Professional Documents
Culture Documents
GPT-9000 Series
USER MANUAL
GW INSTEK PART NO. 82PT-99040EA1
Table of Contents
SAFETY INSTRUCTIONS .................................................. 5
OPERATION .................................................................. 27
Menu Tree ............................................ 29
Test Lead Connection ........................... 33
Manual Testing .................................... 35
Common Utility Settings ...................... 71
Automatic Tests ................................... 78
3
GPT-9000 Series User Manual
4
SAFETY INSTRUCTIONS
SAFETY INSTRUCTIONS
This chapter contains important safety
instructions that you must follow during
operation and storage. Read the following before
any operation to ensure your safety and to keep
the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the
instrument.
5
GPT-9000 Series User Manual
Safety Guidelines
6
SAFETY INSTRUCTIONS
7
GPT-9000 Series User Manual
8
GETTING STARTED
GETTING STARTED
This chapter describes the safety tester in a
nutshell, including its main features and front /
rear panel introduction. After going through the
overview, please read the safety considerations in
the Set Up chapter.
AC / DC Withstanding Voltage / PASS FAIL READY TEST
GPT-9803 Insulation Resistance Tester
START STOP
RETURN
REMOTE
POWER
Appearance .................................................................... 14
GPT-9801/9802/9803 Front Panel .........................................................14
GPT-9000 Rear Panel................................................................................18
Set Up ........................................................................... 20
Line Voltage Connection and Power Up ..............................................20
Installing the Optional GPIB Card .........................................................22
Workplace Precautions ............................................................................23
Operating Precautions .............................................................................24
Basic Safety Checks ..................................................................................26
9
GPT-9000 Series User Manual
Series lineup
The GPT-9000 Series Safety Testers are AC/DC withstanding
voltage, insulation resistance and ground bond safety testers. The
GPT-9801 is an AC withstanding voltage tester, the GPT-9802 is an
AC/DC withstanding voltage tester and the GPT-9803 is an
AC/DC withstanding voltage and insulation resistance tester. The
GPT-9804 includes all the functions of the other models as well as
ground bond testing. All models can operate at up to 5kVAC for
AC withstanding voltage testing. The GPT-9802, GPT-9803 and
GPT-9804 can also operate at up to 6kVDC for DC withstanding
voltage testing.
The GPT-9000 Series can store up to 100 manual tests, as well as
run up to 16 manual tests sequentially as an automatic test,
allowing the safety testers to accommodate any number of safety
standards, including IEC, EN, UL, CSA, GB, JIS and others.
10
GETTING STARTED
Main Features
11
GPT-9000 Series User Manual
Accessories
12
GETTING STARTED
Package Contents
Check the contents before using the GPT-9000.
13
GPT-9000 Series User Manual
Appearance
GPT-9803
AC / DC Withstanding Voltage /
Insulation Resistance Tester
PASS FAIL READY TEST
TEST indicator
CAUTION 5.0 kVAC MAX.
HIGH VOLTAGE 6.0 kVDC MAX.
ESC PAGE
HIGH VOLTAGE
indicator
START STOP
HIGH VOLTAGE
POWER
REMOTE
RETURN
output terminal
POWER MANU/AUTO EDIT/SAVE UTILITY
button RETURN
terminal
STOP
SENSE H RETURN SENSE L &
REMOTE
SENSE H Rx SENSE L
RETURN Terminal
SOURCE H SOURCE L
MANU/AUTO EDIT/SAVE UTILITY
SOURCE H
SOURCE L
PASS FAIL
Pass/Fail The PASS and FAIL indicators
indicators light up upon a PASS or FAIL test
result at the end of a manual test
or automatic test.
14
GETTING STARTED
READY
READY indicator The READY indicator is lit when
the tester is ready to begin testing.
The STOP button is used to put
the tester into READY status.
TEST
TEST indicator The TEST indicator is lit when a
test is on. The START button is
used to put the tester into TEST
status.
WARNING
USE EXTREME CAUTION.
Do not touch the HIGH VOLTAGE terminal
during testing.
15
GPT-9000 Series User Manual
RETURN
RETURN terminal The RETURN terminal is the
GPT-9801/9802 return terminal for all tests.
/9803
For GB testing, the RETURN
terminal is also the SENSE L
terminal.
EDIT/SAVE
EDIT/SAVE key Used to start editing
MANU/AUTO tests as well as
save settings and parameters.
MANU/AUTO
MANU/AUTO The MANU/AUTO key is used to
key select manual tests (MANU) or
automatic tests (AUTO).
