Professional Documents
Culture Documents
2011
°C
%RH
m/s
hPa
CO2
Lux
CO
aW
rpm
mA
V
02-098X-705X-Measurement 13.12.2010 11:06 Seite 2
Probe selection
Measurement and application ranges of flow probes
The flow measurement range 0 to 100 m/s can be divided into three sections:
– Low-speed velocity 0 to 5 m/s
100
– Mid-speed velocity 5 to 40 m/s Thermal Vane measurement Vane/temperature Pitot Pitot
probe probes to +60 °C measurement tube to tube
– High-speed velocity 40 to 100 m/s. to +70 °C probes +600 °C NiCr-Ni
Thermal probes are used for accurate measurements in the range 0 to 5 m/s. Vane -40°C
60 +600 °C
probes are ideal for velocities from 5 to 40 m/s. The measuring range of the Pitot to to
tube depends on the differential pressure probe used. The new 100 Pa probe can +140 °C +350 °C
therefore be used for the exact measurement of flow speed from approx. 1 m/s to
12 m/s. The Pitot tube yields optimum results in the higher velocity range. An 40
additional criterion when selecting the right velocity probe is the temperature.
Thermal sensors can normally be used at up to approx. +70 °C. Special design
vane probes can be used to maximum +350 °C. Pitot tubes are used for
20
temperatures above +350 °C.
15
10
The air vent greatly changes the relatively The vane probe is set exactly if the flow direction
uniform flow inside the duct. Areas of ar is parallel to the vane axis. If the measuring probe
higher flow velocity are created at the Lamin is turned slightly in the air current, the value
flow
free vent surfaces and areas of low in the e of the shown in the instrument changes. The measuring
flow velocity and swirl at the grids. centr t probe is positioned exactly in the air current if the
duc
The flow profile steadies at a value shown is at max.
distance from the the grid When measuring in a duct there should also be a
depending on the grid design but is minimum of ten diameters of straight run before
usually 20 cm. For best accuracy, a Max. values the measuring spot and four diameters of straight
large diameter vane is recommended. run after the spot for best results. By design,
Min. values
The area of the vane helps to get an vanes are less influenced by turbulence than thermal probes or Pitot tubes.
Mean values
average reading of the turbulent flow
from the grid.
Site selection
Direction of flow
D
Measurements at suction apertures using a volume flow funnel
10 x D 4xD
Even without the disturbing effects of a grid in an aperture, the lines of flow are not
directional and the flow profile is irregular. Because a partial vacuum in the duct You should measure in a straight part of the duct, if possible. The duct part should
draws air out of the room in a funnel shape even a short distance from the aperture, have a minimum of ten diameters of straight run before the measuring spot and
there is no defined area in the room over which a measurement could be made. four diameters of straight run after the measuring spot. The flow profile should not
Therefore, only the duct or funnel measurement yields reproducible results. be interrupted in any way by flaps, dips, angles etc.
Measuring funnels of various sizes are available for such applications.
These create defined flow conditions at a known distance from the grid with a fixed
volume. A velocity probe is positioned centrally and secured. The volume flow is
calculated from the velocity multiplied by the funnel factor (e.g. funnel factor 22). Flow measurement in ducts
Measuring ambient air velocity using testo 400 in accordance with DIN 1946 Part 2, ANSI/Ashrae 55-1992
Ambient air velocity is a very important parameter in the thermal comfort of people Ambient air velocity
in rooms. testo 400 supplies the current and mean air velocities. The maximum Air temperature in °F
permissible mean air velocity depends on the air temperature measured by testo 68 70 72 74 76 78 80
1,7 140
Mean air velocity m/s
400 and the amount of turbulence calculated from the air velocity. The example
shows a permissible mean air velocity of 0.26 m/s with an air temperature 1,6 120
0,4 80
Mean air velocity fpm
10%
0,3 60
20%
40%
0,2 40
60%
0,1 20
0 0
20,0 21,1 22,2 23,3 24,4 25,5 26,7
Air temperature in °C
2 Degree of turbulence T
03-098X-705X-Contents 13.12.2010 11:07 Seite 3
Contents
Measuring instruments
Practical multi-function measuring instrument
testo 435-1/-2/-3/-4 All-rounder for ventilation and indoor air quality Page 4
Measurement systems
testo 445 Service instrument for ventilation/air conditioning systems Page 8
testo 400 The reference measuring instrument for A/C and ventilation systems Page 13
Accessories
Software and Accessories Page
ComSoft 3 - Professional Professional Software including Data Filing 24
Printer Page
Testo fast printer Versatile infrared printer for testo 435, 445, 400 26
Ethernet adapter Page
Ethernet adapter Access Ethernet with Testo measuring instruments 27
3
04-098X-705X-435-1-2-3-4 13.12.2010 11:07 Seite 4
testo 435 All-round talent for ventilation and Indoor Air Quality
All measurement parameters for Common product advantages testo 435
air conditioning
Wide selection of probes (optional):
The testo 435 provides the
- IAQ probe for evaluating the indoor air quality via
possibility of analysing the indoor
CO2, air temperature, indoor air humidity and
air. On the one hand, this serves absolute pressure
as an indicator for the well-being
- Thermal probe with integrated temperature and air
of people at their workplaces, and
humidity measurement
on the other hand as an important
and deciding factor in storage - Vane and hot wire probes
and production processes. - Radio probes for temperature
Easy operation with user profiles
In addition to this, the Indoor Air
Quality signals whether the air Printout on the testo fast printer (optional)
conditioning system (HVAC) is
working with as much energy
economy as possible, or whether
it needs to be adjusted with the
help of testo 435.
Further product advantages of the variants
The parameters CO2 , relative
humidity and room temperature Integrated differential pressure measurement (435-
are available for evaluating the 3/-4, not retrofittable)
quality of the air. Absolute - for flow measurement
pressure, draught, Lux, U-value
- for monitoring filters
and surface temperature can
additionally be determined. In Extended instrument function (435-2/-4, not
retrofittable)
order to determine the volume
flow, all the possibilities of flow - Instrument store for 10,000 readings
velocity measurement are - PC software for analysing, archiving and
available, such as thermal probes, documenting measurement data
vane anemometers and Pitot - Humidity probes with radio or wire (optional)
tubes. - Lux probe connection possible
- Comfort level probe connection possible
- Possibility of connecting U-value probe
Versatility with wireless probes
In addition to classical probes on
wires, a wireless measurement up
to a distance of 20 m (without
obstruction) is possible. Damage
to the wire or hindrances in usage
are thus eliminated. A maximum
of three wireless probes can be
recorded and displayed with testo
435. The wireless probes are for
the measurement parameters
temperature and, depending on
the instrument type, humidity. The
optional, easily plugged-in radio
module can be retrofitted at any
time.
435-1/-2/-3/-4 Probes
435-1/-2/-3/-4
IAQ probes Illustration Meas. range Accuracy Part no.
