Professional Documents
Culture Documents
EJX530A
EJX530A
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN Nov. 20, 2019-00
<<Contents>> <<Index>> 3
□ NORMAL OPERATING CONDITION
(Optional features or approval codes may
affect limits.) 100(14.5) Atmospheric
pressure
Ambient Temperature Limits Working
−40 to 85°C (−40 to 185°F) pressure
kPa abs
−30 to 80°C (−22 to 176°F) with LCD display (psia)
Process Temperature Limits
Applicable range
−40 to 120°C (−40 to 248°F)
10(1.4)
Ambient Humidity Limits
0 to 100% RH
Maximum Over Pressure
2.7(0.38)
Pressure
Capsule EJX510A EJX530A
1(0.14)
A and B 4 MPa abs (580 psia) 4 MPa (580 psig) -40 0 40 80 120
C 20 MPa abs (2900 psia) 20 MPa (2900 psig) (32) (104) (176) (248)
(-40)
D 75 MPa abs (10800 psia) 75 MPa (10800 psig)
Process temperature °C (°F)
Working Pressure Limits (Silicone oil) F02E.ai
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN Apr. 01, 2013-00
<<Contents>> <<Index>> 4
Communication Requirements “◊” □ PHYSICAL SPECIFICATIONS
(Approval codes may affect electrical requirements.)
Wetted Parts Materials
BRAIN
Diaphragm, process connector
Communication distance Refer to “MODEL AND SUFFIX CODES.”
Up to 2 km (1.25 miles) when using CEV
polyethylene-insulated PVC-sheathed cables. Non-wetted Parts Materials
Communication distance varies depending on type of Housing
cable used. • Low copper cast aluminum alloy
Load capacitance • Low copper cast aluminum alloy with corrosion
0.22 µF or less resistance properties (copper content ≤ 0.03%,
iron content ≤ 0.15%) (optional)
Load inductance • ASTM CF-8M Stainless steel (optional)
3.3 mH or less
Coating of housing
Input impedance of communicating device [for aluminum housing]
10 kΩ or more at 2.4 kHz. Polyester resin powder coating
EMC Conformity Standards Mint-green paint (Munsell 5.6BG 3.3/2.9 or its
EN 61326-1 Class A, Table2 equivalent)
EN 61326-2-3 [for option code /P or /X2]
EN 61326-2-5 (for fieldbus) Epoxy and polyurethane resin solvent coating
European Pressure Equipment Directive Degrees of protection
2014/68/EU IP66/IP67, Type 4X
Sound Engineering Practice (for all capsules) Pipe
With option code /PE3 (for D capsule) Polypropylene
Cover O-rings
Category III, Module H, Type of Equipment: Pressure Buna-N, fluoro-rubber (optional)
Accessory-Vessel, Type of Fluid: Liquid and Gas,
Group of Fluid: 1 and 2 Name plate and tag
316 SST
EU RoHS Directive
EN IEC 63000 Fill fluid
Silicone, Fluorinated oil (optional)
Safety Requirement Standards
EN 61010-1, C22.2 No.61010-1 Weight
• Installation category: I Capsule A, B and C: 1.2 kg (2.6 lb)*
(Anticipated transient overvoltage 330 V) Capsule D: 1.4 kg (3.1 lb)*
• Pollution degree: 2 *: Without integral indicator and mounting bracket.
• Indoor/Outdoor use Add 1.5 kg (3.3 lb) for Amplifier housing code 2.
Connections
Refer to “MODEL AND SUFFIX CODES.”
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN July 1, 2021-00
<<Contents>> <<Index>> 5
MODEL AND SUFFIX CODES
For EJXC40A Remote Digital Sensor, please refer to GS 01C25W05-01EN.
The ‘#’marks indicate the construction materials conform to NACE material recommendations per MR0175/ISO 15156.
Please refer to the latest standards for details. Selected materials also conform to NACE MR0103.
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN Jan. 31, 2023-00
<<Contents>> <<Index>> 6
[Process Connections Code for Diaphragm Seal System]
Following table shows the code dedicated for EJXC50A Diaphragm Seal System. The code cannot be specified
without a diaphragm seal system. Please also refer to the GS 01C25W01-01EN for EJXC50A.
