Professional Documents
Culture Documents
Department of Food Science and Technology, The University of New South Wales, PO
Box 1, Kensington, NSW 2033, Australia
ABSTRACT
NOTATION
o Percentage error
Subscripts
Copper
F Modified Fitch method
71
Journal of Food Engineering 0260-8774/91/$03.50 - 0 1991 Elsevier Science
Publishers Ltd, England. Printed in Great Britain
72 Md. S. Rahmun
0 Initial
p Probe method
s Surface of the copper rod
INTRODUCTION
Construction
::. ;.
..
.
The apparatus which was a further modification of the Zuritz et al.
(1989) apparatus, is shown in Fig. 2. It consists of a constant-tempera-
c‘. . .
:. I,.
. ..
. i::
.‘: :.
~~~
..
\ i+
,..
,’
. .
.
.
Fig. 2. (1) Vacuum flask. (2) Copper disc assembly. (3) Copper rod assembly. c,
Copper disc; d, polystyrene disc; e, copper rod; i, insulation; r, rubber cork; s, sample
compartment. (All dimensions in mm.)
Theory
The quasi-steady conduction heat transfer through the sample yields the
following simplified equation (Fitch, 1935; Mohsenin, 1980):
Modified Fitch method for thermal conductivity 75
AW-JM
c
c
c
c (1)
1 dt
The thermal conductivity (k) can be calculated from the slope of the plot
of ln[( To- T,)/( T - T,)] versus t.
The precision and error depend on the applicability of eqn (1). There
will be minimum error and higher precision (i.e. quasi-steady state) when
eqn (1) is satisfied. Hence, the error arises due to not satisfying quasi-
steady conduction and depends on A, k,, T,, 1, M,, C, and t. The slope
(S) is the response of the apparatus and can be written as:
Hence, the response will depend on A, I and M,. The values of A, 1 and
M, should be fixed based on practical considerations and limitations. For
example, the response will increase with decreasing thickness (I), which
is good up to a point when I cannot be measured precisely. Again,
conductivity of fresh food is slightly dependent on temperature above
freezing, but below freezing conductivity changes significantly with
temperature. In this case, a lower I will give a higher response, which will
reduce precision. Also, the highest response is limited by the heat
transfer from the copper disc to the polystyrene disc. Zuritz et al. ( 1989)
found that the heat transfer rate between the copper plug and its insula-
tion became significant (i.e. > 10% error) when the temperature change
in the copper plug exceeded 15°C.
The instrument was calibrated, and a correction factor defined as the
ratio of the thermal conductivities measured by the probe and by the
Fitch method:
f = kplkf
The correction factor ( f ) was correlated by multiple regression as:
f=a +bl+ck, (5)
76 Md. S. Rahman
diffusion of the ethylene glycol into the frozen sample during the experi-
ments. The thermal conductivity value was recorded at the mean sample
temperature during the experiment. The duration was always 180 s and
the temperature was recorded at 1O-s intervals.
3.0
A
A apple
A0 0 potato
2.0 -
m A
A
0 AA
4
X 430
A
A
dB
: l.O- 4&g Q 6%
0”
G A 0
h 0
I A 0
0.0 ! I I I I I I
o 2 4 6 8 10 12
Sample thickness(mm)
Fig. 3. Effect of sample thickness on response (slope) for M, = 8.132 x lo-’ kg and
D= 10X 10v3m.
78 Md. S. Rahman
3.0 I
0 apple
A potato
m 2.0 -
0
X
G
; l.O-
v,
0 2 4 6 8 10
Sample thicknessfm m)
Fig. 4. Effect of sample thickness on response (slope) for M, = 8.648 x lo-’ kg and
D=15X10-3m.
Modified Fitch method for thermal conductivity 79
7 0.0
A chicken meat
9.0- 0 ice
8.0-
m
0
7.0 -
2
m
6.0 -
W
a
0 \
u-l 5.0 -
4.0 - \
3.0- 3
Sample thick ness (m m)
Fig. 5. Effect of sample thickness on response (slope) for M, = 8.648 X 10e3 kg and
D= 15 x 10e3 m.
the sample) due to the high conductivity value. For example, ice has a
thermal conductivity value four tunes higher than water. Also, the
thermal conductivity of frozen material changes significantly with
temperature. Another problem might be the heat losses or gains between
the sample and surrounding air due to the high temperature gradient
(35-50°C). The apparatus was calibrated by measuring the thermal
conductivities of ice and frozen chicken meat and the correction factors
are correlated with the sample thickness (range: 4-7 mm) as:
f=3*77-6951+6_91k, (8)
TABLE 1
Percentage Error of Modified Fitch Apparatus at Temperatures Above Freezing
(M,=8~648~10-~kg;D=15x10-~m;Z=4x10-~m)
in using this apparatus below freezing. Hence, eqn (8) is valid only when
the average sample temperature remains around - 25°C during the
experiment. The maximum percentage error for frozen material was 9%
of the conductivity at the average sample temperature during the experi-
ment (Table 2). The rise in the sample temperature could be reduced by
increasing the copper disc mass (M,) or reducing the contact area (A ).
Modified Fitch method for thermal conductivity 81
TABLE 2
F’ercentage Error of Modified Fitch Apparatus at Temperatures Below Freezing
(M,=8~648~10-~kg;D=15x10-~m;Z=7x10-~m)
CONCLUSION
REFERENCES