You are on page 1of 12

Aust N Z J Obstet Gynaecol 2023; 63: 278–289

DOI: 10.1111/ajo.13667

SYSTEMATIC REVIEW

Clinical practice guidelines for intrapartum


cardiotocography interpretation: A systematic review

James Brown1,2 , Dhusyanthy Kanagaretnam1 and Monica Zen1

1
Obstetrics and
Gynaecology, Westmead Hospital, Background: Clinical practice guidelines on intrapartum cardiotocography (CTG)
Sydney, New South Wales, Australia interpretation provide structured tools to detect fetal hypoxia. Despite frequent
2
University of Sydney, Sydney, New use of different guidelines, little is known about their comparable consistency. We
South Wales, Australia
sought to appraise guidelines relevant to intrapartum CTG interpretation and sum-
Correspondence: Dr. James
Brown, Obstetrics and marise consensus and non-­consensus recommendations.
Gynaecology, Westmead Hospital,
Aims: To compare existing intrapartum CTG interpretation guidelines.
Westmead, NSW 215, Australia.
Email: j.h.w.brown@gmail.com Materials and methods: We searched PubMed, CINAHL, Cochrane, Embase,
guideline databases and websites of guideline development institutions using
Conflict of Interest: The authors report
no conflict of interest. terms ‘cardiotocography’, ‘electronic fetal/foetal monitoring’, and ‘guideline’ or
Received: 18 November 2022; equivalent term. The search was restricted to English-­language articles published
Accepted: 20 February 2023 between January 1980 and January 2023 and excluded animal studies. The initial
search yielded 2128 articles with 1253 unique citations. Guidelines were included if
they: used English as the reporting language; included CTG interpretation criteria or
guidelines as a primary objective; were published or updated since 1980; and were
the most recently updated publications when multiple versions were identified.
Results: Nineteen studies were considered for full review, and 13 met inclusion
criteria. Two reviewers independently assessed guideline quality using the AGREE
II instrument, and synthesised consensus and non-­consensus recommendations
using the content analysis approach. Most guidelines employed a three-­tier inter-
pretation framework. There were significant differences between the guidelines
for relative importance of key CTG features such as accelerations, decelerations
and variability, with respect to the outcome of fetal hypoxia.
Conclusions: There are significant differences between key intrapartum CTG in-
terpretation guidelines currently being used. Greater consistency is needed across
CTG interpretation guidelines to improve the quality of data, clinical governance,
monitoring of outcomes, and to support future developments.

KEYWORDS
cardiotocography, practice guidelines, systematic review, intrapartum, electronic
fetal monitoring

INTRODUCTION high income countries.1 Through interpretation of the fetal heart


rate and uterine contractions, clinicians can potentially detect
Cardiotocography (CTG) or electronic fetal monitoring (EFM) is the early signs of fetal hypoxia2 and act accordingly. However, large
most used technique for assessing intrapartum fetal wellbeing in studies of CTG effectiveness are inconclusive—­hampered by high

278wileyonlinelibrary.com/journal/anzjog
© 2023 Royal Australian and New Zealand College of Obstetricians and Gynaecologists.
1479828x, 2023, 3, Downloaded from https://obgyn.onlinelibrary.wiley.com/doi/10.1111/ajo.13667 by Universitat Autonoma De Barcelona, Wiley Online Library on [18/07/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
J. Brown et al. 279

