Professional Documents
Culture Documents
INSTALLATION, OPERATION
AND MAINTENANCE MANUAL
BILGE WATER SEPARATOR
ULTRA-SEP™ MODEL US1000-CJ103
440-480V / 60 Hz / 3PH
120 VAC CONTROL
All vessels and parts thereof that are claimed to be defective by the purchaser must be returned to an
authorized CWWS service depot, at the expense of the purchaser, for inspection prior to the allowance of
any credit or replacement.
This warranty does not apply to damage resulting from accident, neglect, or misuse. The foregoing
warranty is the only warranty extended by CWWS and is in lieu of all other warranties, express or
implied, or for fitness of a particular purpose.
TABLE OF CONTENTS
SECTION 1: INTRODUCTION ..................................................................................................................... Page
1.1 MANUAL DESCRIPTION............................................................................................................ 1-1
1.2 THEORY OF OPERATION ......................................................................................................... 1-2
SECTION 2: GENERAL DESCRIPTION
2.1 GENERAL SYSTEM DESCRIPTION......................................................................................... 2-1
2.2 PROCESS FLOW DESCRIPTION .............................................................................................. 2-1
2.3 SPECIFICATIONS ........................................................................................................................ 2-3
SECTION 3: INSTALLATION
3.1 UNPACKING ................................................................................................................................. 3-1
3.2 INSTALLATION............................................................................................................................ 3-1
3.2.1 SITE SELECTION AND PREPARATION .............................................................................. 3-1
3.2.2 PIPING CONNECTION ............................................................................................................. 3-1
3.3 ELECTRICAL REQUIREMENTS .............................................................................................. 3-3
SECTION 4: OPERATION INSTRUCTIONS
4.1 COMPONENT FUNCTIONAL DESCRIPTION........................................................................ 4-1
4.2 PRESTART UP .............................................................................................................................. 4-2
4.2.1 HELI-SEP FILL .......................................................................................................................... 4-3
4.2.2 SPIR-O-LATOR FILL ................................................................................................................ 4-3
4.3 INITIAL START-UP & OPERATION ........................................................................................ 4-3
4.3.1 NORMAL OPERATION (BELOW 15PPM) ........................................................................... 4-3
4.3.2 NORMAL OPERATION OIL DISCHARGE........................................................................... 4-3
4.3.3 NORMAL OPERATION (OVER 15 PPM) RECIRCULATION TO BILGE ....................... 4-4
4.4. ROUTINE SHUTDOWN PROCEDURES.................................................................................. 4-4
4.5 ALARMS......................................................................................................................................... 4-4
SECTION 5: MAINTENANCE PROCEDURES
5.1 HELI-SEP FLUSH & REMOVAL ............................................................................................... 5-1
5.2 SPIR-O-LATOR® FLUSH ............................................................................................................. 5-1
5.3 FLUSHING THE SMARTCELL (OIL CONTENT DETECTOR) ........................................... 5-2
5.4 FILTER BAG REPLACEMENT .................................................................................................. 5-3
5.5 SPIR-O-LATORS® CLEANING ................................................................................................... 5-4
5.5.1 PREPATATION FOR CLEANING .......................................................................................... 5-4
5.5.2 CLEANING PROCEDURES ..................................................................................................... 5-5
5.5.3 CLEANING DETERGENTS ...................................................................................................... 5-5
5.5.4 CHANGE OUT PROCEDURES (IF NEEDED CONSULT FACTORY FIRST) ................. 5-6
5.5.5 REDUCED RECOVERY ............................................................................................................ 5-5
5.6 SPIR-O-LATOR® PRESERVING ................................................................................................ 5-9
5.6.1 PREPARATION FOR PRESERVING ..................................................................................... 5-9
5.6.2 PRESERVING PROCEDURES ................................................................................................ 5-9
5.7 CLEANING THE OIL SENSING PROBE .................................................................................. 5-10
5.7 SEPARATOR SOLIDS & SLUDGE DUMP ............................................................................... 5-11
5.8 CLEANING THE SEPARATOR.................................................................................................. 5-12
SECTION 6: TROUBLE SHOOTING
6.1 TROUBLE SHOOTING GUIDE .................................................................................................. 6-1
SECTION 7: PARTS LIST
7.1 SPARE PARTS............................................................................................................................... 7-1
7.2 CONSUMABLE SUPPLIES ......................................................................................................... 7-1
APPENDICES
NOTES, CAUTIONS, AND WARNINGS are included in the operating procedures. A NOTE indicates an operating
procedure or condition that is essential to highlight or which provides an explanation for a given procedure. A CAUTION
indicates an operating procedure or practice that, if not strictly observed, may result in damage to, or destruction of
equipment or degradation of the water quality. A WARNING indicates an operating procedure or practice that can result
in bodily injury or loss of life if not correctly followed. NOTES, CAUTIONS, and WARNINGS always precede the
paragraphs to which they apply. Under normal operating conditions, adherence to the procedures described in this manual
will assure trouble-free operation.
It is important that the operator of the ULTRA-SEP® System reads and fully understands this entire manual. By
understanding the functions and normal operation of each component in the system, the operator can diagnose and correct
operating problems as they develop. Please take the time to read this manual, study the process flow diagram, and
understand the system before attempting to install and operate the system.
SPIR-O-LATOR
HELI-SEP MEMBRANES
FILTER CONTROL
BOX &
MOTOR
CONTROL
CENTER
OIL CONTENT
DETECTOR
PROCESS
PUMP
FEED PUMP
1-1
Section 1: Introduction
1.2 THEORY OF OPERATION
The ULTRA-SEP® Bilge Water Separator is a dual stage bilge water separation system. The first stage utilizes a
system generated vacuum, differential specific gravity and high efficiency two stage coalescer for the removal of the
majority of free oil and dirt separation. The second stage uses a micron filtration for removal of trace solids and
small oil droplets followed by ultra-filtration to separate and remove emulsified oils, along with the remaining
minute solids and entrained air from bilge water. The system is designed for continuous and/or intermittent
operation. After the system has been started, it is capable of automatic operation.
The innovative design of the CWWS Bilge Water Separator eliminates the need for flocculants and
chemical additives to remove the emulsified oily bilge water.
Stage 1
Incoming bilge water passes through a basket strainer (customer provided) to remove oversized solid contaminants.
From there it is drawn into the first stage consisting of a well proven coalescing HELI-SEP® Oily Water Separator,
where the free oil droplets coalesce as they pass through the permanent matrix. This oil rises to the top of the vessel
and it is collected for discharge during the oil discharge cycle. The water is then drawn through a pack of oleophillic
polishing beads, which trap the remaining fine oil droplets. These oil droplets are washed out and discharged during
the oil discharge cycle.
Stage 2
Processed water from the HELI-SEP® is then directed through particulate filtration, which further reduces solids
contamination. This semi-treated water is then fed into the process pump; this pump increases the flow and pressure
to the optimum conditions for the SPIR-O-LATORS®. The Spir-O-Lators® are engineered to produce clean water in
compliance with MEPC 107 (49), and a reject stream, which is re-circulated back for re-processing.
The system has been designed to be installed as a single skid assembly or in separate modules to accommodate
various engine room layouts.
OIL TO
RECOVERED
OIL TANK
MONITOR
HELISEP
FILTER
1-2
Section 2: General Description
2.1 GENERAL SYSTEM DESCRIPTION
The ULTRA-SEP® Bilge Water Separator is an ultra-filtration process designed for separating a wide variety of
chemically or mechanically emulsified bilge water solutions. At the heart of the system are the CWWS HELI-SEP®
and the CWWS SPIR-O-LATOR® (which are designed and developed specifically for bilge water and industrial
waste applications and are impervious to most chemicals). The ULTRA-SEP®’s major components are the HELI-
SEP®, Feed Pump, Bag Filter, Process Pump, the SPIR-O-LATORS® and must also include an approved MEPC 107
(49) Oil Content Detector. The system is also available with optional influent heater for enhanced cold water
operation. The ULTRA-SEP® is designed to eliminate the need for attended operation. The controls provide two
basic modes of operation; an run mode and a clean mode. The gauge displays provide operating pressures and visual
indicators of operating conditions. The ULTRA-SEP®' s components are all corrosion resistant for long service life
and are installed on one powder-coated structural steel skid which is arranged for ease of installation and servicing.
The Feed Pump pressurizes the influent for processing through the process Bag Filter before supplying the Process
Pump with a positive suction head. The Bag filter removes suspended solids and oil droplets of approximately 5
microns and larger from the flow.
This semi-cleaned influent is supplied to the Process Pump for processing by the CWWS SPIR-O-LATORS®. The
process pump increases the influent pressure and drives the emulsions across the surface of the SPIR-O-LATOR® at
a high velocity. Water, salts and detergents are forced through the SPIR-O-LATOR® by back pressure created by the
system and the systems operating pressure control valve while oils and other waste larger than or greater than twenty
thousandths (20,000) in molecular weight are rejected. The portion of the influent forced through the SPIR-O-
LATOR® is referred to as the processed water (clean water or permeate) while the remainder is referred to as the
concentrate or reject. The processed water is discharged from the system. This system is being constantly replenished
by the influent supplied by the Feed Pump. The concentrate with the rejected oils and waste is returned to inlet of the
Bag Filter, for de-oiling and polishing before being fed to the Process Pump inlet for reprocessing through the
SPIR-O-LATORS®. This reject reprocessing allows for a higher velocity flow rate across the SPIR-O-LATORS®
which maximizes the process water output by minimizing fouling because the higher velocity shears off the excess
oils and waste of the surface of the SPIR-O-LATORS®. The process flow rate is the combination of the process
water flow and the concentrate flow.
The ULTRA-SEP® is equipped with additional features and functions to maintain optimum operating conditions.
The process water, concentrate and concentrate bleed to bilge flows are factory set and can be manually checked for
verification. Liquid filled pressure gauges are provided and identified to help monitor operating conditions.
2-1
Section 2: General Description
In summary, the processed water, the output of the ULTRA-SEP®, is equal to the systems emulsified influent minus
the small concentrate bleed outlet stream. The concentrated bleed is continuously discharged from the recirculation
line, where concentrated oils can be further coalesced in the bilge, or holding tank, or in the HELI-SEP® when being
reprocessed through the ULTRA-SEP®. This process allows the ULTRA-SEP® to produce one stream of processed
water that is less than 15 parts per million that can be discharged overboard per IMO Resolution MEPC.107 (49).
The system is equipped with a Certified MEPC.107 (49) oil content monitor, which continuously monitors the
processed water for oil content. If the oil content exceeds the alarm set point the water will automatically be re-
directed to re-circulate to the bilge.
2-2
Section 2: General Description
2.3 SPECIFICATIONS
Type: ULTRA-SEP® system with a HELI-SEP® bilge water
separator, Bag Filter for solids, and SPIR-O-LATORS®
for enhanced waste/oil separation.
Clean Water Supply Ranges: (Flow) 0.1 M3/h Min. 0.45 GPM Min.
(Pressure) 275-550 kPa 40-80 PSI
Design Data:
2-3
Section 2: General Description
Piping Connections:
2-4
Section 3: Installation
3.1 UNPACKING
Carefully remove top and sides of the crating. Inspect the system for any damage that may have occurred during
shipment. Unbolt the system from the crate. Padding on fragile items should remain in place to protect fragile items
during rigging of the system into position. Once the system is securely installed, the padding on the fragile items can
be removed.
3.2 INSTALLATION
Installation in accordance with the recommendations that follow will ensure long and trouble-free operation from the
equipment. Refer to Appendix “C” for the Installation drawing for dimensional and orientation information, and to
the Piping & Instruments Drawing for piping arrangements.
NOTE
Allow ample room for servicing units electrical controls, pump motor,
SPIR-O-LATORS® and other components. Reference the drawings in
Appendix “C” for required clearances.
