Professional Documents
Culture Documents
ML032960458
ML032960458
Approved by:
(6n ,J Vt LUc
C. V. McVay, Manager
Analytical and Process
Part I
1.0 PURPOSE .
2.0 APPLICATION
> 2.1 This method is designed for the determination of the density of
aqueous and organic liquids. Exceptions are liquids that attack
the glass oscillator. Hydrofluoric acid is particularly
destructive. The volume -of the oscillator is 0.6 mL; therefore,
the.minimum sample volume is l.mL. In routine~use, the precision
of this method is~approximately i-0.001 density units.
3.0 DISCUSSION
3.3 The major sources of error in the analysis are (a) air bubbles in
the oscillator; (b) incompletely filled oscillator; and (c) air
flowing through:the oscillator when obtaining the air reading for
..calibration. -
GEIJO37O0:E10-118 Page 1 of 7
~DR~
PROJ PDC I
AQK-DEN-1801, Rev. I
Effectdve Date: 06/28/89
4.0 REFERENCES
Part II
5.0 EQUIPMENT
6.4 0.5 percent hydrofluoric acid. (To be used only as cleaning agent
for very short periods of time.)
8.0 RECORDS
> 8.1 All measurement data and sample identification shall be recorded
on the work sheet (Attachment B). The final result shall be
recorded on the analytical request sheet per PRD 5.0.
> 9.2 Calibration of the density meter is done by using air and ultra-
pure water. As both air and water have specific density values,
quality control is assured by calibrating the density meter by
using the known values of each. Constants A and B are recorded on
QC charts each time instrument is calibrated.
> 9.3 Air and.water densities-shall be checked and recorded on each day
the instrument is used. When density values are out of limits for
air and ultrapure water, calibration is necessary and the
constants are recorded.
> 9.4 A quality control sample which has a density within 0.5 g/cc of
the density of samples to be analyzed shall be checked and
recorded on each day the instrument is used or at a rate of one in
twenty samples in each set whichever is more frequent.
Recalibration is required if the QC sample is out of tolerance.
10.0 PROCEDURE
10.1 Calibration
10.2 Measurements
> 10.2.4 With a syringe or eye dropper, slowly inject the sample
into the glass oscillator until it is completely filled.
Use light to be certain no air bubbles are present and
oscillator is completely filled. Turn light off so
instrument is able to reach constant temperature of
20.0 *0.30C.
GE"0370:ENG-118 Page 3 of 7
AQ(-DEK-1801, Rev. 1
Effective Date: 06/28/89
11.0 CALCULATIONS
12.0 ATTACHMENTS
ATTACHMENT A
The apparatus constants, A and B have to be determined and entered into the
constant memories. Both constants are temperature dependent. They have to be
redetermined when the measuring temperature is changed. At no change of the
measuring temperature it is enough to check the calibration value in time
intervals of several weeks.
For the calculation of the constant values, the 4-position switch (display
.selector) is set to IT'; then the period of oscillation IT" is shown on the
display. Now, approximately 0.7 mL distilled water is injected through the
lower opening of the oscillator cell using a plastic tipped hypodermic
syringe. The proper filling of the sample tube may be observed through the
window after the light is switched on., The cell is correctly filled, when the
liquid has passed into the upper (larger) diameter of the sample tube.
Overfilling will have no effect on the proper result.
The syringe may be left in the filling position. The upper opening may be
closed with a teflon plug, for highly volatile substances. The sample must be
; homogeneous and free of even the smallest bubbles.
The light is now switched off, and temperature equilibrium is reached when the
number at the display remains the same within one digit at the fourth place.
This TH 0 value is the period of oscillation of a water-filled oscillator.
At this very first measurement, the sample should remain in the oscillator
cell for about 15 minutes and the T-value should be observed. Now remove the
teflon plug and withdraw the sample into the syringe. rinse the sample tube
with alcohol, acetone, or a similar solvent. Then with the air pump switched
on, connect the air hose to the upper inlet of the sample tube and continue
the flow of air until the displayed value repeats itself several times.
Switch off the pump and compare Tair with its original value. If it is not
identical with the original reading, the oscillator should be rinsed out again
and thoroughly dried to ensure complete removal of all solvent or any other
contamination.
I J-~
Now with the density values for air and water, and the corresponding T-values,
the apparatus constants A and B are calculated as follows:
I.
2 2
air air
These values and their corresponding measuring temperature are marked down on
^ the 'inside of the lid (in penc-il)' which covers the constants storage switches.
The constants A and B'are entered into the memory by setting the corresponding
numerical switches on top of the instrument and checked by setting the display
selector switch accordingly. After setting the display selector switch to "p"
the instrument is ready'to work and the density p of a sample is shown on the
display and may be read out from the digital data output.
1,
I~.I
AQ-DEN-1801, Rev. 1
ACK-DEN-1801, Rev. 1
Effective. Date: 06/28/89
ATTACHMENT B Page _ of _
SPECIAL INSTRUCTIONS
I
SAMPLE ID I
,I
I
WATER DENSITY I I
I I
AIR DENSITY I I
I
I
I
I
I
I
AVERAGE SAMPLE I I
DENSITY I I
(g/cc at 200C) I I
ANALYST DATE
APPROVED DATE
GEW0370:ENG-118 Page 7 of 7