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN July 1, 2021-00
<<Contents>> <<Index>> 7
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN Dec. 29, 2023-00
<<Contents>> <<Index>> 8
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN Feb. 1, 2022-00
<<Contents>> <<Index>> 9
OPTIONAL SPECIFICATIONS
Item Description Code
Painting Color change Amplifier cover only*14 P
Amplifier cover and terminal cover, Munsell 7.5 R4/14 PR
Coating change Anti-corrosion coating*1 X2
316 SST exterior parts 316 SST zero-adjustment screw and setscrews*16 HC
Fluoro-rubber O-ring All O-rings of amplifier housing. Lower limit of ambient temperature: –15°C (5°F) HE
Lightning protector Transmitter power supply voltage: 10.5 to 32 V DC (10.5 to 30 V DC for intrinsically safe type,
9 to 32 V DC for Fieldbus communication type.)
A
Allowable current: Max. 6000 A ( 1×40 µs ), Repeating 1000 A ( 1×40 µs ) 100 times
Applicable Standards: IEC 61000-4-4, IEC 61000-4-5
Status output*2 Transistor output (sink type)
AL
Rating: 30 V DC, 120 mA DC (max) Low level: 0 to 2 V DC
Oil-prohibited use*21 Degrease cleansing treatment K1
Degrease cleansing treatment and fluorinated oilfilled capsule.
K2
Operating temperature −20 to 80°C (−4 to 176°F)
Degrease cleansing treatment K41
Degrease cleansing treatment and fluorinated oilfilled With certificates
capsule. K42
Operating temperature −20 to 80°C (−4 to 176°F)
Oil-prohibited use with Degrease cleansing and dehydrating treatment K5
dehydrating treatment*21
Degrease cleansing and dehydrating treatment with fluorinated oilfilled capsule.
K6
Operating temperature −20 to 80°C (−4 to 176°F)
Degrease cleansing and dehydrating treatment K45
Degrease cleansing and dehydrating treatment with With certificates
fluorinated oilfilled capsule. K46
Operating temperature −20 to 80°C (−4 to 176°F)
Capsule fill fluid*21 Flourinated oil filled in capsule
K3
Operating temperature −20 to 80°C (−4 to 176°F)
Calibration units*3 P calibration (psi unit) D1
bar calibration (bar unit) (See Table for Span and Range Limits.) D3
M calibration (kgf/cm2 unit) D4
Output limits and failure Failure alarm down-scale : Output status at CPU failure and hardware error is −5%, 3.2mA DC
C1
operation*4 or less.
Failure alarm down-scale: Output status at CPU
NAMUR NE43 Compliant C2
failure and hardware error is −5%, 3.2 mA DC or less.
Output signal limits:
3.8 mA to 20.5 mA Failure alarm up-scale: Output status at CPU
C3
failure and hardware error is 110%, 21.6 mA or more.
Gold-plated diaphragm*13*21 Surface of isolating diaphragms are gold plated, effective for hydrogen permeation. A1
Wired tag plate 316 SST tag plate wired onto transmitter N4
Data configuration at factory*8 Data configuration for HART
Software damping, Descriptor, Message CA
communication type
Data configuration for BRAIN
Software damping CB
communication type
Data configuration for HART Software damping, Descriptor, Message,
CJ
communication type External zero adjustment prohibition setting
Data configuration for BRAIN Software damping, External zero adjustment
CK
communication type prohibition setting
Advanced diagnostics*17 Multi-sensing process monitoring
• Impulse line blockage detection *18 DG6
• Heat trace monitoring
European Pressure PED 2014/68/EU
Equipment Directive*15*21 Category: III, Module: H, Type of Equipment: Pressure Accessory-Vessel, PE3
Type of Fluid: Liquid and Gas, Group of Fluid: 1 and 2
Material certificate*6*21 Process Connector M15
Process connector, Diaphragm, Capsule body MA2
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN Jan. 31, 2023-00
<<Contents>> <<Index>> 10
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN Jan. 31, 2023-00
<<Contents>> <<Index>> 11
DIMENSIONS
Unit: mm (approx.inch)
Model EJX510A and EJX530A
● With process connections code 7
95(3.74) 188(7.40)
60(2.36)*2 Integral indicator Zero
(optional) adjustment 110(4.33)
Extenal indicator 12 39
Conduit connection Conduit connection*5
(0.47) (1.54)
(optional)
ø78(3.07)
(2.76)
54
ø70
126(4.96)
110(4.33)
(2.13)
159(6.26)
(6.50)
165
(0.24)
Ground terminal
Electrical connection
for code 5, 9, A, and D. Open to
atmosphere*1
2-inch pipe
Mounting bracket 12*3 (O.D. 60.5 mm) 129(5.08)
(optional) (0.47)
(6.77)