IAQ probe to assess Indoor Air Quality, CO2, 0 to +50 °C ±0.3 °C 0632 1535
0 to +100 %RH ±2 %RH (+2 to +98 %RH)
humidity, temperature and absolute pressure 0 to +10000 ppm CO2 ±(50 ppm CO2 ±2% of mv) (0 to +5000 ppm CO2)
measurement, with desk-top stand ±(100 ppm CO2 ±3% of mv) (+5001 to +10000 ppm CO2)
+600 to +1150 hPa ±3 hPa
Ambient CO probe, for detecting CO in buildings 0 to +500 ppm CO ±5% of mv (+100.1 to +500 ppm CO) 0632 1235
and rooms ±5 ppm CO (0 to +100 ppm CO)
Funnel set consisting of funnel for disc outlets (Ø 200) and funnel 0563 4170
for ventilator (330 x 330 mm) for in- and outgoing air
Fast-action surface probe with sprung thermocouple 115 mm -60 to +300 °C Class 2* 3s 0602 0393
strip, also for uneven surfaces, measurement range
Ø 5 mm
short-term to +500°C, TC Type K
Conn.: Fixed cable Ø 12 mm
Pipe wrap probe for pipe diameter 5 to 65 mm, -60 to +130 °C Class 2* 5s 0602 4592
with exchangeable measuring head. Meas. range
short-term to +280°C, TC Type K Conn.: Fixed cable
Clamp probe for measurements on pipes, pipe -50 to +100 °C Class 2* 5s 0602 4692
diameter 15 to 25 mm (max. 1"), meas. range
short-term up to +130°C, TC Type K Conn.: Fixed cable
Immers./penetr. probes Illustration Meas. range Accuracy t99 Part no.
Waterproof immersion/penetration probe, TC Type K 114 mm 50 mm -60 to +400 °C Class 2* 7s 0602 1293
Ø 5 mm Ø 3.7 mm
Conn.: Fixed cable 1.2 m
435-2/-4
IAQ probes Illustration Meas. range Accuracy Part no.
Comfort level probe for degree of turbulence 0 to +50 °C ±0.3 °C 0628 0109
0 to +5 m/s ±(0.03 m/s +4% of mv)
measurement with telescopic handle (max. 820
mm) and stand, meets EN 13779 requirements max. 820 mm
Lux probe, for measuring light intensity Accuracy to DIN 5032, Part 6: 0635 0545
f1 = 6% = V(Lambda) adjustment
f2 = 5% = cos-like weighting, Class C
* According to standard EN 60584-2, the accuracy of Class 2 refers to -40 to +1200 °C.
** when used with an NTC or wireless humidity probe for measuring outside temperature and 20 K difference between the air inside and outside
Pitot tube, 500 mm long, stainless steel, for 0 to +600 °C 0635 2045
measuring flow velocity 500 mm Ø 7 mm
Pitot tube, 1000 mm long, stainless steel, for 0 to +600 °C 0635 2345
measuring flow velocity 1000 mm Ø 7 mm
Technical data
Probe type NTC Type K Type T Testo humid. sensor, Vane Hot wire Absolute CO2 (IAQ probe)
cap. pressure probe
Meas. range -50 to +150 °C -200 to +1370 °C -200 to +400 °C 0 to +100 %RH 0 to +60 m/s 0 to+20 m/s 0 to +2000 hPa 0 to +10000 ppm CO2
Accuracy ±0.2 °C (-25 to +74.9 °C) ±0.3 °C (-60 to +60 ±0.3 °C (-60 to +60 See probe data See probe data See probe See probe data See probe data
±0.4 °C (-50 to -25.1 °C) °C) °C) data
±1 digit ±(0.2 °C +0.3% of ±(0.2 °C +0.3% of mv)
±0.4 °C (+75 to +99.9 °C)
mv) (remaining range) (remaining range)
±0.5% of mv (remaining range)
Resolution 0.1 °C 0.1 °C 0.1 °C 0.1 %RH 0.01 m/s (60 vane) 0.01 m/s 0.1 hPa 1 ppm CO2
0.1 m/s (16 vane)
Technical data 435-2/-4 Technical data 435-3/-4 Oper. temp. -20 to +50 °C
Probe type Lux Probe type Differential pressure Storage temp. -30 to +70 °C
probe, internal Dimensions 220 x 74 x 46 mm
Meas. range 0 to +100000 Lux Meas. range 0 to +25 hPa Battery type Alkali manganese,
Accuracy See probe data Accuracy ±0.02 hPa (0 to +2 hPa) mignon, Type AA
±1 digit ±1 digit 1% of mv (remaining range) Battery life 200 h (typical vane
measurement)
Weight 450 g
Resolution 1 Lux / 0.1 Hz
Overload 200 hPa Material/Housing ABS/TPE/Metal
Resolution 0.01 hPa Warranty 2 years
435-2/-4
Radio probes incl. humidity probe head Meas. range Accuracy Resolution
0 to +100 %RH ±2 %RH (+2 to +98 %RH) 0.1 %RH
Radio handle for attachable probe heads with -20 to +70 °C ±0.3 °C 0.1 °C
humidity probe head
Part no.
0563 4450
Measurement systems
Cable, 1.5 m long, connects probe with plug-in head to meas. instrument, 0430 0143
PUR coating material Probe type Testo humid. sensor, cap. Vane Thermal
Cable, 5 m long, connects probe with plug-in head to measuring instrument 0430 0145
PUR coating material Meas. range 0 to +100 %RH 0 to +60 m/s 0 to +20 m/s
Extension cable, 5 m long, between plug-in head cable and instrument 0409 0063 Accuracy See probe data See probe data See probe data
PUR coating material ±1 digit
Printer and Accessories
Testo fast printer with wireless infrared interface, 1 roll thermal paper and 4 0554 0549
AA batteries
Fast testo 575 printer, incl. 1 roll of thermal paper and batteries 0554 1775 Resolution 0.1 %RH (0 to +100 0.01 m/s (0 to +60 m/s) 0.01 m/s (0 to +10 m/s)
infrared thermal line printer with graphics function %RH) 0.1 m/s (+10.1 to +20
m/s)
External fast charger for 1-4 AA rech. batteries, incl. 4 Ni-MH rech. batteries 0554 0610
with individual cell charging and charge control display, incl. impulse trickle
charging, integrated discharge function, with built-in international mains
Probe type Pressure CO2 probe
plug, 100-240 V, 300 mA, 50/60 Hz
Spare thermal paper for printer (6 rolls) 0554 0569 Meas. range See pressure probes 0 to +1 Vol. % CO2 0 to +10000 ppm CO2
Accuracy ±0.1% of mv See probe data ±(100 ppm CO2 ±3% of mv)
Spare thermal paper for printer (6 rolls) 0554 0568 (+5000 to +10000 ppm CO2)
±1 digit
measurement data documentation legible for up to 10 years ±(500 ppm CO2 ±2% of mv)
(0 to +5000 ppm CO2)
Label thermal paper (Testo patent) for testo 575 printer (6 rolls), can be 0554 0561
applied directly
Software and Accessories Resolution 0.001 hPa (Sonde 0638 0 Vol. % CO2(0 to +1 1 ppm CO2 (0 to +10000
ComSoft 3 - Professional with data management 0554 0830 1345) Vol. % CO2) ppm CO2)
incl. database, analysis and graphics function, data analysis, trend curve 0.001 hPa (Sonde 0638
1445)
RS232 cable 0409 0178 0.01 hPa (Sonde 0638 1545)
connects instrument to PC (1.8 m) for data transfer 1 hPa (Sonde 0638 1645)
Ethernet adapter, RS232 - Ethernet incl. software driver, mains unit 0554 1711
facilitates data communication in network Probe type CO probe
Calibration Certificates
