Serviceanleitung MELAtronic 23 - EN
Serviceanleitung MELAtronic 23 - EN
®
MELAtronic 23
DANGER
Before opening the device, switch off at the power switch and remove the plug!
CONTENT Page
CONTENT PAGE .......................................................................................................................................................... 2
APPENDIX........................................................................................................................................................................ 7
3.1 EQUIPMENT VIEWS ............................................................................................................................................... 8
3.2 ARTICLE/ SPARE PART NUMBERS ......................................................................................................................... 8
3.3 PIPING PLAN ......................................................................................................................................................... 8
3.4 WIRING DIAGRAMM ............................................................................................................................................. 8
3.5 PROGRAM FLOWCHART ........................................................................................................................................ 8
3.6 OPERATION OF THE ELECTRONIC CONTROLS "MELATRONIC" ............................................................................... 8
3.7 PARAMETER LIST ................................................................................................................................................. 8
3.8 OPERATING INSTRUCTIONS .................................................................................................................................. 8
Customer-side power
supply o.k.? Ensure power supply
permanent short-circuit
Failure immediately upon power switch
Supply voltage at
connections N and L1 activation of the mains wiring harness
of the electronics? switch?
electronic control (20)
Short-circuit at
Failure with program start or
activation of the drying heating (42) or at the
function? overheating protection (44)
Seite 3
1.2 No water feed after pressing the switch "water feed"
If there is no intake of water in the steam sterilizer chamber after activating the switch "water feed" (19),
check the following possible causes:
Solenoid valve "water feed" (38) does not open
- No power supply (water feed switch, wiring harness, connector socket)
- Solenoid valve spool defective (check magnetic force on spool with screwdriver, check spool for
electrical passage)
- Valve mechanically blocked (plunger sticks/jams, open valve housing, clean valve)
Tank filter (41) clogged (unscrew filter with jaw or ring wrench size 13 and clean and possibly replace
mesh filter insert, clean drillings in the filter housing)
Filter water storage tank (27) is clogged (unscrew conduit (28) at the filter, withdraw and clean mesh
filter insert downwards, possibly replace, clean drillings in the filter housing)
Notice: Cleaning the filter without opening the device and dismantling the piping can be attempted by
activating the switch "water feed" during a program sequence pressure at the manometer > 1 bar to
produce a pressure discharge on the water feed line. After completed pressure discharge, the program
must immediately be cancelled manually.
Seite 5
Device error:
Heating (42) defective (device remains cold, check heating for passage)
Solenoid valve "water feed" (38) does not close (water is pressed back over the water feed filter into the
reserve chamber, check with empty reserve chamber)
Vent nozzle (30) washed out (heating permanently on, standard temperature is not achieved)
Spring-operated safety valve (49) untight, opens too early Overheating protection system (44) defective
(switches too early or permanently activated). Check the position of the jacket heating unit. This unit
should be pushed as far in as it can go at the bottom rear fitting of the sterilization chamber.
Power switch (26) defective ( despite actuation, LED lights up, no output voltage )
Electronic control (20) defective
Attention!
The pivot pins must be completely
inserted in the latch bolt or in the
hinge yoke of the pressure locking
bar (left side plug. Check the
retaining rings above and below
for correct seat.
Seite 7
Appendix
3.1 Equipment views ______________________________________________________
see drawing (Z.-No. 23E_0_05 page 1 and 2)