false-­positive rates, increased surgical births, and underpowered published between January 1980 and January 2023 and excluded
3,4
perinatal morbidity analyses. animal studies.
Standardised interpretation of CTG patterns is critical to clin- Search terms used were ‘cardiotocography’, ‘electronic fetal/
ical decision making.3,5 Poor interobserver consistency has been foetal monitoring’ for guideline databases and websites of guide-
highlighted as a critical limitation of the technology.6 In an effort line development institutions, combined with terms of ‘guideline*’,
to improve the efficacy of CTG, a number of national and college ‘guidance*’, ‘recommendation*’, ‘consensus*’, ‘best practice*’,
guidelines have been developed and refined to provide a frame- ‘statement*’, ‘standard*’, ‘practice parameter*’, ‘position paper’
work for more accurately interpreting CTG features.7 and ‘position stand’ for electronic databases. A full list of guide-
In 1986, the International Federation of Gynaecology and line databases and websites of guideline development institutions
Obstetrics (FIGO) developed the first general interpretation crite- that were searched, along with detailed search strategies, are pro-
ria based on key fetal heart rate pattern morphologies and their vided (Appendix S1 and S2). Reference lists of relevant articles and
relationship to the contraction.8 Since then, criteria have been reviews were manually searched for additional reports.
refined and differentiated, but with no significant change to the This systematic review is reported in accordance with the
4,9–­12 4
interobserver variability. In turn, the FIGO-­15 interpretation Preferred Reporting Items for Systematic Reviews and Meta-­
template, created by an international consensus panel, resulted Analyses (PRISMA) criteria. The protocol for this systematic review
in being more prescriptive in aligning features with grades of fetal was prospectively registered on PROSPERO (CRD42018085085).
hypoxia,7 and therefore, more likely to induce clinical action. This Eligibility criteria were predetermined. Guidelines were in-
represented a shift in how CTG patterns were interpreted, offer- cluded if they: (i) used English as the reporting language; (ii)
ing a practical escalation that has been merged, extrapolated and/ were a labelled guideline, or recommendation, or consensus, or
or integrated with local and international guidelines for intrapar- practice parameter, or position paper/stand; (iii) included intra-
tum monitoring.2 Countries have chosen to adopt or adapt the partum CTG interpretation criteria or guidelines as a primary ob-
7
FIGO-­15 guidelines based on local preference and evidence. jective; (iv) were published or updated since 1980; and (vi) were
The fractured and staggered adoption of standardised CTG the most recently updated iteration when multiple serial versions
interpretation criteria has important ramifications. First, it means were identified.
that women giving birth are treated differently based on location Literature search, screening of search outputs, identification
and local consensus. Second, the tendency for local adaptation of eligible guidelines and data extraction were performed by inde-
means that with each minor permutation, the criteria may have pendently two reviewers. A third reviewer checked data for errors
migrated further away from where the current evidence lies. and resolved discrepancies or disagreements through discussions
Third, varied interpretation and consequent management makes or consultations.
even retrospective analysis of intrapartum CTG patterns and in- Two review authors independently extracted the following
terpretation criteria impossible at the scale required to answer data from all included guidelines: title, published year, last up-
some of the historically unanswered concerns about the utility dated year, guideline group, country or region, evidence bases,
and safety of the technology.13 and method of access.
To address the inconsistency, we used systematic review Two review authors extracted all the relevant content on CTG
methods to identify, summarise and compare available intrapar- interpretation from each guideline into predetermined data col-
tum CTG interpretation guidance. Through this we aim to map lection forms. Recommendations were organised in accordance
the current state of CTG interpretation guidelines worldwide and with accepted interpretation algorithms; significant deviations
move toward greater consistency and strategically aligned re- from this scaffold are highlighted in the text. On completion, unan-
search in this clinical area. imous and discrepant recommendations were categorised, sum-
marised and analysed, respectively. During the whole process, a
third reviewer checked the data for errors and resolved discrep-
MATERIALS AND METHODS ancies or disagreements through discussions or consultations.
The AGREE II (Appraisal of Guidelines for Research and
A systematic literature search was performed to identify Evaluation II) instrument was used to critically appraise the meth-
intrapartum cardiotocography interpretation guidelines via odological rigour and to report the quality of included guidelines.14
guideline databases, websites of guideline development The AGREE II consists of 23 items organised into the following six
organisations in midwifery or obstetrics and gynaecology, domains: scope and purpose (three items), stakeholder involve-
government websites, as well as electronic databases including ment (three items), rigour of development (eight items), clarity of
PubMed, CINAHL, Cochrane, and Embase. Clini​calTr​ials.gov presentation (three items), applicability (four items) and editorial
was not included in the search as the review is examining independence (two items). Each of the 23 items was scored on a
clinical practice guidelines and not clinical trial or effectiveness seven-­point Likert scale ranging from one (strongly disagree) to
data. The search was restricted to English-­
language articles seven (strongly agree).
1479828x, 2023, 3, Downloaded from https://obgyn.onlinelibrary.wiley.com/doi/10.1111/ajo.13667 by Universitat Autonoma De Barcelona, Wiley Online Library on [18/07/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
280 Review of CTG guidelines

As recommended by the AGREE II instrument,14 each of the Characteristics of included guidelines


six-­domain scores were calculated independently by summing up Table 1 presents the characteristics of the guidelines included.
all the scores of the individual items in a domain, as well as by The majority (69%) of the guidelines were published or updated
scaling the total as a percentage of the maximum possible score within the last ten years.
for that domain. Domain scores were calculated as conducted in The guidelines originated from the USA (three guidelines), the
similar studies: (obtained score –­minimum possible score)/(max- UK (three guidelines), Oceania (one guideline), Sweden (one guide-
imum possible score –­minimum possible score). The minimum line), Canada (one guideline), Ireland (one guideline), Japan (one
possible score was calculated as 1 × (number of items) × (num- guideline), Germany (one guideline) and Sri Lanka (one guideline).
ber of appraisers). The maximum possible score was calculated All included guidelines were at least partially evidence-­based
as 7 × (number of items) × (number of appraisers). We defined a (ie references to peer-­reviewed research as foundation for pol-
cut-­off of 50%, and scores above this were considered acceptable. icy recommendation) with consensus recommendations (ie policy
The quality assessment of all included clinical practice guidelines recommendations based on the conclusions of a recognised pro-
was independently performed by two appraisers. fessional body in the relevant field), primarily carried over from
previous published criteria or workshops. All were available as
electronic copies.
RESULTS

Systematic search results AGREE II assessment


Our initial search on 31 January 2023 retrieved 2128 records. After Overall scores and domain scores of the included 13 guidelines
removal of duplicate records (n = 875), we screened the titles are shown in Appendices S4 and S5, respectively.
and/or abstracts of 1253 references, of which 19 articles were All but one of the 13 guidelines scored >50% for total scores
subsequently reviewed in full text. Among them, six articles were (59.42–­77.54%). Of the 13 guidelines, the NICE, ACOG and SWE-­
excluded (see Appendix S3) and the final selection yielded a total 17 guidelines scored >50% across all six domains—­and gener-
of 13 clinical practice guidelines (Fig. 1). ally outperformed all the other guidelines in each domain. The

FIGURE 1 Flow diagram illustrating the selection of guidelines.


1479828x, 2023, 3, Downloaded from https://obgyn.onlinelibrary.wiley.com/doi/10.1111/ajo.13667 by Universitat Autonoma De Barcelona, Wiley Online Library on [18/07/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
J. Brown et al. 281

TABLE 1 Characteristics of included guidelines

Abbreviation Guideline group or authors First published Last updated Country/region


