You are on page 1of 8

A Noise-Sensor Light Alarm Reduces Noise

in the Newborn Intensive Care Unit


Ying-Ju Chang, Ph.D., R.N.,1 Ya-Jung Pan, B.S.N., R.N.,2 Yuh-Jyh Lin, M.D.,3
Yan-Zen Chang, B.E.,4 and Chyi-Her Lin, M.D.3

ABSTRACT

This one-group pre-post test design was to evaluate sound distribution and
sudden peak noise frequencies (SPNs) and the associated events after using a noise-sensor

Downloaded by: UPSTATE Medical University. Copyrighted material.


light alarm in a tertiary neonatal intensive care unit (NICU). The alarm is activated as the
sound level reaches  65 A-weighted decibel (dBA). The environmental sound level was
monitored continuously for a period of 1 week before and 1 month after using the alarm.
The mean sound level in the incubator of patients receiving ventilator support before and
after using the device were 58.0  0.6 and 56.4  0.7 dBA (t ¼ 8.619; p < 0.001), whereas
those at the radiant heated bed were 58.0  2.4 and 58.1  2.0 dBA (t ¼ 0.715; p ¼ 0.476).
The percentage of observation time of sound levels < 58 dBA increased by 28% in the
incubator and 4% at the radiant heated bed (p < 0.001). Episodes of SPN decreased from
630 to 185 times/d in the incubator and from 2069 to 748 times/d at the radiant heated
bed after using the device. The noise-sensor light alarm effectively reduces sound level
and episodes of SPN in the NICU. This may alleviate stress of noise for newborns with
critical illness.

KEYWORDS: Noise, alarm, newborn intensive care unit

Noise is regarded as an important source of disturbance, agitation, and auditory disorders.3–9 These
stress in neonatal intensive care units (NICUs).1 The responses aggravate infants’ energy consumption, induce
fetus receives low-pitched sounds derived mainly from physiological instability and can affect growth and de-
maternal rhythmic heart beating and intestinal peristal- velopment. In addition, recent studies reveal that the
sis. In contrast, newborns in the NICU are subjected to sound intensity not only has an adverse effect on the
more intense, arrhythmic sounds.2 Monitoring physio- infants but also can interfere with caregivers’ communi-
logical variations in newboms with critical conditions, cation and job performance.10
using life support equipment, and performing caregiving In 1974, the American Academy of Pediatrics
activities inevitably makes noise. Exposure to excessive (AAP) recommended that noise levels exceeding 58 dB
noise has been associated with infants’ heart rate accel- be avoided in the NICU.1 Furthermore, in 1997, due to
eration, bradycardia, decline in oxygenation, increase in the adverse impacts of noise on newborns, the AAP also
muscle tension, blood and intracranial pressure, sleep recommended that the ideal sound level in the NICU

1
Department of Nursing, College of Medicine, National Cheng Kung Address for correspondence and reprint requests: Dr. Chyi-Her Lin,
University, Taiwan, Republic of China; 2Nursing Department, Chi- Department of Pediatrics, National Cheng Kung University, Taiwan,
Mei Foundation Hospital, Taiwan, Republic of China; 3Department No. 1 Ta-Hsueh Road, Tainan, Taiwan, 701, Republic of China.
of Pediatrics, National Cheng Kung University Hospital, Taiwan, Am J Perinatol 2006;23:265–272. Copyright # 2006 by Thieme
Republic of China; 4Photoelectronics Application Department, Semi- Medical Publishers, Inc., 333 Seventh Avenue, New York, NY 10001,
conductor Process Equipment Division, Mechanical Industry Research USA. Tel: +1(212) 584-4662.
Laboratories, Industrial Technology Research Institute, Taiwan, Re- Published online: May 9, 2006.
public of China. DOI 10.1055/s-2006-941455. ISSN 0735-1631.
265
266 AMERICAN JOURNAL OF PERINATOLOGY/VOLUME 23, NUMBER 5 2006

is < 45 dB.11 For preserving better sleep of infants,


Philbin et al12 recommended that continuous sound in
any patient’s space should not exceed an hourly equivalent
sound level of 50 A-weighted decibels (dBA) and the
duration of sound level exceeding 55 dBA should only be
10% of the time or 6 minutes in an hour.
Studies have shown that human-related factors
accounted for a major source of noise in the NICU, such
as conversations among the staff, bumping metalware,
and forcefully opening and closing trash can lids.4,13,14
These findings indicate that modification of staff behav-
ior may be the most effective way to attenuate NICU
sound. Researches have documented that staff behaviors
can be modified by education.15–17 However, it is diffi-
cult to remember to prevent noise all the time, and staff
may need to be reminded when the sound level increases.
Few studies have focused NICU noise reduction using
an alarm reminder. The objective of this study was to