16
GETTING STARTED
STOP
STOP button The STOP button is used to
stop/cancel tests. The STOP
button will also put the safety
tester in the READY status to
begin testing.
START
START button The START button is used to start
tests.
17
GPT-9000 Series User Manual
WARNING
TO AVOID ELECTRIC SHOCK THE POWER CORD
SIGNAL I / O RS 232 PROTECTIVE GROUNDING CONDUCTOR MUST BE
CONNECTED TO GROUND .
FOR CONTINUED FIRE PROTECTION REPLACE
ONLY WITH SPECIFIED TYPE AND RATED FUSE .
NO OPERATOR SERVICEABLE COMPONENTS INSIDE .
DO NOT REMOVE COVERS . REFER SERVICING TO
QUALIFIED PERSONNEL .
GND
GPIB
AC
POWER MAX.
500VA
SIGNAL I / O
SIGNAL I/O port The SIGNAL I/O port is used to
monitor the tester status (PASS,
FAIL, TEST) and input (START/
STOP signals). It is also used with
the Interlock key.
RS232
RS232 interface Used for remote control and
port firmware updates.
230
18
GETTING STARTED
120
230
100
100V/120V T5A 250V
220V/230V T2.5A 250V
Optional GPIB GPIB
port Optional GPIB interface for remote
control.
19
GPT-9000 Series User Manual
Set Up
Steps 1. Check the line voltage and the fuse Page 139
in the fuse holder.
230
120
arrow on the fuse holder. 100
100
20
GETTING STARTED
S Y S T EM S E L F TEST
S y s t em Ch e c k i n g . . .
Ha r dwa r e Che c k i ng . . .
F i r mw a r e C h e c k i n g . . .
VIEW status
MANU = * * * - 0 0 2 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
21
GPT-9000 Series User Manual
22
GETTING STARTED
Workplace Precautions
23
GPT-9000 Series User Manual
Operating Precautions
24
GETTING STARTED
25
GPT-9000 Series User Manual
1. Ensure all test leads are not broken and are free
from defects such as cracks or splitting.
26
OPERATION
OPERATION
Menu Tree...................................................................... 29
Menu Tree Overview ...............................................................................30
27
GPT-9000 Series User Manual
28
OPERATION
Menu Tree
This section describes the overall structure of the operation statuses
and modes for the GPT-9000 safety testers. The testers have two
main testing modes (MANU, AUTO) and 5 main operation statuses
(VIEW, EDIT, READY, TEST and STOP).
PASS/FAIL result
READY
status
MANU UTILITY
menu (MANU
Press ED I T Press
EDIT/ Save the MANU
UTILITY status SAVE test or AUTO test
mode only)*
* Press EDIT/SAVE to save settings, or ESC to cancel and return to the previous screen.
† Press the STOP key twice for a FAIL result.
29
GPT-9000 Series User Manual
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
EDIT status
MANU = * * * - 0 0 2 MANU _ NAME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC = OF F
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
READY status
MANU = * * * - 0 0 2 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA READY
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
30
OPERATION
0 100 kV 00 33 mA TEST
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
STOP status
MANU = * * * - 0 0 2 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV 00 33 mA S TOP
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
31
GPT-9000 Series User Manual
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB ADD
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
C O MMO N U T I L I T Y
S t a r t C t r l : F RON T P ANE L
Do u b l e A c t i o n : OF F
K e y L o c k : OF F I N T ER L OCK : OF F
L CD BUZ Z I NTER CTRL
MANU = * * * - 0 0 2 MANU UT I L I T Y
ARC MO D E : O F F
P A S S HO L D : OF F
F A I L HO L D : OF F
MA X HO L D : OF F
G R O U N D MO D E : O N
32
OPERATION
DUT
Return terminal
HIGH VOLTAGE
Terminal
RETURN
Terminal
33
GPT-9000 Series User Manual
GB Connection
GB Connection GPT-9000
Source H
Sense H
DUT
Source L
Sense L
SENSE H Rx SENSE L
34
OPERATION
Manual Testing
This section describes how to create, edit and run a single ACW,
DCW, IR or GB safety test. Each manual test can be stored/recalled
to/from one of 100 memory locations. Each stored manual test can
be used as a test step when creating an AUTO test (page 78).