15
ACOG American College of Obstetricians and 2009 2009 USA
Gynaecologists
DGGG16 German Society of Gynaecology and Obstetrics 2014 2014 Germany
FIGO4 The International Federation of Gynaecology and 1986 2015 United Kingdom
Obstetrics
IRELAND17 National Women and Infants Health Programme 2019 2019 Ireland
Ireland
NICE18a National Institute for Health and Care Excellence 2014 2022 United Kingdom
(NICE)
NICHD19 National Institute of Child Health and Human 2008 2008 USA
Development (NICHD)
PARER20 Parer et al.b 2007 2007 USA
21
PCJSOG The Perinatology Committee of the Japan Society of 2010 2010 Japan
Obstetrics and Gynaecology (PCJSOG)
RANZCOG22 Royal Australian and New Zealand College of 2002 2019 Australia
Obstetricians and Gynaecologists (RANZCOG)
SLCOG23 Sri Lanka College of Obstetricians & Gynaecologists 2013 2013 Sri Lanka
(SLCOG)c
SOGC24 Society of Obstetricians and Gynaecologists of 2007 2018 Canada
Canada (SOGC)
SWE-­177 Ekengård et al.b 2009 2017 Sweden
25 d
WALES Wales Maternity Network 2018 2018 Wales (UK)
a
Previously RCOG Electronic Fetal Monitoring guideline.
b
Downloaded from journal.
c
Evidence base not stated.
d
Consensus-­based only.

highest average domain score of all guidelines was ‘Scope and The RANZCOG guideline22 used a four-­category system. The
Purpose’ (88.46%), followed by ‘Clarity of Presentation’ (86.32%), first three categories were similar to the three-­category guidelines
‘Stakeholder Involvement’ (62.39%), ‘Rigour of Development’ described above, with an additional fourth category of CTG fea-
(57.21%), ‘Applicability’ (53.21%), and ‘Editorial Independence’ tures of severe fetal compromise that require immediate man-
(44.87%). All 13 guidelines obtained scores >50% in the domain of agement, which included: bradycardia, absent baseline variability,
‘Clarity of Presentation’. sinusoidal pattern, complicated variable decelerations with re-
duced or absent variability, and late decelerations with reduced
or absent variability. Many of these features are included in the
Summary of interpretation criteria
Abnormal category of the three-­category guidelines, such as: sinu-
soidal pattern, complicated variable decelerations with reduced
Categorisation
or absent baseline variability, or bradycardia.
All but three of the 13 guidelines grouped CTG patterns into PCJSOG21 and Parer et al.20 outlined a five-­category system
three distinct risk categories (Tables 2–­4), with one using four that describes escalating concerns depending on the character
22 20,21
categories and the other two using five categories. of decelerations and their relationship to other features and are
Of those guidelines with three categories, the lowest risk cat- described further below.
egory was labelled either ‘Category I', ‘Normal’ or ‘Reassuring’
(Normal); the middle risk category was labelled ‘Category II',
Features considered
‘Atypical’, ‘Non-­reassuring’ or ‘Suspicious’ (Suspicious); the highest
risk category was labelled ‘Category III', ‘Pathological’ or ‘Abnormal’ All guidelines include consideration of baseline fetal heart rate,
(Abnormal). There was no consistent relationship between the fetal heart rate variability, and decelerations.
labelling used for categories and the features of that category.
Despite inconsistency in labelling most guidelines stratified the re-
Fetal baseline heart rate
quired action for each category similarly: no intervention needed
(Normal), increased vigilance recommended (Suspicious), and ur- All guidelines consider a fetal baseline heart rate between 110
gent medical review needed (Abnormal). and 160 beats per minute (BPM) to be Normal. All guidelines
1479828x, 2023, 3, Downloaded from https://obgyn.onlinelibrary.wiley.com/doi/10.1111/ajo.13667 by Universitat Autonoma De Barcelona, Wiley Online Library on [18/07/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
282 Review of CTG guidelines

TABLE 2 CTG three-­tier interpretation criteria—­low probability of fetal acidosis category (Normal)

Baseline
Guideline Label rate (bpm) Variability (bpm) Decelerations Accelerations

ACOG15 Category I 110–­160 6–­25 Absent late or variable Present or absent


decelerations, OR
Absent or present early
decelerations.
DGGG16 Normal 110–­160 >5 (during the interval when No decelerations. Two accelerations in
no contractions occur) 20 min
FIGO4 Reassuring 110–­160 5–­25 No repetitive decelerations Present or absent
(>50% of contractions).
IRELAND17 Normal 110–­160 5–­25 No repetitive—­>50% Present or absent
contractions
NICE18 White 110 to 160 5–­25 No decelerations, OR Present or absent
Early, OR
Variable decelerations with no
concerning characteristics for
less than 90 min.
NICHD19 Category I 110–­160 6–­25 Absent late or variable Present or absent
decelerations, OR
Absent or present early
decelerations.
SLCOG23 Reassuring 110–­160 ≥5 None Present
24
SOGC Normal 110–­160 6–­25, OR None, OR ≥2 accelerations with
≤5 (absent or minimal) for Occasional variable <30 sec. acme of ≥15 bpm, lasting
<40 min. 15 sec <40 min of testing
SWE-­177 Normal 110–­160 5–­25 No repetitive (>50%) Not included.
decelerations, OR
Uniform early decelerations,
OR
Variable decelerations lasting
<1 min
WALES25 Reassuring 110–­160 5–­25 True early decelerations, OR Present or absent
V-­shaped variable
decelerations with <50% of
contractions, OR
V-­shaped variable
decelerations with >50% of
contractions for less than
90 min, OR
U-­shaped variable
decelerations with <50%
of contractions (if all other
features NAD)

