Thank you for purchasing this Ozeri Blood Pressure Monitor. Please read this
manual thoroughly before first use, and keep this manual handy for future
6755 Mira Mesa Blvd. reference.
Suite 123-239
San Diego, CA 92121
sales@ozeri.com
TABLE OF CONTENTS TABLE OF CONTENTS
Operating Instructions
Applying the Arm Cuff and Positioning 15
Measuring Your Blood Pressure 16
About Blood Circulation and Pressure 17
Irregular Heart Beat? 18
What is Arrhythmia? 19
Recalling Records 20
Deleting Records 21
Troubleshooting the Unit 22
Maintenance 23
2 3
INTRODUCTION IMPORTANT SAFETY INFORMATION
Thank you for purchasing the Ozeri BP3T Fully Automatic To ensure the correct use of this product, basic safety
Blood Pressure Monitor. This device uses the oscillometric measures should always be followed. Read all information
method of blood pressure measurement which detects the in the instruction manual and any other literature included in
movement of blood through your brachial artery, and the box before using this device, and observe all Warnings
converts your blood pressure into a digital reading. The unit and Cautions.
is simple to use because no stethoscope is needed while
using an oscillometric monitor. This device automatically Warnings
stores 60 sets of 2 measurement values with date and time The following are warnings of potentially hazardous
data for 2 users (120 measurements total). You can access situations which, if not avoided, could result in death or
the stored data simply by pressing the MEM button. serious injury:
The BP3T Monitor includes the following items: ● DO NOT confuse self-monitoring with self-diagnosis. This
● Monitor unit allows you to monitor your blood pressure. Contact
● Cuff your physician for specific information about your blood
● Instruction manual printed in English pressure. Self-diagnosis and treatment using measured
● Carrying bag results may be dangerous. Follow the instructions of your
● 4 AAA batteries physician or licensed healthcare provider.
● Consult your physician before measuring blood pressure,
especially if you have any of the following conditions:
Hypertension, Diabetes, Arteriosclerosis, Kidney or
Vascular Disease, or any conditions affecting circulation.
● DO NOT adjust medication based on measurement results
from this blood pressure monitor. Take medication as
prescribed by your physician. Only a physician is qualified
to diagnose and treat High Blood Pressure.
● The monitor is not intended to be a diagnostic device.
4 5
IMPORTANT SAFETY INFORMATION (CONTINUATION) IMPORTANT SAFETY INFORMATION (CONTINUATION)
BEFORE YOU START
● This device is intended for use in measuring blood ● Operate this device only as it is intended. Do not use this
pressure and heart rates for adults only. Do not use on device for any other purpose.
infants or persons who cannot express their intentions. ● Do not use a cellular phone near this device. Cell phones
● This unit is not suitable for continuous monitoring during may inhibit the accuracy of this device.
medical emergencies or operations. ● Do not disassemble or attempt to repair this device or any
● The equipment is not AP/APG equipment and not suitable of its components. Doing so will void the warranty.
for use in the presence of a flammable anesthetic mixture ● Do not use this device where flammable gases (such as
with air, oxygen, or nitrous oxide. anesthetic gas, oxygen or hydrogen) or flammable liquids
● If battery fluid should get in your eyes, immediately rinse (such as alcohol) are present.
with plenty of clean water. Contact a physician ● This device is intended for non-invasive measuring and
immediately. monitoring of arterial blood pressure. It is not intended for
use on extremities other than the upper arm or for
Precautions functions other than obtaining a blood pressure
The following are Precautions of potentially hazardous measurement.
situations which, if not avoided, may result in minor or ● If you are taking medication, consult your physician to
moderate injury to the user or patient or damage to the determine the most appropriate time to measure your
equipment or other property: blood pressure.
● If the cuff pressure exceeds 40 kPa (300 mmHg),the unit
● Read all of the information in the instruction manual and will automatically deflate. Should the cuff not deflate when
any other literature in the box before operating the unit. pressures exceeds 40 kPa (300 mmHg), detach the cuff
● Dispose of the device, components and optional from the wrist and press the stop button to stop inflation.
accessories according to applicable local regulations.
Unlawful disposal may cause environmental pollution.
● Use only Ozeri authorized parts and accessories. Parts
and accessories not approved for use with the device may
damage the unit.
