Convergys® ISE comfort Electrolyte

Analyzer

SERVICE MANUAL
2014/03/24 Version 2.0

Convergent Technologies GmbH & Co. KG


Ringstrasse 14
35091 Coelbe, Germany
Tel. +49 6421 8869948
Fax. + 49 6421 8869958
www.convergent-technologies.de
email: info@convergent-technologies.de
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

Contents
1. Convergys® ISE comfort brief introduction .....................................................................................4
1.1 front panel ................................................................................................................. 4
1.1.1 LCD display with touch keys ........................................................................... 4
1.1.2 Valves.............................................................................................................. 4
1.1.3 Sample probe.................................................................................................. 5
1.1.4 Sample chamber ............................................................................................. 5
1.1.5 Pump set ......................................................................................................... 5
1.1.6 Cal-1 calibrator................................................................................................ 6
1.1.7 Cal-2 calibrator................................................................................................ 6
1.1.8 Waste bottle .................................................................................................... 6
1.1.9 Front cover ............................................................................................................... 7
1.2 Back panel................................................................................................................. 7
1.2.1 Printer.............................................................................................................. 7
1.2.2 RS232 port ................................................................................................................8
1.2.3 Instrument label ........................................................................................................8
1.2.4 Ground connector.....................................................................................................8
1.2.5 Power supply.............................................................................................................8
1.2.6 Back cover .................................................................................................................8
2. Convergys® ISE comfort installation ...............................................................................................9
2.1 Front .......................................................................................................................... 9
2.1.1 Install electrodes ............................................................................................. 9
2.1.2 Install pump..................................................................................................... 9
2.1.3 Install Cal-1 ..................................................................................................... 9
2.1.4 Install Cal-2 ................................................................................................... 10
2.1.5 Install waste bottle ........................................................................................ 10
2.2 Back......................................................................................................................... 10
2.2.1 Connect to host PC....................................................................................... 10
2.2.2 Connect to ground .........................................................................................11
2.2.3 Connect power supply ...................................................................................11
2.2.4 Prepare printer ...............................................................................................11
3. Convergys® ISE comfort change consumable part.................................................................... 12
3.1.1 Change valve A ............................................................................................. 12
tube (here green) ................................................................................................... 12
3.1.2 Change valve B tube (here blue) .................................................................. 12
3.1.3 Change air valve tube (here yellow) ............................................................. 12
3.2 Change sample probe ............................................................................................. 13
3.2.1 Change O-ring .............................................................................................. 13
3.2.2 Change sample tube..................................................................................... 13
3.3 Change electrodes .................................................................................................. 13
3.3.1 Change electrode inner solution ................................................................... 13
3.3.2 Change REF. electrode inner solution .......................................................... 14
3.4 Change pump tube and connect tube..................................................................... 14

2
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

3.5 Change calibrator .................................................................................................... 14


3.5.1 Change Cal-1 calibrator ................................................................................ 14
3.5.2 Change Cal-2 calibrator ................................................................................ 14
3.6 Change printer paper .............................................................................................. 14
3.7 Change fuse ............................................................................................................ 15
4. Trouble Shooting .............................................................................................................. 16
4.1 General requirements ..................................................................................................... 16
4.2 General procedure to set up electrodes ................................................................. 16
4.3 1-point calibration failed .......................................................................................... 17
4.4 2-point calibration failed ................................................................................................. 18
5.1 Connectors on CPU board ...................................................................................... 19
6. Convergys® ISE comfort PCB .............................................................................................. 20
7. Convergys® ISE comfort AMP .............................................................................................. 28

Revision history

Version Section affected Edited by Reviewed by Date


1.0 First edition ZG MG 05.02.2012
2.0 Address changes JW SK 24.03.2014

3
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

1. Convergys® ISE comfort brief introduction


1.1 front panel

1. LCD display
2. Pinch valves
1
3. Sample probe
4. Sample chamber
5. Pump
4 6. Cal-1 calibrator
7. Cal-2 calibrator
8 6 7
8. Waste bottle
2
3

1.1.1 LCD display with touch keys

LCD display: menu, instruction, result


and other information.