considered a fetal heart rate below 100 BPM to be abnormal and consider variability of ≤5 BPM to be a Normal feature if present
a feature of the highest possible risk category. Four guidelines for less than 40 min. There was marked variation in Suspicious
(DGGG, NICE, SLCOG, WALES) considered fetal tachycardia above variability criteria, with most guidelines indicating that vari-
180 BPM to also meet criteria for the highest risk category.16,18,23,25 ability below 5 BPM for between 30 to 90 min indicates mod-
Most guidelines noted a baseline between 100–­110 BPM and 160–­ erate risk of fetal hypoxia. Five guidelines (ACOG, DGGG, NICE,
180 BPM to be a Suspicious feature. Only two guidelines (NICE, NICHD, WALES) indicated increased variability above 25 BPM as
SOGC) included a rise in baseline (without breaching the above a Suspicious feature.15,16,18,19,25 All guidelines agreed on absent
18,24
limits) as a Suspicious feature. variability as an Abnormal feature. Decreased variability was
considered prolonged and abnormal if continuing beyond 50 to
90 min depending on the guideline. A sinusoidal pattern was
Variability
identified in all guidelines as a grossly abnormal pattern, but
All guidelines considered a baseline variability of 5–­25 BPM three allowed for up to 30 min of this pattern before meeting
to be Normal. The SOGC guideline24 was the only guideline to Abnormal criteria.7,17,25
TABLE 3 CTG three-­tier interpretation criteria –­moderate probability of fetal acidosis category (Suspicious)

Guideline Label Advice Baseline rate (bpm) Variability (bpm) Decelerations Accelerations
15
ACOG Category II • <110 not accompanied by absent • Minimal baseline variability, • Recurrent variable decelerations Absence of induced
J. Brown et al.

baseline variability, OR • >160 OR accompanied by minimal or moderate accelerations after


• Absent baseline variability baseline variability, OR fetal stimulation
with no recurrent • Prolonged deceleration more than
decelerations, OR 2 min but less than10 min, OR
• >25 baseline variability • Variable decelerations with other
characteristics, such as slow return to
baseline, ‘overshoots’, or ‘shoulders’
DGGG16 Suspicious At least one criteria • 100–­109, OR • <5 and >40 min but <90 min, • Early decelerations, OR Periodical occurrence
but all others Normal • 161–­180 without simultaneous OR • Variable decelerations, OR with every contraction
accelerations • >25 • Prolonged decelerations (abrupt drop
below baseline for at least 60–­90 sec
but <3 min).
FIGO4 Suspicious Lacking at least one
characteristic of
normality, but with no
pathological features
IRELAND17 Suspicious Lacking at least one
characteristic of
normality, but with no
pathological features
NICE18 Amber Only one non-­ • 100–­109, OR • Less than five for 30–­50 min, • Variable decelerations with no
reassuring feature • 161–­180 OR concerning characteristics for ≥90 min,
AND two reassuring OR • More than 25 for 15–­25 min OR
features • an increase in baseline fetal heart • Variable decelerations with any
rate of 20 bpm or more from the concerning characteristics in ≤50% of
start of labour or since the last contractions for ≥30 min, OR
review an hour ago • Variable decelerations with any
concerning characteristics in >50% of
contractions for <30 min, OR
• Late decelerations in >50% of
contractions for <30 min, with no
maternal or fetal clinical risk factors
such as vaginal bleeding or significant
meconium.
NICHD19 Category II • <110 not accompanied by absent • Minimal baseline variability, • Recurrent variable decelerations Absence of induced
baseline variability, OR • >160 OR accompanied by minimal or moderate accelerations after
• Absent baseline variability baseline variability, OR fetal stimulation
with no recurrent • Prolonged deceleration more than
decelerations, OR 2 min but less than10 min, OR
• >25 baseline variability • Variable decelerations with other
characteristics, such as slow return to
baseline, ‘overshoots’ or ‘shoulders’
283

(Continued)

1479828x, 2023, 3, Downloaded from https://obgyn.onlinelibrary.wiley.com/doi/10.1111/ajo.13667 by Universitat Autonoma De Barcelona, Wiley Online Library on [18/07/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
1479828x, 2023, 3, Downloaded from https://obgyn.onlinelibrary.wiley.com/doi/10.1111/ajo.13667 by Universitat Autonoma De Barcelona, Wiley Online Library on [18/07/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
284 Review of CTG guidelines

Accelerations

• ≤2 accelerations with
Uncertain significance

15 sec in 40–­80 min.


acme of ≥15, lasting
There was considerable variation in the role of fetal heart rate ac-
Accelerations

celerations in intrapartum CTG interpretation. Three guidelines


(DGGG, SLCOG, SOGC) describe accelerations as a necessary fea-
ture of a Normal CTG,16,23,24 whereas the other guidelines deem
that accelerations can be present or absent and/or have an uncer-
tain significance in the interpretation of fetal compromise. This is

shaped late decelerations for <30 min,


• V-­shaped variable decelerations with

• Repetitive (>50% of contractions) U-­


>50% of contractions for >90 min, OR
carried over into the Suspicious and Abnormal categories where

• Single deceleration lasting 3–­5 min.


• Variable decelerations of 30–­60 sec

variable decelerations with >50% of


only the SOGC guideline describe absent accelerations as a fea-
• Repetitive variable >1 min with
normal baseline and variability.
• Typical variable with >50% of

ture of an Abnormal CTG.24

contractions for <30 min, OR


• Non V-­shaped (U-­shaped)
contractions for >90 min.

Decelerations
Decelerations

Both the description and relative risk stratification of decelera-


tions varied dramatically between guidelines. All guidelines agree
duration.

that absence of decelerations was a Normal CTG feature—­and


two (DGGG, SLCOG) indicated that presence of any decelerations
OR

was a Suspicious or Abnormal finding.16,23 That is, only a CTG with


no decelerations could be considered in the Normal category.
• ≤5 (absent or minimal) for

Two guidelines (FIGO, IRELAND) permitted no repetitive decelera-


tions (<50% of contractions).4,17 Five (ACOG, NICE, NICHD, SWE-­17,
• <5 for 30–­50 min, OR
• ≥25 for 15–­25 min.