6 7
HEALTH AND BLOOD PRESSURE VALUES SUGGESTIONS BEFORE MEASURING
What is the standard blood pressure classification? 1. Rest for at least 5-10 minutes before taking your blood
This device refers to a blood pressure classification pressure.
developed by The World Health Organization (WHO) and 2. Avoid eating, smoking, exercising, bathing and drinking
the International Society of Hypertension (ISH). (especially alcohol, caffeine or tea) for at least 20-30
Diastolic Blood Pressur (mmHg) minutes before taking a measurement, and avoid talking
Severe hypertension while measuring your blood pressure.
110
Moderate hypertension
3. Make sure the wrist cuff is at heart level.
100 4. Stress raises blood pressure. Try to keep calm before
Mild hypertension
90 and during the measurement process.
Normal systolic value
85 5. Avoid a very cold environment.
Normal blood pressure
80
Optimal
6. Use the bathroom before taking measurement.
blood pressure
(target value)
7. Avoid excessive talking and hand movement.
120 130 140 160 180 Systolic Blood Pressure (mmHg)
8. Wait at least 20 min after taking bath or shower.
9. Wait 2-3 minutes before retaking your blood pressure
Your monitor will display the blood pressure classifications measurement.
as shown in the table below: 10. Keep the records for your physician and refer to your
physician for guidance. Remember that one reading
Blood
Level
does not provide a true blood pressure value; only
Pressure
(mmHg) long-term records will reveal a more accurate picture
130~139
SYS <120 120~129 140~159 160~179 >180
of your blood pressure. Try to take your
DIA 80~84 85~89 90~99 100~109
<80 >110
measurement at the same time each day to stay
consistent.
CAUTION
Only a physician can tell you your normal blood pressure range
and the point at which you are at risk. If the measurements taken
with this product consistently fall outside your normal range,
consult your physician for further testing.
8 9
LCD FULL DISPLAY DEVICE COMPONENTS & CONTROL BUTTON DESCRIPTIONS
CUFF
AIR HOSE
AC power socket
Deflating Cuff air is deflating
MEM/UP
Time (hour:minute) Currently time TOUCH KEY
START/OFF BUTTON
Memory Displays measurements stored in memory
LCD DISPLAY SET/DOWN TOUCH KEY
Measured unit of blood pressure
mmHg
(1mmHg=0.133kPa)
10 11
BATTERY INSTALLATION / REPLACEMENT PROGRAMMING THE UNIT
Remove and store the batteries if this device will not be used for
save it. The unit will
an extended period of time (i.e., 3 months). The batteries can leak
successively
chemicals which can destroy electronic components in this device.
continue on to
Old batteries are harmful to the environment, so please dispose of
setting the date,
properly and follow your local recycling guidelines.
time, and units.
12 13
PROGRAMMING THE UNIT (CONTINUATION) APPLYING THE ARM CUFF AND POSITIONING
4.Repeat steps 2 and 3 to set the [Month] and [Day] Applying the cuff to your left arm
1.Make sure the air plug is securely
inserted into the main unit.
START START
STOP STOP
14 15
MEASURING YOUR BLOOD PRESSURE ABOUT BLOOD CIRCULATION AND PRESSURE
START
1.Press the STOP to turn on the monitor. START
The unit will begin the entire measurement STOP Blood circulation is responsible for supplying the body with
automatically during which it will cycle through oxygen from the lungs. During each heartbeat cycle, the blood
the following screen displays: pressure exerted on the arteries varies between a maximum
(systolic) and a minimum (diastolic) pressure.
Unit is automatically
The diastolic blood pressure value (lower pressure or lower
performing a zero
adjustment in preparation
value) represents the blood pressure produced in between
for measurement. heartbeats when the heart muscle relaxes.
START
2.Press the STOP to power off
START
the monitor. The unit will also
STOP
turn off after 1 minute of
inactivity.
16 17
IRREGULAR HEARTBEAT? WHAT IS ARRHYTHMIA?