Downloaded by: UPSTATE Medical University. Copyrighted material.


assess the benefits of a bedside noise-sensor light alarm
on reducing the sound level and sudden peak noise
(SPN) episodes in the NICU.
Figure 1 The appearance of the noise-sensor light alarm. The
red bulb lights when the noise level exceeds 65 dBA.
METHODS

Study Design Noise-Sensor Light Alarm


A one-group pre-post test design was conducted from The assembly structure and circuit diagram of the noise-
September 2002 to February 2003 in a 20-bed NICU of sensor light alarm are shown in Figs. 1 and 2, respec-
a tertiary care center in southern Taiwan. The sound tively. The operation principle of the noise-sensor light
level inside incubators and at radiant-heated beds, both alarm uses microphone and amplifier circuits to convert
of which were combined with ventilators, were moni- the environmental noise (the sum of acoustic energy
tored continuously for a period of 7 days before and from a human voice, equipment noise, air conditioner
1 month after introduction of the noise-sensor light noise, etc.) to electrical potential. This electrical poten-
alarm system. The same model of incubator was used tial, when it is higher than the set point of potential, will
throughout the study. During the monitoring period, drive the alarm signal of the red light-emitting diode
day, evening, and night shifts of 5 weekdays were (LED) lamp.
grouped, respectively, and then one sample from each This invention was developed by one of the
shift was selected randomly. Sources of SPN in these authors (Y.-Z.C.) for this study, and costs approximately
24 hours during pretest and posttest periods were ob- $200 U.S. Becuse human ears are more sensitive and
served directly by three trained research assistants. susceptible to the adverse effect of high-frequency noise

Figure 2 The circuit diagram of the noise-sensor light alarm. VCC, electrical potential; X1, microphone; R1-11, electric resistance; C1-2,
electric capacitance; D1, diode; U1-2, integrated circuit; LED, light-emitting diode.
NOISE-SENSOR LIGHTALARMIN THE NICU/CHANG ET AL 267

than to low-frequency noise,3 the A-weighted decibel for comparisons. The hourly mean sound levels of pretest
scale was applied to monitor noise levels in the NICU. and posttest periods were compared with paired t sta-
Based on a previous study indicating that the mean loud tistics. The sources and frequencies of peak noise were
conversation or monitor alarm sound level was approx- categorized and recorded directly from bedside observa-
imately 65 dB,15 the light alarm was then set to an tions. Mean sound levels of SPN were calculated by
electrical potential threshold induced by noise at drawing data from synchronized records in the computer.
65 dBA. When the environmental noise was > 65 dBA,
a red LED light of the alarm was automatically turned on
to remind the staff to lower their voices or modulate the RESULTS
source of noise. After repetitive comparisons and calibra- During the study periods the hourly mean sound level
tions with a sound meter (TES 1336 Soundmeter; TES inside the incubator was 58.0  0.6 (standard deviation)
Electrical Electronic Corp., Taiwan, Republic of China) and 56.4  0.7 dBA in the pretest and posttest, respec-
during a testing period, the alarm was found to have a tively. The paired t test shows that the difference
high sensitivity of 94% and specificity of 92%. between the two periods was significant (t ¼ 8.617;
The NICU consisted of two separated areas. A p < 0.001). The hourly mean sound level on the radi-
previous study found that the area near the nurse station ant-heated bed was 58.0  2.4 and 58.1  2.0 dBA in the
with three opened rooms and a total of 14 beds had a pretest and posttest, respectively, and no significant
higher noise level than the other area.13 Therefore, this difference was found (t ¼ 0.715; p ¼ 0.476). The envi-

Downloaded by: UPSTATE Medical University. Copyrighted material.