35
GPT-9000 Series User Manual
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
MANU # 001~100
(MANU# 000 is a special mode)
MANU number
MANU = * * * - 0 0 2 MANU _ NAME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC = OF F
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
36
OPERATION
V I EW
MANU = * * * - 0 0 2 MANU _ NAME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC = OF F
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
37
GPT-9000 Series User Manual
A CW D CW I R GB
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
test function
38
OPERATION
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
cursor
39
GPT-9000 Series User Manual
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
cursor
40
OPERATION
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
cursor
H I S E T = 0 1 . 0 0mA
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
cursor
41
GPT-9000 Series User Manual
cursor
MANU = * * * - 0 0 2 MANU _ NAME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z L 0 S E T = 0 1 . 0 0mA ARC = OF F
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
42
OPERATION
Test V
Start V
time
RAMP TEST TIME
Initial time Discharge time
(100ms) (200ms)
43
GPT-9000 Series User Manual
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
cursor
ACW 000.5s~999.9s
DCW 000.5s~999.9s
IR 001.0s~999.9s
GB 000.5s~999.9s
Special Manual When in special MANU test mode (page 66) the
Mode Timer can be turned off when using the DCW
or ACW test function.
44
OPERATION
Test V
Start V
time
RAMP TEST TIME
Initial time Discharge time
(100ms) (200ms)
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
cursor
ACW 000.1s~999.9s
DCW 000.1s~999.9s
IR 000.1s~999.9s
45
GPT-9000 Series User Manual
cursor
MANU = * * * - 0 0 2 MANU _ NAME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z L 0 S E T = 0 1 . 0 0mA ARC = OF F
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB H I / L O T I MER
46
OPERATION
MANU = * * * - 0 0 2 MANU UT I L I T Y
ARC MO D E : O F F
P A S S HO L D : OF F
F A I L HO L D : OF F
MA X HO L D : OF F
G R O U N D MO D E : O N
47
GPT-9000 Series User Manual
48
OPERATION
MANU = * * * - 0 0 2 MANU UT I L I T Y
ARC MO D E : O F F
P A S S HO L D : OF F
F A I L HO L D : OF F
MA X HO L D : OF F
G R O U N D MO D E : O N
49
GPT-9000 Series User Manual
MANU = * * * - 0 0 2 MANU UT I L I T Y
ARC MO D E : O F F
P A S S HO L D : OF F
F A I L HO L D : OF F
MA X HO L D : OF F
G R O U N D MO D E : O N
50
OPERATION
MANU = * * * - 0 0 2 MANU UT I L I T Y
ARC MO D E : O F F
P A S S HO L D : OF F
F A I L HO L D : OF F
MA X HO L D : OF F
G R O U N D MO D E : O N
51
GPT-9000 Series User Manual
52
OPERATION
Source H
stray
Sense H
resistance,
Sense L DUT
capacitance
Source L
53
GPT-9000 Series User Manual
Source H
stray
Sense H
resistance,
Sense L DUT
capacitance
Source L
MANU = * * * - 0 0 2 MANU UT I L I T Y
ARC MO D E : O F F
P A S S HO L D : OF F
F A I L HO L D : OF F
MA X HO L D : OF F
G R O U N D MO D E : O N
54
OPERATION
0 050 kV mA 0E D050
I T kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB ACHWI / L OD C
TWI MER I R GB H I / L O T I MER
GROUND GROUND
MODE = OFF MODE = ON
55
GPT-9000 Series User Manual
Background After all test parameters have been set, the test
can be saved. After a test is saved it can be used
when creating an AUTO test.
ED I T
MANU = * * * - 0 0 2 MANU _ NAME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC = OF F
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
56
OPERATION
VIEW status
MANU = * * * - 0 0 2 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
57
GPT-9000 Series User Manual
STOP
2. Press the STOP button to put the
tester into the READY status.
READY status
MANU = * * * - 0 0 2 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA READY
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
START
4. Press the START button when the
tester is in the READY status. The
manual test starts automatically
and the tester goes into the TEST
status.
TEST status
MANU = * * * - 0 0 2 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV 00 00 mA TEST
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
58
OPERATION
0 100 kV 00 37 mA TEST
RAMP =000 . 1S T I ME R = 0 0 3 . 2 S
A CW D CW I R GB
0 100 kV 00 37 mA TEST
RAMP =000 . 1S T I ME R = 0 0 3 . 2 S
A CW D CW I R GB
0 100 kV 00 56 mA TEST
RAMP =000 . 1S T I ME R = 0 0 3 . 2 S
A CW D CW I R GB
03 01 A 200 7 mΩ TEST
GBV = 0 . 6 2 2 V T I ME R = 0 0 3 . 2 S
A CW D CW I R GB
59
GPT-9000 Series User Manual
STOP
Stop the Test 1. To stop the test at any time when it
is running, press the STOP button.
The test will stop immediately.
When the STOP button is pressed,
a judgment is not made on the test.