An irregular heartbeat is defined as a heartbeat that A heartbeat is stimulated by electrical signals that cause
varies by 25% from the average of all heartbeat intervals the heart to contract. Arrhythmia is a condition where the
while the unit is measuring the systolic and diastolic blood
heartbeat rhythm is abnormal due to flaws in the
pressure.
bio-electrical system that drives the heartbeat. Typical
symptoms are skipped heartbeats, premature contraction,
an abnormally rapid (tachycardia) or slow (bradycardia)
pulse. This can be caused by heart disease, aging,
physical predisposition, stress, lack of sleep, fatigue etc.
The appearance of an irregular heartbeat symbol does
If an irregular rhythm is detected more than twice during
indicate arrhythmia. Arrhythmia and other heart conditions
measurement, the irregular heartbeat symbol with three
hearts will appear on the display along with the can only be properly diagnosed by a doctor through a
measurement results. medical examination.
Irregular
Heartbeat
Symbol
18 19
RECALLING RECORDS DELETING RECORDS
1.With the monitor turned off, select user The following procedure will delete all records for a user.
A or B, and then briefly press the
“MEM” button to show the average of 1.With the monitor
the record. turned off, select
user A or B.
The “AVG” sign will MEM Press and hold
show in the right “MEM” for 3 seconds
corner, Indicating that until the display at
the average recorded right appears.
values of Systolic,
Diastolic, and Pulse
rate are displayed 2.If you do not want to delete the
records, press START
STOP
to escape.
2.Press the
“MEM” or “SET” UP
button to scroll Otherwise, press
up or down “SET” to confirm
through the DOWN
deleting and the
records. monitor will turn
off automatically.
The the record,
date, and time
will display Second of eight The corresponding The corresponding
alternately in records. date is October 11th. time is 10:38.
the lower right
corner of the 3.When the user’s
screen. record is empty the
display at right will
CAUTION show.
The most recent record is shown first. Each new measurement is
assigned to the first record. All other records are pushed back one
digit (e.g., 2 becomes 3, and so on), and the last record (60) is
dropped from the list.
20 21
TROUBLESHOOTING THE UNIT MAINTENANCE
The following table includes a list of error messages and In order to get the best performance and to maximize the life
the possible causes and solutions for problems you may of your unit please review the following tips:
encounter with your blood pressure monitor. If the product
is not operating as you think it should, please check here
before arranging for servicing.
No power
Display is dim or
will not light up.
Batteries are dead.
M
EM
The cuff is not secured Refasten the cuff and then
E1 properly. measure again. Store in a dry cool place and Do not drop or shake the unit,
Secure the air plug into avoid direct sunshine. and do not store objects on the unit.
Air plug disconnected.
the main unit.
E2 The cuff over inflated. Refasten the cuff and
then measure again.
22 23
SPECIFICATIONS ACCESSORIES
Accuracy Pressure:
15oC~25oCwithin±0.4kpa(3mmHg)
Adapter
10oC~40oC(out of 15oC~25oC) Type:UE08WCP-060100SPA
within ±0.8kpa (6mmHg) Input:100~240V,50~60Hz,400mA
Pulse value: ±5% Output:6V 1A
Guidance and manufacturer’s declaration – electromagnetic emissions Immunity test IEC 60601 Compliance level Electromagnetic environment –
test level guidance
The BP3T is intended for use in the electromagnetic environment specified Electrostatic ± (2, 4, 6) ± (2, 4, 6) kV Floors should be wood,
below. The customer or the user of the BP3T should assure that it is used in discharge (ESD) kV contact contact concrete or ceramic tile. If floors
such an environment. ± (2, 4, 8) ± (2, 4, 8) kV are covered with synthetic
IEC 61000-4-2 kV air air material, the relative humidity
Emissions Compliance should be at least 30 %
Electromagnetic environment - guidance
test
Electrical fast ±2 kV for Not applicable Main power quality should be
The BP3T uses RF energy only for its internal transient/burst power that of a typical commercial or
RF emissions hospital environment.
function. Therefore, its RF emissions are very low
Group 1 IEC 61000-4-4 supply lines
and are not likely to cause any interference in nearby
CISPR 11 ±1 kV for
electronic equipment.
input/output
RF emissions lines
Class B ±1 kV
CISPR 11 Surge Not applicable Main power quality should be
differential that of a typical commercial or
Harmonic IEC 61000-4-5 mode hospital environment.
emissions ±2 kV common
Not applicable mode
The BP3T is suitable for use in all establishments,
including domestic establishments. Voltage dips, <5 % UT Not applicable Main power quality should be
IEC 61000-3-2
short (>95 % dip in that of a typical commercial or
Voltage interruptions and UT) hospital environment. If the user
fluctuations / voltage variations for 0,5 cycle of the BP3T requires continued
flicker emissions Not applicable on power supply operation during power mains
input lines 40 % UT interruptions, it is recommended
IEC 61000-3-3 (60 % dip in UT) that the BP3T be powered from
IEC 61000-4-11 for 5 cycles an uninterruptible power supply
or a battery.