area with more noise was investigated. A noise-sensor ronmental sound levels were divided into four intensity
light alarm was installed on the central wall. The groups based on the pretest mean of the radiant-heated
researcher informed the NICU staff of the purpose of bed (58 dBA) and with an increment criterion of 2 SD of
explained the alarm and the meaning when the red light the sound level (approximately 5 dBA). The respective
was turned on. percentage of the total observation time, based on the
aggregated noise level per minute, at each level was
calculated (Table 1). The percentage of sound levels
Measurement of Environmental Sound Level < 58 dBA in the incubator increased by 28%, whereas
The environmental sound level was measured using the the percentages for the other three groups decreased
TES 1336 Soundmeter with accuracy an of  1.5 dBA. after using the alarm. The proportional test showed that
The sound meter’s microphone was placed in the in- these differences were all significant (p < 0.001). Similar
cubator and at the radiant-heated bed within 10 cm of an change patterns were found on the radiant-heated bed.
infant’s ear. A computer with data logging software was The percentage of sound levels < 58 dBA increased,
used to record the level of sound monitored second by whereas the percentage of levels for the other three
second. groups decreased. Although amounts of change were
small, they were all statistically significant (p < 0.001).
The peak period of sound levels across 24 hours
Measurement of SPN was further examined by averaging hourly sound level
Consisting of the set point of the noise-sensor light throughout a day. In the incubator, as shown in Fig. 3, a
alarm, SPN in this study was defined as any abrupt noise higher sound level phase occurred once every 3 to
event  65 dBA. Three research assistants were trained 4 hours and the sound levels from 10 A.M. to 4 P.M.
by one of the authors (Y.-J.P.) to observe and record were higher than that in remaining hours before the
sources of the SPN (interrater reliability was 87%). The light alarm was installed. However, the hourly mean
researcher first summarized possible noise sources into a sound levels throughout the day were distributed mostly
number of categories and used abbreviated terms to at lower values after the device was installed. The phase
represent each of them. During bedside observation of peak sound level was found in the period from 8 A.M. to
periods, the corresponding categorical terms were noted
on a chronological table when peak noise events oc- Table 1 Percentages of Total Observation Time by
curred. To allow easy recall of data correlating to Sound Levels and Beds with a Ventilator
particular noise events from the computer records, the Incubator Radiant-Heated Bed
Sound Level
research assistant used timers synchronized with the (dBA) Pretest Posttest Pretest Posttest
computer and noted the time a peak noise occurred
with the precision the 1-second level. < 58 52.0 80.2* 56.2 60.3*
58–62 44.7 17.8* 37.2 34.8*
63–67 2.4 1.5* 4.8 4.0*
Statistical Analysis  68 0.9 0.4* 1.8 0.9*
Environmental sound level data were initially acquired Total 100 100 100 100
second by second and aggregated by minute and hourly *Results of proportional test with p < 0.001.
268 AMERICAN JOURNAL OF PERINATOLOGY/VOLUME 23, NUMBER 5 2006

Figure 3 Noise distribution of the incubator of patients receiving a ventilator support during a 24-hour period.

Downloaded by: UPSTATE Medical University. Copyrighted material.


1 P.M. For the radiant-heated beds, neither a difference alarm installation were conversations (51.3%), monitor
in hourly mean sound levels nor a clear periodic peak alarms (18.7%), nursing activities (9.7%), and movement
phase tendency was observed before and after the alarm of carts or equipment (4.3%). These sources did not
was installed. Similarly, the sound levels from 8 A.M. to change after the light alarm was installed; however,
4 P.M. were relatively higher than that in the remaining occurrences of SPN decreased by 63.8% (Table 3).
hours (Fig. 4).
Direct observation of SPN episodes in 24 ran-
domly selected hours of the weekday showed that an DISCUSSION
SPN inside the incubator before and after the light alarm Data from this study indicate that a noise-sensor light
was installed occurred 630 and 185 times, respectively. alarm intervention has a positive effect on reducing
The four primary sources of SPN before light alarm use environmental sound in the NICU. In the incubators
were nursing activities (57.5%), monitor alarms (19.2%), with ventilators, the percentage of observation time of
staff conversations (10.3%), and oxygen supply proce- sound levels > 58 dBA was decreased after introduction
dures (6.3%). These items remained the main sources of of the sensor-light alarm system. Although the mean
SPN after the device was installed; however, the occur- sound level was only reduced by 2 dBA, a reduction of
rence of SPN was reduced by 71.5% (Table 2). An SPN 3 dBA reflects a 50% of noise reduction perceived by the
at the radiant-heated bed with ventilator before and after human ear because decibel measurements use a logarith-
light alarm use occurred 2067 and 748 times, respec- mic scale to measure sound pressure.11,15 Therefore,
tively. The four primary sources of SPN before light the findings were clinically significant in environmental