WALES) confirmed typical early decelerations also in the Normal


• <5 for 40–­90 min
Variability (bpm)

category,7,15,18,19,25 while one (DGGG) placed early decelerations in


the Suspicious category.16 However, other (NICE, SWE-­17, WALES),
40–­80 min.

guidelines also permitted limited typical variable decelerations


as Normal.7,18,25
Note: Grey cells indicate that this feature was not considered in the guideline for this risk category.

Decelerations qualifying as Suspicious were discrepant be-


tween the guidelines. Some guidelines (NICE, SWE-­17, WALES)
considered recurrent typical variable decelerations Normal if
present for less than 90 min and up to 60 sec in duration,7,18,25
• 100–­109 for >10 min, OR

while others (ACOG, SWE-­17) made explicit that all other features
• >160 for <30 min, OR

(such as variability and baseline) must be in the range expected


Baseline rate (bpm)

• >160 for >10 min.

of the Normal category.7,15 There was more consistency in the


• Rising baseline
• 100–­109, OR

• 100–­110, OR

• 100–­109, OR

Abnormal criteria for decelerations, with unanimous classification


of late decelerations and bradycardia (>5 min) as convincing evi-
• 161–­180

• >160

dence of fetal compromise.


Importantly, despite impacting which risk category the CTG
would be interpreted as, the description of deceleration mor-
phology was inconsistent across guidelines. For example, the
presence of recurrent variable decelerations are considered
Abnormal in ACOG15 and NICHD19 guidelines, but the SOGC
guideline24 classifies variable decelerations on their length
Advice

(>60 sec) rather than their frequency. The Wales PROMPT25


guideline is the most descriptive of deceleration morphology,
classifying decelerations as ‘V-­shaped’ or ‘U-­shaped’ and using
Suspicious
reassuring

reassuring

frequency, recurrence and variability to further clarify which


TABLE 3 (Continued)

Atypical

risk category to classify the CTG.


Label

Non-­

Non-­

Five-­category guidelines
Guideline

SLCOG23

WALES25
SWE-­177
SOGC24

Although the vast majority of guidelines stratify CTG patterns into


one of three categories as described above, two guidelines differ
TABLE 4 CTG three-­tier interpretation criteria—­high probability of fetal acidosis category (Abnormal)

Guideline Label Advice Baseline rate (bpm) Variability (bpm) Decelerations Accelerations
15
ACOG Category III Absent baseline • Bradycardia • Absent baseline variability, OR • Recurrent variable
J. Brown et al.

variability and any • Sinusoidal pattern decelerations, OR


of the following • Recurrent late decelerations
features.
DGGG16 Pathological At least one feature • <100, OR • <5 for >90 min, OR • Atypical variable decelerations, • Absent for >40 min—­
present. • >180 • Sinusoidal pattern OR unclear significance
• Late decelerations, OR
• Isolated prolonged deceleration
for >3 min
FIGO4 Pathological • <100 • Reduced variability, OR • Repetitive late or prolonged
• Increased variability, OR (>3 min) decelerations for
• Sinusoidal pattern >30 min or > 20 min if reduced
variability, OR
• Single prolonged deceleration
with >5 min
IRELAND17 Pathological • <100 for >10 min • Reduced variability <5 for >50 min, • Repetitive late or prolonged
OR (>3 min) decelerations for
• Increased variability >25 for >30 min or >20 min if reduced
>50 min, OR variability, OR
• Sinusoidal pattern for >30 min. • Single prolonged deceleration
with >5 min
NICE18 Red • <100, OR • Reduced variability <5 for >50 min, • Variable decelerations with any
• >180 OR concerning characteristics in
• Increased variability >25 for >50% of contractions for >30 min
>50 min, OR (or less if any maternal or fetal
• Sinusoidal pattern clinical risk factors), OR
• Late decelerations for >30 min
(or less if any maternal or fetal
clinical risk factors), OR
• Acute bradycardia, OR
• Single prolonged deceleration
lasting 3 min or more
NICHD19 Category III Absent baseline • Bradycardia • Absent baseline variability, OR • Recurrent variable
variability and any • Sinusoidal pattern decelerations, OR
of the following • Recurrent late decelerations
features.
SLCOG23 Abnormal • <100, OR • <5 >90 min, OR • Repetitive late or atypical
• >180 • Sinusoidal pattern decelerations for >3 min
SOGC24 Atypical • <100, OR • ≤5 for ≥80 min, OR • Variable decelerations of • ≤2 accelerations with
• >160 for >30 min, OR • ≥25 for >10 min, OR >60 sec duration OR acme of ≥15, lasting
• Erratic baseline • Sinusoidal pattern • Late decelerations 15 sec in >80 min.

(Continued)
285

1479828x, 2023, 3, Downloaded from https://obgyn.onlinelibrary.wiley.com/doi/10.1111/ajo.13667 by Universitat Autonoma De Barcelona, Wiley Online Library on [18/07/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
286

TABLE 4 (Continued)

Guideline Label Advice Baseline rate (bpm) Variability (bpm) Decelerations Accelerations
7
SWE-­17 Pathological • <100 • <2, OR • Repetitive late decelerations for
• <5 for >60 min, OR >30 min, OR
• >25 for >30 min, OR • Repetitive late decelerations for
• Sinusoidal pattern for >30 min >20 min with tachycardia and/or
decreased variability, OR
• Repetitive variable
decelerations lasting >1 min with
tachycardia and/or decreased
variability for >20 min, OR
• Repetitive prolonged
decelerations lasting >3 min, OR
• Single prolonged deceleration
lasting >5 min
WALES25 Pathological • <100, OR • <5 for >50 min, OR • Non V-­shaped (U-­shaped)
• >180 • >25 for >25 min, OR variable decelerations with >50%
• Sinusoidal pattern for >30 min of contractions for >30 min, OR
• Repetitive (>50% of
contractions) U-­shaped late
decelerations for >30 min, OR
• Repetitive (>50% of
contractions) U-­shaped late
decelerations and reduced
variability for >20 min, OR
• Single deceleration lasting
>5 min.

