actions and optimal control
devices, titled “Quality Management for Unit-Use strategies to minimize risk of errors for any specific
Testing” [18]. This document advocates the use of an device in their healthcare setting and use-case scenarios.
error matrix to list and prioritize the frequency and sever- POCT has the potential to provide faster results and
ity of potential errors associated with testing devices. This improve patient outcomes through quicker therapeutic
guideline also emphasizes a shared responsibility between intervention provided that the clinician understands the
the manufacturer and the consumer of the testing device method limitations and when to use the devices in patient
in developing appropriate control strategies to manage care appropriately. POCT is thus a complex system with
errors. Another CLSI document, EP 23-A, titled many factors to consider and manage to obtain optimal
“Laboratory Quality Control Based on Risk Management” benefits.
[19], outlines a strategy to develop an IQCP for the labo-
ratory test based on risk management. An IQCP basically
summarizes many of the quality activities a laboratory is
References
already doing in a more formal manner. The IQCP is dis-
cussed in further detail in Chapter 5, Quality control. [1] J. Greyson, Quality control in patient self-monitoring of blood glu-
Manufacturers have been conducting risk assessments for cose, Diabet. Care 16 (1993) 1306 1308.
years and must submit their data for FDA review when [2] W.A. Rutala, V.A. Kennedy, H.B. Loflin, F.A. Sarrubi Jr., Serratia
marcescens nosocomial infection of the urinary tract associated
seeking market approval of new tests. CLSI EP23 and the
with urine measuring containers and urinometers, Am. J. Med. 70
IQCP introduce these industrial risk management princi-
(1981) 659 663.
ples into the clinical laboratory. [3] R. Webb, M. Currier, J. Weir, K.M. McNeill, E. Bancroft, D.
Dassey, et al., Transmission of hepatitis B virus among persons
undergoing blood glucose monitoring in long-term care facilities—
Summary
Mississippi, North Carolina and Los Angeles County, California,
POCT is an increasingly popular means of delivering fas- 2003 2004, MMWR Morb. Mortal. Wkly. Rep. 54 (2005)
ter test results closer to the patient. As the number of 220 223.
Point-of-care testing Chapter | 19 335

[4] Department of Health and Human Services, Office of Inspector 120pp. , http://www.aacc.org/science-and-practice/practice-guide-
General. Enrollment and certification processes in the Clinical lines/diabetes-mellitus. (accessed 05.03.2020).
Laboratory Improvement Amendments Program. OEI- [12] Clinical Laboratory Standards Institute, Point-of-Care Blood Glucose
05 0000251. Chicago, IL: Office of Evaluation and Inspections, Testing In Acute and Chronic Care Facilities; Approved Guideline
2001, p. 18. ,http://oig.hhs.gov/oei/reports/oei-05-00-00251.pdf. Third Edition POCT12-A3, CLSI, Wayne, PA, 2013. 49pp.
(accessed 05.03.2020). [13] Centers for Disease Control, Laboratory Practice Assessment Br,
[5] CMS. CMS certificate of waiver and provider performed micros- Division of Laboratory Systems, Public Health Practice Program
copy procedures pilot project. Baltimore, MD: CMS, 2003, p. 30. Office. Clinical laboratory performance on proficiency testing
,http://www.cms.hhs.gov/CLIA/downloads/ppmpfr2001.pdf. samples United States 1994. MMWR Morb Mortal Wkly Rep
(accessed 05.03.2020). 1996;45:193 6.
[6] Clinical Laboratory Standards Institute, Point-of-Care [14] R. Belsey, J.I. Morrison, K.J. Whitlow, D.M. Baer, S. Nelson, D.
Connectivity: Approved Standard POCT1-A2, CLSI, Wayne, PA, F. Hardwick, Managing bedside glucose testing in the hospital,
2006. 306pp. JAMA 12 (1987) 1634 1638.
[7] G.J. Kost (Ed.), Principles & Practice of Point of Care Testing, [15] Clinical Laboratory Standards Institute, Use of glucose meters for
Lippincott, Williams & Wilkins, Philadelphia, PA, 2002. 654pp. critically ill patients. POCT17, CLSI, Wayne, PA, 2016. 15 pp.
[8] J.H. Nichols (Ed.), Point-of-Care Testing: Performance [16] IFCC Working Group GMECC. How should glucose meters be
Improvement and Evidence-Based Outcomes, Marcel-Dekker, evaluated for critical care? Milan, Italy: International Federation
New York, NY, 2003. 500pp. of Clinical Chemistry and Laboratory Medicine, 2018, p. 48.
[9] E.T. Chen, J.H. Nichols, S.H. Duh, G. Hortin, Performance evalu- ,http://www.ifcc.org/ifcc-education-division/emd-committees/c-
ation of blood glucose monitoring devices, Diabetes Technol. poct/wg-gmecc/. (accessed 05.03.2020)
Ther. 5 (2003) 749 768. [17] US Department of Health and Human Services, Medicare,
[10] D.B. Sacks, D.E. Bruns, D.E. Goldstein, N.K. Maclaren, J.M. Medicaid, and CLIA programs; laboratory requirements relating
McDonald, M. Parrott, Guidelines and recommendations for labo- to quality systems and certain personnel qualifications; final rule.
ratory analysis in the diagnosis and management of diabetes melli- 42CFR Part 493, Fed. Regist. 68 (2003) 3639 3714.
tus, Clin. Chem. 48 (2002) 436 472. [18] Clinical Laboratory Standards Institute, Quality Management for
[11] D.B. Sacks, M. Arnold, G.L. Bakris, D.E. Bruns, A.R. Horvath, M. Unit-Use Testing: Approved Guideline. EP18-A, CLSI, Wayne,
S. Kirkman, et al., Guidelines and recommendations for laboratory PA, 2002. 38pp.
analysis in the diagnosis and management of diabetes mellitus: [19] Clinical Laboratory Standards Institute, Laboratory Quality
Updated draft guidelines, National Academy of Clinical Control Based On Risk Management: Approved Guideline EP23-
Biochemistry Laboratory Medicine Practice Guidelines, 2011. A, CLSI, Wayne PA, 2011. 73pp.
336 Contemporary Practice in Clinical Chemistry

