You are on page 1of 4

Service Manual

custo spiro mobile

Required tools:
- Screw driver

Required spare parts:


128201 2.5 m USB connection cable
128203 3 m USB connection cable
128205 5 m USB connection cable

Additional available spare parts:


11031 measuring adapter for spiro mobile

Procedural method:
1. Detach measuring adapter

SHB-0053 / DK-1193 / V002 / 23.05.2012


Service Manual

custo spiro mobile

2. Lift the screw cover with a small screw driver


Pay attention that you don’t damage the shell.

SHB-0053 / DK-1193 / V002 / 23.05.2012


Service Manual

custo spiro mobile

3. Remove O-Ring, unscrew both shell screws

SHB-0053 / DK-1193 / V002 / 23.05.2012


Service Manual

custo spiro mobile

4. Open the device

• Note the colours of the single wires


• Loose the strain relief of the USB cable and unsolder the wires.
• Replace the USB cable, solder the single wires,
bolt together the strain relief
• Put the board back into the device, bolt the shell screws,
attach O-Ring and screw cover
• Attach the measuring adapter
• Connect custo spiro mobile to the PC and perform a test
• Remove the sticker with serial number from old USB cable
and attach it to the new one.

SHB-0053 / DK-1193 / V002 / 23.05.2012