Note: Grey cells indicate that this feature was not considered in the guideline for this risk category.
Review of CTG guidelines

1479828x, 2023, 3, Downloaded from https://obgyn.onlinelibrary.wiley.com/doi/10.1111/ajo.13667 by Universitat Autonoma De Barcelona, Wiley Online Library on [18/07/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
1479828x, 2023, 3, Downloaded from https://obgyn.onlinelibrary.wiley.com/doi/10.1111/ajo.13667 by Universitat Autonoma De Barcelona, Wiley Online Library on [18/07/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
J. Brown et al. 287

from the rest and use a distinct five-­table, five-­category classifi- shape, while others highlighted depth and duration as important
20,21
cation system. PCJSOG and Parer et al. describe five discrete features. Signs of immediate and profound fetal hypoxia, like bra-
cross-­tables with the deceleration types on the x-­axis and fetal dycardia and sinusoidal patterns, were much more likely to be
heart rate baseline on the y-­axis. Clinicians choose which of the consistent across guidelines.
five tables to use depending on the baseline heart rate variability Inter-­guideline variability has crucial ramifications on ac-
of the CTG for interpretation. curacy, safety and feasibility of CTG interpretation. Guidelines,
Decelerations are classified as early, variable (mild), variable although superficially may be seen as inert documents, they influ-
(severe), late (mild), late (severe), prolonged (mild), prolonged ence a myriad of consequential actions in delivery wards around
(severe). The guideline defines the severe category of the three the world, on a daily basis. With such disparity in CTG interpreta-
relevant decelerations as below, and if it does not meet this, it is tion, women are being treated differently depending on location
considered mild. of birth, rather than what the best evidence dictates.
With limited data supporting the included guidelines, the vari-
• Late—­decline in heart rate of more than 15 bpm from the ation stems from permutations of consensus that have evolved
baseline to the lowest point. over time to improve specificity and/or sensitivity. The permissi-
• Variable—­when the lowest point is less than 70 bpm and con- bility of such disparity reveals an underlying inaccuracy in all CTG
tinues for more than 30 sec, or when the lowest point is more interpretations, despite the gravity of decisions it informs. There
than 70 bpm and less than 80 bpm and continues for more are few studies comparing inter-­guideline accuracy,7 and with or-
than 60 sec. ganic evolution of guidelines, we need such research to under-
• Prolonged—­Prolonged deceleration is when the lowest point is stand what constitutes actual best practice.
less than 80 bpm. In a broader context, guideline inconsistency has significant
medico-­legal ramifications. The high false-­positive rate of CTG in-
Once all three variables are considered and cross-­tabulated, terpretation allows attribution of intrapartum decisions to grave
the CTG will be classified as one of five colour-­coded ‘variant adverse outcomes, when a similar CTG pattern may not have re-
patterns’ (Normal, Benign, Mild, Moderate, Severe). Each variant sulted in a similar management decision or outcome at another
pattern aligns with a recommended set of actions depending location, including within the same city or country. Guideline vari-
on the severity. For example, the expected action for a category ation codifies otherwise subjective interpretation and may permit
five CTG is ‘conduct caesarean section or operative vaginal de- a legal argument that a specific pattern was considered more
livery, prepare for resuscitation of newborn’, with no alternative severe under the rules of a different guideline, even if not used
option described. in that unit. This may not even be with sinister intentions, with
expert witnesses practising in different locations accustomed to
different guidelines and CTG interpretation. It is feasible to as-
DISCUSSION sume that CTG interpretation guideline inconsistency has been in-
strumental in court decisions that had grave consequences to an
This work gives a comprehensive overview of clinical practice obstetrician's livelihood and wellbeing—­while those practitioners
guidelines for the interpretation of intrapartum cardiotocography were behaving in accordance with their preferred interpretation
and highlights their similarities and inconsistencies. Overall, the algorithm and consequent clinical management.
guidelines were high quality as assessed by the AGREE II tool. The ultimate challenge is balancing sensitivity and specific-
Despite difference in labelling, most guidelines were struc- ity, where over-­reaction to a false-­positive may itself result in
tured into a three-­tier risk stratification of Normal (low), Suspicious unnecessary adverse outcome. CTG interpretation guidelines
(moderate) and Pathological (high). However, significant variation are caught between two opposing forces that seem to sustain
in risk stratification of key CTG features was identified, suggest- and inform the differences. On one hand, more CTG features
ing critical differences in directing clinical practice depending on can be defined as Suspicious or Abnormal to improve the sen-
which guideline a facility or practitioner chooses to use. sitivity of the guideline—­while on the other hand, including too
Difference in clinical importance of specific features occurred many will result in unnecessary intervention and possibly inflate
across all risk categories, including in the Normal category. For ex- the number of unnecessary surgical births, both of which are
ample, some guidelines required accelerations to be present for a not without ramifications. Conversely, considering too few fea-
CTG to be considered Normal, whereas most guidelines thought tures as Suspicious or Abnormal will miss instances of critical
accelerations were a positive feature if present but of unclear clin- fetal acidosis.26 Efforts to therefore introduce more nuance to
ical significance if absent. guidelines impair the effectiveness of the interpretation tool it-
The Suspicious category was the most heterogenous across self. While there is some evidence that the five-­tier system is
the guidelines, with major differences in expectations for variabil- more sensitive than a three-­tier system,27 the intricacy of such
ity and deceleration patterns. Deceleration morphology itself was guidelines may make CTG interpretation a laborious, analyti-
varied across guidelines, with some guidelines prescriptive about cal task that is likely to be unwelcome and impractical in busy
1479828x, 2023, 3, Downloaded from https://obgyn.onlinelibrary.wiley.com/doi/10.1111/ajo.13667 by Universitat Autonoma De Barcelona, Wiley Online Library on [18/07/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
288 Review of CTG guidelines

