OPERATING MANUAL PAGE 1 1) Please Read The Instruction Very Care Fully. 2) Storage Temperature: 0 + 50C. 3) Excitation Voltage : 5 vDC 4) Input : 0 to 25mv DC 5) Dimention : 96(H) x 96)W) x 85(D) mm 6) SCALES SUPPLIED a) 2 Relay O/p Indicator 7) Key Configuration: a) “ E” Escape key – Return from configuration c) “ Enter “ For Entering Date & Configuration b) “ UP/TARE “ For increase the value & TARING c) “ DOWN “ For Decrease the value
8) Set Decimal Point : while showing display 00000
Press Enter key 0000 (Password, factory setting 0000) Press ENTER SETP (set up) DP (Digit point) 0000.0 (set by up key) Press E key. completed . 9) Set Alaram :Cut Off Weight(Relay Set Point) Press Enter key 0000 (Password, factory setting 0000) Press ENTER SETP (set up) DP (Digit point) Press UP key RELY Press Enter RLY 1 H – ON 02000(for 2000 kg) Page 2 HYST 000.1 RLY 1 Press UP key RLY 2 H – ON 03300 (for 3300 kg) HYST 000.1 Press E key. completed 10) Calibration : Press Enter key and turn on Power supply PASS Press ENTER 0000 Press Enter CAL Press Enter Inp Press Enter ZERO (Wait for 10sec and set Zero Load) Press Enter SPAN (Put Maximum capacity weight on Platform & wait for 10 sec.) Press Enter key ***** (set count by up/down key equivalents to Load )(0000.1 for 100kg) Press Enter 00001 Press Enter Imp Press Esc key 0000.1(for 100kg) Calibration completed. 11) Loadcell connection : 5V – Loadcell Red wire 0V – Loadcell Black wire LC + – Loadcell Green wire(As per Loadcell) LC - – Loadcell White wire (As per Loadcell)