ISO calibration certificate velocity ,hot wire, vane anemometer, Pitot 0520 0004 Meas. range 0 to +500 ppm CO
tube; calibration points 1; 2; 5; 10 m/s Accuracy ±5% of mv (+100 to
ISO calibration certificate/Velocity 0520 0034 ±1 digit +500 ppm CO)
hot wire, vane anemometer, Pitot tube; calibration points 5; 10; 15; 20 m/s ±5 ppm CO (0 to +100
ppm CO)
DAkkS calibration certificate/velocity* 0520 0244
hot wire, vane anemometer; calibration points 0.5; 1; 2; 5; 10 m/s
DAkkS calibration certificate/velocity* 0520 0204 Resolution 1 ppm CO (0 to +500
hot wire, vane anemometer, Pitot tube; calibration points 2; 5; 10; 15; 20 m/s ppm CO)
*Successor organization of the DKD
Vane/temperature probe, Ø 16 mm, attachable to Vane +0.4 to +60 m/s ±(0.2 m/s +1% of mv) 0635 9540
0430 3545 handle or 0430 0941 telescopic Type K (NiCr-Ni) -30 to +140 °C (+0.4 to +40 m/s)
180 mm ±(0.2 m/s +2% of mv)
handle Ø 16 mm (+40.1 to +50 m/s)
Vane/temperature probe, Ø 25 mm, can be Vane +0.4 to +40 m/s ±(0.2 m/s ±1% of mv) 0635 9640
attached to 0430 3545 handle or 0430 0941 Type K (NiCr-Ni) -30 to +140 °C (+0.4 to +40 m/s)
telescopic handle 180 mm
Ø 25 mm
Bendable vane probe (can be bent by 90°), Ø 60 Vane +0.25 to +20 m/s ±(0.1 m/s ±1.5% of mv) 0635 9440
mm, attachable to handle 0430 3545 or telescopic Oper. temp. (+0.25 to +20 m/s)
Ø 60 mm 0 to +60 °C
handle 0430 0941, for meas. on ventilation outlets
Bendable vane probe (can be bent by 90°), Ø 100 mm, Vane +0.1 to +15 m/s ±(0.1 m/s ±1.5% of mv) 0635 9340
attachable to handle 0430 3545 or telescopic handle Oper. temp. (+0.1 to +15 m/s)
Ø 100 mm 0 to +60 °C
0430 0941, for measurements on ventilation outlets
Affordable, robust hot bulb probe, Ø 3 mm, for 150 mm
Hot bulb 0 to +10 m/s ±(0.03 m/s ±5% of mv) 0635 1549
measurements in the lower velocity range, with NTC -20 to +70 °C (0 to +10 m/s)
handle Ø 4 mm Ø 3 mm
Robust hot bulb probe, Ø 3 mm, with handle and Hot bulb 0 to +10 m/s ±(0.03 m/s ±5% of mv) 0635 1049
telescopic handle for measurements in the lower NTC -20 to +70 °C (0 to +10 m/s)
velocity range 850 mm Ø 3 mm
Quick-action hot wire probe, Ø 10 mm, with Hot wire 0 to +20 m/s ±(0.03 m/s ±4% of mv) 0635 1041
telescopic handle, for measurements in the lower NTC -20 to +70 °C (0 to +20 m/s)
velocity range with direction recognition 760 mm Ø 10 mm
Vane probe, Ø 16 mm, with telescopic handle, Vane +0.6 to +40 m/s ±(0.2 m/s ±1.5% of mv) 0628 0005
Tmax +60°C (+0.6 to +40 m/s)
700 mm Ø 16 mm
Vane probe, Ø 60 mm, with telescopic handle, for Vane +0.25 to +20 m/s ±(0.1 m/s ±1.5% of mv) 0635 9449
integrating velocity measurement (+0.25 to +20 m/s)
Ø 60 mm
1100 mm
High temperature vane probe, Ø 25 mm, with Vane +0.6 to +20 m/s ±(0.3 m/s ±1% of fsv) 0635 6045
handle for continuous measurements up to Type K (NiCr-Ni) -40 to +350 °C (+0.6 to +20 m/s)
+350°C 560 mm Ø 25 mm
Precision pressure probe, 100 Pa, measures Differential pressure 0 to +100 Pa ±(0.3 Pa ±0.5% of mv) 0638 1345
differential pressure and velocities (in probe (0 to +100 Pa)
combination with Pitot tube)
Pressure probe, 10 hPa, measures differential Differential pressure 0 to +10 hPa ±0.03 hPa (0 to +10 hPa) 0638 1445
pressure and velocities (in combination with Pitot probe
tube)
Measurement systems
Pitot tube, 500 mm long, stainless steel, for Oper. temp. 0635 2045
measuring flow velocity Ø 7 mm 0 to +600 °C
500 mm
Pitot tube, 350 mm long, stainless steel, for Oper. temp. 0635 2145
measuring flow velocity Ø 7 mm 0 to +600 °C
350 mm
Pitot tube, 300 mm long, stainless steel, for Oper. temp. 0635 2245
measuring flow velocity Ø 4 mm 0 to +600 °C
300 mm
Pitot tube, 1000 mm long, stainless steel, for Oper. temp. 0635 2345
measuring flow velocity Ø 7 mm 0 to +600 °C
1000 mm
3-function probe for simultaneous measurement of Hot bulb 0 to +10 m/s ±(0.03 m/s ±5% of mv)(0 to
10 m/s)
0635 1540
temperature, humidity and velocity. With plug-in head, Testo humid. sensor, 0 to +100 %RH
cap. -20 to +70 °C ±2 %RH (+2 to +98 %RH)
0430 0143 connection cable required NTC ±0.4 °C (0 to +50 °C)
270 mm Ø 21 mm
±0.5 °C (remaining range)
Comfort level probe for measuring degree of turbulence, Hot wire 0 to +5 m/s ±(0.03 m/s ±4% of mv) 0628 0009
with telescopic handle and stand. Fulfills EN 13779 NTC 0 to +50 °C (0 to +5 m/s)
Ø 90 mm ±0.3 °C (0 to +50 °C)
requirements
CO2 probe measures indoor air quality and monitors the CO2 probe 0 ... +1 Vol. % CO2 ±(50 ppm CO2 ±2% of mv)(0 to
+5000 ppm CO2) 0632 1240
workplace. With plug-in head, connection cable 0430 0 ... +10000 ppm ±(100 ppm CO2 ±3% of
CO2 mv)(+5001 to +10000 ppm
0143 or 0430 0145 required
CO2)
Ambient CO probe, for detecting CO in buildings 0 to +500 ppm CO ±5% of mv (+100.1 to 0632 3331
and rooms +500 ppm CO)
±5 ppm CO (0 to +100
ppm CO)
Precision pressure dew point probe for 300 mm 0 to +100 %RH ±0.8 °C tpd (-4.9 to +50 °C tpd) 300 0636 9841
-60 to +50 °C tpd ±1 °C tpd (-9.9 to -5 °C tpd) s
measurements in compressed air systems incl. ±2 °C tpd (-19.9 to -10 °C tpd)
cert. with test point -40°C tpd ±3 °C tpd (-29.9 to -20 °C tpd)
±4 °C tpd (-40 to -30 °C tpd)
Super quick-action immersion/penetration probe 150 mm -200 to +600 °C Class 1* 1s 0604 0493
for measurements in liquids
Ø 1.5 mm
Plug-in head. connection cable 0430 0143 or 0430 0145 required
Super quick-action immersion/penetration probe 150 mm 20 mm -200 to +600 °C Class 1* 1s 0604 9794
for measurements in gases and liquids with a Ø 1.4 mm Ø 0.5 mm
low-mass tip Plug-in head. connection cable 0430 0143 or 0430 0145 required
Pipe wrap probe for pipes up to 2" in diameter -60 to +130 °C Class 2* 5s 0600 4593
Fixed cable
Spare meas. head for pipe wrap probe, TC Type K 35 mm -60 to +130 °C Class 2* 5s 0602 0092
15 mm
Globe thermometer to measure radiant heat Accuracy corresponds to 0 to +120 °C ±0.5 °C (0 to +49.9 °C) 0554 0670
ISO 7243, ISO 7726, DIN ±1 °C (+50 to +120 °C)
Ø 150 mm
EN 27726, DIN 33403
requirements
Fixed cable
*According to standard EN 60584-2, the accuracy of Classes 1 / 2 refer to -40 to +1000/+1200 °C.