CAUTION
BE SURE THAT THE BILGE HAS BEEN THOROUGHLY CLEANED
OF ALL DEBRIS, LIQUIDS AND CONTAMINANTS REMAINING
FROM THE SHIPS’ FABRICATION BEFORE START UP.
Securely install the system as level as possible with the interface connections positioned to permit the most direct
route to the piping connections. Eliminate as many 90 degree turns as possible. System should be fully supported and
bolted down through holes provided in skid.
CAUTION
INSTALL A FOOT VALVE (USER SUPPLIED) AT THE SUMP OR
HOLDING TANK END OF THE SYSTEM’S INLET SUCTION LINE.
1. Connect a line to the Oily Water Inlet from the (user supplied) sump or holding tank which is to supply the
oily water to the ULTRA-SEP® for processing. Ascertain that a foot valve (user supplied) is installed at the
(user supplied) sump or holding tank end of the line. A bleed port should be provided as close to the foot
valve end of the suction line as possible for priming purposes. The source for the inlet should be in an area
that will provide the best quality solution possible to the system. Ensure that piping is capable of supplying
the inlet to the system as required by the specifications.
3-1
Section 3: Installation
NOTE
The processed water vertical lift to overboard can not exceed 20 Meters [65 feet]. If
a higher head is required an optional pump maybe required.
2. Connect the Processed Water Outlet to the discharge line going overboard.
3. Connect a line to both Recirculation’s to return the re-circulates preferably to the bilge for complete
reprocessing, or to a holding tank which is acceptable, but connecting these lines right back to the system’s
Oily Water Inlet is un-acceptable and must be avoided.
4. Connect the Clean Water Inlet to a clean pressurized water source (Sea or Fresh) for flushing the HELI-
SEP®, and for flushing or filling the ULTRA-SEP®. PRESSURE MUST NOT BE LESS THAN 1.38 bar
[20 PSI] or exceed 5.5 bar [80 PSI].
5. Connect a line to the Oil Discharge to discharge the Separator’s accumulated oil to a slop tank for waste
disposal or storage.
NOTE
For Gravity Drain, the drain lines must be higher than the (user supplied)
sump or bilge into which they are to discharge.
6. Connect lines to the Drains to gravity drain the system back to the (user supplied) sump or bilge.
3-2
Section 3: Installation
3.3 ELECTRICAL REQUIREMENTS
Refer to Appendix "C" for the appropriate Electrical Drawing for the system being installed. The electrical power
requirements are given on the serial plate.
WARNING
ELECTRICAL INSTALLATION AND CONNECTION TO THE MAIN
POWER SOURCE MUST BE PERFORMED BY A FULLY QUALIFIED
ELECTRICIAN IN ACCORDANCE WITH THE PERTINENT CODES
AND REGULATIONS. TURN OFF FACILITY POWER BEFORE
MAKING ANY ELECTRICAL CONNECTIONS. SYSTEM MUST BE
GROUNDED IN ACCORDANCE WITH APPLICABLE ELECTRICAL
CODES. FAILING TO FOLLOW THE PERTINENT ELECTRICAL
SAFETY REGULATIONS CAN CAUSE THE DAMAGE OF THE
SYSTEM, PERSONAL INJURY OR DEATH.
CAUTION
INSTALL A SUITABLE CIRCUIT BREAKER (USER SUPPLIED) TO
MAIN LINE CONNECTED TO THE SYSTEM. SEE WIRING DIAGRAM
DRAWING IN APPENDIX “C”.
1. Connect incoming power for the system to the main Power Disconnect Switch located in the Motor Control
Center (MCC).
CAUTION
SYSTEM SHOULD BE VENTED AND FULL OF WATER PRIOR TO
STARTUP OR ROTATION CHECK.
CAUTION
CHECK ROTATION OF SYSTEM'S PUMP/MOTORS PRIOR TO
INITIAL START-UP.
2. Prior to start-up, ascertain that the system pump/motors are rotating in the proper direction as indicated by
the rotation arrow located on each pump/motor. The information below is typical but follow the arrows.
3-3
Section 4: Operating Instructions
4.1 COMPONENT FUNCTIONAL DESCRIPTION
This table describes the function of the components.
10. Process Filter Housing (FIL1) Contain replaceable filter, the final polishing filter for the
influent before being processed by SPIR-O-LATOR®.
Lights
11. Power Illuminates when the power switch is in on position.
12. Feed Pump (PU1) Illuminates when the Feed Pump PU1 is on.
13. Process Pump (PU2) Illuminates when the Process Pump PU2 is on.
15. Oil Discharge Illuminates when the OSP (oil sensing Probe) senses and SV4
opens and puts he Heli-Sep® in an Oil Discharge Mode.
IMPORTANT: the same thing will happen if it sees air
instead of oil from leaky inlet lines or strainers from the bilge.
4-1
Section 4: Operating Instructions
16. Alarm • Illuminates when temperatures is over limit (TSH1).
• Illuminates when low pressure switch (PSL2) is sensed.
• Illuminates when Feed Pump (PU1) or Process Pump (PU2)
motor contactor is tripped.
• Illuminates when Oil Sensing Probe (OSP) is sensed
• Blinks when cleaning mode is complete.
4-2
Section 4: Operating Instructions
Operating chart (at a glance) valve, light, pump and switch location for each function. This chart should also be
referenced when reading all the following:
PRE START UP NORMAL OPERATION MAINTENANCE
4.3.3
OPERATING CHART OVER 15 5.1 5.2 5.5.1 & 5.5.2 5.3 5.4
4.2.1 4.2.2 SPIR- 4.3.2
SECTION 4.3.1 PPM HELI-SEP SPIR-O- CLEANING SMART BAG
HELI-SEP O-LATOR OIL
OPERATING RECIRCULA FLUSH & LATOR SPIR-O- CELL FILTER
FILL FILL DISCHARGE
TION TO SLUDGE FLUSH LATOR FLUSH CLEAN
BILGE
BV1
BV2
BV3
/ / /
MANUAL VALVE
BV4
BV5
BV7
BV8 /
BV10 /
GV2
GV3 LEAVE ALL AT FACTORY SETTINGS
GV4
SV1 * * * * * * * * *
AUTOMATIC
* * * * * * * *
VALVE
SV2 *
SV3
SV4 / /
FEED PUMP
OFF ON ON OFF ON OFF ON ON/OFF OFF OFF
PUMP
(PU1)
PROCESS
PUMP (PU2) OFF ON ON ON ON OFF ON ON OFF OFF
POWER ON ON ON ON ON ON ON ON ON OFF
SYSTEM
OFF OFF RUN RUN RUN RUN RUN RUN RUN OFF
SWITCHES
CLEAN /
OFF / RUN
FLUSH
OWS / OFF / SPIR-O- SPIR-O-
OWS OFF OFF OFF OWS OFF OFF OFF
SPIR-O- LATOR LATOR
LATOR
POWER ON ON ON ON ON ON ON ON ON OFF
FEED PUMP OFF ON ON OFF ON OFF ON ON OFF OFF
PROCESS
PUMP OFF ON ON ON ON OFF ON ON OFF OFF
LIGHTS
WATER
DISCHARG ON ON ON ON ON ON ON ON ON OFF
E
OIL
DISCHARG ON ON ON ON ON ON ON ON ON OFF
E
ALARM OFF OFF OFF OFF OFF OFF OFF OFF OFF OFF
* : CONTROLLED BY SMART CELL ONLY. IF OVER 15 PPM SV2 OPEN SV1 CLOSED. IF BELOW 15 PPM SV1 OPEN / SV2
CLOSED
/ : OPEN THEN CLOSE
4-3
Section 4: Operating Instructions
4.2 PRE-START UP
4.2.1 HELI-SEP FILL
1. Put all manual BV valves in the position shown on chart BV1, BV4 & BV8 are all open.
5. When air is gone and the OSP (Oil Sensing Probe) sees water, and when water bleeds out of BV2 & BV8
then close those valves.
6. Note, this will fill in 3 minute cycles only. If all the air is not gone out and the OSP (Oil Sensing Probe)
does not yet see water then put the Flush Switch in Off position and then back to OWS to repeat another
three minutes and continues until all the air is removed.
1. Put all manual BV valves in the position shown on the chart. BV1, BV4 & BV10 are open, all
others closed.
2. Turn Power Switch On, System Switch in Off position and Flush Switch in Spir-O-Lator® position.
3. When air is gone and water bleeds out of BV4 and BV10 then close those valves.
4. This cycle run in 3 minute increments. If all the air does not bleed out repeat cycle by putting Flush
Switch to Off the back to Spir-O-Lator® until all the air is removed.
1. Put all manual valves in position shown on the chart BV1 is open all others are closed.
2. Turn on Power Switch, put System Switch in Run position, put Flush Switch in Off Position.
3. The system will begin producing clean water and discharging it over board.
4. Power, Feed Pump & Process Pump lights will be illuminated. The Water Discharge light will also be
illuminated if SV1 is open and the system is discharging over board.
6. All solenoid and ball valves should be in the positions shown on P&ID and also shown in the operating
chart.
When Oil Sensing Probe (OSP) sees enough oil to discharge it will automatically do the following:
4-4
Section 4: Operating Instructions
1. Shut off Feed Pump PU1.
2. Open SV4.
5. If this discharge runs over 5 minutes before Oil Sensing Probe (OSP) sees water the system will shut
down and illuminate the Alarm light.
6. If the Oil Sensing Probe (OSP) signals the need for 3 consecutive oil discharges within 15 minutes the
system will also shut down and illuminate the Alarm light. Note, this typically mean one of two things:
a. There is excessive oil in the bilge or
b. There is an air leak feeding the system and the Oil Sensing Probe (OSP) sees air and thinks its
water
7. When all oil is discharged to the slop tank and oil sensing probe sees clean water it will revert back to
normal operation.
8. Note, Process Pump (PU2) will continue running during this cycle.
When the “Oil Content Detector” (OCD) reads over 15 PPM the following happens automatically
to prevent oil from being discharge over board.
CAUTION
4-5
Section 4: Operating Instructions
4.5 ALARMS
The following chart lists the alarms, activating device, and possible cause of the alarm. A shutdown alarm must be
reset by the Alarm Reset Switch before the Ultra-Sep can be restarted.
4-6
Section 5: Maintenance Procedures
5.1 HELI-SEP FLUSH & SLUDGE REMOVAL
It is recommended that this procedure be performed every 24-32 hours of operation to remove excess oil build up in
the system and to remove sludge that settle in the bottom of the vessel it should also be performed prior to each
membrane cleaning.
2. Put the System Switch in Off and allow the system to sit for 10 minutes.
6. Close BV3
7. Put Flush Switch in Off position and then back to OWS position to continue flushing to slop tank for 3
more minutes.
8. Put Flush Switch back into Off position the Process Pump (PU2) will go on and system will be back in
normal operation.
• After cleaning the Spir-O-Lator® to flush the cleaning detergent from the system.
3. The System will automatically flush for 3 minutes. Feed Pump (PU1) and Process Pump (PU2) will be
running, SV4 & SV3 open, SV1 & SV2 closed for 3 minutes.
If after a cleaning the smart cell reads over 15 PPM you can repeat the flushing or put the System Switch
in Run position it will recirculate to bilge until it flushes itself clean and then return to discharge over
board when the PPM drops below 15.
5-1
Section 5: Maintenance Procedures
5.3 FLUSHING THE SMARTCELL (OIL CONTENT DETECTOR)
1. On the OCD SMART-CELL use the DOWN and UP Keys to view the Menu options.
2. Scroll down to "Diagnostics" then press the ENTER key. The screen will change as shown:
3. Select “View disp PPM” and press ENTER. The clean water solenoid valve now is activated and clean
water is introduced to the monitor. You must then open (BV7) to allow the clean water to flow thru the
monitor and to drain. This will flush out the SMART-CELL. The display will appear as below and the
ppm reading of the clean water will be displayed. This reading should be less than 2 ppm.
