You are on page 1of 16

Summit Industries, Inc.

CATHODE H.V. CABLE ANODE H.V. CABLE


RECEPTACLE RECEPTACLE

Figure 21: High Voltage Transformer Diagram – AWW430-10


(-) (+)

L
C

H.V. X-FMR
SILICONE
TERMINALS
FILAMENT FILAMENT RECTIFIERS
(LARGE) (SMALL)
X-FMR X-FMR
10-3

Installation and Service Manual


XC XS

P2 P1
G
M1 XL
HIGH VOLTAGE X-RAY
TRANSFORMER

Innovet Select™
A700 HIGH VOLTAGE TRANSFORMER DIAGRAM
J432 J476

0
Tested Tap VAC
A455
Switch
02255-000

ON / OFF
Circuit
W440
Breaker
SNUBBER
A365
120
VAC

A981

SCR BLOCK
00194-000
J425 : Cable Display
Generator (internal)

J424 : Cable Display


COLUMN (external)

Each wire at Auto-


Transformer has 1/4
inch connect

Collimator Supply
J400 Control

H3:14 VAC

Motor Start Board


MOLEX C1
Collimator connects here.
function J300
Collimator Switch 1 TP1
Collimator Switch 2 TP2
C1 - 24VAC L1 TP11
C2 - 24VAC L2 TP12
GND Collimator chassis.
<O>
(J300 is small board inside
the collimator, Lamp Timer)
. . . . . . . . . . .
mA Select Time Select
Input Input
H11 H10

H9
Fil Out

H1
SCR

H8
Fil
Resist

H7
Rotor

H2 H3 J1 H5 H6
KV in 14 VAC Therm. footsw. Display Data
(AC volts) Power input bypass and Therm. Output
COMPONENT LAYOUT for J350 DISPLAY PCBA
00130-000
RC Timing Control
Technical Support Dept
Ph: (800) 729 - 9729
Fax: (773) 588 - 3404

A455 Interfacing to the Generator

12/29/06
Annotations by Roger F, Tech Support

Filament Boost
Unregulated 120VAC at PREP
DC (Chassis F2 Via Switch)
12vdc 20 - 24v

not used Fuse rating reduced to 0.04A

> DC Ground
(-) side C3 Capacitor
same as Chassis F1 fuse

09
Anode
Motor
Common

>
Time Delay
Adjust

Expose Interlock to Timer

Expose
07 Motor Main Ready
Indicator

08 Motor Aux.

Chassis F1.

You might also like