STOP status
MANU = * * * - 0 0 2 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV 00 00 mA S TOP
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
STOP
2. To put the tester back into READY
status, press the STOP button
again.
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
60
OPERATION
0 100 kV 00 37 mA PASS
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
STOP
Pressing the STOP button will
return the tester to the READY
status.
61
GPT-9000 Series User Manual
START
Pressing the START button will
restart the test.
PASS Timing The timing diagrams below show the ACW, DCW,
Diagrams IR and GB timing for the START status, TEST
status and PASS judgment.
ACW PASS
Timing START
TEST
PASS
Output V
time
RAMP TEST TIME
Initial time Discharge time
(100ms) (200ms)
DCW PASS
Timing START
TEST
PASS
Output V
time
RAMP TEST TIME
Initial time Discharge time
(100ms) (200ms)
62
OPERATION
IR PASS Timing
START
TEST
PASS
Output V
time
RAMP TEST TIME
Initial time Discharge time
(100ms) (200ms)
GB PASS Timing
START
TEST
PASS
Output I
time
0.5s
Initial time Total Test Time
(100ms)
0 100 kV 01 37 mA FA I L
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
STOP
2. The FAIL judgment will be held on
the display until the STOP button
is pressed. Pressing the STOP
button will return the tester to the
READY status.
63
GPT-9000 Series User Manual
READY status
MANU = * * * - 0 0 2 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA READY
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
FAIL Timing The timing diagrams below show the ACW, DCW,
Diagrams IR and GB timing for the START status, TEST
status and FAIL judgment.
Output V
time
RAMP TEST TIME
Initial time Discharge time
(100ms) (200ms)
Output V
time
RAMP TEST TIME
Initial time Discharge time
(100ms) (200ms)
64
OPERATION
IR FAIL Timing
START
TEST
FAIL
Output V
time
RAMP TEST TIME
Initial time Discharge time
(100ms) (200ms)
GB FAIL Timing
START
TEST
FAIL
Output I
time
0.5s
Initial time Total Test Time
(100ms)
65
GPT-9000 Series User Manual
MANU = * * * - 0 0 2 MA NU _ N AME R E F # = 0 0 0 . 0 mΩ
FREQ= 6 0 H z H I S E T = 3 0 0 . 0 mΩ
03 01 A mΩ V I EW
GBV = V T I ME R = 0 0 3 . 2 S
A CW D CW I R GB
STOP
3. Press the STOP button to put the
tester into the READY status.
66
OPERATION
START
5. Press the START button to perform
the zero check. The tester will go
into the ZERO status.
Measured Resistance
ZERO status
MANU = * * * - 0 0 2 MA NU _ N AME R E F # = 0 0 0 . 0 mΩ
FREQ= 6 0 H z H I S E T = 3 0 0 . 0 mΩ
03 01 A 200 7 mΩ Z E RO
GBV = 0 . 6 2 2 V T I ME R = 0 0 3 . 2 S
A CW D CW I R GB ZER0
Reference value
MANU = * * * - 0 0 2 MA NU _ N AME R E F # = 2 0 0 . 7 mΩ
FREQ= 6 0 H z H I S E T = 3 0 0 . 0 mΩ
03 01 A mΩ V I EW
GBV = V T I ME R = 0 0 3 . 2 S
A CW D CW I R GB
67
GPT-9000 Series User Manual
I < SET FA I L
GBV = 5 . 6 2 2 V T I ME R = 0 0 3 . 2 S
A CW D CW I R GB ZER0
R=0 Stop the test and perform the zero check again.
R = 0 error message
FAIL status
MANU = * * * - 0 0 2 MA NU _ N AME R E F # = 0 0 0 . 0 mΩ
FREQ= 6 0 H z H I S E T = 3 0 0 . 0 mΩ
R = 0 FA I L
GBV = 5 . 6 2 2 V T I ME R = 0 0 3 . 2 S
A CW D CW I R GB ZER0
68
OPERATION
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
69
GPT-9000 Series User Manual
70
OPERATION
LCD Settings
VIEW status
MANU = * * * - 0 0 0 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
C O MMO N U T I L I T Y
L CD Con t r a s t : 5
L CD B r i g h t n e s s : BR I GH T
71
GPT-9000 Series User Manual
Buzzer Settings
VIEW status
MANU = * * * - 0 0 0 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
72
OPERATION
C O MMO N U T I L I T Y
Pa s s S o u n d : ON T I ME : 0 0 0 . 5 S
Fa i l S o u n d : OF F
Interface Settings
73
GPT-9000 Series User Manual
VIEW status
MANU = * * * - 0 0 0 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
C O MMO N U T I L I T Y
I n t e r f a c e : RS 2 3 2
Baud : 115200
74
OPERATION
Control Settings
75
GPT-9000 Series User Manual
VIEW status
MANU = * * * - 0 0 0 MA NU _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
C O MMO N U T I L I T Y
S t a r t C t r l : F RON T P ANE L
Do u b l e A c t i o n : OF F
K e y L o c k : OF F I N T ER L OCK : OF F
L CD BUZ Z I NTER CTRL
76
OPERATION
settings.