Figure 4 Noise distribution on the radiant-heated bed of patients receiving a ventilator support during a 24-hour period.
NOISE-SENSOR LIGHTALARMIN THE NICU/CHANG ET AL 269

Table 2 Peak Noise Distribution in the Incubator with a Ventilator before and 1 Month after Using Noise-Sensor Light
Alarm
Frequency (%) Noise Level (dBA)*
Source Pretest Posttest Pretest Posttest

Caregiving (rubbing, moving, or wrapping object) in the incubator 362 (57.5) 98 (53.0) 69.3  4.1 68.5  3.5
Alarm of monitors 121 (19.2) 14 (7.6) 71.4  4.6 67.8  3.6
Staff conversation 65 (10.3) 17 (9.2) 68.2  3.1 68.8  4.6
Handling equipment of oxygen supply 40 (6.3) 4 (2.2) 72.9  7.3 66.7  2.7
Other (moving cart or chair, closing trash can lid or drawers, 32 (5.4) 50 (26.7) 67.6  3.7 67.4  3.0
bumping or dropping objects, cleaning apparatus and containers,
tearing plastic or paper bag, reading charts, opening, closing,
bumping, and cleaning the incubator, on or off monitors)
Total 630 (100.0) 185 (100.0) 69.5  4.3 68.1  3.4
*Data expressed as mean  1 standard deviation.

support for newborns. On the radiant-heated beds, the as high as 70 to 80 dB when a ventilator or other oxygen
mean sound level did not change after the noise-sensor supply devices were incorporated.22–24 These study re-

Downloaded by: UPSTATE Medical University. Copyrighted material.


light was introduced. Only a small increase in sults revealed that the sound level in the NICU has great
the percentage of observation time that sound levels variation. The daily mean sound level may not be
< 58 dBA was found. This finding differed from the sufficient to describe the actual clinical situation.
experimental result in the incubator. This was probably Through continuous monitoring, this study used per-
because a radiant-heated bed is an open space without centages of observation time for specific noise level
any shielding, unlike the environment in an incubator. ranges to display newborns’ exposure to noise. This is
Thus, there might be more uncontrollable noise sources. another way to determine the sound context of the
In addition, the noise-sensor light threshold was set NICU and provides a new method for assessing the
at 65 dBA. The red light turned on only if a sound effects of the noise-sensor light alarm.
impulse exceeded the threshold. Therefore, this might Similar to the results from previous reports, this
have modified the clinical personnel’s awareness of more study found that the highest sound level, whether in
intense noise and made them less sensitive to the noise incubators or radiant-heated beds, occurred in the day-
< 65 dBA. time,13,24–26 suggesting that noise is associated primarily
Previous studies indicated that sound levels in the with staff activities. Thus, analysis of staff activities in
NICU ranged from 45 to 135 dB and the mean sound the NICU can be a useful basis for noise reduction.
level varied considerably from 54 to 80 dBA.11,13,18–21 The most significant effect from the noise-sensor
Long et al14 found that incubators could partially block light was to reduce peak noise occurrences. Abrupt
noise and preserve a quieter internal space than the noise may awaken infants and causes startle responses,
external environment by 5 to 18 dB. Other studies which are stressful to the vulnerable infants.12 Because
reported that sound levels inside incubators could reach SPN events in the NICU occurred mainly due to

Table 3 Peak Noise Distribution at the Radiant-Heated Bed with a Ventilator before and 1 Month after Using the
Noise-Sensor Light Alarm
Frequency (%) Noise Level (dBA)*
Source Pretest Posttest Pretest Posttest