4. Let the SMART-CELL flush for 30-60 seconds then close (BV7) and press the ENTER key twice to leave
the diagnostics screen, set the clean water solenoid valve back to sample and return to the operation screen
of the SMART-CELL.
5-2
Section 5: Maintenance Procedures
5.4 FILTER BAG REPLACEMENT
Refer to the Flow Diagram Drawing in Appendix “C” for the location of the referenced valves.
WHEN TO REPLACE
The Process Bag filters require replacing when the differential pressure reaches 1 bar [15 PSID] between the inlet
and outlet of the filter housing (Use PI2 and PI3 to determine this difference). Shut down the system by turning the
power switch (1) off
REPLACEMENT PROCEDURE
When changing bags it is a good practice to flush out the housing to remove accumulated dirt and sediment. The
following procedure should be followed to change the bag filters.
7. Check the internal filter bag support basket for debris, remove and clean if necessary.
9. If removed, re-install the cleaned support basket into the filter housing making sure basket flange is firmly
seated onto the step on the inside of the filter housing.
10. Install a new or cleaned filter bag into the support basket making sure the filter ring is firmly seated against
the top of the basket flange. Be sure that the bag is fully extended to the bottom of the basket.
11. Re-install the cover. When tightening the cover clamp, it may be necessary to light tap the clamp to have it
fully seat and seal the filter housing.
13. Refill the system as described in Spir-O-Lator® fill on section 4.2.2 before restarting. Be sure air is fully
bleed out of BV4 before closing.
5-3
Section 5: Maintenance Procedures
5.5 SPIR-O-LATORS® CLEANING
WHEN TO CLEAN
The Coffin World Water System ULTRA-SEP® Bilge water Separator employs an automatic system flush after
shutdown of the process pump, which is a very effective means of protecting the SPIR-O-LATORS® from fouling.
However, in normal operation the SPIR-O-LATORS® can become fouled by mineral scale, biological matter and
grime. These deposits build up during operation until they cause a loss in processed water production or a loss of
processed water quality, or both. The SPIR-O-LATORS® require a cleaning when the processed water back pressure
drops to 2 Bar [30 PSI], or when processed water quality decreases noticeably.
It should be noted that the processed water output rate will vary depending on when flow rate is checked in
relationship to time from last flush. Also, flow rate will change if the inlet composition changes, the operating
pressure changes or the temperature changes. This is normal and does NOT indicate SPIR-O-LATORS® fouling.
Check all system operating parameters and make comparison with initial operating data recorded during initial
startup, or in log sheets, before attempting to clean the SPIR-O-LATORS®. If after considering all the above factors,
it is determined that the SPIR-O-LATORS® require cleaning or change out, carefully follow the steps outlined on the
following pages.
2. First run a Heli-Sep flush see section 5.1, second run a Spir-O-Lator® flush see section 5.2.
CAUTION
When mixing the cleaning solutions, wear a face mask, rubber apron and
rubber gloves.
3. After the SPIR-O-LATOR® flush has concluded, a batch of detergent should be applied. Pour the
recommended amount of cleaning detergent as given in Section 5.4.4 into the Process Filter Housing’s
(FIL1) Cleaning Canister. Open the Process Filter Drain Valve (BV5) followed by Cleaning Valve (BV6) to
allow for the cleaning detergent to be drawn into the Process Filter by vacuum.
5. Once the Cleaning Canister is emptied of detergent, immediately close the Cleaning Valve (BV6) and
Process Filter Drain Valve (BV5) as to minimize pulling any air into the Process Filter Housing (FIL1).
5-4
Section 5: Maintenance Procedures
5.5.2 CLEANING PROCEDURES
1. Follow preparation for cleaning in Section 5.5.1. to perform a Pre Clean cycle and load a batch of cleaning
detergent.
2. Start the Cleaning cycle by opening BV4 and putting the clean/off/run switch in the clean position. During
this time, approximately 10 seconds, purge the Process Filter (FIL1) of any air accumulated during the
Preparation for Cleaning procedures by opening and closing its Vent Valve (BV4).
3. The Process Pump (PU2) will start running in a Circulation Cycle allowing the detergent to mix and flow
across the SPIR-O-LATORS® to scourer its surfaces. It will run for a minimum of 30 minutes or until the
temperature switch (TSH1) shuts the system down when the recalculating cycle heat the water to 50°C.
Note: Heating the water to between 45°C - 50°C will provide the best cleaning.
4 If after the Circulation Cycle is finished turn the clean switch off and let it soak for a minimum of 4 hours.
Over night would be best. Then do Spir-O-Lator® flush to remove detergent from the system see 5.2
5. After the flush has finished, and Spir-O-Lator cleaned water production has not fully recovered follow the
Preparation For Cleaning 5.4.1 steps 4 and 5 to inject another batch of cleaning detergent for a second and
last Circulation Cycle if needed.
EXCEPTION, if cleaning with chlorine bleach, DO NOT repeat this cycle and DO NOT soak.
6. At the end of the cleaning cycle (or when restarting after a prolonged or overnight soak) you will need to
flush out the SPIR-O-LATORS® thoroughly to remove all the cleaning detergents from the system,
otherwise on start up the smart cell will register higher PPM’s and recalculate to bilge until the detergent is
purged.
7. After cleaning, the process bag filter may need to be replaced with a new filter if the differential pressure is
over 10 PSI. (See Filter Bag Replacement 5.3)
8. The ULTRA-SEP® is now ready to operate. To restart the system after cleaning, refer to the Pre-start
procedures. If a prolonged shutdown is required, refer to the SPIR-O-LATOR® Preserving Section 5.6.
The best procedures to be used will often need to be determined by experience due to bilge water characteristics,
which change from ship to ship.
NOTE
It is recommended to keep the PH of any alkaline cleaning mixture between
10 and 12. See the following PH measurement procedure.
“The PH of the solution can be checked by taking a sample from the Process
Filter (FIL1) Vent Valve (BV4) after the solution has mixed for 1 to 2 minutes
for measurement. If the PH is over 12, drain some out and add water to the
canister. Continue this until the PH gets below 12.
1. If the bilge water is fouled with fair soot, (fine carbon particles) and/or oil, and /or organics it is
recommended to use CWWS Cleaner No. 1 or No. 1A (alkaline types).
5-5
Section 5: Maintenance Procedures
NOTE
Normally cleaning for oil (CWWS Cleaner No. 1 or No. 1A) will provide the
best performance, however a scale may gradually build up that is not
removed by this cleaning procedure. When this occurs, it will appear that
the Alkaline procedure is ineffective. An Acid cleaning is then
recommended to see if the problem is corrected. When using these acidic
compounds keep the PH levels over 2.
2. If the membrane is fouling due to the water hardness and metallic scale (rusty pipes or bilge water), it is
recommended to use CWWS Cleaner No. 2 (acid type).
NOTE
Chlorine bleach is to be strictly used only when biological fouling has
occurred and must be used in the recommended amounts (refer to table on
the following page) and NEVER allowed to soak or circulate for over ten
minutes.
3. If the membrane becomes biologically fouled, typical indications are bad odor, process water discolored to
light green, or it can be blackish. This usually means the unit has not been flushed or run for awhile and the
SPIR-O-LATORS® were allowed to soak in the bilge water. It is recommended to use a short, under 10
minutes, chlorine bleach cleaning cycle followed by a thorough flush and then followed by an alkaline
cleaning.
Also reference our Spare Parts List (7-1) for cleaners available:
NOTE
Other cleaners such as Drew Chemicals “Fresh” or “LAC” can be used to
clean the SPIR-O-LATORS for the short term. But their performance may
not be optimal and may decrease membrane life and void the warranty.
RECOMMENDED CWWS
PART GOOD FOR: TYPE:
CLEANERS NUMBER:
NOTE
As indicated in cleaning procedures, paragraph 5.4, cleaning should take
place when the processed water back pressure drops to 1.7 Bar [25 PSI]
(Reference Pressure Gauge PI5). The time between cleanings will vary
based on how the Bilge is set up, Bilge Management Practices, and the
characteristics of the Bilge Water.
5-6
Section 5: Maintenance Procedures
RECOMMENDED QTY TO BE USED PER CLEANING CYCLE (IN QUARTS)
MODEL CLEANER 1 & 1A / CLEANER 2 / CHOLRINE BLEACH
250 ½ ½ ½
500 ½ ½ ½
1000 ¾ ¾ ¾
2000 1½ 1½ 1½
3000 1½ 1½ 1½
5000 3 3 3
7500 4½ 4½ 4½
10000 6 6 6
TYPICAL SPIR-O-LATOR
P&ID PRODUCTION FLOW CURVE
Line 6
Max flow Clean Point
rating 30 PSI
1. Flush the Spir-O-Lator® (see 5.2) for system for 9-12 minutes before changing out. Then follow that with
the Preserving Procedure (5.6.2) before changing out.
NOTE
®
Refer to the SPIR-O-LATOR Housing Assembly drawing located in
Appendix “C”, for detailed assembly procedures.
2. Remove each complete fouled SPIR-O-LATOR® using the quick disconnects and replace with complete
new or cleaned SPIR-O-LATOR® from your stock.
3. Replace Process Filter (FIL1) if required before restarting the unit (See Filter Bag Replacement 5.3)
5. Pack the wet preserved fouled SPIR-O-LATOR® in the plastic bag the replacement came out of. Seal the
bag tight, then place in the individual SPIR-O-LATOR® box.
6. Obtain a RMA (Return Material Authorization) from CWWS for the special cleaning. When cleaned they
will be returned to you for the next cycle.
5-7
Section 5: Maintenance Procedures
NOTE
If you do see improvement, but not enough. Repeat the cycle extending the
circulation and soak times.
1. FIRST: Try a different cleaner following the normal cleaning procedure. For example, if you have been using
cleaner No. 1, try cleaner No. 1A, if this does not work then try No. 2. If you still have no success, try the
chlorine for Bio-Fouling as outlined in Section 5.4.4 Step 3, and as stated above, if you see improvement repeat
the cycle lengthening circulation and soak times.
2. SECOND: We have found the following two step procedure to be effective in some cases:
Step 1: Do a normal cleaning with cleaner No. 1, extending the cleaning circulation time (5.4.2 Step 3)
to 1 ½ hours. Then extend the final soak (5.4.2 Step 7) for a minimum of 1 hour. Be sure to flush
thoroughly when complete.
Step 2: Repeat the cleaning procedure again using cleaner No. 1A. Also repeat the 1 ½ hour cleaning
circulation time (5.4.2 Step 3)
NOTE
This will heat the solution to about 40°C
Then extend the final soak to overnight , or for a minimum of 4 hours. Follow with a through flushing.
CAUTION
Use the proper amounts of chemicals as outlined in section 5.5.3
and error on the (+) side.
5-8
Section 5: Maintenance Procedures
5.6 SPIR-O-LATOR® PRESERVING
PROLONG SHUTDOWN
The interior of the SPIR-O-LATORS®, being dark and moist, is an excellent breeding ground for microorganisms.
During prolonged shutdowns or storage periods of three (3) or more weeks, the SPIR-O-LATORS® should be
disinfected and sterilized by filling the system with a preserving solution. Preserving the SPIR-O-LATORS® will
save them from premature fouling or early failure by preventing bacteria from forming while the system is shut down
for extended periods of time. If not clean, the membranes should be cleaned before preserving. Refer to Flow
Diagram drawing in Appendix "C" for the location of the referenced valve callouts shown in parenthesis. Refer to
Figure 3, The Control Panel, for the location of the electrical switches and displays.
2. SECOND: Then prepare the SPIR-O-LATORS® per the Preparation for Cleaning procedures 5.4.1 except
on step 5 replace the cleaning detergent with the preserving solution as outlined below.
