Content
Delivered Parts Oxygen Sensor for Softdrinks ................................................................................ 2
Delivered Parts Oxygen Sensor for Beer ........................................................................................ 2
Electrical Wiring .............................................................................................................................. 2
Wiring Power Supply ................................................................................................................... 3
Wiring Oxygen Sensor................................................................................................................. 3
Wiring Analog Output to Analog Input of the mPDS 5 .................................................................. 3
Mount the M400 Evaluation Unit ..................................................................................................... 4
Mechanical Installation .................................................................................................................... 6
Install the Oxygen Sensor Bypass: .............................................................................................. 6
Install the Oxygen Sensor Inline: ................................................................................................. 7
Start-up: .......................................................................................................................................... 8
Replacing the OptoCap ................................................................................................................... 9
Two-Point Sensor Calibration ....................................................................................................... 10
DLI: Dynamic Lifetime Indicator .................................................................................................... 11
Electrical Wiring
Hint
If the Option Oxygen Sensor is ordered together with a Cobrix 5 the electrical wiring is performed
by Anton Paar.
All connections to the M400 evaluation unit are made on the rear panel.
1 Power supply
2 Connection for oxygen sensor cable
3 Analog output
1 brown Sensor +
2 black Sensor -
3 not used
TB 4 5 not used
6 not used
7 blue RS 485 -
8 white RS 485 +
9 not used
Mount the holder for the evaluation unit as described in the picture below.
Clampsfor DIN rail mounted with
2 screws M5x8 and 2 snap rings.
Mount the DIN rail on the thread bolts inside of the Cobrix 5 stainless steel housing.
Mount the evaluation unit on the DIN rail. Connect the power supply cable to the Cobrix 5
distributor and the analog cable to the appropriate mPDS 5 analog input.
Mechanical Installation
Mount the oxygen sensor in the bypass adapter und tighten the nut at the sensor with a
size 19 flat wrench.
Mount the bypass adapter with sensor at the outlet of the Cobrix 5.
Remove the red protection cap and connect the cable to the oxygen sensor.
Mount the oxygen sensor in the Varivent N Process Connection and tighten the nut with a
size 19 flat wrench.
Remove the red protection cap and connect the cable to the oxygen sensor.
Start-up:
(PATH: Menu/Quick Setup)
Hint
If the option oxygen sensor is ordered together with a Cobrix 5 the start-up settings are performed
by Anton Paar.
Channel Selection =
ISM
Parameter = = O2 Opt
Set the parameters for oxygen unit and analog output as described in the picture below.
Press [Enter] to switch in the next window.
Set the scaling for the analog output as described in the picture below.
Press [Enter] to switch in the next window.
Press [Enter] until the evaluation unit asks to save the changes. Select “Yes & Exit” und press
[Enter]
The M400 evaluation unit saves the new settings.
Hint
If the cap sleeve is detached, take care of the inner part of the sensor shaft. Damage and soiling of
the inner parts and optical fibre may influence the signal or destroy the sensor. Clean with a lint-
free cloth.
Hint
Make sure that this maintenance step is carried out in a clean place.
1. Unscrew the cap sleeve from the sensor shaft and carefully pull it off the sensor.
2. Pull off the OptoCap.
3. Place the new OptoCap on top of the inner body of the sensor shaft.
4. Carefully slip the cap sleeve over the fitted OptoCap and screw it down. The cap sleeve
must be clean and dry.
5. After each exchange of the OptoCap, the DLI must be reset manually.
6. After each exchange of the OptoCap, the sensor must be recalibrated with a two-point
sensor calibration.
Notice
From time to time check the lifetime of the OptoCap and change once if the lifetime is zero.