Touch keys: interface

1.1.2 Valves

1. Valve of Cal-1
2. Valve of Cal-2
3. Tube of air Valve( here yellow)
4 4. Tube of Cal-2 valve( here blue)
2 5. Tube of Cal-1 valve (here green)
6. Tube connector
5
3
6
1

4
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

1.1.3 Sample probe

2
1

1. Liquid level sensor


2. Sample tube
6 3. Tube connect to Cal-1
4 4. Tube connect to Cal-2
7 5. Sample probe
5 6. Cal-2 valve
3
7. Cal-1 valve

1.1.4 Sample chamber

1. Shielding cover
5 2. Electrode housing
3. Tighten screw
6 4. Connect tube
5. Electrode label
6. Electrode connector
8 7 3
2 7. Electrodes
4 8. Connect to sample probe

1.1.5 Pump set

5 3

1
1. Pump holder
2. Pump tube
2
3. Connect to waste tube
4. Pump wheel
5. Connect to tighten screw
4

5
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

1.1.6 Cal-1 calibrator

Use for instrument calibration, set the


reference point. Rinse tubing kit.

1.1.7 Cal-2 calibrator

Use for instrument calibration, set


electrode slope.

1.1.8 Waste bottle

Contain waste liquid, such as used


Cal-1, Cal-2 and sample serum.

6
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

1.1.9 Front cover

Front cover

1.2 Back panel

1. Back cover
1
2. RS232port
3. Instrument label
4. Ground connector
3 5. Power supply

2 4
5

1.2.1 Printer

1. Printer

7
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

1.2.2 RS232 port


9-pin RS232 connector (female)
connects to host PC by this connector.

1.2.3 Instrument label


Instrument label, indicate the
information s relate to instrument, such
as Model, S/N, power rating etc.

1.2.4 Ground connector


Connect to ground by this connector.

1.2.5 Power supply

Connect to power net by power cord.


Switch on or off power supply. Fuse is
in built.

1.2.6 Back cover

Back cover

8
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

2. Convergys® ISE comfort installation


2.1 Front
2.1.1 Install electrodes

1. Remove all the air bubble that stays at electrodes


bottom by knock it with finger. Hold electrodes by,
1 2 3
push it to electrodes housing. Make sure the
electrodes are in proper position.
2. You must change the seal cap of REF. electrode
before installing it. There is thin hole in the seal cap
let the air flow pass. Remove all the air bubble in
REF. electrode.
3. Drive to tighten screw by hand.

2.1.2 Install pump

1. Fix left end of pump tube at tube holder 1.


2. Fix other side on tube holder 2.
3. Make sure pump tube is entangled on
1 3
pump wheel.

2.1.3 Install Cal-1

1 1. Remove seal cap of Cal-1, insert tube


2 (here green) into bottle.
2. Drive tighten seal cap and put bottle on.

9
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

2.1.4 Install Cal-2

1
1. Remove seal cap of Cal-2 bottle, insert
2
tube (here blue) into bottle.
2. Drive tighten seal cap and put bottle on.

2.1.5 Install waste bottle

1. Insert connect tube to waste bottle.


2. Put waste bottle on.

2 1

2.2 Back
2.2.1 Connect to host PC

2 1. Put cable on RS232 port.


2. Drive to tighten the 2 screws.
3. Connect other side of cable to host PC.

10
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

2.2.2 Connect to ground

1
1. Switch off power supply and remove power cord
3
before connect to ground.
2. Remove screw at position 1. Entangle ground
wire to ground connector.
2
3. Drive to tighten screws.
4. Connect other end of ground wire to ground.

2.2.3 Connect power supply

1. Insert power cord to power inlet.


3 2. Connect power cord to power net. Switch
on power switch, instrument will start.
1

2.2.4 Prepare printer

1
1. Remove printer cover. Put paper roll
into paper box. Please insert the paper
into the correct position.
2. Make a sharp end of paper, and insert it
into printer.

11
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

3. Convergys® ISE comfort change consumable part


3.1.1 Change valve A
tube (here green)

1. Cut 60mm from tube including in standard packing.


1
2. Remove tube (here green) from valve A.
3. Connect tube to tube connector.

3.1.2 Change valve B tube (here blue)

1 1. Cut 60mm from tube including in standard packing.