Staff conversation 1062 249 (33.3) 68.4  2.9 68.9  3.9


Alarm of monitors 386 (18.7) 108 (14.4) 69.5  3.1 69.3  3.6
Caregiving (rubbing, moving or wrapping object) on the radiant warmer 201 (9.7) 39 (5.2) 67.6  2.1 71.4  4.6
Moving cart, chair, or equipment 90 (4.3) 33 (4.4) 69.3  2.7 67.9  4.9
Opening or closing trash can lid 54 (2.6) 33 (4.4) 68.0  3.1 68.8  4.1
Others (bumping objects or stainless cart, handling equipment of 276 (13.4) 286 (38.3) 69.3  2.8 68.3  3.7
oxygen supply, cleaning the warmer bed, on or off monitors,
tearing plastic or paper bag, closing drawers at the station,
staff walking or running, cleaning apparatus and containers,
reading charts, dropping objects, phone or pager ringing, hanging)
Total 2069 748 (100.0) 68.6  2.8 68.6  3.9
*Data expressed as mean  1 standard deviation.
270 AMERICAN JOURNAL OF PERINATOLOGY/VOLUME 23, NUMBER 5 2006

human-related factors such as caregiving procedures nel, and ventilators used in the unit during the pretest
and staff conversations, these sources have favorable and posttest periods. Whether the noise-sensor light
potential for modification. The staff members involved alarm could have the similar effects in other settings
in this study were not informed about what peak noise warrants further study.
sources were detected during the pretest period. They
only received immediate feedback via the light alarm,
making them aware of noise occurrences after installa- ACKNOWLEDGMENTS
tion of this alarm. One month after noise-sensor light This study was sponsored by the Cheng-Hsing Medical
alarm introduction, SPN events were reduced by 70 and Foundation and Chinese Premature Baby Foundation.
63% in incubators and radiant-heated beds, respec- The authors thank the staff in the NICU of NCKUH
tively. Similar results were shown in a previous study for their support in the noise-sensor light alarm shape
displaying noise intensities from a variety of caregiving design and data collection.
activities on videotape, in-service education for noise
prevention, and a 3-month program of behavior mod-
ification for staff in the ICU. After education, the mean REFERENCES
sound level in the unit decreased,15,27 and the peak
1. American Academy of Pediatrics Committee on Environ-
sound events > 80 dBA were reduced by 28%.27 How-
mental Hazards. Noise pollution: neonatal aspects. Pediatrics
ever, some studies found that education programs may 1974;54:476–479

Downloaded by: UPSTATE Medical University. Copyrighted material.