Accessories for velocity probes, pressure probes Part no. Accessories: Humidity, 3-function probe Part no.
Professional telescopic handle for plug-in vane probes, max. 1 m long 0430 0941 Cable, 1.5 m long, connects probe with plug-in head to meas. instrument 0430 0143
PUR coating material
Extension for telescopic handle, 2 m long 0430 0942 Cable, 5 m long, connects probe with plug-in head to measuring instrument 0430 0145
please also order the 0409 0063 extension cable PUR coating material
Handle for plug-in vane probes 0430 3545 Extension cable, 5 m long, between plug-in head cable and instrument 0409 0063
PUR coating material
Extension cable, 5 m long, between plug-in head cable and instrument 0409 0063 Telescopic handle, max. 1 m, for probe with plug-in head 0430 0144
PUR coating material cable: 2.5 m long, PUR coating material
Connection hose, silicone, 5m long 0554 0440 testo saline pots for control and humidity adjustment of humidity probes, 0554 0660
max. load 700 hPa (mbar) 11.3 %RH and 75.3 %RH with adapter for humidity probe
Cover plugs for test holes (50 off) 0554 4001 Metal protection cage, Ø 12 mm for humidity probes 0554 0755
for measurement in flow velocities of less than 10 m/s
Cap with wire mesh filter, Ø 12 mm 0554 0757
testo 400 The reference measuring instrument for A/C and ventilation systems
Precision reference class
measuring instruments have
everything the professional user VAC module for evaluating the
needs to complete complicated measurement directly on site with
measurement tasks efficiently, integrated inaccuracy calulation
accurately and conveniently. Clear graphics display Attachable printer
3 user defined function buttons Prints readings on site in the
testo 400 includes the matter of seconds
parameters Save up to max. 500,000 readings) or
print at the touch of a button
temperature, CO2, rpm, current, Mains connection/quick battery
voltage, relative humidity, recharge
pressure, flow and volume flow.
Attachable printer (optional)
Clear graphics display
Intelligent electronics ensure the Prints readings on site in a matter of °C
We recommend:
ComSoft 3 - Professional with data management 0554 0830
incl. database, analysis and graphics function, data analysis, trend curve
RS232 cable 0409 0178
connects instrument to PC (1.8 m) for data transfer
We recommend: Attachable printer (securely attached) including 1 roll of thermal paper and 0554 0570
DAkkS calibration certificate/temperature* 0520 0201 batteries
El. resistance thermometer, el. thermometer; cal. points selectable from -80 to +1000°C SoftCase (protects instrument from impact) with carrier strap, magnetic 0516 0401
holder and probe holder
The pro set for assessing workplaces subjected to heat SoftCase for attachable printer (protects printer from dirt/impact) 0516 0411
protects from impact and falls
- testo 400, multi-functional measuring instrument, incl. measurement value store up to 500,000
readings, VAC-module (determination of volume flow with error calculation), battery, Li-cell and System case (aluminium) for measuring instrument, probes and accessories 0516 0410
calibration protocol (Part no. 0563 4001) probes in lid make it easy to find parts in case
- Wet Bulb Globe temperature probe to assess workplaces subjected to heat, in accordance with ISO DAkkS calibration certificate/Velocity for laboratory fume cupboard probe*
7243 or DIN 33403, incl. WBGT case (Part no. 0635 8888)
- Attachable printer (securely attached) including 1 roll of thermal paper and batteries (Part no. ISO calibration certificate/Velocity for laboratory fume cupboard probe
0554 0570)
Measurement systems
Technical data
Probe type Vane Thermal Testo humid. sensor, cap. Pressure aw value
Meas. range 0 to +60 m/s 0 to +20 m/s 0 to+100 %RH 0 to +2000 hPa 0 to +1 aW
Accuracy See probe data for system See probe data for system See probe data Probe 0638 1347 See probe data
accuracy accuracy Probe 0638 1447
±1 digit Probe 0638 1547
Probe 0638 1647
Probe 0638 1747
Probe 0638 1847
±0.1% of mv
Probe 0638 1741
Probe 0638 1841
Probe 0638 1941
Probe 0638 2041
Probe 0638 2141
±0.2% of mv
Resolution 0.01 m/s (for Ø 60/100 mm), 0.01 m/s (0 to +20 m/s) 0.1 %RH (0 to +100 %RH) 0.001 hPa (Probe 0638
0.1 m/s (for rem. probes) 1347)
0.001 hPa (Probe 0638
1447)
0.01 hPa (Probe 0638 1547)
0.1 hPa (Probe 0638 1647)
0.1 hPa (Probe 0638 1747)
0.1 hPa (Probe 0638 1847)
0.01 bar (Probe 0638 1741)
0.01 bar (Probe 0638 1841)
0.01 bar (Probe 0638 1941)
0.01 bar (Probe 0638 2041)
0.01 bar (Probe 0638 2141)
Probe type NTC Pt100 Type K (NiCr-Ni) Type S (Pt10Rh-Pt) Type J (Fe-CuNi)
Meas. range -40 to +150 °C -200 to +800 °C -200 to +1370 °C 0 to +1760 °C -200 to +1000 °C
Accuracy ±0.2 °C (-10 to +50 °C) ±0.1 °C (-49.9 to +99.9 °C) ±(0.3 °C + 0.1% of mv) ±1 °C (0 to +1760 °C) ±0.4 °C (-150 to +150 °C)
±1 digit ±0.4 °C (-40 to -10.1 °C) ±(0.1 °C + 0.1% of mv) ±1 °C (-200 to -150.1 °C)
±0.4 °C (+50.1 to +150 °C) (remaining range) ±1 °C (+150.1 to +1000 °C)
Resolution 0.1 °C (-40 to +150 °C) 0.01 °C (-99.9 to +300 °C) 0.1 °C (-200 to +1370 °C) 1 °C (0 to +1760 °C) 0.1 °C (-200 to +1000 °C)
0.1 °C (-200 to -100 °C)
0.1 °C (+300.1 to +800 °C)
Probe type CO2 probe CO probe Mechanical Current/voltage measurement Current/voltage measurement
Meas. range 0 to +1 Vol. % CO2 0 to +500 ppm CO 20 to 20000 rpm 0 to +20 mA (0554 0007) 0 to +10 V