CAUTION
When adding the preserving solution, wear a face mask, rubber apron and
rubber gloves.
PRESERVATIVE MIXING
Model Teaspoons of Chlorine Mixed in pints of water
US250/500 1 2
US1000 1 1
US2000 2 2
US3000 3 3
US5000 4 4
US7500 6 6
US10000 8 8
2. Start the Preserving Cycle (same as cleaning cycle). During initial start up, approximately 10 seconds, purge
the Process Filter (FIL1) of any air accumulated during the Preparation for Preserving procedures by
opening and closing its Vent Valve (BV4).
2. After the 10 seconds, the Process Pump (PU2) will run in a Circulation Cycle allowing the preservative to
mix and flow across the SPIR-O-LATORS®. Allow the Circulation Cycle to operate for approximately 5 –
10 minutes to fully penetrate the SPIR-O-LATOR surfaces.
3. After the preservative has been circulated, stop the preserving operation by putting the Clean/Off/Clean
Switch to Off position.
5-9
Section 5: Maintenance Procedures
4. After a prolong shutdown, restart the system by first using the Manual Operation Procedures. Following the
pre-start procedures, see 4.2.1, 4.2.2
NOTE
When replacing the Oil Sensing Probe ensure that the threads are properly
aligned. Do not over tighten.
3. Replace Oil Sensing Probe and reconnect.
5-10
Section 6: Troubleshooting
6.1 TROUBLE SHOOTING GUIDE BILGE WATER SEPARATOR
WARNING
System inoperative when Facility power not ON. Turn on facility power.
Power Switch is turned to On.
Rapid cycling between operation Air leak in system or inlet lines. Hydrostatically test suction piping and
and oil discharge. vessel then repair as necessary.
Note: 3 Oil Discharges in 15
minutes will put the system in
alarm condition
Defective Probe Amp. Relay 5CR1. Replace relay.
Discharging excessive water to the Air leak in incoming lines Fix leak
slop tank
Dirty Oil Sensing Probe. Clean probe.
6-1
Section 6: Troubleshooting
No oil discharge. Clean water inlet line is not pressurized with Ascertain that clean water inlet line
water. is pressurized.
PU1Pump Motor will not start. Separator in Oil Discharge Mode. Check for pump run light or oil
discharge run light.
6-2
Section 7: Parts List
SEE Oil Content Detector Manual Spare Parts List for Oil Content Detector Spares
7-1
APPENDIX 'A'
Figures
1 ULTRA-SEP
ISO
2 ULTRA-SEP
ISO
3 ULTRA-SEP
Control Box
APPENDIX 'B'
Major Components
NO. DESCRIPTION
1 Membrane Housings
2 Membrane Elements
3 Process Pump
4 Feed Pump
5 Process Filter
6 Pressure Switches
8 System Valves
9 Pressure Gauges
7-1
8
1. Membrane Housings
7-1
10
4” SPIR-O-LATOR HOUSING: 1 - ELEMENT – MHUP1440
The SPIR-O-LATOR HOUSING has been designed to meet the standards of the
American Society of Mechanical Engineers (ASME), Boiler and Pressure Vessel
Code, Section X. At an additional cost, vessels can be inspected during
fabrication by an ASME Authorized Inspector and Code stamped.
L ±.06
49.1 28 15
S -1
(1247) (711) (7)
14
2. Membrane Elements
7-1
12
SPIR-O-LATOR
CWWS P/N: MUFO0449 Nominal Pore size: 0.01
TYPE: Ultrafilic, Hydrophilic Typical Operating Pressure: 5 bar (70 P.S.I.)
Dimensions: 4" x 40" Max. Pressure Drop Recommended: 0.7 bar (10 P.S.I.)
O il / W ater S eparation
T he s eparation of oil and water by ultrafiltration (U F) is a well-proven technology, but its wide
s pread utilization for was tewater m inim ization or recycling applications has been lim ited by three
com m on problem s :
Small pore Oil/dirt Larger pore Oil/dirt Pore opening Oil & dirt
with open rejected at with open rejected at become imbedded
substructure the surface substructure the surface in the pore &
substructure
Molecular Filtration Ultrafilter With Larger Pores Particle Filtration
Chemical Resistance
& Performance Versatility Membrane Chemical Compatibility
T he chem ical s tability of the U ltraFilic m em brane
CHEMICAL TYPE ULTRAFILIC PVDF PS
m aterial is key to the S pir-O-Lator's success
in the environm ental m arket (Figure 4). It not Aliphatic Hydrocarbons
(e.g., Hexane)
only is com patible with m os t hydrocarbons ,
acids and polar organic solvents used in m any Aromatic Hydrocarbons
(e.g., Xylene)
industrial plants, the U ltraFilic m em branes can
operate at tem peratures to 194 F (90 C) and are Halogenated Hydrocarbons
(e.g., 1,1,1 richloroethane)
T
stable in the pH range of 2 to 10. T his versatility
elim inates the need to neutralize the was te Polar Organic Solvents
(e.g., Acetone, MEK)
stream prior to filtration and m akes it possible
to recycle m os t aqueous cleaners on-line at Acids
their operating tem peratures .
Base
Figure 4
ate
Perme
ntra te
Conc e
Spir-O-Lator
UltraFilic Membranes
3. Feed Pump
(PU1)
7-1
14
Model 11810-Series
Model 11810-Series
SELF-PRIMING PUMPS
FEATURES
Body: Bronze Construction
Impeller: Nitrile Compounds
Shaft: Stainless Steel Sleeve over Motor Shaft
Wearplate: Replaceable
Shaft Seal: Lip Type
Ports: 3/4" (19mm) Garden Hose
External Thread and
1/2" NPT Internal
Motor: 1/3 HP, 115V AC 60 HZ, 1725 RPM,
Single Phase, Capacitor Type,
Open Drip Proof or Totally Enclosed
Fan Cooled, Handle and 8' Cord with
3 prong grounded plug on
ODP motor only
Weight: 20 lb (9.1 kg) with ODP Motor
25 lb (11.4 kg) with TEFC Motor
VARIATIONS AVAILABLE
MODEL NO. DESCRIPTION FOOD, BEVERAGE, & DAIRY: Circulate cooling water,
11810-0003 Nitrile Impeller, ODP Motor defrost refrigeration coils, fill drums, wash down
11810-0013 Nitrile Impeller TEFC Motor equipment and floors, and utility pump.
HOME: Fill and empty tubs and pools, empty
! WARNING basements, cesspools, water systems, water lawns and
Explosion hazard. Motor can spark. gardens, remove water from low areas, and recharge
Explosion & death can occur. Do not use water softeners.
where flammable vapors are present. PLUMBING: Empty clogged sumps, drains, sinks and
tubs, and pump out water closets.
APPLICATIONS FARM: Fill tractor tires, fill water tanks, pump out silo
pits, liquid manures, transfer oil, and flush and clean
MARINE: Bait tank circulation, transfer oil, wash down, animal pens.
and utility dock pumps.
CONTRACTORS: Remove water from excavations, sup-
INDUSTRIAL: Circulate coolant on machine tools, ply water to mortar boxes and cement mixing, and wash-
industrial filling systems, utility shop pump, transfer gum down equipment.
arabic, transfer distilled water, recharge water softeners,
ice water recirculation, hydrostatic pressure testing,
vacuum pump, and aspirating pump.
OPERATING INSTRUCTIONS ASSEMBLY
1. INSTALLATION: Pump may be mounted in any 1. Install slinger and position 1⁄4" from motor boss.
position without loss of efficiency; however, it is 2. Press seal into seal bore with lip facing toward the
suggested that the pump head be down if vertical impeller bore. Apply grease to back of seal lip.
mounting is desired. The rotation of the motor shaft
determines the location of the pump’s intake and dis- 3. Install wearplate, aligning slot with dowel pin.
charge ports. Refer to Dimensional Drawing. Right 4. Apply sealant to top of cam and cam screw threads.
hand port, looking at pump end cover, is Install cam with cam screw.
inlet port.
5. Install body on motor. Tighten nut at clamp. Do not
2. DRIVE: Direct. over tighten or binding of shaft will result.
3. SELF-PRIMING: Pump will produce a suction lift 6. Lubricate impeller bore with water pump grease.
approaching 10 feet (3m) when dry and a lift up to
22 feet (6.7m) when primed. Be sure suction lines 7. Lubricate shaft OD with grease, align drive ding in
are airtight or pump will not self-prime. extended impeller insert with flat on shaft and install
impeller.
4. RUNNING DRY: Unit depends on liquid pumped for
lubrication. Do not run dry for more than 30 seconds. 8. Install gasket, end cover and end cover screws.
Lack of liquid will damage the impeller TO EQUIP WITH VACUUM SWITCH (4732-0000)
5. PRESSURES: For continuous operation, pressure 1. Install switch in tapped hole on top of inlet port.
should not exceed 20 psi (1.4 kg/sq cm). (Seal threads prior to assembly.)
6. TEMPERATURES: Nitrile impeller recommended 2. Connect switch to motor. See diagram.
for liquid temperatures of 50° to 180°F (10° to 82°C).
3. Plug into 115V AC outlet and hold down restart lever
7. FREEZING WEATHER: Drain unit by loosening the on switch until the pump primes.
end cover.
SPECIAL ATTENTION
8. NOTICE: Do not pump solvents, thinners, highly
1. Constant vibration of the restart lever indicates:
concentrated or organic acids. Damage to pump
may result. If corrosive fluids must be handled, pump (A) An air leak. This will usually show up when the
life will be prolonged if pump is flushed with water pump is first started. All suction lines should be
after each use or after each work day. checked for loose connections.
9. GASKET: Use standard pump part. A thicker gasket (B) A worn impeller. Impeller should be replaced.
will reduce priming ability. A thinner gasket will cause 2. Intermittent stopping and starting indicates that the
impeller to bind. Standard thickness 0.010". unit is operating against excessive discharge
10. SPARE PARTS: A Jabsco Service Kit should be pressure.
kept on hand to service all but the most badly worn
11810-Series pump.
VACUUM SWITCH WIRING DIAGRAM
SERVICE INSTRUCTIONS
DISASSEMBLY
1. Remove end cover screws, gasket and end cover.
2. Grasp impeller hub with water pump pliers and
withdraw from body.
3. Loosen cam screw and remove cam. Clean off
sealant. Remove wearplate with screwdriver or
hooked wire.
4. Loosen nut at body clamp. Remove body from motor
shaft.
5. Press put seal from motor mounting end of body
towards impeller bore.
6. Remove slinger from shaft.
NOTE: Inspect all parts for wear or damage and replace
if necessary.
EXPLODED VIEW
PARTS LIST
6590-Series
Key Description Qty. Part No.
1 Screw (End Cover) 6 91002-0020
2 End Cover 1 11830-0000
3 *Gasket 1 3298-0000
4 Neoprene Impeller 1 5929-0001
*Nitrile Impeller 5929-0003
5 Cam 1 490-0001
6 Wearplate 1 2235-0000
7 *Seal (Shaft) 1 92700-0420
8 Screw (Cam) 1 91003-0010
9 Body 1 11814-0000
10 Motor (ODP) 1 18739-0000
Motor (TEFC) 18753-0014
11 Nut 1 9110-0030
12 Bolt 1 91095-0090
13 Slinger (O-Ring) 1 92000-0230
14 Plug (for vacuum 1 92650-0070
switch hole)
4732-0000 - Vacuum Switch Kit:
Automatic shut off for use when pump is left unattended.