birth units. Some guidelines19 now recommend stimulation to All the included guidelines are an attempt to make the best
induce fetal acceleration to further clarify fetal reserves, which use of the available evidence for the best outcomes; however,
acknowledges normally functioning homeostatic mechanisms the lack of evidence remains the rate-­limiting step for improve-
that may temporarily meet abnormal CTG criteria. Beyond ment. The widespread use of CTG criteria is a paradoxically self-­
this, diversity in CTG interpretation, and subsequent decision-­ defeating cycle. To improve criteria, we must first understand the
making, undermine reproducibility of obstetric research, accu- efficacy of the CTG itself for detecting fetal acidosis. To use the
racy of perinatal epidemiology, and staff skill transition between CTG for this purpose, we need guidelines for interpretation of the
units. In essence, it appears we are all speaking the same lan- patterns. However, the guidelines themselves are varied and lack
guage, when we are not. robust evidence. Therefore, it becomes impossible to separate
In practice, these findings position CTG interpretation guide- the efficacy of a guideline from the technology. Even comparison
lines in a new context. Over time, such guidelines have shifted between criteria is skewed by the denominator of the technolo-
from descriptive tools to assist clinician triage, into prescriptive gy's limitations. Moreover, the technology is now so entrenched
action plans based on rigid stratifications—­often mandating that it is ethically irresponsible to withhold or modify its use for
immediate delivery by quickest mode. However, while there research. With these key limitations in mind, the first goal toward
are universally accepted features of Normal, Suspicious and improvement is comparable consistency. That is, we must agree
Pathological CTG features, the differences between the guide- to play from the same rulebook. There are not enough good data
lines allow for the return of clinician interpretation of CTG pat- to justify so many different guidelines being used. This can be
terns in the specific context of their patient. In other words, it achieved two ways. First, disassemble existing guidelines into the
is difficult to justify mandatory action on a CTG pattern that specific patterns they describe and analyse those patterns against
is considered less severe by an equally reputable guideline. neonatal outcomes—­after which, the patterns are recombined ac-
Applicability is further hampered by differences in risk toler- cording to the findings. Or second, a single consensus guideline is
ance between countries and institutions. For example, expecta- created that can be tested and improved by adding, subtracting
tion of maximal sensitivity (at the risk of more false positives) in or modifying key patterns described in research or smaller guide-
a certain context will produce a different pattern of behaviour lines. Until consistency is improved, innovation in CTG interpreta-
in response to an identical CTG pattern in an institution with tion will be hamstrung by subjective variety.
a higher risk tolerance. It must always be acknowledged that
CTG interpretation remains a subjective science, at least for the
time being. CONCLUSION
The consequences of this inter-­interpretation inconsistency
cannot be overstated. There are few other areas of widely prac- This is the first study to identify and evaluate clinical practice
tised, high-­stakes, clinical care where there are well-­documented guidelines relevant to intrapartum CTG interpretation. The meth-
differences in key decision-­making tools. Consider the ramifica- odological quality of guidelines was consistent and generally
tions of 13 accepted guidelines defining different parameters to good; however, most recommendations were based on consen-
diagnose sepsis. Some hospitals would miss fatal cases, while oth- sus. There were significant differences in the features nominated
ers would over-­prescribe antibiotics and contribute to antimicro- as Normal, Suspicious and Abnormal. This inconsistency is likely
bial resistance. This review demonstrates that something similar to have ramifications on sensitivity and specify in detecting fetal
is happening in the interpretation of CTGs—­in some centres criti- acidosis, and almost certainly results in unnecessary or delayed
cal fetal distress might be missed, while others routinely perform intervention. Improving consistency across guidelines is an impor-
unnecessary caesarean sections. Efforts toward evidence-­based tant step toward establishing a global gold standard and better
guideline uniformity are critical for future safety. research data in this field, which is necessary to further improve
The strengths of this systematic review include its large num- the efficacy of intrapartum CTG monitoring.
ber of included guidelines, quality of the guidelines, and rigorous
analysis. Limitations include language and availability. This study
A U T H O R C O NT R IB U T IO NS
only includes guidelines in the English language, and it is likely
there are guidelines available in non-­English languages that have JB and MZ conceived and designed the study. JB and DK per-
not been included. As practical tools foremost, guidelines are formed the literature search, data extraction and guideline
more likely to be in native languages than typical peer-­reviewed quality appraisal. JB, DK and MZ synthesised the guideline rec-
research. With regard to availability, it can be anticipated that ommendations. All authors contributed to drafting and editing
many guidelines are used locally at a district or facility-­level and the manuscript.
are not uploaded for public availability on the internet. Therefore,
it is possible that this study underestimates the number of avail-
F U ND ING INF O R M A T IO N
able guidelines; however, it is likely that these guidelines are
based on the foundational ones included in this study. No funding was received for this systematic review.
1479828x, 2023, 3, Downloaded from https://obgyn.onlinelibrary.wiley.com/doi/10.1111/ajo.13667 by Universitat Autonoma De Barcelona, Wiley Online Library on [18/07/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
J. Brown et al. 289

