Professional Documents
Culture Documents
SERVOPRO MultiExact 4100 Brochure
SERVOPRO MultiExact 4100 Brochure
MultiExact 4100
FLEXIBLE
Comprehensive solution for Digital communications for remote
industrial and medical gas access: RS232/RS485 Modbus and
manufacture and for PROFIBUS (Ethernet Modbus TCP/
pharmacopeia applications IP to follow)
SERVOPRO MultiExact 4100 easures up to four gas streams
M Up to 32 alarms and 32 relays
simultaneously with follow or freeze options
Integrated support for the Four analog outputs and four
The SERVOPRO MultiExact 4100 is a high- AquaXact 1688 Aluminum Oxide analog inputs
performance multi-gas analyzer designed moisture transmitter
to provide up to four simultaneous gas
stream measurements including: O2 (trace,
control, and purity), CO2, CO (% and EASY TO USE
trace), N2O, CH4 (trace), Ar in O2, N2 in Ar,
O2 or air, and He in Ar, O2 or N2.
Intuitive-use icon-driven color
Analyzer configurations can
touchscreen for easy device be easily duplicated using USB
Specifically designed to address the
interaction and configuration thumb drive
needs of industrial and medical gas
manufacturers, the MultiExact 4100 can
USB serial port for data logging
be fitted with a wide range of sensing and software upgrades
technologies – Paramagnetic, Zirconia,
Gas Filter Correlation Infrared (GFx),
single-beam Infrared (SBSW), Aluminum LOW COST OF OWNERSHIP
Oxide, and Thermal Conductivity (TCD).
Uses ultra-stable, non-depleting Auto-calibration function helps to
These proven, digital sensors not only digital sensing technologies that reduce operational costs
provide highly accurate measurements help extend maintenance intervals
but also offer operational flexibility in
a range of industries and applications.
The non-depleting, high-stability sensing UNRIVALLED PERFORMANCE
technologies help extend maintenance
intervals while intelligent functionality
Uses industry-leading, Manufactured by Servomex –
such as independent auto-calibration
ultra-sensitive and reliable more than 60 years’ experience
helps to deliver operational efficiencies
Paramagnetic, GFx Infrared, SBSW innovating and pioneering gas
over a long product life. Infrared, Zirconia, Aluminum Oxide analysis and thousands of units
and TCD sensing technologies used in the field every year
Using the same reliable measurement
technology as the SERVOPRO 4100 and
MultiExact 5400, the new MultiExact
4100 is backward-compatible with
BENCHMARK COMPLIANCE
existing installations, providing trusted
gas analysis while offering the advanced USP compliant method for assay In compliance with Low Voltage,
digital communications that many of medical oxygen and medical air EMC and applicable EU Directives
modern systems require. European Pharmacopoeia compliant
for medical oxygen and medical air
Plasma
KEY APPLICATIONS
Product purity on air separation plant Monitor trace CO2 on scrubbed air inlet
to air separation process
Process control on air separation plant
Validation of medical O2, N2, air and He
Bottling/filling plant applications
TECHNOLOGIES
UV FLUORESCENCE ALUMINUM OXIDE CHEMILUMINESCENCE
PRODUCT DATA: SERVOPRO MultiExact 4100
MONITORING PERFORMANCE
Gas O2 purity O2 control O2 trace Ar, N2, He CO2 (trace) CO2 (trace)†
Thermal
Technology Paramagnetic Paramagnetic Zirconia Infrared (Gfx) Infrared (Gfx)
Conductivity (TCD)
Accuracy
1% of rdg or 1% of rdg or
(intrinsic ±0.01% O2 ±0.1% O2 ±0.1ppm** ±1% span
<0.1ppm* <0.5ppm*
error)
1% of rdg or 1% of rdg or
Repeatability <0.01% O2 <0.1% O2 <0.1ppm** <0.5% span
<0.1ppm* <0.5ppm*
0-50/0-500ppm, 0.25/0.5/1/2.5/5/
Range 0-50/0-500ppm 0-50/0-500ppm 1/2.5/10 %
0-10/0-50ppm† 10/25/50 /100%
Accuracy
1% of rdg or 1% of rdg or 1% of rdg or
(intrinsic <1% FS <1% FS
<0.5ppm* <0.5ppm* <0.5ppm*
error)
Zero drift/
1ppm 1ppm 1ppm <2% FS <2% FS
week
FLAME IONIZATION
DETECTOR
TECHNOLOGIES
PRODUCT DATA:
SERVOPRO MultiExact 4100
Size:
132.5mm (5.2”) high (265.5mm
(10.5”) high with expansion chassis)
481.6mm (19”) wide
544.2mm (21.4”) deep
Weight:
Main unit: approx 14kg (30.9lb)
Expansion chassis: approx 13.7kg
(30.2lb)
(dependent on number and
type of sensors used)
Operating Temperature:
o o o o
5 C – 40 C (41 F – 104 F)
Certifications:
EN 61010-1: Safety requirements
for electrical equipment for
measurement, control and
laboratory use
E N 61326-1: Electrical equipment
for measurement, control and
laboratory use
EMC requirements
These analyzers are not intended for any form Please note: This document was updated in July 2017. While every effort has been made to ensure
of use on humans and are not medical devices accuracy, no responsibility can be accepted for errors or omissions. Data may change, as well as
as described in the Medical Devices Directive legislation, and you are strongly advised to obtain copies of the most recently issued regulations,
93/42EEC. standards and guidelines. This document is not intended to form the basis of a contract.
Servomex has a policy of constant product improvement and reserves the right to change specifications
without notice. © Servomex Group Limited. 2014. A Spectris company. All rights reserved.
SERVOMEX.COM