* Parts contained in Service Kit.
www.jabsco.com
7-1
14
Installation,
Operation and
Model
Maintenance 3656/3756
Instructions
Table of Contents
SUBJECT PAGE
Safety Instructions .......................................................... 2
Description and Specifications ........................................ 2
Engineering Data ........................................................... 2
Installation .................................................................... 2
Location .................................................................... 2
Close-Coupled Units .................................................. 2
Frame-Mounted Units ................................................ 3
SAE Engine Driven Pumps .......................................... 3
Coupling Alignment ....................................................... 3
Frame-Mounted Units Only ........................................ 3
Piping ............................................................................ 4
Suction ...................................................................... 4
Owner’s Information Discharge ................................................................... 4
Wiring and Grounding ................................................... 4
Rotation ........................................................................ 4
Operation ...................................................................... 4
Model Number:
Maintenance .................................................................. 5
Serial Number: Disassembly ................................................................... 5
Dealer: Reassembly .................................................................... 6
Packed Box Instructions ................................................. 6
Dealer’s Phone No.
Trouble Shooting ........................................................... 7
Date of Purchase: Installation Date: Repair Parts ................................................................... 8
SAFETY INSTRUCTIONS Engineering Data
TO AVOID SERIOUS OR FATAL PERSONAL INJURY Maximum Liquid Temperature:
OR MAJOR PROPERTY DAMAGE, READ AND 212° F (100° C) – standard seal or packing
FOLLOW ALL SAFETY INSTRUCTIONS IN THE 250° F (120° C) – Optional high temp. seal
MANUAL AND ON THE PUMP. Maximum Working Pressure (table 1):
– NPT connections, 200 PSI (1379 kPa)
This is a SAFETY ALERT SYMBOL. – 125# ANSI flanged connections, 175 PSI (1207 kPa)
When you see this symbol on the pump or Maximum Suction Pressure: 100 PSI (689.5 kPa)
in the manual, look for one of the following Starts per Hour: 20, evenly distributed
signal words and be alert to the potential for
personal injury or property damage. Group Size Suction Discharge
11⁄2 x 2-6 (H) 2" NPT 11⁄2" NPT
DANGER Warns of hazards that WILL cause serious 1 x 2-7 2" NPT 1" NPT
personal injury, death or major property 21⁄2 x 3-7 3" NPT 21⁄2" NPT
damage. S
3 x 4-7 4" Flange 3" Flange
WARNING Warns of hazards that CAN cause serious 1 x 2-8 2" NPT 1" NPT
personal injury, death or major property 11⁄2 x 2-8 2" NPT 11⁄2" NPT
damage. 2 x 2-5 2" NPT 2" NPT
LH 21⁄2 x 21⁄2-5 21⁄2" NPT 21⁄2" NPT
CAUTION Warns of hazards that CAN cause personal 3 x 3-5 3" NPT 3" NPT
injury or property damage. 21⁄2 x 3-8 3" NPT 21⁄2" NPT
NOTICE: INDICATES SPECIAL INSTRUCTIONS 3 x 4-8 4" Flange 3" Flange
WHICH ARE VERY IMPORTANT AND 4 x 5-8 5" Flange 4" Flange
MUST BE FOLLOWED. 11⁄2 x 2-10 2" NPT 11⁄2" NPT
21⁄2 x 3-10 3" Flange 21⁄2" Flange
THIS MANUAL IS INTENDED TO ASSIST IN THE M
3 x 4-10 4" Flange 3" Flange
INSTALLATION AND OPERATION OF THIS UNIT. 4 x 6-10 6" Flange 4" Flange
THOROUGHLY REVIEW ALL INSTRUCTIONS AND 21⁄2 x 3-13 3" Flange 21⁄2 " Flange
WARNINGS PRIOR TO PERFORMING ANY WORK 3 x 4-13 4" Flange 3" Flange
ON THIS PUMP. 4 x 6-13 6" Flange 4" Flange
6 x 8-13 8" Flange 6" Flange
MAINTAIN ALL SAFETY DECALS.
L 8 x 10-13 10" Flange 8" Flange
NOTICE: INSPECT UNIT FOR DAMAGE AND 4 x 6-16 6" Flange 4" Flange
REPORT ALL DAMAGE TO CARRIER
IMMEDIATELY. Installation
WARNING UNIT NOT DESIGNED FOR USE LOCATION
WITH HAZARDOUS LIQUIDS OR • Locate the pump as near liquid source as practical; below
FLAMMABLE GASES. level of liquid for automatic repriming capability.
• Allow adequate space for servicing and ventilation. Protect
the unit from weather and water damage due to rain,
flooding or freezing temperatures.
Hazardous fluids
can cause fire,
burns or death.
• Protect the pump and piping from freezing temperatures.
• Allow adequate space around the unit for service and
DESCRIPTION and SPECIFICATIONS ventilation.
CLOSE-COUPLED UNITS
The series 3656/3756 are single stage, end suction, centrifugal
pumps for general liquid transfer, booster applications, • Units may be installed horizontally, inclined or vertically with
irrigation and general service pumping. Pumps are available in the motor above the pump.
three different materials of construction: all iron, bronze-fitted • The motor feet MUST be bolted to a substantial surface
and all bronze (“S” group only). (horizontal or vertical) that is capable of complete and rigid
Pump impellers are fully enclosed, key driven and held in support for the pump and motor. For L-Group pumps, the
position by an impeller bolt and washer. Casings are full volute motor adapter feet must also be bolted to the supportive
in design with replaceable wear rings. surface.
Dependant on the pump size, the suction and discharge • For vertical operation, the motor should be fitted with a drip
connections will be threaded or flanged. Shafts are protected cover or otherwise protected against liquid entering the
with stainless steel shaft sleeves. motor (rain, spray, condensation, etc.)
Close-coupled units have NEMA standard JM or JP motors NOTICE: DO NOT INSTALL WITH MOTOR
with C-face mounting and key driven shaft extension. SAE BELOW PUMP. ANY LEAKAGE OR
drive units bolt directly to the engine flywheel housing for SAE CONDENSATION WILL AFFECT
sizes 1, 2, 3, 4 or 5. Optional elastomer element couplings are THE MOTOR.
available for 61⁄2, 71⁄2, 8, 10, 111⁄2 and 14 inch flywheels. Frame
mounted units can be coupled to motors through a spacer
coupling, or belt driven.
2
FRAME-MOUNTED UNITS • For L-Group pumps it is recommended that the pump is
• A flat substantial foundation surface MUST be provided to supported beneath the two feet cast into the motor adapter
avoid distortion and/or strain when tightening the (108). These feet must be bolted to the support.
foundation bolts. A rubber mounting is acceptable to reduce • Begin the assembly of the pump to the engine by thoroughly
noise or excessive vibration. cleaning the engine flywheel and flywheel housing.
• Tighten motor hold-down bolts BEFORE connecting piping • For Goulds Pumps optional couplings – Ensure that the
to pump. coupling hub set screw is backed out enough to ensure
clearance for the shaft key during assembly. Mount the
(1⁄4”) Grout hole Base
Finished grouting coupling assembly to the engine flywheel using the bolts
provided torqued as follows in a crossing sequence:
(1⁄2 to 3⁄4”)
Grout
61⁄2" or 71⁄2" Flywheel – 11 lbs.-ft. (15 N. m)
Allowance
for leveling
Leveling wedges or shims – left in place
8", 10" or 111⁄2" Flywheel – 20 lbs.-ft. (27 N. m)
Top of foundation – 14" Flywheel – 50 lbs.-ft. (68 N. m)
clean and wet down
(For other couplings follow manufacturers
Wood
frame Sleeve
recommended installation procedure.)
• Place the pump shaft key into the pump shaft (122) and align
the shaft to the coupling. Slide the pump into the coupling
Washer until the engine adapter ring (340) contacts the engine
Lug flywheel housing.
• Mount the pump to the engine using the bolts and lock-
Figure 1 washers provided, torqued as follows in a crossing sequence:
• Baseplate MUST be grouted to a foundation with solid SAE #2, #3, #4, #5 – 20 lbs.-ft. (27 N. m)
footing. Refer to Figure 1. SAE #1 – 50 lbs.-ft. (68 N. m)
• Place unit in position on wedges located at four points, two • Install coupling guards (501N).
below approximate center of driver and two below • Bolt motor adapter (108) to the rigid support described
approximate center of pump. Adjust wedges to level unit. above.
Level or plumb suction and discharge connections.
Coupling Alignment
• Make sure bedplate is not distorted and final coupling
alignment can be made within the limits of movement of WARNING FAILURE TO DISCONNECT AND
motor and by shimming, if necessary. LOCKOUT ELECTRICAL POWER
BEFORE ATTEMPTING ANY
• Tighten foundation bolts finger tight and build dam around
MAINTENANCE CAN CAUSE
foundation. Pour grout under bedplate making sure the areas
SEVERE PERSONAL INJURY.
under the pump and motor feet are filled solid. Allow grout
to harden 48 hours before fully tightening foundation bolts.
Hazardous machinery
• Tighten pump and motor hold-down bolts before aligning can cause personal
injury or death.
shaft or connecting the piping to pump.
• Allow grout to harden for 48 hours before tightening 4 FRAME-MOUNTED UNITS ONLY
foundation bolts. • Alignment MUST be checked prior to running. See Figure 2.
Operation
Figure 3 Figure 4 WARNING DO NOT OPERATE FRAME
H
16 MOUNTED OR SAE UNITS WITHOUT
15
14 SAFETY GUARDS IN PLACE OR
H = Min. Submergence in feet
13
12 SEVERE PERSONAL INJURY MAY
11
10 RESULT.
9
8
3.0D 7
6 Hazardous machinery
min. 5 can cause personal
4 injury or death.
1.5D 3
H min. 2
min.
1
D min. 1 2 3 4 5 6 7 8 9 10111213141516 V
D 2 D
V = Velocity in feet per second
= GPM x 0.321 GPM x 0.4085
Area D2
Figure 5 Figure 6
4
WARNING SPLASHING OR IMMERSING OPEN FRAME-MOUNTED UNITS
DRIP PROOF MOTORS IN FLUID CAN • Model 3756 S-group has greased for life bearings.
CAUSE FIRE, SHOCK, BURNS OR No regreasing is possible or necessary.
DEATH.
• Model 3756 M or L-group bearing frame and SAE drive
bearing frame should be regreased every 2,000 hours or at a
three month interval, whichever occurs first. Use a #2
Hazardous voltage
can shock, burn or
sodium or lithium based grease. Fill until grease comes out of
cause death. relief fittings, or lip seals, then wipe off excess.
• Follow motor or engine and coupling manufacturer’s
WARNING OPERATION AT OR NEAR ZERO lubrication instructions.
FLOW CAN CAUSE EXTREME HEAT,
PERSONAL INJURY OR PROPERTY • Recheck alignment.
DAMAGE. SEASONAL SERVICE
• To REMOVE pump from service, remove drain plug and
drain all unprotected piping.
Extreme heat can • To RETURN pump to service, replace drain plug using
cause personal injury
or property damage. Teflon™ tape or equivalent on male threads.
• Reconnect suction line if removed, examine union and repair
NOTICE: NO NOT RUN PUMP DRY OR SEAL
if necessary.
DAMAGE WILL RESULT.
• Refer to OPERATION section of manual.
• After stabilizing the system at normal operating conditions,
check the piping. If necessary, adjust the pipe supports.
Disassembly
• On frame-mounted units, coupling alignment may have
changed due to the temperature differential between pump • Follow ALL warnings and instructions in the
and motor. Recheck alignment following procedures and “MAINTENANCE” section of this manual.
hazard warnings in “COUPLING ALIGNMENT” section • Close-coupled units: Remove motor hold-down bolts.
of this manual. • Frame-mounted units: Remove coupling guard, spacer,
coupling and frame hold-down bolts.
Maintenance LIQUID END
WARNING FAILURE TO DISCONNECT AND 1. Remove casing bolts (371).
LOCKOUT ELECTRICAL POWER
BEFORE ATTEMPTING ANY 2. Remove back pull-out assembly from casing (100).