0 to +10000 ppm CO2 0/4 to 20 mA (0554 0528)
Accuracy See probe data ±5% of mv (0 to +500 ppm CO) ±1 digit ±0.04 mA (0 (0554 0007) ±0.01 V (0 to +10 V)
to +20 mA)
±1 digit
Measurement systems
Thermocouple, made of fibre-glass insulated 2000 mm -200 to +400 °C Class 1* 5s 0644 1109
thermal pipes, pack of 5 Insulation: twin conductor, flat, oval, opposed and covered with fibre-glass, both
conductors are wrapped together with fibre-glass and soaked with lacquer,
Please order adapter 0600 1693 Ø 0.8 mm please order adapter 0600 1693
Quick-action surface probe with sprung 150 mm -200 to +300 °C Class 2* 3s 0604 0194
thermocouple strip, measuring range short-term Ø 10 mm
to +500 °C Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
Super quick-action surface probe, probe tip at -200 to +300 °C Class 2* 3s 0604 0994
50 mm
90° angle, with sprung thermocouple strip Ø 10 mm
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
100 mm
Robust surface probe 150 mm -200 to +600 °C Class 1* 25 s 0604 9993
Ø 4 mm
Ø 4 mm
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
Robust surface probe with sprung thermocouple 200 mm -200 to +700 °C Class 2* 3s 0600 0394
strip for high temperature range up to +700°C Ø 15 mm
Conn.: Fixed cable, coiled
Roller surface probe for measurements on rollers 274 mm -50 to +240 °C Class 2* 0600 5093
and rotating drums, max. circumferential velocity Ø 33 mm
18 to 400m/min Conn.: Fixed cable, coiled
Magnetic probe, adhesive power approx. 20 N, 35 mm -50 to +170 °C Class 2* 0600 4793
with magnets, for measurements on metal Ø 20 mm
surfaces Conn.: Fixed cable
Magnetic probe, adhesive power approx. 10 N, 75 mm -50 to +400 °C Class 2* 0600 4893
with magnets, for higher temperatures, measures Ø 21 mm
on metal surfaces Conn.: Fixed cable
Adhesive thermocouple, pack of 2, carrier Diameter extension 2 x
-200 to +200 °C Class 1* 0644 1607
material: aluminium foil 0.2 mm, 0.1 mm thick
Is fixed at the measuring point using conventional adhesives or silicone heat paste 0554 0004
Fast response immersion/penetration probe 150 mm -200 to +400 °C Class 1* 3s 0604 0293
Ø 3 mm
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
Super quick-action immersion/penetration probe 150 mm -200 to +600 °C Class 1* 1s 0604 0493
for measurements in liquids Ø 1.5 mm
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
Super quick-action immersion/penetration probe 470 mm -200 to +1100 °C Class 1* 1s 0604 0593
for high temperatures Ø 1.5 mm
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
Robust immersion/penetration probe made of 150 mm -200 to +400 °C Class 1* 3s 0600 2593
V4A stainless steel, waterproof and oven-proof, Ø 3.5 mm
Ø 3 mm
e.g. for the food sector Conn.: Fixed cable
Smelting probe for measurements in non-ferrous 1100 mm -200 to +1250 °C Class 1* 60 s 0600 5993
melting baths, with exchangeable measuring tip Ø 6.5 mm
Conn.: Fixed cable
Pipe wrap probe for pipes up to 2" in diameter -60 to +130 °C Class 2* 5s 0600 4593
* According to standard EN 60584-2, the accuracy of Classes 1 / 2 refer to -40 to +1000/+1200 °C.
Plug-in measuring tip, 550mm long, flexible, for 550 mm -200 to +1100 °C Class 1** 4s 0600 5793
high temperatures, outer casing: Inconel 2.4816 Ø 3 mm
Please order handle with Part no. 0600 5593
Plug-in measuring tip, 1030mm long, flexible, for 1030 mm -200 to +1100 °C Class 1** 4s 0600 5893
high temperatures, outer casing: Inconel 2.4816 Ø 3 mm
Please order handle with Part no. 0600 5593
Precision air probe 150 mm -100 to +400 °C 1/10 Class B (0 to 100°C) 1/5 75 s 0628 0017
Class B (rem. range) to EN
Ø 3 mm Ø 9 mm 60751***
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
Robust surface probe 150 mm -50 to +400 °C Class B*** 40 s 0604 9973
Ø 4 mm Ø 9 mm
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
Velcro probe for pipes with diameter of max. 75 280 mm -50 to +150 °C Class B*** 40 s 0628 0019
mm
Conn.: Fixed cable
Standard immersion/penetration probe 200 mm Stainless Steel -200 to +400 °C Class A*** 20 s 0604 0273
Ø 3 mm
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
Standard immersion/penetration probe 200 mm Nickel -200 to +550 °C Class A*** 20 s 0604 0274
Ø 3 mm
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
Highly accurate immersion/penetration probe incl. 295 mm Stainless Steel -40 to +300 °C ±0.05 °C (+0.01 to +100 °C) 60 s 0614 0240
±(0.05 °C ±0.05% of mv)
certificate Ø 4 mm (-40 to 0 °C)
±(0.05 °C ±0.05% of mv)
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required (+100.01 to +300 °C)
Highly accurate immersion/penetration probe 200 mm -100 to +400 °C 1/10 Class B (0 to 100°C) 1/5 30 s 0628 0015
Class B (rem. range) to EN
Ø 3 mm 60751***
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
Flexible precision immersion probe, cable heat- 1000 mm 50 mm -100 to +265 °C 1/10 Class B (0 to 100°C) 1/5 Class 80 s 0628 0016
proof up to +300°C B (rem. range) to EN 60751***
Ø 3.5 mm Ø 6 mm
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
Robust immersion/penetration probe with 150 mm -200 to +400 °C Class A*** 30 s 0604 2573
sharpened measuring tip, waterproof and oven- Ø 3.5 mm Ø 3 mm
proof Conn.: Fixed cable
*with EEPROM: Precision adjustment for each probe at a measuring point; measuring range limits are saved in probe; t95 extrapolation; surface allowance in surface probe can be adapted to measuring task