172
ø6
5(0.20)
(0.24)
F04E.ai
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN Jan. 31, 2023-00
<<Contents>> <<Index>> 12
Connection Description
Analog output EJX electrical terminal
Distributor
+ + 24V DC
SUPPLY
– 250Ω
*1
CHECK A *1: Either A or + –
– – 250Ω
If shield cable is not
*1
used, communication ALARM A
is not possible. *1: Either A or +
Magnetic
External power supply valve
30V DC, 120mA max
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN May 20, 2016-00
<<Contents>> <<Index>> 13
< Ordering Information > “◊” < Factory Setting > “◊”
Specify the following when ordering
Tag number As specified in order
For output signal code –J, refer to GS 01C25T01-
01EN. Software
‘2.00 s’ or as specified in order
damping *1
1. Model, suffix codes, and option codes
Calibration range
2. Calibration range and units lower range value
As specified in order
1) Calibration range can be specified with range
Calibration range
value specifications up to 5 digits (excluding any upper range As specified in order
decimal point) for low or high range limits within value
the range of -32000 to 32000. When reverse
[EJX530A]
range is designated, specify Lower Range Selected from mmH2O, mmH2O(68°F),
Value (LRV) as greater than Upper Range Value mmAq*2, mmWG*2, mmHg, Pa, hPa*2,
(URV). kPa, MPa, mbar, bar, gf/cm2, kgf/cm2,
2) Specify only one unit from the table, ‘Factory inH2O, inH2O(68°F), inHg, ftH2O,
Settings’ when shipped.’ ftH2O(68°F) or psi. (Only one unit can be
3. Display scale and units (for transmitters equipped Calibration range specified)
with integral indicator only) units [EJX510A]
Specify either 0 to 100 % or engineering unit scale Torr, Pa abs, hPa abs*2, kPa abs,
and ‘Range and Unit’ for engineering units scale: MPa abs, mbar abs, bar abs, kgf/cm2 abs,
Scale range can be specified with range limit mmH2O abs, mmH2O abs(68°F),
specifications up to 5 digits (excluding any decimal mmHg abs, inH2O abs, inH2O abs(68°F),
point) for low or high range limits within the range inHg abs, ftH2O abs, ftH2O abs(68°F),
of -32000 to 32000. The unit display consists of psia, atm.
6-digit, therefore, if the specified unit is longer than Display setting
Designated value specified in order.
7 characters excluding ‘/’, the first 6 characters will (%, or user scaled value.)
be displayed on the unit display.
*1: To specify these items at factroy, /CA, /CB, /CJ or
4. TAG NO (if required) /CK option is required.
Specified characters (up to 16 characters for *2: Not available for HART protocol type.
BRAIN, 22 characters for HART, or 16 characters
for /N4 tag) are engraved on the stainless steel tag
< Material Cross Reference >
plate fixed on the housing.
5. SOFTWARE TAG (for HART only. if required) ASTM JIS
Specified characters (up to 32 characters) are set
grade 316 SUS316
as “Tag” (the first 8 characters) and “Long tag”*1
(32 characters) in the amplifier memory. Use grade 316L SUS316L
alphanumeric capital letters. grade 304 SUS304
When the “SOFTWARE TAG” is not specified,
specified “TAG NO” is set as “Tag” (the first 8 <Information on EU WEEE Directive>
characters) and “Long tag”*1 (22 characters) in the EU WEEE (Waste Electrical and Electronic Equipment)
amplifier memory. Directive is only valid in the EU.
*1: applicable only when HART 7 is selected.
6. Other factory configurations (if required) Specifying This instrument is intended to be sold and used only
option code /CA, /CB, /CJ or /CK will allow further as a part of equipment which is excluded from WEEE
configuration at factory. Following are configurable Directive, such as large-scale stationary industrial
items and setting range. tools, a large-scale fixed installation and so on, and,
[/CA, /CJ : For HART communication type] therefore, subjected to the exclusion from the scope
1) Descriptor (up to 16 characters) of the WEEE Directive. The instrument should be
2) Message (up to 30 characters) disposed of in accordance with local and national
3) Software damping in second (0.00 to 100.00) legislation/regulations.
[/CB, /CK : For BRAIN communication type]
1) Software damping in second (0.00 to 100.00)
All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01EN Jan. 31, 2023-00
Subject to change without notice.