SY STEM ATIC REVI E W R E G I S TR ATI O N registry No. 015/036. Geburtshilfe Frauenheilkd 2014; 74(8):
721–­7 32.
PROSPERO, CRD42018085085. 17. National Women's and infants health Programme. National
Clinical Guideline for intrapartum fetal heart rate monitoring: Ireland.
Ireland: National Women's and infants health Programme, 2019.
REFERENCES 18. O'Heney J, McAllister S, Maresh M, Blott M. Fetal monitoring in
labour: Summary and update of NICE guidance. BMJ 2022; 379:
1. Trochez RD, Sibanda T, Sharma R, Draycott T. Fetal monitoring
o2854.
in labor: Are accelerations good enough? J Matern Fetal Neonatal
19. Macones GA, Hankins GD, Spong CY et al. The 2008 national insti-
Med 2005; 18(5): 349–­352.
tute of child health and human development workshop report on
2. Alfirevic Z, Gyte GM, Cuthbert A, Devane D. Continuous cardioto-
electronic fetal monitoring: Update on definitions, interpretation,
cography (CTG) as a form of electronic fetal monitoring (EFM) for
and research guidelines. Obstet Gynecol 2008; 112(3): 661–­666.
fetal assessment during labour. Cochrane Database Syst Rev 2017;
20. Parer JT, Ikeda T. A framework for standardized management of
2: CD006066.
intrapartum fetal heart rate patterns. Am J Obstet Gynecol 2007;
3. Jia Y-­J, Chen X, Cui H-­Y et al. Physiological CTG interpretation: The
197(1): 26.
significance of baseline fetal heart rate changes after the onset of
21. Okai T, Ikeda T, Kawarabayashi T et al. Intrapartum management
decelerations and associated perinatal outcomes. J Matern Fetal
guidelines based on fetal heart rate pattern classification. J Obstet
Neonatal Med 2021; 34(14): 2349–­2354.
Gynaecol Res 2010; 36(5): 925–­928.
4. Ayres-­de-­Campos D, Spong CY, Chandraharan E, FIFMEC P.
22. The Royal Australian and New Zealand College of Obstetricians
FIGO consensus guidelines on intrapartum fetal monitoring:
and Gynaecologists. Intrapartum Fetal Surveillance. Clinical
Cardiotocography. Int J Gynecol Obstet 2015; 131(1): 13–­24.
guideline—­Fourth edition 2019. Australia: The Royal Australian and
5. Melman S, Schoorel ENC, Dirksen C et al. SIMPLE: Implementation
New Zealand College of Obstetricians and Gynaecologists, 2019.
of recommendations from international evidence-­based guide-
23. Editor T. Guideline on fetal monitoring in labour. Sri Lanka J Obstet
lines on caesarean sections in The Netherlands. Protocol for a
Gynaecol 2014; 35(4): 140–­141.
controlled before and after study. Implement Sci 2013; 8(1): 3.
24. Liston R, Sawchuck D, Young D. No. 197a-­fetal health surveil-
6. Lotgering FK, Wallenburg HCS, Schouten HJA. Interobserver and
lance: Antepartum consensus guideline. J Obstet Gynaecol Can
intraobserver variation in the assessment of antepartum cardio-
2018; 40(4): e251–­e271.
tocograms. Am J Obstet Gynecol 1982; 144(6): 701–­705.
25. Welsh Government. Implementation of all Wales Intrapartum
7. Ekengård F, Cardell M, Herbst A. Impaired validity of the new
Fetal Surveillance Standards for Maternity Services. Wales: Welsh
FIGO and Swedish CTG classification templates to identify fetal
Government, 2019.
acidosis in the first stage of labor. J Matern Fetal Neonatal Med
26. Ekengård F, Cardell M, Herbst A. Low sensitivity of the new FIGO
2021; 1–­8: 4853–­4860.
classification system for electronic fetal monitoring to identify
8. Rooth G, Huch A, Huch R. Guidelines for the use of fetal monitor-
fetal acidosis in the second stage of labor. Eur J Obstet Gynecol
ing. FIGO news. Int J Gynecol Obstet 1987; 25: 159–­167.
Reprod Biol X 2021; 9: 100120.
9. Chudáček V, Spilka J, Burša M et al. Open access intrapartum CTG
27. Coletta J, Murphy E, Rubeo Z, Gyamfi-­Bannerman C. The 5-­tier
database. BMC Pregnancy Childbirth 2014; 14(1): 16.
system of assessing fetal heart rate tracings is superior to the 3-­
10. Ayres-­de-­Campos D, Bernardes J, Costa-­Pereira A, Pereira-­Leite
tier system in identifying fetal acidemia. Am J Obstet Gynecol 2012;
L. Inconsistencies in classification by experts of cardiotocograms
206(3): 226.
and subsequent clinical decision. BJOG 1999; 106(12): 1307–­1310.
11. Paneth N, Bommarito M, Stricker J. Electronic fetal monitoring
and later outcome. Clin Invest Med 1993; 16(2): 159–­165.
12. Blackwell SC, Grobman WA, Antoniewicz L et al. Interobserver and
intraobserver reliability of the NICHD 3-­tier fetal heart rate inter- SUPPORTING INFORMATION
pretation system. Am J Obstet Gynecol 2011; 205(4): 378.
13. Spencer JA. Clinical overview of cardiotocography. BJOG 1993; Additional supporting information may be found online in the
100: 4–­7.
Supporting Information section at the end of the article.
14. AGREE Next Steps Consortium. The AGREE II Instrument (Electronic
Version), 2017. [Accessed 4 April 2021]. https://www.agree​trust.
org/ Appendix S1. Website searches of guidelines.
15. American College of Obstetricians and Gynecologists. Practice Appendix S2. Database searches of guidelines (30 June 2021).
bulletin no. 116: Management of intrapartum fetal heart rate
Appendix S3. Excluded articles and reason for exclusion.
tracings. Obstet Gynecol 2010; 116(5): 1232–­1240.
16. Maternal Fetal Medicine Study Group. S1-­g uideline on the use Appendix S4. AGREE II overall scores of the included guidelines.
of CTG during pregnancy and labor: Long version—­A WMF Appendix S5. AGREE II domain scores of the included guidelines.

You might also like