**According to standard EN 60584-2, the accuracy of Class 1 refers to -40 to +1000 °C.
***According to standard EN 60751, the accuracy of Class A and B refer to -200 to +600 °C.
Globe thermometer to measure radiant heat 0 to +120 °C ±0.5 °C (0 to +49.9 °C) 0554 0670
Ø 150 mm ±1 °C (+50 to +120 °C)
Accuracy corresponds to ISO 7243, ISO 7726, DIN EN
Conn.: Fixed cable 27726, DIN 33403 requirements
Measurement systems
CO2 probe measures indoor air quality and 0 ... +1 Vol. % CO2 ±(50 ppm CO2 ±2% of mv)(0 to +5000 ppm 0632 1240
monitors the workplace. With plug-in head, 0 ... +10000 ppm CO2 CO2)
connection cable 0430 0143 or 0430 0145 ±(100 ppm CO2 ±3% of mv)(+5001 to
Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required +10000 ppm CO2)
required
Mechanical rpm probe with plug-in head 20 to 20000 rpm ±1 digit 0640 0340
Included
2 probe tips Ø 8 and Ø 12 mm Conn.: Plug-in head. connection cable 0430 0143 or 0430 0145 required
1 hollow cone Ø 8 mm
1 surface speed disc Ø 19 mm to measure rotational speed: rpm = rotational speed in mm/s
Current/voltage cable (±1 V, ±10 V, 20 mA) 0 to +1000 mV ±1 mV (0 to +1000 mV) 0554 0007
0 to +10 V ±0.01 V (0 to +10 V)
0 to +20 mA ±0.04 mA (0 to +20 mA)
Silicone heat paste (14g), Tmax = +260°C, improves heat transfer in 0554 0004
surface probes
Spare measuring tip for smelting probe 0363 1712
Duct humidity/temperature probe 180 mm 0 to +100 %RH ±2 %RH (+2 to +98 ±0.4 °C (-10 to +50 °C) 12 s
-20 to +70 °C %RH) ±0.5 °C (remaining range)
0636 9715
Telescopic handle 0430 9715, see Ordering data/Accessories
Ø 12 mm
Fixed cable
Thin humidity probe incl. 4 attachable protection 250 mm 0 to +100 %RH ±2 %RH (+2 to +98 ±0.4 °C (-10 to +50 °C) 15 s 0636 2130
caps for ambient air measurements, measurements -20 to +70 °C %RH) ±0.5 °C (-20 to -10.1 °C)
Ø 4 mm ±0.5 °C (+50.1 to +70 °C)
in exhaust air ducts and equilibrium moisture Plug-in head. connection cable 0430 0143 or 0430 0145 required
measurements
Highly accurate reference humidity/temp. probe 0 to +100 %RH ±1 %RH (+10 to +90 * ±0.2 °C (+10 to +40 °C) 12 s 0636 9741
Ø 21 mm -20 to +70 °C %RH)* ±0.4 °C (remaining range)
±2 %RH (remaining
Plug-in head. connection cable 0430 0143 or 0430 0145 required range)
Humidity/temperature probe 0... +100 %RH ±2 %RH (+2... +98 ±0.4 °C (+0.1 to +50 °C) 12 s 0636 9742
Ø 21 mm -20 to +70 °C %RH) ±0.5 °C (-20 to 0 °C)
±0.5 °C (+50.1 to +70 °C)
Plug-in head. connection cable 0430 0143 or 0430 0145 required
Precision pressure dew point probe for 300 mm 0 to +100 %RH ±0.8 °C tpd (-4.9 to +50 °C tpd) 300 0636 9841
-60 to +50 °C tpd ±1 °C tpd (-9.9 to -5 °C tpd) s
measurements in compressed air systems incl. ±2 °C tpd (-19.9 to -10 °C tpd)
cert. with test point -40°C tpd ±3 °C tpd (-29.9 to -20 °C tpd)
Plug-in head. connection cable 0430 0143 or 0430 0145 required ±4 °C tpd (-40 to -30 °C tpd)
High humidity level probe w/ heated sensor 300 mm 0 to +100 %RH ±2.5 %RH (0 to ±0.4 °C (-10 to +50 °C) 30 s 0636 2142
element, no humidity on sensor -20 to +85 °C +100 %RH) ±0.5 °C (-20 to -10.1 °C)
Ø 12 mm ±0.5 °C (+50.1 to +100 °C)
Plug-in head. connection cable 0430 0143 or 0430 0145 required
Robust high temperature/humidity probe up to 300 mm 0 to +100 %RH ±2 %RH (+2 to +98 ±0.4 °C (+0.1 to +50 °C) 30 s 0628 0021
+180°C -20 to +180 °C %RH) ±0.5 °C (remaining range)
Ø 12 mm
Plug-in head. connection cable 0430 0143 or 0430 0145 required
Flexible humidity probe (does not retain shape) 1500 mm 100 mm 0 to +100 %RH ±2 %RH (+2 to +98 ±0.4 °C (+0.1 to +50 °C) 30 s 0628 0022
for measurements in inaccessible places -20 to +180 °C %RH) ±0.5 °C (-20 to 0 °C)
Ø 12 mm ±0.5 °C (+50.1 to +180 °C)
Plug-in head. connection cable 0430 0143 or 0430 0145 required
Probes Material and equilibrium moisture Illustration Meas. range Accuracy t99 Part no.
Flexible humidity probe with mini module for 0 to +100 %RH ±2 %RH (+2 to +98 ±0.4 °C (-10 to +50 °C) 20 s 0628 0013
meas. e.g. on material testing rigs, module cable -20 to +125 °C %RH) ±0.5 °C (remaining range)
length 1500mm, probe tip 50x19x7mm
Plug-in head. connection cable 0430 0143 or 0430 0145 required
Sword probe for measuring humidity and 320 mm 0 to +100 %RH ±2 %RH (+2 to +98 ±0.4 °C (-10 to +50 °C) 12 s 0636 0340
temperature in stacked material -20 to +70 °C %RH) ±0.5 °C (-20 to -10.1 °C)
18 mm x 5 mm ±0.5 °C (+50.1 to +70 °C)
Plug-in head. connection cable 0430 0143 or 0430 0145 required
Robust humidity probe e.g. for measuring 300 mm 0 to +100 %RH ±2 %RH (+2 to +98 ±0.4 °C (-10 to +50 °C) 30 s 0636 2140
equilibrium moisture or for measurements in -20 to +120 °C %RH) ±0.5 °C (remaining range)
Ø 12 mm
exhaust ducts to +120°C
Plug-in head. connection cable 0430 0143 or 0430 0145 required
Material moisture probe Free scaling, reference 0636 0365
measurement, no water level
Material/building moisture cable 0 to 100 k Ohm = Display values in instrument 0636 0565
100 to 0 % display mean: 100 to 66 wet; 0
to 1 very dry
Pressure probes Illustration Meas. range Accuracy Overload Static pressure Zeroing Part no.
Precision pressure probe, 100 Pa, in robust metal 0 to +100 Pa ±(0.3 Pa ±0.5% of 50 hPa 100 hPa up to 20 Pa 0638 1347
housing with impact protection, incl. magnet for fast mv)
attachment, to measure differential pressure and flow
speeds (in combination with Pitot tube) Plug-in head. connection cable 0430
0143 or 0430 0145 required
Pressure probe, 10 hPa, in robust metal housing with 0 to +10 hPa ±0.03 hPa 50 hPa 1000 hPa to 0,4 hPa 0638 1447
impact protection incl. magnet for fast attachment, to
measure differential pressure and flow speeds (in
combination with Pitot tube) Plug-in head. connection cable 0430
0143 or 0430 0145 required
Pressure probe, 100 hPa, in robust metal housing with 0 to +100 hPa ±0.5% of mv (+20 to 300 hPa 1000 hPa to 4 hPa 0638 1547
impact protection, incl. magnet for fast attachment, to +100 hPa)
measure differential pressure and flow speeds (in ±0.1 hPa (0 to +20
combination with Pitot tube) hPa)
Plug-in head. connection cable 0430
0143 or 0430 0145 required
Pressure probe, 1000 hPa, measures differential 0 to +1000 hPa ±1 hPa (0 to 200 2000 hPa 1000 hPa to 20 hPa 0638 1647
pressure, in robust metal housing with impact protection, hPa)
incl. quick-closing coupling (M8 x 0.5), magnet for fast ±0.5% of mv (200 to
attachment 1000 hPa)
Plug-in head. connection cable 0430
0143 or 0430 0145 required
Pressure probe, 2000 hPa, measures differential 0 to +2000 hPa ±2 hPa (0 to 400 3000 hPa 1000 hPa to 40 hPa 0638 1747
pressure, in robust metal housing with impact protection, hPa)
incl. quick-closing coupling (M8 x 0.5), magnet for fast ±0.5% of mv (400 to
attachment 2000 hPa)
Plug-in head. connection cable 0430
0143 or 0430 0145 required
Pressure probe, 2000 hPa, measures absolute pressure, 0 to +2000 hPa ±5 hPa (0 to +2000 4000 hPa – – 0638 1847
in robust metal housing with impact protection, incl. hPa)
quick-closing coupling (M8 x 0.5), magnet for fast
attachment Plug-in head. connection cable 0430
0143 or 0430 0145 required
Measurement systems
Caps for humidity probes Ø 12m and 21mm Illustration Part no.