READY
I N T ER L OCK OP EN
0 100 kV mA
RAMP = 0 0 0 . 1 S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
77
GPT-9000 Series User Manual
Automatic Tests
This section describes how to create, edit and run automatic tests.
Automatic tests allow you to link together up to 16 different
MANU tests and run them sequentially. Each stored MANU test is
used as a test step when creating an AUTO test.
Choose/Recall an Automatic Test→ from page 78
Edit Automatic Test Settings → from page 80
Adding a Step to the Automatic Test → from page 81
Creating an AUTO Test File Name → from page 82
Saving and Exiting EDIT Status → from page 83
Automatic Test Page View → from page 84
Running an Automatic Test → from page 87
Automatic Test Results → from page 91
78
OPERATION
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
V I EW
AUTO # 001~100
AUTO number
AUTO= 0 0 1 - 1 0 0 A U T O _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
79
GPT-9000 Series User Manual
V I EW
AUTO= 0 0 1 - 0 0 1 AU TO _ NAME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC = OF F
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB ADD
80
OPERATION
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB ADD
0 100 kV mA FUL L
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB ADD
81
GPT-9000 Series User Manual
0 100 kV mA ED I T
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB ADD
82
OPERATION
ED I T
AUTO= 0 0 1 - 0 0 1 AU TO _ NAME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC = OF F
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB ADD
83
GPT-9000 Series User Manual
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
84
OPERATION
85
GPT-9000 Series User Manual
86
OPERATION
87
GPT-9000 Series User Manual
VIEW status
AUTO= 0 0 1 - 1 0 0 A U T O _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
STOP
2. Press the STOP button to put the
tester into the READY status.
READY status
AUTO= 0 0 1 - 1 0 0 A U T O _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA READY
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
START
4. Press the START button when the
tester is in the READY status. The
AUTO test starts automatically and
the display changes to TEST status.
88
OPERATION
0 100 kV 00 37 mA TEST
RAMP =000 . 1S T I ME R = 0 0 3 . 2 S
A CW D CW I R GB
HOLD status
AUTO= 0 0 1 - 0 0 1 AU TO _ NAME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV 00 37 mA HO L D
RAMP =000 . 1S T I ME R = 0 0 3 . 2 S
A CW D CW I R GB
START
3. To continue to the next test after
HOLD is displayed on-screen,
press the START button.
89
GPT-9000 Series User Manual
STOP
4. To stop the test when HOLD is
displayed on-screen, press the
STOP button.
STOP
Stop a Running 1. To stop the AUTO test at any time
Test when it is running, press the STOP
button. The AUTO test will stop
immediately. When the STOP
button is pressed, a judgment is
not made on the current test and
any remaining tests are aborted.
AUTO= 0 0 1 - * * * AU TO _ NAME
#01 : FA I L # 0 2 : PASS # 0 3 : S TOP # 0 4 : - - - -
#05 : - - - - #06 : - - - - #07 : - - - - # 0 8 : - - - -
#09 : #10 : #11 : # 1 2 :
#13 : #14 : #15 : # 1 6 :
STOP
2. To put the tester back into READY
status, press the STOP button
again.
90
OPERATION
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
91
GPT-9000 Series User Manual
AUTO= 0 0 1 - * * * AU TO _ NAME
# 0 1 : PASS # 0 2 : PASS # 0 3 : PASS # 0 4 : PASS
# 0 5 : PASS # 0 6 : PASS # 0 7 : PASS # 0 8 : PASS
#09 : #10 : #11 : # 1 2 :
#13 : #14 : #15 : # 1 6 :
AUTO= 0 0 1 - * * * AU TO _ NAME
# 0 1 : PASS # 0 2 : PASS # 0 3 : PASS # 0 4 : PASS
# 0 5 : PASS #06 : FA I L #07 : FA I L # 0 8 : PASS
#09 : #10 : #11 : # 1 2 :
#13 : #14 : #15 : # 1 6 :
0 100 kV 00 37 mA PASS
RAMP =000 . 0S T I ME R = 0 0 3 . 2 S
A CW D CW I R GB
PASS/FAIL result
92
OPERATION
STOP
2. To put the tester back into READY
status, press the STOP button
(twice for a fail result).