2. Remove tube (here blue) from valve B.
3. Connect tube to tube connector.

3.1.3 Change air valve tube (here yellow)

1. Cut 60mm from tube including in standard packing.


2. Remove tube (here yellow) from air valve.
3. Connect tube to tube connector.

12
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

3.2 Change sample probe


3.2.1 Change O-ring

1. Head up sample probe, remove probe hosing by


2
screw it at counter-clockwise.
2. Remove used O-ring, and change new one.
3. Screw probe housing.

3.2.2 Change sample tube

1
2

1. Head up probe, remove probe housing by screw it


at anticlockwise.
2. Remove sample used tube and change new one.
3. Screw probe housing.

3.3 Change electrodes


3.3.1 Change electrode inner solution

1. Remove metal screw on the up of electrodes,


remove the used inner solution.
2. Pipette about 0.5ml inner solution from bottle by
syringe, and pipette into electrode, makes sure the
liquid level is about 3/4 of electrode height.
3. Screw the metal screw, clean the electrode surface.
Remove
13
air bubble.
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

3.3.2 Change REF. electrode inner solution

2
1. Remove seal cap, remove used inner solution.
2. Pour inner solution to electrode from bottle
directly. Make sure the liquid level is about 1/2 of
electrode height.
3. Screw seal cap, remove air bubble, clean
electrode surface.
3.4 Change pump tube and connect tube

1. Remove tube from pump, remove connect tube.


2. Change new tube, and connect to connect tube.
3. Fix pump tube to holder.
3.5 Change calibrator
3.5.1 Change Cal-1 calibrator
Please refer to 2.1.3

3.5.2 Change Cal-2 calibrator


Please refer to 2.1.4

3.6 Change printer paper


Please refer to2.2.4

14
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

3.7 Change fuse

2
1
1. Remove fuse holder by small ‘-’ type screw driver,
remove used fuse, change new one.
2. Put on fuse holder.

15
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

4. Trouble Shooting
The Convergys® ISE comfort is a highly sensitive instrument. To fix problems that might occur
after installation, please follow the guide below step by step. Send the resulting report to your
technical support.

4.1 General requirements

Checked

• The ISE comfort has to be properly grounded


• Working at 10°C~40°C. Avoid temperature shifts after calibration. Avoid
placing the instrument in direct sunlight or in front of a source of heat or
drafts.

• Avoid electrical interference.

• Power for this instrument should be AC 220 V ±22 V and 50 Hz ±1 Hz.


Ensure stable supply of voltage (use a UPS with integrated sinus voltage
stabilizer).

• Ensure that there are no air bubbles inside of all of the electrodes.
• Ensure that the aspirated liquid is covering the tube left and right of the
electrode chamber and the tube channel between the electrodes is filled
properly.

• Check Cal-1 and cal-2 solution, usage of another batch

4.2 General procedure to set up electrodes

Checked
Use fresh plasma to activate the electrodes till they are stable. It might need
approximately 2 hours and the process has to be executed in the “sensor” mode.

The equipment may report inconsistent results in case the electrodes are
contaminated. Please use anti-protein solution to clean regular contamination. In
case of heavier contamination the operator may consider using a thin cotton
thread and moving it carefully through the sample channel.

The electrodes may face attrition in case filling solution is not adequate in the
inner chamber of electrodes. Please remove the left out filling solution from the
electrodes and fill up with the fresh Filling Solution.

REMARKS:

16
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

4.3 1-point calibration failed

Caused by electrode unstable Checked


Check if the electrode is new. If yes, it will take 90 minutes to be stable, it is
strongly recommend using serum to activate the electrodes.
Check if one or more electrodes are unstable
There is air bubble on the membrane of the electrode. Click the electrode with
fingers until the air bubble goes out.
Something is wrong with the O ring: take away and put back the electrode again,
or change O ring.
Electrode connection tube dirty: rinse with anti-protein solution.
Electrode quality fade: change electrode.

Cal out of range in all the electrodes Checked


There is air bubble in the reference electrode: click reference electrode with
fingers several times until the air bubble goes out.
Short of reference filling solution: add or/and change filling solution.
The orifice on the top of the reference electrode is clogged: open it with needle.
The pipeline of reference electrode is dirty: rinse it with anti protein cleaning
solution
The AgCl on the reference electrode internal element drops. Change reference
internal element.
The membrane of reference electrode broken: change reference electrode.