Metal protection cage, Ø 12 mm for humidity probes, material: stainless steel V4A. Quick 0636 9740, 0636 9715 0554 0755
adjustment time, robust and temperature-proof. Used when measuring velocities of less than 10 Ø 12 mm
m/s.
Cap with wire mesh filter, Ø 12 mm All humidity probes with Ø 12 mm 0554 0757
PTFE sintered filter, Ø 21 mm, PTFE. Not affected by condensation, water-repellent, resistant to All humidity probes with Ø 21 mm 0554 0666
corrosive substances. Applications: compressed air measurements, high humidity range Ø 21 mm
(continuous measurements), high flow velocities
Sintered PTFE filter, Ø 12 mm material PTFE. Favourable behaviour in condensation, water 0636 9769, 0636 9740, 0636 9715 0554 0756
repellent, high resistance to aggressive media. Applications: Compressed air measurements, high Ø 12 mm
humidity range (long-term measurements), high flow velocities.
PTFE sintered filter, Ø 12 mm, PTFE. Not affected by condensation, water-repellent, resistant to 0628 0021, 0628 0022, 0636 2140, 0554 0758
corrosive substances. Applications: compressed air measurements, high humidity range Ø 12 mm 0636 2142
(continuous measurements), high flowvelocities
Stainless steel sintered cap, Ø 21 mm, can be screwed onto humidity probe All humidity probes Ø 21 mm 0554 0640
protection in case of high mechanical load and high velocities Ø 21 mm
Stainless steel sintered cap, Ø 12 mm, material: stainless steel V2A. Very rugged, suitable for 0636 9740, 0636 9715 0554 0647
penetration, can be cleaned with compressed air, mechanical sensor protection. Applications: High Ø 12 mm
mechanical loads, high flow velocities.
PTFE cap, Ø 5 mm, attachable, PTFE material, (5 off). Applications: dust protection, high humidity 0636 2130 0554 1031
level measurements, high flow velocities Ø 5 mm
Accessories: Humidity probes Part no. Accessories: Pressure probes Part no.
Cable, 1.5 m long, connects probe with plug-in head to meas. instrument 0430 0143 Connection cable, 2.5 m long, for pressure probes 0638 1741/1841/1941 0409 0202
PUR coating material
Cable, 5 m long, connects probe with plug-in head to measuring instrument 0430 0145 Cable, 1.5 m long, connects probe with plug-in head to meas. instrument 0430 0143
PUR coating material PUR coating material
Extension cable, 5 m long, between plug-in head cable and instrument 0409 0063 Cable, 5 m long, connects probe with plug-in head to measuring instrument 0430 0145
PUR coating material PUR coating material
Telescopic handle, max. 1 m, for probe with plug-in head 0430 0144 Connection hose, silicone, 5m long 0554 0440
cable: 2.5 m long, PUR coating material max. load 700 hPa (mbar)
Adapter for surface humidity measurement, for humidity probes Ø 12 mm 0628 0012 Connection hose set, 2 x 1 m, coiled, incl. 1/8" screw connection 0554 0441
locates damp spots on walls, for example pressure-tight up to 20 bar, for probe 0638 1647
Cap for bore holes, for humidity probe Ø 12 mm 0554 2140 Adapter for pressure probes, 1/2" outer thread, 1/4" inner thread 0699 3127
Measures equilibrium moisture in bore holes
testo saline pots for control and humidity adjustment of humidity probes, 0554 0660
11.3 %RH and 75.3 %RH with adapter for humidity probe
Vane/temperature probe, Ø 16 mm, attachable to Vane +0.4 to +60 m/s ±(0.2 m/s +1% of mv) 0635 9540
0430 3545 handle or 0430 0941 telescopic Type K (NiCr-Ni) -30 to +140 °C (+0.4 to +40 m/s)
180 mm ±(0.2 m/s +2% of mv)
handle (+40.1 to +50 m/s)
Ø 16 mm
Vane/temperature probe, Ø 25 mm, can be Vane +0.4 to +40 m/s ±(0.2 m/s ±1% of mv) 0635 9640
attached to 0430 3545 handle or 0430 0941 Type K (NiCr-Ni) -30 to +140 °C (+0.4 to +40 m/s)
telescopic handle 180 mm
Ø 25 mm
Bendable vane probe (can be bent by 90°), Ø 60 Vane +0.25 to +20 m/s ±(0.1 m/s ±1.5% of mv)0635 9440
mm, attachable to handle 0430 3545 or telescopic Oper. temp. (+0.25 to +20 m/s)
Ø 60 mm 0 to +60 °C
handle 0430 0941, for meas. on ventilation outlets
Bendable vane probe (can be bent by 90°), Ø Vane +0.1 to +15 m/s ±(0.1 m/s ±1.5% of mv)0635 9340
100 mm, attachable to handle 0430 3545 or Oper. temp. (+0.1 to +15 m/s)
telescopic handle 0430 0941, for measurements Ø 100 mm 0 to +60 °C
on ventilation outlets
Vane probe, Ø 16 mm, for stationary assembly, 3 +0.4 to +60 m/s ±(0.2 m/s ±1% of mv) 0628 0036
m cable (PVC) Ø 16 mm Oper. temp. (+0.4 to +60 m/s)
0 to +70 °C
250 mm
High temperature vane probe, Ø 25 mm, with Vane +0.6 to +20 m/s ±(0.3 m/s ±1% of fsv) 0635 6045
handle for continuous measurements up to Type K (NiCr-Ni) -40 to +350 °C (+0.6 to +20 m/s)
+350°C 560 mm Ø 25 mm
Thermal probes Illustration Probe type Meas. range Accuracy Part no.