READY status
AUTO= 0 0 1 - 1 0 0 A U T O _ N AME RE F # = 0 0 . 0 0mA
FREQ= 6 0 H z H I S E T = 0 1 . 0 0mA ARC= OF F
0 100 kV mA READY
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
0 100 kV mA V I EW
RAMP =000 . 1S T I ME R = 0 0 1 . 0 S
A CW D CW I R GB
93
GPT-9000 Series User Manual
PASS Timing
Diagram START
TEST
PASS
Output V
time
step 1 ~ step 16
FAIL Timing
Diagram START
TEST
FAIL
Output V
time
step 1 ~ step 16
94
EXTERNAL CONTROL
EXTERNAL CONTROL
The External Control chapter covers the
REMOTE terminal and the SIGNAL I/O port.
95
GPT-9000 Series User Manual
96
EXTERNAL CONTROL
97
GPT-9000 Series User Manual
Pin Assignment 6 7 8 9
1 234 5
Pin name Pin Description
INTERLOCK1 1 When INTERLOCK is ON, a test is only allowed
INTERLOCK2 2 to start when both INTERLOCK pins are shorted.
INPUT_COM 3 Common input line
INPUT_START 4 Start signal input
INPUT_STOP 5 Stop signal input
OUTPUT_TEST 6 Indicates that a test is in progress
OUTPUT_FAIL 7 Indicates that a test has failed
OUTPUT_PASS 8 Indicates that a test has passed
OUTPUT_COM 9 Common output line
Interlock
connection PIN 1 INTERLOCK1
PIN 2 INTERLOCK2
Input Connection
PIN 3 INPUT_COM
PIN 4 INPUT_START
PIN 5 INPUT_STOP
Output
Connection PIN 6 OUTPUT_TEST
PIN 7 OUTPUT_FAIL
PIN 8 OUTPUT_PASS
PIN 9 OUTPUT_COM
98
EXTERNAL CONTROL
99
GPT-9000 Series User Manual
100
REMOTE CONTROL
REMOTE CONTROL
This chapter describes basic configuration of
IEEE488.2 based remote control. The remote
interface supports USB, RS232 and GPIB.
101
GPT-9000 Series User Manual
Interface Configuration
Data bits 8
Stop bit 1
102
REMOTE CONTROL
RS232
Panel operation 1. Connect the Null modem cable to
the rear panel RS232 port.
GPIB
Panel operation 1. Connect the GPIB cable to the
rear panel GPIB port.
103
GPT-9000 Series User Manual
*idn?
104
REMOTE CONTROL
0 100 kV mA RM T
RAMP =000 . 1S T I ME R = 0 0 3 . 2 S
A CW D CW I R GB
STOP
Steps 1. When RMT is on the display, press
the STOP button. The panel goes to
the READY status.
105
GPT-9000 Series User Manual
Command Syntax
IEEE488.2 Partial compatibility
Compatible
Standard SCPI, 1999 Partial compatibility
MANU MANU:ACW:VOLTage
ACW
Example *IDN?
106
REMOTE CONTROL
Example MANU:ACW:VOLTage?