70 % UT
Guidance and manufacturer’s declaration – electromagnetic immunity (30 % dip in UT)
for 25 cycles
The BP3T is intended for use in the electromagnetic environment specified below.
The customer or the user of the BP3T should assure that it is used in such an <5 % UT
(>95 % dip in
environment.
UT)
for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic
(50/60 Hz) fields should be at levels
magnetic field characteristic of a typical
location in a typical commercial
IEC 61000-4-8 or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
26 27
FCC STATEMENT FCC STATEMENT
Guidance and manufacturer’s declaration – electromagnetic immunity NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
The BP3T is intended for use in the electromagnetic environment specified is affected by absorption and reflection from structures, objects and people.
below.
The customer or the user of the BP3T should assure that it is used in such an Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
environment.
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
Immunity IEC 60601 Compliance Electromagnetic environment – assess the electromagnetic environment due to fixed RF transmitters, an
test test level level guidance electromagnetic site survey should be considered. If the measured field strength in the
Portable and mobile RF communications location in which the BP3T is used exceeds the applicable RF compliance level above,
equipment should be used no closer to the BP3T should be observed to verify normal operation. If abnormal performance is
any part of the BP3T, including cables, observed, additional measures may be necessary, such as reorienting or relocating the
than the recommended separation BP3T.
distance calculated from the equation
applicable to the frequency of the Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3
transmitter. V/m.
28 29
FCC STATEMENT WARRANTY INFORMATION
Recommended separation distances between Your Ozeri Blood Pressure Monitor, excluding the arm
portable and mobile RF communications equipment and the BP3T cuff, is warranted to the original owner for 5 years from the
The BP3T is intended for use in an electromagnetic environment in which date of purchase against defects in material and
radiated RF disturbances are controlled. The customer or the user of the
workmanship under normal use. The arm cuff is warranted
BP3T can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications to the original owner for 1 year to be free from defects in
equipment (transmitters) and the BP3T as recommended below, according to material and workmanship under normal use. Should your
the maximum output power of the communications equipment.
blood pressure monitor and/or arm cuff appear defective
Rated maximum
output power of Separation distance according to frequency of transmitter within the applicable period from the date of purchase,
transmitter
return the defective unit, freight prepaid along with an
W explanation of the claim to Ozeri. Please pack carefully to
Electrical fast 150 kHz to 80 80 MHz to 800 800 MHz to 2,5 avoid damages in transit. Under this warranty, Ozeri will
transient/burst MHz MHz GHz
d = 1,2 P d = 1,2 P d = 2,3 P
repair or replace any parts found to be defective due to a
IEC 61000-4-4
manufacturer’s defect. This warranty extends only to
0.01 0.12 0.12 0.23
personal use and does not extend to any product that has
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
been used for commercial use, or use which is not
10 3.8 3.8 7.3
intended. There are no warranties other than that
100 12 12 23 expressly set forth herein.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
30 31
WARRANTY INFORMATION CUSTOMER SUPPORT
This warranty is not transferable. Ozeri is not responsible Ozeri Customer Service
in any way for damages, losses or inconveniences caused Customer service: 1-877-299-1296 or
by equipment failure caused by user negligence, abuse, or Email: support@ozeri.com
use not in accordance with the “User Guide” or any (OzeriTM and its subsidiaries assume no liability for
additional safety, use, or warnings included in the product damage caused by the use of the Ozeri Blood Pressure
materials. Further, Ozeri is not liable for incidental or Monitor other than for its intended use or as instructed
consequential damages of any nature resulting from the above and in the User Manual.)
use of this product and any liability shall not exceed the
purchase price of the product.
32 33