Robust hot bulb probe, Ø 3 mm, for Hot bulb 0 to +10 m/s ±(0.03 m/s ±5% of mv) 0628 0035
measurements in the lower velocity range, 2m Ø 3 mm NTC -20 to +70 °C (0 to +10 m/s)
cable (PVC) 150 mm
Affordable, robust hot bulb probe, Ø 3 mm, for Hot bulb 0 to +10 m/s ±(0.03 m/s ±5% of mv) 0635 1549
measurements in the lower velocity range, with NTC -20 to +70 °C (0 to +10 m/s)
handle 150 mm Ø 3 mm
Robust hot bulb probe, Ø 3 mm, with handle and Hot bulb 0 to +10 m/s ±(0.03 m/s ±5% of mv) 0635 1049
telescopic handle for measurements in the lower NTC -20 to +70 °C (0 to +10 m/s)
velocity range 850 mm Ø 3 mm
Quick-action hot wire probe, Ø 10 mm, with Hot wire 0 to +20 m/s ±(0.03 m/s ±4% of mv) 0635 1041
telescopic handle, for measurements in the lower NTC -20 to +70 °C (0 to +20 m/s)
velocity range with direction recognition 760 mm Ø 10 mm
Thermal anemometer probe, Ø 10 mm, w. Hot wire 0 to +5 m/s ±(0.02 m/s ±5% of mv) 0635 1047
telescopic handle, measures air flow in lab fume NTC 0 to +50 °C (0 to +5 m/s)
cupboards to DIN EN 14175 760 mm Ø 10 mm
Differential pressure probes Illustration Probe type Meas. range Accuracy Overload Static pressure Zeroing Part no.
Precision pressure probe, 100 Pa, in robust metal Differential pressure 0 to +100 Pa ±(0.3 Pa ±0.5% of 50 hPa 100 hPa to 20 Pa 0638 1347
housing with impact protection, incl. magnet for fast probe mv)
attachment, to measure differential pressure and flow
speeds (in combination with Pitot tube)
Pressure probe, 10 hPa, in robust metal housing with Differential pressure 0 to +10 hPa ±0.03 hPa 50 hPa 1000 hPa to 0,4 hPa 0638 1447
impact protection incl. magnet for fast attachment, to probe
measure differential pressure and flow speeds (in
combination with Pitot tube)
Pressure probe, 100 hPa, in robust metal housing with Differential pressure 0 to +100 hPa ±0.5% of mv (+20 to 300 hPa 1000 hPa to 4 hPa 0638 1547
impact protection, incl. magnet for fast attachment, to probe +100 hPa)
measure differential pressure and flow speeds (in ±0.1 hPa (0 to +20
combination with Pitot tube) hPa)
Measurement systems
300 mm
Pitot tube, 350 mm long, stainless steel, for Oper. temp. 0635 2145
measuring flow velocity Ø 7 mm 0 to +600 °C
350 mm
Pitot tube, 500 mm long, stainless steel, for Oper. temp. 0635 2045
measuring flow velocity Ø 7 mm 0 to +600 °C
500 mm
Pitot tube, 1000 mm long, stainless steel, for Oper. temp. 0635 2345
measuring flow velocity Ø 7 mm 0 to +600 °C
1000 mm
Straight Pitot tubes Illustration Probe type Meas. range Part no.
Pitot tube, stainless steel, 360 mm long, measures flow Type K (NiCr-Ni) -40 to +600 °C 0635 2040
speed and temperature, for pressure probes 0638
1347/..1447/..1547 360 mm Ø 8 mm
Pitot tube, stainless steel, 500 mm long, measures flow Type K (NiCr-Ni) -40 to +600 °C 0635 2140
speed and temperature, for pressure probes 0638
1347/..1447/..1547
500 mm Ø 8 mm
Pitot tube, stainless steel, 1000 mm long, measures flow Type K (NiCr-Ni) -40 to +600 °C 0635 2240
speed and temperature, for pressure probes 0638
1347/..1447/..1547
1000 mm Ø 8 mm
Comfort level measurement Illustration Probe type Meas. range Accuracy Part no.
3-function probe for simultaneous measurement of Hot bulb 0 to +10 m/s ±(0.03 m/s ±5% of mv)(0 to
10 m/s) 0635 1540
temperature, humidity and velocity. With plug-in head, Testo humid. sensor, 0 to +100 %RH ±2 %RH (+2 to +98 %RH)
0430 0143 connection cable required cap. -20 to +70 °C ±0.4 °C (0 to +50 °C)
270 mm Ø 21 mm NTC ±0.5 °C (remaining range)
Comfort level probe for measuring degree of turbulence, Hot wire 0 to +5 m/s ±(0.03 m/s ±4% of mv) 0628 0009
with telescopic handle and stand. Fulfills EN 13779 NTC 0 to +50 °C (0 to +5 m/s)
Ø 90 mm ±0.3 °C (0 to +50 °C)
requirements
890 mm
Wet Bulb Globe temperature probe to assess workplaces Ø 150 mm 0 to +120 °C In accordance with ISO 0635 8888
subjected to heat, in accordance with ISO 7243 or DIN 7243 or DIN 33403
33403, incl. WBGT case ID No.
0699 4239/1
Multi-function
instruments Use measuring instrument for the data to:
• The locations can be selected via barcode labels using the pen.
Save Delete
• The measurement is completed when...
- the predefined number of readings is reached.
- date/time is reached.
- the memory is full.
- ended manually.
0.22 m/s
WINDOW 1 WINDOW 1
m3/h
• The individual measurement protocol can be either saved or deleted following analysis. 870 1070 1 m/s2 m3/h
- - - - - - - - -
3 °C
- -
Mean: 3 27.70 °C
- - - - - - - - - - - - - - - - - - - - - -
1 0.22 m/s
Info:
2 965.00 m3/h Req. value 1000 m3/h
3 27.70 °C
- - - - - - - - - - - 16.10.02
15:29:32
Info:
Req. 1000 m3/h
16.10.02
15:29:32
Multi-function
instruments Use ComSoft 3 software on measurements to:
• The readings are determined using the measuring instrument and can also be displayed
online using the software.
• Analysis:
- with calculation functions
- with crosshairs
- with mean calculation
- with calculation of standard deviation
- taking all conventional refrigerants into consideration (refrigeration module, optional)
• Display:
- as table or as graphic
- as digit field or as histogram
- with analog display
- Measurement channels can be activated or deactivated at the touch of a button
• Documenting:
- Data is transferred to Excel table using “Copy and Paste”.
• Functions can be selected from the function list and the finished profile can be saved.
ComSoft 3 - Professional for: ComSoft 3 - Professional with data Accessories Part no.
management RS232 cable 0409 0178
incl. database, analysis and graphics connects instrument to PC (1.8 m) for data transfer
· Monitoring instrument testo 445
· Reference instruments testo 400 function, data analysis, trend curve
and 454
Part no.
0554 0830
Additional
Weitere
information
Infos unter
at www.testo.com 25
26-098X-705X-Printer 13.12.2010 11:30 Seite 26
Testo fast printer Versatile infrared printer for testo 435, 445, 400
The universal printer with IRDA System compatibility with other Testo
and infrared interface saves you measuring instruments (also
time since it stores the print data downward compatibility)
prior to printing. Data transfer is Fast data transfer, the measuring
completed within 2 seconds. The instrument is ready for use again
instrument is then immediately within 2 sec.
ready for operation. Fast print function thanks to newest
line printer
The readings are saved black on
Energy-saving Auto-off/Wake-up
white with date and time.
function
Testo design with integrated magnetic
plate
Robust housing (adapted to testo
327)
Mains operation possible (same
Testo fast printer with wireless
mains unit as for testo 327/330)
infrared interface, 1 roll thermal
paper and 4 AA batteries
Part no.
0554 0549
26 Additional
Weitere
information
Infos unter
at www.testo.com
27-098X-705X-Ethernet 13.12.2010 11:30 Seite 27
Accessories
Ethernet offers:
· Fast transmission of readings
· Use of an existing network
without additional cabling
· Long transmission distances
· Identification of measuring
instruments in system network
Part no.
0554 1711