107
GPT-9000 Series User Manual
Message
CR, LF Carriage Return + Line feed code
Terminator
108
REMOTE CONTROL
Command List
109
GPT-9000 Series User Manual
110
REMOTE CONTROL
System Commands
SYSTem:LCD:CONTrast .......................................................................111
SYSTem:LCD:BRIGhtness .....................................................................111
SYSTem:BUZZer:PSOUND ...................................................................112
SYSTem:BUZZer:FSOUND ...................................................................112
SYSTem:BUZZer:PTIMe ........................................................................112
SYSTem:BUZZer:FTIMe ........................................................................113
SYSTem:ERRor .......................................................................................113
Set
SYSTem:LCD:CONTrast Query
111
GPT-9000 Series User Manual
Set
SYSTem:BUZZer:PSOUND Query
112
REMOTE CONTROL
Set
SYSTem:BUZZer:FTIMe Query
SYSTem:ERRor Query
SYSTem:GPIB:VERSion Query
113
GPT-9000 Series User Manual
Function Commands
Set
FUNCtion:TEST Query
114
REMOTE CONTROL
MEASure<x> Query
115
GPT-9000 Series User Manual
Manual Commands
MANU:STEP ........................................................................................... 117
MANU:NAME........................................................................................ 117
MANU:RTIMe ........................................................................................ 117
MANU:EDIT:MODE.............................................................................. 118
MANU:ACW:VOLTage ........................................................................ 118
MANU:ACW:CHISet............................................................................. 119
MANU:ACW:CLOSet ............................................................................ 119
MANU:ACW:TTIMe ............................................................................. 120
MANU:ACW:FREQuency .................................................................... 121
MANU:ACW:REF .................................................................................. 121
MANU:ACW:ARCCurrent ................................................................... 122
MANU:DCW:VOLTage ........................................................................ 122
MANU:DCW:CHISet ............................................................................. 123
MANU:DCW:CLOSet ............................................................................ 123
MANU:DCW:TTIMe ............................................................................. 124
MANU:DCW:REF .................................................................................. 124
MANU:DCW:ARCCurrent ................................................................... 125
MANU:IR:VOLTage .............................................................................. 125
MANU:IR:RHISet ................................................................................... 126
MANU:IR:RLOSet .................................................................................. 126
MANU:IR:TTIMe ................................................................................... 127
MANU:IR:REF ........................................................................................ 127
MANU:GB:CURRent ............................................................................. 127
MANU:GB:RHISet ................................................................................. 128
MANU:GB:RLOSet ................................................................................ 128
MANU:GB:TTIMe .................................................................................. 129
MANU:GB:FREQuency ......................................................................... 129
MANU:GB:REF ...................................................................................... 130
MANU:GB:ZEROCHECK ..................................................................... 130
MANU:UTILity:ARCMode .................................................................. 131
MANU:UTILity:PASShold.................................................................... 131
MANU:UTILity:FAILhold .................................................................... 131
MANU:UTILity:MAXHold ................................................................... 132
MANU:UTILity:GROUNDMODE ....................................................... 132
MANU<x>:EDIT:SHOW ....................................................................... 132
116
REMOTE CONTROL
Set
MANU:STEP Query
117
GPT-9000 Series User Manual
118
REMOTE CONTROL
Set
MANU:ACW:CHISet Query
For example:
HI SET value: 12.34
LO SET value1: 0.005 error
LO SET value2: 0.053 no error
In the example above LO SET value1 will produce
an error as all digits are outside the range of HI
SET. LO SET value2 will not produce an error, but
will return 0.05, not 0.053.
119
GPT-9000 Series User Manual
Syntax MANU:ACW:CLOSet<NR2>
Query Syntax MANU:ACW:CLOSet?
Parameter/ <NR2> 0.000 ~ 041.9
Return parameter
Example MANU:ACW:CLOS 20.0
Sets the ACW LO SET current to 20 mA.
Set
MANU:ACW:TTIMe Query
120
REMOTE CONTROL
Set
MANU:ACW:FREQuency Query
121
GPT-9000 Series User Manual
Set
MANU:ACW:ARCCurrent Query
122
REMOTE CONTROL
Set
MANU:DCW:CHISet Query
123
GPT-9000 Series User Manual
Syntax MANU:DCW:CLOSet<NR2>
Query Syntax MANU:DCW:CLOSet?
Parameter/ <NR2> 0.000 ~ 010.9
Return parameter
Example MANU:DCW:CLOS 2.00
Sets the DCW LO SET current to 2mA.
Set
MANU:DCW:TTIMe Query
124
REMOTE CONTROL
125
GPT-9000 Series User Manual
Example MANU:IR:VOLT 1
Sets the IR voltage to 1 kV.
Set
MANU:IR:RHISet Query
126
REMOTE CONTROL
Set
MANU:IR:TTIMe Query
127
GPT-9000 Series User Manual
128
REMOTE CONTROL
Set
MANU:GB:TTIMe Query
129
GPT-9000 Series User Manual
Set
MANU:GB:REF Query
130
REMOTE CONTROL
Set
MANU:UTILity:ARCMode Query
131
GPT-9000 Series User Manual
MANU<x>:EDIT:SHOW Query
132
REMOTE CONTROL
Auto Commands
AUTO:STEP ............................................................................................133
AUTO<x>:PAGE:SHOW.......................................................................134
AUTO:NAME .........................................................................................134
TESTok:RETurn ......................................................................................135
Set
AUTO:STEP Query
133
GPT-9000 Series User Manual
AUTO<x>:PAGE:SHOW Query
134
REMOTE CONTROL
Set
TESTok:RETurn Query
Common Commands
*CLS ..........................................................................................................135
*IDN .........................................................................................................135
*CLS Set
*IDN Query
135
GPT-9000 Series User Manual
Error Messages
136
FAQ
FAQ
• The tester will not turn on.