MAINTENANCE CAN CAUSE SHOCK, 3. Remove casing wear ring (103) if excessively worn.
BURNS OR DEATH. NOTICE: DO NOT INSERT SCREWDRIVER BETWEEN
IMPELLER VANES TO PREVENT ROTATION.
Hazardous voltage 4. On close-coupled units, remove motor end plug or cover
can shock, burn or
cause death. to expose screwdriver slot or flats on end of motor shaft.
FAILURE TO RELIEVE SYSTEM 5. While restraining shaft with an appropriate tool (close-
CAUTION coupled units) or with a strap wrench (frame-mounted
PRESSURE AND DRAIN SYSTEM
BEFORE ATTEMPTING ANY units) remove impeller bolt (198). Impeller bolt may need
MAINTENANCE CAN CAUSE to be heated with torch to remove. Discard.
PROPERTY DAMAGE, PERSONAL NOTICE: EXERCISE CAUTION WHEN HANDLING
INJURY OR DEATH. HOT IMPELLER BOLT.
Hazardous pressure
can cause personal 6. Remove impeller washer (199). Discard.
injury or property
damage. 7. Insert two pry bars, 180° apart, between impeller (101) and
seal housing (184). CAREFULLY pry off impeller.
WARNING IF PIPING HAZARDOUS OR TOXIC 8. Remove impeller key (178).
FLUIDS, SYSTEM MUST BE FLUSHED
PRIOR TO PERFORMING SERVICE. 9. Remove seal housing bolts (370B) and seal housing (184)
pulling with it the mechanical seal assembly. Discard seal
assembly and seal housing o-ring (513). For packed box
pumps see “PACKED BOX INSTRUCTIONS”.
10. Remove adapter bolts (370) and adapter (108).
Hazardous fluids can
cause personal injury
or property damage. 11. Inspect shaft sleeve (126). If badly scored, remove by
heating with torch. Discard.
CLOSE-COUPLED UNITS NOTICE: EXERCISE CARE IN HANDLING HOT
• Bearings are located in and are part of the motor. For SHAFT SLEEVE.
lubrication information, refer to motor manufacturer’s
instructions.
5
12. Push out the mechanical seal stationary seat from the 8. Fully and squarely install the rotary assembly of seal
seal housing. Discard. against the stationary seat.
13. On units equipped, remove seal housing wear ring (203) NOTICE: REPLACE IMPELLER BOLT AND WASHER
if excessively worn. WHENEVER IMPELLER IS REMOVED.
DISASSEMBLY OF BEARING FRAME OR 9. Install impeller key in shaft keyway. Mount impeller on
SAE BEARING FRAME shaft and push until it bottoms. Hold in place.
1. Remove deflector (123) from shaft. 10. Install new impeller washer.
2. Remove bearing cover (134). 11. Apply LOCTITE® #262 or equivalent, to new impeller bolt
3. Remove shaft assembly from frame. threads and tighten to:
4. Remove lip seals (138, 139) from bearing frame (228) and
3
⁄8"-16 bolts 20 lbs.-ft. (27 N. m)
bearing cover (134) if worn. Discard..
1
⁄2"-13 bolts 38 lbs.-ft. (51 N. m)
5. Remove retaining ring (361). 12. Replace casing wear ring, if removed.
6. Use bearing puller or arbor press to remove ball 13. Replace casing bolts and tighten, in a crossing sequence,
bearings (112, 168). to torque values indicated below:
3
⁄8"-16 bolts (bronze casing) 25 lbs.-ft. (34 N. m)
Reassembly 3
⁄8"-16 bolts (cast iron casing) 37 lbs.-ft. (50 N. m)
1
⁄2"-13 bolts (cast iron casing) 90 lbs.-ft. (122 N. m)
• All parts should be cleaned before assembly. 3
⁄4"-10 bolts (cast iron casing) 175 lbs.-ft. (237 N. m)
NOTICE: O-RING SHOULD BE REPLACED AFTER ANY 14. Check reassembled unit for binding by rotating shaft with
DISASSEMBLY OF UNIT. appropriate tool from motor end.
BEARING FRAME 15. If rubbing exists, loosen casing bolts and proceed with
1. Replace lip seals if removed. tightening sequence again.
2. Replace ball bearings if loose, rough or noisy when rotated. 16. Replace motor hold-down bolts and motor end plug or
3. Check shaft (122) for runout. Maximum permissible is cover on close-coupled units.
0.002" (0.05 mm) TIR. 17. Replace coupling, spacer, coupling guard and frame
4. Refer to the “MAINTENANCE” section of this manual hold-down bolts on frame-mounted units.
for M-group bearing frame regreasing instructions. NOTICE: ALWAYS RECHECK BOTH ALIGNMENTS
LIQUID END AFTER MAKING ANY ADJUSTMENTS.
1. Inspect shaft removing any debris or burrs. 18. Refer to the “COUPLING ALIGNMENT” section of
this manual to realign shaft.
2. Treat shaft with LOCQUIC® Primer “T”, or equivalent,
following manufacturer’s instructions carefully. 19. Assembly is complete.
3. When replacing shaft sleeve, spray new shaft sleeve’s bore Packed Box
with LOCQUIC® Primer “T”, or equivalent. Let parts dry
and then apply LOCTITE® #262 on the same surfaces. 1. Make sure stuffing box is free of foreign materials and
Slide new sleeve over shaft with a twisting motion, wipe clean before beginning packing of packed box. Refer to
off excess. Let cure according to instructions. Sectional Assembly in the repair parts section.
NOTICE: MECHANICAL SEAL MUST BE REPLACED 2. Take special care during installation of packing rings
WHENEVER SEAL HAS BEEN REMOVED. because they are die-formed. To install, twist the ring
FOLLOW SEAL MANUFACTURER’S sideways just enough to fit it around the shaft sleeve.
INSTRUCTIONS CAREFULLY. FOR PACKED DO NOT ATTEMPT TO PULL RINGS STRAIGHT
BOX PUMPS SEE “PACKED BOX OUT. See Figure 7.
INSTRUCTIONS”. CORRECT WRONG
6
3. Install the two piece Teflon lantern ring supplied as shown Trouble Shooting Guide
in figure 8. Note: two pieces make one ring. Notches on
ring must face each other, but alignment is not necessary. WARNING DISCONNECT AND LOCKOUT
ELECTRICAL POWER BEFORE
ATTEMPTING ANY MAINTENANCE.
FAILURE TO DO SO CAN CAUSE A
SHOCK, BURN OR DEATH.
Hazardous voltage
can shock, burn or
cause death.
SYMPTOM
MOTOR NOT RUNNING
Teflon Lantern Ring See Probable Cause – 1 through 5
Figure 8 LITTLE OR NO LIQUID DELIVERED
4. Install the packing rings and lantern ring in the following See Probable Cause – 6 through 13
sequence to pack the packed box. Install two rings of EXCESSIVE POWER CONSUMPTION
packing, then the lantern ring, followed by the final three See Probable Cause – 3, 13, 14, 15
rings of packing. Install each ring separately and firmly EXCESSIVE NOISE and VIBRATION
seat. The use of a wooden split bushing is recommended See Probable Cause – 3, 6, 7, 10, 12, 14, 16, 17
to accomplish this. See Figure 9. Use gland to jack the
bushing and ring into the box. Stagger joints in each ring PROBABLE CAUSE
90°. Make sure the flush tap in the packed box lines up 1. Motor thermal protector tripped
with the center of the lantern ring. Any extra rings are 2. Open circuit breaker or blown fuse
spares. 3. Impeller binding
4. Motor improperly wired
5. Defective motor
WOODEN “SPLIT BUSHING” 6. Pump is not primed, air or gases in pumpage
7. Discharge, suction plugged or valve closed
8. Incorrect rotation (3 phase only)
9. Low voltage or phase loss
GLAND 10. Impeller worn or plugged
11. System head too high
12. NPSHA too low – Excessive Suction lift or losses
STUFFING 13. Incorrect impeller diameter
BOX SHAFT
14. Discharge head too low – excessive flow rate
15. Fluid viscosity, specific gravity too high
Figure 9 16. Worn bearing
17. Pump, motor or piping loose
5. Tighten gland nuts evenly, but not tight. When the pump
is started, slowly tighten the gland nuts until the leak rate
is between 40 and 60 drops per minute. A grease lubricant
can be used when the pumpage contains abrasive particles
or for a suction lift condition.
REMOVAL OF PACKED BOX
• Follow these steps to remove the packing from the
packed box.
1. Remove gland assembly.
2. With a “packing hook” remove packing.
3. Insert a wire hook into the ring on the outer edge to
remove the lantern ring.
4. Clean the packed box.
7
Repair Parts Series 3656/3756
4 6 7 8 2 5 9 10 3 11
13
3
10 17 18
12 1 14
19
20
21
8
Performance Curves – 60 Hz, 3500 RPM
Curvas de desempeño – 60 Hz, 3500 RPM
These curves show the performance of the 3656 and 3756 at 3500 RPM and Estas curvas ilustran el desempeño de los modelos 3656 y 3756 operando a
1750 RPM, 60 Hz, and at 2900 RPM and 1450 RPM, 50 Hz. Standard impeller 3500 RPM y 1750 RPM en 60 Hz, y a 2900 RPM y 1450 RPM en 50 Hz. Se
trims are shown. muestran los diámetros de impulsor estándar.
Model 3656/3756 S-Group 3500 RPM NOTE: Not recommended for operation beyond Optional Impeller
printed H-Q curve. Impulsor optativo
9 BF / Size (Tamaño) 1 x 2 – 8 ODP & TEFC NOTA: No se recomienda la operación más allá de la
Imp. Dwg. 120-67 curva impresa de H-Q (carga-capacidad). Ordering Code Dia.
METERS FEET
METROS PIES Código de pedido Diá.
100 A 8 1⁄16"
NPSHR-FT (PIES) 3' 4' B 75⁄8"
90 300 25 30 35 40 6'
C 7 5⁄16
45
D 7
80 A 81/16'' DIA. 50
E 61⁄2
250 10'
B 7 /8''
5
TOTAL DYNAMIC HEAD (CARGA DINÁMICA TOTAL)
F 63⁄16
70 50 % EFF
G 53⁄8
C 7 /16''
5
H 51⁄8
60 200
D 7''
NOTE: Pump will pass a
50 10 sphere to 3⁄16" diameter.
E 61/2'' HP
150 NOTA: La bomba dejará
F 63/16''
40 7.5
HP pasar una esfera de hasta 3⁄16
de pulgada de diámetro.
G 5 /8''
3
5H
30 100 P
H 51/8''
20 3H
P
50 2H
P
10
0 0
0 10 20 30 40 50 60 70 80 90 100 110 120 U.S. GPM
0 5 15 20 25 m3⁄hr
CAPACITY (CAPACIDAD)
7-1
14
M O D E L L C O - 6 A N D 8 1 2 5 P S I B A G F I L T E R S
Features
• Permanently piped housings are
opened without special tools
• Quick opening cover
• Carbon or stainless steel housings
• Large area bag and basket for greater
dirt-holding capacity
• O-ring seals: Buna N, EPR and Viton®
• Adjustable-height tripod legs
• 125 psi rated housing
Pipe size: 1-1/2"
• Basket material is compatible with housing 4.0
• Bag surface area is 5.6 square feet (LCO8), 3.0
and 2.0 square feet (LCO6) 2.5
• Uses number 12 size bags (LCO8), and 2.0
number 8 size bags (LCO6) 1.5
Pressure drop ( P) , psi
*Based on housing only. Fluid viscosity, filter bag used, and ex- Flow, gpm
pected dirt loading should be considered when sizing a filter.
