Kuster Company
cen om CALIBRATION
=e QUALITY CONFORMANCE
Model: K10 Geothermal Strain Gauge
Serial Number(s): GS5043
Calibration Date: 03/15/2016
Pressure Range: 0 to 5,000 psig
Temperature Range: 20° to 325°C
Max. Pressure Error: +0.024%F.S.
Max. Temperature Error: +0.25%F.S.
Calibration Standards
Pressure:
* DH Instruments PG7302-AMH (Uncertail
ity: [ + 40Pa + (25 ppm + 0.04ppm/Mpa)])
Temperature:
+ Hart Scientific 1502A Tweener (Accuracy: +0.018 °C)
* Hart Scientific 5615-12 PRT (Uncertainty: 0.029 °C)
+ Hart Scientific 9172 Metrology Well (Stability: £0.01 °C / Accuracy: +0.016 °C)
‘Tractability
* Kuster Company local gravity has been determined through the National Geodetic
Survey gravity prediction.
All working standards are traceable to the National Institute of Standards and Technology,
NIST, United States.