not be effective in noise reduction. Although staff 2. Gottfried AW, Hodgman JE. How intensive is newborn
members might be aware of the sources of noise, they intensive care? An environmental analysis. Pediatrics 1984;74:
may not be aware of when to control the noise.18,27 This 292–294
indicates the importance of an immediate alarm for 3. Catlett AT, Holditch-Davis D. Environmental stimulation of
noise reduction. A similar recommendation of a sound- the acutely ill premature infant: physiological effects and
nursing implications. Neonatal Netw 1990;8:19–26
detection device that sends a signal to a vision alarm
4. DePaul D, Chambers SE. Environmental noise in the
when the sound level is considered too high has been neonatal intensive care unit: implications for nursing practice.
addressed for the purpose of infants’ developmental J Perinat Nurs 1995;8:71–76
supportive care.28 5. Osfeld B, Krawciw N, Burke S, et al. Impact of environment
Although the mean sound level of incubators on the high risk infant: the effect of monitor alarm. Pediatr
after using a noise-sensor light alarm was reduced, it is Res 1990;27:218A
not ideal. This is probably because the light alarm was 6. Trapanotto M, Benini F, Farina M, et al. Behavioral and
physiological reactivity to noise in the newborn. J Paediatr
set to a high threshold ( 65 dBA). It is possible that
Child Health 2004;40:275–281
if the noise-sensor light were set to a threshold of 7. Van Reempts PJ, Wouters A, De Cock W, Van Acker KJ.
60 dBA and the sound level values under this threshold Clinical defense response to cold and noise in preterm
are simultaneously displayed to the staff, the observa- neonates after intrauterine conditions associated with chronic
tion time for environmental sound levels > 58 dBA stress. Am J Perinatal 1996;13:277–286
would become longer. This issue deserves further 8. Wharrad HJ, Davis AC. Behavioral and autonomic responses
study. The current study indicated that staff caregiving to sound in pre-term and full-term babies. Br J Audiol 1997;
31:315–329
activities and conversations were the main sources
9. Zahr LK, Balian S. Responses of premature infants to routine
of peak noise, in agreement with previous stud- nursing interventions and noise in the NICU. Nurs Res 1995;
ies.13,14,24,29 Therefore, further studies focusing on 44:179–185
the context and factors involving these peak noise 10. Thomas KA, Martin PA. NICU sound environment and the
events will be crucial to sound reduction. If a staff potential problems for caregivers. J Perinatol 2000;20:S94–
education program on sound reduction is conducted S99
with noise-sensor light use, the combination may 11. American Academy of Pediatrics Committee on Environ-
mental Hazards. Noise: a hazard for the fetus and newborn.
create a synergistic effect.
Pediatrics 1997;100:724–727
Reducing sound levels in the NICU is a way to 12. Philbin MK, Robertson A, Hall JW. Recommended permis-
assist and support newborn development and should also sible noise criteria for occupied, newly constructed or renovated
be the common objective of NICU personnel. The hospital nurseries. The Sound Study Group of the National
noise-sensor light alarm described in this study de- Resource Center. J Perinatol 1999;19:559–563
creased NICU sound level and episodes of SPN. How- 13. Chang Y-J, Lin C-H, Lin L-H. Noise and related events in a
ever, whether long-term use of noise-sensor light can neonatal intensive care unit. Acta Paediatr Taiwan 2001;42:
212–217
bolster the effect of noise reduction or lead to a state of
14. Long JG, Lucey JF, Philip AG. Noise and hypoxemia in the
neglect requires further investigation. Furthermore, this intensive care nursery. Pediatrics 1980;65:143–145
study is limited by the fact that the observation and 15. Elander G, Hellstrom G. Reduction of noise levels in
manipulation were done in a single NICU. The results intensive care units for infants: evaluation of an intervention
could be confounded by the number of patients, person- program. Heart Lung 1995;25:376–379
NOISE-SENSOR LIGHTALARMIN THE NICU/CHANG ET AL 271

16. Hale DR. Noise in the hospital: a quality improvement 24. Chen H-F, Chang Y-J. Noise distribution of an incubator
approach. J Nurs Admin 1996;26:4–8 with nebulizer at neonatal intensive care unit in southern
17. Philbin MK, Gray L. Changing levels of quiet in an intensive Taiwan. J Nurs Res 2001;9:25–32
care nursery. J Perinatol 2002;22:455–460 25. Moore MM, Nguyen D, Nolan SP, et al. Intervention to
18. Horgan MJ, Eston C. Sources of environmental noise: reduce decibel levels on patient care units. Am Surgeon 1998;
perceptions by NICU staffs. Pediatr Res 1999;45:201A 64:894–900
19. Levy GD, Woolston DJ, Browne JV. Mean noise amounts in 26. Robertson A, Cooper-Peel C, Vos P. Peak noise distribution in
level II vs level III neonatal intensive care units. Neonatal the neonatal intensive care unit. J Perinatol 1998;18:361–364
Netw 2003;22:33–38 27. Kahn DM, Cook TE, Carlisle CC, et al. Identification and
20. Thomas K. How the NICU environment sounds to a preterm modification of environmental noise in an ICU setting. Chest
infant. Am J Matern Child Nurs 1989;14:249–251 1998;114:535–540
21. Weibley T. Inside the incubator. Am J Matern Child Nurs 28. Bowie BH, Hall RB, Faulkner J, Anderson B. Single-room
1989;14:96–100 infant care: future trends in special care nursery planning and
22. Bess FH, Peek B, Chapman J. Further observations on noise design. Neonatal Netw 2003;22:27–34
levels in infant incubators. Pediatrics 1979;63:100–106 29. Gorski P, Hole W, Leonard C, Martin J. Direct compu-
23. Blennow G, Svenningsen N, Almqist B. Noise levels in infant ter recording of premature infants and nursery care: distress
incubators (adverse effects?). Pediatrics 1974;53:92–32 following two interventions. Pediatrics 1983;72:198–202

Downloaded by: UPSTATE Medical University. Copyrighted material.


Downloaded by: UPSTATE Medical University. Copyrighted material.

You might also like