17
14
6. Pressure Switches
(PSL2)
7-1
18
Installation And Maintenance Instructions Pressure/ Vacuum Switches
Pressure Switches
Series 96200, 96201, 96210, 96211
Vacuum Switches CWWS P/N: SWAP0411 2.5 - 15 PSI
Series 96220, 96221
Description CWWS P/N: SWAP0413 8.5 - 50 PSI
These switches are miniature size pressure/vacuum switches
having factory set or field-adjustable set-points, fixed deadbands
(differentials), and have piston or diaphragm piston sensors.
The switches are designed to provide long life and maintain
excellent set-point accuracy despite environmental conditions. Wiring
Materials wetted by the process fluid include pressure
connections of brass or stainless steel and internal elastomers Wiring must comply with local and national electric codes.
of Buna-N and Teflon as standard. Caution: Electrical rating must be within range stated on the
switch nameplate. Failure to stay within the rating
Caution: Do not use these switches for hazardous of the switch may result in damage to, or premature
or corrosive service, and do not use for oxygen
failure of, the electrical contacts.
service without proper degreasing and proper
preparation. Check nameplates on switch for Standard switch rating: 5 amps, 250 VAC/28 VDC
maximum proof (overpressure) pressure or SPDT
vacuum limits.
Standard free leads color code:
Installation Pressure switch Vacuum switch
purple = COMMON purple = COMMON
Caution: This switch should be installed by a trained red = NO red = NC
service person. A media filter should be in the blue = NC blue = NO
system to protect the switch.
Mounting: Switch may be mounted in any position. Optional termination:
Piping: Support adequate piping and mount the pressure -T1: 1/4" male quick connect terminals. C, NO, NC markings
switch to avoid excessive shock or vibration. To are on the switch.
minimize the effect of vibration on switch, mount
perpendicular to vibration. Apply pipe compound -T2: DIN 43650 type. Use 5 to 8 mm O.D. cable size. After
sparingly to male pipe threads only. Avoid pipe the cable hook-up, tighten the gland squeeze nut to
strain on switch by properly supporting and ensure firm environmental seal.
aligning piping. Pressure switch Vacuum switch
Tamper Field-adjustable models with a “T” prefix have position 1 = C position 1 = C
Resistant two set screws to lock-in the set-point position 2 = NC position 2 = NO
Option: adjustment sleeve. Use a 5/64 allen wrench. position 3 = NO position 3 = NC
Overtightening will result in switch damage. position 4 = NOT USED position 4 = NOT USED
Set-point adjustment for field-adjustable models -T4: 1/2" NPT male conduit connection with free leads.
-T5: 1/2" NPT female conduit connection with free leads.
For pressure switch (96201 and 96211 models):
Connect applicable conduit to the switch (plastic conduit
1. After connecting swi tch, turn adjustment sleeve clockwise is recommended). Generally, handtight is adequate for
to increase the set-point and counter-clockwise to decrease a conduit enclosure. Caution: use conduit hex only when
the set-point. The switch can be set either on “increasing tightening. Do no t exceed 80 lb in tightening or bending
pressure” mode or “decreasing pressure” mode. torque. Failure to follow this instruction may result in
2. For exact pressure set ting, cycle pressure swi tch and switch damage.
make fine adjustments with sleeve.
Testing of the switch
For vacuum swi tch (96221 model):
Testing of the switch may be done before or af ter final
1. After connecting swi tch, turn adjustment sleeve counter- installat ion. If bench tested, the switch should be re-tested
clockwise to increase the set-point and clockwise to
when installed in the final application. Be sure switch can be
decrease the set-point. The swi tch can be set either on
“increasing vacuum” mode or “decreasing vacuum” mode. tested without affecting other equipment. Check nameplates
for electrical rating and circui try (normally closed or normally
2. For exact vacuum setting, cycle vacuum switch and make
open) of switch. Cycle switch a few times to check operation.
fine adjustments with sleeve.
Troubleshooting OPTIONAL ELECTRICAL TERMINATIONS
Warning : Disconnect electrical power supply to swi tch
before removal or inspection.
Important: The switch is not field-repairable. In case of
damage, replace entire swi tch. NO
Causes of improper operation: C NC
1. Incorrect electrical connection: check leads to swi tch. Be
sure they are properly connected. See “Wiring” section for
circuitry color code.
2. Faulty control circuit: check electrical power supply to switch
Check for loose or blown fuses, open-circuited or grounded
wires, loose connections at switch.
3. Incorrect pressure: check pressure in system.
Pressure/vacuum must be within range specified on
-T1 SUFFIX -T2 SUFFIX
nameplate.
4. External leakage: replace pressure/vacuum switch.
5. Excessive vibration or surges: check for pressure fluctuations
in system. Check switch mounting and be sure there is no
excessive vibration.
If the operation of the pressure/vacuum switch cannot be
corrected by the above means, consult factory or authorized
factory representative.
All products of the company are sold and all services are offered
subject to the company's standard conditions of sale.
SET SCREW
(5/64 ALLEN WRENCH)
DO NOT OVERTIGHTEN
FREE LEADS
STANDARD
PREFIX "T"
SET-POINT TAMPER RESISTANT
ADJUSTMENT ADJUSTMENT SLEEVE
SLEEVE
1 1/4 HEX
(APPLY WRENCH
HERE ONLY)
PRESSURE
PORT
1/4 NPT
MALE
FACTORY SET FIELD-ADJUSTABLE -P1 SUFFIX
MODEL MODEL 7/16-20UNF MALE
96200, 96210, 96201, 96211, PRESSURE PORT
96220 SERIES 96221 SERIES WITH O-RING
7. Oil Content Detector
(OCD)
7-1
20
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
109080
INSTRUCTION MANUAL
BILGE OIL ALARM
CWWS
1732 Mc Gaw Avenue
Irvine, California, 92614
TELEPHONE: 949-222-5777 FAX: 949-222-5770
WEB: http://www.cworldwater.com
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Section 1
Introduction
1.1 Important Notes 2
1.2 SMART CELL - BILGE Overview 2
1.3 Specification Listing 3
Section 2
Installation
2.1 Construction details 5
2.2 Mounting 6
2.3 Pipe work and Connections 7
2.4 Electrical Connections 8
Section 3
Commissioning
3.1 Piping 11
3.2 Electrical 11
3.3 Functional testing 11
Section 4
Operating Instructions
4.1 Getting Started 11
4.2 Clean Water Calibration 13
4.3 Set Alarm Points 14
4.4 Set Alarm Delays 15
4.5 Set Analogue out 16
4.6 View IMO Data 16
4.7 Diagnostics & Clean Water Flushing 17
Section 5
Maintenance and Spares
5.1 Maintenance (weekly) 19
5.2 Maintenance (Yearly) 19
5.3 Spares 19
Section 6
Trouble Shooting & Contact
6.1 Trouble Shooting Guide 20
6.2 Tech Support registration 21
Annexe
Certificates
EC MED Certificates 22/23
CWWS Page 1 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Section 1
Introduction
1.1 Important Notes
§ All National or local codes of practice or regulations must be observed and, where
applicable, are deemed to take precedence over any directive or information
contained in this manual.
§ This equipment must be installed and operated in strict accordance with the instructions
contained in this manual. Failure to do so will impair the protection provided.
§ Installation and retro fitting of parts listed in Section 5.4 must be undertaken by a
competent and suitably skilled person.
§ The unit must be isolated from the electrical supply before any maintenance of the
equipment is attempted.
§ Hazardous voltages are present within the monitor enclosure and solenoid valve
assembly.
§ The SMART CELL enclosure must not be opened. If opened a cell breach error will be
displayed on the screen (this error cannot be reset) and the “VOID” tamper proof labels
on either side of the enclosure lid will be destroyed.
Analysis of the oily water sample is undertaken by a unique detector array arrangement. A
narrow beam of light is emitted across a glass cell tube which the oily water sample passes
through. Light scattered and transmitted is detected across the array and passed through
a complex oil content calculation algorithm. The resulting oil content is then displayed on
the LCD screen.
Two oil alarm relays and a 0-20mA / 4-20mA output are provided to control the overboard
discharge and to view / record oil content levels remotely.
To meet the new requirements of MEPC 107(49) an on board flash memory, Real time
clock, dual switch input and clean water / sample selection valve are fitted. These are
described in more detail in the operating instructions (Section 4).
CWWS Page 2 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Measurement
Oil types: Fuel Oil, Diesel and Mixture C (IMO defined)
Alarms
Oil alarm 1 setpoint: 1 – 15 ppm user adjustable
User Interface
LCD display: 4 x 16 alphanumeric back lit LCD display
Input / Output
Analogue output: Loop powered 4-20mA / 0 - 20mA 24V
IMO required data retrieval: Via LCD display or RS 232 communication link
CWWS Page 3 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Protection class: IP 65
software lock
Weight: 3.2 Kg
CWWS Page 4 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Section 2
Installation
2.1 Construction Details
The SMART CELL - BILGE Monitor is split into two logical modules. The display / control
module (2) (left) contains the membrane keypad, LCD display and the Smart Bilge Control
PCB. Contained on this PCB are all the control and data storage associated electronics in
addition to all the customer terminations which can be accessed via the cable glands(5).
This PCB links a 5 volt supply and RS485 communications to the Cell electronics via a 4 wire
interface cable.
The SMART CELL (1), contains the optoelectronics and mechanics used in the calculation of
the sample oil content. UNDER NO CIRCUMSTANCE IS THIS ENCLOSURE TO BE OPENED.
Sample (7) and clean water (8) lines are connected to the solenoid valve to allow
automatic selection of clean water for clean water calibration and diagnostics.
Sample/Clean water is exhausted from the monitor via the outlet (6) after passing through
the measuring sensor. A manual cell clean unit (3) is provided which must be used in
accordance with the maintenance requirements in Section 5. The throttle (9) can be used
to increase or decrease the back pressure in the cell. This is especially useful if air is present in
the sample.
Fig 2.1
Construction details
FIXING HOLES
FOR M6 3 1 SMART - MEASURING CELL
TYPICAL 4 POSNS 6
2 SMART - DISPLAY/CONTROL MODULE
3 MANUAL CLEAN
WARNING - HAZARDOUS VOLTAGE INSIDE ! !
4 3 WAY SOLENOID VALVE
THROTTLE ADJ 9
SMART - BILGE
RTE
RTE
CALIBRATION VOID
WARNING
2 SMART - CELL
6 SAMPLE OUPUT CONNECTION
289
RTE
235.0
5
4 WEIGHT:- 3.2 Kg
7 8
275.0
148.5
295.0
CWWS Page 5 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Mounting Note: if the monitor is sold with Bilge Water Separator, It is pre-installed.
Mount the SMART CELL - BILGE by means of 4 x M6 bolts on to a rigid vertical surface and
preferably with the display panel of the monitor at operator eye level.
Fig 2.2
Mounting Details
FIXING HOLES
FOR M6
TYPICAL 4 POSNS
DO NOT OPEN
WARNING - HAZARDOUS VOLTAGE INSIDE ! NO SERVICEABLE PARTS INSIDE !
THROTTLE ADJ
SMART - BILGE
RTE
RTE
CALIBRATION VOID
CALIBRATION VOID
WARNING
WARNING
SMART - CELL
RTE
RTE
205.0
275.0
CWWS Page 6 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Note: If the monitor is sold with Bilge Water Separator, the system is pre-wired and
pre-piped.
Fig 2.3
Pipe Fittings
2 Metre
Minimium
3 200mm
SAMPLE OUT Min
10mm PIPING
DO NOT OPEN
WARNING - HAZARDOUS VOLTAGE INSIDE ! NO SERVICEABLE PARTS INSIDE !
THROTTLE ADJ
SMART - BILGE
SMART - CELL
1 2
DRAIN
ENSURE LEVEL
IS CLEAR OF
BILGE WATER
CWWS Page 7 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Please read the following points before attempting any electrical work.