One electrode voltage is out of range Checked


Low or short of electrode filling solution: change the filling solution.
The pipeline of reference electrode is dirty: rinse it anti protein cleaning solution
The AgCl on the electrode internal element drops. Change internal element.
Electrode quality fade: change electrode.
Na electrode is unstable, check Na and Reference electrode

Caused by Sample fault or there is air bubble in the electrode Checked


Check if the pipe is clogged or broken.
Check the O-ring in the sample system.
Check the pipe to see if it is clogged or if it is broken step by step
Check if there is enough solution in the calibration bottle.
Check if the solution is polluted.

REMARKS:

17
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

4.4 2-point calibration failed

Caused by electrode unstable Checked


Check if the electrode is new. If yes, it will take 90 minutes to be stable. Use the
electrode installation Guide to install the electrodes.
If all electrode or only Na electrode is unstable, it is most might caused by
reference electrode.
There is air in the conjunction of reference electrode, shake the electrode
several times until the air bubble goes out.
The orifice on the top of the reference electrode is clogged: open it with needle.
The AgCl on the reference electrode internal element drops. Change reference
internal element.
The membrane of reference electrode broken: change reference electrode.
Check if one or more electrodes are unstable
There is air bubble on the membrane of the electrode. Click the electrode until
the air bubble goes out.
Something is wrong with the O ring:take away and put back the electrode again,
or exchange O-ring.
Electrode pipeline is dirty: calibration with cleaning solution.
Electrode quality fade: change electrode.

Caused by slope out of range Checked


If all electrodes show slope are out of range, please check if your calibration
solution 1 and 2 are in the right place or if they are polluted.
If only one electrode shows the slop out of range. Please check if the electrode’s
membrane is polluted, if yes rinse with anti protein cleaning solution several
times.
Please check if the quality of the electrode were fade, if yes, change the
electrode.

Possible cause for aspiration failure of solution Checked


Please check if the pipe be clogged or broken, if the inlet probe is clogged or if
the O-ring is broken.
Check the pipe to see if it is clogged or if it is broken step-by-step refereeing to
the pipe design picture.
Check if there is enough solution in the calibration bottle.
Check if the solution is polluted.

REMARKS:

Should you still have trouble with the function of the instrument, please contact

18
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

Convergent Technologies: info@convergent-technologies.de. And sent the obtained


results together with this report to your technical support.

Print outs: Cal-1 calibration


Cal-2 calibration
Print out of the calibration Data
Prepare a list with the k, d, L, Δ L and Δ H parameters

Send a copy of the printouts and remarks to your technical support.

5. Convergys® ISE comfort CPU board


5.1 Connectors on CPU board

1. connect electrodes housing


12 1
2. connect to LCD
2 3. connect to printer
4. connect to valves
5. connect to sample probe sensor
6
9 6. connect to liquid level sensor
3 5
7. connect to AC\DC adaptor.
4
8. memory battery
8
9. ROM (software programmed in)
10. connect to RS-232
11. connect pump motor
7
10 11 12. connect to key pad

19
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

6. Convergys® ISE comfort PCB


LCD :

Printer:

20
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

Sensor and waste:

Stepping motor control circuit:

21
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

ISE comfort Pump motor circuit:

Buzzer and power LED circuit:

Stepping motor 1 circuit:

22
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

ISE comfor setting and sample position:

CPU Connector bars:

23
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

Power in:

RS232:

Other:

24
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

ISE comfort CPU:

25
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

26
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

27
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

7. Convergys® ISE comfort AMP


The ISE comfort pre-Amplifier contains independent circuits for the different electrodes.

28
Convergys® ISE comfort E l e c t r o l y t e A n a l y z e r Service Manual

Manufactured by:

Convergent Technologies GmbH & Co. KG


Ringstrasse 14
D-35091 Coelbe, Germany
Tel: +49 6421 8869948
Fax: +49 6421 8869958
Homepage: www.convergent-technologies.de
e-mail: info@convergent-technologies.de

29

You might also like