Self-assessment questions b. Stopping reimbursement from Medicare and Medicaid


c. Closing the laboratory or physician’s office perform-
1. What is the definition of point-of-care testing? ing testing.
a. Testing conducted close to the site of patient care d. All of the above
b. Bedside testing 7. POCT device connectivity and the POCT1 standard
c. Portable testing in an ambulance or helicopter are important for what reason(s)?
d. All of the above
a. Allows multiple devices to connect to a single data man-
2. Which of the following tests are available from POCT ager and share an interface with a laboratory or HIS
devices? b. Allows collection and review of data from multiple
a. Growth hormone sites
b. Dihydrotestosterone c. Facilitates implementation of new POCT devices
c. HIV antibody through plug-and-play connections
d. Renin d. All of the above

3. What are the advantages of POCT? 8. Limitations of POCT devices include which of the
following?
a. Larger sample
b. Faster turnaround time for test results a. Ability of physicians to rapidly treat patients
c. Less expensive than core lab testing b. Environmental and drug interferences with rapid
d. More precise than core lab instrumentation methods
c. Ease of use and training of the operator
4. Which regulations apply to POCT devices? d. Small sample volumes
a. Federal CLIA regulations 9. Which of the following is a QC process?
b. State law
c. Private accreditation agency standards like The Joint a. A whole blood control above the medical decision
Commission limit for a test
d. All of the above b. A colored disk to check the optics of a spectrophoto-
metric device
5. Which of the following are true? c. A separate control line that develops with each test on
a. POCT results can lead to wrong treatment, and care a pregnancy strip test
should be taken to guarantee quality. d. All of the above
b. POCT devices are so simple that they pose no risk to 10. What is risk management?
the patient if incorrectly performed or a wrong answer
is generated. a. A requirement for FDA approval of new devices
c. POCT results are used only for screening patients, and b. An industrial process for defining potential errors in
results are always confirmed by a lab method. the use of a device
d. Clinicians can purchase POCT devices from a local c. A means for clinical laboratories to prevent and detect
pharmacy without having to comply with laboratory errors
regulations if the tests are used only on their own d. All of the above
patients in an office practice.
6. The DHHS enforces CLIA standards in which way(s)?
a. Limiting the physician’s medical license to practice
Answers
1. d
2. c
3. b
4. d
5. a
6. b
7. d
8. b
9. d
10. d

You might also like