User terminations are located by removing the lid from the SMART Display/Control module
via the four screws. A protection plate is situated above the terminals which must be
removed via 4 M3 screws.
Fig 2.4.1
User Terminations
JP1 TO BE LINKED
JP1
IF 24VAC IS REQUIRED
J1 J2 J3 J4 J5
L 1 2 3 1 2 3
FUSE F1 E N L E N L E N
SUPPLY IN SAMPLE-CLEAN
250V 1.25A ANTI SURGE
~
ALARM 1 ALARM 2
SMARTBILGE PCB
CUSTOMER OUTPUT
INPUT TO SOLENOID
230-110 VAC VALVE
OR 24VAC CONNECTION
50 - 60 Hz
NORMALLY CLOSED
IN ALARM MODE
OUTPUTS FOR
CUSTOMER
CWWS Page 8 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Table 2.4.2
User Terminal Connections
J2 - L Live 24/115/230 V AC
Open Terminal. Note:
J2 - E Earth for optional features not 250V AC , 500mA.
used with this application.
J2 - N Neutral 24/115/230 V AC
J3 - L Live 24/115/230 V AC
Sample clean control. Note:
J3 - E Earth wiring to this terminal is not to 250V AC, 500mA.
be altered.
J3 - N Neutral 24/115/230 V AC
J9 - 1 5V DC from PCB
Back-flush Switch Input 2 5V DC 20mA
J9 - 2 Switch In
J11 - 1 Tx
J11 - 3 Rx
Precise wiring details will vary dependant upon the control system to be employed but
Alarm relay 1must be used for control of the overboard discharge valve and Alarm relay 2
for alarm annunciation only.
CWWS Page 9 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Section 3
Commissioning
3.1 Piping
3.2 Electrical
§ Check that correctly rated isolator is connected to the mains input (see Section 2.4.).
§ Check the SMART CELL - BILGE is correctly earthed.
§ Check all cabling conforms to specifications set out in Table 2.4.
CWWS Page 10 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Section 4
Operating instructions
4 x 16 Alphanumeric LCD
Display
CWWS Page 11 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
On power up the user will be presented with a screen as shown in Fig 4.1.2.
§ At the top of the display the user will see the oil concentration in Parts Per Million (PPM).
§ The status of the two oil alarms will be displayed.
§ The date and time is displayed. (time is set to GMT / UTC ).
§ If either of the alarm set points is passed the LCD backlight will flash.
Fig 4.1.2
Screen Shot - Main Information Screen
Shows Oil
Concentration in
PPM
To interrogate the SMART CELL - BILGE Monitor settings the user can press any of the Interface
Keys to bring up the Menu Options Screen as shown in Fig 4.1.3
Fig 4.1.3
Screen Shot - Menu Screen Options
CWWS Page 12 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
A clean water calibration can be performed to account for differences in clean water
between the factory calibrated clean water and the ships own clean water supply. On
activation of the clean water calibration the clean water / sample solenoid will select
clean water. The SMART CELL - BILGE will then average the clean water reading over the
period of 30 seconds after which a further 10 seconds will lapse for the user to verify the
calibration in monitor mode before the solenoid selects sample once more.
To set the clean water calibration enter the Menu screen as shown in Fig. 4.1.3 and scroll
down until the cursor reaches "Clean wtr cal" then press the ENTER key. The screen will
change as shown in Fig 4.2.2.
Fig 4.2.2
Screen Shot - Menu Screen Options
CWWS Page 13 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Note: The alarms have a 0.5 ppm hysteresis included to prevent relay chatter. e.g if an
alarm is set at 15 ppm and the oil concentration exceeds 15 ppm, the alarm will be
activated. The alarm will not deactivate until the oil concentration drops below 14.5 ppm.
To set the required Alarm points enter the Menu screen as shown in Fig. 4.1.3 and scroll
down until the cursor reaches "Set alrm point" then press the ENTER key. The screen will
change as shown in Fig 4.3.1.
Fig 4.3.1
Screen Shot - Alarm Set Options
Fig 4.3.2
Screen Shot - Alarm PPM Value
CWWS Page 14 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Note: The main information display screen will show if an oil alarm is less or more than the
current oil reading regardless of the alarm delay.
To set the required alarm delays go the menu screen as shown in Fig. 4.1.3 and scroll down
until the "Set alrm delay" then press the ENTER key. The screen will change as shown in Fig
4.4.1
Fig 4.4.1
Screen Shot - Alarm Delay Options
Fig 4.4.2
Screen Shot - Alarm Set Delay Value
CWWS Page 15 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Fig 4.5.1
Screen Shot - Analogue Output Options
The separator status is recorded via the SW1 and SW2 inputs. The following configuration is
used.
SW1 Separator ON/OFF status.
SW2 Separator Back-flush (If fitted)
Note. SW2 will select clean water and activate alarms permanently when closed. To be
connected to Separator backflush control circuit via a potential free contact.
The configuration of the switches must be recorded on board and date stamped.
To view or download the IMO required data go to the Menu screen as shown in Fig. 4.1.3
and move the curser down until "View IMO data" then press the ENTER key. The screen will
change as shown in Fig 4.6.1.
Fig 4.6.1
Screen Shot - View IMO Data
CWWS Page 16 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Fig 4.6.2
Screen Shot - IMO Data
Fig 4.6.3
Screen Shot - Dump IMO Data
NOTE: Dump IMO data command will NOT erase any data
Fig 4.7.1
Screen Shot - Diagnostics
CWWS Page 17 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Fig 4.7.2
Screen Shot – View detectors
Fig 4.7.3
Screen Shot – View disp PPM
CWWS Page 18 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Section 5
Maintenance and Spares
5.1 Weekly Maintenance
§ Manual Clean Cell Unit, plunge and release 10 times
durring clean water flush Section 4.7
§ At yearly intervals (or where necessary) the wiper ring on the manual clean assembly
should be replaced. This can be ordered from CWWS directly as part no. SBWR9039 .
5.3 Spares
When ordering spares, it is important to supply details of the type of monitor, ser ial number
of monitor, part number of each spare required, its description and any other relevant
information.
OTHER SPARES
CWWS Page 19 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Section 6
Trouble Shooting Guide
Monitor stays in alarm or Dirty measuring cell. Clean cell and perform Clean
reading is inaccurate. water calibration. Section 5.1
CWWS Page 20 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
Display shows SYS ERROR, RS 485 communications Ensure 4-pin plug is tightly
Communications failure cable is loose / not inserted into J15. Fig 2.4.1
connected
Display shows SYS ERROR, Smart Bilge electronics Contact CWWS for replacement
RTC – I2C failure failure. Smart Bilge PCB (Section 5.4).
Display shows SYS ERROR, Measuring Cell has been Contact CWWS for replacement
Cell breach contact CWWS opened. measuring cell (Section 5.4).
CWWS Page 21 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
CWWS Page 22 of 23
SMART CELL - BILGE - Instruction Manual
Doc. 109080 Rev. B 10.01.05
CWWS Page 23 of 23
14
8. System Valves
22
7-1
SOLENOID VALVES
(SV1-SV4)
The following procedure is for disassembly and reassembly of the solenoid valves. An exploded view is
provided for clarification. (See illustration B1 for ½” thru 1” Solenoid Valves and illustration B2 for 1
½” Solenoid Valves on the following pages.)
1. Drain water from the system and turn off electrical power to the system.
2. Pull up on coil and remove. If coil stem is separate item than valve cover, unscrew to remove
core cover, core “O”-ring, core and core spring.
3. Unscrew valve cover and remove cover screws, and cover. If coil stem is integral to valve
cover, remove core and core spring along with cover.
4. Remove exposed diaphragm spring (1 ½” valve only), diaphragm, and all body “O”-rings when
applicable.
NOTE
Refer to the spare parts list, page 7-1, when ordering replacement
parts for the solenoid valves.
5. All parts are now accessible for cleaning or replacement. Replace worn or damaged parts with
complete Spare Parts Kit for best results.
6. Reassemble in reverse order of disassembly, paying careful attention to the exploded view provided
for placement of all parts.
7. After maintenance, operate valve a few times to be sure of proper opening and closing.
24
Solenoid Valves: ½” thru 1”
The items marked with * are included in the valve repair kit.
COIL
COVER SCREWS
COVER
* DIAPHRAGM
BODY
Illustration B1
Exploded View ½” thru 1”
Solenoid Valves
7-1
Solenoid Valves: 1 ½”
The items marked with * are included in the valve repair kit.
COIL
CORE COVER
COVER BOLTS
COVER
* SPRING
* DIAPRAGHM
BODY
Illustration B2
Exploded View 1 ½”
Solenoid Valves
26
SWING CHECK VALVES
(CV1-CV8)
GLOBE VALVES
(GV1-GV4)
MODEL
!! !#
! "
# $
" #$%$& " #'%#( " )$ " )' " *) " ($ " '$
$'
#
#'
)
)'
*
*'
(
!
" ## $ % & %
$ # '&
(( ) ##
*## $ % +, -
!"
#" $
#
( %& '. %
BALL VALVES
(BV1-BV8)
The Drain Valves and the Sample Valves are provided to manually drain and sample the contents of the
separator vessel. The standard valves used to perform these functions are typical brass ball valves.
(See illustration below).
Series 530C 2
6
Sizes: 1/2" or 3/4" (15 or 20mm)
5
Series 530C Calibrated Pressure Relief Valves are spring operated
bronze valves designed for use only as protection from the build up 4
of excessive pressure in systems containing water, oil or air. Series
530C valves incorporate a calibrated adjustment feature for setting 3
the valve to the relief pressure required. These valves are ideally
suited for bypass thermal expansion relief. 1
Model 530C
Features
Materials
• Calibrated adjustment feature for setting valve to relief pressure
required 1. Body Bronze
• Adjustable range 50 – 175psi (3.45 – 12.06 bars) 2. Bonnet Brass
• All bronze construction 3. Disc Holder Brass
• All stainless steel springs 4. Disc Buna-N
• Buna-N disc on machined body seat 5. Adjustable Spring Stainless Steel
• Inlet (bottom), male NPT threaded 6. O-ring Buna-N
• Outlet (side), female NPT threaded 7. Spring Washer Brass
Pressure – Temperature
Capacity (530C)
psi bars
Maximum Temperature: 180° F (82° C) 300 20.7
Maximum Operating Pressure: 300psi (20.67 bars) 275 18.9
250 17.2
Spring Ranges 225 15.5
Inlet Pressure
3/4"
1 2
/ " or 3/4" (15 or 20mm): 50 – 175psi (3.45 – 12.06 bars) 200 13.8
(20mm)
1 2
/ " (15mm): 100 – 300psi (6.89 – 20.67 bars) 175 12.1
1/2"
150 10.3
Application Note: The Watts Series 530C are not ASME approved (15mm)
125 8.6
safety relief valves and should not be used in system application
100 6.9
with this requirement.
75 5.2
50 3.4
25 1.7
0 5 10 15 20 25 30 35 40 gpm
19 38 57 76 95 114 133 152 lpm
7-1
28
GAUGES
Model 40
Industrial Gauges
(Dual Scale)
STANDARD RANGES
Standard gauge scales - Dual.
Outer scale is PSI.
Inner scale is metric (kPa & bar.)
1 /2 , 2 , 2 /2 , or 4 Case
1 ” ” 1 ” ” 30”-0-200 psi
30”-0-300 psi
-1-0-14 bar
-1-0-20 bar
✔
✔
✔
✔
✔
✔
✔
✔
50
50
5
5
B A SIZE A B C D E
Style 1
B A
Style 2
A
C
Style 3
A C
Style 4
B
2-1/2” 2.620” 1.165” 2.270 2.470” 1.109”
E
4” 4.200” 1.265” 2.734” 3.969” 5.203”
APPENDIX 'C'
Drawings
J103-US1000-C - INSTALLATION
Certificates