CERTIFICATE OF CALIBRATION “ent Ati Nason
[ABORATORIUM KALIBRAS!
UG 12LIDN
Issued By _—~PT. Caltest Engineers Cerlificate Number
Date of Issue 12-Nov-19 2076891152128
Page 2 of2
‘Traceability Information
The Stindard used to calibrate is traceable to SI unit through the Calibration Lab
contained in the column traceability as below:
Instrument Description SN Traceability Due Date
Pressure Indicator 3312671 LK-121-IDN 11-Mar-20
Calibration Result
: Reading of Standard Contion
Leicel AsFomd ‘Aste As Found Aaah Ascoaar
Insure | Aaoending | Dessoning Tacending | Descending oa
Ascending | Descends ‘Ascending | Descending} si
psi psi psi i "S| psi si a ad
0 00 00 | same [same | 0.0 7 : 30 | pass
100 100.1 | 100.1 | same } same | out - : 5.0 | pass
200 1999 | 2001 | same | same | 01 - : 50 | pass
300 2996 | 3000 | same | same | 04 - . 50 | pass
400 309.1 | 3998 | same | same | 09 - 2 50 | pass
500 498s | 4985 | same } same | -15 : : 5.0 | pass
Remarks
~ The estimated measurement uncertainty associated in this calibration is withi 2.7 psi at 95% confidence level
with coverage factor of k= 1.96
ceasorom 36