DELMARK

THE TRUTH OF WEIGHING


OPERATING MANUAL
PAGE 1
1) Please Read The Instruction Very Care Fully.
2) Storage Temperature: 0 + 50C.
3) Excitation Voltage : 5 vDC
4) Input : 0 to 25mv DC
5) Dimention : 96(H) x 96)W) x 85(D) mm
6) SCALES SUPPLIED
a) 2 Relay O/p Indicator
7) Key Configuration:
a) “ E”
Escape key – Return from configuration
c) “ Enter “
For Entering Date & Configuration
b) “ UP/TARE “
For increase the value & TARING
c) “ DOWN “
For Decrease the value

8) Set Decimal Point : while showing display 00000


Press Enter key
0000 (Password, factory setting 0000)
Press ENTER SETP (set up)
DP (Digit point)
0000.0 (set by up key)
Press E key. completed
.
9) Set Alaram :Cut Off Weight(Relay Set Point)
Press Enter key
0000 (Password, factory setting 0000)
Press ENTER SETP (set up)
DP (Digit point)
Press UP key RELY
Press Enter RLY 1
H – ON
02000(for 2000 kg)
Page 2
HYST
000.1
RLY 1
Press UP key RLY 2
H – ON
03300 (for 3300 kg)
HYST
000.1
Press E key. completed
10) Calibration :
Press Enter key and turn on Power supply
PASS
Press ENTER 0000
Press Enter CAL
Press Enter Inp
Press Enter ZERO
(Wait for 10sec and set Zero Load)
Press Enter SPAN
(Put Maximum capacity weight on Platform & wait for 10 sec.)
Press Enter key *****
(set count by up/down key equivalents to Load )(0000.1 for 100kg)
Press Enter 00001
Press Enter Imp
Press Esc key 0000.1(for 100kg)
Calibration completed.
11) Loadcell connection :
5V – Loadcell Red wire
0V – Loadcell Black wire
LC + – Loadcell Green wire(As per Loadcell)
LC - – Loadcell White wire (As per Loadcell)