• The panel keys are not working.
• When I press the START button the tester will not start testing?
• The accuracy does not match the specification.
Ensure the power cord is connected. Ensure the line input is set to
the correct line voltage. Check to make sure the fuse is not blown.
See page 139.
When I press the START button the tester will not start
testing?
The tester must first be in the READY status before a test can be
started. Ensure the tester displays READY before pressing the
START button, page 57 (manual test), 87(automatic test).
If “Double Action” is enabled, the START button must be pressed
0.5 seconds after the STOP button is pressed, otherwise the tester
will not start testing.
137
GPT-9000 Series User Manual
138
APPENDIX
APPENDIX
Fuse Replacement
off.
230
120
100
100
100
139
GPT-9000 Series User Manual
220
230
120
220 100
230
120
100
Error Messages
140
APPENDIX
141
GPT-9000 Series User Manual
GPT-9000 Specifications
The specifications apply when the GPT-9000 is powered on for at
least 30 minutes at 15˚C~35˚C.
Specifications
Environment
Range Temperature Humidity
Warranty 15˚C ~ 35˚C ≤70% (No
condensation)
Operation 0˚C ~ 40˚C ≤70% (No
condensation)
Storage -10˚C ~ 70˚C ≤85% (No
condensation)
Installation Location Indoors at an amplitude of up to 2000m.
AC Withstanding Voltage
Output Voltage Range 0.100kV~ 5.000kV
Output Voltage Resolution 2V
Output Voltage Accuracy (1% of setting +5V) with no load
Maximum Rated Load(Table1) 200 VA (5kV/40mA)
Maximum Rated Current 40mA
0.001mA ~ 10mA(0.1kV≤V≤0.5kV)
0.001mA ~ 40mA(0.5kV<V≤5kV)
Output Voltage Waveform Sine wave
Frequency 50 Hz / 60 Hz
Voltage Regulation 1% +5V
[Maximum rated load → no load]
Voltmeter Accuracy (1% of reading+ 5V)
Current Measurement Range 0.001mA~040.0mA
Current Best Resolution 1uA
0.001mA(0.001mA~0.999mA)
0.01mA(01.00mA~09.99mA)
0.1mA(010.0~040.0mA)
Current Measurement (1.5% of reading +30uA)
Accuracy
Window Comparator Method Yes
ARC DETECT Yes
Rise-time Control Function Yes
RAMP (Ramp Time) 0.1~999.9S
Continued…
142
APPENDIX
DC Withstanding Voltage
Output Voltage Range 0.100kV~ 6.000kV
Output Voltage Resolution 2V
Output Voltage Accuracy (1% of setting +5V) with no load
Maximum Rated Load(Table1) 50W (5kV/10mA)
Maximum Rated Current 10mA
0.001mA ~ 2mA (0.1kV≤V≤0.5kV)
0.001mA ~ 10mA (0.5kV<V≤6kV)
Voltmeter Accuracy (1% of reading+ 5V)
Voltage Regulation 1% +5V
[Maximum rated load → no load]
Current Measurement Range 0.001mA~010.0mA
Current Best Resolution 1uA
0.001mA(0.001mA~0.999mA)
0.01mA(01.00mA~09.99mA)
0.1mA(010.0mA)
Current Measurement (1.5% of reading +30uA)
Accuracy
Window Comparator Method Yes
ARC DETECT Yes
Rise-time Control Function Yes
RAMP (Ramp Time) 0.1~999.9S
TIMER (Test Time) OFF*, 0.5S~999.9S
GND RETURN/GUARD
* The timer can only be turned off under special MANU mode (MANU=***-000)
143
GPT-9000 Series User Manual
Interface
REMOTE (Remote terminal) Yes
SIGNAL IO Yes
RS232 Yes
USB (Device) Yes
GPIB Yes (OPTION)
General
DISPLAY 240 x 64 dot matrix LED back light LCD
MEMORY AUTO/MANU mode 100 memory blocks total
POWER SOURCE AC100V/120V/220V/230V 10%
50Hz/60Hz
Continued…
144
APPENDIX
145
GPT-9000 Series User Manual
GPT-9000 Dimensions
395.9
AC / DC Withstanding Voltage / PASS FAIL READY TEST
GPT-9803 Insulation Resistance Tester
START STOP
RETURN
REMOTE
POWER
322.0
452.0
329.7
146
APPENDIX
Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No. 7-1, Jhongsing Rd, Tucheng Dist., New Taipei City 236, Taiwan
147
GPT-9000 Series User Manual
148
INDEX
149